SCIENTZ-09 LAB BLENDER
Operator Manual
ISO9001: 2008
ISO13485--2003
NINGBO SCIENTZ BIOTECHNOLOGY CO., LTD
ADD:No.65 Mujin Road, National Hi-tech Park, Ningbo, China
TEL:0086-574 88352466 FAX: 0086-574-87113393
M.B: 0086-13967868094 Post Code:315013
E-mail: [email protected]
Website:www.scientz.com www.scientzbio.com
《医疗器械生产许可证》编号:浙食药监械生产许 20090011 号
No.of Production License for Medical Device: No. 20090011, Zhejiang
shi yao jian xie
医疗器械注册证书编号:浙甬食药监械(准)字 2009 第 1410019号
No. of Registration Certificate for Medical Device: No. 1410019 (2009)
ZY shi yao jian xie (zhun) zi
Product Standard No.:YZB/ZY 090-2009
1. Operation Principle
Generally, unprocessed microbial samples are mushy, farinose or
block-like. To facilitate analyzing and inoculating samples, samples shall
be refined to be liquid-like. Basic principle of operation: original samples
(bigger ones shall be cut to be block-like. Dimension: 10×10mm) shall be
put into homogenize bag together with certain liquid or solvent.
Homogenizer will hammer sample bag to generate pressure. The pressure
will arouse surge and accelerate mixing, so microbial components in the
solvent can be equally distributed. This homogenizer can be used to
homogenize brain, kidney, liver, spleen etc.
Homogenize bag made from special materials and used to store
original samples and solvent has gone through asepsis. Seal the bag
with the rail on the top of the homogenizer door can ensure no
leakage of liquid while raped by relevant plate and balanced mixing
of the sample in the bag by the pressure from the rapping. Generally,
the homogenize speed range is 3~~~~12 T/Sec and the time is 0.1~~~~
99min, however, customers can adjust them to fit actual situations.
180 sec and 8 T/Sec are acceptable to the homogenizing of livers of
mouse and rat, for instance.
I. Application and Scope
It is widely used in the asepsis of animal tissues, biological samples
and food (such as pig stomach, meat, fish, vegetables, fruits and
biscuit), and cosmetics. Thanks to its easy operation, convenience and
low time-consuming, it is especially qualified to be used in the
preparation for test samples of various microbes. As single-use
sterilized homogenize bag is soft, pollution-free & damage-free, with
no heating-up, no need to carry out asepsis and wash vessels,
containing glass windows for easy observation, and adopting
qualified and environmental stainless steel or similar materials, it can
avoid cross contamination and protect the operators. Besides, by
homogenizing various tissues of tumor, such as liver cancer, intestinal
cancer, gastric cancer and breast cancer, etc, abundant individual
cells can be obtained at the speed of 2×105
in every 2 min, as well as
adjusting the process of softening these tissue cells.
II. Major Technical Data
01.Timing Range: 0.1~~~~99 min (as 0.1 min)
02.Rap Speed: 3.0~~~~12T/sec (as 0.1T/sec), above 10T/sec (as 1T/sec)
---The system will adjust the rap speed automatically as long as
the actual strikes are displayed.
03. Data Memory: Store in 10 groups (for the convenience of
customers)
04. Adjustable Temperature : Room temperature~~~~60℃℃℃℃
05. Sterilization of the Trunk: Rap the ultraviolet sterilization lamp
inside the trunk.
06. Capacity: 3~~~~400ml
07. Material of Rap Plate: Stainless Steel
08. Rated Power: 120W
09. Dimension: 420×245×340mm
10. Sterilized Homogenize Bag: 17*30cm
11. Weight: 18.5kg
12. Power: 220V/50Hz
III. Operation Procedures and Setting
Open the packing case, and carry out the homogenizer, then put
it on a stable working table near power source.
1. Plug in (AC220V, 50Hz) first, then power on.
2. Start the equipment, the boot screen will appear.
Fig. 1 Boot Screen
(1) Set of Storage Group
Fig.2 Set of Storage Group
a. Press key, “L1” showed on
the display window refers to the
first storage group, and then
press key to switch to L1~~~~
L10. (See Fig. 2)
Fig.3 Set of temperature
b. Press key to select
storage group, and set the
temperature by keys and ,
after that, press key to
store the data you set.
c. Adjustable Temperature:
Room Temperature~~~~60℃℃℃℃. (See
Fig. 3)
(2) Set Time
Fig. 4 Set Time
a. Press keys and to set
the time.
b. Timing Range: 0.1~~~~99min
(3) Set Rap:
Fig.5 Set Rap
a. Press keys and to set
the rap speed.
b. Rap Range: 3.0~~~~12T/sec (as
0.1T/sec), above 10T/sec (as
1T/sec).
3. Standby Screen (Press key)
Indication of the Temperature Control:
1. Green: The set temperature has been
reached.
2. Red: Heating
Fig.6 Standby Screen
(1) Modify the timing and rev (for method, refer to aforesaid).
(2) Sterilization (can only be started by pressing key on the
standby screen).
The automotive sterilization period is 5min.
Press key to stop the sterilization process.
(3) The temperature displayed is the actual temperature inside the
trunk and heat preservation begins when the temperature set is
reached.
(4) Press key to the transfer screen of data stored.
4. Short press key to the standby screen, and press it again
to work. Remember: Long press is not effective.
(1) Start on, the indicator light is on ( )
(2) Adjust the rev (rap speed) at real-time.
(3) Short press to pause ( indicator light
flashes), press again to return.
(4) If you long press during running, the equipment will
stop work and back to the standby screen.
5. Check whether the under edge of the homogenizer door has been
buckled to the rail at the bottom of homogenize trunk. Remember,
the under edge of the homogenizer door should be buckled perfectly
to the rail.
6. Hold the door switch and open the door fully with your right hand,
then, put the homogenize bag with samples and solvent into the
homogenize trunk with your left hand. The under edge of the
homogenize bag should be no lower than the bottom of rap plate.
Emerge the edge of the homogenize bag opening at the door cover
levelly. After that, turn the door switch with you left hand until you
hear “Click”. By this time, the door and top rail seal the homogenize
bag by clamping it.
7. When the equipment finishes running, take out the homogenize
bag at its upper edge with your left hand and open the homogenizer
door with your right hand. Choose the rev again as you open the
homogenizer door.
8. You’d better stop all the panel operations or cut off the power for
your security.
ⅣⅣⅣⅣ Attentions:
1. Make sure the plug has been inserted into the socket tightly, any
looseness may result the computer control panel in a halt. If it is
halted, please switch off the rear panel and restart after 3min.
2. Do not attempt to open the homogenizer door when the rap plate is
running, otherwise, the sample liquid will splash. If you open the
door by force, the equipment would stop running automatically and
the display screen will stay still at that time. Then, close the door and
press the key , it will go on the remaining tasks automatically.
3. Please cut off the power and pull out the plug when you finish
using the equipment.
4. The bottom of the homogenizer door is empty to cope with
incidents, such as breakage of homogenize bag, and to wipe away
spilth easily, therefore, put your hands into the equipment from the
bottom while it is running is absolutely forbidden as your hands may
be injured.
5. Check whether the homogenize trunk is clean with no other
unnecessary matters before it goes to work, so that any accident and
damage to the homogenize bag can be avoided.
6. Any material of hard lump, bone or ice type are not suitable to be
put into homogenize bag as they can do damage to it.
7. Do not expose the equipment and homogenize bag to the sunshine
directly. To protect them from aging, please store in places with no
sunshine or no UV, especially homogenize bag.
6. Maintenance:
Clean the equipment, hammer plate, especially the homogenize bag
with clean cloth. Cleaning shall be carried out when power is switched
off.
7. Malfunction and Trouble Shooting
(1) If the machine fails to work after the door is soundly closed and
there is no display on the display screen, users shall examine the power
line to make sure it is connected normally and the fuse is in good
condition.
(2) If the machine sounds abnormal, examine bolts and hammer
plate to see whether it is soundly connected. If lubricant dries out, users
shall clean the hammer pole and hammer cover in time and lubricant
grease shall be re-filled in these parts.
8、、、、Warranty Scope:
(1) Under normal service and maintenance condition, end-users can
enjoy one year’s free maintenance and service from the day the machine
is delivered.
(2) When malfunction occurs, please deliver the machine to the HQ of
Ningbo Scientz Biotechnology Co., Ltd. Delivery payment shall be
prepaid.
As to functional disfigurement resulted from inappropriate use, or
operating and maintaining against the instrument or caused by
unauthorized staff, Ningbo Scientz Biotechnology Co., Ltd. will not take
on the responsibility to repair for free. Payment for repairing will be
charged according to specific conditions.
ⅨⅨⅨⅨ. Packing List
Main Unit 1 Set
Power Cord 1
Sterilized Homogenize Bag 500
Sample Holder (Stainless Steel) 1
3A Fuse 2
Rap glove 1 Pair
Instruction Book 1
Certificate of Qualification/ Warranty Card 1/1
Examples (I)
Test Report of Granular Chinese Medicine
Add: NINGBO SCIENTZ BIOTECHNOLOGY CO., LTD
Date: October 13th, 2008
Equipment: scientz-09 sterilized homogenizer, rap plate coated with
rubber sleeve, silicon rubs of 2mm put inside the sides of glazed door.
The test includes two programs:
No. 1
1. Dip two of the granular medicine into boiled water for 10 min, and
then put them (medicine and water) together into a homogenize bag, after
that, put another homogenize bag outside the first one.
2. Finally, place the medicine of step 1 into the homogenizer to be raped
between the rubber sleeve and silicon rub for 2.5 min.
3. Rev: 6T/Sec
Result: No leakage occurs to both two homogenize bags.
No. 2
1. Dip two of the granular medicine into cold water for 10 min, and then
put them (medicine and water) together into a homogenize bag, after that,
put another homogenize bag outside the first one.
2. Finally, place the medicine of step 1 into the homogenizer to be raped
between the sleeve and silicon rub for 3 min.
3. Rev: 6T/Sec
Result: The inside bag is broken, outside one is normal, with uneven
surface, color of Program 1 is darker than Program 2.
Conclusion: Homogeneous effect can be achieved by dipping into cold
water for 10 min, with two homogenize bags and coating a silicon rub of
2mm.
Examples (II)
Obtain Large Scale of Individual Cells by Sterilized Homogenizer
Add: NINGBO SCIENTZ BIOTECHNOLOGY CO., LTD
Date: May 27th, 2008
Equipment: Scientz-09 Sterilized Homogenizer
Place the following tissues of tumor procured from operations on the
plate, such as gastric cancer, colon cancer and breast cancer, etc. Wash
them by physiological saline with 100U/ml gentamicin for three times to
remove bacteria on the tissues, and then add 10ml physiological saline
per 2g. Tear them into small pieces, as 2-5mm, afterward, put these pieces
into homogenize bags.
Operate as the instructions in the manual of sterilized homogenizer, as
8T/Sec for 2-8min, rest 1min after 1min’s operation, and record the
accumulative time. Different tissues may need different period of time.
Intake these tissue mixtures into test tubes for a standing of 3-5min, after
that, select out the clear cell suspension which includes a variety of
scattering individual cells, separate it for 5min with a centrifuge at the
speed of 1000 r/min. Add in 20% calf serum nutrient solution for culture
(directly is allowable) and control the cell population around 2×105ml.
Cells for this purpose are enough although the total population may have
change with the size of tissue.
These individual round cells in the following picture is the scattered
individual cells we needed, including glandular cells and epithelioid cells,
these irregular spindle cells are scattered myofibroblast, also known as
scattered individual cells. Both kinds of cells are intact and plump, with
excellent activity.
Picture of Pig Stomach in 8min