Sizing software for cerebral devices - Neurovascular Exchange
Transcript of Sizing software for cerebral devices - Neurovascular Exchange
Sizing software for cerebral devicesWLNC Cases
Mathieu SANCHEZ (PhD), CEO
www.sim-and-cure.com
16/06/2017 Sim&Cure 1
Fusifrom Aneurysm:Pipeline
16/06/2017 Sim&Cure 2
16/06/2017 Sim&Cure 3
Very Large Fusiform aneurysm = 4 Telescopic Pipelines (5/35mm)
Major problem = Anchor the Pipeline in Distal and Proximal
Distal Telescopic Pipeline 1
Option 1
Color legend:(vessel cut)
Distal Telescopic Pipeline 1
Option 2
“Individual” view
16/06/2017 Sim&Cure 4
Telescopic Pipeline 2 Telescopic Pipeline 3 Telescopic Pipeline 4:
Proximal anchor
“Individual” view
16/06/2017 Sim&Cure 5
Combined view
Basilar Tip Aneurysm:WEB vs PulseRider
16/06/2017 Sim&Cure 6
16/06/2017 Sim&Cure 7
WEB SLS 8
Green = Wall Apposition
Red = No Wall Apposition
16/06/2017 Sim&Cure 8
WEB SL 8X5
Green = Wall Apposition
Red = No Wall Apposition
16/06/2017 Sim&Cure 9
PR 201
16/06/2017 Sim&Cure 10
PR 211
16/06/2017 Sim&Cure 11
PR 301
16/06/2017 Sim&Cure 12
PR 303
16/06/2017 Sim&Cure 13
Giant ICA Aneurysm:Pipeline
16/06/2017 Sim&Cure 14
Pipeline 3/30 mm
Warning : Due to the
injected drugs the day of
the treatment, we could
observe a dilation of the
artery diameter the day of
the surgery. To validate an
accurate Pipeline sizing is
better to perform a
simulation with the
« fresh » Dicom
(Acquisition on the day of
the intervention)
1 Pipeline 3/30 mm : To obtain a good anchor in proximal and
distal and to be able to start and finish before the curvatureColor legend:(vessel cut)
For information:
Distal vessel diameter: 2.5-2.7 mm
Proximal vessel diameter: 2.8-3 mm
Sim&Size do not provide vessel measurement
Sizing software for cerebral devicesWLNC Cases
Mathieu SANCHEZ (PhD), CEO
www.sim-and-cure.com
16/06/2017 Sim&Cure 15