Slit lamp fundoscopy

51
Slit Lamp Fundoscopy Dr. Anton Vurdaft 2014, edited in 2015 FCOphth(ECSA), FICO

Transcript of Slit lamp fundoscopy

Page 1: Slit lamp fundoscopy

Slit Lamp Fundoscopy

Dr. Anton Vurdaft

2014, edited in 2015

FCOphth(ECSA), FICO

Page 2: Slit lamp fundoscopy

Slit Lamp Fundoscopy

Biomicroscopy with simultaneous fundus examination.

Using high-diopter lens (+60D, +78D, +90D, +120D and others).

Ophthalmoscopy with slit lamp is the most difficult skill in ophthalmic examination.

Ophthalmoscopy using slit lamp and the lens gives the most clear picture of the fundus: magnified, 3D, very bright).

Page 3: Slit lamp fundoscopy

Slit Lamp Fundoscopy

Slit lamp with teaching piece

Page 4: Slit lamp fundoscopy

Slit Lamp Fundoscopy

It is possible to do fundoscopy with wide AND with narrow pupil most of the times.

The latter is surely more difficult for inexperienced examiner.

Moreover, it is difficult and sometimes not possible to examine fundus periphery with narrow pupils.

Page 5: Slit lamp fundoscopy

Slit Lamp Fundoscopy Principles

Use slit lamp as you do for anterior segment.

The task is to focus the beam light on the fundus.

The beam is narrow. Examine fundus stepwise.

Everything what is not into the beam light focus – is not seen or seen poorly.

Page 6: Slit lamp fundoscopy

Slit Lamp Fundoscopy Principles

With slit lamp fundoscopy we only see SLIT image, we do not see the picture as a whole at one time. We have to move the slit in certain directions. NOTICE the glare, which is moved out of the picture with a mild tilting of the lens.

Page 7: Slit lamp fundoscopy

Slit Lamp Fundoscopy Technique

Prepare for the examination:

• Make the room as dark as possible• Fixate your arms comfortably, using the

armrest• Dilate the pupils of the patient (important if

you are not experienced) – use tropicamide or combination of tropicamide with phenylephrine

Page 8: Slit lamp fundoscopy

Slit Lamp Fundoscopy Technique

DILATE THE PUPILS!

Page 9: Slit lamp fundoscopy

Slit Lamp Fundoscopy TechniqueThe armrest you can make yourself of any suitable material or involving carpenters.

This is especially important for those who have short hands.

Sometimes you can use the case of high-diopter lens

Page 10: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Set Slit Lamp

First use magnification of x15 or less (i.e. x10). The bigger the magnification, the harder it is for inexperienced examiner to set the focus on the fundus and to orient within the fundus.

Set the light beam as narrow as possible. The narrower the pupil, the narrower the beam should be. The height of the beam is not so important. The width can be regulated during the examination.

Page 11: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Set Slit Lamp

Set the brightness of the beam to the level most comfortable for you (usually it is the brightest) and for the patient (usually patients are more comfortable with darker beam).

Height, width and brightness of the beam can be regulated during the procedure, that is especially important with uncooperative patients and patients with photophobia.

The light-source and the pieces of the slit lamp should be aligned into one axis (90 degree to the patient).

Page 12: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Start Exam

Fixate a chin and a forehead of the patient (using chinrest and headrest).

Ask patient to look straight or in other needed gauze position.

Focus the beam on the center of the pupil.Fix your hand with the lens on a cheek of a patient or

using a headrest. It is more suitable to use left hand for a right eye of a

patient, and vice versa.

Page 13: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Technique

DO NOT USE ONE HAND FOR BOTH EYES. SWITH YOUR HANDS. IT IS EASY TO GET USED TO ONE HAND, AND IT WILL SPOIL YOUR EXAMINATIONS LATER.

Page 14: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Technique

Grab the lens with big and index finger, the rest of the fingers will help you to tilt the lens and pull the upper eyelid up.

Set the center of the lens on the center of the pupil (use the beam to check it).

The working distance of the lens is 5-10 mm.It does not matter what side of the lens should be faced

to the patient, if the lens is double aspheric (volk, ocular instruments). Otherwise – the letters on the lens rim should be directed towards the patient (see the foto).

Page 15: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Technique

The picture is floating in the air actually. Notice the direction of letters of the lens rim.

Page 16: Slit lamp fundoscopy

That is how you focus a beam on the center of a pupil. Notice how the fingers are fixed.

Slit Lamp Fundoscopy: Technique

Page 17: Slit lamp fundoscopy

Slit Lamp Fundoscopy: TechniquePatient should look in the needed direction.

Light source of the microscope moved closely to the patient.

Slowly you pull the microscope back to you.

Focus of the beam is moving from the cornea to the lens, than we see the image of the fundus, it is formed somewhere in the air.Stepwise examination follows.

Page 18: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Technique

Practice makes perfect!

Do at least one dilated and one non-dilated fundoscopy of the disc EVERY DAY.

After 2-3 weeks you may add the macula to your daily MUST DO list.

Page 19: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The ImageThe image which you see – is inverted. It means that you see it upwards down and left-to-right. If you see something in the upper part of your picture – it is actually in the lower part of that part of the fundus, which you are examining. The same goes to “left or right” parameters.

The illustration of that fact is the relation of optic disc and macula. The latter is always temporally (closer to the temple), the disc is nasally (closer to the nose). But with the slit lamp fundoscopy we see the picture inverted – disc is always temporally there.

Page 20: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The Image

Fundus of the left eye (egg), as if we see the eye in horizontal cross-section. As it should be with the left eye – the optic disc is closer to the nose than macula.

Page 21: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The Image

Fundus of the same left eye (egg) with the using of high-diopter “+” lens. Equally to the +90D or +78D, the image gets inverted. Macula and disc have changed their places and got upside-down.

Page 22: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The Image

When you are drawing the fundus picture in the file – there are two ways you can do it. The first one implies mental gymnastics – you may try to picture the findings immediately in the way they are really located in the fundus (direct picture). The other way is to rotate your file on 180 degree and draw the picture JUST exactly how you’ve seen it in inverted way. If you’ll rotate this file back to 180 degree – the findings will be in correct position and direct.Use mental gymnastics! It is useful for a brain and will help you to understand the topography of the fundus better.

Page 23: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The Image

The area of fundus which you can see depends on refractive power of the lens. The more powerful is the lens, the bigger part of fundus you see at a one time. For example, with +60D you will struggle trying to see the periphery with narrow pupil, but +120D lens will easily allow you to see it.

Page 24: Slit lamp fundoscopy

Slit Lamp Fundoscopy: The Image

The field of fundus which is seen – depends not only on the lens, but also on the patient’s cooperation. The more precisely patient follows your instructions, the bigger piece of fundus you will see.It is wise to do the fundoscopy in the next order: optic disc (patient fixates the contralateral ear of the examiner), macula (patient fixates straighth), periphery (patient fixated every needed direction).

Page 25: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

It is most comfortable for the beginner to examine periphery of the retina starting from 12 o’clock meridian going clockwise and seeing the 1-30, 3, 4-30, 6, 7-30, 9 and 10-30 o’clock meridians.

Page 26: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

For the beginner it is difficult to orient within the fundus in a certain gauze direction of a patient. It is quiet easy. Patient looks at the direction of the piece of fundus which you see. Patient looks down – we see the lower periphery. Patient looks up – we see the upper periphery. Etc.

Page 27: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient fixated straight. We receive macula in the field of view.

Page 28: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient fixated the contralateral ear of examiner. We receive optic disc in the center of a fundus picture.

Page 29: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient looked at the ceiling, we see upper periphery (12 o’clock).

Page 30: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient looked at the floor, we see lower periphery (6 o’clock).

Page 31: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient looked left, we see temporal periphery (3 o’clock).

Page 32: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Patient looked right, we see nasal periphery (9 o’clock).

Page 33: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Inversion of the picture of upper periphery

Page 34: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Retinal Periphery

Inversion of the picture of upper periphery

Page 35: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Fundus Structure

Page 36: Slit lamp fundoscopy

Slit Lamp Fundoscopy: Fundus Structure

Page 37: Slit lamp fundoscopy

Fundus Structure

Page 38: Slit lamp fundoscopy

Picturing The Fundus

Optic disc is always nasally, while macula is temporally

Page 39: Slit lamp fundoscopy

Optic Disc Structure

Page 40: Slit lamp fundoscopy

Measuring the Cup of the Disc

Measurement is not so precise, do not struggle

Page 41: Slit lamp fundoscopy

Measuring the Cup of the Disc

For our practice we need approximate cup/disc ratio

Page 42: Slit lamp fundoscopy

Measuring the Cup of the Disc

This cup/disc ratio is approximately 0,7. Notice the absence of the neuroretinal rim from the temporal side

Page 43: Slit lamp fundoscopy

Measuring the Cup of the Disc

This cup/disc ratio is 0,9, almost total (1,0). Notice the absence of the neuroretinal rim everywhere except nasal side of the disc.

Page 44: Slit lamp fundoscopy

Measuring the Cup of the Disc: Stereo

The color is not enough to assess the real cup-disc ratio. Use stereopsis!

Page 45: Slit lamp fundoscopy

Notice the absence of neuroretinal rim at the lower part of the disc

Measuring the Cup of the Disc: Stereo

Page 46: Slit lamp fundoscopy

Measuring the Cup of the Disc: Stereo

Vessels bayoneting demarcates the borders of cupping

Page 47: Slit lamp fundoscopy

Measuring the Cup of the Disc: Stereo

Normal disc picture (10)

Page 48: Slit lamp fundoscopy

Lenses Care:

Page 49: Slit lamp fundoscopy

Lenses Care:

1)Do not touch the lens itself, touch only the rim!

2)Occasionally you may clean the lens surface gently with a smooth tissue without fibers.

3)Put the lens into the case immediately after use!

Page 50: Slit lamp fundoscopy

References:(1) S.Barnard, A.Field, "Slit Lamp Indirect Ophthalmoscopy", American Academy of

Optometry, http://www.academy.org.uk/tutorials/volklens.htm (2) "Practical Ophthalmology - A Manual for Beginning Residents", 5th ed., American

Academy of Ophthalmology, 2005 (3) James Benjamin, "Ophthalmology Investigation and Examination Techniques", 2006 (4) Контактные диагностические линзы и фиксация локтя:

http://terra-ophthalmica.blogspot.com/2012/06/blog-post_21.html (5) Rosalind A. Stevens, Patrick J. Saine, Marshall E. Tyler, "Stereo atlas of fluorescein

and indocyanine green angiography", 1999 (6) William L. Jones, "Peripheral Ocular Fundus (Third Edition)", 2007 (7) Vortex Veins and Blonde Fundus, http://www.optos.com/RecognizingPathology/pages/Visualization/VortexVeins-

8.html(8) S.A. Rizvi, "A simple method to view stereo-images of the optic nerve head on

ordinary computer monitors", Eye (2009)(9) My own fotos

Page 51: Slit lamp fundoscopy

Thank you!