Semi-invasive biopotential & activity in rodents.
-
Upload
lexus-tooke -
Category
Documents
-
view
222 -
download
1
Transcript of Semi-invasive biopotential & activity in rodents.
on each subject the following 8 signals are available:
2 eeg, each sampled at 1kHz
2 emg/ecg, each sampled at 500Hz
3 axis acceleration, sampled at 80Hz on each axis
1 global acceleration module available at 80Hz, computed from 3 axes
rodentPACK is a semi-invasive telemetry device to monitor biopotentials and activity on rats
rodentPACK consists of 3 main parts
electrodes implanted in the subject and linked to a connector permanently glued onto the skull
transmitter housing amplifiers, transmission electronics and batteries, to be coupled to the head connector prior to experiment
receiver to read transmistted data and to provide calibrated analog outputs
radio transmission range, 5m or more
transmission duration with same batteries: 70 hours(cost of batteries: 3€ ($4US) / pair)
group housing: multiple animals in same cage
20 to 40 animals in same room with no required shielding
analog outputs: connects at no extra cost to any existing software
features
noise free, high quality eeg signaleeg electrodes are rigidly welded on the head connector which is itself solidly linked to the transmitter, therefore animal movements trigger no flexible wire movement and cause no artefact
3-axis acceleration allowing activity and posture assessment
very low recurring costno transmitter refurbishment is ever necessary batteries cost 3€ per recording
advantages
low investment costto record simultaneously from 4 rats, your complete hardware investment is less than 13k€ ($18kUS)
even lower investment costby scheduling your recordings, you may further decrease cost
for example, conducting a study on 16 subjects split in 4 groups of 4, only requires a total hardware investment of 14k€ ($20kUS)
open system if you already own a preferred acquisition/analysis software, you may use it at no extra cost : rodentPACK is provided with analog outputs and immediately connects to any recording platform
more advantages
does the device alter subject behaviour?
after an acclimation of under 24 hours, the rats ignore the device. They do not « touch » it nor do they try to rub it on the cage sides
how long can the implanted electrodes and glued connector stay in place ?
rats equipped with this types of electrodes and connectors have been routinely used for many years in tethered systems
rats used for our tests were implanted 12 months ago and are still in good health. No special precautions were taken during this period to keep the skin-to-glue junction clean or sterile
frequently asked questions
power in up to 6 adjustable frequency ranges
absolute and relative power values
versatile scheduling of analysis, epoch lengths, overlap factor, windowing
power spectrum analysis with eeg module within ecgAUTO software
3-D display of average spectrums over time
power spectrum analysis with eeg module within ecgAUTO software
sleep scoring moduledefine threshold values applicable to any parameter from any input
software applies logical rules to attribute 1 out of 4 sleep stages to each epoch
trend graph displays parameters as well as sleep stages
user may, on the fly, adjust thresholds and see the effect on sleep scoring
from top:cortexhippocampusmuscle
sleep scoreactive wakequiet wakeREM sleepSW sleep
sleep scoring module
visit our websites for complete product descriptions, case studies, a knowledge base, & frequently asked questions:
www.emka.fr
website