Service Manual - Frank's Hospital...

52
Le marquage CE qui se trouve sur ce produit indique que celui-ci est en conformité avec la directive européenne des dispositifs médicaux 93/42/CEE This product bears a CE marking in accordance with the provisions of the 93/42/EEC MDD dated June 14, 1993. 0459 Technical Publication SM-0521R3 Service Manual BRS Basic Radiographic System STEPHANIX rue Jean Moulin Zone Industrielle du Bayon 42150 La Ricamarie FRANCE Tel. : 00 33 4 77 47 81 60 ; Fax : 00 33 4 77 37 55 19

Transcript of Service Manual - Frank's Hospital...

Le marquage CE qui se trouve sur ce produit indique que celui-ci est en conformité avec la directive européenne des dispositifs médicaux 93/42/CEEThis product bears a CE marking in accordance with the provisions of the 93/42/EEC MDD dated June 14, 1993.0459

Technical PublicationSM-0521R3

Service Manual

BRSBasic Radiographic System

STEPHANIX

rue Jean Moulin

Zone Industrielle du Bayon

42150 La Ricamarie

FRANCE

Tel. : 00 33 4 77 47 81 60 ; Fax : 00 33 4 77 37 55 19

Basic Radiographic System

Service Manual

SM-0521R3

REVISION HISTORY

REVISION DATE REASON FOR CHANGE

0 April 22, 2003 First edition

1 30 March, 2004 Installation and dimensions modified.

2 07 May, 2005 Installation improvements.

3 12 Jul, 2005 New schematics and updated photos.

This Document is the English original version, edited and supplied by the manufacturer.

The Revision state of this Document is indicated in the code number shown at the bottom of this page.

ADVISORY SYMBOLS

The following advisory symbols will be used throughout this manual. Theirapplication and meaning are described below.

DANGERS ADVISE OF CONDITIONS OR SITUATIONS THATIF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUSPERSONAL INJURY OR DEATH.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOTHEEDEDORAVOIDEDCOULDCAUSESERIOUSPERSONALINJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT ORDATA.

Advise of conditions or situations that if not heeded oravoidedcould causepersonal injury or damage to equipmentor data.

Note . Alert readers to pertinent facts and conditions. Notes representinformation that is important to know but which do not necessarilyrelate to possible injury or damage to equipment.

Basic Radiographic System

Service Manual

SM-0521R3

SAFETY SYMBOLS

The following safety symbols will be used in the equipment.Their meaning are described below.

Attention, consult accompanying documents.

Ionizing radiation.

Type B equipment.

Dangerous voltage.

Ground.

This symbol indicates that the waste of electrical andelectronic equipment must not be disposed as unsortedmunicipal waste and must be collected separately. Pleasecontact an authorized representative of the manufacturer oran authorized waste management company for informationconcerning the decommissioning of your equipment.

Basic Radiographic System

Service Manual

SM-0521R3 i

TABLE OF CONTENTS

Section Page

1 INTRODUCTION 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.1 Tools 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.2 Pre-Installation Checks 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2 UNPACKING 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3 INSTALLATION 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.2 Positioner Installation overview 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4 ADJUSTMENTS 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.1 Adjustment tools 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.2 Alignment of X-ray Beam 18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.2.1 Alignment of Light Field with X-ray Field 22. . . . . . . . . . . . . . . . . . . . . . . . .

4.2.2 Perpendicularity Adjustment of X-ray Beam with Image Receptor 25. . .

4.2.3 Centering of X-ray Field and Image Receptor Bucky Assembly 27. . . . .

4.3 Field Size Indicator Test 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.4 X-ray Tube or Collimator Replacement 30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.5 Adjustment of rotation Plate 30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5 RENEWAL PARTS 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6 MANTEINANCE 33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7 INTERCONNECTION MAPS 35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Basic Radiographic System

Service Manual

SM-0521R3ii

Basic Radiographic System

Service Manual

SM-0521R3 1

SECTION 1 INTRODUCTION

1.1 TOOLS

The following hand tools are required for the Installation:

• Standard service engineers tool kit.

• Electric drill motor and assorted bits.

1.2 PRE-INSTALLATION CHECKS

Prior to beginning installation it is recommended to inspect the site and verifythat the X-ray room complies with requirements such as:

• Themaximum dimensions and systemmovements to plan theminimumspace required in the room:

Maximum Height 2500 mm

Maximum Width 1550 mm

Maximum Length 1340 mm

• Conduits and walls are ready to install the System.

• Electricity installation:

Power Line (for Collimator Lamp): 24 VAC, 50/60 Hz, 6.5 A

Power Line (for Movement Locks): 24 VDC, 0.2 kVA

ACCORDING TO MDD93/42/CEE, THIS UNIT IS EQUIPPEDWITH EMC FILTERS. THE LACK OF THE PROPERGROUNDING MAY PRODUCE ELECTRICAL SHOCK TO THEUSER.

Basic Radiographic System

Service Manual

SM-0521R32

Illustration 1-1Dimensions

Max.H

eight=

2500

mm.

AlturaMax.=

2500

mm.

Max.Travel=

1400

mm.

RecorridoMax.=

1400

mm.

Colum

nMax.height=

2240

mm.

AlturaMax.C

olum

na=2240

mm.

700

707 mm.

2148

566

885

1400

SID = 1400 mm.

Max. Width = 1595 mm.Anchura Max. = 1595 mm.

(*) Dimensions Subject to change depending on the type of tube

(*) Dimensiones sujetas a cambio dependiendo del tipo de tubo

1795

440

610

497

Basic Radiographic System

Service Manual

SM-0521R3 3

SECTION 2 UNPACKING

The Unit is shipped in one box to facilitate transport and installation. Uponreceipt of the X-ray unit and associated equipment, inspect all shippingcontainers for signs of damage. If damage is found, notify the carrier or hisagent immediately.

1. Place the shipping pallet near its final site in the room and remove all itslaterals. Do not discard any packing material such as envelopes, boxes,bags until all parts are accounted for as listed on the packing list.

AT LEAST THREE PEOPLE ARE REQUIRED TO REMOVEALL HEAVY COMPONENTS FROM THE SHIPPING PALLET.

2. When the equipment is unpacked, check part numbers and serialnumbers of each component with its identification labels. Inspect allpieces for visible damages. If any damaged part is found, repair it ororder its replacement to prevent unnecessary delay in installation.

3. Verify that all items on the customer order are present.

4. Leave a working area around equipment until its final installation iscomplete.

Basic Radiographic System

Service Manual

SM-0521R34

This page intentionally left blank.

Basic Radiographic System

Service Manual

SM-0521R3 5

SECTION 3 INSTALLATION

3.1 POSITIONER INSTALLATION OVERVIEW

2.-- Install the Wall Support at the back of the Column(the Column laid on the pallet)

1.-- Install the Support of the Tube--Collimator Assemblyand the Control panel (in some cases it is factoryinstalled)

3.-- Drill WALL HOLES and anchor the Wall Support .

4.-- Install the Collimator and the X-ray tube.

5.-- Remove the Safety Locking Rod/Central carriage Screw.

7.-- Install the Oscillating Grid Assembly or Fixed GridAssembly. (level it)

9.-- Connect cables (HV, Tube, Stator, Bucky or Grid)and guide them .

10.-- Check levels on Column, Tube-Collimator Assemblyand Receptor Assembly.

TICK BOXES WHEN DONE

1

2

4

7

9

5

10

6.-- Install all the counterweight plates supplied.

6

8.-- Drill FLOOR HOLES and anchor the unit to the floor.

Basic Radiographic System

Service Manual

SM-0521R36

Perform Installation following the order described below.

DO NOT REMOVE SAFETY LOCKING ROD FROM THECOUNTERWEIGHT CARRIAGE AND/OR THE CENTRALCARRIAGE SCREWS (2) BEFORE SPECIFICALLYINSTRUCTED IN THIS DOCUMENT.

Due to packaging requirements, some Units may not include theTube Collimator Assembly and the Control panel factorypre-installed. If that is the case, follow steps 1 and 2, if not, go tostep 3.

1. Assemble the Support of the Tube-Collimator Assembly to the Upperarea of the Swivel Arm using the four Safety Screws installed in the Arm.

SAFETY SCREWS

SUPPORT OF TUBE-COLLIMATOR ASSEMBLY

2. Assemble theControl Panel in the Support with the two screws, washersand nuts attached to the Control Panel.

SUPPORT OF TUBE-COLLIMATOR ASSEMBLYCONTROL PANEL

SCREWS, WASHERS AND NUTS

Note .

Note .

Basic Radiographic System

Service Manual

SM-0521R3 7

3. Assemble the Upper Wall Support to the Column. Lift the Column asmuch as necessary as to mount the Upper Wall Support in its upperpart.

SAFETY SCREWS

At least two people are required to perform the next operation.

4. Position the column against the wall on its final site in the room. Whileone person is holding up the Column, the other one should level itvertically on both lateral sides and on the front. Also use the levelingscrews at the base of the column.

5. Mark its anchoring positions on the wall (not to the floor yet).

Distances shown on Illustration 3-1 may also be taken asreference to mark the anchoring holes position.

Note .

Note .

Basic Radiographic System

Service Manual

SM-0521R38

Illustration 3-1Drill Template

1839

160

FOUR FLOOR MOUNTING HOLESFOR SCREWS10 mm.

FOUR WALL MOUNTING HOLES

FOR SCREWS 8 mm.

Basic Radiographic System

Service Manual

SM-0521R3 9

6. Move the Column and prepare the anchorages.

BEFORE ANCHORING THE UNIT, CHECK THAT THE WALLAND THE ANCHORING SYSTEM ARE STRONG ENOUGH(2000 NEWTON TRACTION FORCE) TO ENSURE A SAFEINSTALLATION. SOME NON BRICK WALLS MAY REQUIREADDITIONAL ANCHORAGE INSTALLATION.

7. Drill holes and anchor the Wall Support. Make sure the Column is firmlyanchored and leveled at front, sides and base.

8. Install the X-ray Tube in the Upper Support of the Column using theCollimator Adaptation Ring and its four Safety Screws (Allen).

9. Before installing the Collimator, unscrew the same number of turns thefour Centering Adjustment and Safety Screws (Allen) to allow theCollimator installation in the Collimator Adaptation Ring. Adjust theCollimator Blades to their widest setting and carefully install theCollimator centering it with the X-ray Tube window.

Tighten carefully the four Centering Adjustment and Safety Screws(Allen) equally (same number of turns) until Collimator is centered andheld firmly on the Coupling Ring (support). (Also, refer to CollimatorManual).

Illustration 3-2Installation of X-ray Tube and Collimator

X-RAY TUBE CABLE

X-RAY TUBE COVER

COLLIMATOR COVER

COLLIMATOR CABLE

CENTERING/ADJUSTMENT SCREWS

10. Remove the X-ray Tube and Collimator covers and connect the wiresfrom the upper harness to the X-ray Tube and Collimator according totheir identification (refer to 54301065).

Basic Radiographic System

Service Manual

SM-0521R310

11. Remove now the Safety Locking Rod from the Counterweight Carriageand the Central Carriage Screws (2). The carriage may go down.

Safety locking Rod

Central Carriage Screw

12. Add all the counterweigh pieces to both entrances at the back side of theColumn and secure them with the provided nuts. The unit is factorycounterweighted, so the counterweights provided are to be installed.

Counterweight entrancesat Column Back

13. Add all the counterweigh pieces to both entrances at the back side of theArm and secure them with the provided nuts. The unit is factorycounterweighted, so all the counterweights provided are to be installed.

Counterweight entrancesat Arm Back

14. Now install the Oscillating Grid Assembly or the Fixed GridAssembly. Follow steps 14 to 17 for Oscillating Grid Assembly or steps18 to 21 for Fixed Grid Assembly.

Basic Radiographic System

Service Manual

SM-0521R3 11

15. Oscillating Grid Assembly installation:Take out Table-Top of theOscillating Grid Assembly -- Four screws at the corner of the OscillatingGrid Assembly.

Table-Top

Corner Screw

16. Install the Oscillating Grid Assembly in the Swivel Arm with the 2 sets ofscrews (4 screws in each set). A set of gauges (included in the package)can be installed for perpendicularity adjustment.

Oscillating Grid Assembly Installation Screws

17. Connect J1 (only for Oscillating Grid Assembly with motorized Grid), J2and GND from the Harness to the Oscillating Grid Assembly.

J1

GND

18. Reinstall the Oscillating Grid Assembly Table-top.

Basic Radiographic System

Service Manual

SM-0521R312

19. Fixed Grid Assembly Installation: Remove the Safety Screws installedin the Fixed Grid Assembly.

20. Remove the Back Cover of the Swivel Arm end.

21. Install the FixedGrid Assembly in the Swivel Armwith the Safety Screwspreviously removed. (Before connect J2).

22. Level the Assembly with the leveling screws and re-install the BackCover.

Safety Screws

Back Cover at Swivel Arm End

Leveling Screws

J2

23. Connect the cables from the main harness (Stator, Bucky, Power Line,Collimator and ground (GND) cables routed directly from the Column) tothe Generator (for these connections refer to Interconnection Map54301065 and also Generator Service Manual).

24. Connect directly the ground cables (GND) from the Column to the Mainground stud (GND) of the Generator.

25. Connect HV Cables of the X-ray Tube.

Terminal Pins of HV Cables are extremely delicate and theyare easily damaged. Take particular care to handle themcarefully. Make sure that they remain straight and that thesplits in the pins are open (parallel to sides).

Basic Radiographic System

Service Manual

SM-0521R3 13

Prepare theHigh Voltage terminals that will be installed in the X-ray Tubereceptacles. Apply Silicone Paste over the entire surface of the Plugincluding the Pins.

Carefully connect cables to their related receptors of the Tube and fixtheir nuts tightly.

26. Fasten HV cables to lateral side of Pivoting Arm with the clamp on thecarriage of the Tube Collimator Assembly. Give the cable length enoughto enable movements of the of Pivoting Arm.

H V Cables

Fasteners

Tube HV Cables may also be tied up to a ceiling pole to avoidcollisions and to easy movements of the system.

27. Mark and drill the anchoring positions on the floor. Check leveling againin different sides of the unit and correct if necessary before fixing the unitdefinitely.

28. Anchor the Column to the floor.

Distances shown on Illustration 3-1 may also be taken asreference to mark the anchoring holes position.

29. Snap in the decorative base of the Columm (pressing it).

30. Turn the system on and check that all controls and movements operatecorrectly.

Note .

Note .

Basic Radiographic System

Service Manual

SM-0521R314

This page intentionally left blank.

Basic Radiographic System

Service Manual

SM-0521R3 15

SECTION 4 ADJUSTMENTS

4.1 ADJUSTMENT TOOLS

The following special Tools have been used for adjustment of the X-ray System(these tools are not included with the System):

• Collimator Test Tool (Model RMI 161B9).

• Beam Alignment Test Tool (Model RMI 162A).

• SID Test Stand Tool (Inside Case: RMI Model 175).

• Light Meter (Standard).

Basic Radiographic System

Service Manual

SM-0521R316

Illustration 4-1Checkings overview

1

2

3

CENTERING ADJUSTMENTSCREWS

Check /Correct levels of Column and tube

Check /Correct levels of Receptor

Check / Correct alignment of the light field with the X-ray field

Leveling Screws

Fixed Grid Assembly

Oscillating Grid Assembly

Basic Radiographic System

Service Manual

SM-0521R3 17

Illustration 3-1 (cont)Checkings Overview

CENTERING ADJUSTMENTSCREWS

5

Check / Correct the perpendicular adjustmentof the X-Ray Beam with the Image Receptor

Check / Correct centering of X-Ray field andImage Receptor(same adjustments as step 4)

4

7

6 Field Size Indicator Test

Collimator Lamp Brightness Test

A B C

DEVIATION

Leveling Screws

Fixed Grid Assembly

Oscillating Grid Assembly

Basic Radiographic System

Service Manual

SM-0521R318

4.2 ALIGNMENT OF X-RAY BEAM

1. Check with a level the vertical/horizontal position of the Column. Ifnecessary, modify the position of the Column with the screws attachedto the Base of the Column, a correct installation starts with a goodleveling of the column.

Levels

2. Check the level of the Tube-Collimator Assembly. The mechanicalinstallation provides a levelled Assembly, anyway check parallelism. Ifnecessary reinstall the Tube-Collimator Assembly. (Refer to installationsection).

3. Check the Receptor, first in horizontal position (thorax), then in verticalposition (Undertable). If necessary, move up or down the Receptorassembly with the Receptor Assembly leveling screws.

4. Place the Pivoting Arm in vertical position. The Tube-CollimatorAssembly has to beperfectly alignedwith theReceptor Assembly. Checkwith a level correct horizontal position of both assemblies.

Illustration 4-2Vertical Position of Pivoting Arm with Alignment Tools

Basic Radiographic System

Service Manual

SM-0521R3 19

5. Turn on Collimator light and center the Collimator in relation to theReceptor Table-Top. Horizontal and transversal position of the light axesprojected by the Collimator Lamp must be in line with the axes or filmsizes marked on the Table-Top.

If needed, modify Collimator position by carefully unscrewing andscrewing the four Centering Adjustment and Safety Screws (Allen) ofCollimator.

Illustration 4-3Collimator Screws

CENTERING ADJUSTMENT AND SECURITY SCREWS

6. Position the Collimator Test Tool (RMI model 161B) on the ReceptorTable-Top.

7. Turn on Collimator light and by means of the Collimator Control Knobs,center the Collimator Test Tool with the light axes projected by theCollimator Lamp.

Basic Radiographic System

Service Manual

SM-0521R320

8. Adjust the Light Field of the Collimator Lamp to the rectangle drawninside the Collimator Test Tool.

9. Place centered the Beam Alignment Test Tool (RMI model 162A) on theCollimator Test Tool.

Illustration 4-4Alignment Test Tools

BEAM ALIGNMENT TEST TOOL -- RMA MODEL 162A

COLLIMATOR TEST TOOL -- RMA MODEL 161B

10. Load Cassette film Tray with cassette film 24x30 and insert it.

11. Make an exposure at 60 kVp / 5 mAs.

12. Process film and:

a. Check that the X-ray Field falls just within the image of the innerrectangle of the Collimator Test Tool.

If an edge of the X-ray Field falls out of the inner rectangle meansa misalignment of the Light Field respect to the X-ray Field. Themaximummisalignment allowed is 2% of SID (for SID 1.4m = 2.8cm tolerance).

Refer to Section 4.2.1 for alignment of Light Field with X-ray Field.

Basic Radiographic System

Service Manual

SM-0521R3 21

b. Check that the X-ray Beam is perpendicular to the plane of theImageReceptor. If the Image receptor is parallel to theTable-Top,the perpendicularity of the X-ray Beam can be checked using theBeam Alignment Test Tool with the Collimator Test Tool.

Based on next illustration, the criteria for SID at 1.4 meter is:

-- If the image of the two balls overlap (A) the X-ray Beam isperpendicular to within 0.5o.

-- If the image of the top ball (larger shadow) intercepts the firstcircle (B), the X-ray Beam is about 1.5o away from theperpendicular.

-- If the imageof the top ball (larger shadow) intercepts the secondcircle (C), the X-ray Beam is about 3o away from theperpendicular.In cases (A) and (B) perpendicularity is within tolerance for SID at1 meter (top ball is within or intercepting the first circle). The thirdcase (C) needs readjustment.

Refer to Section 4.2.2 for perpendicularity adjustment.

A B C

c. Check that the X-ray Beam is properly centered with the ImageReceptor. To determine the center of the Image Receptor, drawdiagonal lines from corner to corner of the X-ray Film. (Alternately,the film can be folded in half and creased at the center). The twolines will cross in the center of the Image Receptor (film). Thendraw diagonal lines from the corners of the imaged X-ray Field.

If the center of the X-ray Field and Image Receptor is the same,the diagonals of both sets of lines should cross at the same point.The maximum misalignment allowed is 2 % of SID (for SID 1.4m= 2.8 cm tolerance).

Refer to Section 4.2.3 for centering of X-ray Field and ImageReceptor.

Basic Radiographic System

Service Manual

SM-0521R322

4.2.1 ALIGNMENT OF LIGHT FIELD WITH X-RAY FIELD

1. Place the Pivoting Arm in vertical position. The Tube-CollimatorAssembly has to beperfectly alignedwith theReceptor Assembly. Checkwith a level correct horizontal position of both assemblies.

2. Position the Collimator Test Tool (RMI model 161B) on the ReceptorTable-Top.

COLLIMATOR TEST TOOL

3. Turn on Collimator light and center the Collimator Test Tool with the lightaxes projected by the Collimator Lamp.

4. Adjust the Light Field of the Collimator Lamp to the rectangle drawninside the Collimator Test Tool.

Basic Radiographic System

Service Manual

SM-0521R3 23

5. Check on the processed film the adjustment required to meet the LightField with the X-ray Field. Identify the deviation on the axes imaged onthe film.

PROCESSED FILM

PROCESSED FILM

COLLIMATED LIGHT FIELD

COLLIMATED X-RAY FIELD

LIGHT FIELD MISALIGNMENT

GOOD FIELDS ALIGNMENT

COLLIMATED LIGHT FIELD = COLLIMATED X-RAY FIELD

Basic Radiographic System

Service Manual

SM-0521R324

6. Do not remove the Collimator Test Tool from its original position andadjust the Light Field bymoving theCollimator Unit and/or theCollimatorLamp.

a. Collimator Lamp must be ON during Light Field adjustment.

b. Modify Collimator position by carefully unscrewing and screwingthe four Centering Adjustment and Safety Screws (Allen) until theLight Field coincides with the axes reference imaged (numbersand dots) on the processed film.

Illustration 4-5Collimator Screws

CENTERING ADJUSTMENT AND SECURITY SCREWS

c. If adjustment is still necessary, modify position of Collimator Lamp(refer to Collimator Manual).

7. Repeat exposure and procedure until the result is satisfactory. Themaximum misalignment allowed is 2 % of SID (for SID 1.4m = 2.8 cm oftolerance).

Basic Radiographic System

Service Manual

SM-0521R3 25

4.2.2 PERPENDICULARITY ADJUSTMENT OF X-RAY BEAM WITH IMAGE RECEPTOR

In case that perpendicularity is out of tolerance (top ball is out of first circle),adjust perpendicularity as follows:

1. Place the Pivoting Arm in vertical position. The Tube-CollimatorAssembly has to beperfectly alignedwith theReceptor Assembly. Checkwith a level correct horizontal position of both assemblies.

The Receptor Assembly is factory adjusted (slightly up) and it isrecommended do not perform any additional correction. Duringprocedure, it must behorizontally placedat 0o (check positionwith a leveland with its indicator plate).

2. Position the Collimator Test Tool (RMI model 161B) on the ReceptorTable-Top.

3. Turn on Collimator light and center the Collimator Test Tool with the lightaxes projected by the Collimator Lamp.

4. Place centered the Beam Alignment Test Tool (RMI model 162A) on theCollimator Test Tool and observe if shadow of the Beam Alignment TestTool is projected in equal proportion around it.

5. Check on the processed film the adjustment required to center the topball mark. Shadow around the Beam Alignment Test Tool can also helpto make a first correction.

6. Loosen slightly the four Safety Screws (Allen M8) of the Tube-CollimatorAssembly. If required, perform the following adjustments:

G For horizontal correction move horizontally the Tube-CollimatorAssembly before tightening the four Safety Screws.

G For vertical correction loosen or tighten carefully the four LevelingScrews (Allen M6) of the Tube-Collimator Assembly beforetightening the four Safety Screws.

7. Repeat exposure and procedure until the result is satisfactory (top ballmust be inside of the first circle).

Basic Radiographic System

Service Manual

SM-0521R326

Illustration 4-6Adjustments in the Tube-Collimator Assembly

1.-- Leveling Screws: The Tube-Collimator Assy. moves slightlyup or down from one side.

2.-- Positioning screws and Fix-ing Screws:

2a: Loose first the Fixing Screws,then adjust the tube-collimatorwith the positioning screws2b: then tighten the FixingScrews.

3.-- Shims (factory installed/ad-justed, and extra units includedin package). These plates can beadded or removed in order to ex-tend/reduce the length of theTube-Collimator Assembly--.

12b

3

2a

Basic Radiographic System

Service Manual

SM-0521R3 27

4.2.3 CENTERING OF X-RAY FIELD AND IMAGE RECEPTOR RECEPTOR ASSEMBLY

The error between centers of the X-ray Field and the Image Receptor shouldnot be greater than 2% of the SID. (for SID 1.4m = 2.8 cm tolerance).

CENTER OF THE X--RAY FILM CENTER OF THE IMAGE

DEVIATION

If the deviation is greater than 2% of SID, perform the following procedure:

1. Check on the processed film the correction required for centering theX-ray Field with the Image Receptor. Adjustments will be performed asthe samewaydescribed for Perpendicularity correction, soonly re-adjustit if really is necessary.

Centering adjustments may affect to perpendicularitycorrections.

2. Repeat exposure and check centering until the result is satisfactory(centers position are within tolerance).

Basic Radiographic System

Service Manual

SM-0521R328

4.3 FIELD SIZE INDICATOR TEST

Before starting with the Field Size Indicator Test, the Alignment ofX-Ray Beam Test, the Alignment of Light Field with X-Ray FieldTest and the SID Indicator Test should be performed.

1. Place the Pivoting Arm in vertical position. The Tube-CollimatorAssembly has to beperfectly alignedwith theReceptor Assembly. Checkwith a level correct horizontal position of both assemblies.

2. Open the Collimator blades to set a Field Size of 24 x 30 cm for SID1,4 m.

3. Turn on the Collimator Light and center the Collimator in relation to theReceptor Table-Top. Horizontal and transversal position of the light axesprojected by the Collimator Lamp must be in line with the axes or FilmSize marked on the Table-Top.

Note .

Basic Radiographic System

Service Manual

SM-0521R3 29

4. Load on the Cassette Film Tray with a Cassette Film of 35 x 43 cmcentered and insert it.

5. Make an exposure at 60kVp, 1mAs.

6. Process film and check the following measurements known (refer toIllustration 4-7):

a. Measure the length of the X-ray Field on the Processed Film(identified as L)

b. Measure the width of the X-ray Field on the Processed Film(Identified as W)

Illustration 4-7X-ray Field Size

X-RAY FIELD SIZE ON THE FILM

FILM 35x43 cm

PROCESSED FILM

LENTH DIMENSION (L)

WIDTHDIMENSION(W

)

7. With the measurements obtained and according to the field sizeindication selected, the results should be:

G Length dimension (L) = 300 mm

G Width dimension (W) = 240 mm

8. The difference between the indicated Field Size and the obtained FieldSize may not exceed 1.5 % (rejection limit) of the SID in either direction.Therefore the deviation in any direction should not be higher than 21 mm(rejection limit) for SID at 1.4 meter.

9. If the deviation of Field Size Indication is out acceptance limits, it isnecessary to readjust the index of Collimator Blades Control Knobs. Forthat, loosen each Collimator Control Knob and position it according todeviation. Repeat the complete tests until the X-Ray Field Size selected(24 x 30 cm) is obtained.

Basic Radiographic System

Service Manual

SM-0521R330

4.4 X-RAY TUBE OR COLLIMATOR REPLACEMENT

This section only applies to systems where the X-ray Tube or theCollimator has to be replaced by a new different X-ray Tube orCollimator (different weight) or when the Swivel arm is notvertically balanced.

1. Place swivel Arm in horizontal position.

2. Remove the top cover of the Tube-Collimator Assembly Support.

3. Add or remove counterweight plates until the Arm is balanced.

Counterweight entrancesat Arm Back

4. Re-install the top cover.

4.5 ADJUSTMENT OF ROTATION PLATE

This section only applies to systemswhere the rotation plate doesnot match with the real angle.

1. Place swivel Arm in horizontal position.

2. Loose Safety Screws and adjust correct angle with a level.

3. Tighten Safety Screws.

CENTERING ADJUSTMENT AND SECURITY SCREWS

STEEL CABLE

Note .

Note .

Basic Radiographic System

Service Manual

SM-0521R3 31

SECTION 5 RENEWAL PARTS

Item Designation Qty FRU Rep Mfg. Ref. Second Ref. Remarks

Positioner

1 Switch button (black) 3 2 N 50613013

2 Handle 3 2 Y S02.05.012

3 Solenoid 24 VDC 1 1 Y 52301006

4 Electrobrake 24 -- 35 VDC (positive) 5 1 N 52301011

5 Basic Mill. Control PCB 1 1 Y A3510--03

6 Bucky Midwest 1 1 Y 6685-01

7 Collimator 1 2 N 6693--06

8 Lamp (Collimator) 1 2 N 54203005

18

7

7

6

5

3

4

21

1

Basic Radiographic System

Service Manual

SM-0521R332

This page intentionally left blank.

Basic Radiographic System

Service Manual

SM-0521R3 33

SECTION 6 MAINTENANCE

The Optima BRS has been designed as a long term X--ray positioner withminimum maintenance.

The mechanical parts of the Optima BRS require a yearly basis servicemaintenance from the installation date in the following way:

Visual Check: Covers, Pushbuttons, Table-top, Receptor Assembly, SteelCables appareance (from the Column Slot side), connectors and electricalcables.

Functional Check: Push-buttons, movements of the carriageand arm, cassettetray, perform an exposure and check correct alignment of film, Collimator andCollimator light (refer to Section 4).

Yearly Functional Check

ITEM ACTION

Push-buttons Visual Check, correct contact and corresponding movement.

Movement of Carriage Check complete travel of carriage, it should be soft and noiseless

Movement of Arm Check complete angle travel of the Arm, it should be soft and noiseless.

Both Steel Cables Move carriage to the lowest position and check appareance of both steel cables from the side slotof the column. Check connection plate and disassembly top cover to check the whole cable travel.They should be in without stranded or broken hairs.

Cassette Tray Check correct positioning inside Receptor assembly, correct clamping of film.

Collimator and Tube Refer to Section 4 in this Manual and Generator maintenance Section.

Basic Radiographic System

Service Manual

SM-0521R334

This page intentionally left blank.

Basic Radiographic System

Service Manual

SM-0521R3 35

SECTION 7 INTERCONNECTION MAPS

Refer to the following maps for details of the wire connections.

• Optima BRS-Who interconnections 54301083. . . . . . . . . . . .

• BRS System IM-339. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Basic Radiographic System

Service Manual

SM-0521R336

This page intentionally left blank.

A A

B B

C C

D D

E E

44

33

22

11

INTE

RC

ON

EX

ION

ES

OP

TIM

A B

RS

-WH

O(O

PTI

MA

BR

S-W

HO

INTE

RC

ON

NE

CTI

ON

S)

BA

SIC

MIL

CO

NTR

OL

boar

d(A

3510

-03-

rev.

A)

GN

D C

able

0 V

DC

+24

VD

C

GEN

ERA

DO

R C

OM

PAC

T(C

OM

PAC

T G

ENER

ATO

R)

Tarje

ta F

reno

s(L

ocks

Boa

rd)

NO

TA.-

Par

a m

odel

os d

e G

ener

ador

Alto

,co

nect

ar e

l Cab

le A

limen

taci

ón a

la re

glet

aTS

1 (+

24 V

DC

aTS

1-31

, 0 V

DC

a T

S1-

32, 2

4V

AC

a T

S1-

19, y

0 V

AC

a T

S1-

20).

NO

TE.-

For T

all G

ener

ator

mod

els,

con

nect

the

Sup

ply

Cab

le to

term

inal

stri

p TS

1 (+

24V

DC

to T

S1-

31, 0

VD

C to

TS

1-32

, 24

VA

C to

TS1-

19, a

nd 0

VA

C to

TS

1-20

)

Man

guer

a C

olim

ador

(Col

limat

or C

able

)

A31

11-X

X

CO

LIM

ADO

R R

alco

(Ral

co C

ollim

ator

)

24 V

0 V

Man

guer

a A

limen

taci

ón F

reno

s(L

ocks

Sup

ply

Cab

le)

A

7622

-XX

CO

LUM

NP

AN

EL

VE

RTI

CA

L C

ON

TRO

LR

OTA

TIO

N C

ON

TRO

L

GN

D

24 V

0 V

Opt

ima

Buc

ky C

able

(

A33

42-0

1)

Buc

ky C

able

( A

3343

-01)

to G

ener

ator

(see

she

et 2

)

BU

CK

YPA

NEL

Buc

ky

Buc

ky P

anel

Cab

le

( A

7655

-01)

Vert

/Rot

Cab

le

( A

7654

-01)

CO

LLIM

ATO

RP

AN

EL

Col

limat

orPa

nel C

able

( A76

53-0

1)

Con

trol

Cab

le( A

7620

-01)

J1

Con

nect

ion

Cab

le

( A67

73-0

1)

Sole

noid

Cab

le

( A76

21-0

1)

+ -

+24

VD

C

RO

T S

OLE

NO

ID

SE

TW

1-B

5430

1083

1/2

01/0

1/04

F.G

AR

CIA

16/0

3/04

A.D

IAZ

RE

VD

ES

CR

IPT

ION

ISS

UE

D B

YD

AT

E

DR

AW

ING

RE

VIS

ED

NA

ME

DA

TE

SH

EE

T /

OF

RE

V

SE

DE

CA

L

+24V

+24V

+24V

+24V

+24V

+24V

CR

71N

4007

CR

81N

4148

C2

100

uF35

V

C1

100

uF35

V

CR

31N

4148

CR

51N

4148

RO

TATI

ON

CR

21N

4007

CR

11N

5408

CR

61N

4148

J2 1 2 3

CR

91N

4148

C3

100

uF35

V

J1 1 2 3 4 5 6 7 8 9 10 11 12

J2

1 2 3 4 5 6 7 8 9 10 11 12

R6,

1K

R7,

1K

R8,

1K

K3

RO

T

51

873

642

TB7 5 6 4 3

CR

11N

4007

SO

LEN

OID

RO

T

CR

41N

4007

CR

21N

4148

K1

LON

G

59 8

1 1612

K2

VE

RT

59 8

1 1612

RO

TATI

ON

VE

RTI

CA

L

K4

TELE

SC

OP

IC 59 8

1 1612

TS2

1 2 3 4

R2

511

TS1

1 2 3 4 5 6

C5

470

uF50

V

TS4

1 23 45

LAM

P

TIM

ER

1 2 34 56

AB

W1

CR

101N

4007

R3

160

2w

SW

R4

511

VE

RT

LOC

K

R1

511

R5

1K+2

4VTP

2

GN

DTP

1

C4

100

uF35

V

VE

RTI

CA

L

A A

B B

C C

D D

E E

44

33

22

11

Aña

dir P

uent

e(A

dd J

umpe

r)

AR

MA

RIO

GEN

ERA

DO

R(G

ENER

ATO

R C

AB

INET

)

MAN

GUE

RA B

UCKY

(BUC

KY C

ABLE

)

(A33

43-0

1)

BU

CKY

1 M

OTI

ON

BU

CKY

1 M

OTI

ON

RTN

BU

CKY

1 D

R

115/

220

VAC

0 VA

C

INTE

RC

ON

EX

ION

ES

OP

TIM

A B

RS

-WH

O(O

PTI

MA

BR

S-W

HO

INTE

RC

ON

NE

CTI

ON

S)

GN

D

Buc

kyCa

ble

Buck

y O

ptim

a(O

ptim

a Bu

cky

Cabl

e)

(A

3342

-01)

OPT

IMA

BR

S

Not

a.- E

ste

plan

o es

apl

icab

le c

uand

o se

util

ize

el b

ucky

y/o

la c

ámar

a io

niza

ción

.N

ote.

- Thi

s dr

awin

g is

app

licab

le w

hen

usin

g th

e bu

cky

and/

or th

e io

n ch

ambe

r.

MA

NG

UER

A C

AM

AR

A IO

NIZ

AC

ION

(IO

N C

HAM

BER

CAB

LE)

(A

3253

-XX)

Shi

eld

IC O

UTP

UT

Con

ecto

r IC

pue

de s

er J

1, J

2, o

J3,

depe

ndie

ndo

de la

Cám

ara

Ioni

zaci

ónse

lecc

iona

da.

(Con

nect

or IC

cou

ld b

e J1

, J2,

or J

3,de

pend

ing

on th

e Io

n C

ham

ber

sele

cted

)

CA

MAR

A IO

NIZ

ACIO

N(IO

N C

HA

MB

ER)

FLD

1 D

R (L

EFT)

FLD

3 D

R (R

IGH

T)

GN

D

FLD

2 D

R (C

ENTE

R)

STR

T D

R (R

ES

ET)

+12

VD

C

Shi

eld

-12

VDC

Tarj

eta

AE

C A

DA

PTA

TIO

N(A

EC

AD

AP

TATI

ON

Boa

rd)

Shi

eld

Usa

r la

Man

guer

a C

ámar

a Io

niza

ción

A32

34-0

2 cu

ando

no

se u

tiliz

a la

tarj

eta

AEC

AD

APT

ATI

ON

.Ve

r esq

uem

a I/F

-003

par

a in

terc

onex

ione

s.(U

se Io

n C

ham

ber C

able

A32

34-0

2 w

hen

the

AEC

AD

APT

ATI

ON

Boa

rd is

not

use

d. S

ee I/

F-00

3sc

hem

atic

for i

nter

conn

ectio

ns)

5430

1083

2/2

01/0

1/04

F.G

AR

CIA

16/0

3/04

A.D

IAZ

RE

VD

ES

CR

IPT

ION

ISS

UE

D B

YD

AT

E

DR

AW

ING

RE

VIS

ED

NA

ME

DA

TE

SH

EE

T /

OF

RE

V

SE

DEC

AL

J1 5 641 32

151 2 6 9873 4 2 3 4 8 6

TS1 5 6 4 1 3 2 7

J5

51 2 6 9873 4

A A

B B

C C

D D

E E

44

33

22

11

BR

S SY

STEM

See

sch

emat

ic 5

4301

083

for i

nter

conn

ectio

ns

BR

S w

ith F

IXED

GR

ID

LOC

KS S

UPP

LY C

ABLE

(A76

22-X

X)TS

1

CO

LLIM

ATO

R C

ABLE

(A31

11-X

X)

GEN

ERA

TOR

CA

BIN

ET

TB7

LOC

KS

BO

AR

D

TB7

CO

LLIM

ATO

R

GEN

ERAT

OR

SU

PPLY

CAB

LE (S

ee N

ote

1)

TUBE

STA

TOR

CAB

LE (S

ee N

ote

2)TS

2X-

RA

YTU

BE

GN

D C

ABLE

(A63

83-X

X)

Cha

ssis

GN

DC

hass

is G

ND

CA

BLE

ASS

EMB

LY

AC L

INE

VA

C S

UP

PLY

Not

e 1.

- Cab

le c

ode:

A841

0-01

for C

ompa

ct G

ener

ator

- si

ngle

pha

seA8

411-

01 fo

r Com

pact

Gen

erat

or -

thre

e ph

ase

A841

2-01

for B

atte

ry G

ener

ator

A841

3-01

for C

apac

itor G

ener

ator

THIS

SH

EET

ON

LY A

PPLI

ES T

O B

RS

with

FIX

ED G

RID

Not

e 2.

- Cab

le c

ode:

A305

2-XX

for o

nly

Low

Spe

edA7

014-

XX fo

r Low

/Hig

h Sp

eed

J1 J3

POW

ER C

ABLE

(A70

21-X

X)

J3

INTE

RFA

CE

CAB

LE (A

7069

-XX)

GEN

ERA

TOR

CO

NSO

LE

ATP

CO

NS

OLE

BO

AR

D

GN

D C

ABLE

TS1/

TS3

CO

MM

UN

ICAT

ION

CAB

LE (A

7022

-XX)

J2

J2

AAN

OD

E C

ABLE

C

HV

TAN

KC

ATH

OD

E C

ABLE

AC

IM -

339

1/3

07/0

7/05

F.G

AR

CIA

07/0

7/05

A.D

IAZ

RE

VD

ES

CR

IPT

ION

ISS

UE

D B

YD

AT

E

DR

AW

ING

RE

VIS

ED

NA

ME

DA

TE

SH

EE

T /

OF

RE

V

SE

DE

CA

L

A A

B B

C C

D D

E E

44

33

22

11

BR

S SY

STEM

J1

GEN

ERA

TOR

CA

BIN

ET

AC L

INE

J2

TS1

BR

S w

ith O

SCIL

LATI

NG

GR

ID

TB7

J3

Cha

ssis

GN

D

X-R

AY

TUB

E

GN

D C

ABLE

(A63

83-X

X)

TS2

J2

CO

LLIM

ATO

R C

ABLE

(A31

11-X

X)

J3

LOC

KS

BO

AR

D

BU

CK

Y

LOC

KS S

UPP

LY C

ABLE

(A76

22-X

X)

J1

TB7

TS1/

TS3

INTE

RFA

CE

CAB

LE (A

7069

-XX)

TS1

BUC

KY C

ABLE

(A33

43-X

X)

GEN

ERA

TOR

CO

NSO

LE

VA

C S

UP

PLY

Cha

ssis

GN

D

POW

ER C

ABLE

(A70

21-X

X)

GN

D C

ABLE

CO

MM

UN

ICAT

ION

CAB

LE (A

7022

-XX)

CO

LLIM

ATO

R

See

sch

emat

ic 5

4301

083

for i

nter

conn

ectio

ns

ATP

CO

NS

OLE

BO

AR

D

CA

BLE

ASS

EMB

LY

Not

e 1.

- Cab

le c

ode:

A841

0-01

for C

ompa

ct G

ener

ator

- si

ngle

pha

seA8

411-

01 fo

r Com

pact

Gen

erat

or -

thre

e ph

ase

A841

2-01

for B

atte

ry G

ener

ator

A841

3-01

for C

apac

itor G

ener

ator

GEN

ERAT

OR

SU

PPLY

CAB

LE (S

ee N

ote

1)

THIS

SH

EET

ON

LY A

PPLI

ES T

O B

RS

with

OSC

ILLA

TIN

G G

RID

Not

e 2.

- Cab

le c

ode:

A305

2-XX

for o

nly

Low

Spe

edA7

014-

XX fo

r Low

/Hig

h Sp

eed

TUBE

STA

TOR

CAB

LE (S

ee N

ote

2)

AAN

OD

E C

ABLE

C

HV

TAN

KC

ATH

OD

E C

ABLE

AC

IM -

339

2/3

07/0

7/05

F.G

AR

CIA

07/0

7/05

A.D

IAZ

RE

VD

ES

CR

IPT

ION

ISS

UE

D B

YD

AT

E

DR

AW

ING

RE

VIS

ED

NA

ME

DA

TE

SH

EE

T /

OF

RE

V

SE

DE

CA

L

A A

B B

C C

D D

E E

44

33

22

11

Cha

ssis

GN

D

BU

CK

Y

CO

LLIM

ATO

RTB

7

LOC

KS

BO

AR

D

AEC

INTE

RFA

CE

CAB

LE (A

3251

-01)

CA

BLE

ASS

EMB

LY

X-R

AY

TUB

E

ATP

CO

NS

OLE

BO

AR

D

GEN

ERA

TOR

CA

BIN

ET

See

sch

emat

ic 5

4301

083

for i

nter

conn

ectio

ns

J1/J

2/J3

GEN

ERA

TOR

CO

NSO

LE

See

sch

emat

ic IM

-010

TB7

Cha

ssis

GN

D

AEC

CAB

LE (A

3253

-XX)

AE

C A

DA

PTA

TIO

N B

OA

RD

TS1

ION

CH

AM

BE

R(P

ream

plifi

er)

TS2

TS1

J3

CO

LLIM

ATO

R C

ABLE

(A31

11-X

X)

BUC

KY C

ABLE

(A33

43-X

X)

J1

GN

D C

ABLE

(A63

83-X

X)

J1

LOC

KS S

UPP

LY C

ABLE

(A76

22-X

X)

J4J5

BR

S w

ith O

SCIL

LATI

NG

GR

ID a

nd A

EC

BR

S SY

STEM

J2 J3

POW

ER C

ABLE

(A70

21-X

X)

CO

MM

UN

ICAT

ION

CAB

LE (A

7022

-XX)

J2IN

TER

FAC

E C

ABLE

(A70

69-X

X)TS

1/TS

3

GN

D C

ABLE

VA

C S

UP

PLY

AC L

INE

THIS

SH

EET

ON

LY A

PPLI

ES T

O B

RS

with

OSC

ILLA

TIN

G G

RID

and

AEC

GEN

ERAT

OR

SU

PPLY

CAB

LE (S

ee N

ote

1)

Not

e 1.

- Cab

le c

ode:

A841

0-01

for C

ompa

ct G

ener

ator

- si

ngle

pha

seA8

411-

01 fo

r Com

pact

Gen

erat

or -

thre

e ph

ase

A841

2-01

for B

atte

ry G

ener

ator

A841

3-01

for C

apac

itor G

ener

ator

TUBE

STA

TOR

CAB

LE (S

ee N

ote

2)

Not

e 2.

- Cab

le c

ode:

A305

2-XX

for o

nly

Low

Spe

edA7

014-

XX fo

r Low

/Hig

h Sp

eed

AAN

OD

E C

ABLE

C

HV

TAN

KC

ATH

OD

E C

ABLE

AC

IM -

339

3/3

07/0

7/05

F.G

AR

CIA

07/0

7/05

A.D

IAZ

RE

VD

ES

CR

IPT

ION

ISS

UE

D B

YD

AT

E

DR

AW

ING

RE

VIS

ED

NA

ME

DA

TE

SH

EE

T /

OF

RE

V

SE

DE

CA

L