Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf ·...

90
Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation

Transcript of Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf ·...

Page 1: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Seminar: Sanger Sequencing

04. August 2010Dr. René Gödde

Produktspezialist Laborautomation

Page 2: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Seminar: Sanger Sequencing• Beckman Coulter (Genomics)• Agencourt CleanSEQ kit• Requirements for a good sequencing

result• Troubleshooting Sanger sequencing

Page 3: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Beckman Coulter (Genomics)

Page 4: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Worldwide Organization

• HQ: Orange County, CA• 10,000+ employees• 250,000 placements

in 130 countries• Dedicated to

biomedical testing

Page 5: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

SIMPLIFY • AUTOMATE •INNOVATE

19351935 20102010

Page 6: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Founders developed SPRI nucleic acid purification chemistry at Whitehead which was used to sequence over 1/3rd of the Human Genome, 1998

Agencourt Bioscience incorporated, July 2000

Began offering Sequencing Services, April 2001

Began offering NA purification kits, June 2003

Acquired by Beckman Coulter, June 2005

Acquisition of Cogenics, April 2009

Renamed to Beckman Coulter Genomics, August 2009

Today the world‘s largest commercial provider for genomic services and nucleic acid purification products !

Agencourt History

Page 7: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Beckman Coulter Genomics

Genomic Services Products

• Regulated Services• DNA Sequencing• NextGen Sequencing• Gene Expression• Genotyping• Biologis Testing• Nucleic Acid Extraction• Biorepository• Sample Submission• Service Facilities

• SPRI Laboratory Automation• DNA Purification and Cleanup• Accessories• Software• SPRIworks Fragment Library System I & II• SPRI-TE Nucleic Acid Extractor

Page 8: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Proven ExperienceMammalian Gene Collection (MGC)

cDNA library screening Identified full length clones for all human/mouse genes Submitted over 1,250,000 sequences to GenBank

Human Genome ProjectGenerated > 70 Mb of draft sequence Finished > 46 of Chromosome 10 BACs

Tetraodon nigroviridis Genome ProjectSequenced > 1 billion high quality bases Collaboration with the Whitehead Institute

Mus musculus Genome ProjectSequenced > 1 billion shotgun sequencing bases 180,000 fosmid clones end sequenced

Rattus norvegicus Genome ProjectSequenced >3 billion high-quality bases Performed disease gene analysis Collaboration with the Baylor HGSC

Methanobacterium Genome ProjectFinished 1.75 Mb genome sequence Detailed comparative analysis and annotation

Ciona savignya Genome ProjectSequenced 160,000 fosmid clones Collaboration with the Whitehead Institute

Cancer Genome Anatomy Project (CGAP)SAGE sequencing Determined gene expression profiles for > 20 distinct tumor types

Rickettsia sibirica Genome ProjectCreated high and low copy genome libraries Sequenced complete genome to 12X coverage Initiated protein-protein interaction studies

Clostridium Genome ProjectFinished 4.1 Mb genome sequence Detailed comparative analysis and annotation

Felis catus Genome ProjectSequenced >4 billion high-quality basesWhole genome assembly of 5 strains

Chrysosporium lucknowense Genome ProjectSequenced to 6X coverage and assembledUtilized by Dyadic for strain improvement

Page 9: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

LiquidLiquidHandlerHandler

ProcessProcessWorkstationWorkstation

AssayAssayWorkstationWorkstation

IntegratedIntegratedSystemSystem

The Biomek® Continuum:Flexible and Upgradeable Automation

Page 10: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Automated dye removal at the CCG

Biomek FXp DualHybrid Target Prep Express

Page 11: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Agencourt CleanSEQ kit

Page 12: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

SPRI Technology• SPRI - Solid Phase Reverse

Immobilization

• Patented nucleic acid isolation process - Developed by Whitehead and exclusively licensed to Agencourt

• Paramagnetic bead-based - Uses Seradyn’s Sera-Mag™ microparticles

• Reversible immobilization of DNA on bead surface – Based on buffer conditions

- nucleic acids bound in presence of proprietary buffer

- no chaotrophic salts required- DNA eluted in aqueous solution

Page 13: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

COO-

COO-

-

COO-

How Does SPRI Work?

Page 14: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

COO-

COO-

-

COO-

EtOH

H2O

H2OH2O

H2O

H2O

H2O

H2O

H2O

EtOH

EtOH

How Does SPRI Work?

Page 15: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Typical SPRI Protocol

Step 1: Nucleic Acid Immobilization

Step 2: Contaminant Removal & Wash

Step 3: Nucleic Acid Elution

Page 16: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

CleanSEQ Competitive Quality

pGEM® sequenced using M13 primers and 1/8th dilution BigDye™ v3.0 and sequenced on an ABI PRISM® 3700 Sequencer

Page 17: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

CleanSEQ vs. Ethanol

Lower % Failure

Higher Average Phred20

Lower Standard Deviation

Phred 20 Read Length

Per

cent

age

of R

eads

~1,500 EtOH readsx

~1,500 CleanSEQ reads

Page 18: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Requirements for a good sequencing result

Page 19: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

How does Sanger sequencing work?

Page 20: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Template DNA *• PCR• Plasmide• Fosmide/Cosmide• BAC

Primer *

dNTPs *

ddNTPs *Sequenase *

Buffer salts *

What‘s inside a sequencing reaction?

Water

* specific for each reaction* contained within BigDye kit

It‘s all about balance !

Additives• DMSO• Biotine• BDT buffer

Page 21: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Component requirements:• Template DNA

Concentration- Gel- Nanodrop- Picogreen- Bioanalyzer

Purity- proper purification methods

Specifity- no multiple templates

• PrimerConcentration

- dilute lyophilized primers properlyHPLC purified

- non-HPLC may lead to n-1 artifactsDesign

- use Primer 3 software or comparable

• AdditivesConcentrationsOnly use when necessary (except DMSO)

Page 22: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Typical reaction setups:

ABI recommended Agencourt recommended

BigDye 8 µl 0.25 to 0.5 µl(1-2 µl for large/difficult templates)

Template DNA 50 ng to 1 µg 50 ng to 1 µg

Primer 1.6 to 3.2 pmol 1.6 to 5 pmol(probably 10 pmol for BACs)

BDT buffer - 1x

Water d.s. d.s.

Total 20 µl 10 µl

Page 23: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Considerations when diluting BigDye

• When diluting BigDye, you‘re also diluting enzyme, salts, dNTPs and ddNTPs !

• May lead to disbalance in sequencing reaction !

• Fix by adding sequencing buffer and/or dNTPs !(see CCG web page for recipes)

Page 24: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Typical reaction setups:

ABI recommended Agencourt recommended

BigDye 8 µl 0.25 to 0.5 µl(1-2 µl for large/difficult templates)

Template DNA 50 ng to 1 µg 50 ng to 1 µg

Primer 1.6 to 3.2 pmol 1.6 to 5 pmol(probably 10 pmol for BACs)

BDT buffer - 1x

Water d.s. d.s.

Total 20 µl 10 µl

Page 25: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Regarding template amounts

Concentration [ng/µl]

Number of molecules [mol]

Page 26: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Why?

100ng/µl of 0.5 kb PCR product

100ng/µl of 7 kb PCR product

Page 27: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Problem: [mol] hard to determine

• [mol] is sometimes hard to determine and prone to calculation failures

• Fix by using concentration with following formula:

ng Template = 100ng x No. of kilobases

• Example: 0.5 kb PCR product

100ng x 0.5 = 50ng of DNA template per rxn

Page 28: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Typical reaction setups:

ABI recommended Agencourt recommended

BigDye 8 µl 0.25 to 0.5 µl(1-2 µl for large/difficult templates)

Template DNA 50 ng to 1 µg 50 ng to 1 µg

Primer 1.6 to 3.2 pmol 1.6 to 5 pmol(probably 10 pmol for BACs)

BDT buffer - 1x

Water d.s. d.s.

Total 20 µl 10 µl

Page 29: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Troubleshooting Sanger sequencing

Page 30: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Ideas and tools• Identify sequencing problems by looking at raw data, not analyzed data

!

• Use trace viewer such as FinchTV !

• Use internet resources to indetify problems !

• In 99% of all instances the problem is IN the sequencing reaction !

Page 31: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Failed DNA sequencing reaction:

How to identify • The trace chromatogram has noisy or “messy” sequence peaks with low quality scores.

• No or little signal in the raw data channels except for leftover dye at the beginning of the trace.

• Signal strength in the raw channel usually below 100.

• The trace sequence does not match either the expected sequence or other sequences in GenBank.

Page 32: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Failed DNA sequencing reaction:

Traces

Figure 1. Example of a failed reaction trace. (A) Processed channel data.(B) Raw channel data at the same base location.

Page 33: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Failed DNA sequencing reaction:

Causes • Poor quality DNA. Very common when sequencing plasmid miniprep templates.

• Loss of the reaction during clean up. This can be a particular problem when using ethanol precipitation clean up protocols.

• Too much template DNA. Excess template can kill the sequencing reaction.

• Wrong primer used. More common than you might think!

• Bad water. The water used contains a sequencing inhibitor.

• Degraded or failed synthesis primer. Oligonucleotide synthesis is chemically complex and primer synthesis failures are fairly common.

• Dead sequencing chemistry. Can occur if the BigDye chemistry is stored under the wrong conditions or is freeze-thawed too many times. Either the Taq DNA polymerase or dye labeled nucleotides can be degraded.

• Blocked capillary. Every trace using a particular capillary fails. Can be identified by tracking trace quality on a trace by trace basis.

Page 34: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Failed DNA sequencing reaction:

Troubleshooting • Poor quality DNA. The best way of avoiding this problem is to not sequence plasmid DNA

and sequence a PCR amplified fragment of the plasmid insert. If this is not possible then it is recommended that a plasmid miniprep kit is used. One tip is to perform a final ethanol precipitation on the kit purified plasmid DNA. This often solves problems with the quality of the template. “Dirty cleanups” like ExoSAP digestions may also cause purity problems.

• Loss of sequencing reaction during clean up. This can be avoided by not using an ethanol precipitation protocol to clean up the sequencing reaction.

• Too much template DNA was used. This can be avoid by checking the concentration of the template before sequencing. This may also allow you to see the purity of the template DNA and if there is a significant amount of contaminating genomic DNA or RNA present. Do not rely on a spectrophotometer reading to calculate the template concentration!

• Wrong primer. This is simple to solve, but can be difficult to detect. Check the sequence of the primer and template to make sure that the primer binding site is present. This can be a particular problem with some "universal" forward and reverse primer sequences which do not work with some common plasmids. Do not trust other people's working stock solutions (especially those of your unreliable lab colleagues!) and make your own. It might take 5 minutes longer, but it will save you a lot of future headaches.

Page 35: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Failed DNA sequencing reaction:

Troubleshooting • Bad water. Inhibitors can end up in lab water stocks that can kill DNA sequencing reactions.

If you think this may be a problem then throw out the water and use a fresh stock - remember water is cheap.

• Degraded primer. Don't use old diluted primer stocks. Store the primers in 10mM Tris/ 0.1 mM EDTA (pH 8.5) rather than water. Don't use other peoples stocks. If you have any doubt about how the primer quality is discard it and make up a fresh working solution from the primer stock.

• Failed oligonucleotide synthesis. If you suspect that the primer is poor quality check in a polymerase chain reaction (PCR). Alternatively if you have a control template that you know should work with the primer then this can be a good way of identifying primer problems.

• Dead sequencing chemistry. This is a relatively rare problem, however, if a batch of BigDye chemistry has not been used for some time, or there is any doubt about how it has been stored, then it is advisable to perform a control sequencing reaction before undertaking a large number of experimental reactions. Many problems with dead chemistry can be prevented by storing the BigDye chemistry in small aliquots and avoiding repeated freeze/thaw cycles.

• Blocked capillary. Can be identified by tracking trace quality on a trace by trace basis. While this can be done manually we recommend using our QualTrace software. If you suspect that a sequencer capillary is blocked then inform the operator.

Page 36: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Noisy sequence due to overload signal :

How to identify

• The sequencing trace shows “noisy” signal, preferably at the beginning of the trace.

• Raw data signals are saturated at approx. 32000 units.

Page 37: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Noisy sequence due to overload signal :

Traces

Figure 1. Example of a trace with overload signal. (A) Processed data channel. (B) Raw channel data.

A B

Page 38: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Noisy sequence due to overload signal :

Causes

• Too much yield from the sequencing reaction and sequencing cleanup.

• Injection times are too long.

Page 39: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Noisy sequence due to overload signal :

Troubleshooting

• Troubleshooting overloaded signals is easiest achieved by reducing the number of cycles in the cycle sequencing reaction or by adding additional EDTA in the eluted sequence.

• Injection times may be reduced in a re-run of the same samples.

Page 40: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

“n-1” primers:

How to identify

• The sequence shows a pre-peak to every peak of the same colour of the following peak.

Page 41: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

“n-1” primers:

Traces

Figure 1: Tracefile showing n-1 peaks

Page 42: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

“n-1” primers:

Causes

• „n-1“ primers are present in the sequencing primer, consisting of primers that are one base shorter than the original primer.

• “n-1” primers are a common artifact of primer synthesis, but normally do not exceed 1% of all synthesized primer.

• If the number of „n-1“ primers represent more than 20-30% of the overall primers, this will result in measurable numbers of „n-1“ cycle sequencing products.

Page 43: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

“n-1” primers:

Troubleshooting

• Change your oligo vendor or order only HPLC purified primers (recommended).

Page 44: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Mixed DNA templates :

How to identify • The sequencing trace has two or more sequence peaks at the same location. The secondary

peaks may be the same height as the primary peaks down to about 20%. Lower levels of mixed signal (below 20%) are normally base called well by the KB.

• The trace signal strength is good (>200U) in the raw channel.

• The quality scores are low with often less than 100 Q20+ bases.

• The base called sequence does not match the expected sequence or any known sequence in GenBank.

Page 45: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Mixed DNA templates :

Traces

Figure 1. Example of a trace with mixed template. (A) Raw channel data. (B) Processed data channel at the same base location.

Page 46: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Mixed DNA templates :

Causes • Two or more templates were present in the reaction. This is the most

common cause of mixed template.

• A "double pick" of two colonies. This can occur when the colonies are too close together on the colony plate.

• Two primers were present in the sequencing reaction. This can occur when using premixed PCR reagents for sequencing where the primer stock is actually a mix of universal forward and reverse oligonucleotide primers.

• The PCR fragment was not purified from PCR primes before sequencing.

• Two priming sites are present in DNA template. This can occur when a PCR product with universal priming tails is cloned into a plasmid or if there are sequence similarities in the templates.

• Poor quality PCR template containing multiple DNA fragments was used.

• Too low primer annealing temperature was used in the sequencing reaction.

Page 47: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Mixed DNA templates :

Troubleshooting • Makes sure only one DNA template is present. Prepare a new plasmid prep making sure

that only one colony is selected. If sequencing a PCR product check that only one PCR product is present by running an agarose gel. Remember that even a relatively low amount of a small PCR product can cause mixed template problems and that gels cannot determine between fragents that are at least 10bp apart.

• Check the template for possible multiple priming sites. If two sites are present use a different primer. This can often occur when a fragment that contains the priming site is sub-cloned into a vector that also contains the priming site. Be aware of this problem when using the M13 universal primers.

• Insure that you use a PCR clean-up protocol that remove leftover PCR primers. Even low levels of the PCR primers can cause mixed signal problems, especially if they have a high annealing temperature.

• Check the predicted melting temperature of the sequencing primer. If it is more than 5˚C above the annealing temperature used in the sequencing reaction then raise the annealing temperature. Because of the inclusion dITP in the BigDye sequencing mix the annealing/extension temperature can't be raised above 60˚C. If your primer is still misanneals at 60˚C then synthesize a new primer (this is easily done by removing bases from the 5' end until the Tm is below 60˚C).

Page 48: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Short DNA Trace Read Lengths :

How to identify • Sequence provide less than 650 Q20+ bases for traces run on 36cm arrays and less than

750 Q20+ bases for traces collected on 50cm arrays under standard run conditions.

• End of the trace is "messy" with "rough" looking peaks

• Trace signal ends abruptly before base 800 on 36cm array traces and before base 1000 on 50cm array traces.

Page 49: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Short DNA Trace Read Lengths :

Traces

Figure 1. Example of a short/poor trace. (A) KB processed data channel data at approximately base 550. (B) Raw channel data at the same base location.

Page 50: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Short DNA Trace Read Lengths:

Causes • Too much or too little amount of template DNA

• Excessive dilution of the BigDye reagent

• Too much or too little amount of primer

• "Dirty" template DNA has been used

• Unsequencable region reached (eg. homopolymer G)

Page 51: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Short DNA Trace Read Lengths:

Troubleshooting • Check the concentration of the template DNA by gel electrophoresis. Do not rely on

spectrophotometer readings alone as spec reading are often inaccurate.

• Check to see if an unsequencable region has been reached. If an unsequencable or “hard stop” region is the cause then it may be possible to sequence though it by PCR amplifying the region using 7-deaza-deoxy guanosine triphosphate (7-deaza-dGTP), then sequencing the PCR product directly. The 7-deaza-dGTP analog disrupts the hoogsteen base pairing between successive guanosine bases and allows the sequencing DNA polymerase to walk through the high G+C region.

• Check that the oligonucleotide primer concentration is correct. Do not rely on old primer stocks!

• Check if the template is clean. Consider sequencing a PCR amplified insert or switching to using a commercial plasmid miniprep kit if you are finding that many of your sequencing reactions are failing.

Page 52: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with poorly resolved or "blurry" trace chromatogram peaks:

How to identify• The trace peaks are very broad or "blurry". This blurriness occurs before base 500 for trace

collected using 36cm arrays, or before base 650 for traces collected on 50cm arrays.

• The nucleotide runs in the sequence appear as indistinct blobs rather than as individually separate peaks.

Page 53: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with poorly resolved or "blurry" trace chromatogram peaks:

Traces

Figure 1. Example of a "blurry" or poorly resolved trace chromatogram collected using a ABI 3730 50cm array.

Page 54: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with poorly resolved or "blurry" trace chromatogram peaks:

Causes

• Capillary overload. This is normally caused by running "dirty" samples with large amounts of template (or other) DNA, proteins or salt.

• High sequencing run voltages. The higher the run voltage the worse the resolution of the peaks in the trace file.

Page 55: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with poorly resolved or "blurry" trace chromatogram peaks:

Troubleshooting

• Use clean DNA. Avoid using templates that contain a significant amount of Genomic DNA, protein or salt.

• Check the concentration of the template. If the template concentration was too high then resequence using less template.

• Check that the correct run voltage was used. This can be determined by examining the run protocol contained with the ABI trace file.

Page 56: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing problems caused by excess free dye terminators:

How to identify

• The trace has one or more large, broad peaks between bases 50 and 140.

• Each large peak color appears at different base positions as separate peaks.

Page 57: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing problems caused by excess free dye terminators:

Traces

Figure 1. Example of a trace with a moderate dye blob problem. Note that the base call is incorrect at approximately base 65.

Figure 2. Example of a minor dye blob problem. Note that the base call is not compromised by this small dye peak.

Page 58: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing problems caused by excess free dye terminators:

Causes

• Poor post-sequencing clean up. Does most frequently occur with EtOH precipitation and sephadex purifications. Not a problem with Agencourt CleanSEQ.

• Lost pellet. When the DNA pellet is lost during EtOH precipitation the remaining small amount of Big Dye can appear as large dye blob peaks.

• Failed reaction. When a reaction works inefficient or fails completely, the amount of leftover dye that was not incorporated during cycle sequencing can appear as large dye blob peaks.

• Note that with Agencourt CleanSEQ dye blobs commonly indicate that the sequencing reaction ran inefficiently!

Page 59: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing problems caused by excess free dye terminators:

Troubleshooting

• Check the ethanol. 100% ethanol absorbs water from the air.

• Check to see if the sequencing reaction has failed by examining the raw trace data using software like FinchTV. If the reaction ran inefficient or failed then the dye blob problem is mostly likely only an artifact of having no or very little trace signal present.

Page 60: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with noisy or weak signal :

How to identify

• The trace peaks are misshapen and/or contain secondary peaks under the primary peak.

• The sequencing peaks are very weak (under 150 units) in the raw channel before base 300.

• The total quality scores counts are generally low.

Page 61: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with noisy or weak signal :

Traces

Figure 1. Example of a noisy or weak trace. (A) Processed data channel.(B) Raw channel data. Note the misshapen secondary peaks in the processed channel and the very low signal strength (~100 units) in the raw channel.

Page 62: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with noisy or weak signal :

Causes

• Partially failed sequencing reaction.

• Too much or too little DNA.

• Any of the other causes listed in the “failed DNA sequencing reaction” chapter.

Page 63: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with noisy or weak signal :

Troubleshooting

• The solutions for solving weak signal traces are basically the same as those for solving failed reactions. For help on this problem see the solutions listed “failed DNA sequencing reaction”chapter.

Page 64: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA Sequencing Traces With Primer Dimers :

How to identify

• The trace signal is mixed in the early regions (normally before base 400) yet the late regions are not mixed.

• The signal drops from very high early in the trace (usually in the first 100 bases) to much lower levels. Examination of the raw signal shows a high signal "block“.

• The total quality scores counts are generally low to moderate depending on the region affected.

Page 65: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA Sequencing Traces With Primer Dimers :

Traces

Figure 1. DNA Sequencing Traces With Primer Dimers . Processed data channel. Note that the sequence is mixed up to base 80, then continues asa normal sequence.

Page 66: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA Sequencing Traces With Primer Dimers :

Causes

• Contamination of the template, primer stock or other sequencing reagents with primer dimers.

• Too low annealing temperature.

• The sequencing primer is self-annealing.

• Sequencing TempliPhi templates. For reasons unknown TempliPhi templates are much more prone to primer dimer problems.

Page 67: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA Sequencing Traces With Primer Dimers :

Troubleshooting

• Change all the sequencing reagents and water for a fresh batch. Also clean the pipettors as they can also be the cause of the primer dimer contamination.

• Raise the annealing temperature and/or use a shorter annealing time.

• For primer design, use a software that checks for self-annealing (like Primer 3).

Page 68: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sharp peaks in the trace signal :

How to identify

• The trace file contains sharp spikes (from 1 to 9 data points wide) in the raw data channel that can have intensities as high as 32,000 (detector limit).

• The sharp peaks are normally higher than the surrounding sequence peaks.

• The sharp peaks are generally more common towards the end of the trace than in the beginning but they can occur anywhere in the trace.

Page 69: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sharp peaks in the trace signal :

Traces

Figure 1. Example of sharp signal spikes. (A) KB processed data channel data. (B) Raw channel data at the same location. (C) Large signal spike (raw channel). Note that the signal reaches an intensity over 32,000.

Page 70: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sharp peaks in the trace signal :

Causes

• The cause of sharp signal spikes is pretty much a mystery. The current best theory is that these peaks are caused by small gas bubbles forming in the capillaries during electrophoreses which refract the laser light into the CCD camera when they pass by the detector zone. The frequencies of sharp spikes seems to be operator and machine dependent.

Troubleshooting • The best solution is to purge buffers and load fresh matrix. Avoiding excessive shaking of the

buffers may also help.

Page 71: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Multinucleotide repeat sequence slippage sequencing problems :

How to identify

• The trace data becomes mixed after a long mononucleotide (single base) run in the DNA template.

• Can also occur on long dinucleotide repeats in a manner analogous to stutter bands that occur in microsatellite polymerase chain reaction (PCR) amplification.

Page 72: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Multinucleotide repeat sequence slippage sequencing problems :

Traces

Figure 1. Sequence slippage on a mononucleotide T run.

Figure 2. Sequence slippage on a mononucleotide C run.

Page 73: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Multinucleotide repeat sequence slippage sequencing problems :

Causes

• Long runs of a mono nucleotide base causes the DNA polymerase to "slip" on the template. This occurs by either the template or extension product looping out and rehybridizing and resulting in the generation of sequencing products of varying size. These sequencing products then appear as mixed signal in the trace downstream of the mono nucleotide run.

• Insertion and deletions in one allele. This can occur when two alleles are sequenced together (PCR templates) and one of the two mononucleotide runs has an InDel polymorphism.

Page 74: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Multinucleotide repeat sequence slippage sequencing problems :

Troubleshooting

• Sequence the DNA template from both directions. For example, if the template shows slippage after using the forward primer then sequence the template from the reverse direction. You may need to make a custom primer to do this if the template is large.

• Use a custom primer designed to hybridize just outside the mononucleotide or dinucleotide run region.

• Use the sequence-by-mutagenesis (SAM) approach to avoid have long mononucleotide runs in your templates.

Page 75: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sudden ending of DNA sequence signal (hard stops) :

How to identify

• Good quality trace signal suddenly stops, or rapidly declines, in region of the trace that should be well resolved.

• The trace peak signal falls to under 150 units (on ABI 3730 instruments) in the raw channel after the stop region.

• The average quality scores after the "hard stop" region are generally low.

• Template has a high percentage of G and/or C nucleotides, especially in the region where the stop occured.

Page 76: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sudden ending of DNA sequence signal (hard stops) :

Traces Figure 1. Example of a sudden "hard stop" at a region of strong template secondary structure. (A) Processed data channel. (B) Raw channel data.

Page 77: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sudden ending of DNA sequence signal (hard stops) :

Causes

• Regions of strong secondary structure in the DNA template. These region can fold back on themselves and form hairpin structures that the sequencing polymerase can not pass through.

• Long runs of G or C nucleotides in the template. These can not only form hairpins with strong secondary structure, but can also form single stranded conformational structures that the the sequencing polymerase has problems passing through. Long template regions of guanidine (G runs) are particular problematic because of the substitution of dITP for dGTP in the BigDye sequencing mix. The BigDye sequencing DNA polymerase does not efficiently extend runs of multiple inosine residues causing the polymerase to stop in the G run regions.

Page 78: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Sudden ending of DNA sequence signal (hard stops) :

Troubleshooting

• Use a 1:4 ratio mix of BigDye to dGTP Sequencing premix. This mix will help for high G+C templates.

• Add a small amount of dGTP nucleotide (~40µM) to the BigDye mix. This helps the sequencing polymerase pass through long G runs without causing major compression problems.

• PCR amplify the template using 7-deaza-GTP or dITP to reduce the strength of the secondary structure. This allows the sequencing polymerase to be able to pass through the region causing the hard stop.

• Use the sequence-by-mutagenesis (SAM) approach to avoid to have long mononucleotide runs in your templates.

Page 79: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with indels :

How to identify

• The trace trace is excellent before the indel region and mixed after the indel region.

• The trace often has leading or trailing 'shifted' peaks where a small peak is found before or after a large peak.

Page 80: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with indels :

Traces

Figure 1. Example of indel trace. Note the mixed sequencing after the seven T run.

Figure 2. Example of 'shifted' peaks due to a template indel. Note the small G peak preceding the large single G peak present after the nine T run.

Page 81: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with indels :

Causes

• Direct sequencing of PCR products derived from polymorphic templates. The most common cause of this is when directly sequencing PCR products amplified from diploid templates with heterozygous regions.

• Random mutation has occurred during the cloning process or during plasmid mini prep process (usually during the colony picking stage). This is much more likely to occur if the cloned template is toxic or unstable in the E. coli host.

Page 82: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

DNA sequencing traces with indels :

Troubleshooting

• Clone the PCR products before sequencing if the template is heterozygous.

• If the template is unstable in E. coli try cloning into a low copy vector or using a different strain.

• Sequence from the other direction using a reverse primer. This option is more viable for small inserts.

Page 83: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces containing chimeras and sequence rearrangements :

How to identify

• The trace signal is high quality up to a point where it becomes mixed. Sometimes the mixed signals appear to be related to each other in a way that the peaks present are 'echoed' latter in the trace.

• The trace sequence data aligns to the expected data up to a point and then shows no further homology.

Page 84: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces containing chimeras and sequence rearrangements :

Traces

Figure 1. Example of a chimeric trace.

Page 85: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces containing chimeras and sequence rearrangements :

Causes

• Cloning artifacts. Chimeras can result from cloning two or more unrelated DNA fragments at once in the same vector.

• 'Unstable' sequences like repeats or long mononucleotide runs. This is a common problem when cloning cDNA sequences as the long poly A tails can be quite unstable in many vectors or E. coli strains.

• 'Toxic' sequences. This can occur if the gene cloned is expressed in E. coli and is toxic to the cell. This tends to occur more commonly in high copy number vectors and templates with low a G+C ratio.

• Strong secondary structure. Deletion can occur in regions of strong secondary structure. This is commonly seen when PCR products are cloned as the PCR process selects for these deletions artifacts.

Page 86: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces containing chimeras and sequence rearrangements :

Troubleshooting

• Isolate and sequence another clone. This can solve the problem if the particular clone you have selected has undergone a rare mutation event. Sometime going back to the original plate and reelecting the same colony can solve the problem as the mutation may have occurred in liquid culture.

• Use a low copy number vector or grow the cells at 30˚C. This can help maintain the stability of unstable or toxic sequences by reducing the selective pressure for mutation or rearrangement.

• Don't over-grow your cultures. The long the cells are grown for the higher the selection pressure for a deletion or rearrangement in the desired sequence.

• Try cloning using a different E. coli strain. This can sometime work especially if you original used a non-cloning strain such as BL21.

• Try adding some glucose to the growth media to inhibit expression of the LacZ promoter. This can help if the instability is being caused by leaky expression of the lacZ operon.

Page 87: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with delayed sequencing signal start :

How to identify

• The first signal peak in the raw channel is found after scan 3200 for 50cm traces and scan 2700 for 36cm traces. Traces normally start between scan 1800 and 2200 for 50cm traces and scan 1500 and 1900 for 36cm traces.

• The signal in the processed channels is often of low resolution with wide peaks.

Page 88: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with delayed sequencing signal start :

Traces

Figure 1. Example of a good (A) and a delayed start (B) 50cm trace. The raw channel is shown.

Page 89: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

Traces with delayed sequencing signal start :

Causes

• Capillary overload. Unlike a lot of DNA sequencing problems delayed peak signal has only one cause - capillary overload. The two most common causes of capillary overload are too much template DNA or dirty template DNA contaminated with proteins an/or salt.

Troubleshooting

• Use less template DNA. Modern ABI capillary sequencers require far less template DNA than the old slab based sequencers. However, many researchers are still using protocols developed for the slab gel based system.

• Use cleaner DNA. Try changing or at least improving your template preparation protocol.

Page 90: Seminar: Sanger Sequencing - CCG | Homeportal.ccg.uni-koeln.de/.../Sequencing-WS...030810.pdf · Seminar: Sanger Sequencing 04. August 2010 Dr. René Gödde Produktspezialist Laborautomation.

THX!