DEEPSOUND P1
www.seongsanlab.com
Sdf sdf sdf sdf sdf
®
Complete User Manual
Portable Phased Array
UT System
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 1
Getting Started Guide
Manual release 1.1
This instruction manual supports the usage and methodology of the DEEPSOUND P1 device, its software and functions.
Audience
This document and the information contained in its strictly reserved for current Seongsan Lab personnel, Seongsan Lab licensed representatives and Seongsan Lab customers who have purchased a valid service agreement for use by the customer’s designated in-house service employee on equipment located at the customer’s designated site.
General Precautions
Before using and turning on the device, please carefully read the instructions in this user manual. Keep this manual as the priority for future reference. Please follow the installation and operation procedures carefully.
User manual format
This manual is in Portable Document Format (PDF), for viewing on a laptop computer or any other electronic device compatible with Acrobat Reader® . A list of bookmarks functions as an additional table of contents.
Conventions of the user manual
This user manual is designed to be easy to understand and user-friendly, but to take full benefit of the information provided, conventions are useful to know easily.
Various typographical conventions explained below were defined to standardize and simplify the look and feel of this documentation:
Hypertext links are bold.
All procedures are numbered, and sub-procedures are lettered.
Control names and menu items or titles are spelled as they are in the device, and they appear in bold text.
Symbols, menu bars, and images appear as they appear in the software part
An italic typeface is used to indicate on a specific word or phrase (for example: should always).
Small capitals are generally used when reference is made to inscriptions found on the device (indicator lights, buttons, etc.)
Safety and Other Indications
The purpose of the various safety indications is to ensure operator safety and device integrity. In order to maintain this condition and to ensure a safe operation, it is absolutely necessary that you read the following safety information before putting the instrument into operation.
This sign indicates a hazardous action. It denotes that if not correctly performed or adhered to, could result in material damage or loss of data.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 2
Important: The Important sign denotes to a note that provides important information, or information necessary to the completion of a task with the device.
Note: The Note sign indicates an operating procedure, a practice that requires special attention. It also denotes related information, but not imperative ones.
Questions or comments about this manual
If you have questions about this manual, please contact Seongsan Lab,
Research and Development team, at the address below.
ATTENTION
Seongsan Lab, 2018. Rev 1.0. All rights reserved. All the information herein is subject to change without prior notification.
Contact us:
Seongsan Lab
Seung-il venture tower 802,
Gasan Digital Road 2-gil 136
(Gasan-dong 371-42)
Geumcheon-gu, Seoul, Korea
Seoul, South Korea
P: 02-2039-5725
F: 02-2039-5726
www.seongsanlab.com
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 3
List of Abbreviations
DAC- Distance amplitude curve
CLB – Calibration Method
GD - global delay
GN – echo gain
PA - Phased array
PRF - pulse repetition frequency
Rx – receiver
RNG – range
Ref – reference
PD – probe delay
PW-pulse width
ID – inside diameter
IDX – index
Neg - negative
Pos - positive
SNR - signal-to-noise ratio
SCN – scan
TOFD - time-of-flight diffraction
TCG - time corrected gain
Tx – transmitter
UT – ultrasound testing
US – ultrasound
USB – Universal serial bus
VLC – material velocity
WD - wedge delay
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 4
Icons & button function explanations
Icons in the status field
Icon Meaning
Pitch-and-Catch/TOFD probe
Single/Conventional probe
Battery charging status
Using an adapter status
Freeze: All Pulse & Receive operations are disabled
Locks a functions, except the
main menu
Device input options
Input methods Action
Touch screen Touch to select, zoom, drag gestures
Control buttons Scroll without menu list, select particular action
Rotary knob switch Scroll through menu list
Key functions
Button Main Function
Turn On/Off the device
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 5
Table of Contents
PART I: P1 Device Guide .................................................................................................................................... 9
1. Device Overview ......................................................................................................................................... 9
1.2 SPECIAL FEATURES ................................................................................................................................ 9
2. Accessories and packages ............................................................................................................................. 11
3. DEEPSOUND P1 Device specifications ...................................................................................................... 12
3.1 General specifications: ............................................................................................................................ 12
4. Control buttons and parts of the device ..................................................................................................... 14
5. Setting and connector references ................................................................................................................. 16
5.1 Turning ON/OFF the device: ................................................................................................................ 16
5.2 Setting up the P1 device: ........................................................................................................................ 17
5.3 Probe connections .................................................................................................................................... 18
5.3.1 Installing a probes ............................................................................................................................ 18
5.3.2 Removing a probes .......................................................................................................................... 18
5.4 Batteries and function ................................................................................................................................. 18
5.4.1 Installing a Lithium-Ion battery ......................................................................................................... 19
5.4.2 Charging the battery ............................................................................................................................ 20
5.4.3 Battery Status Indicators ...................................................................................................................... 20
6. Encoder connector (receptacle) .................................................................................................................... 21
6.1 Installing the encoder connector ................................................................................................................ 22
6.2 Removing the connector ............................................................................................................................. 22
7. LED indicator status ...................................................................................................................................... 23
8. Other connector references ........................................................................................................................... 23
9. Safety information ......................................................................................................................................... 24
9.1 Maintenance of the P1 device ................................................................................................................ 24
10. Troubleshooting and error handling ......................................................................................................... 25
1. P1 Device Software Functions ................................................................................................................. 27
1.0 User Interface and functions .................................................................................................................. 27
1.2 Input methods .......................................................................................................................................... 29
2. Menu explanations.................................................................................................................................... 31
2.1 Main Window .......................................................................................................................................... 31
2.2 Toolbar explanations: .............................................................................................................................. 32
2.3 Scan View Screen ......................................................................................................................................... 34
2.5 Submenu: Toolbars and Wizard ................................................................................................................ 38
2.5.1 Mini Toolbar- PULSER ........................................................................................................................ 39
2.5.2 Mini Toolbar- GATE ............................................................................................................................ 40
2.5.3 Mini Toolbar- Calibration.................................................................................................................... 41
2.5.4 Mini Toolbar- DAC/TCG .................................................................................................................... 44
2.5.5 Mini Toolbar- ENCODER Scan .......................................................................................................... 46
2.5.6 Mini Toolbar- BEAM ............................................................................................................................ 47
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 6
2.5.7 Mini Toolbar- Data ............................................................................................................................... 48
2.5.8 Mini Toolbar- Channel Group ............................................................................................................ 48
2.5.9 Mini Toolbar– WIZARD ...................................................................................................................... 49
2.5.9-1: Configuring the Specimen & Weld ................................................................................................ 50
2.5.9-2: Configuring the Channel Group .................................................................................................... 53
2.5.9-3: Configuring a Focal Law ................................................................................................................. 64
2.5.9-4: Calibration ......................................................................................................................................... 67
2.5.9-4.1: Velocity Calibration .................................................................................................................. 71
2.5.9-4.2: Wedge Delay Calibration ......................................................................................................... 68
2.5.9-4.3: Sensitivity Calibration .............................................................................................................. 71
2.5.9-4.4: Encoder Calibration .................................................................................................................. 77
2.6 Submenu: Settings ....................................................................................................................................... 81
2.6.1 Submenu SUB TOOLBARS: CHANNEL SETTING ........................................................................ 81
2.6.2 Submenu SUB TOOLBARS: PULSER ................................................................................................ 83
2.6.2-1 Sub Menu: PULSE ......................................................................................................................... 83
2.6.2-2 Sub menu: RECEIVE .................................................................................................................... 87
2.6.3 Submenu SUB TOOLBARS: BEAM ................................................................................................... 90
2.6.3-1 Sub-Menu: BEAM ......................................................................................................................... 90
2.6.3-2 Sub-Menu: FOCUS ....................................................................................................................... 91
2.6.4 Submenu SUB TOOLBARS: PROBE .................................................................................................. 93
2.6.4-1 Sub Menu: PROBE ........................................................................................................................ 93
2.6.4-2 Sub Menu: ELEMENT .................................................................................................................. 94
2.6.3-4 Sub Menu: OFFSET....................................................................................................................... 95
2.6.5 Submenu SUB TOOLBARS: WEDGE ................................................................................................ 97
2.6.6 Submenu SUB TOOLBARS: TRANSMITTER/RECEIVER ............................................................ 97
2.6.6-1 Sub Menu: Tx DELAY .................................................................................................................. 97
2.6.6-2 Sub-Menu: Rx DELAY ................................................................................................................. 98
2.6.7 Submenu SUB TOOLBARS: MECHANICAL SETTING ............................................................... 100
2.6.7-1 SCAN: ........................................................................................................................................... 100
2.6.7-2 INDEX: ......................................................................................................................................... 101
2.6.8-3 ENCODER ................................................................................................................................... 103
2.6.8 Submenu SUB TOOLBARS: SPECIMEN/WELD SETTING ........................................................ 104
2.6.8-1 Sub-Menu: SPECIMEN .............................................................................................................. 104
2.6.8-2 Weld .............................................................................................................................................. 105
2.6.9 Submenu SUB TOOLBARS: SETTING LOAD / SAVE ................................................................ 106
2.6.9-1: Setup (file format: SET) ............................................................................................................. 106
2.6.9-2: Image (file format: PNG) .......................................................................................................... 107
2.6.9-3: Data (file format: IQV) .............................................................................................................. 107
2.6.9-4: Layout.......................................................................................................................................... 108
2.6.10 Submenu SUB TOOLBARS: DEVICE SETTING .......................................................................... 109
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 7
2.6.10-1: Display ...................................................................................................................................... 110
2.6.10-2: Gate Line ................................................................................................................................... 113
2.6.10-3: Device ........................................................................................................................................ 114
2.6.10-4: Preference ................................................................................................................................. 115
2.7 Submenu: Gate option toolbar ................................................................................................................. 116
2.8 Submenu: Scan view toolbar .................................................................................................................... 118
2.9 Submenu: Shortcut toolbar (horizontal) ................................................................................................. 119
APPENDIX B .................................................................................................................................................... 123
APPENDIX C ................................................................................................................................................... 130
Important NOTICE .......................................................................................................................................... 133
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 8
DEEPSOUND P1 DEVICE GUIDE
Revision 1.1
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 9
PART I: P1 Device Guide
1. Device Overview
The DEEPSOUND P1
Seongsan Lab’s DEEPSOUND P1 device is targeted for providing a high quality – low
cost, portable and highly efficient, user-friendly, ultrasound inspection using PAUT
technology with various functions.
The P1 is a mobile Phased-Array device that well suited for:
Ultrasound Testing (UT) is used for a flaw detection, size measure, and
imaging. PAUT technology has the ability to modify electronically the acoustic
probe characteristics.
The encoder saves collected data and improves the difficulty of data storage
when the UT is used. The device also enables data collection in the form of 2D-
images along availability of A-Scan, B- Scan and C-Scan.
1.2 SPECIAL FEATURES
Special features of the DEEPSOUND P1:
Phased Array UT technology with simple UT
Time of Flight Diffraction Technique (TOFD) included
7.0 inch Touch screen with high-resolution LCD
Bipolar Pulse Wave + Burst Wave
Easy scan setup through various Wizard functions (Channel Group, calibration etc)
2-D image analysis through B-Scan and C-Scan
Storage of control and acquired data sets enabling quick calibration and reproducibility of the test
Various filter type availability (Band pass, High pass, Low pass)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 10
Acquisition of data by encoder and Sectorial scan view, B scan and C scan view – In UT technique, it is quite difficult to save test data and illustrate B and C scan view using an encoder. The advantage of DEEPSOUND P1 device improved this point, thus the user may save acquired data collected by an encoder in computer file format, copy the saved data through USB and read the test result from PC.
Due to disadvantages of PAUT, ultrasound signals are provided to examine in a one-dimensional form. Therefore in order to improve the need for much time, knowledge, and hands-on experience in determination of defects, the encoder is input and one Scan and C-Scan data in real-time. These features are very helpful in determining defects and helping to make defective analysis over cracks easier for even inexperienced users.
Note: In this document, hardware controls that you press to activate are referred to as keys, the word button is reserved for software control.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 11
2. Accessories and packages
This chapter guides you about the accessories and packages available for the P1 device.
2.1 Accessories
Product Description
Device case Carrying case for P1 device only
Strap for wrist Can use for mobile activity
Strap for shoulder Can use for traveling, mobile activity
2.2 Available packages
Part Description
Rechargeable battery x 1 ea Removable battery with 6400 mAh/7.5 V
AC/DC adapter 12V / 8.3A adapter
User Manual Korean/English version available
Device calibration certificate Calibration document available in Korean
Quality assurance documents Quality assurance proof document available in Korean
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 12
3. DEEPSOUND P1 Device specifications
This chapter will provide more information about specifications of the P1 device.
Main body description of P1 device
Size (H x W x D): 23 x 16 x 6 cm
Net weight: 1.6 kg (without battery)
2.0 kg (with battery)
Battery life
Up to 6 hours (battery replacement, adapter use without battery supported)
3.1 General specifications:
Display: 7.0 inch LCD with Touch Interface (diagonal)
Display resolution: 1024 pixels x 600 pixels
PAUT channel: 16 : 64 PR
Probe connector: IPEX 80 x1 ea (PAUT)
Pulse wave type: Bipolar
Max burst count: 5
Dynamic sample focusing: Yes
Digitizing frequency: up to 50 MHZ
Amplitude Resolution Signal: 16 bit
Maximum PRF: 20 KHz
Maximum pulse voltage: 160V
Bandwidth: 0.25 ~ 20 MHz
Real-time data compression: Yes
Rectification: Full wave, Positive unipolar, Negative unipolar, RF wave
Scan view: A Scan view, Sectorial scan view, C scan view, B scan view
Filtering: Analog/Digital
Encoder Interface: 2 Axes /D-Sub pin 3 row
Max gate: 3 gate (A,B,I) (supports beep sound)
Interface:
Ethernet connector interface: 1 GBps/s
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 13
USB: High speed 3.0
Adapter and Battery:
Input: 12.0 V / 8.3A
Battery: 2S2P (6400 mAH) x 1 ea
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 14
4. Control buttons and parts of the device
Front panel – Display, knob switch and power button
Top panel – probe connector (locker and socket), encoder and USB connector
Right side panel – Ethernet (RJ-45) and DC adapter connector
Left side panel – handler
Rear side panel – Cooling fan, standing, battery cover lid
Detailed descriptions:
1. Display- 7.0 inch touch screen LCD display
2. Power button/ LED indicators – On/Off the device
3. Rotary (dial + push) knob switch – adjust the parameter value
4. PAUT probe connector – used to connect phased array probe (Mini DLP)
5. Encoder connector – used to connect encoders
6. USB connector – used to connect external USB storage device
7. Ethernet connector- used to connect with LAN cable
8. DC adapter connector – connect device with power source/charge batteries
9. Battery lid with screw: contains battery (1 ea) and lid with screw (2x)
10. Stand: used to switch position to tilt
11. Hanger strap: used to carry the device with hand
12. Rear cooling fan : Part of the cooling system
13. Heat sink
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 15
Figure: P1 device front and left side panel
Figure: P1 device rear and left side panel
1
7
8
2
3 4
5 6
11
10 9
12
13
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 16
5. Setting and connector references
This section explains the device general settings and its various connector references in detail. Furthermore, appendix section will provide more information about probes, which will lead you to obtain even better acquisition scanning result.
We strongly recommend user to read all this manual thoroughly before real-time application or unintended hazardous action, since this document
provides all information you may need.
5.1 Turning ON/OFF the device:
To turn on the P1, press the power button, and the display will be on. The settings of all function values and basic settings are the same as they were before the software was turned off.
We strongly recommend users to do not turn on the P1 device until it has reach room temperature, if P1 has carried under extremely low temperature
for a long-time, immediate turning on may lead P1 to the damage.
Turning ON
To turn on the P1, press the power button for 1 second, then red LED will lit on.
You will first see the start display of the P1 showing the current the device name and logo, software version (figure a) and the main screen (figure b)
a) start screen b) the main screen
After a few seconds of a period, the normal screen display appears, and the device will be ready for operation (figure b).
The settings of all function values and the basic settings (units/time/date) are the
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 17
same as they were before the device was turned on/off
Turning OFF
To turn off the P1, press the power button and “Shut Down” notification will appear. Then red LED will lit which indicates a shutting down procedure.
But shutting down process is bit different, you have to still check the red LED is turning down.
NOTE*: Make sure check the red LED is turned off, which indicates the complete shutdown of the P1
Figure : Window when power button is pressed
5.2 Setting up the P1 device:
The P1 device is lighter than other similar products, therefore the device can be used on the desk, floor or other flat spaces or you can carry the device with proper carrying strap in any direction or condition.
Since P1 meets the safe and light performance in hardware matter, but safety is still essential. Therefore we recommend users to use P1 device always makes sure positioned away from heat sources, and if the device has been carried during cold or harsh weather for a long-time, wait until has been adapted to the proper temperature before its get used.
Important: It is recommended to that device should positioned close to a power outlet that could avoid disconnect unexpectedly in some manner.
Also, we encourage you to not block cooling fan for while intentionally or not.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 18
5.3 Probe connections
In order to prepare P1 for UT inspection, you can create an Ultrasonic (UT) Channel Group for UT inspection using the UT Connector or PA Connector by selecting the Probe Type in the settings.
Example of probe connections:
Important: Switched or incorrectly installed probe causes a mismatch which can lead to distorted echoes or power losses.
5.3.1 Installing a probes
To install a probe:
Remove the adapter from the storage.
Align the adapter on Pulser/Receiver, and gently push it in. The adapter should slide easily.
Secure the probe with turning tightening screws in direction of closing. Device is now ready to work with probes.
5.3.2 Removing a probes
To remove connector:
Turn tightening screws in opening direction to remove a probes.
Gently pull out to remove connector from the socket.
We recommend users to pull out the adapter in a straight direction. Otherwise, connector pin could bend and render the adapter unusable.
5.4 Batteries and function
The P1 device uses single powerful, long-lasting Lithium-Ion storage, that has a high capacity, and supports battery replacement.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 19
Recommendation:
In case of store batteries, store at below 20 C, low humidity, no dust and no corrosive gas atmosphere. Store the battery with a state charge between 50-70%.
5.4.1 Installing a Lithium-Ion battery
The Lithium On storage battery can be temporary installed in a device battery tray.
Before installing batteries:
To insert a new battery, the lid located rear side of the device must be removed using two screws and remove the battery compartment lid and insert batteries one by one.
Figure: Battery compartment lid (2 ea screws) Battery position inside of the body
Loose the two screws of the battery lid
Insert the battery into battery holder and connect the wire to connector
Close the battery lid and tighten up screws
Note: P1 equipment has a caution between battery replacement and combination.
[1] You need to check the direction of the connector between replacing and engaging the battery
[2] Make sure you do not have any water on your hands (it may affect the PCB).
You can charge the Lithium-Ion battery during operation directly on the P1 using the external power supply/charger unit of P1. As soon as you connect the power supply unit (plug-in power supply unit) to the P1, the battery usage is stopped immediately.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 20
5.4.2 Charging the battery
To charge the Lithium-Ion storage batteries, the P1 AC/DC power supply/charger unit must connect, it will automatically recharge any battery requiring charging, starting with the weakest battery. Since the P1 is running, less power is available from the DC source to recharge the battery.
The P1 DC power adaptor is designed to charge P1 batteries only. Do not attempt to charge any other type of battery, or use any other chargers/adaptors. Doing so could cause an explosion or serious injury to both user and the device. Do not attempt to power or charge other electronic equipment with the P1 DC power adaptor, as doing so could cause an explosion resulting in serious personal damage. We strongly recommend user to read all this manual thoroughly before real-time application or unintended hazardous action, since this document provides all information you may need.
Note: [1] The batteries can be charged during normal operation. You do not have to remove the battery lid for this purpose.
[2] You can charge a separate Lithium-IoN storage battery by connecting to the charger unit.
5.4.3 Battery Status Indicators
There is a single charge indicator for the P1 device. The indicator show the amount of power remaining in the battery or batteries (see the figure below).
Figure: P1 device adapter using /battery charging status
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 21
6. Encoder connector (receptacle)
The encoder connector is fully compatible with the scanning mechanism. Following image shows that encoder connection port of the P1, which compatible with manual and automatic driven scanning mechanisms. P1 encoder interface has 4 quadrature type and 2 differential type.
Description: D-SUB DS1033-15F connector
Figure: Encoder connection on side panel /connector socket
Pin Signal
3 Output 5V
9 ENC_Q0
10 ENC_I0
12 ENC_I1
11 ENC_Q1
15 GND
Table: Pinout for encoder scanner connector
The important thing that working with a probe is make sure that the connector pins are properly aligned and that the adapter is not positioned in the wrong side. Also when manipulating the adapter, never use
excessive force, that may bend the pins and render the adapter unusable.
Encoder connector port
Sdf sdf sdf sdf sdf
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 22
6.1 Installing the encoder connector
To install the encoder connector:
Open the rubber cover
Align the encoder on port and gently push it in. Please keep in your mind that do not use excessive force to push the encoder to a port, it may cause the bend of pins or malfunction.
6.2 Removing the connector
To remove connector:
Turn tightening screws counter clockwise until they are completely out of the sockets.
Gently pull out to remove it from the connector.
Close back the rubber cover
We strongly recommend users to pull out the adapter in a straight direction. Otherwise, connector pins could bend and render the adapter unusable.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 23
7. LED indicator status
The light indicator let you know an alarms or activity on the current status of the P1 device.
Gate -> Echo [ Gate A(Red), B(Orange), I(Green) ]
System booting
System shutdown
Charge indication
Figure: Alarm LEDs indicator
8. Other connector references
Gigabit Ethernet Connector
P1 uses auto-configurable Gigabit Ethernet port, a user may use any type of ethernet cable to connect the P1 device to the internet.
Description: RJ-45
Figure: Ethernet connector reference
Parameters Specifications
Description RJ-45 , four pair
Alarm LEDs Sdf sdf sdf sdf sdf
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 24
9. Safety information
The P1 is designed and tested according to safety standards and was technically in perfectly safe and faultless condition when leaving the manufacturing works.
The P1 is a device for materials testing. Any use for other applications such as medical or other purposes is not allowed.
9.1 Maintenance of the P1 device
There is no maintenance required, a few recommendations are listed below:
Cleaning the device surface
The P1’s external cases (front and side panel) can be cleaned under some circumstances:
To clean the device:
Always make sure the device power cord is disconnected and turned off completely.
Keep in mind that device body surface is better to clean with a soft clothing material.
We strongly recommend users to not clean the device with liquid, especially when the device is turned on. The liquid could make the device partially or wholly inoperable, or when connectors stay wet without immediate drying process, it may produce a short circuit when
plugging cables.
Wait until the device dries completely before plugging into the power cord or any other cables.
We recommend you to always use carrying case when the device is not used.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 25
10. Troubleshooting and error handling
This section will provide basic troubleshooting if device malfunctioned or problem arises. In case of problem detected, please refer the following table.
Problem Cause Recommended solution
P1 does not turn ON Power cord disconnection Check the power cord and replace it with the
proper size.
P1 does not turn ON Battery on empty state Make sure the adapter
has connected with power supply.
The P1 shuts down unexpectedly
The device has been used for extreme temperatures
for an extended time
Wait few minutes before turning on the device.
Change the working location or reduce the ambient temperature.
Encoder signal are not detected
The wrong setup configured in the settings
page.
Verify pin-out and refer to this software manual.
Table: Basic troubleshooting
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 26
DEEPSOUND P1 Device Software Manual
Revision 1.1
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 27
PART II: P1 Device Software
According to the current state of the art, software is never completely free from errors. Before using any software-controlled test equipment, please make sure that the required functions operate perfectly in the intended combination of usage.
If you have any questions about the use of our device and software, please contact our software department using the address provided in Page 2.
Important notes on software
Please read the following notes before using DEEPSOUND P1 device. It is important that you understand and observe this information to avoid any operator errors that might lead to false test results. This could result in damages to property.
Operator pre-knowledge
The operation of an ultrasound test device requires a proper knowledge of ultrasound test methods,
• Non-destructive inspection method
• How to operate NDT
Lack of such knowledge could lead to false test results with unwanted consequences.
1. P1 Device Software Functions
This chapter presents the general features of the P1 system software user interface. The shortcut menus that are available in full-screen modes are explained in detail with “Shortcut menus”.
1.0 User Interface and functions
In the following hierarchy, if a button has a pale background, it is selected, and the focus is on that particular menu. The main list temporarily appears vertically over the data screen area while the focus is on the menu level. The submenu buttons appear vertically on the left side of the screen, below the menu button. The parameter buttons appear at the top of the screen.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 28
Menu > Submenu > Parameter = Value
Figure: The menu hierarchy and identification syntax
The following procedure provides a detailed description on how to select menu, submenu, and parameter buttons in the user interface of the P1 software.
To navigate in the P1 software:
Choose the desired Menu button
Proceed as follows to choose the desired item from the menu list:
Tap the selected item on the touch screen
Dial or push the Rotary Knob to adjust
The software immediately reassigns the functions of the submenu buttons to those corresponding to the chosen menu item.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 29
1.2 Input methods
The display touch screen acts as pointing device by tapping and other gesture.
To select an interface element, touch the screen surface gently with your single finger. To drag an interface element, just slide your fingertip across the screen surface.
Figure: Touch gesture example
Display functions also work with dragging and shifting gestures. In a scan view window, gesture works as the following chart:
Action Description Usage
Single tap
Select the button or highlight the action
Double tap
Entering the selected function or button
Drag
Drag the gate lines in scan view window
Pin-to zoom
Pin to zoom-in/zoom-out the scan view window or enlarge/shrink the
wave and gates
Long press
Activate other auxiliary functions.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 30
Note: The icon in this document indicates the touch feature is available.
Rotary knob (dial and push)
The P1 is equipped with the rotary knob switch. The rotary serves for setting the currently selected function, and enables you to directly set the gain. Also, knob switch enables to select function as pushing on it. This enables users to select functions as not just tapping on the screen, also by pushing on the knob switch.
NOTE: Rotary knob: Since the device has intended for a touch interface, icon bars can be user accessible with a double tap.
*These icons will guide through rest of function reference.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 31
2. Menu explanations
2.1 Main Window
This part will guide you through toolbars with functions of the software. The main window contains the following elements.
Figure: Toolbar components of the main window
Shortcut toolbar Sdf sdf sdf sdf sdf
Mini toolbar Sdf sdf sdf sdf sdf
Data Screen Sdf sdf sdf sdf sdf
Acquisition toolbar Sdf sdf sdf sdf sdf
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 32
2.2 Toolbar explanations:
The user can choose a variety of toolbars in the main setting file. In the top toolbar, there is six submenus are available to choose.
NOTE: Every tap on top-right corner button (with touch icon in the figure) will change options. See next pages.
Every submenu is accessible by first selecting by touch the screen, then use rotary knob switch to change values.
2.3 Data screen:
Displays scan view, the graphical lines of gates, its position and displacements.
2.4 Top menu: acquisition toolbar
Displays the measurement values, the function identification, as well as control buttons on the mini-toolbar.
2.6 Submenu: WIZARD toolbar
Displays various options of the wizard, as well as sub-options in mini-toolbar.
2.7 Submenu: Settings toolbar
Displays various settings and explanations of settings.
2.8 Submenu: Gate option toolbar
Displays various options of gate toolbar-options in mini-toolbar. Displays the
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 33
2.9 Submenu: Scan view toolbar
Right
Shortcut toolbar
Displays the measurement values, the function identification, as well as control buttons on the mini toolbar.
Mini toolbar
Displays the names of the various menu.
Submenu: Optional setting
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 34
2.3 Scan View Screen
The Scan view screen area displays the various ultrasound data views and layouts.
A scan is a 2-D graphical representation of ultrasound data with a ruler or scale corresponding to the horizontal and vertical axis. For example, an A-scan and a C-scan are two different types of scans.
A view is a volumetric representation of a part, which includes signal overlays. Like a scan, a view has two axes. However, instead of being related to a specific group, a view is linked to the part. A signal that originates from a single group or from multiple groups can be displayed without affecting the view dimensions.
A layout is an organization of one or more scans and views. For example, the A-B-C layout is a graphical arrangement of an A-scan, a B-scan, and a C-scan.
View Point of view Axis content
A Scan Looking down on the
material Amplitude versus US axis
B Scan Side Scan axis versus US axis
C Scan Top Index axis versus Scan axis
D Scan End Index axis versus US axis
S Scan Side Index axis versus US axis
Table: Scan view screen explanation
A-scan View
This is the view window that underlies all other view windows. Generates an ultrasonic beam to indicate the received ultrasonic amplitude or waveform. In the signal, the peak corresponds to a reflector within the specimen or a discontinuous reflected signal. Generally, the peaks at the beginning and end of an ultrasonic axis generally correspond to the reflected signals on the surface and bottom of the specimen.
B-scan
2-D side representation of the part showing US data with the scan length on one axis, and the US path on the other axis.
C-scan
2-D side representation of the part showing US data with the scan length on one axis, and index length on the other axis. One of the available parameters (for example,
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 35
amplitude) is projected on the index-scan plan for each point (pixel).
D-scan
2-D side representation of the part showing index and US data the scan length on one axis, and the US path on the other axis.
S-scan (Phased Array only)
2-D representation of US data presetting all the A-scans generated by the focal laws in an angular sector or sectorial range in order to create a cross section of the part. The A-scans are represented by lines on which the amplitude is color-coded, and they are corrected for delay and true depth so that their positions are accurate relative to the US axis.
Rulers
Each ruler/scale in filled with a dedicated color to help you identify the axis on the various views. Following table provides a list of the available rulers/scales, and their associated colors and functions.
The basic color is attributed to each axis. The axis appears in various tints of the basic color. The lightest tint corresponds to the raw data representation. Progressively darker tints correspond to increasing data correction complexity relative to the axis. A darker tint is also used for an axis that appears as a reference, in which case the zoom bar is not available.
Axis Color Appearance Note
Amplitude Light green Uncorrected
MetalPath Pink No zoom
Ultrasound Purple True depth units
Scan Blue Volumetric correction
Index Green N/A
Ultrasound Pale pink Sound path units
Degree Pale blue Degree
measurement
NOTE: The colors in this document differ slightly from those on the P1 device screen.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 36
2.4 Top menu: Acquisition toolbar
Tap on top-right corner icon with down arrow button, the following sub-option will appear.
SubMENUs (button options):
Button Unit Description
Gain dB gain value ranges between 0 to 90 dB
Gate n/a Gate A, Gate B and Gate I are available
A%, B%, I% % Amplitude in A,B,I gate
in (%) screen height
SA, SB, SI mm Sound path of echo in A,
B,I gates
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 37
DA, DB, DI mm Reflector depth in A, B, I
gates
RA, RB, RI mm Reduced projection
distance in the A, B,I gates
PA, PB, PC mm Projection distance of
the echo in A, B,C gates
AdBA,BdBB dB Amplitude (dB) of the signal detected in Gate
A, B
AdBr, BdBr dB
Difference between the current signal amplitude and the reference signal amplitude detected in
Gate A, B
%(r), %(m), %(m-r)
% Amplitude value at the
Reference/Measurement bar and their difference
U(r), U(m),U(m-r)
mm
Defines the position value of
Reference/Measurement bar and their difference
S(r), S(m), S(m-r)
mm
Defines position value of Reference/Measurement bar and their difference
on Scan axis
I(r), I(m) I(m-r)
mm
Defines position value of Reference/Measurement bar and their difference
on Index axis
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 38
2.5 Submenu: Toolbars and Wizard
Figure: Mini toolbar-Wizard
Top-menu left corner with button option allows you to change the Submenu Wizard toolbar. Submenu Wizard toolbar are main in menu change with the user. Tap on top-left corner icon with down arrow button, and following menu will appear:
Figure: wizard-mini toolbar
Submenu buttons, rotating the Rotary knob push, clockwise or counter clockwise allow you to a desired software button, or change a parameter value.
Long-press to main instant menu change with the position you can drag.
Figure: Virtual keyboard with numeric keypad (right) or system built-in virtual keyboard (left)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 39
2.5.1 Mini Toolbar- PULSER ( )
Tap on ( ) button to open UT Group menu bar. Tap on buttons, then change the value or options with rotary knob switch.
Wave Type: Used to select the wave type (LW or SW)
WD (us): Set Wedge Delay in units of usecs. Wedge Delay is the delay time of the ultrasonic waves between the wedge and the probe. Before adjusting the Wedge Delay, set the sound velocity and calibration range according to the dimensions of the test specimen.
RNG (mm): In AScan, maximum sample count can be adjusted up to 16K (16384), you can read and adjust max range in mm.
VLC (m/s): Select the function US delay velocity and adjust the value by using the rotary knob switch.
VLT (V): Pulser voltage can be set through this function by using the rotary knob to select the required output voltage between 25 V ~ 200 V.
PRF-Recurrence (Hz): Recurrence value indicates the number of times an initial pulse is triggered, which 1000 Hz is generated and collected a pulse once every 0.001 (1/1000) seconds. By using this function, you can determine the value of the pulse repetition frequency.
Pulse Width (mm): Width value settings of Bipolar pulse wave. Pulse width can be defined as 1/Pulse frequency (Hz)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 40
2.5.2 Mini Toolbar- GATE ( )
Tap on ( ) button to open Gate menu bar. Tap on buttons, then change the value or options with rotary knob switch.
GATE: Select the gate type (A, B, I)
Gate Start (mm): Used to set the starting position of the selected gate. This position is related to the gate synchronization. The real position of the gate is the synchronization position plus the gate start position.
Figure: Gate start example (1mm (left) and 5mm (right))
Gate Range (mm): Set the range of gates. Adjust the value by using the rotary knob switch. Refer the following image.
Figure: Gate range example (1mm (left) and 5mm (right))
GH (%): Gate height, indicates the height position in percent. Adjust the value by using the rotary knob switch.
Figure: Gate Height (GH) example (50mm (left) and 80mm (right))
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 41
2.5.3 Mini Toolbar- Calibration ( )
Tap on ( ) button to open Calibration menu bar. Tap on buttons, then change the value or options with rotary knob switch.
Type: Select the calibration type: Metal Path, Depth, Velocity, and adjust the value by using the rotary knob switch. Settings become different according to calibration type selection. See more from Appendix B: Calibration section.
Settings of Metal Path
WD (us): Measures and displays the Wedge Delay in usecs. Used to add ultrasound speed delay cause by the wedge. This delay line in the wedge can be adjusted with wedge delay function.
Ref A (mm): Displays a horizontal line on the A-scan at the height of the reference-amplitude value. (Available when Metal path, Velocity method selected)
Auto FSH (%): Automatically adjust a gain value to the registered setting value (80%) of FSH (Full Screen Height). See the following figure.
Figure: example of performing auto FSH
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 42
Envelope:
Displays a curve with the maximum and minimum amplitudes on the A-scan view.
Clear Envelope:
To clear the envelope, tap and hold the A-scan view, then tap Clear envelope on the shortcut menu of Calibration.
Calibrate: After configuring the parameters, tap on Calibrate button to finalizes the calibration.
Settings of Depth mode
The selection of this method is similar to “settings of Metal Path” except:
WD (us): Measures and displays the Wedge Delay in usecs. Used to add ultrasound speed delay caused by the wedge. This delay line in the wedge can be adjusted with wedge delay function.
Ref Depth A: Measures the depth of amplitude at which the curve is located if the actual gain is same as the reference gain (available when Depth mode selected).
Auto FSH (%): Automatically adjust a gain value to the registered setting value (80%) of FSH (Full Screen Height). See the following figure.
Clear Envelope:
To clear the envelope, tap and hold the A-scan view, then tap Clear Envelope on the shortcut menu of Calibration.
Calibrate:
After configuring the parameters, tap on Calibrate button to finalize the calibration.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 43
Settings of Velocity mode
The selection of this method is identical to “settings of Metal Path”, except:
VLC (m/s): Used to calibrate the velocity of the sound propagation in the material of the inspected part. Select the function and adjust the value by using the rotary knob switch.
Ref A (mm): Measures the amplitude at which the curve is located if the actual gain is same as the reference gain (available when Metal path, Velocity method selected).
Ref B (mm): Measures the amplitude at which the curve is located if the actual gain is same as the reference gain (available when Velocity method selected)
Auto FSH (%): Automatically adjust Gain to the registered setting value (80%) of FSH (Full-Screen Height).
Clear Envelope: To clear the envelope, tap and hold the A-scan view, then tap Clear envelope on the shortcut menu of Calibration.
Calibrate: Tap on Calibrate (Wizard < Calibration < Calibrate) button to finish the procedure.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 44
2.5.4 Mini Toolbar- DAC/TCG ( )
Tap on ( ) button to open DAC/TCG menu bar. Tap on buttons to change the setting.
DAC mode
The echo height of reflectors of size depends on which probe using, mainly cause of the sound attenuation in the material.
DAC – Distance-Amplitude Curve record using defined references, represents these effects graphically.
The DAC function enables:
Visualize the DAC of the reference echoes.
Peak the echo of a probable flaw. If necessary, increase the gain until the echo exceeds the gate threshold.
Activate the TCG as described above, however, with a simultaneous display of the TCG/DAC function.
For about DAC/TCG methods, please see the Appendix C.
Activate the TCG (Time-Corrected Gain) to increase the gain in distance-dependent mode so that all reference echoes reach the same screen height.
Index menu appears as the following figure. Tap on DAC button to open further tool options. Peak the echo of a probable flaw, and move the gate so that the echo data are evaluated.
Determine the sound path in the display boxes for readings and, in addition, the reduced projection distance as well as the flaw depth if you are using angle beam probes. Whether or not the echo is recordable, can be immediately seen from the A-scan.
Determining can be done through dB amount by which the response threshold is exceeded: Adjust the gate threshold exactly to the intersection point of the curve with the probable flow indication.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 45
TCG mode
An echo amplitude evaluation as the measurement of the difference relative to the reference echo is only possible if the reference height stored during the recording reference echo still applies. If another echo is stored as reference echo with dB Ref after recording the curve, the original reference echo is lost.
After switching on the TCG mode from DAC mode, all echoes of the reference reflectors reach the height of the first reference echo.
Figure: DAC /TCG menu buttons (see the Table below)
Button Description Appearance
DAC (Add Point & Remove Point)
Add or remove DAC curve line in scan view (purple)
TCG (Add Point & Remove Point)
Add or remove TCG curve line in scan view (green)
Table: DAC/TCG menu-function button explanations
This button function is ideal to previous Settings 1 function group.
NOTE: Rest of selection toolbar is identical to Setting 1 menu (above figure) See the previous page.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 46
2.5.5 Mini Toolbar- ENCODER Scan ( )
Tap on ( ) button to open Encoder Scan menu bar. Tap on buttons, then change the value or options with rotary knob switch.
Scanner Type:
Defines scanner types including: Encoder 1, Encoder 2 and Time. h the curve is located if the actual gain is same as the reference gain (available when Velocity method selected)
SCN (mm): Indicates the measurement of scan direction. The unit is in mm.
IDX (mm): Indicates the measurement of index direction. The unit is in mm.
( ): Scan resume: Resumes the scanning mode.
( ): Scan start: Note that when the scan is started, Display lock function will start, due to no interruption except resume/stop action
Figure: Scan and Index measurement indicator
(start/stop/step) in mm respectively
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 47
2.5.6 Mini Toolbar- BEAM ( )
Tap on ( ) button to open Beam menu bar. Tap on buttons, then change the value or options with rotary knob switch.
Scan Type: Tap on this button to select scan type: Sectorial and Linear (see the figure below)
Figure: Scan type
Angle Start (deg): Indicates the minimum angle value for the ultrasound beams (see the figure below)
Angle Stop (deg): Indicates the maximum angle value for the ultrasound beams (see the figure below)
Resolution: Indicates the number of degrees between each ultrasound beams (see the figure below)
Aperture Size: Indicates a size of the group of active elements (Available when Linear scan type is selected) (see the figure below)
Figure: Settings in beam menu illustration
Focus Position: Defines the position where beam is focused. Unit: mm
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 48
2.5.7 Mini Toolbar- Data ( )
Tap on ( ) button to open Data menu bar. This menu intends only to demonstrate the IQV file that loaded. This option only available in Analysis Mode.
Soft Gain: You can set the Echo height arbitrarily on the screen of the IQV file.
Figure: Before setting up Soft Gain Figure: After setting up Soft Gain
Close Data: Tap on this button to close the data view.
2.5.8 Mini Toolbar- Channel Group ( )
Tap on ( ) button to open Channel Group menu bar. Tap on buttons, then change the value or options with rotary knob switch.
Channel Group: Select available Channel Group here. You find the Channel Group settings from here.
View Type: Select Scan view type -> AScan, BScan, B Scan Unc, C Scan Unc, DScan, SScan and FFT Scan
Add to: Select the position of Scan view windows.
Add: Add the selected Channel Groups to the Scan view.
Remove: Remove the selected Channel Groups from Scan view.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 49
2.5.9 Mini Toolbar– WIZARD ( )
The wizard menu offers step-by-step assistance to quickly and easily create a complete setup for your application (see the following figure). The step-by-step approach ensures that you do not miss a parameter change during the setup. The help context of each wizard step provides specific information regarding the parameter to be set.
To perform a complete application setup, use the different wizards available from the submenu buttons, in the following order:
Specimen & Weld (2.5.9-1)
Channel Group (2.5.9-2)
Focal Law (2.5.9-3)
Calibration (2.5.9-4)
Help context
BeamTracing
illustration
Parameters of
the step
Step-by-step navigator
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 50
BeamTracing
The BeamTracing view is a dynamic cross-section illustration of the part, the propagation of the US beam in the part, the limits of gate A, the location of the indications in the part, and the weld drawing.
Figure: The Beam Tracing example
2.5.9-1: Configuring the Specimen & Weld
To configure the specimen and weld
Select Wizard > Specimen & Weld > Next to start the Specimen & Weld
Follow the on-screen help context to complete the remaining steps of the wizard.
Tap on Finish to finalize the Specimen & Weld wizard
Note: You can exit the wizard at any time by pressing the other key.
Specimen & Weld > Specimen
Tap or using rotating knob switch, select the specimen material, shape and its thickness.
Thickness:
Specify the thickness of specimen. Change the value to thickness of the specimen.
Measurement unit: mm
Figure: Specimen thickness
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 51
Material:
Used to select the specimen material. This will set the velocities for both L-Wave/S-Wave velocities for the selected material.
Figure: Specimen material list example
Shape:
Plate: A flat surface (while tube is type of curved surface, including OD and ID).
OD: The inspection is performed on the Outer Diameter of a tube. When this mode selected, the unit of axis can be set to a degree.
ID: The inspection is performed on the Inner Diameter of a tube. When this mode selected, the unit of axis can be set to a degree.
Note: In the program user interface, the visualization of the Beam Tracing only supports flat geometries.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 52
Specimen & Weld > Specimen>Weld > Size
Used to define the weld orientation on the pipe. Define the weld type from 6 different types, including V, Reverse V, Bevel, Reverse Bevel, Double V, Double V offset
Figure: Weld types
Each weld shape can be changed or adjusted by following parameters:
Select Weld – V: Cap-Height, Cap-Width, Neck-Width
Select Weld – Reverse V: Cap-Height, Cap-Width, Neck-Width
Select Weld – Double V: Cap-Height, Cap-Width
Select Weld – Double V offset: Left Cap-Width, Left Root-Width, Left Cap-Width, Left Root-Height
Select Weld – Bevel: Cap-Width, Root-Width, Cap-Height, Root-Height
Specimen & Weld > Specimen>Weld > Size > Finish
After configuring the setting, select Finish to complete Specimen & Weld wizard.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 53
2.5.9-2: Configuring the Channel Group
Selecting the Probe and the Wedge
The P1 device needs to obtain the necessary information about the probe and wedge models to be used for your inspection. When you use a P1 connector and a different brand of a probe, you need to manually select the probe and wedge.
To manually select a probe:
Select Wizard > Channel Group > Probe Mode > Next > Probe
Select Manufacturer category > Frequency parameter list > PartNo parameter list
Figure: Select Probe Manufacturer (category) > Frequency (parameter list) > PartNo (parameter list)
Note:
In Phased Array mode, only the wedges that are dedicated to the probe are available by default. The list of dedicated wedges speeds up the wedge selection process.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 54
Defining a Probe (SETTINGS MENU)
This section describes how to define a probe in the P1 device.
To define a probe, go to the Probe Submenu
Navigate to Probe button ( ) and Go to > Probe>Element>Offset
Select Probe Name > Tap the button ( ) to open the list
Figure: Select Probe Manufacturer (category) > Frequency (parameter list) > PartNo (parameter list)
Note: Probe Name> Tap the button ( ) to open the list.
( ) Button to save the set list automatically. ( )
Probe Recent:
When you set up the probe, it brings up a list of frequently used ones.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 55
Skew angle:
Used to define the difference on the index axis between the 0, 0 position of the inspection and the front face of the probe. Negative for skew 90, and positive for skew 270.
Figure: skew angle: 0, 90, 180, 270 degree
Frequency:
Insert the value for Frequency of probe (MHz).
Pulse width:
Used to select the bipolar pulse width value.
Element pitch:
Enter the value for probe-element pitch, which indicates the center to center distance between two adjacent elements.
Figure: Probe-element pitch
Probe separation:
In a case of Pitch & Catch probe, the distance between the element for Tx and the center of the element for Rx.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 56
Figure: Pitch and Catch probe, distance between pitch (left) and catch (right)
A number of element:
Used to define a total number of probe elements. For example: 32, 64 etc. See the following figure.
Figure: N number of probe element number
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 57
Type:
Select the probe type. See the following table as a reference.
Probe type Figure
Single (PAUT)
Conventional
1
Pitch and Catch
TOFD
Transmitter Start/Stop:
Set the transmitter value for Start/Stop according to probe selection. See the following figure, which indicates example of N number of elements.
Figure: Transmitter Start (red), Stop (red)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 58
Receiver Start/Stop:
Set the receiver value for Start/Stop according to probe selection. See the following figure, which indicates the example of N number of elements.
Figure: Receiver Start (red), Stop (red)
Offset:
Scan offset:
Set the probe reference position setting from origin point coordinate in mm unit. See the following figure for reference.
Figure: Scan offset
Index offset:
Set the probe reference position setting from origin point coordinate in mm unit. See the following figure for reference.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 59
Figure: Index offset
DB
b
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 60
Defining a Wedge (SETTINGS Menu)
This section describes how to define a wedge in the P1 device.
To define a probe, go to the Wedge Submenu in Settings.
To manually select a wedge
1. Select Wizard > Channel Group > Probe Mode > Next > Probe
2. Select Manufacturer category > Frequency parameter list > PartNo parameter list
Enable:
Used to allow a user to a setting of the wedge. Tab to check/uncheck ( ) the bar to activate the wedge settings.
Wedge name:
Navigate to Probe button ( ) and Go to > Wedge>Wedge Name
Select Wedge Name > Tap the button ( ) to open the list
Figure: The list of predefined wedges
Select in following order: Manufacturer > Group> Part
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 61
Wedge angle:
Used to specify the angle of the wedge. Select the Wedge Angle and then enter the angle of the wedge. See the following figure for reference.
Figure: Wedge angle
Roof angle:
An angle between probe and wedge in Pitch and Catch probe type only. Enter the value manually.
Figure: Roof angle
L-Wave velocity:
Used to set the longitudinal wave velocity according to the material selected. Value can be automatically changed according to the selected wedge. For example: Olympus –ABW1105 5L128-I3 wedge model: L-wave velocity: 2330 m/s
S-Wave velocity:
Used to set the shear wave velocity according to the material selected. Value can be automatically changed according to the selected wedge. For example: Olympus –ABW1105 5L128-I3 wedge model: S-wave velocity: 1157 m/s
Shape
Height of the first element:
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 62
Select height of the first element, and enter the value (See the figure). This parameter is usually provided by the wedge manufacturer.
Figure: Height indicator (red arrow)
Primary axis position:
Select primary axis position, which indicates distance from front-end of the wedge to the first element of the probe in direction of primary axis, and enters the value (See the figure below).
Figure: Primary axis position (red arrow)
Secondary axis position:
Select secondary axis position, which indicates the distance from front-end of the wedge to the first element of the probe in direction of secondary axis, and enters the value (See the figure below).
Figure: Secondary axis position (red arrow)
Wedge length
Used to specify the length of the wedge to generate accurate focal laws according to
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 63
this specific geometry (See the figure). This value will be automatically filled for generic wedges, but must be manually entered for custom wedges.
Figure: Wedge length
Wedge width
Used to specify the width of the wedge to generate accurate focal laws according to this specific geometry (See the figure). This value will be automatically filled for generic wedges, but must be manually entered for custom wedges.
Figure: Wedge width
Wedge Height
Used to specify the height of the wedge to generate accurate focal laws according to this specific geometry (See the figure). This value will be automatically filled for generic wedges, but must be manually entered for custom wedges.
Figure: Wedge height
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 64
2.5.9-3: Configuring a Focal Law
Selecting the Beam Type
The P1 device needs to obtain the necessary information about the probe and wedge models to be used for your inspection. When you use a P1 connector and a different brand of probe manually select the probe and wedge.
To manually select a beam type and law legs:
Select Wizard > Focal Law > Beam Type > Next
Select Wizard > Law legs > Next
Beam Type:
Sectorial: Provides a multiple-angle scan (see the figure below)
Linear: Provides a linear scan at a configurable angle (see the figure below)
Figure: Beam type – Sectorial / Linear
Law legs:
Set the Legs value according to the number of beam legs in the part.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 65
Figure: Law leg example with Beam Tracing
Note: The Law Legs parameter is only available when the Beam Tracing option is activated.
Material:
See the Material for reference.
Wave:
To select appropriate wave type for your application, select Wave-Longitudinal or Shear.
Note: Since the Material was previously defined (in the Channel Group wizard or by selecting Group/Probe & Specimen), it only appears as a reference in the Focal Law wizard.
Select Wizard > Focal law > …> Beam Angle > Next Beam
Select Angle Start and then enter the minimum angle value for the US beams.
For sectorial scans:
Select Angle Start and then enter the maximum angle value for the US beams
Select Resolution, and then enter the number of degrees between each US beam
Select Focus Position, and them enter the distance at which the US beams focus in the inspected material
Select Aperture Size, and enter the size of the group of active elements.
Figure: Beam angle/resolution/focus position illustration
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 66
Select Wizard > Focal law > ... > Gate A > Next
Select Start position for gate A (according to depth)
The range of gate A appears as line segments over the US beam limit lines.
Set the Range of gate A (according to depth)
Figure: Gate range illustration
Select Wizard > Focal law > ... > Complete > Next/Continue
Select Finish to complete the Focal Law wizard or Select Continue to quit the Focal Law wizard and proceed to the next wizard (Calibration)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 67
2.5.9-4: Calibration
Calibration methods:
Calibration must be performed using a calibration block made of the same material as the probe and the inspected object before performing the test.
Metal path, Depth, Velocity method (see the Appendix B)
The device offers following calibration methods, path, depth and velocity. Select the function calibration method and choose the method by using the rotary knob switch.
Before starting an inspection, you must perform several calibration procedures using a probe, wedge, and calibration block composed of the same material as the part to be inspected.
Following figure shows which calibration mode is currently enabled
Figure: Calibration state indicator of Wizard > Calibration
Icon Description
W Wizard > Calibration > Wedge Delay
V Wizard > Calibration > Velocity
S Wizard > Calibration > Sensitivity
T DAC/TCG > TCG (Encoder)
Go to side menu -> Wizard -> Calibration -> Velocity
Figure: Calibration main menu
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 68
2.5.9-4.1: Wedge Delay Calibration
A wedge delay calibration is a process that enables you to identify the face of the wedge contacting the part, and that establishes a zero position for the entry surface of the part. The calibration procedure requires a calibration block with a reflector.
Step 1: Place the probe on calibration block
Perform the following steps to starts the Wedge Delay calibration wizard
The procedures apply to the following groups, reflectors, and scans.
Step 2: Adjust the parameters of each pages, then click Next to move next page.
Select Wizard > Calibration > Wedge Delay > Gain (dB) > Next
Adjust the Gain (dB) value using the rotary knob or tap hold to change the value.
Select Wizard > Calibration > Wedge Delay > Ref A (mm) > Next
Select (Ref.A, Ref.B) and change the value using rotary knob switch or tap- hold to insert the value manually.
Select Wizard > Calibration > Wedge Delay > Tolerance (mm) > Next
Select Tolerance to specify the tolerance for the measurement of the reference reflector.
Figure: Tolerance line in wedge-delay (green line)
Example: Tolerance line in wedge-delay (green line)
6mm and 14 mm respectively
Tolerance
lines
Envelope
lines
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 69
Select Wizard > Calibration > Wedge Delay > RNG (mm) > Next
Select Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > Wedge Delay > Gate Start (mm) > Next
Used to set the starting position of the selected gate. This position is related to the gate synchronization. The real position of the gate is the synchronization position plus the gate start position (see the example below).
Select Wizard > Calibration > Wedge Delay > Gate Range (mm) > Next
Select Gate Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > Wedge Delay > GH (%) > Next
Select Gate Height, and then enter the height of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch. Tap the NEXT button move on next calibration page.
Select Wizard > Calibration > Wedge Delay > Clear Envelope > Next
Displays a curve with the maximum and minimum amplitudes on the A-scan view. To clear the envelope, tap and hold the A-scan view, then tap Clear envelope on the shortcut menu of Calibration.
Select Wizard > Calibration > Wedge Delay > Reset Wedge Delay > Next
Select Reset Wedge Delay to clear the calibration (wedge delay). This parameter is only available with wedge delay calibration.
Select Wizard > Calibration > Wedge Delay > Apply Wedge Delay > Next
Used to specify the scope for the Wedge Delay and Sensitivity calibration wizard. Reset Wedge Delay has no effect after the Apply button has been tapped.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 70
Step 3: After each parameters are adjusted, move the probe over calibration block to adjust. Check the SA bar is set to 100.0 mm according to 100R method (refer the image below). Once a calibration done, click Finish to finalize the calibration.
Figure: Successful Wedge Delay calibration example
Note: If the calibration is correct, the envelope is between the two red dotted lines.
If the calibration is incorrect, select Clear Envelope to clear the envelope, and select Reset Wedge Delay to start over.
Step 4: Click Next to move to the next calibration section.
Select Wizard > Calibration > Wedge Delay > FINISH > Next
Finish: After configuring the parameters, tap on Finish button to finalize the calibration and move to the next wizard: Velocity.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 71
2.5.9-4.2: Velocity Calibration
The purpose of the ultrasound velocity calibration is to measure the real velocity of ultrasound waves in the material to be inspected. The calibration block must have two known reflectors and be made of the same material as the part to be inspected.
Step 1: Place the probe on calibration block
Perform the following steps to start the velocity calibration wizard
The procedures apply to the following groups, reflectors and scans.
Step 2: Adjust the parameters of each pages, then click Next to move next page.
Select Wizard -> Calibration -> Velocity -> Gain (dB) > Next
Adjust the Gain (dB) value using the rotary knob or tap hold to change the value.
Select Wizard -> Calibration -> Velocity -> Ref A (mm) -> Next
Select Wizard -> Calibration -> Velocity -> Ref B (mm) -> Next
Select (Ref.A, Ref.B) and change the value using rotary knob switch or tap hold to insert the value manually.
Select Wizard -> Calibration -> Velocity -> Angle Start -> Next
Defines angle start value and display at S-Scan view with a position move.
Select Wizard -> Calibration -> Velocity -> Angle Stop -> Next
Defines the Wedge delay with Range, and then enter range with position move Range, and then enter the range of velocity.
Select Wizard -> Calibration -> Velocity -> RNG -> Next
Select Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Tap on NEXT to move on to next menu bar.
Select Gate selection at first stage in order to set the Gate Start, Gate Range, GH parameters.
Select Wizard > Calibration > Velocity > GATE A (B, I) > Next
Select Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > Velocity > Gate Start > Next
Used to set the starting position of the selected gate. This position is related to the gate synchronization. The real position of the gate is the synchronization position plus the gate start position (see the example below).
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 72
Select Wizard > Calibration > Velocity > Gate Range (mm) > Next
Select Gate Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Figure: example) Gate Range: 10.0 mm and 50.0 mm
Select Wizard > Calibration > Velocity > GH (%) > Next
Select Gate Height, and then enter the height of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Figure: example): GH (%): 20% and 40%
Select Wizard > Calibration > Velocity > Clear Envelope > Next
Displays a curve with the maximum and minimum amplitudes on the A-scan view. To clear the envelope, tap and hold the A-scan view, then tap Clear envelope on the shortcut menu of Calibration.
Select Wizard > Calibration > Velocity > VLC (m/s) > Next
Select VLC (m/s) to select modify the velocity parameter value using the rotary knob or tap hold and insert the value (see the example below).
Step 3: After each parameters are adjusted, move the probe over calibration block to adjust. Observe the Measurement bar to make sure SA and SB are correctly set (refer
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 73
the image below). Once a calibration done, click Finish to finalize the calibration.
Select Wizard > Calibration > Velocity > CALIBRATE > Next
Calibrate: After configuring the parameters, tap on Calibrate button to finalize the automatic calibration.
Figure: Correct Velocity calibration example
Note: If the calibration is correct, SA is set to 50.0 mm, SB is set to 100.0 mm.
If the calibration is incorrect, select Clear Envelope to clear the envelope, and start over.
Step 4: Click Next to move to the next calibration section.
Select Wizard > Calibration > Velocity > FINISH > Next
Finish: After configuring the parameters, tap on Finish button to finalize the calibration and move to next Calibration type (Sensitivity Calibration).
Figure: example): perform Finish and move to next Calibration -> Sensitivity
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 74
2.5.9-4.3: Sensitivity Calibration
The sensitivity calibration normalizes the gain for all focal laws to ensure that all focal laws produce a similar amplitude signal for the reference reflector. The calibration procedure requires a calibration block with the reflector.
Step 1: Place the probe on calibration block
Perform the following steps to start the Sensitivity calibration wizard
The procedures applies to the following groups, reflectors and scans.
Step 2: Adjust the parameters of each pages, then click Next to move next page.
Select Wizard > Calibration > Sensitivity > Gain (dB) > Next
Adjust the Gain (dB) value using the rotary knob or tap hold to change the value.
Select Wizard > Calibration > Sensitivity > Ref Amp (%) >Next
Select Ref.Amplitude to specify the level at which the peak amplitude of the signal crossing gate A is plotted when you select the Set the xx.x% parameter.
Select Wizard > Calibration > Sensitivity > Amp Tolerance (%) >Next
Select Amp.Tolerance to specify the position at which the amplitude of the signal gate A is plotted when you select the Set the xx.x% parameter.
Select Wizard > Calibration > Sensitivity > RNG (mm) >Next
Select Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > Sensitivity > Gate Start (mm) >Next
Used to set the starting position of the selected gate. This position is related to the gate synchronization. The real position of the gate is the synchronization position plus the gate start position.
Select Wizard > Calibration > Sensitivity > Gate Range (mm) >Next
Select Gate Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 75
Select Wizard > Calibration > Sensitivity > GH (%) >Next
Select Gate Height, and then enter the height of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > Sensitivity > Clear Envelope >Next
Displays a curve with the maximum and minimum amplitudes on the A-scan view. To clear the envelope, tap and hold the A-scan view, then tap Clear envelope on the shortcut menu of Calibration.
Select Wizard > Calibration > Sensitivity > Reset Sensitivity >Next
Select Reset Sensitivity to clear the calibration (sensitivity delay). This parameter is only available with sensitivity calibration.
Step 3: After each parameters are adjusted, move the probe over calibration block to adjust. Observe the Auto Sensitive Calibration window (refer the image below). Once a calibration done, click Finish to finalize the calibration.
Figure: Building the envelope for the Sensitivity calibration
Note: If the calibration is correct, the envelope is between the two green dotted lines.
If the calibration is incorrect, select Clear Envelope to clear the envelope, and select Reset Sensitivity to start over.
Select Wizard > Calibration > Sensitivity > Apply Sensitivity >Finish
Used to specify the scope for the Wedge Delay and Sensitivity calibration wizard. Reset Sensitivity has no effect after the Apply button has been tapped.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 76
Correct calibration should look like the following example. SA is now set to 100.0 mm.
Figure: Successful Sensitivity calibration example
Step 4: Click Next to move to the next calibration section.
Select Wizard > Calibration > Sensitivity > FINISH >Next
Finish: After configuring the parameters, tap on Finish button to finalize the sensitivity and move to the next wizard: TCG calibration.
Figure: perform Finish and move to next wizard ->TCG
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 77
2.5.9-4.4: TCG Calibration:
This section describes how to calibrate the TCG (Time-Corrected Gain) sizing function. To perform this procedure, a calibration block with a number of identical holes at known depths or thicknesses.
Step 1: Place the probe on calibration block. The procedures applies to the following groups, reflectors and scans.
Step 2: Adjust the parameters of each pages, then click Next to move next page.
Selec Wizrard > Calibration > TCG > Gain(DB) > Next
Use the knob switch or tap hold to adjust the Gain (dB) value to change the value.
Select Wizard > Calibration > TCG > Echo Type >Next
Select Echo Type to determine the echo type either Metal Path/Depth.
Select Wizard > Calibration > TCG > Ref Amp >Next
Select Ref.Amplitude to specify the level at which the peak amplitude of the signal crossing gate A is plotted when you select the Set the xx.x% parameter.
Select Wizard > Calibration > TCG > Amp Tolerance > Next
Select Amp.Tolerance to specify the position at which the amplitude of the signal gate A is plotted when you select the Set the xx.x% parameter.
Select Wizard > Calibration > TCG > Reset TCG > Next
Select Reset TCG to restart the TCG calibration procedure in case of incorrect calibration or settings.
Select Wizard > Calibration > TCG > Gate Start (mm) >Next
Used to set the starting position of the selected gate. This position is related to the gate synchronization.
Select Wizard > Calibration > TCG > Gate Range (mm) >Next
Select Gate Range, and then enter the length of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 78
Select Wizard > Calibration > TCG > GH (%) >Next
Select Gate Height (%), and then enter the height of the time-base range according to the weld thickness. Reduce or increase the range using the rotary knob switch.
Select Wizard > Calibration > TCG > Interval >Next
Select Interval (mm) which defines a distance between the selected Ref.A and next Ref.A. Adjust an value using the rotary knob or insert an value manually.
Select Wizard > Calibration > TCG > Follow Peak >Next
Select Follow Peak to change an angle of maximum peak of signal in AScan. Select On/Off to enable or disable the function.
Select Wizard > Calibration > TCG > Prev / Next
Gate displays the maximum peak signal. Apply TCG # 1 and move to # 2. If you need to modify # 1, you can # 1 Prev to fix it.
Select Wizard > Calibration > TCG > #1, #2, #3...
It shows the part where TCG shows current position.
Select Wizard > Calibration > TCG > Apply
Set the red line to match the green point in the green range.
Select Wizard > Calibration > TCG > Reset >Next
Select Reset to start a calibration process from beginning. This function is also available in Wedge Delay/Sensitivity/TCG calibration.
Select Wizard > Calibration > TCG > Clear Envolope
Displays the maximum and minimum amplitude curves in the Scan View. Select Clear Envelope when the amplitude curve is created incorrectly while calibrating.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 79
Step 3: After each parameters are adjusted, move the probe over calibration block to adjust. Observe the Auto Sensitive Calibration window (refer the image below). Once a calibration done, click Finish to finalize the calibration.
Figure: Building the envelope for the TCG calibration
Note: If the calibration is correct, the envelope is between the two green dotted lines.
If the calibration is incorrect, select Clear Envelope to clear the envelope, and select Reset TCG to start over the process.
Select Wizard > Calibration > TCG > Apply Reflector >Next
Correct calibration should look like the following example.
Figure: Successful TCG calibration example
Step 4: Click Next to move to the next calibration section.
Select Wizard > Calibration > TCG > FINISH >Next
Finish: After configuring the parameters, tap on Finish button to finalize the sensitivity and move to the next wizard: Encoder.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 80
2.5.9-4.5: Encoder Calibration
The following section describes how to perform calibration when encoder is used.
Note: Before starting this wizard, an encoder must be properly connected to the device properly.
Step 1: Perform the following steps:
Select Wizard > Encoder Type
Used to select the desired type of inspection. The available choices are:
Scan: Used to specify the source of the data positioning for the scan axis.
Index: Used to specify the source of the data positioning for the index axis.
Select Wizard > Calibration > Encoder > Step Start
Adjust the Step Start (physically move the encoder to a marked point) value using the rotary knob.
Select Wizard > Calibration > Encoder > Step End
Adjust the Step End (current encoder position) value using the rotary knob.
Select Wizard > Calibration > Encoder > Distance
Distance (mm): physically move the encoder to a second marked point, and note the distance traveled. Adjust the Distance value using the rotary knob or tap hold to change the value.
Step 2: Select Wizard > Calibration > Encoder > CALIBRATE
The encoder calibration is performed.
Select Wizard > Calibration > Encoder > Scan Resolution
Scan resolution (steps/mm): Number of steps per unit distance of scan axis encoder. Adjust the Scan Resolution value using the rotary knob or tap hold to change the value.
Select Wizard > Calibration > Encoder > Finish
If the resolution is correct, tap on FINISH.
Select Wizard > Calibration > Encoder > Reset
If resolution is not correct, tap on RESET to recalibrate.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 81
2.6 Submenu: Settings
Figure: Submenu SUB TOOLBARS Menu
The right menu MAIN TOOLBARS appears as following figure. Tap on button to open Submenu SUB TOOLBARS. By tapping each button will display different page opens respectively.
2.6.1 Submenu SUB TOOLBARS: CHANNEL SETTING ( )
Tap on ( ) button to open channel settings page. Tap on buttons, then change the value or options with rotary knob switch with open channel settings
Channel setting contains following functions:
Channel: Reset
Channel: Add
Channel: Modify
Channel: Delete
Channel: Modify
Channel: Upload
Channel: Sequence Up
Channel: Sequence Down
Reset: Tap on Add ( ) button to reset the channel settings. When you tap on the button, following message box will appear. Tap ok to reset x to cancel.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 82
Figure: Example of resetting the channel group
Add: Tap on Add ( ) button to add a new channel. When you tap on the button, following message box will appear. Tap ok to add x to cancel.
Figure: add channel group
Channel delete: Tap on Delete ( ) icon delete your selection. When you tap on the icon, following info box will appear. Tap ok to delete, x to cancel.
Figure: channel delete confirmation
Channel modify: Tap on Modify ( ) icon to current settings on external storage. When you tap on the icon, same set file save box will appear. Tap ok to save a change.
Figure: Channel modify
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 83
2.6.2 Submenu SUB TOOLBARS: PULSER ( )
Tap on ( ) button to open Pulser various settings, including General, Pulse,
Receive and Digitizing. Then change the value or options with a rotary knob switch.
2.6.2-1 Sub Menu: General
Gain: Used to modify the gain value to optimize the echo amplitude.
ReF.Gain: Used to activate and modify Ref.Gain
Figure: General Setting List
Ref.Gain If you want activity, click on to activate.
Figure: Not Applied state Figure: Applied state
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 84
Start range: Start position (mm) and the end position of data to collect.
Figure: Illustration of start range
Range: Last position (mm) and the end position of data to collect.
Figure: Illustration of range
Wedge Delay: Delay time of the ultrasound waves between the wedge and probe. Unit : nsec
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 85
2.6.2-2 Sub Menu: PULSE
Pulse Width: Width can be set from 50 nsec to 2000 nsec value. Width can be s
Select the pulse width function and set the value using knob rotary switch.
Figure: Example of pulse width
Voltage: Pulser voltage can be set through this function by using the rotary knob to select the required output voltage between 25 V ~ 200 V.
Figure: Example of voltage
Burst count: The burst count selects the number of cycles in each burst. Select the value by using rotary knob range between 1 to 5.
Figure: Example of burst count
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 86
PRF-Recurrence (Hz): Recurrence value indicates the number of times an initial pulse is triggered, which 1000 Hz is generated and collected a pulse once every 0.001 (1/1000) seconds. By using this function, you can determine the value of the pulse repetition frequency.
Max recurrence input method: automatically sets the Max Recurrence value.
Wave type: Allows you to select the type of wave. LW (longitudinal) and SW(shear) are available.
Figure: Classification of waves (LW and SW)
Auto FSH: Automatically adjust a gain value to the registered setting value (80%) of FSH (Full Screen Height)
Figure: example of performing auto FSH
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 87
2.6.2-3 Sub menu: RECEIVE
Digitizing Frequency (with data compression): Set the frequency used when converting analog signal to digital signal. Performs data compression automatically when digitizing and selects the largest value of the sample that skips to digitizing changes.
Input range: 50 MHz, 25MHz, 12.5 MHz, 6.25 MHz and 3.125 MHz
Performs data compression automatically when digitizing, and selects the largest value among the samples that skip to digitizing changes.
Figure: Illustration of digitizing frequency
Compression: Displays the compression factor of the current group. Ex) 1,2,4,8 and 16. The Compression factor is related to the range and quantity of point.
Filter: Allows you to select the filter type to apply, according to receiver.
Low pass filter: Passes all signals lower than Xx MHz
High pass filter: Passes all signals higher than X Mhz
Band pass filter: Passes only signal between higher than X Mhz and lower than Y MHz.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 88
Figure: Examples of 2MHz low pass, 5MHz high pass, and 1 to 5 MHz band pass filters
Rectifier: This function lets you choose the rectification mode of the echo pulses according to your application. The following options can be selectable.
Fullwave: All half-waves are displayed above the baseline on the screen.
RF: No rectification
POS halfwave: Only positive half-waves are displayed on the screen
NEG halfwave: Only negative half-waves are displayed on the screen
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 89
2.6.2-3 Sub menu: DIGITIZING
Samples: Number of samples collected from single angle. Sample count calculation:
Sample count = Range (mm)/(US velocity/Sample frequency/2)
Figure: Illustration of Samples
Resolution: Distance of one sample resolution.
Resolution = (End Range – Start Range)/Samples
Sample Size: The size of the collected sample 1 bit, 16 bits = 2 bytes.
Frame Size calculation: Example: 30 to 61 degrees, 2500 samples per vector, total number of vectors is 32. Therefore, 32 * 2500 * 2 bytes = about 16k bytes
Acquisition rate: The maximum rate (Hz) at which ultrasonic signals can be collected. If you move the scanner faster than the Acquisition Rate, you will not be able to acquire the acquisition signal.
Data Throughput:
Data Throughput = Frame Data Size x Acquisition Rate
Example) 500 Kbyte x 20 Hz = 10 Mb/s. Also this value (10 Mb/s) defines the maximum data throughput range of P1 device.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 90
2.6.3 Submenu SUB TOOLBARS: BEAM ( )
2.6.3-1 Sub-Menu: BEAM
This sub-menu allows you to set the values for refraction, resolution and aperture size. See the following figure.
Figure: Beam types-Sectorial/Linear
Type: Setting for selecting beam type
Sectorial: Collecting beams of different angles
Linear: Collecting beams of same angle
Refraction angle: This setting allows you to start angle, stop angle and resolution for a sectorial type. For example: Start from 30 degree, Stop at 61 degrees of 1.0-degree resolution generates 32 Beam Focal Laws. See the following figure.
Figure: Beam formation example (Start/Stop/Resolution)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 91
Aperture size: When beam type is set to linear mode, the number of elements that make up the beam.
Example) When 16 elements are set to 16 apertures, the total 16 beam focal laws will generate. See the following example and figure.
1 Index: 1~16
2 Index: 2~17..
16 Index: 16~32
Figure: aperture size of 32
2.6.3-2 Sub-Menu: FOCUS
This sub-menu allows you to set the values for focus type and focus position.
Focus Type:
True Depth: All focal laws are focused at Depth value
Half Path: All focal laws are focused at Distance value
Projection: All focal laws are focused on a predefined focus plane (Offset)
Figure: Focus type
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 92
Focus position: The US depth at the highest position. Unit is in mm.
Focus Plane Position Offset: A position (from wedge reference point to the focus plane offset) of the focus plane. Unit: mm
Figure: Illustration of focus plane offset
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 93
2.6.4 Submenu SUB TOOLBARS: PROBE ( )
Tap on ( ) button to open probe setting page. Tap on buttons, then change the value or options with rotary knob switch.
2.6.4-1 Sub Menu: PROBE
Probe Name: By pressing ( ) button to select the probe name using the existing database.
Select the probe in following order: Manufacturer > Frequency > PartNo
For example: Manufacturer (Olympus) > Frequency (1.5)>PartNo (1.5L16-A10)
Skew Angle: Maximum collectable velocity
Figure: Illustration of skew angle
Frequency: Frequency input setting of a probe. Unit: MHz
Element Pitch: Center to center distance between two adjacent elements.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 94
Figure: Element pitch
Probe Separation: Center to center distance between two adjacent probes. Example: Pitch-and-Catch probe, a distance between pitch (left) and catch (right). See the following figure.
Figure: Example (Probe Separation)
2.6.4-2 Sub Menu: ELEMENT
Element setting for Transmitter & Receiver according to the probe type. Probe Element Index starts from 1 and supports up to 128 elements.
Number of element: Defines the total number of a probe element. Example: 32, 64, …
Figure: Number of element
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 95
Type: Options for probe type. Types: Single (PAUT), Pitch and Catch, Conventional and TOFD. See the figure.
Figure: Probe types
Connector: Select connector to select external converter. Connector types: PA (Phased-Array) and BNC. The option available when Conventional, TOFD probe types are selected.
Transmitter: Set the configuration of transmitter element index. For example) Start 1, Stop 16
Receiver: Set the configuration of receiver element index. For example) Start 16, Stop 32
2.6.4-3 Sub Menu: OFFSET
Scan Offset: Probe reference position setting from origin point coordinate
Figure: Illustration of scan offset
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 96
Index Offset: Probe reference position setting from origin point coordinate
Figure: Illustration of scan offset
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 97
2.6.5 Submenu SUB TOOLBARS: WEDGE ( )
This sub-menu allows you to set the values for refraction, resolution and aperture size. See more from following section: Wizard -> Defining a Wedge.
2.6.6 Submenu SUB TOOLBARS: TRANSMITTER/RECEIVER ( )
Tap on ( ) button to open TX/RX setting page. Tap on buttons, then change the value or options with rotary knob switch.
2.6.6-1 Sub Menu: Tx DELAY
Transmit delay of each transmitter for beam focusing. The unit is in nanosec. Transmitter active select (Check to Activate all). The user may also check in custom order.
Azimuthal: Meaning of refracted angle. Also grid in this option indicates tx delay of the selected refracted angle in nano seconds.
Exit Point: Ultrasound signal generated from the probe passes through the wedge and the starting point of the specimen, also indicates the starting position of the actual vector. See the following figure.
Figure: illustration of exit point
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 98
Focal Point Z: The point where the beam is focusing. Unit is in mm. See the figure for reference.
Figure: Illustration of focal point
2.6.6-2 Sub-Menu: Rx DELAY
Receive delay of each transmitter for beam focusing. Unit is in nanosec. Receive active select (Check to Activate all). User may also check in custom order.
Figure: Illustration of Rx Delay
Azimuthal: Meaning of refracted angle. Also grid in this option indicates Rx delay of the selected refracted angle in nano seconds.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 99
2.6.6-3 Sub-Menu: TCG
After setting TCG Calibration, you can easily see the exact position and DB value of TCG point.
Picture: TCG point list indicating
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 100
2.6.7 Submenu SUB TOOLBARS: MECHANICAL SETTING ( )
Tap on ( ) button to open mechanical setting page. Tap on buttons, then change the value or options with rotary knob switch.
2.6.7-1 SCAN:
Scan Unit type (mm/deg): This indicates the scan unit type, which measured in mm/deg in case of flat/cylindrical specimen.
Select the scan page-Scan unit type, and adjust the value using the rotary knob switch.
Scan position: Set the value for specifying the position of a scan before performing. Select the scan-page scan unit type, and adjust the value using the rotary knob switch.
Figure: Scan position (red arrow: scan direction)
Scan start: Set the value for specifying the start position of a scan before performing.
Select the scan page-Scan start, and adjust the value using the rotary knob switch.
Figure: Scan start (red arrow indicates scan start)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 101
Scan stop: Set the value for specifying the stop position of a scan before performing.
Select the scan page - Scan stop, and adjust the value using the rotary knob switch.
Figure: Scan stop (red arrow indicates scan stop position)
Scan step: Set scan page-Scan step, and adjust the value using the rotary knob switch.
Figure: Scan step (red arrow indicates the scan step)
2.6.7-2 INDEX:
Index setting values can be changed in index page. Select the axis (axs) function and select each bar, and you can change required values using the rotary knob switch and by tapping to check/uncheck the option:
Index Unit type (mm/deg): This indicates the index unit type, which measured in mm/deg in case of flat/cylindrical specimen. Select the index page-index unit type, and adjust the value using the rotary knob switch. This indicates the index unit type, which measured in mm/deg in case in
Index position: Set the value for specifying the position of an index before performing.
Select the index page-Index position, and adjust the value using the rotary knob switch.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 102
Figure: Index position (red arrow indicates index position)
Index start: Set the value for specifying the position of an index before performing.
Select the index page-Index position, and adjust the value using the rotary knob switch.
Figure: Index position (red arrow indicates index position)
Index stop: Set the value for specifying the position of an index before performing.
Select the index page-Index position, and adjust the value using the rotary knob switch.
Figure: Index stop (red arrow indicates the index position)
Index step: Set the value for specifying the position of an index before performing.
Select the scan page-Scan position, and adjust the value using the rotary knob to adjust
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 103
the values.
Figure: Index step (red arrow indicates the index step)
2.6.8-3 ENCODER
Encoder setting values can be changed here, in Enc tab page, where unit is steps/mm. Select the Encoder (enc) function and select each bar, and you can change required values using the rotary knob switch.
Axis Reverse: When Scan/Index encoder is recognized as a different axis, changing Axis reverse, encoder axes can be changed to each other.
Figure: Axis reverse
Scan Reverse: Enable to change scan encoder direction (+, -)
Scan resolution: Adjust the value of many steps performed per mm in scan direction. Select the encoder page-Scan resolution, and adjust the value using the rotary knob switch.
Scan resolution type: steps/mm
Index Reverse: Enable to change index encoder direction (+, -)
Index resolution: Adjust the value of many steps performed per mm in scan direction. Select the encoder page-Scan resolution, and adjust the value using the rotary knob switch.
Index resolution type: step/mm
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 104
2.6.8 Submenu SUB TOOLBARS: SPECIMEN/WELD SETTING ( )
Tap on ( ) button to open mechanical probe settings page. Tap on buttons, then change the value or options with rotary knob switch.
Figure: Specimen / Weld setting pages
2.6.8-1 Sub-Menu: SPECIMEN
Name
Select the specimen name from predetermined list or insert the name
Material
Define the type of material that used for specimen, from predetermined list
L-Wave velocity
Ultrasound velocity of specimens (Longitudinal wave)
Figure: Longitudinal Wave
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 105
S-Wave velocity
Ultrasound velocity of specimens (Shear wave)
Figure: Shear Wave
Shape type
Select Flat to change the value of width, height, length
Select Cylindrical (inspection from ID) to change the value of specimens Inside Diameter, Outside Diameter, Length.
Ref.angle (deg): The refracted angle in degree can be calculated by material sound velocity and Snell’s law, which not exceeds than 90 degrees.
Parameter adjustment range: -85 to 85 degrees
Select the Ref.angle function and use the rotary knob switch to set the value.
2.6.8-2 Weld
Enable/disable the wedge list and choose the weld from list
Check ( ) to enable setting menu with external
Weld type: Click here to refer more about weld type.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 106
2.6.9 Submenu SUB TOOLBARS: SETTING LOAD / SAVE ( )
Tap on ( ) button to open setting load and save page. Settings can be easily saved as .SET file on-device storage or external USB storage. Tap on buttons, then change the value or options with rotary knob switch.
2.6.9-1: Setup (file format: SET)
Device storage selection ( ): Select as main storage as files to be saved. Also this function shows how much storage used and available for save/load.
External USB selection ( ): Select as external storage as files to be saved. Also this function shows how much USB storage used and available for save/load.
File save ( ): Tap on save button to save settings.
Add/copy SET file ( ): This function allows to add and copy SET file
Modify the selected folder and file name ( ): This function allows to add and copy SET file
Delete the selected folder and file ( ): This function allows you to eliminate particular folder and files.
Load the SET file ( ): Load the predetermined SET file to the device
Refresh the list ( ): When external USB devices or file/folders not displayed properly, refresh the list using this function.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 107
2.6.9-2: Image (file format: PNG)
Device storage selection ( ): Select as main storage as files to be saved. Also this function shows how much storage used and available for save/load.
External USB selection ( ): Select as external storage as files to be saved. Also this function shows how much USB storage used and available for save/load.
Modify the selected folder and file name ( ): This function allows to add and copy PNG files.
Delete the selected folder and file ( ): This function allows you to eliminate particular folder and files.
Refresh the list ( ): When external USB devices or file/folders not displayed properly, refresh the list using this function.
2.6.9-3: Data (file format: IQV)
Device storage selection ( ): Select as main storage as files to be saved. Also this function shows how much storage used and available for save/load.
External USB selection ( ): Select as external storage as files to be saved. Also this function shows how much USB storage used and available for save/load.
Modify the selected folder and file name ( ): This function allows to add and copy IQV files.
Delete the selected folder and file ( ): This function allows you to eliminate particular folder and files.
Refresh the list ( ): When external USB devices or file/folders not displayed properly, refresh the list using this function.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 108
2.6.9-4: Layout
Device storage selection ( ): Select as main storage as Layout files to be saved. Also this function shows how much storage used and available for save/load.
External USB selection ( ): Select as external storage as files to be saved. Also this function shows how much USB storage used and available for save/load.
Modify the selected folder and file name ( ): This function allows to add and copy Layout files.
Delete the selected folder and file ( ): This function allows you to eliminate particular folder and files.
Refresh the list ( ): When external USB devices or file/folders not displayed properly, refresh the list using this function.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 109
2.6.9-5: Setting
Figure: File Name Setting
Note: Save Probe, Wedge and NO to make it easier to find
Data File Preset: Data Save name set
Image File Preset: Image Save name set
Danger: Problems may arise when you clear the set command.
Note - Create the desired name #SET #_#DATE #_#NO # Example of
Settings: Probe (5L32A) Wedge (N60S) Setting up
Example: PAUTP1 #SET #_#DATE #_#NO # Save
PAUTP1 (5L32AN60S_20180530_1)
2.6.10 Submenu SUB TOOLBARS: DEVICE SETTING ( )
This part contains general procedures that use of function from the File and Device menus, and it describes commonly performed tasks that have an overall impact on the P1 device software interface.
Tap on ( ) button to open device setting page. Tap on buttons, then change the value or options with rotary knob switch.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 110
2.6.10-1: Display
User interface color options
The user interface contains two selectable theme mode, which default (dark), green (light) and white (light) modes. More reference from page 64.
Theme
Choose from drop-down list of Theme to change the UI color. Available colors: Black, Green and White.
Figure: UI theme color options
A Scan color
Choose from drop-down list of A Scan color to change the color. Available colors: Yellow, Blue, Green, Cyan, Red, Magenta, Red, White and Black
Figure: A Scan color options
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 111
A Scan BG color
Choose from drop-down list of A Scan Background color to change the background color. Available colors: Black, White and Gray
Figure: A Scan background color options
C-Scan merge
( ) Displays the maximum value of the entire Scan view in real-time mode.
Figure: C-Scan merge Check list
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 112
Color map
Choose from drop-down list of Color map color to change the S-Scan color. Available colors: Corrosion, Gray, Rainbow, RFRectified and TOFD
Figure: Color map options
Grid
Select a type of Grid for A Scan view background appearance.
Off: no grid background
Grid 1: Wide dotted grid background type 1
Grid 2: Narrow dotted grid background type 2
Figure: Grid options
LCD Brightness: Slide the bar shift -> right or shift-> left to adjust the LCD brightness.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 113
2.6.10-2: Gate Line
Horn
The Horn contains the selection used to define the horn conditions. Alarms are useful for highlighting the occurrence of specific conditions.
In scan view window, when amplitude curve passes over by certain threshold and LED alarms will turn on/Beep sound feedback given by once or a certain amount of period. Select the setting horn and choose the setting by using the rotary knob switch.
Off: No beep sound is generated
Once: Generates beep sound only once when gate is detected
Duration: Generates continuous beep sound when gate is detected
Gate Detection:
Choose from drop-down list of Gate Detection to change the way to detect a gate. You can select among two kinds of gate detection (See the figure below):
Peak: Detect the highest signal position of the gate
Edge: Detect edge signal position of a gate
Figure: Gate detection (Peak and Edge)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 114
2.6.10-3: Device
Display with Alarm gate shows the Version/Date/Time the rotary knob to adjust the date settings with Version/Date and Device > Category
Information
Displays the detailed information of device and software version. Information type: Software version / Max channel/ Max element
Date
This part describes how to set date and time on the P1 device. Tap on Date button and use the rotary knob to adjust the device’s date settings.
Figure: Date setting example
Time
Tap on time and use the rotary knob to adjust the gate shows the Version/Date/Time the rotary knob to adjust the time.
Figure: Time setting example
Save button
Tap on Save button, and select the save button type. By setting this feature, push the physical SAVE button to save particular file mode.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 115
Figure: Example of Save button option
Saving formats and types:
Setting file: Save all settings. The file format is SET.
Screen capture: Capture the screen activities. The file format is PNG.
Time: Automatically saves files without encoder scan. The file format is IQV.
Language
Change the display language. Available languages: English/Korean
2.6.10-4: Preference
Figure: Preference list
Default Setting : Restore settings to their original defaults
Minimize P1 : Select minimize button to minimize a program window to the taskbar
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 116
2.7 Submenu: Gate option toolbar
The Gate option toolbar explained following table. Tap on button to open Submenu: Gate option toolbar.
Submenu GATE option toolbar:
Following chart explains the OPTION in option toolbar.
Button Description Scan view
Switch between Amplitude / US (% to mm)
Flip scan view window Vertically
Flip scan view window Horizontally
Specimen overlay Visible/Invisible
Choose the Maximum A-Scan
Show/hide the Reference bar
Show/hide the Measurement bar
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 117
Show/hide the Gate
Smoothing filter When selected, this option
enables the smoothing filter.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 118
2.8 Submenu: Scan view toolbar
The right menu Scan view TOOLBARS explained in following table. Tap on button to open Submenu Scan view TOOLBARS. Tap on buttons, then change the value or options with rotary knob switch.
Following chart demonstrates the four types of scan view window.
Scan view Description Appearance
Single A scan view window
Single C scan view window
Displays A and C scan view at same time
Drag the gate lines in scan view window and zoom-in/zoom out of
scan view window
Displays the Drag the gate lines in scan view window and zoom-
in/zoom out of scan view window
Adjust number of scan view to show
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 119
2.9 Submenu: Shortcut toolbar (horizontal)
The right menu Gate Submenu toolbar appears as the following figure. Tap on button to open: Gate sub-menu. Tap on buttons, then change the value or options with rotary knob switch. The right menu Main Toolbars appears as following
Submenu: Shortcut toolbar >>Gate sub-menu:
Tap to add Gate line in scan view
Tap to delete Gate line in scan view
Tap to choose the Screenshot* feature.
Tap to choose the amplitude value settings
Tap to Zoom-in the scan view
Tap to Zoom-out the scan view
Tap to Scale to fit
Tap to pause all pulse & receive features
Tap to the maximum A-scan (parameter settings)
*This feature enables you to save image in device and external USB device is inserted
To capture the screen of P1 device:
Tap on ( ) button to take a screenshot, and make sure USB device is connected to the P1 device. Since the saving location predefined in Settings >> File management >>
Image >> Device / USB storage. You can explore your saved images from File Management setting to copy, delete or modify. Note that image file format is .PNG.
Figure: Screen capture and save
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 120
Switch to P1 Scan pane screen:
It operates the program for a more efficient work environment ( )
The program is operated to prevent unnecessary touches and to make the scan window look bigger during scan operation.
Figure: Switch to full screen when clicked
: You can move the window arbitrarily when clicked.
: Switch from full screen mode to default screen mode when clicked
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 121
Submenu: Shortcut toolbar >>Gate sub menu:
Tap to Reference bar (with long press)
Tap to Measurement bar (with long press)
Tap to zoom, drag (with long press)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 122
APPENDIX A
DIGITIZING METHOD
A digitizing frequency default value initially set to 50 MHz, 25 MHz, 12.5 MHz, 6.25 MHz and 3.125 MHz are available. Since the maximum sample count in A-scan is default as 16K (16384).
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 123
APPENDIX B
MANUAL CALIBRATION METHODS
The P1 device can be calibrated manually either automatically (Wizard). Following methods are used for calibrating the P1 device manually. The sound velocity and calibration range have to be set and the probe delay taken into consideration depending on the material and dimensions of the test piece.
Methods (click on title):
Metal Path calibration method
Depth calibration
Velocity calibration
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 124
Metal Path calibration method:
Metal path calibration method uses IIW type II calibration block in order to find the most peak point. (Tap on Calibration method to choose the metal path selection from manual calibration in Settings, see the page).
Step 1: Tap on button and select Type >> Metal Path calibration method (Refer the Figure)
Figure: 100R / 50R method on IIW type II calibration block
Step 2: Tap on button to open scan view window, then tap on to add gate on scan view.
Figure: Example of adding a gate (red)
To find the Peak point of frequency:
Following types of will display in following scheme. Adjust Ref.A to 100.0 mm (100
R) can be calibrated in following way.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 125
Step 3: Tap on Auto FSH (%) button to automatically adjust a gain value to the registered setting value (80%) of FSH (Full Screen Height).
80%
Figure: Perform Auto FSH %
Step 4: Tap on Calibrate button to measure reference position. In case of adjusting 100mm (100R) in Reference A.
Figure: Perform Calibration
As shown figure above, SA (100.0 mm) value is automatically calculated after pressing Calibrate button.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 126
Depth calibration method:
While metal path calibration method used IIW Type II calibration block, the depth calibration method uses Navship test block regarding to find a depth.
Figure: 50R depth hole example on Navship test block
It is a similar method to each other, only difference is metal path method uses 100R whereas depth calibration sets depth (For example, 50mm). See the example in the figure on next page.
Step 1: Tap on button and select Type >> Depth calibration method.
Step 2: Tap on button to open scan view window, then tap on to add a gate on scan view. Adjust Ref Depth A to 50.0 mm in terms of 50R method.
Figure: Example of adding a gate (red)
Step 3: Tap on auto Auto FSH (%) button to automatically adjust a gain value to the registered setting value (80%) of FSH (Full Screen Height).
80%
Figure: Perform Auto FSH %
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 127
Step 4: After adjusting Ref.Depth A to 50 mm (50R), then tap the button to perform Calibrate.
Figure: Perform depth calibration
As shown figure above, DA (25.0 mm) value is automatically calculated after calibration
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 128
Velocity calibration method
Reference Position A (mm): Locating position that accessed as Reference A for calibration (note that the ex. 50.0 mm (50 R) applied)
Reference Position B (mm): Locating position that accessed as Reference B for calibration (note that the ex.100.0 mm (100R) applied) (Refer figure below)
Figure: 100R/50R method on IIW type II calibration block
Metal path Calibration : 100 R for Ref.A
Depth Calibration : 50 R for Ref.B
Velocity Calibration : 50R for Ref.A/100R for Ref.B
Step 1: Tap on button and select Type >> Velocity calibration method (refer figure below)
Step 2: From scan view window, then tap on to add two gates on scan view. Then set Ref.A to 50.0 mm, Ref.B to 100.0 mm.
Figure: Example of adding a two gate line (red, green)
Step 3: Tap on Auto FSH (%) button to automatically adjust a gain value to the registered setting value (80%)
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 129
80%
Figure: Perform Auto FSH % (adjust to 80%)
Step 4: After signal adjusted by FSH function, tap on Calibrate button.
Figure: Perform calibration
As shown figure above, SA (50.0 mm), SB (100.0 mm) value is adjusted, calculated VLC is 3238 m/s after calibration.
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 130
APPENDIX C
MANUAL TCG/ ADC CALIBRATION
Methods (click on title):
TCG calibration
DAC calibration
Example)
Assume there is 20 mm, 40 mm, 60 mm hole on test block (refer the figure below). Since scan view signal set to 100% to default, selecting Auto FSH function adjust each signal curve to 80% for TCG/DAC curve drawing.
40mm
60mm
20mm 80%
80%
80%
Figure: IIW Type II calibration block/ Peak point to FSH for DAC/TCG
Calibration with example)
To perform TCG calibration:
1. Place the probe on calibration block (Navship test block recommended)
2. Select ( ) button from Wizard, then select
a) Select Wizard -> Calibration -> DAC/TCG
b) Select Mode DAC/TCG -> TCG
c) Add Gate Line on A Scan
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 131
Figure: Set Gate on A Scan for TCG peak point
d) Choose ( ) to perform Auto FSH (adjust to 80%)
Figure: Auto FSH for TCG point
e) Choose (Add Point) to perform TCG drawing
Figure: TCG drawing on A Scan
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 132
To perform ADC calibration:
1. Place the probe on calibration block (Navship test block recommended)
2. Select ( ) button from Wizard
a) Select Wizard -> Calibration -> DAC/TCG
b) Select Mode DAC/TCG -> DAC
c) Add Gate Line on A Scan
Figure: Set Gate on A Scan for DAC peak point
d) Choose ( ) to perform Auto FSH (adjust to 80%)
Figure: Auto FSH for DAC point
e) Choose (Add Point) to perform DAC drawing
Figure: DAC point drawing on A Scan
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 133
DEEPSOUND P1 Complete User Manual Release 1.0
Important NOTICE:
Seongsan Lab reserve the right to make corrections, modifications, enhancements, improvements, and other changes to its products and services at any time and to discontinue any product or service without notice. Customers should obtain the latest relevant information before placing orders and should verify that such information is current and complete. All products (hardware and software) are sold subject to Seongsan Lab’s terms and conditions of sale supplied at the time of order acknowledgement.
Any questions with DEEPSOUND P1 device?
Here at Seongsan lab, we understand that sometime there may some problem may arise. In that case, we strongly recommend that:
Read this document deeply, since this document provide all the information you will need.
If you do not find any proper solution from this document, feel free to send us e-mail or mail to addressing below, since testing and other quality control techniques are used to the extent Seongsan Lab deems necessary to support this manual.
Mailing Address:
Seung-il venture tower 802, Gasan Digital Road 2-gil 136,
Geumcheon-gu, seoul, Korea
E-mail: [email protected]
Copyright © 2018, Seongsan Lab
DEEPSOUND P1: Complete Guide Release 1.1, 2018
Copyright⒞ Seongsan Lab. All rights Reserved 134