sensION+ PH1 Technical data sheet - DelAgua · . sensION+ PH1 Parameters pH mV Temperature...

2
Kit versions are available that include carrying case, accessories and sensor. A wide range of high quality sensors that cover all types of applications, from low ionic strength to viscous, dirty samples. Ergonomic and robust instrument, IP67 waterproof; resistant in harsh working conditions. PH1 highlights with easy operation, high quality and excellent price. The carrying case includes all necessary accessories to ensure accurate measurement. Instrument can be seated at an angle for convenient viewing of the display; sensor and the calibration flasks are held in an upright position for easy calibration and measurement. sensION+ PH1 Technical data sheet Large graphic display Simultaneous display of all the parameters Carrying case with all required elements for work Features and benefits www.hach-lange.co.uk

Transcript of sensION+ PH1 Technical data sheet - DelAgua · . sensION+ PH1 Parameters pH mV Temperature...

Kit versions are available that include carrying case, accessories and sensor.A wide range of high quality sensors that cover all types of applications, from low ionic strength to viscous, dirty samples.

Ergonomic and robust instrument, IP67 waterproof; resistant in harsh working conditions.PH1 highlights with easy operation, high quality and excellent price.

The carrying case includes all necessary accessories to ensure accurate measurement.Instrument can be seated at an angle for convenient viewing of the display; sensor and the calibration flasks are held in an upright position for easy calibration and measurement.

sensION+ PH1 Technical data sheet

Large graphic display Simultaneous displayof all the parameters

Carrying casewith all required elements for work

Features and benefits

www.hach-lange.co.uk

sensION+ PH1

Parameters pH mV TemperatureMeasuring range -2.00…19.99 ±1400-20…150˚C(-4…302˚F) Measuringerror(±1digit) Reproducibility(±1digit)pH <= 0.01 pH ± 0.01mV <= 1 mV ± 1Temperature<=0.2°C(0.4°F) Temperature±0.1°C(0.1°F)

Temperature compensation: Withbuilt-inPt1000temperaturesensor.

pHcalibrationWith1,2or3buffersselectablebetween:2.00,4.01,7.00,9.21and10.01(at25˚C).AutomaticrecognitionofpHvaluesofbuffersatdifferenttemperatures.ManualcalibrationatanypHvalue.Theoreticalcalibration(factorysettings).

mVcalibrationWith1standard,220mV(at25˚C).AutomaticrecognitionofmVvaluesofstandardatdifferenttemperatures.ManualcalibrationatanymVvalue.Theoreticalcalibration(factorysettings).

Temperature readjustment Correctionofthetemperatureprobedeviation(A.T.C.)at25˚Cand85˚C.Canbeusedasthermometerofprecision.

Measuring mode: Stabilityandcontinuous.

Connector: MP-5, 5 contacts multipin.

Display: Liquidcrystal,backlit,withpictograms.

Physical parameters: MaterialABS.DegreeofprotectionIP67.Weight300g.Size186x73x38mm.

AmbientconditionsOperating temperature: 0 °C to 50 °C.Storage temperature: –15 °C to 65 °C.80%relativehumidity(non-condensing).

Directives low voltage and EMC AccordingtoCE,UNE-EN61010-1andUNE-EN61326-1.

Power supply3x1.5V,typeAAbatteries.Over 400 hours operation.

Automaticpoweroffafter5minutesofinactivity(thisoptioncanbedisabledbytheuser).

5050T, for general use.CombinedpH,plasticbody.Withbuilt-intemperaturesensor.0to14pH/0to80˚C.

5051T, for general use.Opendiaphragm,glassbody. Withbuilt-intemperaturesensor.2to14pH/0to80˚C.

5052T, for low conductivity.PTFEjunction,glassbody.Withbuilt-intemperaturesensor.0to14pH/0to100˚C.

5055, Redox.Platinumelectrodes(Pt),plasticbody.±2000mV/0to80˚C.

Productnumber Sample type Description Included accessories

LPV2500.98.0002 PH1, with accessories, without electrode

LPV2550T.98.002 Kit for general use PH1, with accessories & electrode 5050T •Carryingcase•Solutions:pH4,pH7,pH9,KClandRedox220mV•Calibrationtubes•Flasksforsampleandsensorcleaning•Workingprotector(onlywithelectrodes5050T,5051T&5052T).•Usermanual

LPV2551T.98.002 Kit for difficult media PH1, with accessories & electrode 5051T

LPV2552T.98.002 Kit for low conductivity PH1, with accessories & electrode 5052T

LPV2555.98.0002 KitforRedoxmeasure PH1, with accessories & electrode 5055

5051TCode:LZW5051T.97.002

5052TCode:LZW5052T.97.002

5055Code:LZW5055.97.0002

5050T with storage protectorCode:LZW5050T.97.002

Technical data sheet

Specifications

Recommended electrodes

How to order

Hach Lange LtdPacific Way - SalfordManchester - M50 1DL Tel: (0)161 872 1487Fax: (0)161 848 [email protected]

www.hach-lange.co.uk