SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63...

41
SELECTOMAT SL Service instructions Steam sterilizer SELECTOMAT SL_ns_en 0911_V2.02

Transcript of SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63...

Page 1: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

SELECTOMAT SL

Service instructionsSteam sterilizer

SELECTOMAT SL_ns_en 0911_V2.02

SELECTOMAT SL

GebrauchsanweisungDampfsterilisator

SELECTOMAT SL_np_de 0703_V2.01

Page 2: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless
Page 3: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

CONTENTS:6. 11. 2009

1 INTRODUCTION......................................................................................................................................3

2 DESCRIPTION OF THE UNIT...........................................................................................................................3

2.1 STERILIZER CHAMBER........................................................................................................................................32.2 STEAM GENERATOR...........................................................................................................................................32.3 PIPEWORK.............................................................................................................................................3

2.3.1 SCHEME OF PIPEWORK CONNECTION - APPLIED ELEMENTS................................................................................32.3.2 STEAM TRAPS............................................................................................................................................3

2.4 COVERAGE.............................................................................................................................................42.5 ELECTRIC CONNECTION......................................................................................................................................4

2.5.1 BLOCK DIAGRAM........................................................................................................................................42.5.2 DESCRIPTION OF THE ELECTRIC CONNECTION DIAGRAM.....................................................................................4

2.5.2.1 NETWORK PART...............................................................................................................................42.5.2.2 SWITCHBOARD............................................................................................................................42.5.2.3 FRONT PANEL (ON THE LOADING SIDE)..................................................................................................52.5.2.4 REAR PANEL (ON THE UNLOADING SIDE)..............................................................................................52.5.2.5 INPUTS........................................................................................................................................52.5.2.6 OUTPUTS...........................................................................................................................62.5.2.7 HARDWARE PROTECTION....................................................................................................................6

2.6 DESCRIPTION OF THE AUTOMATICS FROM THE VIEWPOINT OF SOFTWARE....................................................................62.7 LOADING OF BIOS, APLICATIONS AND LANGUACES..................................................................................................7

2.7.1 LOADING OF BIOS.......................................................................................................................................72.7.2 LOADING OF THE APPLICATION.......................................................................................................................82.7.3 RULES FOR WORK WITH THE PROGRAM LOADER.EXE........................................................................................9

3 FUNCTION OF THE UNIT....................................................................................................................................9

3.1 PIPING DIAGRAM..............................................................................................................................................9

4 ADJUSTENT OF THE UNIT, SETTING THE PARAMETERS........................................................................10

4.1 MENU - SETTING AND SERVICES ACCESSIBLE FOR THE UTILITY TECHNICIANS.............................................................104.1.1 TEST OF SAFETY ELEMENTS.....................................................................................................................104.1.2 DISPLAY OF BLOCKS................................................................................................................................114.1.3 SUMMARY OF INPUTS - OUTPUTS...............................................................................................................114.1.4 DEVICE CONFIGURATION............................................................................................................................124.1.5 SETTING OF THE DEVICE............................................................................................................................12

4.1.5.1 SERIAL NUMBER AND CHARGES COUNTER.........................................................................................124.1.5.2 OFFSET OF PRESSURE...................................................................................................................124.1.5.3 OFFSET OF TEMPERATURES..............................................................................................................124.1.5.4 SETTING THE DOORS.....................................................................................................................134.1.5.5 UNITS.............................................................................................................................134.1.5.6 INTERNAL GENERATOR......................................................................................................................134.1.5.7 WATER SAVING TEMPERATURES.......................................................................................................134.1.5.8 ADMINISTRATOR - PRESET (ACCESS SAFETY AND DOOR BLOCKING PASSWORD)............................................134.1.5.9 SERVICE PLANNING (NEXT MAINTENNANCE)..........................................................................................13

4.1.6 DOWNLOAD OF PROGRAMS........................................................................................................................144.1.7 PROGRAMMES JONING..............................................................................................................................144.1.8 SYSTEM VERSION.....................................................................................................................................154.1.9 LCD PANEL OFFSET.................................................................................................................................15

4.2 REPLACEMENT OF THE PARTS OF ELECTRONICS....................................................................................................154.2.1 REPLACEMENT OF THE ELECTRONICS BOARDS................................................................................................154.2.2 BATTERY REPLACEMENT............................................................................................................................164.2.3 REPLACEMENT OF THE CONTACT PANEL - FRONT PANEL...................................................................................16

Page 4: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

4.2.4 REPLACEMENT OF THE TOUCH PANEL AND DISPLAY - REAR PANEL.......................................................................164.3 DEFINITION OF THE PROGRAM PARAMETERS........................................................................................................16

4.3.1 THE WAY OF DEFINITIO OF THE PROGRAMME PARAMETERS.............................................................................164.4 DESCRIPTION OF THE STERILIZATION PROGRAMME PARAMETERS.......................................................................17

4.4.1 SENSE OF INDIVIDUAL PARAMETERS...........................................................................................................174.4.2 USE OF PT SENSORS ACCORDING TO THE UNICONFIG DATA...........................................................................18

4.5 VACUUM TEST...............................................................................................................................................194.6 AIR DETECTOR ............................................................................................................................................19

SUPPLEMENTS:SETTING THE PRINTER PARAMETERS.........................................................................................................................1 SHEET

STERILIZATION CYCLE – DRAWING NO: KV_202.............................................................................................................1 SHEET

CONNECTION SELECTOMAT SL 4-6X 3X380-430V PE Z468466......................................................................................14 SHEETS

ERROR REPORTING................................................................................................................................................2 SHEETS

Page 5: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

1 INTRODUCTIONThis Service manual is valid for SELECTOMAT SL models and it contains the summary of data concerning the product, necessary to make the repair works, maintenance and device check.It assumes knowledge of both the Instructions for use, written as a separate document, and annexed connection diagrams.

2 DESCRIPTION OF THE UNIT

The SELECTOMAT SL unit belongs on the category of large steam sterilizers. It is designed to be used in the health service facilities, in central sterilization stations, in the hospitals and generally in all branches where sterile goods are used.

2.1 STERILIZER CHAMBERIn the sense the sterilizer chamber is a pressure vessel. Another standard, holds for operation, maintenance, revisions and tests of pressure vessels.The chamber is made from chrome-nickel-molybdenum stainless steel of grade AISI 316L and ASI 316Ti. The inner surfaces of the chamber are surfaced. The sterilizer chamber is closed by a vertically slidable, electrically or manually operated door. The door and chamber are thermally insulated with mineral fibre (thickness 50 -130 mm).The door and the chamber are thermally insulated by mineral wool. An electrically operated door is fitted with a protective bar with four switches situated under the door cover. The door drive coupling is adjusted to develop a force of 150 N when closing the door. A manually operated door is fitted with a locking magnet with respective position switches avoiding opening the door in case that the pressure in the chamber has not yet reached the atmospheric level. The basic sealing element of the door system is a hose-like seal put into a groove in the vessel. After starting the program, the seal is pushed against the door by steam pressure, what secures tightness between the vessel and the door. After finishing the the program, the seal is pulled back into the groove by vacuum. In models 666 – 6612 the left pair of the door travelling rollers is fitted with an excenter for adjusting the clearance between the guide bars.The sterilizer chamber is provided with two test connections with pipe threads ISO 228-G1/2 A and G1-A allowing to introduce checking sensors into the chamber. The chamber contains shelves for placing the goods being sterilized, or a multi-story frame for the model BW loading trucks.

2.2 STEAM GENERATORThe sterilizers are delivered in three variants FD, ED, FED. The built-in electric heated steam generator (variants FD, FED, or possibly FDT) is designed to generate clean steam from completely demineralized water. The generated steam shall not contain hydrazine (N2H4) or other volatile, corrective chemicals and corrosion inhibitors.

The steam generator may be supplied only with respectively treated, totally demineralized water. A treatment on the principle of reverse osmosis is duly recommended.The power input as per sterilizer models amouns to:446/636 22,5 kW666 36 kW669/6612 45 kW9612 63 kW9618 72 kWThe steam generator is made from chrome-nickel-molybdenum stainless steel of grade ASI 316Ti.The boiler is fed by means of the feeding pump from a storage tank with controlled water level. The water level in the boiler itself is controlled by level switches.There is a double protection of the heater elements. The lowest permissible water level NW is secured by the safety level switch B91 (locking the switching-on of heating power). The operational height of the water level is maintained by the level switch B90, having also a protective function. In case that water has not been refilled within a certain time limit (approx. 40 s), the heating power is reversely switched off. The control of the steam generation is performed by a pressure gauge in dependence on the steam pressure. Excessive values of the steam presure are avoided by built-in safety pressure limiter. The heater elements are switched on and off successively in blocks with approximately 1s delay.

2.3 PIPEWORK

2.3.1 SCHEME OF PIPEWORK CONNECTION - APPLIED ELEMENTS

Y01 – Y600, - active valvesV1-V120 - safety valves, hand-operated, reduction, Back check valves, thermo regulatingB20-B31 - pressure switchesB86,B91 - level switchesC1 - feeding pumpF1-F30 - filtersPE1-PE200 - pressure sensorsP1 - flow limiterO1-O200 - steam traps (condensate discharge)VY1 - water-ring vacuum pumpPT1 - PT20 - temperature sensorsTT - heater elementsCHL - cooler.

2.3.2 STEAM TRAPSTwo steam traps Spirax-sarco types are used standardly in the devices.1. “direct” type BTP 13S - 0350210, which is for MMM fitted by a liner with “G” marking2. “angular” type BPT 13A - 0350211, which is for MMM fitted by a liner with “E” markingThe liners are delivered in the form of a set or as a spare part too and they are, by recommendation, bound with the corresponding trap (see the spare parts list).The marking is also in the liner body.liner 0350208, “G” type (for the 0350210 trap)liner 0350207, “E” type (for the 0350211 trap)

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 3

Page 6: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

4 SELECTOMAT SL_ns_en 0911_V2.02

The “G” type liner is set “more sharply” (near the temperature/steam pressure balance) and it is therefore dedicated for traps connected to the outlet from the sterilisation chamber (now newly also for the direct trap of the steam/steam exchanger - FDD).This liner is not so much suitable for condensate draining from jackets as the draining is too rapid, noisy... in some cases.The “E” type liner is set to “standard” (i.e. a little bit under the balanced state, than the previous type) and it is dedicated for other traps, where the softer drain is required. E.g. the condensate drain from the jackets, supply pipe etc.This liner is not suitable for the chamber, as it drains the condensate for this not enough purpose and it can cause a distortion of the measured temperature on the main PT sensor, problems with the residual air, non-condensed gases etc.The liners are mechanically changeable and it is necessary, during the replacement, to replace them by a suitable type and with certain knowledge.

2.4 COVERAGEThe unit is covered with stainless panels of 1.4301 material quality. there is also a lockable service-door in the covering.

2.5 ELECTRIC CONNECTION

2.5.1 BLOCK DIAGRAMThe apparatus is controlled with micro processor automatics. From the following chart it is obvious the basic arrangement of the electrical equipment of the apparatus:

The basic interface „attendance – apparatus“ is the contact panel on the loading side of the sterilizer, which is controlled with the processor MASTER. This processor also ensures the contact with peripheries (PC, printer, chip-card, MMC).In the unloading side there is a smaller contact panel, which is controlled with the processor BACK.Further the apparatus contains the processor SLAVE, which processes the information from the digital (switches) and analog (temperatures and pressures) of inputs and controls the outputs.In order to secure the safety as well as in the case of a failure of the processor SLAVE, the apparatus contains a block of hardware protection, which is controlled with the processor MASTER and some inputs (see the chapter of hardware protection).

2.5.2 DESCRIPTION OF THE ELECTRIC CONNECTION DIAGRAM

The electric connection diagram is in the supplement of these instructions.The connection diagram is attached to these Instructions. The following page numbers correspond to the diagram Z468466 v.14 and diagram Z471105 v.4 (UL variant). In higher versions, the numbers can differ.

2.5.2.1 NETWORK PART The connection of the sterilizer wiring is shown on pages 1 and 2. Page 3 shows the secondary voltage distribution of the T1 transformer to the individual electronic boards. Meaning of the respective components is following:

E1-E8 HeatersF501,F503,F505, F507

Circuit breakers of heaters

K1-K7 ContactorsQ1,Q2,Q3 Circuit breaker relay of motorsVY1 Vacuum pump motorC1 Pump motor - generatorC2 Pump motor - methyleneFL1 Fluorescent lamp

with electronic ballast T1 TransformerF1 Cut-out fuse for supply of

electronics T3,15A

F2 Cut-out fuse for display backlight in the loading side

T3,15A

F3 Cut-out fuse for supply of alternate outputs and contract of the panel in the unloading side

T6, 3A

F4 Cut-out fuse for supply of the inputs and direct-current outputs

T6, 3A

F5 Fuse for T1 and FL1 T1A for 380-480VT2A for 220VT3A for 208V (UL)

M5 Switchboard fan

2.5.2.2 SWITCHBOARDIn the sheet No. 4 it is drawn the arrangement of the respective components in the switchboard.The detailed connection of the X1 and X3 terminal boxes is on the sheet No. 5. This is applied in a device with maximum configuration. In simpler devices, the unused terminal connectors are not fitted. The flexible supply cable is connected to the L1, L2, L3, and PE terminal connectors. Page 5 shows the left version of the distribution board for the UL variant.The interconnection of the VP_CPUS (SLAVE) processor board with the VP_PDC (DC outputs) and VP_PAC (AC outputs) power boards is drawn on the sheet No. 8. The number of power boards depends on the equipment configuration.To the connectors XS1 and XS2 of the board VP_CPUS it is led out the internal busbar RS485. if there are connected both two connectors (two-door version), the switch S1-1 (between the connectors XS1and XS4) must be in position „OFF“, if it is connected only one connector (one-door version), S1-1 must be in position „ON“.The connector XP8 serves for connecting of PC (recording of bios).

CONTACTPANEL

(unloading side) CONTACT

PANEL (loading side)

DIGIT.+ANALOG INPUTS

OUTPUTS

CPU SLAVE

CPU BACK

CPU MASTERRS485RS232

USB

MMC

PC

ETHERNET

CHIPCARDPRINTER

EXTENDING MODULES

HW PROTECTION

Page 7: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 5

The connectors XP3 and XP4 serve for interconnecting of the sterilizers between them (control of consumption). If the cables are connected to both connectors, the switch S1-2 must be in position „OFF“, if there is connected only one connector, S1-2 must be in position „ON“.

2.5.2.3 FRONT PANEL (ON THE LOADING SIDE)Pages 6, 7 and 14 show the interconnection of the individual front panel elements.The tiltable panel for the SELECTOMAT SL models consists of the touch panel glued by a double-sided adhesive tape to the display, of the VP_M_LB121 board, inserted into the display, of the converter for CFL (display illumination) and of VP_CPUM (MASTER) processor board, for the SP HP E and SP HP S models it consists of the touch panel with display (one part), of the VP_LVDS board, of the converter for CFL (display illumination) and of VP_CPUM processor board. On the rear side of the front panel, the VP_KON connector board is placed and the chip card reader, printer and servo (for SELECTOMAT SL models) are connected to it.When making connection of the servo it is important to take care of the position of the black wire – see the diagram. Furthermore on it there are connectors for connection to PC (RS232, USB and ETHERNET). To the connectors XS1 and XS2 it is led out the internal busbar RS485. If the cable is connected only to one of them, it is necessary that the both two switches S1 (beside the connectors XS1 and XS2) would be in position „ON“, otherwise both the switches S1 must be in position OFF.To the connectors XS1 and XS2 of this board it is led out the internal busbar RS485. If it is connected only the cable only to one of them, it is necessary that the both switches S1 (beside the connectors XS1 and XS2) would be in position „ON“, otherwise both the switches S1 must be in position OFF.If the VP_CPUME.. processor panel is used in SP HP E or SP HP S models, it replaces the VP_CPUM, VP_KON and VP_LVDS panels – see page 14.

2.5.2.4 REAR PANEL (ON THE UNLOADING SIDE)In the sheet No. 9 and 10 it is drawn the mutual interconnection of the respective elements of the rear panel.The rear panel consists, for SELECTOMAT SL models,of the touch panel, display, converter for the display illumination and VP_CPUB processor board, for the SELECTOMAT SL models only of VP_CPUBE processor board. The internal RS485 bus is brought on the XS1 and XS2 connectors of the processor boards. If the cable is connected only to one of them, it is necessary that both S1 switches (close by XS1 and XS2 connectors) are in the “ON” position, otherwise, both S1 switches must be in the “OFF” position.

2.5.2.5 INPUTS

Page 11 shows the connection of all inputs.Digital inputs are connected to XP1 - XP4 connectors on VP_CPUS panel (there is a LED at each input lighted up when the switch is on) and analogue inputs are connected to XP5 - XP7 connectors.

Meaning of the respective components connected to the inputs and their function is following:

DIGITAL INPUTSIdentification Meaning Function within

switched contact

Q1 Thermal relay – vacuum pump motor overloaded

Q2 Thermal relay – generator pump motor overloaded

Q3 Thermal relay – pump methylene

motor overloaded

B20 Pressure switch 1.5 Bar - slot ( BS) under pressureB20.1 Pressure switch 1.5 Bar - slot ( ES) under pressureB31 Pressure switch 3 Bar

– generator without pressure

B86 Bottom lever gauge – demiwater floodedB89 Upper level gauge - demiwaterr floodedB90 (HW) Upper level gauge - generator floodedB91 (NW) Bottom level gauge - generator floodedB100 Upper level indicator - methylene floodedB101 Bottom level indicator - methylene floodedB102 External level indicator -

methylene flooded

S11 Entry door (BS) closed closedS11.1 Entry door (BS) opened openedS11.2 Input service door (BS) (9x model) openedS11.3 Input door (BS) closed (9x model) openedS12 Safety strip – entry door (BS) not pressedS13 Output door (ES) closed closedS13.1 Output door (ES) opened openedS13.2 Output service door (ES) (9x model) openedS13.3 Output door (ES) closed (9x

model)opened

S14 Safety strip – output door (ES) not pressedL1 Blocking – lack of electricity blocked

ANALOG INPUTSIdentification Meaning Evaluated by

processorPE1.1 (SLV) Pressure transducer - chamber SLAVEPE1.2 (MST) Pressure transducer - chamber MASTERPE2 Pressure transducer - casing I. SLAVEPE20 Pressure transducer - casing II. SLAVEPE3 Pressure transducer - generator SLAVEPE101 Pressure transducer - air SLAVEPE102 Pressure transducer – entry steam SLAVEPE103 Pressure transducer – water for

vacuum pumpSLAVE

PE104 Pressure transducer – entry demiwater

SLAVE

PE105 Converter pressure – softened water (EW)

SLAVE

PE 106 Pressure converter - heating steam MASTERPT1.1 (SLV) PT sensor – chamber SLAVEPT1.2 (MST) PT sensor – chamber MASTERPT2 PT sensor – water saving SLAVEPT3.1 (SLV) PT sensor – solutions, reserve SLAVEPT3.2 (MST) PT sensor – solutions MASTERPT4 PT sensor – generator SLAVEPT5 PT sensor - demiwater SLAVEPT6 PT sensor – detection of air SLAVE

Page 8: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

6 SELECTOMAT SL_ns_en 0911_V2.02

2.5.2.7 HARDWARE PROTECTION

Due to safety purposes the outputs, which in case of undesirable switching can endanger the user, or possibly to damage the apparatus, are blocked by means of the relays, which are switched with the switches being independent to the automatics of the apparatus. The connection of the relay coils is obvious from the sheet No. 9. The appropriate contacts of these relays are represented in the sheets No. 10 and 11.The moment, when it is possible to switch the outputs, is obvious from the following table:Name of the output

Identifica-tion

Relay Switch

Valve – steam to the chamber

Y07 K9 , K13K9.1*,K13.1*

B20 – slot is sealedB20.1* - slot is sealedS11 - door BS is closedS13 - door ES is closed

Valve - steam to slots BS+ES

Y09 K13 S11+S13 - doors BS +ES are closed

Valve – steam to slot BS*

Y09 K13 S11 - door BS is closed

Valve – steam to slot BS*

Y09.1 K13.1 S11 - door BS is closed

Motor of the BS door

M2 S11.2 – service door closed (for 9x)

Motor of the ES door

M3 S13.2 – service door closed (for 9x)

Valve – steam to slot ES*

Y09.1 Y09.1 S13 - door ES is closed

Connec-tor for vacuum pump

K1 Q1 Q1 in case of overloading of the motor disconnects the coil K1

Connec-tor for the pump C1

K2 Q2 Q2 in case of overloading of the motor disconnects the coil K2

Connec-tor for the pump C2

K3 Q3 Q3 in case of overloading of the motor disconnects the coil K3

Connec-tors for heating

K4, K5, K6, K7

K502 NW – bottom level gauge of the generator is floodedB31- in the generator there is pressure < 3,0 bar(K502 is blocking K4, K4-K5, K5-K6 and K6-K7)

All AC outputs

K60 Electronic board VP_CPUM (through VP_KON)

All AC outputs

K61 Electronic board VP_CPUM (through VP_KON)

* for independent door seal

2.6 DESCRIPTION OF THE AUTOMATICS FROM THE VIEWPOINT OF SOFTWARE

All control activites are performed by two independent microprocessors. The microprocessors cooperate in relation of master and slave. Communication between the Master

2.5.2.6 OUTPUTS

The connection of the AC outputs is shown on the sheet No. 12 and of the DC outputs on the sheet No. 13.The switching is indicated on the power board of the electronics VP_P… with appropriate LED diodes.Meaning of the respective components connected to the outputs is following:Identification Meaning NoteY03 Valve for filling the jacket I Y103 Valve for filling the jacket IIY07 Valve for steam transfer into the

chamber Y09 Valve of the filling slot (entry door BS) Y09.1 Valve of the filling slot (output door

ES) Y20 Valve for discharging of the chamberY23 Valve for discharging of the slot (entry

door BS)Y23.1 Valve for discharging of the slot

(output door ES)Y24 Cooling water valve for the air

detector Y27 Air detector valve - condensateY28 Air detector valve - air Y30 Cooling water valve for vacuum pump Y50 Valve for blocking of the condensate

effluent conduitY80 Valve for vent Y80.1 Sterilizable aeration filter valveY81 Valve for vacuum breaking of the

vacuum pumpY99 Valve for desludgingY150 Valve for degasification Y67 Valve for deaeration of the jacket Y70 Valve for jacket pressurizing with air

(counter-pressure)Y35 Valve for jacket spraying Y21 Valve for discharging the chamber Y66 Valve for discharging the jacket Y02 Valve of waste steaming Y501 Valve of water showering Y502 Valve MethyleneY503 Valve of chamber aerating

(methylene)Y504 Valve Methylene - pump bridging V16 Valve Heating steam K1 Contactor for M1 – vacuum pump VY1M2 Entry door (BS) opening

Entry door (BS) reversal (closing)M3 Output door (ES) opening

Output door (ES) reversal (closing)K2 Contactor for M5 – pump C1

generatorK3 Contactor for M6 – pump C2

methyleneK4 Contactors for heating K5 Contactors for heating K6 Contactors for heating K7 Contactors for heating

Page 9: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

transition to the initial state from the master, i.e. in the chamber it is atmospheric pressure and the slot will be de-sealed. The processor back receives from the attendance of the apparatus the instructions for the requested activity and hands over them to the processor master, which further processes them.

2.7 LOADING OF BIOS, APLICATIONS AND LANGUAGES

For recording – download of the software equipment into the sterilizer the program „Loader.EXE“ serves. Software is transmitted on the serial line RS 232 z PC by means of the cross cable:

Cannon 9 female

If we want to download the software, then we start the program „Loader.EXE“. on the upper bar we select with a mouse the file „File->Open the file“ and choose the file, which we want to download. On the surface of the program there will appear a window with the listing of the program. (The files are marked in a mnemonics manner such as BiosMltMst…). Then on the upper bar we select „Download->Load the file/process the batch“.When starting the download it is necessary that the window with the listing should have a blue bar. After starting the download it is possible to switch over the window. But it must not be closed.

2.7.1 LOADING OF BIOS Bios is loaded into every processor board separately (VP_CPUM, VP_CPUS, VP_CPUB). For this purpose each of these boards has a connector for connecting of a serial line RS232 (for VP_CPUM it is located on the board VP_KON). With respect to the fact that the switches, which enable to load the Bios, are located on the boards of electronics, before its loading it is necessary to uncover the boards. In loading the bios there is erased the processor memory, in which there are stored all the information on the apparatus (apart from the system version), therefore it is necessary, before you record – load the bios on the boards of electronics VP_CPUM, to backup the following information (see. print,…):- Configuration of the device

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 7

and the Slave takes place via RS485 bus, fully controlled by the Master. In moments, when the situation asks for, the Master sends required data to the Slave, in reverse direction pieces of periodically sampled information are transferred in form of data packets with fixed structure.If the apparatus is the two door type, to the busbar RS485 it is connected another microprocessor (back), which ensures the control of the apparatus form the side of the material unloading.

The following table shows the access of both microprocessors to the peripherals.Access to external initiation or possibility to perform activity

Master Slave Back

Communication with the PC – operation parameters

YES NO NO

Measurement of the temperature in the chamber

YES YES NO

Measurement of the pressure in the heating jacket

NO YES NO

Measurement of the pressure in the chamber

YES YES NO

Control of switches for: electromagnetic valves, feeding pump, vacuum pump, heating, door-driving motor

NO YES NO

Information on the status of the following sensors: the door electromagnet, the door-end switch, water level switches in the heating jacket and the water storage tank

NO YES NO

Information on the status of pressure switches

NO YES NO

Access to the LCD display, LED display YES NO YESAccess to the keyboard YES NO YESSound signalling YES NO YESPrinting in the printer YES NO NOInformation on the power supply fall-out YES NO NORAM backup YES NO NOCheck of current flow in the heating jacket water pump

NO YES NO

Check of current flow in the vacuum pump

NO YES NO

Software of the apparatus has two main layers, BIOS and APLIKACE. BIOS ensures the communication between the appropriate processors. And further the upgrade of the application.The application ensures the correct function of the sterilizer. The processor master receives the instructions from the attendance for the requested activity and at the same time for the control it assesses from the processor slave the states of sensors, values of measured variables and also the information regarding to the fact, within which phase of sterilization cycle the process is found. Many of necessary data are obtained independently, or it compares them to the parallel obtained data from the slave. If the attendance gives instruction to the commencement of the activity, the master sends in direction to the slave the packet of necessary parameters. The processor slave is entrusted the course of the sterilization cycle on the basis of the information from the processor master. Slave has a direct access to the outputs (valves, contactor, ...) and to the inputs (thermal and pressure sensors, switches, ...). If the slave detects the error, it notifies of the fact the processor master and freezes its activity, i.e. all the action variables take up the safe state and it is waiting. Only after deleting the error by the attendance the slave receives the instruction to

(2)

(3)

(4)

(6)

(7)

(8)

(5)

(2)

(3)

(4)

(6)

(7)

(8)

(5)

Page 10: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

8 SELECTOMAT SL_ns_en 0911_V2.02

Service instructions

- Setting of the device- Offsets

Bios is loaded in the following method:1. Start the Loader.exe2. Open the file „BiosMltxxx.fls“ (so as it would have a blue

bar) xxx:Mst for VP_CPUM Slv for VP-CPUS Bck for VP_CPUB3. Select „Download->Load the file/process the batch“4. To observe the instruction on the screen: -„Switch over the processor into the boot mode“: - Switch over the both switches DIP on the electronics board into position ON

- For VP_CPUM (inside the front hinged panel) DIP RESET

- For VP_CPUS (in the switchboard)

- For VP_CPUSB (the rear panel)

- Reset the processor by the button RESET- Confirm the dialog5. Wait until on the PC in the window „Listing of the process“

it is listed „End „- „Switch over the processor from the boot mode“- Switch over the both switches DIP on the electronics

board into position OFF- Reset the processor by the button RESETAfter loading the bios it is necessary again:- To load the application- To perform the complete setting of the device including

the offsets- To load the sterilization programsNote:If on the electronics board VP_CPUM or VP_CPUB there is recorded – loaded only the bios, the appropriate screen is black.

2.7.2 LOADING OF THE APPLICATIONThe device must be in the sleeping state (the process is not running) and there must not be selected the utility service „Download of programs“. When downloading the application there is erased the RAM backup memory, therefore it is necessary to record the user´s settings under the menu items „Setting of the display“, „Print“ and as well as the number of the performed batches.When downloading the application, the PC is connected to the sterilizer via the connector RS232 on the electronics board VP_KON, and also for the electronics boards VP_CPUS and VP_CPUB. As to be first it must be downloaded the application (including the languages) into the electronics board VP_CPUM.The application is loaded in the following method:1. Start the Loader.exe2. Open the file „VpsMltxxx.fls“ (so as it would have a blue

bar) xxx: Mst pro VP_CPUM Slv pro VP-CPUS Bck for VP_CPUB

Page 11: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 9

The pressure is scanned by the pressure gauge PE20. The back-flow valves V13 are fitted only within the version FED. After starting the sterilization program the door is sealed via the valve Y09. The pressure under the seal is checked by pressure switch B20. There is started the vacuum pump VY1, which via the valve Y20, filter F4, cooler CHL and the back-flow valve V12 exhausts the air from the working space. The pressure in the chamber is scanned by the pressure gauge PE1.1 and PE1.2. After gaining the necessary under pressure the valve Y 20 is closed and it is opened the valve Y07, through which the steam is filled into the chamber. The condensate trap O2 realizes the condensate exhaust from the chamberse. In an analogic way, the sterilization cycle proceeds till the phase of pressure releasing, when the steam is intermittently discharged from the chamber via the cooler CHL. Cooling water is supplied into the cooler via valve Y30, filter F6 and flow limiter P1. Water is further supplied into the cyclone and via an orifice plate it feeds the vacuum pump VY1. Excess water flows away via the overflow. The temperature of water in the cyclone is scanned by a PT2 temperature sensor.In case that the water temperature is lower than the preset value, the valve Y30 in some phases of the sterilization cycle closes and the vacuum pump is then feeded from a closed circuit to save water. It is possible to set the controlling temperatures within the automatics.Valve Y80 opens when aerating the chamber. The air flows into the chamber via bacteriological filter F1 and non-return valve V10. Steam is sucked off the groove in the door by vacuum pump VY1 via opened valve Y23, filter F5 and non-return valve V12.Compressed air for pneumatic valves is reduced onto approximately 5 bar pressure by the reduction valve V3. Temperature in the chamber is scanned by double sensor PT1.1, PT1.2. The condensate from the jacket I is conducted via the filter F3, condensate trap O1. The condensate from the jacket II is conducted via the filter F103, back-flow valve V111, condensate trap O101.Additives:-Solution program with self-cooling.Within this version in addition to the device is fitted with a movable sensor PT3, which indicates the temperatures of solutions in the reference bottle. The valve Y50 blocks the condensation valve O2 in case of the started solution program – in the phase of cooling down.-Bacteriologic filter - decontaminationWithin this additive the device is supplied with the bacteriologic filter FIL1. The valve Y02 for lower steam filling into chamber – sterilization of condensate. Another valve is the valve Y21 for the condensate exhaust after the sterilization exposure and evacuation of the chamber.

Sterilizer may be alternatively equipped with a series of other additives:• Forced cooling with the air counter-pressure • Independent control of doors• Sterilisable filter (air vent)• Monitoring of media• Waste aftercooling• Air Detector etc.

3. Select „Download-> Load the file/process the batch „4. Wait until on the PC in the window „Listing of the

process“ there is listed „End „5. When loading into the electronics boards VP_CPUM

open the language file „VpsMltMstLng“, for the electronics board VP_CPUB „VpsMltBckLng“

6. Select „Download-> Load the file/process the batch„7. Observe the instructions on the screen - Select maximally 4 languages – they must be the

same as for Mst and Bck !

9. Wait until on the PC in the window „Listing of the process“ it is listed „End „

2.7.3 RULES FOR WORK WITH THE PROGRAM LOADER.EXE

Rule no. 1: If you want to download some file into the sterilizer, then before the selecting „Download“ the window of this file must have a blue bar!Rule No. 2: If you want to close the Hitachi Loader and it does not work, please, select „Download->Stop“ !Rule No. 3 : Do not close the window for the file, which is being downloaded !Rule No. 4 : If some action is not successful, please, repeat it (also for several times) !

3 FUNCTION OF THE UNIT

3.1 PIPING DIAGRAMIf within the automatics it is activated the generator, the pump C1 refills with demineralised water into the generator from the storage tank via the filter F7, back-flow valve V14 and valve V7. Reaching the necessary level is secured by the level switch B90. After starting the generator the heater elements TT begin to heat. The boiler is brought onto the operating pressure, which is being scanned by the pressure gauge PE3. The valve Y99 opens for approximately 2s after every water refilling to decrease concentration of salt in the boiler. The level switch B91 secures protection of the heater elements. The pressure switch B31 serves as protection against overpressure. Valve V6 is used for manual mud discharge from the boiler. The generator is equipped with thermal degasification of the feed water. Water in the storage tank is heated by a heating coil. Heating is regulated with the valve Y150. The temperature of demineralised water is scanned by the temperature sensor PT5. The minimal level of demineralised water in the storage tank is scanned by level gauge B86. In case of shortage of water the feed pump C1 is blocked.

After selecting the sterilization program the steam of the heating jacket I is filled via the filter F2, valve Y03 for the pressure being set by the selected program. The pressure is scanned by the pressure gauge PE2. Into the jacket II the steam is filled via the valve Y103.

Page 12: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

4 ADJUSTENT OF THE UNIT, SETTING THE PARAMETERS

The unit is adjusted and tested by the manufacturer.Selected adjustment and checking values of the SELECTOMAT SL sterilizer:Name Designa-

tionValue Note

Door coupling 150N Door closing forceSet in upper third of the door lift

Steam generator manostat

B31 +3,0 bar

Float valve of the water storage tank

cca 80 mm

Height of water level in the storage tank

Sterilizer safety valve V1 3,2 barSteam generator safety valve

V2 3,2 bar

Cooling water temperature

T1=27°C, T2= 33°C

Automatics

Compressed air reduction valve

V3 cca 5bar

Compressed air reduction valve

V40 cca 2,6

Compressed air reduction valve

V4 3,2 bar

4.1 MENU - SETTING AND SERVICES ACCESSIBLE FOR THE UTILITY TECHNICIANS

The menu enables for the utility technician a detailed setting of the apparatus test of safety elements, controlling of the appropriate outputs independently on the automatics and other functions, which are not accessible for the attendance.In this chapter there are described only the functions, which are not described in the instructions for use.Most of changes, which are described in this chapter, require the entry into the internal memory of the apparatus.

This will occur after pushing the return button . And after confirming this operation by pushing the button “Store”

For the period of storage the apparatus does not react to the touch contact.If you are not sure that you have performed the mentioned operation correctly, please, push the button „Do not store“, by which you will arrive at the screen by one level higher, or push possibly the button „Cancel“ for the return to the set screen.If it is possible to set the mentioned parameter only on certain limits, these limits are indicated in the brackets. In order for the inactive buttons to become the functional ones, it is necessary to log in the following manner:

- To push the button

- The screen appears:- into the line marked as “Jmeno”, write the name,

assigned by the service MMM department, e.g. serv001 (the line is activated by the touch)

- into the line marked as “Heslo”, write the assigned twelve-character password.

- pushing the button

4.1.1 TEST OF SAFETY ELEMENTS After pushing the button „Test of safety elements“ there appears the screen:

- By pushing the button „K60+K61“ it is possible to verify the functionality of the relay K60 and K61, which in case of a critical error will disconnect the supply of outputs (see the connection diagram).

10 SELECTOMAT SL_ns_en 0911_V2.02

Service instructions

Page 13: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 11

- Function of the safety pressure switch B31, which creates the independent protection of the generator, can be checked by pushing the button „B31“. There appears the screen:

Push the button „GENERATOR“ – the generator works in normal mode, i.e. it regulates the pressure within the set pressure values (see the setting of the apparatus). The actual pressure value is displayed by a colour column and concurrently in figures in the left upper part of the screen. After attaining the upper pressure limit of the generator (in this case 370 kPa) push the button „GENERATOR++“. By this the generator is switched on for the period of 9 seconds, which is indicated by the number beside the button. At the moment, when the pressure switch B31 disconnects, the screen will appear with the message „Pressure switch of the generator“. Pushing the button „OK“ you will arrive to the preceding screen se, in which there was changed the message „B31 without pressure“ to „B31 Under the pressure“.In the case that B31 will not disconnect within nine seconds, again push the button „GENERATOR“.On the screen there are also the information on the fact, in which state there are the inputs and outputs being necessary for the function of the generator.If it occurs that sooner than B31 activates the safety valve, it is necessary to adjust B31 to lower pressure, or to replace the defected safety valve or the B31.- Functionality of the bottom level gauge of the generator

can be verified after pushing the button „B91“. Proceed according to the instructions described on the screen. After confirming one step, the following one will automatically highlight.

For the safety reasons, before the test the generator must be in cold state and without pressure.

4.1.2 DISPLAY OF BLOCKS After pushing the button „Display of blocks“, there appears the screen, on which there is schematically displayed the selected block of sterilizers:There are displayed only these parts of the device, which are configured - see „Configuration of the device“.

By means of the buttons on the lower bar it is possible to switch over between the following blocks:

- Block of the chamber

- Block of the generator

- Block of the vacuum pump

- Block of the doors

- summary of digital inputs and outputs All the displayed outputs can be controlled by touch, however, in dependence on the process, which is just running, some outputs return to the state, which is required by the given process.Furthermore on the screen there are displayed all the analog and digital inputs related to the given block. Colour of the digital inputs expresses, if the given input is connected or disconnected (see „block the summary of digital inputs and outputs “), depends also on the selection of the colour palette. The last block „the summary of digital inputs and outputs“ displays all the digital inputs and outputs, but these ones, which are not configured, have different colour and can be inactive ones.

4.1.3 SUMMARY OF INPUTS - OUTPUTS

- After pushing the button there appears the screen “The summary of digital inputs and outputs”- see above.

Page 14: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

- After pushing the button sfirst column provides the identification of the mentioned input.- Second, the measured value without offset.- Third, the actual offset for this value.- Fourth, the measured value plus offset.- Other columns, the set offset see „Setting of the device“.- Under the line there are displayed the values of the

supply voltage for the electronics. It is especially important to follow the voltage of the battery „Ubat“ and in case of decrease to 2.7 V, it is necessary to replace it.

4.1.4 DEVICE CONFIGURATIONThis service allows the device configuration according to its production characteristics. It is very important to choose the correct device type and to configurate only such parts that are embedded in the device; otherwise, the device may not work correctly and can display errors.

After pressing the Device Configuration button, the screen Device Configuration 0/4 is displayed. There you can select the device type; the content of the following screens will be modified accordingly.Use arrows to change over the subsequent screens.The Device Configuration 1/4 screen allows the chamber volume setting.Use the Device Configuration 2/4 screen to set the appropriate content of the device:- Solutions – if yes, the correct PT sensors must be

chosen (PT 3.1 is a standard)- Bacteriological filter- Jacket cooling- Methylene- Air detector- Waste aftercooling- Energetic maximum control- Parameters modified by the user- Ergonomically adjustable LCD panel- Automatic start- Thermal degassing

Use the Device Configuration 3/4 screen to set the following:- Number of doors- Coupled or separated door sealing; use of gas or air for

the sealing- Door type – “right door” means that the door on the

loading side is positioned to the right of the display- Steam supply

- FD – external source of the clean steam- FDT – external source of the heating steam- ED – integrated electric steam generator- FDD – integrated “steam-steam” generator

- Safety valve V1- Sensor B89 – upper level gauge in the demineralised

water tank.

Use the Device Configuration 4/4 screen to set the following:- Chip card sensor- Input media monitoring sensors- Output data configuration - Aeration filter type – if a sterilizable filter is used, we

must find the power board to which the Y 80.1 valve is connected, and to set the given power board

- Y 80 valve position.

12 SELECTOMAT SL_ns_en 0911_V2.02

Service instructions

4.1.5 SETTING OF THE DEVICE

4.1.5.1 SERIAL NUMBER AND CHARGES COUNTER

The serial number and charges counter are set during the equipment production and the serviceman does not have any right to change them. These data are stored on up to nine independent places in the electronics and therefore they are not lost in case of replacement of individual boards.

4.1.5.2 OFFSET OF PRESSURE

Pressure sensor offsets are set here:

You can choose either one-, two- or three-point offset. The one-point offset is set in the extreme only if you do not have the device for pressure sensors setting.The two- and three-point offset is set in the following way:Use the arrows to select the appropriate pressure sensor. Press the Point A button and insert the pressure of the first measurement. Then press the Offset A button and insert the measured deviation including the sign. Proceed in the same way for points B and C.Proceed equally for other pressure sensors.Offsets are measured in the following way: The sensor is screwed off the device (it remains connected electrically to the device) and is connected to the calibration vacuum pump (recommended vacuum of 10 kPa). The difference calculated as the set value minus the pressure on the device without offset is the resulting offset. Repeat the procedure with the sensor connected to the calibration pressure air pump on which overpressure is set (205 kPa is recommended).If you do not have the possibility to measure the offsets and you still have to measure them, use the Preset button. After you press this button, the offsets of all pressure sensors will be set to the basic values.

Store the change by pressing the return button It is recommended to check the pressure offsets during the annual device overhaul.

4.1.5.3 OFFSET OF TEMPERATURES

Offsets of temperatures are set similarly as the offsets of pressures. Temperature offsets are measured similarly to the pressure offsets.

Page 15: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

SELECTOMAT SL_ns_en 0911_V2.02 13

Service instructions

4.1.5.9 SERVICE PLANNING (NEXT MAINTENANCE) This service allows the service technician to plan the service in detail.The device then notifies of an up-coming service activity by means of a message displayed on the screen, or of not meeting the deadline for a preset service activity, if any, by means of a printer output (in the report heading).Up to 15 service activities can be planned (in 5 groups), both according to the time or number of batches processed, either separately, or according to an earlier achieved criterion.• Press the button Menu / Device Setting / Next

Maintenance Use arrows up and down to select one of five pages (groups) of the next maintenance service activities.The next maintenance page can be activated/deactivated by means of the Service Activity button.The indicators in the upper right corner show the number of the actual batch and the activation state of the individual next maintenance pages. If not a single page is On, the function of service activities planning will be disabled at all.In the Service Activity Timing field, select the time or batch criterion, or both criteria together.The Was Set buttons set the initial state of the criterion in the moment of setting; the Interval buttons set the appropriate interval between the service activities.By means of the Notification buttons, you can set the number of days/batches by which the device will notify the operator in advance of the upcoming day of a given service activity.The “speaker” symbol shows the active state of a notification. (The appropriate notification for the operator of the upcoming service day has already been displayed.)The Preset button will reset the actual Next Maintenance page.

The service activity can be defined on the individual lines in the field Service Activities Description. If all three description lines on the page remain empty, the page will be excluded from the service activities planning function.The field Will Be Performed Exceeding shows any potential not observing (exceeding) of a specified date / batch number. (The specified dates are displayed in the field Service Activity Timing in the Will Be Performed indicator.)

Note:Information about the specified maintenance date not observing will also be shown in the printed cycle report!

The only difference is that the PT sensor is inserted in a dry calibration oven (recommended calibration temperature: 121 and 134 °C).It is recommended to check the temperature offsets during the annual device overhaul.

4.1.5.4 SETTING THE DOORS

This selection is accessible only in case of the two door device and it is possible to set one of the three selections:- Selection „12“ sets the door on the side of the main panel as to be strictly loading one and the other door as to be strictly unloading one, it refers to a classical configuration of the two door version, automatics watch the logics of opening the doors.

- Selection „21“ sets the door on the side of the main panel as to be strictly unloading one and the other door as to be strictly loading one, automatics watch the logics of opening the doors.

- Selection „XX“it is not determined, which side of the device is loading one and which the unloading one, automatics do not watch the logics of opening the doors, it is only secured that both the doors cannot be opened at the same time.

4.1.5.5 UNITS

This service allows to set the temperature units to either °C or °F and pressure units to kPa, bar, or psia.It also allows the pressure displaying on the lower LCD panel bar either in relative or absolute units.

4.1.5.6 INTERNAL GENERATOR

This service enables to set the limits of switching off and on of the generator (as standard 340 and 370 kPa).Furthermore here is set the temperature of degasification of demineralised water (as standard 80 °C)

4.1.5.7 WATER SAVING TEMPERATURES

This service enables to set, under what temperature in the cyclone (water storage tank for the vacuum pump), the cooling water is filled. Setting 1 is for the phases, when the vacuum pump comes to a deep under-pressure (the cooler the water, the higher suction capacity of the vacuum pump is – as standard 27 °C). Setting 2 is for other phases (as standard 33 °C).

4.1.5.8 ADMINISTRATOR-PRESET (ACCESS SAFETY AND DOOR BLOCKING PASSWORD)

The serviceman does not have any right to change the access name and administrator password or individual users, he is only allowed to press the button “Přednastavit” in the “Administrátor” menu and so he set the administrator name to “admin” and the password to “heslo”.Also the password for door unblocking could be seen on this display.

Page 16: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

14 SELECTOMAT SL_ns_en 0911_V2.02

Service instructions

Now in the counter, please, confirm the data export. Program was downloaded to the appropriate program location and you have a possibility to confirm or to

prohibit , In the other case the attendance will not be allowed to start it. Similarly it is possible to select, if the program will be in the offer of a quick selection.

By means of the button the program can be partially modified.

4.1.7 PROGRAMMES JOININGIf the programme “Crash test” (eventually other special programme) is downloaded in the equipment, it is possible to join it to some sterilisation programmes. Then, this test is performed automatically after the finishing of given programme.The joining is made as follows:

Choose, in random sequence, one programme from the middle column one from the right column. In the left column, you can see which programmes are joined. The programmes decoupling is made by the gradual pressing of both joined programmes.

Example of a device operator notification (Message) of service activities:

Example of notification of the maintenance interval exceeding in the printer output:

SELECTOMAT SL 080511PS Bowie+Dick test, 131,0 °C, 3.5 minUser serv001Start 08:55:22 2008-06-30T = 61.8 °C, p = 97.5 kPa

Batch 000820Next maintenance 1: 2008-06-24/000825 - exceeding 6/-- Door sealing groove cleaningNext maintenance 2: 000810 - exceeding --/10 Door sealing replacementNext maintenance 3: 2008-05-15/000808 - exceeding 46/17 Aeration filter replacement Bacteriological filter-FBA replacement

4.1.6 DOWNLOAD OF PROGRAMSThis service enables to download the programs from PC to the device, or possibly to modify them.In the PC there must be installed the program UNICONFIG, which controls the download.Please, interconnect PC to SELECTOMAT SL by means of a serial cable RS 232 (on the side of the device the cable is inserted into the connector XP3 on the electronics board VP_KON ), start the program UNICONFIG and select the appropriate program. On the side of the device by pushing the appropriate button, select the location, where the program is to be downloaded.

Page 17: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

SELECTOMAT SL_ns_en 0911_V2.02 15

Service instructions

4.1.8 SYSTEM VERSIONAfter the pressing of this button, the following screen is displayed with the information concerning the hardware and software versions of each boards, serial number of the equipment and the charge number.

Automatics watches that in the device there would be fitted the components mutually compatible in terms of hardware and software. In the case that some part of the hardware or software is incompatible, this component is marked in red colour. Information about the version of hardware is downloaded on the boards at the time of debugging, so the spare parts have this information in them.Abbreviation meaning: - Mst – board of electronics VP_CPUM- Slv – board of electronics VP_CPUS- Bck – board of electronics VP_CPUB- Pwr – board of electronics VP_P (AC, DC see the connection diagram).

The first line of each board gives the hardware version, number of the equipment and the charge number, the second line gives the bios version (only for Mst, Slv and Bck), the third gives the application version (only for Mst, Slv and Bck) and the other gives the language version (only for Mst and Bck).

4.1.9 LCD PANEL OFFSETThis service allows the service technician to align the position of points on the touch panel and its image on the LCD panel.After pressing the LCD Panel Offset button, the following screen displays:

When you touch the touch screen at any point, you will see a cross on the LCD panel showing the touched place. Align the touched place with its image by means of gradual pressing the “+” and “-” buttons both in vertical and horizontal direction. The Redraw button renews the screen without crosses displaying. Red button contours show the button position on the touch panel stored in the device memory; the green contours show the position actually being set. It is recommended to set the offset according to the central Test button.

Store the change by pressing the return button

4.2 REPLACEMENT OF THE PARTS OF ELECTRONICS

4.2.1 REPLACEMENT OF THE ELECTRONICS BOARDS

The appropriate electronics boards are obvious from the connection diagram, sheet 4-6 and their order numbers from the spare parts list. In case of their replacement it is necessary to observe the following:1. Before the replacement of the board VP_CPUM it is

necessary to backup:- The setting of the device- Configuration of the device- OffsetsThis information is included in the test report, which is delivered with the equipment. In case that the serviceman changes any parameter, it is suitable to print and store the changed parameters immediately.

2. In case of replacement of the VP_CPUM, VP_CPUS and VP_CPUB processor boards, it is necessary to record there a SW compatible with other boards inclusive bios (the activated board has still recorded bios, therefore it should be overwritten only if an older version is stored on the board).

3. After the replacing and programming the board VP_CPUM it is necessary to perform:- The setting of the device- Configuration of the device- Setting of the offsets

Page 18: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

If it was not possible to read it from the replaced electronics board VP_CPUM, it is necessary to obtain the information from the documentation for the mentioned device.

4. After replacing and programming of the board VP_CPUS it is necessary to store the configuration of the device and offsets of the pressures. In this operation there is not set anything – by performing the storage, the data from the memory of the board VP_CPUM are copied into the memory of the board VP_CPUS.

5. When replacing the other boards of electronics there is not set anything.

6. If it is necessary to replace more boards, they must be replaced gradually and the equipment must be turned on after the replacement of each board!.

4.2.2 BATTERY REPLACEMENTThe battery is placed in the electronics board VP_CPUM (VP_CPUME). It is necessary to replace it, if its voltage decreases to the value of 2.7 V. Before its replacement, please, write down the state of the batch counter. Set the date and time and all user settings (language, colour palette, …) after the replacement.

4.2.3 REPLACEMENT OF THE CONTACT PANEL - FRONT PANEL

The contact panel is glued with a bonding tape into the frame of the hinged panel. The contact surface is on the side of the strip tap, this is inserted into the viewport of the frame. When replacing the contact panel first it is necessary to glue a bonding tape into the frame and then to insert it in such a method so as the active surface of the contact panel would not touch the bonding tape (The active surface is transparent).Further the bonding tape is glued on the other side of the display and on its other side there is left the cover foil (this tape serves for sealing between the contact panel and the display). The display is laid on the contact panel without gluing and it is screwed by means of the raising posts to the frame of the hinged panel.

4.2.4 REPLACEMENT OF THE TOUCH PANEL AND DISPLAY - REAR PANEL

The contact panel is glued with bonding tape to the display. When replacing it first it is necessary to separate by scalpel or a thin knife the old contact panel from the display and then to glue a new contact panel so as the active surface of the contact panel would cover with the active surface of the display.After the replacement of the display, the display may not be legible due to an inappropriately set contract. The contrast can be adjusted roughly by pressing the right upper (contrast increase) or left upper (contrast decrease) corner of the touch panel just after the device switching-on (a clock is displayed). Fine adjustment is carried out after pressing the Menu and Contrast buttons.

4.3 DEFINITION OF THE PROGRAM PARAMETERS

4.3.1 THE WAY OF DEFINITION OF THE PROGRAMME PARAMETERS

The automatics of the SELECTOMAT SL unit is designed so that the actual courses of the individual sterilization programmes are defined in every ready made unit by means of hardware and software tools.Hardware tools:

computer compatible with the IBM PCinterconnecting cable

Software tools:UNICONFIG.EXESet of templates xxx.sab- KlasicCykl.sab (for sterilization data files *.bp1)- Roztoky.sab (for sterilization solutions data files *.li1)- Air detektor.sab (for AirDetector data files *.ad1)- Vakuum test.sab (for vacuum test data files *.vt1)- Unicrash.sab (for UNICRASH data files *.uc1)- Data files

UNICONFIG.exe, templates and data files must be in one directory! The actual form of the program is defined in the following method:By means of the mentioned-above procedure in the device in the menu „Menu“, to select „Download of programs“ and confirm. Then it is necessary to set the program, which we want to alternate. In case of PC it is necessary to proceed as follows:To start the program UNICONFIG.EXE. To select the menu „File->Open“, to open the directory, in which there are stored the appropriate data files, in the line „files of the type“ to select the correct type of the file, to choose the required file (it must appear in the line „name of the file“) and to push the button “Open”. On the PC screen there appears the definition field with parameters. By means of the cursor arrows, or by means of the mouse, the appropriate parameters are selected and then they are modified. The key Enter in general opens the dialog window for editing the value of the selected parameter. A double click on the left pushbutton of the mouse substitutes the key Enter.The tabulator changes the active column of parameters. A finished packet of parameters can be stored on the computer disc.By this there are defined the files of concrete parameters, how they are referred to in the paragraphs about software aids. If we should define a program, we select within the program UNICONFIG.EXE menu „Communication->Data export”. By confirming the program is transferred to the device and there is set the following program. After downloading the programs it is necessary to store the changes. For the transfer it is possible to utilize also the icons on the auxiliary bar.Warning: The parameter changes may unfavourably influence the sterilization cycle and therefore they may be changed only by the authorized person. For this purpose it is better to utilize only the programs created in the manufacturing plant. The manufacturing plant does not hold any responsibility for the damages caused by improper setting of the parameters.

16 SELECTOMAT SL_ns_en 0911_V2.02

Page 19: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Service instructions

Time of sterilization/ Fo

s Duration time of sterilization exposure/value of the parameter Fo

Temperature of sterilization

°C Required temperature of sterilization

Temperature correction (reg. pressure)

°C Add to the required sterilization temperature and from the resulting sterilization temperature it is determined the regulating pressure

Temperature overrun

°C When starting at resulting sterilization temperature the minimum temperature will excess by this value, subsequently the sterilization exposure will start.

Number of teeth when keeping vacuum

Parameter „n“,see annexe KV_202

Pressure at the point B10

kPa see annexe KV_202

Pressure at the point B110

kPa see annexe KV_202

Pressure at the point B16

kPa see annexe KV_202

Time T20 s see annexe KV_202Number of 2.evacuations

Parameter „m“,see annexe KV_202

Pressure at the point B11

kPa see annexe KV_202

Pressure at the point B111

kPa see annexe KV_202

Time T21 s see annexe KV_202Pressure at the point B15

kPa see annexe KV_202

Pressure at the point B17

kPa see annexe KV_202

Time T22 s see annexe KV_202Increase speed of J - pressure

kPa Pressure increase, speed controlling

Increase speed of J - time

s Time increase, speed controlling

Increase speed of D - pressure

kPa Pressure increase, speed controlling

Increase speed of D - time

s Time increase, speed controlling

Increase speed of H - pressure

kPa Pressure increase, speed controlling

Increase speed of H - time

s Time increase, speed controlling

Parallel stairs upwards-time steam

s Not used

Parallel stairs upwards-time pause

s Not used

Time stabilizer s Timeout before the sterilization exposure

Period of exhaust of condensate heating - pause

s Exhaust of condensate by opening the valve Y20 during the heating

Period of exhaust of condensate heating-outlet

0,1s Exhaust of condensate by opening the valve Y20 during the heating

Period of exhaust of condensate sterilization - pause

s Exhaust of condensate by opening the valve Y20 during the exposure

Period of exhaust of condensate heating-outlet

0,1s Exhaust of condensate by opening the valve Y20 during the exposure

SELECTOMAT SL_ns_en 0911_V2.02 17

4.4 DESCRIPTION OF THE STERILIZATION PROGRAMME PARAMETERS

The annexe KV_202 shows the general image of a standard sterilization cycle. The cycle comprises the following phases:

0 -9 Initial state, tests, preparation10 Pressurization of the generator11 Pressurization of the jackets12 Door sealing13,14 1st evacuation - exhaustion15 1st evacuation - keeping vacuum16 1st evacuation - steam filling17 1st evacuation - pressure kept up18 2nd evacuation - exhaustion19 2nd evacuation - keeping vacuum 20 … 2nd evacuation - steam filling…63,64 Steaming - steam filling65… Steaming - steam discharging…93… Heating…98 Sterilization99,100 Pressure release normal programs101 Drying - exhaustion102 Drying - pressure kept down103 Drying - aeration104… Drying - pressure kept up…140,141 Annexeary aeration142 Door unsealing

Note:At the customer´s request some devices can dispose with special programs with a partially different course. For example: Solutions program with self-cooling, solutions program with forced cooling etc.

4.4.1 SENSE OF INDIVIDUAL PARAMETERS Individual parameters of sterilisation programmes are described in the following table, as they are displayed in the UNICONFIG programme for the classical non-solution cycle.Parameter of sterilization program

Units Description

Description The name of the program as it will displayed on the display and will be printed in the printer

Change-over switch of sterilization subtype

Change-over switch of the sterilization cycle type – standard/BD

Switch Type of controlling PT

Adjustment of the temperature controlling sensor, see below

Switch Fo Switching of cycle controlling according to the Fo, see below

Switch Type of drying

Not used

Switch Air detector

Turns on the air detector

Switch Bacteriological filter

Turns on the bacteriological filter

Switch Door logic

0 - through chamber, 1- non through chamber

Number of steaming extrusions

Number of pressure lifts in the steaming phase

Page 20: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

18 SELECTOMAT SL_ns_en 0911_V2.02

Service instructions

Decrease speed downwards of G-pressure

kPa Controlled evacuation

Decrease speed downwards of G-time

s Controlled evacuation

Rychlost Decrease speed downwards of A-pressure

kPa Controlled evacuation

Decrease speed downwards of A-time

s Controlled evacuation

Vacuum pump switching pressure

kPa see annexe KV_202

Jacket pressure I before sterilization

kPa Lower limit of the jacket pressure regulation band in the program phases before sterilization

Jacket pressure II before sterilization

kPa Lower limit of the jacket pressure regulation band in the program phases before sterilization

Control jacket hysteresis

kPa Control jacket hysteresis in all phases of the cycle

Jacket pressure I after sterilization

kPa Lower limit of the jacket pressure regulation band in the program phases after finishing sterilization

Jacket pressure II after sterilization

kPa Lower limit of the jacket pressure regulation band in the program phases after finishing sterilization

Delta of jacket I

kPa Pressure difference between the chamber pressure and the lower limit of the jacket pressure regulation band = parallel working of the chamber and jacket pressure. Chamber pressure is primary, jacket pressure is adapting by the constant Delta P, see annexe KV_202

Delta of jacketII kPa Pressure difference between the chamber pressure and the lower limit of the jacket pressure regulation band = parallel working of the chamber and jacket pressure. Chamber pressure is primary, jacket pressure is adapting by the constant Delta P, see annexe KV_202

Delta of jacket I-S

kPa see annexe KV_202

Delta of jacket II-S

kPa see annexe KV_202

Number of drying cycles

Parametr „k“, see annexe KV_202

Pressure at the point B18

kPa see annexe KV_202

Pressure at the point B19

kPa see annexe KV_202

Time of drying - down

min. Time t5, see annexe KV_202

Time of drying - up

s Time t5, see annexe KV_202

Maximum time of pressure increase in steam generator

s If the phase lasts longer, a fault occurs

Maximum time of pressure increase in the jacket

s If the phase lasts longer, a fault occurs

Maximum time of filling the groove

s If the phase lasts longer, a fault occurs. Function treats the filling and evacuation of the groove.

Maximum time of evacuation

s If the evacuation (pressure down, „n“ teeth, pressure up) lasts longer, a fault occurs

Maximum time of steaming

s If the steaming lasts longer, a fault occurs

Maximum time of heating

s If the heating lasts longer, a fault occurs

Maximum time of pressure releasing into atmosphere

s If pressure releasing lasts longer, a fault occurs.

Maximum time of drying - down

s If pressure releasing lasts longer, a fault occurs.

Maximum time of drying up

s If aeration lasts longer, a fault occurs

Note: If minimum or maximum time for checking the time limits is set null, the function of checking the respective limit is suppressed.

4.4.2 USE OF PT SENSORS ACCORDING TO THE UNICONFIG DATA

Classical non-solution cycleThe equipment may have the accessible PT3, if it is configured as solution

Type of controlling PT (UNICONFIG))

Fo Switch (UNICONFIG)

Cycle controlling from PT

Fo Integration

from PT

Note

0 0 PT1 - Standard0 1 PT1 PT31 0 PT3 - 1 1 PT3 PT32 0 PT1 - 2 1 PT1 PT1

Cycle of SolutionsType of

controlling PT (UNICONFIG)

Fo Switch (UNICONFIG)

Cycle controlling from PT

Fo Integration

from PT

Note

0 0 PT3 - Standard0 1 PT3 PT31 0 PT1 - 1 1 PT1 PT1

Page 21: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

4.5 VACUUM TESTThe vacuum test is a special type of the sterilization program. Its course is drawn in Fig.5.

Pressure in the chamber

Atm. pressure

Phase 1 Phase 2

Testing pressure

Prepara- Testing time tory time

Permitted change in pressure

O Abs.

Fig.5.The definition of vacuum test parameters is performed analogically to the definition of a standard sterilization programme by means of the software UNICONFIG.EXE.

Vacuum test parameter

Unit. Description

Description The name of the program as it will displayed on the display and will be printed in the printer

Evacuation pressure

kPa Pressure at the beginning of the 1st phase of vacuum test

Time of phase 1 VT

min Duration of the 1st phase of vacuum test

Time of phase 2 VT

min Duration of the 2st phase of vacuum test

Pressure releasing time of the groove

s Time of pressure releasing and exhausting the groove of the door seal

Permissible pressure change during the VT

kPa Permissible pressure change during the vacuum test; surpassing this limit during the test results in „VT not passed“.

Maximum time of pressurization of the generator

s If the pressure obtaining in the generator has not been finished within the determined time limit, an error occurs.

Maximum time of pressurization of the jacket

s If the pressure obtaining in the jacket has not been finished within the determined time limit, an error occurs.

Maximum time of filling the groove

s If filling of the groove has not been finished within the maximum time limit, a fault occurs.

Maximum time of evacuation

s If evacuation lasts longer, a fault occurs.

Approximate lenghts of the cycles for individual tests according to the standard EN285:P1 with metal charge cca 14 min.P2 with a small textile charge cca 35 min.P3 with a full textile charge cca 65 min.P4 with a rubber charge cca 45 min.VT cca 18 minBD test cca 18 min.

4.6 AIR DETECTOR The Air detector itself is within the diagram of the piping connection of the device. To the Air detector there belongs a pit with the temperature sensor PT6, valve Y27.Air detector works automatically and in case of possible exceeding the limit value of the non-condensed gasses in the sterilizer chamber, it will report the appropriate error reporting, which interrupts the sterilization cycle.(A condition for the Air detector functionality, similarly as in case of other additives, is the configuration in corresponding service.)

In the service of the Air Detector it is possible to set:• Sterilization programs, within which the Air detector is

active. • Modes of the Air detector

Working mode – principal function mode Testing mode without error – it is the auxiliary testing mode, when during the started sterilization cycle there is open the auxiliary valve Y28. Within this mode into the chamber it is sucked a defined amount of the additional air via the hand adjustable butterfly valve V100 as to verify the proper function of the Air detector.Within this mode the Air detector does not report the appropriate error reporting so as the cycle could continue and it would be possible to assess for example the influence of the air on the temperature distribution in the testing packet.Testing mode with error – it is the auxiliary testing mode, when during the started sterilization cycle there is opened the auxiliary valve Y28. Within this mode into the chamber it is sucked a defined amount of the additional air via the hand adjustable butterfly valve V100.

• Setting of the Air detector – setting of the air penetrationClose the door of the sterilizer.

Service instructions

SELECTOMAT SL_ns_en 0911_V2.02 19

Page 22: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Push the Start button.

The sterilizer performs the evacuation and after a short stabilizing pause it is measured the increase of pressure in the chamber in dependence on the setting of the butterfly (vacuum breaker) valve V100.The pressure increase is continuously displayed in the shape: momentary value/ average value (kPa/min).In case of need it is possible to repeat the measuring cycle by repeated pushing the button Start.The initial value for the setting of the air penetration is 1 kPa/min.The total setting of the Air detector, see EN 285.Additive - the vacuum breaking valve for the setting of the air penetration via the valve Y28 and the butterfly valve V100 has order number: R469696. This additive serves only as the auxiliary device for more comfortable setting and thus it is not necessary for the function itself of the Air detector.The additive of the vacuum breaking valve (R469696) is connected into the orifice„VT“ of the sterilization chamber.Note: For start of this auxiliary program for measuring (setting) of the pressure increase it is necessary to have downloaded the appropriate program in the device. (Air Detector.ad1).• Assessment temperature – enables to set the decisive

temperature of the Air detector.The increase of the decisive temperature it is increased the sensitivity of the Air detector and vice versa (orientation range 50 -100 °C).

Service instructions

20 SELECTOMAT SL_ns_en 0911_V2.02

Page 23: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

SUPPLEMENTS:

Set the printer in the following way:

SETTING THE PRINTER PARAMETERS:GPT- 4454 - V.24-DC/DC-RS232 M.NO. 0338421

Press simultaneously the buttons for paper feed ◄ / ENTER / and the button marked ■ /NEXT/. The printer enters the setting mode and prints information about the internal voltage of the printer (approx. 7, 8 V), firmware, actual parameters setting, and procedure for these parameters modification.

The correct parameters setting for SELECTOMAT SL is as follows:

Density: 30Speed/Quality: med(104)/lowInterface: RS232/USB/BlueCom: 115200, n, 8, Tx+Sleep time: OFFFont #: 1Char.format: D0, W0, H0, S0, 104

If any of the parameters differs, it can be modified in the following way:

1. Press the ◄ button – the first modifiable parameter is printed.2. By repeated pressing of the ◄ button, set the given parameter to the required value.3. Go to the next parameter by pressing the ■ button.4. Repeat points 2 and 3 until all parameters are set correctly.5. To terminate the setting procedure, press the buttons ◄ and ■ simultaneously.6. Connect the printer to the device and print the check copy.

Note:The Com parameter shall be set in two steps: Baudrate: 115200 and Com-params: n, 8, Tx+. The Char. Format parameter shall be set in three steps: Text orientat: Textmode (D0), Char. Size: W0/H0, and Char. Sparing: 0.

Page 24: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

SELECTOMAT

Page 25: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

1/14

Page 26: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

2/14

Page 27: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

3/14

Page 28: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

4/14

Page 29: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

5/14

Page 30: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

6/14

BOARD VP_KON…

CH

IP C

AR

D

SER

VO

RS2

32

ETH

ERN

ET

VP_

CPU

S- X

S1

USB

SWIT

CH

BO

AR

D

BOARD VP_CPUM…

PRINTER

INVERTER

TOUCHSCREEN 12,1“

BOARD VP_M-LB121-AT DISPLAY 12,1“

SWIT

CH

BO

AR

D

BLACK

Control panel front (back) side SEL SL

Page 31: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

7/14

BOARD VP_KON…

CH

IP C

AR

D

SER

VO

RS2

32

ETH

ERN

ET

VP_

CPU

S- X

S1

USB

SWIT

CH

BO

AR

D

BOARD VP_CPUM…

PRINTER

INVERTER

SWIT

CH

BO

AR

D

BLACK

DISPLAY 8,4“ BOARD VP_LVDS…

Control panel front side SEL SL – for VP_CPUM (for VP_CPUME page 14)

Page 32: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

8/14

VP_

CPU

B(B

E)-X

S1

VP_

KO

N(C

PUM

E)-X

S1

24V AC

24V DC

+24V

BOARD VP_CPUS…

1AC BOARD VP_PAC…

2AC BOARD VP_PAC…

3AC BOARD VP_PAC…

2DC BOARD VP_PDC…

1DC BOARD VP_PDC…

RS 232

3DC BOARD VP_PDC…

Page 33: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

9/14

BOARD VP_CPUB…

DISPAY 5,7“ TOUCHSCREEN 5,7“

INV

ERTE

R

SWITCHBOARD

VP_CPUS-XS2

Control panel back side SEL SL

Page 34: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

10/14

BOARD VP_CPUBE…

SWITCHBOARD

VP_CPUS-XS2

Control panel back side SEL SL

Page 35: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

11/14

* For single door ** For independently sealed door *** For steam generator FDD

Page 36: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

12/14

* Y80,Y80.1 - AC

Page 37: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

13/14

* For independently sealed door ** For 9xx *** Y80,Y80.1 – DC (air)

Page 38: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Drew Pleva Date 20.09.2009 CONNECTION SELECTOMAT Version 17 Revision 09/758 Z468466 Page/pages

14/14

Control panel front side SEL SL – for VP_CPUME

PRINTER

RS232

ETHERNET

USB

VP_CPUS-XS1

SWITCHBOARD

BOARD VP_CPUME…

DISPLAY 8,4“

INV

ERTE

R F

OR

CFL

CH

IPC

AR

D

SWITCHBOARD

SWITCH FOR BIOS

Page 39: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

ERROR REPORTING – SELECTOMAT SL If on the display there repeatedly appears the error reporting, it is necessary, after setting the password for the error confirmation, to switch off and again switch on the electricity supply. If the error continues to exist, please, proceed according to the following table.

After pressing the button the service technician can suppress the errors for up to 999 seconds (repeated pressing).

Error Description Debugging WhoFatal error – see “System versions”

Incompatible SW or HW 1. Record correct SW into all the electronics boards

2. Replace the faulty electronics board

Service

Discharged Battery The voltage of the battery decreased below 2,7 V

Change the battery Service

Faulty configuration of the device

After recording the bios, or after replacing the boards VP_CPUM or VP_CPUS

Set the correct configuration of the device Service

Faulty offsets of analog values

After recording the bios, or after replacing the boards VP_CPUM or VP_CPUS

Set the correct offsets Service

There is not any connection between the boards MST and Slv (Bck)

Faulty communication between VP_CPUM and VP_CPUS or VP_CPUB

1. Control the cables for RS 485 2. Control SW in all the electronics boards 3. Replace the faulty electronics board

Service

Pressure switch of generator

High pressure in the generator 1. Control the switching of the heating 2. Control the pressure switch B31

Service

Water level in the generator Switch B91 is not flooded with water for longer period than 250 sec.

1. Control the demi water supply 2. Control B913. Control the VP_CPUS electronics board

Operator/Service

Door is not sealed Door is not sealed 1. Control the door sealing 2. Control the valve Y09 3. Control the pressure switch B20

Operator/Service

Door 1 is not sealed The door on the loading side is not sealed

1. Control the door sealing 2. Control the valve Y09 3. Control the pressure switch B20

Operator/Service

Door 2 is not sealed The door on the unloading side is not sealed

1. Control the door sealing 2. Control the valve Y09 (Y09.1) 3. Control the pressure switch B20 (B20.1)

Operator/Service

High Steam Generator Temperature

High Steam Generator Temperature

1. Control the pressure sensor PE32. Control the temperature sensor PT43. Control the printed circuit board VP_CPUS4. Control the contactors K4-7

Service

High temperature in the chamber

High temperature in the chamber

1. Control the door sealing 2. Control the inlet steam 3. Control the valve Y074. Control temperature and pressure sensors

Operator/Service

Low temperature in the chamber

Low temperature in the chamber

1. Perform the vacuum test. Start again the cycle

2. Remove untightness 3. Control the condensate trap 4. Control the process parameters 5. Control the temperature and pressure

sensors

Operator/Service

“Name phase” – Long phase

The specified phase of the sterilization cycle is longer that it is permitted

1. Control the input media 2. Control the inputs and outputs

Operator/Service

The air in the chamber – inconvenient

The quantity of gases unable to condensate is above the limit

1. Control the sealing2. Control the tightness of joints

Operator/Service

Voltage 24 V The voltage 24V decreased below 17V

1. Control the fuse F42. Control the door seal3. Control the door motors4. Control the pressure sensors5. Control the printed circuit board VP_CPUS and VP_PDC

Service

PE12 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE12 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

Page 40: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

PE101 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE101 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE102 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE102 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE103 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE103 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE104 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE104 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE11 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE11 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE2 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE2 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE20 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE20 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE3 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PE3 Pressure sensor is above limit

Pressure sensor is above limit 1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT11 Pressure sensor is disconnected

Pressure sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT12 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT32 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT2 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT2 Temperature sensor is short-circuited

Temperature sensor is short-circuited

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT31 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT4 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT4 Temperature sensor is short-circuited

Temperature sensor is short-circuited

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT5 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT5 Temperature sensor is short-circuited

Temperature sensor is short-circuited

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT6 Temperature sensor is disconnected

Temperature sensor is disconnected

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

PT6 Temperature sensor is short-circuited

Temperature sensor is short-circuited

1. Control the sensor and its connection 2. Control the printed circuit board VP_CPUS

Service

Page 41: SELECTOMAT SL - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/...9612 63 kW 9618 72 kW The steam generator is made from chrome-nickel-molybdenum stainless

Manufacturer: Distributor:

MMM Münchener Medizin Mechanik GmbHSemmelweisstrasse 6 D-82152 Planeggtel.: +49 089/89918-0, fax: +49 089/89918-118http://www.mmmgroup.com

BMT Medical Technology s.r.o, Cejl 50, CZ 656 60 Brno tel.: +420 545 537 111 fax: +420 545 211 750 e-mail: [email protected] http://www.bmt.cz