Section 11 Advanced Sampling

download Section 11 Advanced Sampling

of 23

Transcript of Section 11 Advanced Sampling

  • 8/10/2019 Section 11 Advanced Sampling

    1/23

  • 8/10/2019 Section 11 Advanced Sampling

    2/23

    Advanced Sampling Page 1

    REV C 134515

    I. nX Vector

    A) nX Vector Configuration

    In addition to 1X and 2X tracking filters, you can use nX Vector to track auser configured third component. N can have nonintegral values between

    0.100 and 1.0 (in intervals of 0.025) or integral values between 1 and 15.

    One nX Vector is configurable for each DAIU. Typical applications of the

    nX vector are to track a particular frequency such as 0.5X from a rub, or a

    higher order of running speed like 5X from vane passage frequency on a

    pump. Differential Phase also uses the nX vector (see below).

    Note: In order to acquire nX values below 0.200, a 208 firmware version of

    1.02 is required.

    You must set n in nX before you begin data acquisition. Configuration of

    the nX Vector is done from the main Configuration dialog box.

    Figure 1. Configuring the nX Vector

  • 8/10/2019 Section 11 Advanced Sampling

    3/23

    Page 2 ADRE for Windows

    REV C 134515

    B) Using nX With Plots

    The nX Vector can be selected as the main variable in many plot

    configurations. The example below shows an nX Trend. If the nX Vector

    is an integer order of running speed (5X for example), nX phase angle willbe available for plotting along with amplitude.

    Figure 2. nX Trend of 5X Component

    Figure 3. Selecting nX Slow Roll Vector Compensation

  • 8/10/2019 Section 11 Advanced Sampling

    4/23

    Advanced Sampling Page 3

    REV C 134515

    C) nX Vector Slow Roll Compensation

    You can select Slow Roll compensation on most plots that have nX as a

    main variable option. Be sure that you have selected nX Vectors from the

    Edit Reference - Slow Roll Vector dialog box (see Section 6, EditingReference Data). Keep in mind that nX Slow Roll vector subtraction can be

    performed only if the nX vector is an integer order of running speed.

    II. Differential Phase

    A) Using Differential Phase

    Differential Phase allows relative phase measurement of nX vectors when

    one of the channels is chosen as a reference. The system subtracts the

    absolute nX phase lag angle of the reference from the absolute nX phase lag

    angles of all the channels and then displays the difference in tabular list and

    current values plots on a sample-by-sample basis. The reference has no

    effect on other frequency components of real-time or saved data.

    Though the nX Vector must be selected prior to sampling, Differential

    Phase can be invoked either before or after sampling. Differential Phase is

    configured from the main Configuration dialog box.

    Figure 4. Configuring Differential Phase

  • 8/10/2019 Section 11 Advanced Sampling

    5/23

    Page 4 ADRE for Windows

    REV C 134515

    B) Differential Phase Configuration

    To configure differential phase:

    1) Choose DAIU 1 from the DAIU group box. The "n" value of thenX vector will be displayed in the text boxes of this group, as

    specified in nX Vector Configuration.

    2) Choose a Reference channel from channels 1 through 8.

    3) Repeat Steps 1 and 2 for DAIU 2 and Channels 9 through 16 if you

    are using a second DAIU.

    4) Click Enable.

    5) Press OK to confirm your settings and return to Configuration

    dialog box. Or if you prefer, use one of the Go To push buttons

    to exit; they will also confirm your differential phase configuration.

    ! T a b u l a r L i s t - D i f fe r e n t ia l P h a s e D is a b le d

    ! T a b u l a r L i s t - D if fe r e n t ia l P h a s e E n a b le d

    n X V e c to r

    n X V e c to r

    R e fe r e n ce C h a n n e l

    Figure 5. Differential Phase Example in Tabular List

  • 8/10/2019 Section 11 Advanced Sampling

    6/23

    Advanced Sampling Page 5

    REV C 134515

    Three rules apply to Differential Phase channels:

    1. The reference channel must be an active channel. (See Channel

    Configuration.)

    2. Select only one reference channel from each active DAIU.

    3. The transducer for the selected channel cannot be a process

    variable transducer.

    Note: You can choose a channel that is associated with a Simulated

    Keyphasor as a reference even though phase lag angle data from a

    Simulated Keyphasor is unreliable. This is true because relative and not

    absolute phase lag angles are subtracted from the data.

    Use Disable to temporarily disable your Differential Phase setup. By

    disabling the setup you can view data without using a reference channel.

    III. Zoom Channels

    A Using Zoom Channels

    When enabled, Zoom Channels configures up to two channels per DAIU for

    high resolution (3200 line) spectrums. Zoom Channels must be configured

    prior to sampling.

    The system takes longer to collect data for Zoom Channels than for

    channels receiving data at normal resolutions. Use this formula to determine

    the time it will take to collect data:

    Time (sec) = 8192 / (FS X 2.56)

    where FS is the frequency span. For example, if you set a frequency span of

    50 Hz, the DAIU will take more than a minute to collect data in zoom mode.

  • 8/10/2019 Section 11 Advanced Sampling

    7/23

    Page 6 ADRE for Windows

    REV C 134515

    Figure 6. Half Spectrum Plot with Zoom Channel Enabled

    B) Zoom Channels Configuration

    Any channel can be configured as a zoom channel. The Zoom Channels

    configuration is accessed through the main Configuration dialog box. Toconfigure zoom channels:

    1) Find the channel you want to configure as a zoom channel using

    the scrollbar if necessary.

    2) Pick the checkbox of the channel(s) you want to configure as a

    zoom channel.

    !High Resolution 3200-Line Spectrums

    !(2) Zoom Channels / DAIU

    !(1) Zoom Channel / Channel Pair

  • 8/10/2019 Section 11 Advanced Sampling

    8/23

    Advanced Sampling Page 7

    REV C 134515

    Figure 7. Zoom Channels Configuration

    Zoom channel configuration must conform to these three rules:

    1) Only active channels can be zoom channels.

    2) Each DAIU can have two channels configured as zoom channels,

    so you can have up to four zoom channels if you are using two

    DAIUs.

    3) Only one of the two channels in a channel pair can be configured

    as a zoom channel. So if channel 3 is a zoom channel, channel 4

    cannot be.

    Caution: When zoom channels are enabled, spectrum plots can only be

    displayed for the channels configured as zoom channels. Also,synchronous waveform plots cannot display direct data for zoom

    channels. And, if only one zoom channel is configured with the

    first keyphasor association, the second keyphasor association will

    not collect any direct waveform data.

  • 8/10/2019 Section 11 Advanced Sampling

    9/23

    Page 8 ADRE for Windows

    REV C 134515

    C) Enabling and Disabling Zoom Channels

    Use Disable to deactivate all zoom settings without deleting them. With

    zoom channels deactivated, the system collects data at the standard

    resolution set in Spectrum Sample Rate Configuration. You can reactivatezoom settings by pressing Enable Zoom.

    IV. Trigger Event Sampling

    A) Trigger Event Options

    Use this dialog box to choose a trigger source and trigger threshold. A

    trigger source is the type of event that starts data collection. A trigger

    threshold is the level of the trigger source that starts the data collection.

    Once a trigger event threshold requirement has been met, data acquisitionwill start and will continue according to the sampling scheme set in Sample

    Mode configuration. There are five options which can be configured as a

    trigger event. Only one trigger event can be configured at a time to

    automatically start the data collection process in Store Enable. There is the

    option to manually trigger (start) the data collection when you are in the

    Store Enable menu.

    Figure 8. The Trigger Event Configuration Dialog Box

  • 8/10/2019 Section 11 Advanced Sampling

    10/23

    Advanced Sampling Page 9

    REV C 134515

    The most often used trigger source is Speed. The default setting of Over

    100 and Under 60,000 initiates sampling as soon as the rpm of

    Keyphasor 1 is between 100 rpm and 60,000 rpm after entering Store

    Enable.

    Over and Under setpoints of any trigger event can be configured and

    saved with a database configuration for later use even if they are not

    enabled.

    B) Trigger On Speed Configuration

    Select Speed as the trigger source and select Over Setpoint, Under

    Setpoint, or both. Press the Configure button to bring up the Trigger

    On Speed Configuration dialog box. Enter a level as an integer between

    100 and 60,000 (rpm) in the Over edit box, the Under edit box, or both.

    Sampling will begin when machine speed moves above or below the

    enabled setpoints.

    Figure 9. Configuration Example for Trigger on Speed

  • 8/10/2019 Section 11 Advanced Sampling

    11/23

    Page 10 ADRE for Windows

    REV C 134515

    C) Trigger on Time Configuration

    To start collecting data at a preset date and time, select Time as the trigger

    source and press the Configure button to bring up the Trigger on Time

    Configuration dialog box. Enter the date and time you wish to start dataacquisition in these six edit boxes. The time must be a future event. You

    will receive an error message if the start time you select is earlier than the

    computer clock time.

    Figure 10. Trigger On Time Configuration Example

    The settings for Trigger on Time must be within these ranges:

    Hour: 0-23

    Minute: 0-59

    Second: 0-59

    Day: 1-31

    Month: 1-12

    Year: 1980-2037

  • 8/10/2019 Section 11 Advanced Sampling

    12/23

    Advanced Sampling Page 11

    REV C 134515

    D) Trigger On Amplitude Configuration

    Use this dialog box to start collecting data when the amplitude of a trigger

    moves above or below a set amplitude. Each active channel can have one or

    two trigger thresholds or setpoints. Over specifies an amplitude abovewhich data acquisition will begin and Under specifies an amplitude below

    which data acquisition will begin. "Over" and Under" edit boxes are

    available for each of the 16 channels. The second group of eight channels is

    accessible by using the scrollbar.

    Select Amplitude as the trigger source, then select "Under setpoint"

    and/or "Over setpoint" check boxes in the Trigger On group. Press the

    Configure push button to bring up the Trigger on Amplitude

    Configuration dialog box.

    Figure 11.Trigger On Amplitude Configuration Example

  • 8/10/2019 Section 11 Advanced Sampling

    13/23

    Page 12 ADRE for Windows

    REV C 134515

    To configure trigger on amplitude:

    1) Select the type of amplitude from the Amplitude Trigger Source

    group: Direct, 1X, 2X, nX, or DC Gap/Process

    Variable. The n in nX was specified in nX VectorConfiguration.

    2) Check the Trigger checkbox for the channels desired and enter

    over and under setpoints in the selected channel edit boxes.

    The following rules apply to trigger on amplitude:

    a) TorXimitor and Process Variable transducers cannot trigger

    on Direct, 1X, 2X, or nX.

    b) Acceptable transducer types for DC Gap/Process Variable in

    the Amplitude Trigger Source group are displacement,

    TorXimitor, or Process Variable.

    c) The Under and Over set points must lie within

    configured Full Scale Ranges for each channel and variable.

  • 8/10/2019 Section 11 Advanced Sampling

    14/23

    Advanced Sampling Page 13

    REV C 134515

    E) Trigger On Phase Configuration

    Use this dialog box to start collecting data when the phase lag angle moves

    outside the Over and Under setpoints. Each active channel can use

    either or both phase lag angle setpoints. Over specifies a phase abovewhich data acquisition will begin and Under specifies a phase below

    which data acquisition will begin. "Over" and "Under" values are available

    for each of the 16 channels. The second group of eight channels is

    accessible by using the scrollbar.

    Figure 12. Trigger On Phase Configuration Example

  • 8/10/2019 Section 11 Advanced Sampling

    15/23

    Page 14 ADRE for Windows

    REV C 134515

    To configure trigger on phase:

    1) Select Phase as the trigger source.

    2) Check either or both of the Over Setpoint and Under Setpointcheckboxes.

    3) Press the Configure push button to bring up the Trigger On

    Phase Configuration dialog box. Choose 1X, 2X, or nX

    from the Trigger Source group. Note: Specify the "n" in nX Vector

    configuration; n must be an integer multiple of running speed to

    have an absolute phase angle.

    4) Check the Trigger checkbox for one or more channels and

    specify a range in degrees for each selected channel. Sampling will

    begin when the phase angle for a particular channel goes outside its

    range. Minimum and maximum values for setpoint edit boxes are

    0 and 359.

    A channel may not be available if the channel associated with the checkbox

    is not active (see Channel Configuration), or the channel associated with

    the checkbox was one you assigned a Simulated Keyphasor during Channel

    Configuration (an external Keyphasor is required for absolute phase

    measurements).

  • 8/10/2019 Section 11 Advanced Sampling

    16/23

    Advanced Sampling Page 15

    REV C 134515

    F) Trigger On Contact Configuration

    Figure 13. Trigger On Contact Configuration

    Use this dialog box to use remote relays connected to the DAIU to start

    collecting data. Each DAIU has a terminal block with four connections that

    can be connected to a remote relay: Earth Gnd, Lo Volt, Return, and Hi

    Volt. The Lo Volt handles 5V to 30V. The Hi Volt handles 90V to 240V.

    WARNING: Do not put high voltage in the low voltage circuit.

    To use a remote relay to signal data collection:

    1) Select Contact as the trigger source, then press the Configure

    command button to bring up the Trigger On Contact

    Configuration dialog box.

    2) Select Normally Closed if you want data acquisition to startwhen the relay is open, or:

    Select Normally Open if you want data acquisition to start when the

    relay is closed.

  • 8/10/2019 Section 11 Advanced Sampling

    17/23

    Page 16 ADRE for Windows

    REV C 134515

    V. Sample Management

    A) Manual Samples

    Although you have a buffer of 2560 vector samples (256 waveformsamples) to work with, it is good practice to have extra samples available

    for manual sampling if needed. You never know when a sudden upset

    might occur which you will want to document. If you are sampling at

    regular delta time intervals, a few manual samples may be needed to ensure

    such an intermittent event gets captured.

    Note: Versions of ADRE for Windows before 4.60 have a maximum of

    1280 vector samples (128 waveform samples) to work with instead of 2560.

    Make sure you know which version you are using before assuming your

    limit. Click Help from the main menu bar, then About Adre to

    double check which version you are using.

    Though each manual sample uses only one vector sample for each

    waveform sample, the number of samples you set aside should still be based

    on a 10:1 ratio of vector to waveform samples:

    Manual Samples (Vectors) = Number of Waveform Samples x 10

    Example:You want to set aside 30 manual samples for direct waveforms and

    spectrums.

    Solution:

    Manual Samples = 30 Waveform Samples x 10

    = 300 Vector Samples

    2560 - 300 = 2260

    Set Sample Mode parameters to use no more than 2260 Vector Samplesduring normal programmed (delta time/delta rpm) data acquisition.

  • 8/10/2019 Section 11 Advanced Sampling

    18/23

    Advanced Sampling Page 17

    REV C 134515

    Note: Normally you will not need to use more than 400 samples or so for a

    typical data acquisition exercise (see Data Acquisition Configuration

    in Section 5). If Sample Mode parameters are set conservatively to

    acquire data with a reasonable sample density (approximately 400

    samples for a startup or shutdown), there will be plenty of unusedsamples available for manual samples if needed.

    B) Vector / Waveform Sample Ratio

    Manual samples will skew the normal 10:1 ratio of vector samples to

    waveform samples to a smaller ratio. When this happens, care must be

    taken to make sure there are enough samples remaining to capture the data

    you need. Though the DAIU Control Panel displays the number of samples

    both used and remaining, the actual number of usable vector samples

    remaining may be significantly less than the displayed value because

    sampling will stop when all waveform samples have been used. See

    example below. If you have turned on Continuous Data Collection in the

    Preferences, the data will be saved to disk when a sample limit is reached

    and then ADRE!for Windows will recommence acquiring data. However,

    the computer is not acquiring data while it is saving the run to disk so you

    will have a hole in your data base and the data run will be split between

    different databases.

    Two good rules of thumb to remember when using Manual Samples are:

    1) Usable Vector Samples Remaining = Waveform Samples

    Remaining x 10

    2) Set Sample Mode Parameters Conservatively to start with!!

  • 8/10/2019 Section 11 Advanced Sampling

    19/23

    Page 18 ADRE for Windows

    REV C 134515

    Figure 14. Sample Limits Dialog Box Showing Depleted Waveform Samples and

    End of Sampling

    C) Sample Mode Adjustments During Data Acquisition

    Sometimes you may need to make adjustments to your sample mode

    parameters during sampling to conserve remaining samples. You can make

    the following changes from the DAIU Control Panel:

    Delta Time

    1) Enable/Disable

    2) Adjust Delta Time interval

    Delta RPM

    1) Enable/Disable2) Turn On or Off Increasing or Decreasing

    3) Adjust Delta RPM interval

    Keep in mind that sampling will be paused while you are making these

    changes.

    !!!!DONT LET THIS HAPPEN TO YOU!!

    !!!!Set Sample Mode Parameters CONSERVATIVELY

  • 8/10/2019 Section 11 Advanced Sampling

    20/23

    Advanced Sampling Page 19

    REV C 134515

    VI. Non-Standard Data Acquisition

    A) Steady State Trending + Trip / Coastdown Coverage

    The following example is one possible scenario where you may employboth Delta RPM and Delta Time simultaneously.

    Example:

    Lets say you need to automatically collect trend data overnight (7:00

    PM to 7:00 AM) on a 12,000 rpm compressor. You want to

    automatically capture a coastdown if the machine train is somehow

    tripped any time during the night. Assume no manual samples.

    Solution:

    In Sample Mode:

    - Enable Delta Time andDelta RPM

    - 600 Sample records dedicated to potential shutdown

    - 1960 Sample records dedicated to steady state trend

    Delta RPM =(12,000rpm-100rpm) / 600 =19.8 rpm

    Monitoring Time = 12 hours x 60 minutes/hour = 720 minutes

    Delta Time = 720 minutes / 1960 records = 0.37 minutes/record =23 seconds

    Be conservative:

    - Set Delta RPM 25 rpm Decreasing only

    - Set Delta Time 30 seconds

  • 8/10/2019 Section 11 Advanced Sampling

    21/23

    Page 20 ADRE for Windows

    REV C 134515

    B) Speed Oscillations

    For startups on machine trains where speed fluctuations are common, it is

    easy to use up all 2560 samples prematurely because of the total Delta Rpm

    that takes place. There are several different strategies to use to avoid usingup your samples this way:

    1) Set Delta Rpm for Increasing only.

    - or -

    2) Set a large Delta Rpm interval and use both Increasing and

    Decreasing.

    - or -

    3) Set Delta Rpm for Increasing only with a conservative Delta

    Rpm interval. Set Delta Time with a conservative (long) interval

    so that occasional data will be taken during hold points.

    C) Fast Startups/Shutdowns

    The most frequent sampling problems when taking data during rapid

    machine speed changes are tracking filter response time and spectrum

    sampling speed. A slow tracking filter may miss the full extent of amplitude

    and phase changes if the changes occur quickly. A high resolution spectrum

    may not have enough time at a particular speed to accurately sample a

    spectrum. Sampling speed can usually be improved, but the tradeoff ispoorer resolution. Still, a poorer resolution spectrum is better than no

    spectrum at all. To speed up tracking filter response time and spectrum

    sampling:

    1) Use the 120 rpm bandwidth filter (set in Channel Configuration).

    This is a much faster tracking filter than 12 rpm bandwidth.

    2) Use Delta RPM only. Delta RPM sampling is potentially much

    faster than even the smallest Delta Time interval (0.1 seconds).

    3) Do not use Zoom Channels.

    4) Set Spectral Lines to 200 or even 100 to speed up spectrum

    sampling.

    5) Increase the Spectrum Frequency Span setting.

  • 8/10/2019 Section 11 Advanced Sampling

    22/23

    Advanced Sampling Page 21

    REV C 134515

    D) Repeating Data Collection

    You can configure ADRE to continue collecting data even after a database

    fills up. From the File menu, select Preferences to bring up the

    Preferences dialog box.

    Figure 15. Preferences dialog box

    Select the Repeating Data Collection checkbox to configure the system to

    continue collecting data after the first database fills up and is saved to disk.

    If you select the Reset Trigger checkbox, after a database is filled and

    saved to disk, the system will reset the data acquisition trigger. If the

    machine conditions meet the triggered criteria that you set, data collection

    will resume automatically. If the machine conditions do not meet the

    triggered criteria, ADRE will wait until data acquisition is triggered before

    collecting any more data.

    For example, if you configured your system to trigger when the machinegoes above 350 rpm, to sample with delta time every second, and you

    selected Repeating Data Collection and Reset Trigger, the system will

    begin collecting data when the machine increases above 350 rpm. The

    database will then take a vector sample every second until the database is

    full with 2560 vectors. At this point, ADRE will automatically save the

    database and prepare to continue collecting data. Because you selected

  • 8/10/2019 Section 11 Advanced Sampling

    23/23

    Page 22 ADRE for Windows

    Reset Trigger, the system will see if the machine is still operating above

    350 rpm. If the machine is above 350 rpm, data acquisition will be

    triggered and ADRE will fill another database and save it. This process will

    repeat over and over until the free space on the hard disk drops below 28.2

    MB or the machine is below 350 rpm at the moment the system resets thetrigger.

    Note: If you select Repeating Data Collection but do not check Reset

    Trigger, once data acquisition is triggered the first time, it will

    continuously collect and save databases (according to the sampling mode

    parameters) until you manually stop it or until the free space on the hard

    drive drops below 28.2 MB. This will occur even if the machine no longer

    meets the triggered criteria.