Search Line Optima Plus

19
Point Infrared Gas Detection Point Infrared Gas Detection Searchpoint Searchpoint OPTIMA OPTIMA The story continues !!! The story continues !!!

Transcript of Search Line Optima Plus

Page 1: Search Line Optima Plus

Point Infrared Gas DetectionPoint Infrared Gas Detection

SearchpointSearchpoint OPTIMA OPTIMAThe story continues !!!The story continues !!!

Page 2: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

2

Document control number

Fixed Point Fixed Point Infra-Red Infra-Red

Gas Detection Gas Detection

Page 3: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

3

Document control number

Improve Searchpoint Optima to:

• Increase lamp lifetime• Reduce zero drift• Improve diagnostics• Eliminate Manufacturing issues

• Resist environment– Rapid temp change– humidity

• Installation issues– Rubber caps– Wiring– Noise– Orientation– Inappropriate applications

Page 4: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

4

Document control number

What has been the aim

• Provide upgrade package:– Further improve reliability of design– Reduce unexplained events– Increase availability– Increase bulb life expectancy

• New Optima will have– Reduced manufacturing cost– Improved documentation– Customer focused features

Page 5: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

5

Document control number

• Hazardous area certification PLUSPLUS– European (ATEX)

• Hardware PLUSPLUS• Accessories PLUSPLUS• Diagnostics PLUSPLUS• Software PLUSPLUS• Signal processing PLUSPLUS

OPTIMA OPTIMA PLUSPLUS

The PLUSPLUS in

Page 6: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

6

Document control number

Hardware PLUSPLUS

• New quartz windowquartz window– increased signal transmission

• New component component interconnectionsinterconnections– increased performance integrity

• Dynamic Heating controlDynamic Heating control– Reduced optical condensation

• Bulb drive circuitBulb drive circuit– extended life expectancy

• Back coated mirrorBack coated mirror– increased corrosion resistance and reliability

• Link selectable current sink / sourceLink selectable current sink / source

Page 7: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

7

Document control number

• New deluge protection / sunshadedeluge protection / sunshade– Thermal shock protection.

• New dust barrierdust barrier– Faster response time, less likely to block.

• New gassing capgassing cap– Simplified, more reliable gassing.

• New optional remote gassing celloptional remote gassing cell– Easy to clean

Accessories PLUSPLUS

Page 8: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

8

Document control number

Enhancements

Upgrade

package

Run bulbs less hard

Replace bulbs

Replace crimps

Deluge / Sunshade protection

Dirty optics warning

Integration Algorithm

Configurable output

Configurable output

Noise removal algorithm

Bulb pre-failure detection

Extended event log

Spike removal algorithm

Page 9: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

9

Document control number

Extra’s offered

Searchpoint Optima II

Increased heating of optics

Bulb current monitoring

Multi-point temperature

compensation -40 to +65°C

Eliminate crimps

Deluge / Sunshade protection

as standard

New Adhesive

Plastic optics block Remove

flow cell as standard

Monitor 4-20mA &Supply voltage

Page 10: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

10

Document control number

Spurious events

• All declared events are logged (D&T)• Detector noise - filter algorithm added• Temperature warning added• Deluge / sunshade protects against extreme

conditions• Low supply voltages generate warnings before fault

Page 11: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

11

Document control number

Service request

• Indicates via the configurable 4-20mA output:– Dirty Optics– Low lamp intensity– Negative zero drift– Temperature out of range– Power supplies out of range (internal and external)– Detector noise

Page 12: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

12

Document control number

Diagnostics PLUSPLUS

• “Obscured optics”• “Negative gas reading”• “Lamp output”• “Internal failure”

• “Temp limit exceeded”– Fixed limits, temp shocktemp shock

• ““Bad 24v supply”Bad 24v supply”• ““Bad 4-20mA loop”Bad 4-20mA loop”• ““Signal quality”Signal quality”• ““Alarm”Alarm”

New diagnostic featuresNew diagnostic features

Maintenance on demandMaintenance on demand

Discreet mA Warning & fault levelDiscreet mA Warning & fault level

Simplified interrogator messagesSimplified interrogator messages

Page 13: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

13

Document control number

Software PLUSPLUS

• Multipoint temp compensationMultipoint temp compensation– 43 points from 2 on Optima

• Configurable mA outputConfigurable mA output– Fault 0-mA

– Inhibit 1-4mA

– Warning 1-4mA

– Over range 20-21.5mA

• Event logEvent log– All warnings, faults & alarms

• Advanced signal processingAdvanced signal processing

Page 14: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

14

Document control number

Lamp PLUSPLUS • Extended lamp life

– Drive current reduced - 98% to 85% (7x)

• Low intensity warning

Page 15: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

15

Document control number

Upgrade PLUSPLUS

• Enhancements– Configurable output– Maintenance on demand– Improved diagnostics– Bulb, noise and flicker

algorithms– Integration algorithms

• Software – CENELEC + UL

Q1 2000 – CSA + DMT (performance)

Q2 2000

• Hardware– not retro-fitable

• Not available– Multi point temp compensation– Current monitoring– Improved heating

Page 16: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

16

Document control number

OPTIMA OPTIMA PLUSPLUS

• Standard Configuration– Optima Plus

• methane, propane• butane, ethane• propylene.

– Weather protection– Deluge protection– Dust Barrier

• Gassing cell +

Page 17: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

17

Document control number

OPTIMA OPTIMA PLUSPLUS

• Flow housing Configuration– Optima Plus

• methane• propane• butane• ethane• propylene.

– No change

Page 18: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

18

Document control number

Availability

Authority Detail Deliveries

ATEX EExd IIC T5 –40o to 55oCEExd IIC T4 –40o to 65oC(II 2 G)

End March

UL Class 1 Div 1 B,C & D End March

CSA Class 1 Div 1 B,C & DE79 Exd IIC T5 & T4

Sept 2000

DMT Performance to EN 50054:CH4 + others

Sept 2000

OPTIMA OPTIMA PLUSPLUS

NB: CO2 & FM units no longer available

Page 19: Search Line Optima Plus

Honeywell Proprietary

Honeywell.com

19

Document control number

Optima 1 exit plan

Type Last deliveries

CENELEC: HC & Ethylene End March

UL: HC & Ethylene End March

CSA: HC & Ethylene Sept 2000

DMT: HC Sept 2000

FM: HC & Ethylene Discontinued

CENELEC & UL : CO2 Discontinued