• In this tutorial you will learn:– Why sterilization monitoring?– Steps of Instrument Processing– How SporeCheck works– Features and Benefits of SporeCheck
![Page 3: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/3.jpg)
Instrument Processing
1. Cleaning
2. Inspection
3. Packaging
4. Sterilization
5. Storage & Delivery
6. Quality Assurance (Sterilizer Monitoring)
There are six (6) recommended steps for Instrument Processing. Sterilization monitoring is part of Quality Assurance.
![Page 4: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/4.jpg)
Sterilization Monitoring
Let’s take just a few minutes to review sterilization monitoring before we get into the details of Hu-Friedy’s SporeCheck.
Sterilization Monitoring 101
![Page 5: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/5.jpg)
Why use sterilization monitoring?
• Various factors may potentially cause the failure of instruments to be properly sterilized:
– Poor office procedures & safeguards– Lack of staff training – Improper cleaning of instruments– Improper packaging or incorrect packaging materials – Overloading or improper loading– Failure to allow adequate separation in the sterilizer– Improper time and/or temperature– The sterilizer may not be functioning properly
![Page 6: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/6.jpg)
What is Monitoring?
According to CDC Guidelines …
• Monitoring of sterilization procedures should include a combination of process parameters, including mechanical, chemical, and biological.
• These parameters evaluate both the sterilizing conditions and the procedure's effectiveness.
![Page 7: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/7.jpg)
Types of Sterilization Monitoring
Mechanical• Mechanical techniques for monitoring sterilization
include assessing cycle time, temperature, and pressure by observing the gauges or displays on the sterilizer and noting these parameters for each load prior to unloading the sterilizer.
• Correct readings do not ensure sterilization, but incorrect readings can be the first indication of a problem.
![Page 8: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/8.jpg)
Types of Sterilization Monitoring
Chemical• Chemical indicators, internal and external, use sensitive
chemicals to assess physical conditions (e.g., time and temperature) during the sterilization process.
• Although chemical indicators do not prove sterilization has been achieved, they allow detection of certain equipment malfunctions, and they can help identify procedural errors. – It is recommended that each instrument pack have both
internal and external indicators.
IMS-1240
![Page 9: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/9.jpg)
Types of Sterilization Monitoring
Biological Monitors (Spore Tests)• Biological indicators (BIs) are the most accepted method
for monitoring the sterilization process because they assess it directly by killing known highly resistant microorganisms.
• Because spores used in BIs are more resistant and present in greater numbers than the common microbial contaminants found on patient-care equipment, an inactivated BI indicates other potential pathogens in the load have been killed.
SporeCheck
![Page 10: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/10.jpg)
When do I use Biological Monitors?
• Correct functioning of sterilization cycles should be verified for each sterilizer by the periodic use (at least weekly) of BIs.
• Every load containing implantable devices should be monitored.
• In-office biological monitoring and mail-in sterilization monitoring services can also be used to test both the BI and the control.
![Page 11: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/11.jpg)
Biological Monitoring
Who recommends Biological Monitoring?
• CDC (Center for Disease Control)• OSHA (Occupational and Safety Health Administration)• ADA (American Dental Association)• OSAP (Organization of Safety and Asepsis Procedures)• VA (Veteran’s Affairs)• AAMI (Assoc. for the Advancement of Medical
Instrumentation)• AORN (Association of Operating Room Nurses)• Many state and local agencies
![Page 12: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/12.jpg)
State Requirements For Spore Testing
Some state dental boards mandate spore testing at an indicated frequency:
Weekly:
Alabama
California
Georgia
Hawaii
Idaho
Kansas
Louisiana
Massachusetts
Minnesota
Mississippi
North Carolina
New Mexico
Ohio
Oregon
Pennsylvania
South Carolina
South Dakota
Texas
Washington
Wisconsin
Monthly:
Arkansas Florida
Indiana
Mandated (no frequency indicated):
Oklahoma Tennessee
![Page 13: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/13.jpg)
How Do I Use An In-Office Biological Monitoring System?
![Page 14: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/14.jpg)
Step 1
Note the sterilizer number and load number on the BI vial label.
![Page 15: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/15.jpg)
Steps 2-4
• Step 2 - Place the BI vial inside an instrument tray, peel pouch, cassette, or challenge pack.
• Step 3 - Test the most challenging area in the sterilizer (the bottom self near the door or over the drain of a large sterilizer)
• Step 4 – Process the load according to sterilizer manufacturer’s instructions.– BI testing should be done in
a fully loaded sterilizer.– Never run a BI test on a
sterilizer that has not been properly warmed up.
![Page 16: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/16.jpg)
Step 5-6
• Step 5 - Retrieve the BI Vial and confirm that the chemical indicator has turned brown. If not, re-run with new BI vial.
• Step 6 – After the vial has cooled, activate the processed BI vial and a control (unprocessed) BI vial by using the vial crusher.
![Page 17: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/17.jpg)
Steps 7-8
• Step 7 – Incubate for 24 hours both the control and the processed BI vials.
• Step 8 – Test Results– No color change in
purple media after processing indicates proper sterilization.
– Color change to yellow indicates a positive result and should be reported immediately. Control vials should be yellow.
Negative (good)
Positive (bad)
(-)Good
(+)Bad
![Page 18: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/18.jpg)
Step 9
• Record any negative (purple) results in the Sterilizer Record Notebook.
• Any positive (yellow) results should be recorded and reported immediately to a supervisor.
![Page 19: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/19.jpg)
What if a test fails?
• If the chemical indicator fails to change color or the spore test result is positive (yellow), the following steps are recommended:– Take the sterilizer out of service– Ascertain that the sterilization process was
followed in order to rule out “operator error”– Correct any identified procedural errors and
retest the sterilizer using biological, mechanical, and chemical indicators.
• If the repeat tests are negative (purple), the sterilizer can be returned to service.
![Page 20: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/20.jpg)
Additional Testing
• Additional testing is recommended when any changes are made in sterilization procedures.
• When a new sterilizer is used, a long duration of time exists between uses, or a repaired unit has been returned to operation.
• Every time an implant is run in a load.
• Anytime there is a positive test result. (color change from purple to yellow)
![Page 21: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/21.jpg)
Introducing Hu-Friedy’s SporeCheck
Record Notebook
Dry Block Incubator
Vial Crusher
Self-ContainedSteam BiologicalIndicators (25 vials)
What is SporeCheck?
A quick and accurate approach to steam sterilizer testing. SporeCheck is an easy-to-use, in-office, biological monitor testing system with results in 24 hours, the fastest on the market. It is the most efficient and reliable way to validate the proper functioning of your steam sterilizer.
![Page 22: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/22.jpg)
Features and Benefits
Features BenefitsIncubation test results in only 24 hours.
Fastest results in biological monitoring. Mail-in tests take over a week and other in-office tests can take 48 hours or more.
Testing and results in office. Complete control over all processes and gets the results first-hand. In-office testing supports high quality standards.
Meets all federal, state and local guidelines for sterilizer testing.
Regulatory compliance is an important element of a smooth-working, quality practice.
![Page 23: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/23.jpg)
Features and Benefits
Features BenefitsSimple to use. Any staff member can be
easily and quickly trained in only minutes to conduct the BI testing and record results.
Reliable testing results. Everyone has peace of mind that the equipment used to sterilize instruments is working properly. Good records are the best insurance policy that can be purchased.
Convenience. In-office processing saves the hassle of mailing, waiting, and not promptly knowing the results.
![Page 24: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/24.jpg)
Features and Benefits
Features BenefitsSupports all steam sterilization processes.
One vial does it all. SporeCheck can be used with all --- gravity, pre-vacuum, and flash (Statim) sterilizers.
Visible color change when results are positive.
The results are clear and require no interpretation. When the vial turns from purple to yellow after incubation, the test results are positive.
Fits existing dry block incubators.
Conversion of current users of in-office BI testing is easy since our vials will fit existing dry block incubators.
![Page 25: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/25.jpg)
Summary
• Hu-Friedy’s SporeCheck is a quick and accurate approach to in-office steam sterilization monitoring.
• It is the fastest biological indicator (BI) monitoring system on the market, providing reliable results in 24 hours while being in compliance with local, state and federal guidelines.
• The SporeCheck starter kit includes a record notebook, self-contained steam biological indicators, vial crusher, and dry block incubator.
![Page 26: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/26.jpg)
Conclusion
As a clinician, SporeCheck:
• Helps to optimize office flow which reduces stress• Helps to manage assets and operational costs• Allows for safer practices• Creates a professional and marketable
environment• All while promoting a positive and dedicated dental
team
![Page 27: SporeCheck](https://reader035.fdocuments.us/reader035/viewer/2022062513/5577880dd8b42acf398b4c15/html5/thumbnails/27.jpg)
Qualifying Question
• Thank you for taking the time to learn more about Hu-Friedy’s SporeCheck.
• To earn your Hu-Points towards a product evaluation, please exit and answer the following question on the Member Resources Page:
In this presentation, we illustrated the steps in the SporeCheck Monitoring process. Which step did we use the vial crusher?
a) Step 2b) Step 3c) Step 6d) Step 7e) Step 9