SimSlim TM Patient Simulator
Operator's Manual
PReNK TECHNOLOGIES
www.pronktech.com
©2014
SimSlim, Pronk, and Pronk Technologies are trademarks of Pronk Technologies Inc.
©2012 Pronk Technologies Inc. All rights reserved.
CAUTION:
Risk of explosion if lithium battery is replaced by an incorrect type. Dispose
of used batteries according to country and state specific laws and
regulations.
PRC:,NK TEOHNC:,LC:,C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 2
Compliant products conforming to the relevant standards and directives required for CE have a CE Mark label, as shown below, affixed to the unit. CE Marked products also comply with the WEEE directive for category 9 Monitoring and Control Instrumentation products. Do not dispose of this product as unsorted waste. To dispose, contact Pronk Technologies at [email protected]. A Return Material Authorization will be issued to return the unit for disposal.
Getting Started 4
Detailed ECG Operation 6
Detailed Invasive Blood Pressure Operation 8
Detailed Cardiac Output Operation 10
Detailed YSI Temp Operation 12
Troubleshooting Tips 13
Services 13
Warranty and Repair Information 14
Speci�cations 15
Revision L 07/18/2017
Introduction
The S/mSl/m ,.., simulator is a compact, rugged, easy to use multi-parameter patient simulator
with the following features:
• 10 Snap/Banana plug ECG, Resp and Pacer Simulation.
• 4 Invasive Blood Pressure Channels via four 6 pin mini-din connectors.
• Cardiac Output Simulation.
• YSI 700 and 400 Temp Simulation via two 3.5mm stereo jacks.
• Up to 1 O Year Battery Life (no charging or changing required).
• Simple, single button operation.
• 4mm and 1/8" banana plug grooves for easy hook up to 12 lead carts.
• 50+ different selectable simulations.
Getting Started • please refer to each parameter section in this manual, for more
detailed operational information
1. Turning the SlmSllm Simulator on
• You never have to. The SimSlim simulator runs all day, every day for up to 10 years (depending
how often the LEDs are lit and the duration of various modes selected) on its internal battery.
2. Mode Selection
• The label on the front face has 5 blue boxes, each with an LED. Each box represents a different
mode.
• To show the current mode, click the MODE button.
• To change the mode, double click the MODE button.
• After two hours with no button presses, the SimSlim simulator will return to
the ECG NSR/ADULT mode.
F'Rc:>NK TE=HNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 4
3. Waveform Selection Within a Mode
• Some of the SimSlim simulator modes contain multiple waveforms.
• An example of this is ARR MODE 1 which has NSR with PVCs, VTACH, VFIB and
Asystole waveforms.
• These waveforms are listed on the bottom label.
• To change between waveforms within a mode, click the mode button once.
4. Zeroing Invasive Blood Pressures
• Click the mode button once to zero all blood pressures for 15 seconds, then zero monitor within
that 15 seconds. After 15 seconds the blood pressures will return to their prior values. You may
shorten the 15 second wait time to resume IBP simulation mode by pressing the MODE button
once any time after zeroing the monitor.
5. Cardiac Output Readings
• Double click the mode button to select the desired C.O. mode.
• Enter into the monitor the appropriate computational constant determined by referencing the
injectate temperature listed on the CC table on the bottom label of the SimSlim
• Hold the mode button for 2 seconds to start a simulated C.O. injection.
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 5
Detailed ECG Operation
• Connecting the ECG Leads
ECG lead wires with snaps can be connected directly to the SimSlim simulator snaps. The
snaps are clearly labeled with the name and color code of the lead. If your lead wires have
banana plugs rather than snaps, the banana plugs are slid into the grooves under the snap
such that the metal part of the plug is against the underside of the snap. Two different size
grooves are provided for different size banana plugs.
• ECG Size
ECG size is 1mV, Lead II, in ECG NSR MODE. Resp size is 4.0 Ohms in ECG NSR, ECG
NEO and HR STEPS MODES.
• ECG NSR MODE
Simulates a patient with a NSR waveform at 60 8PM. Resp is at 30 8PM.
• ECG NEO MODE
Simulates a patient with a NSR waveform at 140 8PM. Resp is at 70 8PM.
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page6
1/2
MODE
Revison L 07/18/2017
Detailed Invasive Blood Pressure Operation
• Connecting the Cables
You will need a monitor specific cable to connect the SimSlim simulator IBP simulation ports to
your monitor. A variety of these cables are available from Pronk Technologies and other
sources. Pronk's cables are supplied as a 6' (1.8 m) extension cable which connects to a 6"
(15cm) adapter cable. Most simulator cables with 6 pin mini-din connectors should work, as the
6 pin min-din simulator IBP pin-out is fairly standard. Wiring is as follows: + Excit = pin 1, - Excit
= pin 4, + Sig = pin 3, -Sig = pin 6. Contact Pronk Technologies Inc. at 800/541-9802 for more
information.
The IBP simulation ports are labeled on the bottom of the unit.
• Zeroing Simulated Blood Pressures
In all modes except IBP STEPS, clicking the mode button will force all four IBP channels to
OmmHg for 15 seconds. If you desire to return to normal operation prior to the 15 second
timeout, a second MODE button click will 'unzero' all channels. (Note that this 'click to unzero
function' is not reliable in Beta software prior to B.3)
• Sensitivity
The SimSlim simulator simulates 5uVN/mmHg transducers.
• ECG NSR MODE
Simulates a NSR adult patient at 60 BPM with:
• ART pressure on P1 of 120/80
• PA pressure on P2 of 30/1 O
LV Pressure on P3 of 125/15
• CVP Pressure on P4 of (6) mean
PRc:>NK T'EOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Pages
• ECG NEO MODE
Simulates a NSR neonatal patient at 140 BPM with:
• ART pressure on P1 of 60/30
PA pressure on P2 of 20/10
LV Pressure on P3 of 60/5
CVP Pressure on P4 of (2) mean
• 100 mmHg MODE
Simulates static 1 OOmmHg on all four channels.
• 200 mmHg MODE
Simulates static 200mmHg on all four channels.
• IBP STEPS MODE
In this mode, a selection of static blood pressures are available on all four channels. The
pressures available are shown on the bottom of the unit and are 0, 25, 50, 100, 150, 200, and
250mmHg. When entering the mode, the pressure will be OmmHg, so zero the monitor while in
this step. Click the MODE button to advance to the next pressure step.
PRC)NK TEOHNC)LC)C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 9
Detailed Cardiac Output Operation
• Connecting the Cables
The SimSlim simulator is designed to operate without a custom C.O. cable. Instead, use the
monitor's standard C.O. cable. The connections are as follows:
1. Connect the monitor's C.O. cable to the monitor.
2. Connect the catheter connector on the C.O. cable to the C.O. port on the SimSlim simulator.
3. Make sure the normal injectate temp sensor is connected to the C.O. cable. The injectate
temp sensor should be left at room temperature. It does not need to be be placed in a bath of
any kind, but for best results, place it somewhere the temperature will be room temperature and
relatively stable (i.e. not directly, under a hot lamp or in your hand).
• Reading the lnjectate Temp
With the cables connected correctly, you should be able to read the room temperature as
measured by the injectate probe from the monitor.
• Looking up the Computation Constant (C.C.) Value
Once you know the current injectate temp, you can look up the C.C. value from the table on the
back of the SimSlim simulator. Once you have the correct C.C. value, enter it into the monitor.
The granularity of the C.C. table on the SimSlim simulator can introduce up to a 5% error in the
readings. This should be fine for most applications, but if you want better precision you can
interpolate between values. For example, if your injectate temp was 25.5C, you would look up
the value for 25 (0.500) and the value for 26 (.545) and split the difference (0.522).
If you want to use an injectate temp value that is not in the table, you can calculate the correct
C.C. value according to C.C.=6/(37-Ti).
• Selecting the Simulation mode
Referring to the label on the side of the SimSlim, navigate the modes by double clicking the
MODE button to select the proper mode to simulate the intended 3, 5 or 7 1/m simulation.
PRC:,NK TECHNC:,LC:,C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 10
• Starting a Simulated Injection
Hold the MODE button down for 2 seconds to start a simulated injection. When the injection is
started, the LED will flash. If you are doing C.O. simulations for a long period of time, keep in
mind that the injectate temperature may drift and the C.C. value may have to be updated from
time to time for best accuracy.
• Using an lnjectate Temp Simulator
It is possible to use an injectate temp simulator rather than the actual injectate temp probe.
Unfortunately, different monitor vendors use such a wide variety of thermistors and connectors
for injectate temperature that Pronk Technologies cannot supply these. One solution would be to
fabricate an injectate temp simulator locally from an injectate probe and some resistors. Some
vendors use YSI 400 or YSI 700 thermistors for injectate temp. In these cases, the internal YSI
temp simulator in the SimSlim simulator can be used.
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 11
1/4”
1/4”
monitors
cable and a 3.5mm to 1/4”stereo plug converter.
Revision L 07/18/2017
Troubleshooting Tips
TROUBLESHOOTING TIPS
SYMPTOM SOLUTION
Not calling VTACH or VFIB Some arrhythmia systems will not call VTACH on a straight jump out of clean NSR. At the start of ARR MODE 1 the NSR section has some PVCs and runs mixed in. In some cases, it may be necessary to let the arrhythmia system see some of these PVCs prior to transitioning into VT ACH.
I am trying to advance For rapid single clicking, watch the LED. If you click a second quickly through the time while the LED is still on, this will be seen as a double click, waveforms, but which will change the mode. If you click a second time just after sometimes I the LED has gone out, it will be seen as a second single click, accidentally change the which will advance the waveform. mode
Misshapen or undersized With Hardware E6 and prior, some monitors will occasionally ECG waveform have interaction between the IBP and ECG. Unplug IBP to
work around the problem. On E7 hardware, touching the metal shell of an IBP connector and an ECG snap at the same time can cause waveform interaction.
Not picking RESP Most monitors look for respiration drive across RNLA leads.
waveform Some do not. If RESP waveform is not picking, try swapping the RL and LL leads. Please keep in mind that RESP is zero in ARR MODE 1 and ARR MODE 2.
No Heart rate on EASI When viewing AVR or V2 lead, change the V lead to V6 snap. configured telemetry
Unable to resolve Contact Pronk Technologies' Technical Support at: problem (800) 541-9802
Services (Prices subject to change without notice)
SERVICE ITEM NUMBER DESCRIPTION PRICE (USO$)
Annual SimSlim Calibration Full checkout, adjustment and calibration certificate
*Pronk Technologies recommends an annual calibration interval.
PRC:,NK TEOHNC:,LC:,GIES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 13
Call for Pricing
SimSlim Simulator Limited Warranty
The SimSlim simulator is warranted against defects in materials and workmanship for a period
of forty eight (48) months from the date of shipment to the original purchaser. Warranty is valid
only to the original buyer. Defective equipment should be returned freight prepaid to Pronk
Technologies. Equipment returned with defective parts and assemblies shall be either repaired
or replaced at the manufacturer's sole discretion. This warranty is not applicable if the unit has
been opened, if repair has been attempted, if the unit has been damaged due to operation
outside the environmental and power specifications for the product, or due to improper handling
or use.
If any fault develops, notify Pronk Technologies (see Returns and Repairs, below) giving full
details of the difficulty, and include the model and serial number of the device. Upon receipt of
shipping instructions, forward the device prepaid and repairs will be made at the factory.
The foregoing warranty is in lieu of all other warranties expressed or implied, including but not
limited to any implied warranty or merchantability, fitness or adequacy for any particular purpose
or use. Pronk Technologies shall be liable only for repair or replacement of the SimSlim
Simulator and optional features. Pronk Technologies shall not be liable for any incidental or
consequential damages.
Order Cancellation and Refund Policy
You may return your item within 14 days of delivery for a full refund. We are unable to exchange
items (however, if you received a defective or incorrect item, we will be happy to make an
exchange). ltem(s) returned for refund must be in its original condition, undamaged and with no
missing parts, packed in its original packaging, and include both the original receipt and an RMA
number.
We will notify you via e-mail or fax of your refund once we have received and processed the
returned item. You can expect a refund in the same form of payment originally used for purchase
within 7 to 14 business days of our receiving your return.
Returns and Repairs
Please call Pronk Technologies' Service Department at 800-541-9802 to obtain a Return
Merchandise Authorization (RMA) number and the shipping address. Returns should be
packaged securely in the original packaging materials. The RMA number should be clearly
marked on the packaging. If the return is for a new item and is a result of our error, we will make
arrangements for payment of return shipping. Otherwise, items should be returned freight
prepaid to Pronk Technologies.
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 14
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 15
PRc:>NK TEOHNc:>Lc:>C31ES
Revision L 07/18/2017 www.pronktech.com (800) 541-9802 Page 16