Sizing of powders in inhalers with an AerosizerJ according to.pdf

download Sizing of powders in inhalers with an AerosizerJ according to.pdf

of 19

Transcript of Sizing of powders in inhalers with an AerosizerJ according to.pdf

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    1/19

    Aerosol Science 35 (2004) 509527

    www.elsevier.com/locate/jaerosci

    Sizing of powders in inhalers with an AerosizerJ according toa mixed experimental factorial design

    P.J. Mendesa;b, A. Raposoa;b, J.M.M. Sousab, J.F. Pintoa ;

    aDept. de Tecnologia Farmaceutica, Faculdade de Farm acia da Universidade de Lisboa,

    Av. Prof. Gama Pinto 1649-003 Lisboa, PortugalbDept. de Engenharia Mecanica, Instituto Superior Tecnico, Universidade Tecnica de Lisboa,

    Av. Rovisco Pais 1, 1049-001 Lisboa, Portugal

    Received 18 March 2003; received in revised form 18 September 2003; accepted 3 October 2003

    Abstract

    In the present work, the AerosizerJ with the AerodisperserJ is used to assess the size and size distribution

    of two common dry powder inhalers: Ventilan RotacapsTM (salbutamol sulfate) and Bricanyl TurbohalerTM

    (terbutaline sulfate), according to a mixed factorial design. The dierent parameters of the equipment (voltage

    applied to the detector, shear force, deagglomeration, feed rate and pin vibration) were the factors considered

    in the design, set to dierent levels. Altogether, a full 41 31 23 factorial was considered and the results

    of the experiments analyzed according to the Montgomery technique. Results have shown that samples with

    a single population of particles were less aected by the changes in the parameters, whereas for mixtures

    incorporating dierent populations results were aected upon change of the levels of the factors. The voltage

    at the detector was the most important factor, followed by the shear force. By contrast, pin vibration and feed

    rate were less signicant. The interactions between variables were also responsible for signicant changes in

    the results and should not be neglected throughout the measurements.

    ? 2003 Elsevier Ltd. All rights reserved.

    Keywords: Aerosizer; Factorial design; Montgomery technique; Dry-powder inhaler; Salbutamol sulfate; Terbutaline sulfate

    1. Introduction

    In recent years, a great development on aerosols and related technology has been observed and

    the dispersion of ne powders or liquids found application in the administration of drugs to humans.

    Corresponding author. Tel.: +351-217-946-434; fax: +351-217-937-703.

    E-mail address: [email protected] (J.F. Pinto).

    0021-8502/$ - see front matter? 2003 Elsevier Ltd. All rights reserved.doi:10.1016/j.jaerosci.2003.10.005

    mailto:[email protected]:[email protected]
  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    2/19

    510 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Aerosols have been used in the treatment of diseases, namely those aecting the respiratory tract.

    The main advantages of the respiratory route, as compared to the oral one, are the lower dose of

    drug required to achieve the same therapeutic eect, in turn leading to decreased side eects, and

    an earlier onset of the drugs eect (Wierlik, Diepenmaat, & Damhuis, 2002).Dry powder inhalers (DPIs) play a major role in aerosol therapy because they are chlorouoro-

    carbon (CFC) free (Wierlik et al., 2002; Dunbar, 2002), less dependent on patients coordination on

    inhalation, easy to formulate and less aected by environmental changes (e.g. temperature) (Ramon

    et al., 2000). Ramon et al., 2000 provided that their components have adequate properties (Hickey,

    2002). However, the size of the particles and the way they interact are critical for a proper deposition

    in the alveoli (Gonda, 1992), and sizing of the particles is, therefore, of paramount importance for

    a good therapeutic eect.

    There are many in vitro methods for sizing DPI particles. Examples of compendial methods are

    the cascade impactor (e.g. Andersen multi-stage impactor), and the multistage liquid impinger (e.g.

    twin impinger) (US Pharmacopoeia, 26). Unfortunately, these methods are time consuming (up to1 h per measurement) by contrast to other analyzers which give measurements in minutes (Peters,

    Vanderpool, & Wiener, 2001). Examples of such equipment are the time-of-ight (TOF) based

    aerodynamic particle size analyzers, such as the AerosizerJ, the Aerodynamic Particle SizerJ spec-

    trophotometer or the E-SpartJ (Mitchell & Nagel, 1999). These analyzers are often used in the

    assessment of medicinal aerosols, either dry powder or pressurized metered dose inhalers (pMDIs).

    The particle size distribution spectrum obtained reects the dierences in the aerodynamic diameters

    of particles present in the aerosol. For instance, a DPI containing lactose (larger particles), as the

    drug (salbutamol sulfate, smaller particles) carrier, exhibits a spectrum with two peaks that could

    be cautiously related to the individual symmetrical spectra produced for both lactose and salbutamol

    sulfate particles (Srichana, Brain, Marriott, & Martin, 2000). However, care must be observed when

    measurements are carried out, since when suspensions of pMDIs are considered, the measured valuesdier from those obtained by other techniques (Mitchell, Nagel, & Archer, 1999). This reects the

    vulnerability of the method when disperse systems are considered, with some preferential removal

    of larger mass median aerodynamic diameter (MMAD) particles, due to the evaporation of solvent

    and/or inertial gravitational deposition (Nagel, Wiersema, Bates, & Mitchell, 2002), thus leading

    to dierent results. Moreover, when DPIs are considered, the ability of the equipment to disperse

    particles is highly dependent on the properties of the materials (Laitinen & Juppo, 2003). In a less

    conventional application of this technique, the use of a third laser beam has been suggested in order

    to avoid the particle density dependency of the method (Oskouie, Noll, & Wang, 2003). The use of

    adequate mathematical models, that allow the calculation of the particles lumped density and shape

    factors (Oskouie, Noll, & Wang, 2002) have also been suggested.The technique described above has several advantages as compared to others: (a) the large number

    of particles counted per second, (b) the easy operating procedure and maintenance and (c) the small

    sample size required. However, some limitations have been pointed out (Mitchell & Nagel, 1996)

    such as: (a) the particles are detected on a one at a time basis, therefore some particles may not

    be detected particularly at high feed rates; (b) the equipment assumes that all the particles are

    spherical and (c) the equipment is unable to distinguish between particles belonging to dierent

    populations (e.g. drug and excipients), so all the particles are detected and treated identically and

    (d) the density of the sample must be known. The rst limitation aects seriously the counting

    eciency and can create false peaks. Since the equipment determines the aerodynamic diameter and

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    3/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 511

    then calculates the geometric diameter using the density of the material, an incorrect value for the

    density of the material produces wrong values for the geometric diameters, without aecting the

    aerodynamic diameter because the AerosizerJ measures the TOF of individual particles which is

    dependent on their mass. Moreover, non-spherical particles present dierent velocities of ight fromspherical particles due to dierences on the drag coecients. For the above-mentioned limitations,

    special care should be taken in the analysis of the results since a DPI can have more than two

    dierent populations of particles with dierent drag forces, densities and shapes, leading to inaccurate

    measurements, a drawback shared with other techniques. It follows that the AerosizerJ provides more

    accurate results for single populations of particles rather than for mixtures of dierent populations of

    particles, i.e., populations made of dierent chemical identities, and therefore showing multimodal

    distributions. The equipment may also underestimate the aerodynamic sizes of airborne particles from

    both powder and pressurized inhalers, often requiring a comparison with other measuring techniques

    (Nagel et al., 2002).

    The aim of the present work was the assessment of the performance of the AerodisperserJ

    AerosizerJ on the characterization of two DPIs when all the operating parameters were considered

    in a factorial experimental design. This design, selected on the ground of its robustness and because

    the variables (equipment settings) can be considered simultaneously, allows for the evaluation of the

    possible interactions between variables after establishing the relative importance of each parameter,

    by the Montgomery method (Montgomery, 1996).

    2. Material and methods

    In this work, VentilanTM RotacapsTM (GlaxoWellcome) and BricanylJ TurbohalerJ (AstraZeneca)

    were used as model DPIs, whose particle size aect their therapeutic ecacy. The rst is presentedto patients as a hard gelatin capsule containing 200 g of salbutamol sulfate (in 25 mg of powder),

    whereas the second contains 500 g of terbutaline sulfate per blister. For calibration purposes, two

    sets of spheres were considered: one of nylon (mean diameter standard deviation of 5 1:5 m,

    Dantec Measurement Technology GmbH, Germany), and another of glass (mean diameterstandard

    deviation of 10:0 1:0 m, Duke Corporation, Palo Alto, CA, USA).

    In these experiments, the two types of DPIs were analyzed by using an AerodisperserJAerosizerJ

    (TSI, GmbH, Aachen, Germany). Briey, this equipment operates by suspending the particles in the

    air, using a disperser, a uidizer and a deagglomerating device (AerodisperserJ), before this air

    ow is expanded into a vacuum. This vacuum draws the particles through a ow nozzle, produc-

    ing a precisely controlled acceleration to a high-speed (subsonic) aerosol jet (Allen, 1990) in theAerosizerJ. Fig. 1 presents the assembly considered in the present study. The velocity of the air

    remains constant in the measurement region but the particles accelerate at dierent rates, inversely

    related to their aerodynamic size (Barber, 1993). Two laser beams (at a known distance between

    them) placed in the nozzles exit allow the measurement of the time required for the particles to

    travel between them and, thus, the velocity. The scattering of the beams by the particles induces the

    emission of an electrical signal by the detector, made of photomultiplier diodes, starting/stopping a

    clock. The particles aerodynamic diameter is calculated from the time elapsed between the two sig-

    nals (Laitinen & Juppo, 2003). Traveling times are in the order of hundreds of nanoseconds for small

    particles and thousands of nanoseconds for larger particles. The measurement of the aerodynamic

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    4/19

    512 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Fig. 1. Schematic representation of the AerodisperserJ connected to the AerosizerJ.

    diameter of particles is automatically compensated for changes in shape, surface irregularities, mass

    and refractive index. Visual inspection of the samples was made by scanning electron microscopy(Phillips Analytical, UK).

    Prior to the analysis of the DPIs, according to a factorial experimental design, the equipment was

    calibrated with both types of spheres and validated for reproducibility and eectiveness of counting.

    These characteristics were assessed by the run time and presentation of the data according to size

    distribution by number and by volume, as a function of the aerodynamic diameter. To check the

    counting eciency of the AerosizerJ, lactose monohydrate (Granulac 200, Meggle, Germany) and

    microcrystalline cellulose (Avicel PH-101, FMC Corp., USA) were used as received, in the same

    quantity (6 mg) as that collected from the DPIs. The distribution of the populations was characterized

    by the mode, or, if otherwise referred in the text, by another measure of central tendency.

    2.1. Experimental design

    Experiments were carried out according to a mixed factorial design and the results were analyzed

    by the method described by Montgomery (1996). Briey, this statistical technique allows the trans-

    formation of an n-level factor (with n 2) into a k2 factorial design (kfactors at two levels each),

    and the assessment of the eect of each factor and their interactions on the results. In the present

    investigation, a 413123 mixed factorial design was considered, where the exponents represent the

    equipment variables and the bases stand for the levels of each variable in the experiments (Table 1).

    According to this method, the variables with three or four levels can be converted into two variables

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    5/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 513

    Table 1

    Equipment variables and levels considered in the experiments

    Variables Levels

    Photomultiplier voltage (PMV) (V) 850 1100

    Shear force (SF) (Psi) 0.5 1.5 3.0 4.0

    Deagglomeration (D) 0 1

    Feed rate 103 (FR) (counts/s) 15 10

    Pin vibration (PV) 0 1

    with two levels each (22) (Myers & Montgomery, 1995). Therefore, for a variable with two levels,

    this technique indicates the eect of that variable on the results when it changes from the lowest to

    the highest level. For a variable with three levels, it indicates the average eect of that variable foreach level (levels 1 to 2 or levels 2 to 3). Thus, when the variable changes from the lowest to the

    highest level, the magnitude of the eect doubles the value found for the intermediate variations.

    With factors at four levels, Montgomerys technique produces three results: the rst corresponds to

    the average eect when the factor changes from levels 1 to 2 or 3 to 4, the second corresponds to

    the average eect of the change from levels 1 to 3 or 2 to 4 and the third reects the tendency of

    the change. Consequently, the total eect is the sum of the rst and second results.

    The dierent magnitudes of the eects observed were compared by an analysis of variance

    (ANOVA) of the factorial design with more than two levels (Armstrong & James, 1996), and

    the levels of signicance assessed by an F-test. The smallest values of the mean squares were aver-

    aged and considered as the experimental error in experiments. Appendix A presents a more detailed

    explanation of the method.

    3. Results and discussion

    3.1. Preliminary experiments

    At rst, the good performance of the measuring equipment was ascertained with the calibration

    spheres. The mean geometric diameters by number measured with the AerodisperserJAerosizerJ

    conguration were 4.97 and 10:45 m for the spheres of nylon and glass, respectively. The dierences

    to the nominal values of 5.00 and 10:00 m were, therefore, 0:03 m (i.e., 0.6%) and 0:45 m (i.e.,4.5%), respectively. These dierences were deemed acceptable and indicative of the ability of the

    AerosizerJ to measure the diameter of the particles accurately. Moreover, as shown in Fig. 2,

    the narrow distribution of the geometric diameters by number is identical to that reported by the

    manufacturers for the calibration spheres.

    In a separate set of experiments, it was observed that the run time of the equipment was pro-

    portional to the number of particles counted and dependent on the feed rate: the higher the feed

    rate, the shorter the run time (Fig. 3). A change from the largest to the lower feed rate decreased

    the number of counts per second from 1950 (higher) to 1460 (medium) and 590 (lower). The

    eect of the number of particles counted on the diameter measured has been studied by using the

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    6/19

    514 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Fig. 2. Distribution by number of counts for the calibration spheres: () spheres of nylon, (- - - - - -) spheres of glass.

    Fig. 3. Relationship between the run time and the number of particles counted as a function of feed rate: ( ) low, ( )

    med, () high.

    Fig. 4. Relationship between the aerodynamic diameters (mode) of particles and the number of particles counted: ()

    distribution by number, ( ) distribution by volume.

    pharmaceutical VentilanTM RotacapsTM (salbutamol sulfate). Once the parameters of the equipment

    have been set, it was realized that for a number of counted particles larger than 120,000, the geo-

    metric aerodynamic diameter (modes shown) was more uniformly distributed (Fig. 4). This gure

    takes 54 runs into consideration and shows the results for the aerodynamic diameters obtained in

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    7/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 515

    Fig. 5. Size distributions for microcrystalline cellulose samples detected at 850 V (a) and 1100 V (b): () 1 mg,

    ( ) 4 mg, () 8 mg, () 16 mg, () 24 mg.

    distributions by number and by volume, and the number of counted particles in each run. As an-

    ticipated, as the sample size increases (reected by the number of counts), the precision of the

    measurement also increases (reected by a decrease of the standard deviation). Above 120,000 par-

    ticles, the dispersion of the mode values was small enough for the measurement to be considered

    acceptable. For samples of other materials, a dierent number of counts was required (results not

    shown) for the readings to become reproducible. For instance, when microcrystalline cellulose was

    used, a minimum sample quantity of 8 and 4 mg were required when results were detected at 850

    and 1100 V, respectively (Fig. 5).

    As previously discussed, when DPIs are considered, one should be aware that more than one

    population of particles may be present in the sample. There is, usually, one population of drugparticles (small in size) and another of the carrier (larger particles). The proportion between the two

    populations has an impact on the therapeutic eect of the drug and, consequently, it is important

    to characterize them adequately. In this part of the study, carried out using VentilanTM RotacapsTM

    (containing particles of both salbutamol sulfate and lactose), the evaluation of both populations was

    possible by considering the particle size distribution both by number and by volume. It was possible

    to estimate the approximate geometric diameters (modes) of both components in the formulation,

    due to the large dierence in size between the two populations of particles. In short, the averages of

    the modes found in 54 runs were calculated and the values of 1:48 m (by number) and 68:9 m

    (by volume) were estimated for the drug and carrier particles, respectively. The validity of the

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    8/19

    516 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Fig. 6. Relationship between the diameter (mode) expressed by volume and by number of counts.

    procedure is supported on the ground that dierent measurements have shown that the diameters by

    number and by volume are not related (Fig. 6). In fact, in this gure, a scattered distribution of

    the readings is observed, with no relationship between the geometric diameters by number and by

    volume, as expected.

    Thornburg, Cooper, and Leith (1998) observed that the counting eciency (ratio of the number

    of particles detected to the number of particles passing through the detector) of the AerosizerJ is

    variable. The same authors (Thornburg et al., 1998) stated that counting eciencies below 100%

    or, variations dependent upon particle size, would give inaccurate size distributions. Furthermore,

    factors, such as the concentration of the sample, the model of the equipment, the voltage at thedetector, the measurement of particles with diameters below 7 m or acceptance of counting rates

    above 20,000 particles per second, were also described as aecting the counting eciency of the

    equipment. These observations were conrmed in the present work, when particles of lactose were

    considered. In fact, for samples with the same mass and equivalent number of counts, the size

    distribution of the particles varied at small number of counts.

    After performing a few tests with VentilanTM RotacapsTM, it was noticed that the size distribution

    of the particles was not constant between runs performed under the same conditions, even for sam-

    ples taken from the same capsule, although the modes of the distributions (peaks) remained constant.

    This fact shows that the counting eciency changes with the size of the particles in dierent runs

    and samples. Furthermore, we noticed that the values read in the equipment were discontinuous,i.e., identical values were registered several times, whereas in between values were never found.

    This happens because the AerosizerJ collects the TOF of the particles in 2048 linearly spaced time

    channels, which are converted by the software into 500 logarithmically spaced sized bins (Amherst

    Process Instruments). Consequently, the equipment presents the rounded value in the bin instead

    of showing the true measurement. This eect of rounding the results is dicult to assess be-

    cause the software of the equipment shows the results already in bins and, therefore, that some

    lack of consistency in the measurements is expected. Results are presented in a more discrete way

    than the measurements, and consequently, some lack of accuracy, reproducibility or sensitivity can

    be anticipated.

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    9/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 517

    Fig. 7. Eect of the voltage in the detector over the size of the particles. Insets show scanning electron photomicrographs

    of BricanylJ TurbohalerJ and VentilanTM RotocapsTM, () BricanylJ TurbohalerJ (1100 V), (- - -- - -) BricanylJ

    TurbohalerJ (850 V), ( ) VentilanTM RotacapsTM (1100 V), (- - - - - -) VentilanTM RotacapsTM (850 V).

    Results present in both Figs. 4 and 6 stress these observations since the results are organizedaccording to horizontal and vertical lines, rather than a scattered distribution. This problem may aect

    the analysis of the samples, especially of those populations with large particles. When the results

    are represented as percentages this dierence is less signicant, but dierences in the aerodynamic

    and geometric diameters can still be observed, even though the samples have been prepared under

    the same conditions.

    3.2. Factorial design

    The performance of the AerosizerJ changed with the voltage applied to the detector, as observed

    when the voltage was changed between measurements, both for the calibration spheres and theaerosols. A narrower distribution curve was obtained when the voltage was increased (Fig. 7), and

    in all cases, the number of counted particles was higher for the highest voltage, as observed already

    byThornburg et al. (1998).In close agreement with the work ofMitchell and Nagel (1996),dierent

    sensitivities of the detector did not aect the mean geometric diameter for the calibration spheres

    (nylon) with a very narrow size distribution as opposed to the results observed for the inhalers (large

    sizes distributions).

    In the case of BricanylJ TurbohalerJ, the distributions had identical shapes for the two recom-

    mended values of the photo-multiplier voltage (850 and 1100 V), although dierent geometric mean

    diameters and smaller base lines of the peaks were observed for the highest voltage (Fig. 7, open

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    10/19

    518 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    circles). Like the calibration spheres, this DPI presents only one population of particles (Fig. 7 and

    micrograph associated with curves), but a wider particle size distribution. In both cases, only one

    peak was observed. When the sensitivity of the equipment was increased, smaller particles were

    detected more easily, and, thus, a shift of the distribution of the curve to smaller sizes was observed(mean diameters diered by 34%, whereas the mode remained constant). However, when the shear

    force was changed from peak to low, the aerodynamic diameters (modes) increased from 0.76

    to 0:79 m (0.3% increase). As this product does not contain any excipient, it can be assumed that

    the shear force only promotes a minor separation of the particles of terbutaline, the only component

    in the formulation (Fig. 7, open circles and micrograph).

    When VentilanTM RotacapsTM was considered, the aerodynamic diameters expressed by the number

    of counts originated three peaks within the distribution. This type of distribution is due to the presence

    of several groups of particles present in the product (Fig. 7, open squares and micrograph). For this

    product, a higher voltage led to a shift of the peaks towards smaller sizes (Fig. 7). On the other

    hand, when the results (aerodynamic diameter) were expressed by volume, only a slight shift of theresults towards smaller sizes was observed, suggesting a higher sensitivity of the detector to smaller

    particles, as mentioned previously. For VentilanTM RotacapsTM, the dierences between the means

    (aerodynamic diameter) of the particles observed for high and low voltages on the detector were

    0:52 m (33%) and 23:02 m (33%) for number and volume distributions, respectively (Table 2).

    When all the results of the factorial design were considered, it was possible to express them in

    terms of modes of the aerodynamic diameters, either by number or by volume. Results of the eects

    are shown both in Tables2and3. Table2 presents the variations on the distribution by number and

    by volume of the aerodynamic diameters observed for the dierent samples of the pharmaceuticals

    considered. The upper part refers to the main eects or individual eects when a single variable

    was changed at a time. Therefore, it is possible to assess the impact of each operational parameter

    (variable) over the size distribution of samples. Table3 presents the results of the ANOVA performedon the experimental data emphasizing their degree of signicance as reected by the F-values. An

    increase on the voltage of the photomultiplier promoted a decrease on the aerodynamic diameter

    (mode) of 0.52 and 23:02 m, respectively by number and by volume (Table 2). This variable is

    the most signicant one with a value ofp 0:001 (Table 3). The sensitivity of the detector aects

    the number of counts, and, thus, the size of the sample. In order to obtain consistent and reproducible

    results, the magnitude of this parameter must be carefully chosen and, thus, special care should be

    taken on the selection of the photo-multiplier voltage.

    When the shear force applied (Fig. 8a and b) was increased from low to peak, an increase on

    the measured diameter (by number) was observed. In fact, when two levels were considered (low

    to high and medium to peak), the eect was signicant (SF2, Table 2 and p 0:01, Table 3).This increasing tendency is aected only in a few cases by an increase on the shear force from

    high (3 Psi) to peak (4 Psi), resulting in measured diameters that were equal or smaller (Fig. 9a

    and b). Results in Table 2 do not show this linear tendency, but it should be borne in mind that

    the dierence between levels low and medium and high and peak is only 1 Psi, whereas the

    dierence between medium and high is 1:5 Psi. On the other hand, an increase on the shear force

    (Fig. 8a and b) often leads to a decrease on the measured diameter for larger particles (upper parts

    of the graphs in Figs. 8 and 9). Although the observation is not signicant (Table 3, distribution by

    volume) and the results do not show a particular tendency (Fig. 9a and b), it is possible to suggest

    that by increasing the shear force, the separation between the particles was promoted, as mentioned

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    11/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 519

    Table 2

    Results of the main eects and interactions from the calculations, carried out according to the Montgomery technique

    Variables and interactions Distribution by number (m) Distribution by volume (m)

    Main eects

    Photomultiplier voltage (PMV) 0.52 23.0

    Shear force

    Total (SFtotal) 0.17 1.15

    1 level (SF1) 0.05 0.49

    2 levels (SF2) 0.12 0.66

    Shape of tendency line 0.01 0.52

    Deagglomeration (D) 0.00 1.65

    Feed rateTotal (FRtotal) 0.02 1.92

    1 level (FR1) 0.01 0.96

    Shape of tendency line 0.00 3.51

    Pin vibration (PV) 0.02 0.23

    Binary interactions

    PMVSFtotal 0.12 0.26

    PMVD 0.03 0.19

    PMVFRtotal 0.01 0.06

    PMVPV 0.01 0.05

    SFtotalD 0.03 0.58

    SFtotalFRtotal 0.03 0.90

    SFtotalPV 0.02 0.20

    DFRtotal 0.02 1.06

    DPV 0.01 0.66

    FRtotalPV 0.01 1.06

    elsewhere (Houzego, 2002). Because larger particles present higher contact areas, the adhesion and

    cohesion between them are larger too. Consequently, the eect of the shear force is observed mainly

    on the smaller particles of the drug with a shift of the peak observed in the distribution by number to

    larger sizes. This variable is particularly important for samples containing more than one populationbut, even for a single population, the shear force changes the values measured, as a consequence of

    the additional deagglomeration produced.

    Regarding the deagglomeration, when this parameter is set to high, the particles are transported

    from the sample cup to the dispersion pin at a higher velocity. Hence, a stronger impact on the pin

    occurs generating a larger force to separate the particles (Houzego, 2002). It can be assumed that

    the same sort of explanation given for the eect of the shear force applies to the deagglomeration

    variable. For smaller particles, a change from normal to high in deagglomeration failed to produce

    a major dierence between the results (Table 2), whereas for larger particles a reduction on the

    measured sizes was observed (Fig. 8a and b), although not statistically signicant (Table 3).

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    12/19

    520 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Table 3

    Results of the ANOVA for the dierent experiments, according to the Montgomery technique

    Degrees of Distribution by number Distribution by volume

    freedom

    Sum of Mean F Sum of Mean F

    squares squares squares squares

    Variables

    Photomultiplier voltage (PMV) 1 6.82 6.82 9443.0 11806.4 11806.4 1793.5

    Shear force (SF) 3 0.53 0.17 245.0 41.21 13.73 2.09

    Deagglomeration (D) 1 0.00 0.00 0.00 26.15 26.14 3.97

    Feed rate (FR) 2 0.01 0.00 4.85 323.0 161.5 24.54

    Pin vibration (PV) 1 0.01 0.01 11.08 24.63 24.63 3.74

    Binary interactions

    PMVSF 3 0.13 0.04 59.08

    52.15 17.38 2.64PMVD 1 0.03 0.03 42.92 13.39 13.39 2.04

    PMVFR 2 0.02 0.01 15.23 136.1 68.08 10.34

    PMVPV 1 0.00 0.00 5.54 3.95 3.94

    SFD 3 0.07 0.02 31.85 119.8 39.92 6.07

    SFFR 6 0.05 0.01 12.00 242.8 40.47 6.15

    SFPV 3 0.09 0.03 39.23 9.87 3.28

    DFR 2 0.01 0.00 3.46 88.41 44.20 6.72

    DPV 1 0.00 0.00 18.62 18.61 2.83

    FRPV 2 0.03 0.01 20.08 80.80 40.39 6.14

    Blank spaces represent results associated with the experimental error.p 0:05,p 0:01,p 0:001.

    The feed rate directly inuences the number of particles counted by unit of time. The

    AerodisperserJ adjusts automatically the air jet to satisfy the value selected for the aforementioned

    parameter. Table 2and Fig. 8(a and b) show a general decrease on the size of both small and large

    particles when the feed rate was increased. However, the dierences in the particles characterized

    by smaller diameters are not very signicant, although this variable promoted a signicant decrease

    (p 0:01, Table 3) on the aerodynamic diameter by volume, i.e., over particles with larger sizes.

    Eventually, a synergistic eect occurred at high feed rates due to the presence of more particles,

    promoting cluster formation and the detection of apparently larger particles. Fig. 10(ad) shows thediameter of the particles as a function of feed rate for the dierent values of the voltage applied to

    the detector when the pin vibration was on and o. Pin vibration aects the deposition of particles

    on their way to the detector. Data clearly show that a more consistent pattern was observed when a

    distribution by number of counts was considered, as compared to the distribution by volume (see,

    respectively, lower and upper parts of Fig. 10). Pin vibration did not produce a major eect on the

    size distribution (Fig. 8a and b, Table 2) and distribution by number was barely aected (Table 3)

    by a change on the pin vibration status. Even though, if a distribution by volume is considered, two

    remarks have to be made: (a) some of the observed dierences were not due to the AerosizerJ but

    to variations in the content of the capsules and (b) the most important component in the capsules

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    13/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 521

    Fig. 8. Aerodynamic diameter of the particles as a function of parameters level: low (a) and high (b) voltage in the

    detector. () shear force, () deagglomeration, () feed rate, ( ) pin vibration.

    (salbutamol sulfate) was assumed to show the smallest particles, while the excipient (lactose) was

    associated with the largest particles and less uniform size distribution.

    The interactions of the variables at two levels (binary interactions) were more relevant for the

    results presented by number of counts (smaller particles) than for the results presented by volume,

    suggesting that the distribution by number is more sensitive to changes of these variables than the

    distribution by volume. These interactions can occur between 2, 3, 4, or 5 parameters (results not

    shown). However, as the number of interactions increases, the interpretation of the eect becomesmore dicult because the variables contribute dierently to the nal result. Results not presented

    have shown that as the level of the interaction increases, the module of the result tends to decrease

    even if the result is highly signicant. Thus, interactions with higher number of levels (more than

    3) have proved to be more dicult to discuss based solely on the data available.

    Interesting to point out are the results for the interaction between PMV and SF (p 0:001,

    Table 3). A positive value for the distribution by number (Table 2) suggests an interaction between

    the two variables, whereby the overall decrease on the value of the mode (balance between the

    eect of PMV and SF) is about half of the value expected. For the interaction between PMV and

    D and PMV and FR although an interaction was expected this time, the value is smaller for the

    rst interaction and larger for the second when the balance between the main eects of the variablesare considered. These interactions are signicant at p 0:001 and 0:01, respectively (Table 3).

    It can be assumed that these variables change the sample in the way it is presented to the detector.

    Also signicant are the interactions between the shear force and other variables (SF and D, FR and

    PV). On the contrary, interactions such as PMVPV, DFR orDPV are not signicant. A possible

    explanation for these observations is that the shear force transforms agglomerates in individual

    particles that are measured by the detector. Since neither the pin vibrator nor the deagglomeration

    directly aect the process of counting, their interactions are not signicant.

    Due to the fact that the signicant interactions were observed for the measurements by number,

    the results presented by volume were not discussed in detail.

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    14/19

    522 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Fig. 9. Eect of the shear force on the measured diameter of the particles with pin vibrationOn (a) and O (b).

    ( - - -- - -) deagglomeration: normal; feed rate: low; () deagglomeration: high; feed rate: low; (- - -- - - )deagglomeration: normal; feed rate: med; () deagglomeration: high; feed rate: med; (- - - - - -) deagglomeration:

    normal; feed rate: high; () deagglomeration: high; feed rate: high.

    4. Conclusions

    The study has proved the ability of the AerosizerJ to determine the aerodynamic and geometric

    diameters both by number and by volume of counts. Although the equipment performs in a fairly

    constant way, its calibration is convenient prior to measurements. As anticipated, the measurements

    made by the equipment are more accurate for powders containing single populations of particles,

    rather than with samples made up of two or more populations, which are more sensitive to changesin the parameterization of the equipment. However, if two populations of particles with distinct size

    distribution are present, it is still possible to characterize each population in the mixture, provided

    adequate analysis of the results is made.

    It was observed that the larger the number of particles counted, or the lower the feed rate, the

    larger the required run time of the equipment. The counting eciency and reproducibility were

    aected by the number of counts performed. For the two pharmaceuticals considered in the study,

    for example, a minimum of 120,000 counts were necessary. Since a dierent number of counts

    is required when other materials are considered, a preliminary evaluation of these parameters is

    recommended before starting the analysis.

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    15/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 523

    Fig. 10. Measured diameter of the particles as a function of feed rate at dierent voltages applied to the detector: (a)

    850 V and pin vibrationOn; (b) 850 V and pin vibrationO; (c) 1100 V and pin vibrationOn; (d) 1100 V and pin

    vibrationO. (- - - - - -) Shear force: low; deagglomeration: normal; (- - - - - -) shear force: med; deagglomeration:

    normal; (- - - - - -) shear force: high; deagglomeration: normal; (- - - - - -) shear force: peak; deagglomeration: normal;

    () shear force: low; deagglomeration: high; () shear force: med; deagglomeration: high; ( ) shear force:

    high; deagglomeration: high; () shear force: peak; deagglomeration: high.

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    16/19

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    17/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 525

    Table 4

    Layout of a table for a 23 factorial design

    A B C AB AC BC ABC

    (1) + + +

    A + + +

    B + + +

    C + + +

    AB + + +

    AC + + +

    BC + + +

    ABC + + + + + + +

    Table 5

    Conversion of a three level factor into factors A and B

    Three level Two level

    X A B

    x1

    x2 +

    x2 +

    x3 + +

    eects. Eects for the factors and interactions are found by application of the signs to the set of

    experimental results.

    To nd the main eect or the interaction eect, the signs in the corresponding column are

    multiplied by the corresponding combination, added and the sum divided by 2k1.

    For example, the eect of A is

    A = 1

    231[(a + ab + ac+ abc) ((1) + b + c+ bc)]

    and the interaction ABC is

    ABC= 1

    231[(abc+ a + b + c) ((1) + ab + ac+ bc)]:

    A.2. More than two levels for each factor

    The method just described cannot be used with three or four levels for each factor. To overcome

    the problem, Montgomery has developed a technique that allows the analysis of these factorials.

    Let X be a three level factor (x1; x2; x3), it can be decomposed into two factors with two levels

    each, factors A and B (Table 6):

    Let Y be a four level factor (y1; y2; y3; y4), it can be decomposed into the factors C and D,

    (Table 5):

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    18/19

    526 P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527

    Table 6

    Conversion of a four level factor into factors C and D

    Three level Two level

    Y C D

    y1

    y2 +

    y3 +

    y4 + +

    In practice, these new factors substitute the three or four levels of individual factors and can be

    used as mentioned previously, including the calculation of the eects and interactions.

    In the analysis of the results the eect of A has to be added to the eect of B to nd the eect

    of X, and the eect of C to the eect of D to show the eect of Y.

    References

    Allen, T. (1990). Sampling and sizing from the atmosphere. In: T. Allen (Ed.), Particle size measurement (4th edn)

    (pp 112115). New York, USA: Chapman & Hall.

    Amherst process instruments, API Aerosizer LD, User Manual (V7.02), Hadley, MA, USA.

    Armstrong, N. A., & James, K. C. (1996). Pharmaceutical experimental design and interpretation (pp. 131167). London,

    UK: Taylor & Francis.

    Barber, T. A. (1993). Particle population analysis. In: T. A. Barber (Ed.), Pharmaceutical particulate matter, analysis

    and control (pp. 281298). Bualo Grove, IL, USA: Interpharm Press.

    Dunbar, C. (2002). Dry powder formulations for inhalation. Drug Delivery Systems & Science, 2, 7880.Gonda, I. (1992). Targeting by deposition. In A. J. Hickey (Ed.), Pharmaceutical inhalation aerosol technology

    (pp. 6182). New York, USA: Marcel Dekker Inc.

    Hickey, A. J. (2002). Delivery of drugs by the pulmonary route. In G. S. Banker, & C. T. Rhodes (Eds.), Modern

    pharmaceutics (pp 484490). New York, USA: Marcel Dekker Inc.

    Houzego, P. (2002). Deaggregation mechanisms in dry powder inhaler. Drug Delivery Systems & Science, 2, 8185.

    Laitinen, N., & Juppo, A. M. (2003). Measurement of pharmaceutical particles using a time-of-ight particle sizer.

    European Journal of Pharmaceutics and Biopharmaceutics, 55, 9398.

    Mitchell, J. P., & Nagel, M. W. (1996). An assessment of the API AerosizerJ for the real-time measurement of medical

    aerosols from pressurized Metered-Dose Inhaler (pMDI) systems. Aerosol Science and Technology, 25, 411423.

    Mitchell, J. P., & Nagel, M. W. (1999). Time of ight aerodynamic particle size analyzers: Their use and limitations for

    the evaluation of medical aerosols. Journal of Aerosol Medicine, 12, 217240.

    Mitchell, J. P., Nagel, M. W., & Archer, A. D. (1999). Size analysis of a pressurized metered dose inhaler delivered

    suspension formulation by the API Aerosizer time-of-ight aerodynamic particle size analyzer. Journal of Aerosol

    Medicine, 12, 255264.

    Montgomery, D. C. (1996). Factorials with mixed levels. In: D. C. Montgomery (Ed.), Design and analysis of experiments

    (4th ed.) (pp. 461465). New York, USA: Wiley.

    Myers, R. H., & Montgomery, D. C. (1995). Response surface methodology: Process and product optimization using

    designed experiments (pp. 3256). New York, USA: Wiley.

    Nagel, M. W., Wiersema, K. J., Bates, S. L., & Mitchell, J. P. (2002). Size analysis of a pressurized metered dose inhaler

    delivered solution formulation by an Aerosizer-LD time-of-ight aerosol particle size spectrometer. Journal of Aerosol

    Medicine, 15, 7585.

    Oskouie, A. K., Noll, K. E., & Wang, H. C. (2002). Determination of particle diameter and lumped density and shape

    factor using a three beam time-of-ight (TOF) instrument. Journal of Aerosol Science, 33, 11251138.

  • 7/26/2019 Sizing of powders in inhalers with an AerosizerJ according to.pdf

    19/19

    P.J. Mendes et al. / Aerosol Science 35 (2004) 509 527 527

    Oskouie, A. K., Noll, K. E., & Wang, H. C. (2003). Minimizing the eect of density in determination of particle

    aerodynamic diameter using a time-of-ight instrument. Journal of Aerosol Science, 34, 501506.

    Peters, T. M., Vanderpool, R. W., & Wiener, R. W. (2001). Methodology for measuring PM 2.5 separator characteristics

    using an Aerosizer. Aerosol Science & Technology, 34, 398406.Ramon, M., Juan, G., Ciscar, M. A., Amrti-Bonmati, E., Morcillo, E. J., & Main, J. (2000). Inuence of low temperature

    on bronchodilatation induced by terbutaline administered by metered dose or dry powder inhalers in asthmatics.

    Fundamentals of Clinical Pharmacology, 14, 133138.

    Srichana, T., Brain, A., Marriott, C., & Martin, G. P. (2000). A study of drug-carrier interactions in dry powder inhaler

    formulations using the Andersen cascade impactor, X-ray microanalysis and time-of-ight aerosol beam spectrometry

    (TOFABS). Chemical Pharmaceutical Bulletin, 48, 167174.

    Thornburg, J., Cooper, S. J., & Leith, D. (1998). Counting eciency of the API Aerosizer. Journal of Aerosol Science,

    30, 479488.

    US Pharmacopoea 26The United States Pharmacopoeial Convention (2003). USP Convention, Rockville, MD, USA.

    Wierlik, H., Diepenmaat, P., & Damhuis, R. (2002). Formulation of lactose for inhaled delivery systems. Pharmaceutical

    Technology Europe, 14, 4752.