Solara® LubeFree Handpiece Cleaning & SterilizationCleaning Diamond Burs: - After each use, clean...

2
Air and Water Adjustment 1. Air Pressure – WARNING: To prevent premature damage to the turbine, do not exceed 34 psi when using a HiFlo swivel. Using an air pressure gauge, StarDental Part #262591 for 4/6-Line,set pressure for to 30-34 psi for HiFlo swivels and for all non-HiFlo swivels set pressure to 34-42 psi. If a gauge is not available,set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing.(See Specifications for recommended operational pressure.) For example, if you want to operate at 34 psi and you are using either 5 feet or 12 feet of tubing, then the pressure should be set at the delivery head to compensate for the air pressure losses. 5 feet tubing 34 psi + 5 psi = 39 psi 12 feet tubing 34 psi + 12 psi = 46 psi 2. Water Pressure – Adjust water volume at the delivery head to obtain a satisfactory spray mist. Handpiece Operation Cleaning & Sterilization CAUTION: Prior to cleaning, let lamp cool a few minutes, then remove. After cleaning, let lamp dry completely before reinstalling. Note: Do not attempt to disassemble handpiece or swivel. The Solara Handpieces contain a sophisticated LubeFree ceramic turbine bearing. Lubrication of the bearing is never required.This LubeFree feature insures that residual lubricant is never expelled into the oral cavity. CLEANING External Surfaces To remove debris from external surfaces,clean using a small toothbrush with isopropyl alcohol.Dry thoroughly. PLEASE NOTE: to ensure that your High-Speed chuck continues to be in optimum working condition, clean once a week with Junk-Out High-Speed Chuck Cleaner Kit P/N 263790: Shake well before use - Install brush to handle. See diagram at right. - Use one new brush per cleaning application. - Do not reuse brushes. - Remove bur from chuck. - Dip brush into bottle. - Remove excess cleaner from the brush. - Insert brush into bur end of turbine. - Clean gently by rotating and inserting the brush in and out of the turbine. - Repeat process if necessary. Reseal cap on bottle immediately after use to reduce evaporation. CLEANING Optical Surfaces Every exposed optical surface of the Solara Fiber Optic Handpiece accumulates debris which ultimately decreases light output.To maintain maximum intensity,clean each surface with an alcohol soaked cotton swab.Use firm rubbing motion to remove debris and mineral deposits.See the following illustration denoting each of the seven surfaces to clean. CLEANING the Mating Parts of the Swivel To optimize operation of the swivel disconnect mechanism, swab all internal surfaces of handpiece with a cotton swab soaked in isopropyl alcohol until clean. Dry using syringe air. Clean area from bearing/spring assembly at rear of swivel using isopropyl alcohol and a toothbrush. Dry using syringe air. CLEANING the Swivel 1. Disconnect handpiece from swivel;disconnect swivel from tubing by rotating coupler in opposite direction as shown in Handpiece Installation. Depress foot control; verify water exits tubing. 2. Connect swivel to tubing by following steps in Handpiece Installation. Depress foot control; if almost no water exits front of swivel,clean as follows. 3. Remove swivel and find water tube on front face of swivel. See illustration. 4. Insert cleaning wire, StarDental Part #256719,into water tube.Move wire back and forth to loosen debris at blockage. 5. Remove wire and apply a blast of air into water tube with syringe air. Bearing/Spring Assy. Clean Clean Internal Surfaces Water Tube 4-line HiFlo F.O. Swivel 7 Fiber Optic Surfaces Swivel (1) Fiber Optic Lamp (7) (6) (5) (4) (2) (3) Solara ® High Speed Handpieces are designed to fit a StarDental ® 360° quick disconnect swivel. The Solara ® connects to standard 4-line ISO fiber optic tubing and standard 4/6-line ISO fiber optic tubing. • Specifications • Handpiece Installation • Handpiece Operation • Safety Rules • Cleaning & Sterilization • Troubleshooting Contents Type: High speed fiber optic with a swivel & replaceable turbine.(Swivel sold separately.) Operating Air Pressure: 30 to 34 psi Air Consumption: 2.0 SCFM @ 34 psi No Load Speed: 430,000 rpm Connection: 360° quick disconnect swivel with 4-line ISO configuration & fiber optics. 4/6 line ISO fiber optic tubing. Power: 18 watts Weight: 32 g Specifications (Nominal) Note: Filter air and water supplies, recommended filtration:25 microns.Soften and deionize water.Be sure filters operate efficiently. Handpiece Installation 1. Grasp the swivel in front of the bearing assembly between the thumb and index finger as shown. 2. Slide tubing coupler until the end of the ISO connector is exposed; hold in hand as shown with fingers securing the tubing. 3. Align and insert swivel tubes into the ISO connector holes. 4. While applying slight pressure to hold fittings in the connector holes, slide coupler forward and engage threads. Continue threading coupler clockwise until hand tight against the swivel. 5. Using the wrench as shown, tighten the swivel to tubing coupler until snug. Be sure to turn tubing coupler, not swivel. DO NOT OVER TIGHTEN. Connect handpiece to swivel. Note: This joint loosens over time, resulting in loss of power. Check for tightness periodically. Swivel ISO Connector Tubing Coupler Tubing Bearing Assembly Wrench Tubing Coupler Turn tubing coupler while holding swivel in place 1 2 3 CLEANING Burs NOTE: Burs Must Be Cleaned and Sterilized After Each Patient. Cleaning Diamond Burs: - After each use, clean the diamond bur by running it against a wet diamond cleaning stone. -To mechanically clean,the diamond bur should be held in a solution of either enzymatic presoak or water. - Rinse the bur thoroughly and allow it to dry. - Sterilize – See Sterilization Cleaning Carbide Burs: - After each use,presoak the bur in solution of either enzymatic presoak or water. - Rinse the bur thoroughly and allow it to dry. -The bur may be cleaned with an autoclavable bur brush to remove any resistant debris from the bur. - Sterilize – See Sterilization • Turbine Replacement • Replacement Part Numbers • Warranty Sterilizable up to 135°C 135°C WARNING: Always insert the bur full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use and cause injury to you or your patient. CAUTION: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level.This is an indication of a failing turbine.A failing turbine will reduce the effiency of the handpiece and may result in end cap heat up. See "Turbine Replacement". Use standard length concentric diamond and carbide burs with .0628" nominal diameter friction grip shanks designed for high speed handpieces.Every StarDental Diamond Bur meets specifications below for recommended lengths. Contact your dealer for a complete list of StarDental Diamond and/or Carbide Burs. StarDental autochuck turbines have the following bur requirements: Autochuck Bur Changing System Maximum 24.0 mm OVERALL LENGTH Minimum 7.3 mm Shank Length Use only burs with hardened, tempered steel shanks or carbide shanks. Every StarDental Bur maintains required hardness. Before use, verify concentric rotation by operating handpiece outside the oral cavity in a safe place. Note: Do not use burs that produce excessive noise or do not run concentrically. Do not use burs that are modified, eccentrically designed, or have bent or soft(non-heat treated) shanks. Autochuck Bur Changing System 1. Insert bur into autochuck until resistance is felt. 2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops. 3. Release end cap button. Now,apply light inward force to the dental bur.This will help set the chucking mechanism for initial start up. (Always tug on bur after it has been installed to confirm that it is secured). CAUTION: Never depress the autochuck end cap button or rest the end cap against any part of the oral anatomy during handpiece operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and result in end cap heat-up and/ or loss of bur retention. 1. To remove, grasp handpiece firmly so your thumb touches the end cap and push until a slight ‘click’ is noticed. 2. Pull bur straight out. 3. Release end cap. Safety Rules PLEASE OBSERVE SAFETY RULES 1. Always tug on bur after chuck is loaded to confirm that the bur is secured. 2. Before use, verify concentric rotation by operating the handpiece outside the oral cavity in a safe place. Do not use burs that do not run concentrically or produce excessive noise. 3. Never operate handpiece without a bur installed. 4. Always use burs with an overall length of less than 24 mm.The shank length must be at least 7.3 mm. This includes most short shank burs. 5. When using the autochuck, never depress end cap while bur is rotating. 6. Use only recommended burs to prevent damage to autochuck and in some cases to prevent burs from improper or incorrect placement.An improperly clamped bur may dislodge during use,causing injury to you or your patient.A sudden high-pitched sound or increased vibration while handpiece is operating alerts the operator to the possibility that the chuck is not clamped sufficiently to retain the bur or a bent bur is inserted. STERILIZATION Handpiece and burs Must be cleaned and sterilized after each patient. Note: To Prevent Damage To the Bur, Use an Autoclavable Bur Block. A wrapped handpiece or bur should be steam sterilized at 132° C for a minimum of 10 minutes or at 121° C for a minimum of 50 minutes. Do not exceed 135° C. An unwrapped handpiece or bur should be steam sterilized at 132° C for a minimum of 8 minutes or at 121° C for a minimum of 30 minutes. Do not exceed 135° C. Note: An unwrapped handpiece or bur needs to be used immediately or sterility is not maintained. Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or paper/plastic bags. Always use complete dry cycle of sterilizer. Always allow product to cool in bag to room temperature prior to using. Contact your sterilizer manufacturer if moisture remains inside bag and on product after drying. CAUTION: Use steam sterilization exclusively for complete sterility. Never dry heat sterilize, ultrasonically clean, or expose handpiece, bur or swivel to chemical disinfection or cold sterilization by: • Immersion techniques or Surgical milk • Surface wipes, Hand soaps, or Detergents • Surface sprays • Sterilization bags containing disinfectants MAKES YOUR PRACTICE PERFECT® Solara® LubeFree Handpiece Instruction Manual ® ® Division of DentalEZ Group Lancaster, PA 17601 U.S.A. Printed in the U.S.A. ©2010 DentalEZ, Inc. DentalEZ Dental Products (GB) Ltd., Cleveland Way, Hemel Hempstead, HP2 7DY, ENGLAND Caution: Federal law restricts this device to sale by or on the order of a dentist. 866.DTE.INFO (383.4636) www.DentalEZ.com 0344 DO NOT USE FOR ORAL SURGERY

Transcript of Solara® LubeFree Handpiece Cleaning & SterilizationCleaning Diamond Burs: - After each use, clean...

Page 1: Solara® LubeFree Handpiece Cleaning & SterilizationCleaning Diamond Burs: - After each use, clean the diamond bur by running it against a wet diamond cleaning stone. - To mechanically

Air and Water Adjustment1. AirPressure–

WARNING: To prevent premature damage to the turbine, do not exceed 34 psi when using a HiFlo swivel.

Usinganairpressuregauge,StarDentalPart#262591for4/6-Line,setpressureforto30-34psiforHiFloswivelsandforallnon-HiFloswivelssetpressureto34-42psi.

Ifagaugeisnotavailable,setpressureatdeliveryheadallowing1psipressurelossforeveryfootoftubing.(SeeSpecifications forrecommendedoperationalpressure.)

Forexample,ifyouwanttooperateat34psiandyouareusingeither5feetor12feetoftubing,thenthepressureshouldbesetatthedeliveryheadtocompensatefortheairpressurelosses.

5feettubing 34psi+5psi=39psi

12feettubing 34psi+12psi=46psi

2. WaterPressure–

Adjustwatervolumeatthedeliveryheadtoobtainasatisfactoryspraymist.

Handpiece Operation

Cleaning & Sterilization

CAUTION: Prior to cleaning, let lamp cool a few minutes, then remove. After cleaning, let lamp dry completely before reinstalling.Note:Do not attempt to disassemble handpiece or swivel.

TheSolara HandpiecescontainasophisticatedLubeFreeceramicturbinebearing.Lubricationofthebearingisneverrequired.ThisLubeFreefeatureinsuresthatresiduallubricantisneverexpelledintotheoralcavity.CLEANING External SurfacesToremovedebrisfromexternalsurfaces,cleanusingasmalltoothbrushwithisopropylalcohol.Drythoroughly.

PLEASE NOTE: to ensure that your High-Speed chuck continues to be in optimum working condition, clean once a week with Junk-Out High-Speed Chuck Cleaner Kit P/N 263790:Shake well before use-Installbrushtohandle.Seediagramatright.-Useonenewbrushpercleaningapplication. - Do not reuse brushes.-Removeburfromchuck. -Dipbrushintobottle.-Removeexcesscleanerfromthebrush. -Insertbrushintoburendofturbine.-Cleangentlybyrotatingandinsertingthebrushinandoutoftheturbine.-Repeatprocessifnecessary.Reseal cap on bottle immediately after use to reduce evaporation.

CLEANING Optical SurfacesEveryexposedopticalsurfaceoftheSolara FiberOpticHandpieceaccumulatesdebriswhichultimatelydecreaseslightoutput.Tomaintainmaximumintensity,cleaneachsurfacewithanalcoholsoakedcottonswab.Usefirmrubbingmotiontoremovedebrisandmineraldeposits.Seethefollowingillustrationdenotingeachofthesevensurfacestoclean.

CLEANING the Mating Parts of the SwivelTooptimizeoperationof theswiveldisconnectmechanism,swaball internalsurfacesofhandpiecewithacottonswabsoakedinisopropylalcoholuntilclean.Dryusingsyringeair.Cleanareafrombearing/springassemblyatrearofswivelusingisopropylalcoholandatoothbrush.Dryusingsyringeair.

CLEANING the Swivel1. Disconnecthandpiecefromswivel;disconnectswivelfromtubingbyrotatingcouplerinopposite

directionasshowninHandpiece Installation.Depressfootcontrol;verifywaterexitstubing.2. ConnectswiveltotubingbyfollowingstepsinHandpiece Installation.Depressfootcontrol;if

almostnowaterexitsfrontofswivel,cleanasfollows.3. Removeswivelandfindwatertubeonfrontfaceofswivel.

Seeillustration.4. Insertcleaningwire,StarDentalPart#256719,intowatertube.Movewirebackandforthtoloosen

debrisatblockage.5. Removewireandapplyablastofairintowatertubewithsyringeair.

Bearing/Spring Assy.Clean

Clean Internal Surfaces

Water Tube

4-line HiFlo F.O. Swivel

7 Fiber Optic Surfaces

Swivel

(1)

Fiber Optic Lamp

(7) (6) (5)

(4)

(2)

(3)

Solara®HighSpeedHandpiecesaredesignedtofitaStarDental®360°quickdisconnectswivel.TheSolara® connectstostandard4-lineISOfiberoptictubingandstandard4/6-lineISOfiberoptictubing.

•Specifications•HandpieceInstallation

•HandpieceOperation•SafetyRules

•Cleaning&Sterilization•Troubleshooting

Contents

Type: Highspeedfiberopticwithaswivel&replaceableturbine.(Swivelsoldseparately.) OperatingAirPressure: 30to34psi AirConsumption: 2.0SCFM@34psi NoLoadSpeed: 430,000rpm Connection: 360°quickdisconnectswivelwith4-lineISOconfiguration&fiberoptics. 4/6lineISOfiberoptictubing. Power: 18watts Weight: 32g

Specifications (Nominal)

Note: Filter air and water supplies, recommended filtration: 25 microns. Soften and deionize water. Be sure filters operate efficiently.

Handpiece Installation

1. Grasptheswivelinfrontofthebearingassemblybetweenthethumbandindexfingerasshown.

2. SlidetubingcoupleruntiltheendoftheISOconnectorisexposed;holdinhandasshownwithfingerssecuringthetubing.

3. AlignandinsertswiveltubesintotheISOconnectorholes.

4. Whileapplyingslightpressuretoholdfittingsintheconnectorholes,slidecouplerforwardandengagethreads.Continuethreadingcouplerclockwiseuntilhandtightagainsttheswivel.

5. Usingthewrenchasshown,tightentheswiveltotubingcoupleruntilsnug.Besuretoturntubingcoupler,notswivel.DONOTOVERTIGHTEN.Connecthandpiecetoswivel.

Note:This joint loosens over time, resulting in loss of power. Check for tightness periodically.

Swivel ISO Connector

Tubing Coupler

Tubing

BearingAssembly

Wrench

Tubing Coupler

Turn tubing coupler while holding swivel in place

1

2

3

CLEANING BursNOTE:BursMustBeCleanedandSterilizedAfterEachPatient.CleaningDiamondBurs: -Aftereachuse,cleanthediamondburbyrunningitagainstawetdiamondcleaningstone. -Tomechanicallyclean,thediamondburshouldbeheldinasolutionofeitherenzymaticpresoakorwater. -Rinsetheburthoroughlyandallowittodry. -Sterilize–SeeSterilizationCleaningCarbideBurs: -Aftereachuse,presoaktheburinsolutionofeitherenzymaticpresoakorwater. -Rinsetheburthoroughlyandallowittodry. -Theburmaybecleanedwithanautoclavableburbrushtoremoveanyresistantdebrisfromthebur. -Sterilize–SeeSterilization

•TurbineReplacement•ReplacementPartNumbers•Warranty

Sterilizableupto135°C135°C

WARNING:Alwaysinserttheburfulllengthofthechuck.Ifyouattempttoextenditfromthechuck,theburmaydislodgeduringuseandcauseinjurytoyouoryourpatient.

CAUTION: Replacetheturbinewhenthehandpiecefailstoachievefullspeedand/orincreasessharplyinnoiselevel.Thisisanindicationofafailingturbine.Afailingturbinewillreducetheeffiencyofthehandpieceandmayresultinendcapheatup.See"TurbineReplacement".

Use standard length concentric diamond and carbide burs with .0628" nominal diameterfrictiongripshanksdesignedforhighspeedhandpieces.EveryStarDentalDiamondBurmeetsspecificationsbelowforrecommended lengths.Contactyourdealer foracomplete listofStarDentalDiamondand/orCarbideBurs.StarDentalautochuckturbineshavethefollowingburrequirements:

Autochuck Bur Changing SystemMaximum 24.0 mm

OVERALL LENGTH

Minimum7.3 mm

Shank Length

Useonlyburswithhardened,temperedsteelshanksorcarbideshanks.EveryStarDentalBurmaintainsrequiredhardness.Beforeuse,verifyconcentricrotationbyoperatinghandpieceoutsidetheoralcavityinasafeplace.

Note: Do not use burs that produce excessive noise or do not run concentrically. Do not use burs that are modified, eccentrically designed, or have bent or ‘soft’ (non-heat treated) shanks.

Autochuck Bur Changing System1. Insertburintoautochuckuntilresistanceisfelt.

2.Depressautochuckendcapbuttonandcontinuetofeed-inburuntilitclicksandstops.

3.Releaseendcapbutton.Now,applylightinwardforcetothedentalbur.Thiswillhelpsetthechuckingmechanismforinitialstartup.(Alwaystugonburafterithasbeeninstalledtoconfirmthatitissecured).

CAUTION: Never depress the autochuck end cap button or rest the end cap against anypart of the oral anatomy during handpiece operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and result in end cap heat-up and/or loss of bur retention.

1. To remove,grasphandpiecefirmlysoyourthumbtouchestheendcapandpushuntilaslight‘click’isnoticed.

2.Pullburstraightout.

3.Releaseendcap.

Safety RulesPLEASE OBSERVE SAFETY RULES1.Alwaystugonburafterchuckisloadedtoconfirmthattheburissecured.2.Beforeuse,verifyconcentricrotationbyoperatingthehandpieceoutsidetheoralcavityinasafeplace.Donotusebursthatdonot

runconcentricallyorproduceexcessivenoise.3.Neveroperatehandpiecewithoutaburinstalled.4.Alwaysuseburswithanoveralllengthoflessthan24mm.Theshanklengthmustbeatleast7.3mm.Thisincludesmostshort

shankburs.5.Whenusingtheautochuck,neverdepressendcapwhileburisrotating.6. Useonlyrecommendedburstopreventdamagetoautochuckandinsomecasestopreventbursfromimproperor incorrect

placement.Animproperlyclampedburmaydislodgeduringuse,causinginjurytoyouoryourpatient.Asuddenhigh-pitchedsoundorincreasedvibrationwhilehandpieceisoperatingalertstheoperatortothepossibilitythatthechuckisnotclampedsufficientlytoretaintheburorabentburisinserted.

STERILIZATIONHandpieceandbursMustbecleanedandsterilizedaftereachpatient.Note:To Prevent Damage To the Bur, Use an Autoclavable Bur Block.Awrappedhandpieceorburshouldbesteamsterilizedat132°Cforaminimumof10minutesorat121°Cforaminimumof50minutes.Donotexceed135°C.Anunwrappedhandpieceorburshouldbesteamsterilizedat132°Cforaminimumof8minutesorat121°Cforaminimumof30minutes.Donotexceed135°C.Note:An unwrapped handpiece or bur needs to be used immediately or sterility is not maintained.

Donotexceedthesterilizermanufacturer’sguidelinesforloadcapacity.Useonlypaperorpaper/plasticbags.Alwaysusecompletedrycycleofsterilizer.Alwaysallowproducttocoolinbagtoroomtemperaturepriortousing.Contactyoursterilizermanufacturerifmoistureremainsinsidebagandonproductafterdrying.CAUTION: Use steam sterilization exclusively for complete sterility. Never dry heat sterilize, ultrasonically clean, or expose handpiece, bur or swivel to chemical disinfection or cold sterilization by:

• Immersion techniques or Surgical milk• Surface wipes, Hand soaps, or Detergents

• Surface sprays• Sterilization bags containing disinfectants

M A K E S Y O U R P R A C T I C E P E R F E C T ®

Solara® LubeFree HandpieceInstruction Manual

® ®

Division of DentalEZ GroupLancaster, PA 17601 U.S.A.Printed in the U.S.A.©2010 DentalEZ, Inc.

DentalEZ Dental Products (GB) Ltd.,Cleveland Way,Hemel Hempstead,HP2 7DY, ENGLAND

Caution: Federal law restricts this device to sale by or on the order of a dentist. 866.DTE.INFO (383.4636)

www.DentalEZ.com0344

DO NOT USE FOR ORAL SURGERY

Page 2: Solara® LubeFree Handpiece Cleaning & SterilizationCleaning Diamond Burs: - After each use, clean the diamond bur by running it against a wet diamond cleaning stone. - To mechanically

Problem: IMPROPER SPRAY Cause: Cloggedfilters,tubesorwaterports Correction: Inspectandcleanwaterfilterindeliveryhead.Disconnecthandpiece. Depressfootcontrol;verifywaterexitstheswivel. (Ifwaterdoesnotcomefromswivel,refertoCleaning the Swivel.)

Cause: Brittlehandpiecesiliconerubberseal Correction: Replacesiliconerubberseal

Cause: SwivelO-ringsworn Correction: ReplaceO-rings,StarDentalPart#259884orStarDental Part#263771forHiFloswivels.

Cause: Cloggedsprayports Correction: Useclean-outwiretounclogports.

See "Turbine Replacement" for Replacement Instructions

Problem: BUR LOCKS IN AUTOCHUCK Cause: Debrisbuild-upinchuckingmechanism Correction: CleanwithJunk-OutChuckCleanerPartNumber263790

Cause: Useofnon-ISOburs Correction: AlwaysuseISOburs.

Problem: BUR SLIPS IN AUTOCHUCK Cause: Debrisbuild-upinchuckingmechanism Correction: CleanwithJunk-OutChuckCleanerPartNumber263790

Cause: Useofnon-ISOburs Correction: AlwaysuseISOburs.

Note:Always use ISO burs to prevent locking and slipping of burs in autochuck. If problems of locking and slipping continue, replace worn autochuck immediately with LubeFree Autochuck Turbine Replacement Kit, StarDental Part #263739. If any problem persists, contact dealer for further assistance.

Troubleshooting

Turbine ReplacementNote:Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level. The replacement cartridge is available from your dealer and is easily replaced in the office. When ordering a replacement, be sure to order StarDental Autochuck Turbine Replacement Kit, StarDental Part #263739.

1.Removeoldendcapandturbineorcanisterassembly.

2. Inspecthousingbeforecleaning.Removeanyobviousdebristopreventdamageandtoinsureperformanceofturbine.

3. Connecthandpieceandrunairthroughturbinehousingwithoutturbineinstalledfor5seconds.

4. Cleanhousingthoroughlywithcottonswabandisopropylalcohol.Drywithsyringeair.

5. Placeturbineassemblyintohousingandgentlyinsertasshown.

6. Positionandengagethreadsofendcapintothehousingbyhanduntilsnug.

7. TightenendcapwiththecombinationendcapwrenchemployingthewrenchopeningmarkedACasshown.Positionthewrenchwithmarkedsidefacingendcapforbestwrenchcontrol.

Note:Swivel assemblies sold separately.

SOLARASolaraFiberOpticHandpiecewithLubeFreeAutochuckTurbine...................................................................................263718LubeFreeVortexAutochuckTurbineReplacementKit.....................................................................................................263739SolaraReplacementEndCap...........................................................................................................................................263744EndCapWrench...............................................................................................................................................................203048

HiFlo 4-LINE SWIVELSwivelAssembly:HiFlo4-lineFiberOptic........................................................................................................................263758RearSeal,HiFlo4-lineFiberOpticSwivel(Packageof3)................................................................................................263734SwivelO-RingKitforHiFloSwivel(2Large,2Small)......................................................................................................263771

HiFlo 6-PIN SWIVELHiFlo6-PinSwivel,3.3volts.............................................................................................................................................263772HiFlo6-PinSwivel,3.1volts.............................................................................................................................................263773RearSeal,HiFlo6-pinFiberOpticSwivel(Packageof3)................................................................................................263770SwivelO-RingKitforHiFloSwivel(2Large,2Small)......................................................................................................263771

ADDITIONAL REPLACEMENT PARTSLampwithspring(Packageof2)....................................................................................................................................260986ReplacementLamp,6-pin 3.3VdcforStarBrightF.O.(Packageof2)...............................................................................262522ReplacementLamp,6-pin 3.1Vdcfornon-Startubing(Packageof2).............................................................................262962SwivelO-RingKitfor259716,262530,and263255(2Large,2Small)............................................................................259884AirPressureGauge(ISO4/6-line).....................................................................................................................................262591CleaningWire(Gauge:0.008,Packageof2).....................................................................................................................256719ISOTubingSetScrew(Packageof2)................................................................................................................................261355AllenWrench.035............................................................................................................................................................280830Junk-OutHigh-SpeedChuckCleaner...............................................................................................................................263790

Always insist on StarDental Parts for best performance.

Replacement Part Numbers

Description Item Number

For new products, manuals and technical information, call 1-866-DTE-INFO or visit our web site @ www.dentalez.com

Problem: NO LIGHT OR INTERMITTENT LIGHT Cause: Electricalconnections Correction: Verifyconnections:walloutlet-powerpack,powerpack-lampcontrol,andlampcontrol-tubingconnector.

Cause: Crimpedsignalairlinetolampcontrol Correction: Straightensignalairlinetoallowairflow.

Cause: COLORight® Tubing,oilonlampcontactpoints Correction: Removelampfromtubing;cleanspring,frontandrearsocketsandlampcontactswithisopropylalcohol.

Cause: COLORight Tubing,lampburnedout Correction: ReplaceKrypton-HalogenlampwithStarDental®Part#260986.

Cause: COLORight Tubing,wirebrokenattubingcoupler Correction: Re-stripwire1/8"andlockwithsetscrew,StarDentalPart#261355.Use.035Allenwrench, StarDentalPart#280830.

Cause: StarBright® Tubing,corrosiononlampcontactpins Correction: Removelamp(locatedintheStarBrightswivelconnector)andcleancontactpinswithisopropylalcohol.

Cause: StarBright Tubing,lampburnedout Correction: ReplaceXenon-HalogenlampwithStarDentalPart#262522,or,whenoperatingwithanon-StarDentalF.O. system,useStarDentalPart#262962.(ConsultyourOperator’sManualforcorrectvoltage.)

Problem: YELLOWISH LIGHT Cause: Debrisonopticalsurfaces

Correction: Cleanallexposedopticalsurfaces.(RefertoCleaning.)

Problem: REDUCED LIGHT Cause: Lowvoltage

Correction: RefertoCOLORightorStarBright Fiber Optic Electrical System & Handpiece Tubing Instruction Manual or contactStarDentalCustomerServiceat1-866-DTE-INFO.

Cause: COLORightSystem,lamp

Correction: ReplaceKrypton-HalogenlampwithStarDentalPart#260986.

Cause: StarBright System,lamp Correction: Replacelampin6-pinswivelwithStarDentalPart#262522,or,whenoperatingwithanon-StarDentalF.O. system,useStarDentalPart#262962.(ConsultyourOperator’sManualforcorrectvoltage.)

Problem: LACK OF POWER Cause: Airpressuretoolow

Correction: Verifyproperhandpieceairpressure.(RefertoHandpiece Installation.)

Cause: Crimpedtubing

Correction: Straightentubinginsideandoutsidedeliveryhead.

Cause: Debrisbuildup

Correction: Cleaninsideofhousingandendcapwithswabandalcohol.LubricateinsideofhousingwithDentaLube II beforeassembly.

Note:Before cleaning, check for and remove any debris from turbine housing to prevent damage and to insure turbine performance.

Cause: Wornoutturbine

Correction: RefertoTurbine Replacement.

Cause: Swivelnotthreadedfullyontotubing

Correction: RefertoHandpiece Installation.

Problem: WATER LEAKAGE Cause: Brittlehandpiecesiliconerubberseal

Correction: Replacesiliconerubbersealfor4-line,“Green”StarDentalPart#202897,andfor6-line, “Green”StarDentalPart#262520.For4-lineHiFloswivel,replace“Black”StarDental Part#263734andfor6-pinHiFloswivel,“Black”StarDentalPart#263770.

Cause: SwivelO-ringsworn Correction: ReplacebothO-ringsonanyswivelwithStarDentalPart#259884orStarDentalPart #263771fortheHiFloswivels.RefertoO-ring and Rear Seal Replacementfrom HiFlo Swivel Connector Manual, StarDentalPart#063776.

Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE Cause: Buildupofdebris Correction: PlacetwodropsofDentaLube IIintoswivel/tubingconnectorarea.Rotateswiveltocirculatelube.Disengage handpiece.Contactdealerifunabletodisengageswivel.Topreventrecurrence,refertoCleaning

Clean debris from inside of housing

Bearings

Swivel O-rings at the front of the swivel

DentalEZ Dental Products (GB) Ltd., Cleveland Way, Hemel Hempstead, HP2 7DY, ENGLAND063766 C Date of Issue 10/01/2010

Division of DentalEZ Group Lancaster, PA 17601 U.S.A.

Printed in the U.S.A.©2010 DentalEZ, Inc.

Limited WarrantyStarDental® Solara® Handpieces, including swivel andLubeFreeTurbinewithAutochuck, arewarrantedagainstdefects inmaterial andworkmanshipfor12monthsfromdateofpurchase.Thefiberopticsarewarrantedforaperiodof5years.Thiswarrantyissubjecttothefollowingconditions:Handpieces,swivelandLubeFreeTurbinewithAutochuckmustnotbeabusedandmusthavebeenoperatedinaccordancewithproceduresdetailedinthismanual.AnyreplacementpartsmustbegenuineStarDentalparts.Warrantyvalidationcardmustbecompletedandreturnedwithin15daysofpurchase.StarDentalwill,withoutcharge,repairorreplace(atitssolediscretion)anydefectivepart(s)coveredinthiswarranty,providingaboveconditionsaremetandtheproductisreturnedthroughanauthorizedStarDentaldealer.Youmusthaveacopyofyouroriginalinvoicetoqualify.

THISWARRANTYISINLIEUOFALLOTHERWARRANTIES,EXPRESSEDORIMPLIED,INCLUDINGTHEWARRANTIESOFMERCHANTABILITYANDFITNESSFORUSE.

Product complieswith MDD 93/42/EEC

0344

Center Spray Tube

LubeFree Autochuck

Wrench Opening Housing

Spray port

Remove before inserting wire

LubeFree Autochuck

Important Notice About Returning Product To The DentalEZ Group

Anyone wishing to return any DentalEZ Group product to a DentalEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must appear clearly marked on both the outside and inside of the returned product carton. When calling for a return authorization number you will be asked to provide the following information:

To Obtain A Return Authorization

Number Call 1-866-DTE-INFO

• Product Name, Model Number or Part Number• Serial Number of Product• Dealer Return Purchase Order Number• Reason For Return• Copy of Invoice Showing Purchase

Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will help us expedite processing of any returns.