Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result...

21
Stability Studies in the IVD Industry Jeffrey Budd, PhD Principal Biostatistician Beckman Coulter [email protected] PS1e - CMC Session: Statistical Considerations When Assessing Product Stability and/or Shelf Life

Transcript of Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result...

Page 1: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Stability Studies in the IVD Industry

Jeffrey Budd, PhD

Principal Biostatistician

Beckman Coulter

[email protected]

PS1e - CMC Session: Statistical Considerations When Assessing Product Stability and/or Shelf Life

Page 2: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

In Vitro Diagnostics (IVD)

• We supply laboratory test results that doctors use to diagnose, treat and monitor patients

• We need to ensure that tests remain accurate even as material used in these tests age – reagents used to detect and measure substances

– calibrators used to convert from instrument signal to substance concentration

– control material used to monitor proper system operation

9/26/2017 IVD Stability Studies 2

Page 3: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

IVD Stability Guidelines

• The two most active organizations in providing guidelines for the IVD industry are

– International Organization for Standardization (ISO)

– Clinical and Laboratory Standards Institute (CLSI)

• The most influential guideline for IVD stability is EP251, published in 2009

– This guideline is currently being revised

9/26/2017 IVD Stability Studies 3

Page 4: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Types of Stability Studies

• Shelf life

– Original packaging, specified storage conditions

• In-use

– After opening, reconstituting, thawing

• Transportation simulation

– Product exposed to potential extreme conditions

• Performance monitoring

– Is stability behavior maintained over life cycle?

9/26/2017 IVD Stability Studies 4

Page 5: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Stability Considerations

• Product storage conditions

– Maximize stability (room temp, refrigerate, freeze)

– If range of temperature what is test temperature?

• Acceptance criteria

– What is clinical need, considering intended use?

• Number of lots (3?)

• Mix of shelf life, in-use, transport simulation

– Beginning or end of shelf life?

9/26/2017 IVD Stability Studies 5

Page 6: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Types of Stability Studies

• Classical

– eg, result measured each month over 13 months

• Isochronous

– eg, each month test material placed in stability condition, all measured together at 13 months

• Matching

– eg, each month, test compared to reference

• Accelerated (Arrhenius, other options)

9/26/2017 IVD Stability Studies 6

Page 7: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Time Point Value Assignment • Factors to consider

– Within run*

– Between vial*

– Between run

– Between day

– Calibration to calibration

– Reagent lot to lot

– Calibrator lot to lot

– Instrument to instrument

– Drift over time* * Currently considered in EP25-A

9/26/2017 IVD Stability Studies 7

Page 8: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Designing a stability study

• Minimize systematic influences – Use same instrument(s), reagent lot(s), calibrator

lot(s) across the study period

– Be aware of potential drift due to these factors

• Sample random factors (eg, calibrations, runs)

• Determine uncertainty at each time point – CVadj = sqrt(CVcal

2/#cals + CVBR2/#runs + CVWR

2/#reps)

• Determine sample size given proposed # points

• Use mean of results at each time point cal = calibrations, BR=between run, WR=within run, reps=replicates

9/26/2017 IVD Stability Studies 8

Page 9: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Plot Replicates or Not?

Slope = -0.0223 At 365 days: Fit = -8.16% 95%CI = -9.48% 10% at 384 days

Slope = -0.0223 At 365 days: Fit = -8.16% 95%CI = -9.12% 10% at 399 days

Not! 9/26/2017 IVD Stability Studies 9

Page 10: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

How to determine baseline

• Some suggest that more robust testing be conducted at day zero to establish baseline

• However, there is no more robust method than using all the data in the study (via the regression): set baseline = zero intercept Akbas (2016)

• This modifies the determination from measuring change from a set value to measuring the percent change over time

9/26/2017 IVD Stability Studies 10

Page 11: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Control Material Matching

• Place control test material in intended use condition (eg, 4°C)

• Place additional control reference material in known, unchanging state (eg, -70°C)

• At each time point measure the difference in results between the two conditions

• Compare drift in this difference to %criteria

• Eliminates the effects of factors: run, day, calibration, instrument, reagent lot, cal lot

9/26/2017 IVD Stability Studies 11

Page 12: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Calibrator Matching

• Use unchanging, internal working calibrators as reference calibrator set at each time point

• Place product calibrator set to be tested in their standard storage condition

• At each time point run both reference and test calibrator sets – In same run, on same instrument, using the same

reagent lot

– Can do multiple repeats if needed

9/26/2017 IVD Stability Studies 12

Page 13: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Calibrator Matching

020000

0

Sy

ste… Reference calibrator points

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

0 20 40 60 80 100 120

Syst

em

Sig

nal

Concentration

Reference calibrator pointswith fitted calibration curve

Test calibrator pointswith fitted calibration curve

Direction ofdegradation

1) Set concentrationof interest (eg, medical

decision point)

2) Find signal level ofmedical decision point from reference curve

3) Using signal level, read concentration off test calibration curve

9/26/2017 IVD Stability Studies 13

Page 14: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Example:

9/26/2017 IVD Stability Studies 14

Page 15: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Current Practice (EP25, Ed. 1)

• Determine sample size based mainly on repeatability (within-run imprecision)

• Plot all replicates (y-axis typically in units)

– Difference from T0 point determined

• Fit regression line to data

– If regression p-value < 0.05 then stability is good

– If regression p-value ≥ 0.05 then use 95% CI of the regression fit

9/26/2017 IVD Stability Studies 15

Page 16: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Example of Current Practice

9/26/2017 IVD Stability Studies 16

95% CI on Regression Fit

Page 17: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Future Practice (EP25, Ed. 2)*

• Determine sample size based on all relevant variance components

• Plot one estimate per time point

– y-axis can be in units or percentage

– Difference from intercept (β0) determined

• Fit regression line to data

– Use 95% CI on percent of change from T0 = β0

9/26/2017 IVD Stability Studies 17

* Based on Holland (2017)

Page 18: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Same Example: Future Practice

9/26/2017 IVD Stability Studies 18

95% CI on %Change

Page 19: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Outcome Predictability

9/26/2017 IVD Stability Studies 19

Page 20: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

Old versus New

• Difference from T0

• p-value >0.05 (pass)

• No confidence statement on outcome

• Unpredictable outcome

• Plot all replicates at Ti

• Underpowered study by missing variance components

• % difference from β0

• Equivalence test

• Can state confidence in outcome

• Outcome is predictable

• Plot mean at each Ti

• Fully powered study that covers all variance components

9/26/2017 IVD Stability Studies 20

Page 21: Stability Studies in the IVD Industry · Types of Stability Studies •Classical –eg, result measured each month over 13 months •Isochronous –eg, each month test material placed

References

9/26/2017 IVD Stability Studies 21

1) Pierson-Perry, J. et al. (2009), “Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline”, CLSI EP25-A.

2) Akbas, N., Budd J., and Klee, G. (2016), “Multiple Calibrator Measurements Improve Accuracy and Stability Estimates of Automated Immunoassays”, Scandinavian Journal of Clinical and Laboratory Investigation, 76, 177-180.

3) Holland, M., Kraght, P., Akbas, N., Budd J., and Klee, G. (2017), “Improved Statistical Methods for Evaluation of Stability of In Vitro Diagnostic Reagents”, Statistics in Biopharmaceutical Research, http://dx.doi.org/10.1080/19466315.2017.1305287 (Accepted)