INTRODUCTION

Page 3: Slit lamp biomicroscope

THE SLIT-LAMP BIOMICROSCOPE is a high-power binocular microscope with a slit-shaped illumination source, specially designed for viewing the different optically transparent or translucent tissues of the eye.

It enhances the examination by :

Excellent image quality

Binocular stereoscopic view

Flexible illumination

Flexible magnification

Providing room for specialized attachments and lenses for detailed examination and diagnostic tests

Page 4: Slit lamp biomicroscope

The science of examination with a slit lamp is

called Biomicroscopy as it allows in vivo study of living tissues at high magnification.

Cont…

Page 5: Slit lamp biomicroscope

HISTORICAL LANDMARKS

Page 6: Slit lamp biomicroscope

• 1911, the first concept of slit-lamp was introduced by ALLVAR GULLSTRAND, it was named as “Large reflection-free ophthalmoscope” manufactured by Zeiss.

Prof. Allvar Gullstrand Large Gullstrand Opthalmoscope (1911)

Page 7: Slit lamp biomicroscope

• 1919, VOGT introduced KOEHLER illumination system to Gullstrand slit-lamp.

• 1930, LETZ introduced telescopic loupe built on the principle of Galilean Telescope which helps in increase working distance.

Haag-Streit-1920 Bausch & Lomb slit lamp (1926)

Cont…

Page 8: Slit lamp biomicroscope

• 1933, GOLDANN developed three control element for cross-slide stage and also common swivel axis for microscope and illumination system in HAAG-STREIT Slit-lamp.

• 1938, Joystick was added to Haag-streit.

Haag-Streit-1933 Haag-Streit-1936

Cont…

Page 9: Slit lamp biomicroscope

• 1950, LITTMANN with ZIESS combined joystick control from Goldmann, illumination path from Comberg prism instrument and also added stereo telescope system with magnification changer.

Cont…

Zeiss slit lamp (Littmann) (1950) Haag-Streit-1959

Page 10: Slit lamp biomicroscope

Fig. : Modern slit-lamp

Cont…

• Onward modern slit-lamp system began to come into light.

Page 11: Slit lamp biomicroscope

TYPES

Page 12: Slit lamp biomicroscope

• On the basis of illumination system it is of 2 types :

1. Zeiss type

In the Zeiss type the illumination comes from below

Page 13: Slit lamp biomicroscope

2. Haag-streit type

In the Haag Streit type the illumination comes from above

Cont…

Page 14: Slit lamp biomicroscope

• On the basis of compound microscope system it is also 2 types :

1. The Grenough type

Flip lever to change magnification

Cont…

Page 15: Slit lamp biomicroscope

2. The Galilean changer type

Knob to change magnification (3 or 5 step)

Cont…

Page 16: Slit lamp biomicroscope

Also different types of HAND-HELD slit-lamp devices available for –

• Handicap

• Infant

• Bedbound patients

Cont…

Page 17: Slit lamp biomicroscope

OPTICAL PRINCIPLES

Page 18: Slit lamp biomicroscope

Illumination system Principle :

Kohler illumination principle is used in both Zeiss & Haag-streit type illumination system.

The light filament (F) is imaged on to the objective lens (O) but the mechanical slit (MS) is imaged on to the patient’s eye (S)

Page 19: Slit lamp biomicroscope

Microscope system Principle :

The eyepiece (Ocular) uses Astronomical telescope system.

Here two convex lenses used one in front of other separated by the distance of their focal length.

Fig. : Slit-lamp biomicroscope optical principles

Page 20: Slit lamp biomicroscope

A Galilean telescope system is used for even higher and changeable magnification.

Here a convex and a concave lens is used in line separated by the distance of their focal length.

Cont…

Fig. : Slit-lamp biomicroscope optical principles

Page 21: Slit lamp biomicroscope

Both telescope systems produce inverse image of an object so an inverting prism named Porro-Abbe prism is used to produce right way erect image.

Here two triangular prisms use the total internal reflection principle of optics.

Cont…

Porro-Abbe Prism

Fig. : Slit-lamp biomicroscope optical principles

Page 22: Slit lamp biomicroscope

The telescope system provide considerable distance between the microscope and the patient’s eye, so that certain maneuver like foreign body removal from the cornea or using extra lenses for fundus examination can be done.

Cont…

Page 23: Slit lamp biomicroscope

Parfocality : The point at which the microscope is focused corresponds to the point on which the light is focused, this coupling effect is called parfocality.

This is achieved by the microscope and the illumination system, having a common focal plane and their common axis of rotation also lies in that focal plane.

Cont…

Page 24: Slit lamp biomicroscope

PARTS OF THE SLIT-LAMP

Page 25: Slit lamp biomicroscope

Basic Design :

1. Mechanical system

2. Illumination system

3. Biomicroscope /Observation system

Associated instruments : • Applanation Tonometer

• Gonioscopic Lens

• Fundoscopy Lens

• Micrometer Eyepieces

• Image archiving device

• Laser delivery system

Page 26: Slit lamp biomicroscope

It concern with :

• Positioning & adjustment of patient and observer

• Maneuvering the illumination and microscope system together with joystick

• Providing base to other parts

Mechanical system : Basic 3 parts :

1. Motorized table (Base)

2. Patient positioning frame

3. Joystick

Page 27: Slit lamp biomicroscope

Parts of mechanical system :

Cont…

Fixation target

Chin rest adjustment

knob

Joystick

Power switch

Table height adjustment

Forehead band

Canthus alignment

Chin rest

Hand grip for patient

Lock for slit lamp base

Low friction plate

Page 28: Slit lamp biomicroscope

Provides a bright, evenly illuminated, finely focused, adjustable slit of light at the eye.

Contains the following components: • Light source • Condenser lens system • Slit and other diaphragms • Filters • Projection lens • Reflecting mirror or prisms

Illumination system :

Page 29: Slit lamp biomicroscope

Parts of illumination system :

Cont…

Scale for slit height

Slit height control

Inclined mirror

Latch to tilt

light column

Light source

Filter control

Centering screw

Slit width control

Page 30: Slit lamp biomicroscope

• Essentially these are compound microscopes composed of optical elements providing enlarged right way stereoscopic image to observer.

• Provides larger working distance in front of microscope for manipulation on patient’s eye.

• Magnification changer for detail view.

Observation system :

Page 31: Slit lamp biomicroscope

Parts of observation system :

Cont…

Eyepiece

Marked ring for

Adjustment of examiner’s

refractive error

Prism housing

Knob to change

magnification

Objective lens

Page 32: Slit lamp biomicroscope

Associated instruments :

Page 33: Slit lamp biomicroscope

CLINICAL PROCEDURE

Page 34: Slit lamp biomicroscope

Prerequisites :

• Switch on power & unlock base screw

• Cleaning the forehead band

• Changing paper strip from chinrest

• Comfortable sitting of pt. and the examiner

• Counseling the patient

• Proper positioning of the pt.

• Target fixation

• Adjust eyepieces to correct for examiner’s refractive error and interpupillary distance

• Children may need to stand, or they can sit on parent’s lap or kneel on a stable chair

Fig. Correct positioning at the slit lamp.

Page 35: Slit lamp biomicroscope

Chronology for slit-lamp examination :

Eyelids Eyelid

margins Tear film

Conjunctiva Cornea Aqueous humor

Iris Lens Vitreous

(ant.)

Page 36: Slit lamp biomicroscope

Illumination techniques :

Slit Width control knob:

Narrow to fully open slit illumination achieved by rotating this knob

Page 37: Slit lamp biomicroscope

Illumination techniques :

Slit height control knob :

Short to long slit illumination achieved by rotating this knob

Cont…

Page 38: Slit lamp biomicroscope

Illumination techniques :

Slit angle rotation :

Cont…

Page 39: Slit lamp biomicroscope

Illumination techniques :

Tilting light column : Illumination column can be tilted 5-20°, with vertical axis which gives extra plane and minimize reflection during posterior segment examination with condensing lens.

Cont…

Page 40: Slit lamp biomicroscope

Illumination methods :

Direct illumination • Direct diffused illumination

• Direct focal illumination

i. Parallelepiped

ii. Optical section

iii. Conical beam

• Specular reflection

• Tangential illumination

Indirect illumination • Indirect proximal illumination

• Retro-illumination

• Sclerotic scatter

Oscillatory illumination

Page 41: Slit lamp biomicroscope

Direct diffused illumination :

• illumination of the eye with a broad, unfocused light beam

• directed obliquely between 30-450

• magnification 6x to 10x

Application :

• General gross overview of eyelids, lashes, conjunctiva, sclera, pattern of redness, iris, pupil, gross pathology, ant capsule of lens

Cont…

Page 42: Slit lamp biomicroscope

Direct diffused illumination :

Cont…

Page 43: Slit lamp biomicroscope

Direct focal illumination :

• Illumination and observation are focused in the same plane

• slit width narrow to broad

• Illumination angle 45° to 60°

• Magnification 10x-40x

Application :

• Cornea in detail

• Anterior chamber

• Crystalline lens

• Anterior part of vitreous

• Grading cell and flare in anterior chamber

Cont…

Page 44: Slit lamp biomicroscope

Direct focal illumination :

Cont…

Page 45: Slit lamp biomicroscope

Direct focal illumination :

i. Parallelepiped

• Slit width 2-4 mm obliquely focusing on the cornea so that a quadrilateral block of light illuminate the cornea

Application :

• To examine corneal surface, stroma

• To ascertain depth (FB, abration)

Cont…

Page 46: Slit lamp biomicroscope

Direct focal illumination :

ii. Optical section

• Slit width 1mm or less

• Illumination angle 45-60° or more

• High illumination & magnification

Application:

• Corneal depth, layers, scars, vessels

• Lens opacity

Cont…

1. Tear film 2. Epithelium 3. Ant. Stroma 4. Post. Stroma 5. Descemet’s mem. & Endothelium

Page 47: Slit lamp biomicroscope

Direct focal illumination :

iii. Conical beam

• Assessment of particles floating in the A/C

• Narrow, short & bright slit of light

• 45°-60° light source directed to pupil

• Magnification 16x-25x

Application :

• Inflammatory cells, flare, pigmented cells, metabolic wastes

Principle : Tyndall phenomenon

Cont…

Page 48: Slit lamp biomicroscope

Specular reflection : • Angle of incidence = angle of

reflection

• Observation and illumination system have same angle with perpendicular axis to each other

• Slit width < 4mm

• The light reflected from the anterior or posterior corneal surface

• Best view with one eye

Application : • assessment of surfaces

Corneal epithelium

Corneal endothelium

Lens surface

• assessment of tear film

Cont…

Corneal endothelium

Page 49: Slit lamp biomicroscope

Tangential illumination :

• A narrow light beam is projected almost parallel along the structure to be observed

• Elevated structures are visible by shadowing

• Illumination angle 90°

• Magnification 10-25x

Application :

• elevated abnormities or changes in the iris

• tumors, cysts

Cont…

Page 50: Slit lamp biomicroscope

Indirect illumination : • The beam is focused in an area

adjacent to ocular tissue to be observed

• Decentered beam

• Illumination 2 to 4mm slit

• Magnification: Low to medium (depending upon object size)

Application: • infiltrates

• corneal scars

• deposits

• epithelial and stromal defects

Cont…

Page 51: Slit lamp biomicroscope

Indirect illumination :

Cont…

Page 52: Slit lamp biomicroscope

Retro-illumination :

• Light reflected on iris or fundus

• Microscope focused on cornea

• Two types: direct and indirect

Application :

For better visualization of

• Epithelial cysts

• Keratic precipitates

• Small blood vessels

• Small scar

Cont…

Page 53: Slit lamp biomicroscope

Direct retro-illumination :

• Observed feature is viewed in direct pathway of reflected light

• with this illumination findings are made visible with high contrast

• medium slit width 2 to 4mm

• Illumination angle 45-60°

Application :

• Infiltrations, small scars, corneal vessels, etc.

Cont…

Page 54: Slit lamp biomicroscope

Indirect retro-illumination :

• Observer at right angle to the observed structures

• Illumination angle greatly reduced or increased

• Feature on the cornea is viewed against a dark background

• medium slit width 2 to 4mm

Application :

• Infiltrations, small scars, corneal vessels, micro cysts, vacuoles

Cont…

Page 55: Slit lamp biomicroscope

Retro-illumination :

Cont…

Indirectly illuminated

Vacuole

Direct retro-illumination

Page 56: Slit lamp biomicroscope

• This technique is used to observe media clarities and opacities.

• The pupil is dilated

• the slit beam and microscope are made co-axial and light strikes the fundus and creates a glow behind the opacity in the media

• The media opacity creates a shadow in the glow

Applications

• abnormities in the anterior vitreous, lens, anterior chamber, cornea

Cont…

Retro-Illumination from the Fundus :

Page 57: Slit lamp biomicroscope

Sclerotic scatter : • Light incident on the limbus

with 2-4mm slit at an angle of 45° - 60°

• Decentered slit

• The microscope focused centrally

• Total internal reflection of the incoming light at inner corneal boundaries (endothelium and epithelium)

Applications

• scars, foreign bodies, corneal defects

• irregularities in the cornea

Cont…

Fig. : Corneal scar

Page 58: Slit lamp biomicroscope

Iris-transillumination :

• Transillumination of the iris by indirect light reflected from the fundus

• Mid dilated pupil (3 to 4mm)

• Illumination and observation at coaxial position

Applications

• Visualization of defects of the pigment layer of the iris

Cont…

Fig. : Transillumination in Albinism

Page 59: Slit lamp biomicroscope

Oscillatory illumination :

• Moves the beam (usually an optical section) from nasal to temporal limbus

• Viewed alternately by direct and indirect illumination

• Often possible to see minute objects or filaments, specially in the aqueous

• Reveals subtle areas of corneal thinning.

Cont…

Page 60: Slit lamp biomicroscope

Van Herrick’s Technique : • Used to evaluate anterior chamber angle without

gonioscopy • Narrow slit beam close to limbus with Illumination angle

60° • Medium magnification

Principle : • Compare the width of cornea seen by optical section with the

dark section seen between ant. surface of iris & back of cornea

Interpretation : Grade 4 – open anterior chamber angle 1:1 ratio Grade 3 – open anterior chamber angle 1:2 ratio Grade 2– narrow anterior chamber angle1:4 ratio Grade 1– risky narrow anterior chamber angle less than 1:4 ratio Grade 0– closed anterior chamber

Page 61: Slit lamp biomicroscope

Van Herrick’s Technique : to assess anterior chamber angle

Cont…

Page 62: Slit lamp biomicroscope

Filters : • White filter

• Overview of ocular surface tissues • Examining intraocular structures

• Cobalt blue filter • Coneal abrasion, ulcer • Applanation tonometry

• Red free filter • Rose-bengal staining

Page 63: Slit lamp biomicroscope

Applanation Tonometer :

The Goldmann Applanation Tonometer is the most common tonometer that usually mounted on the standard slit-lamp biomicroscope.

It’s easy to use and measure the IOP of a seated patient with high accuracy in most clinical situations.

Parts :

1. Tonometer tip (biprism)

2. Metal rod

3. Tonometer housing

4. Force adjustment knob

Associated instruments :

Page 64: Slit lamp biomicroscope

Principle :

• Applanated force (AF) equals the intraocular pressure (IOP)

• diameter of the applanated area: 3.06 mm here counteracting forces of corneal rigidity and capillary attraction cancel each other

Application :

• Measurement of intraocular pressure (IOP)

Cont…

Page 65: Slit lamp biomicroscope

Procedure : 1. Proper positioning of the patient

2. Instill topical anesthetic and flurescein in pt.‘s eye

3. Place the cobalt blue filter with broad slit illumination at 60-90° angle

4. The dial is preset between 1 and 2 (i.e. 10–20 mmHg)

5. Prism is advanced until it just touches the apex of the cornea

6. Viewing into the oculars a pattern of two semicircle will appear which represent the fluorescein-stained tear film touching the prism

Cont…

Page 66: Slit lamp biomicroscope

Procedure : 7. Then gently rotate the force

adjustment knob of tonometer to align the inner margins of the semicircles

8. The reading on the dial, multiplied by 10, gives the IOP. Modern digital display system provides actual reading without multiplication

9. Repeat the procedure for other eye

10. At last wash B/E with antibiotic eye drop

Cont…

Video clip : GAT

Page 67: Slit lamp biomicroscope

Fundus Observation and Gonioscopy :

• Different types of Contact and Non-contact lenses used for examination, diagnostic and therapeutic purpose in fundoscopy and gonioscopy

• With classic three mirror Goldmann lens post. Segment and irido corneal angle can examined simultaneously

Page 68: Slit lamp biomicroscope

Cont…

Fundoscopy :

Contact Glasses :

• Additional tool for fundus observation with the slit lamp

• Mostly direct; erect and non mirrored image of the fundus

• Required :

• dilated pupil

• use of coupling liquid (Hypromellose 0.3%)

• Coaxial angle (0°)

• Illumination – low with 2-4 mm slit

• Magnification – 10-16 x

Page 69: Slit lamp biomicroscope

Cont…

Non-contact lenses :

• Additional tool for fundus observation with the slit lamp

• Mostly indirect; upside-down and mirror image of the fundus (convex optics)

• non contact

Required: • dilated pupil

• Coaxial angle (0°)

• Illumination - low with 2-4mm slit

• Magnification – 10-16 x

Fundoscopy :

Page 70: Slit lamp biomicroscope

Cont…

Fundus view through 90D non-contact lens :

Page 71: Slit lamp biomicroscope

Cont…

Goldmann 3-Mirror gonioscopic lens :

• Procedure same as contact fundus lens

Observation :

• Central lens (0) : Posterior pole

• 73° mirror (1) : Equator

• 67° mirror (2) : Ora serrata

• 59° mirror (3) : Iridocorneal angle

Gonioscopy :

Page 72: Slit lamp biomicroscope

Cont…

Application :

• Grading of A/C angle width (Shaffer system)

• Peripheral anterior synechiae

• Neovascularization

• Hyperpigmentation

• Trauma

Angle recession

Trabecular dialysis

Cyclodialysis

Foreign bodies

• Blood in the Schlemm canal

Gonioscopy :

Fig.: Shaffer system grading

Fig.: Blood in chamber angle

Fig.: Neovascularization

Page 73: Slit lamp biomicroscope

• Slit-lamps have the capability of providing laser delivery with a conjunct laser system installed on it

• Lasers are used for the treatment procedures like photocoagulation, iridotomy or capsulotomy

• Commonly delivered lasers – Nd:YAG

Laser delivery system with slit-lamp :

Fig.: Laser system mounted Slit-lamp

Page 74: Slit lamp biomicroscope

The slit lamp is one of the most commonly used and valuable diagnostic instruments

for an ophthalmologist of today.

So user level care & maintenance is very much important to get optimum performance & long life from it.

Page 75: Slit lamp biomicroscope

Video clip : Slit-lamp demonstration

Page 76: Slit lamp biomicroscope
Page 77: Slit lamp biomicroscope

• Clinical ophthalmology A systemic approach - Jack J Kanski - 7th ed.

• OPHTHALMIC OPTICS – Professor M.A. Matin 2nd ed.

• Clinical Optics – Andrw R. Elkington 3rd ed.

• BASIC OPHTHALMOLOGY – Professor Md. Shamsul Haque – 2nd ed.

• AMERICAN ACADEMY OF OPHTHALMOLOGY (2012-2013) – Sec.-3

• AMERICAN ACADEMY OF OPHTHALMOLOGY (2012-2013) – Sec.-8

• Essentials of Ophthalmology – Samar K Basak – 5th ed.

• Internet – • https://www.google.com - Search Links

• http://en.wikipedia.org

• https://www.youtube.com