Sampling, Sieving, Imaging, & Correlating the Data Between the Two Presenters: Kyle James – Verder Scientific Inc. Gert Beckmann – Retsch Technology

Transcript of Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for...

Page 1: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sampling, Sieving, Imaging, &

Correlating the Data Between the Two

Presenters:Kyle James – Verder Scientific Inc.

Gert Beckmann – Retsch Technology

Page 2: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Areas of DiscussionTOPICS

2© Retsch GmbH

1.) Sample Acquisition: Sampling Methodology & Techniques

2.) Sieve Analysis: Principles / Techniques / Concerns

3.) Dynamic Image Analysis: Methodology / Technique / Correlation

Page 3: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Quality Control / R&DApplications

3© Retsch GmbH

Defined product properties

Detected product properties

Agreement

Certain properties are to be achieved when products are manufactured.These depend on the particle size and the particle distribution.

Particle size/distribution = product property

Page 4: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Particle Sizing ExampleWhy it Matters

4© Retsch GmbH

Example: CoffeeThe particle size determines important taste properties

Too coarsely ground coffee: the brewing process is accelerated and gives a watery cup of coffee

Too finely ground coffee: too many aromatics, acids and bittering agents are dissolved, the filter could be blocked

Page 5: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Acquistion

5© Retsch GmbH

The bigger the particle size, the more difficult it is to attain the representative part sample

Page 6: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sampling Process

6© Retsch GmbH

Transports(belt,container, train and truck)

Accumulation(filling, feeding)

Page 7: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Amount

7© Retsch GmbH

Qmin = 0.07 * dmax / ρ

Qmin = mass of a single sample [dm3]

dmax = maximum particle size [mm]

0.07 = factor [kg/mm]

ρ = bulk density [kg/dm3]

Minimum volume (Qmin) of a single samplewith grain sizes (dmax) < 120 mm

DIN 51701 part 2 - Sampling of solid fuels -

Page 8: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Standard Deviations of Various Sample Divisions Methods

8© Retsch GmbH

qualitative variation

Sample divider PT 100

Random sampling

Sample splitter

Cone and quartering

qualitative variation0 1 2 3 4 5 6 7 8 9 10 %

Page 9: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Coning and Quartering

9© Retsch GmbH

Page 10: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Stream of material

Single sample

Sample Extractor Chute

10© Retsch GmbH

Page 11: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Extractor Bucket

11© Retsch GmbH

Stream of material

Single sample

Page 12: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Extractor

12© Retsch GmbH

Stream of material

Single sample

Page 13: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Dividers from Retsch

13© Retsch GmbH

Page 14: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample Splitters

14© Retsch GmbH

Page 15: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Rotary Tube Divider PT 200

15© Retsch GmbH

Page 16: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Laboratory Sample Divider PT 100

16© Retsch GmbH

Page 17: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

4 accurate particle size analyses= 4 different results!

Importance of sample divisionSample division

17© Retsch GmbH

xc_min [mm]0.5 1.0 1.5 2.0 2.5 3.00

10

20

30

40

50

60

70

80

90

Q3 [%]

randomsampling

sample material: standard sand

*,* = nominal values of standard sand

Page 18: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

xc_min [mm]0.5 1.0 1.5 2.0 2.5 3.0 3.50

10

20

30

40

50

60

70

80

90

Q3 [%]

Importance of sample divisionSample division

18© Retsch GmbH

sample splitter

sample material: standard sand

*,* = nominal values of standard sand

Page 19: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Importance of sample divisionSample division

19© Retsch GmbH

xc_min [mm]0.5 1.0 1.5 2.0 2.5 3.00

10

20

30

40

50

60

70

80

90

Q3 [%]

rotating sample divider

PT 100

4 accurate particle size analyses= 4 similar results!

sample material: standard sand

*,* = nominal values of standard sand

Page 20: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sieveability of particlesSieving

20© Retsch GmbH

Free flowing particles

Van-der-Waals forces

Fluid bridges

Electrostatic Forces

Agglomerated particles

+ -

Page 21: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sieving aidsSieving

21© Retsch GmbH

• Aerosil

• talcum powder

• aluminum oxide

solid

• wet sieving

• degreasing:

benzinealcohols

liquid

• chain rings

• brushes

• cubes

• rubber balls

• agate balls

• steatite balls

mechanical

Page 22: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Test sievesthat comply with standards Test sieves

22© Retsch GmbH

If sieve analysis is used for quality controlwithin the context of DIN EN ISO 9000:2000

then both the sieve shaker and thetest sieves must be subjected to

test agent monitoring.

w = mesh widthd = wire diameter

w

w

Ø d

Ø d

Tolerance for mean value (Y):The mean value of the mesh widthmust not differ from thenominal value w by more than thetolerance ± Y.

Technical requirements & testingaccording to ISO 3310

Page 23: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

100 200 300 40040 50 60 70 80 90particle size x[µm]

10

20

30

40

50

60

70

80

90

Q3 [%]

10

20

30

40

50

60

70

80

90

Q3 [%]

44%

+Y 66,4

36%

-Y 59,6

Importance of mesh widthConsequences of tolerances Test sieves

23© Retsch GmbH

63 µmtolerance ± Y = 3,4 µm

40%

Page 24: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Real Mesh WidthReal Mesh

24© Retsch GmbH

x [µm] 200 400 600 800 1000 1200 0

10

20

30

40

50

60

70

80

90 Passing [%]

Sample-1__xc_min_002.rdf Sieving-upper-range-S1.ref

Excellent correlation

when using the real mesh opening sizes

Page 25: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Influence of Mesh WidthMesh measurements

25© Retsch GmbH

1400µm 1400µm 1429.5µm

Nominal Sieve Mesh = 1400µm Real Sieve Mesh >1400=1455

only beads < 1400µm

will pass the sieve mesh

beads > 1400µm will not pass the sieve mesh

Upper mesh size range ~1455µmsieve No. 03033531 (nominal 1400µm)

Theory: Reality:

Mesh sizes warp Mesh sizes weft

Page 26: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Calibrationcertificate

26© Retsch GmbH

Nomial mesh width

Tolerances

Number ofmeasured apertures

Mean mesh width

Standard deviation σ

Wire diameter

Page 27: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sample AmountSieve Analysis

27© Verder Scientific Division

4 mm2 mm1 mm

500 µm250 µm125 µm

63 µm45 µm

collecting pan

Page 28: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sieving Methods forparticle size determination Basics

28© Retsch GmbH

Horizontal sievingVibratory sieving Air jet sievingTap sieving

dry wet

vibratory

horizontal

tap

air jet

AS 200 jet AS 200 tap AS 200 AS 300 AS 400 AS 450 control

Page 29: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Selection of amplitudeVibratory sieve shakers

29© Retsch GmbH

0

10

20

30

40

50

60

70

80

90

100

Q3(

x) /

%

10 50 100 500 1000 5000Particle size / µm

Quartz,sieving time: 5 min.

AMP- 2 mmAMP- 0.5 mmAMP- 1.2 mm

Page 30: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Statistical resonanceVibratory sieve shakers

30© Retsch GmbH

T = periodic time of sieve bottom vibration

particle

sieve bottom

t

A

T T

Page 31: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Sieve analysis procedureProcess

31© Retsch GmbH

Page 32: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Potential Concerns & IssuesSampling & Sieving

32© Retsch GmbH

1.) Sampling problems:Sampling Technique (pouring and “spooning“)Material Behavior & Composition

2.) Sieving problems: Sieve OverloadSieving Technique / Technology UsedNominal Size Real Mesh SizeMaterial Behavior & Composition

3.) Scale Problems:Low Resolution (0.1g of sample)Not Sensitive Enough (sieves vs actual)

Page 33: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingImage Analysis

33© Retsch GmbH

x [µm]200 400 600 8000

10

20

30

40

50

60

70

80

90

Q3 [%]

RT669_3993_Z_LB_05%_xc_min_001.rdfRT669_RT_3993.ref

Page 34: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Particle SizeMorphology

34© Retsch GmbH

xcmin

xc min

“width”

A

A‘ = Ax a

rea

“diameter overprojection surface”

xarea“length”

xFe max

xFemax

CAMSIZER results are

compatible with

sieve analysis

Page 35: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Image ProcessingComparison

35© Retsch GmbH

x [mm]0.1 10

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Q3

Tinovetin-B-CA584A_BZ_xc_min_002.rdfSyngenta-1mm-2min-Sieb.ref

--- width measurement

-*- Sieving

comparisonCAMSIZER-measurement xc min (red)and sieving * (black)

xc min

Page 36: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Ellipsoid ParticlesComparison

36© Retsch GmbH

x [mm]1.0 1.25 1.5 1.75 2.00

10

20

30

40

50

60

70

80

Q3 [%]

rice

xcmin

A‘ = A

x are

a

A

Red CAMSIZER curve of particle width gives excellent correlation

at the black sieve points

Page 37: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Lenticular ParticlesComparison

37© Retsch GmbH

x [mm]0.2 0.4 0.6 10

10

20

30

40

50

60

70

80

Q3 [%]

Sample A_BZ_0.2%_xc_min_001.rdfSample A_.ref

Page 38: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingLenticular Particles

38© Retsch GmbH

x [mm]0.2 0.4 0.6 10

10

20

30

40

50

60

70

80

Q3 [%]

Sample A_BZ_0.2%_xc_min_001.rdfSample A_.ref

Page 39: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingCubes / Angular

39© Retsch GmbH

x [µm]200 400 600 8000

10

20

30

40

50

60

70

80

90

Q3 [%]

RT669_3993_Z_LB_05%_xc_min_001.rdfRT669_RT_3993.ref

Page 40: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Angular ParticlesCoal, Sand, Sugar

40© Retsch GmbH

x [µm]200 400 600 8000

10

20

30

40

50

60

70

80

90

Q3 [%]

RT669_3993_Z_LB_05%_xc_min_001.rdfRT669_RT_3993.ref

Page 41: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Angular ParticlesShell Limestone

41© Retsch GmbH

xc_min [mm]0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.80

10

20

30

40

50

60

70

80

90

Q3 [%]

Muschelkalk_xc_min_001.rdfRT3204_Muschelkalk_Sieb.ref

*Sieve Analysis -CAMSIZER

Page 42: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingAngular

42© Retsch GmbH

x [µm]200 400 600 8000

10

20

30

40

50

60

70

80

90

Q3 [%]

RT669_3993_Z_LB_05%_xc_min_001.rdfRT669_RT_3993.ref

x [µm]200 400 600 8000

10

20

30

40

50

60

70

80

90

Q3 [%]

RT669_3993_Z_LB_05%_xc_min_001.rdfRT669_RT_3993.ref

angular particles without fitting

CAMSIZER-measurement xc min (red)sieve analysis * (black)

angular particles with Q3-fitting

Page 43: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingDistribution

43© Retsch GmbH

Two samples with different width of distributionbut …

… with similar shape

(= same product type)

Page 44: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Limitations of Old ProceduresOld vs New

44© Retsch GmbH

xc_min [mm]1.0 1.5 2.0 2.5 3.00

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Q3

CAMSIZER Elementary – Fittingxc_min [mm]1.0 1.5 2.0 2.5 3.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Q3

xc_min [mm]1.0 1.5 2.0 2.5 3.00

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Q3

Old fittingmethods

Page 45: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingElementary Fitting

45© Retsch GmbH

Single class:Taken from the sieve stackand measured in CAMSIZER

For creating a CAMSIZER Elementary

fitting file use the more narrow sample

(green)

Elementary fitting with single (narrow) sieve class.It can be the sieve with the highest amount, one above or one below

Page 46: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Digital Imaging SievingElementary Fitting

46© Retsch GmbH

xc_min [mm]1.0 1.5 2.0 2.5 3.00

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Q3

Elementary - Fitting

New elementary fitting with single (narrow) sieve class and entire distribution

{

Page 47: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Measurement of Single ClassSingle Size

47© Retsch GmbH

xc_min [mm]1.0 1.5 2.0 2.5 3.0 3.50

10

20

30

40

50

60

70

80

Q3 [%]

Try to geta single sieve class

as narrow as possiblein the middle

of the distributionof the sample

Page 48: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Elementary Fitting in PracticeSet-up

48© Retsch GmbH

Page 49: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Elementary Fitting in PracticeSet-up

49© Retsch GmbH

Try to get a singlesieve class as narrowas possible ~ in the

middle of thedistribution of the

sample

Try to get a sample ofyour product as

narrow as possible

Try to get a sample ofyour product as

narrow as possible

Try to get a single sieve classas narrow as possible

in the middle of the distributionof the sample

Page 50: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Elementary Fitting in PracticeSet-up

50© Retsch GmbH

Page 51: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Shell LimestoneExample

51© Retsch GmbH

xc_min [mm]0.4 0.6 0.8 1.0 1.2 1.40

10

20

30

40

50

60

70

80

90

Q3 [%]

Muschelkalk_xc_min_001.rdfRT3204_Muschelkalk_Sieb.ref

Page 52: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Multimodal DistributionExample

52© Retsch GmbH

x [mm]0.2 0.4 0.6 0.8 1.0 1.2 1.40

10

20

30

40

50

60

70

80

90

Q3 [%]

SU_Demo02-05_Standaard_BZ_LB_Gl15_03%_xc_min_001.rdfStandaard.ref

x [mm]0.2 0.4 0.6 0.8 1.0 1.20

10

20

30

40

50

60

70

80

90

Q3 [%]

SU_Demo02-05_Mix_BZ_LB_Gl15_03%_xc_min_001.rdfMix-Opzak-and-stMG4502.ref

Excellent sievecorrelation evenwith bimodal or

multimodal distributions

Page 53: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

SummaryCorrelation Wrap-Up

53© Retsch GmbH

CAMSIZER® Elementary Fitting:• For samples with similar shape• Fitting of different width of distribution possible

(even multimodal distributions)• Applications: Sand, Sugar, Fertilizer, Minerals, Plastics,

Foodstuffs ... and many more• Creating a CAMSIZER® correlation method

only takes 20 minutescompared to 3 hours with competitive instruments

Samples with varying particle shape, e.g. abrasiveswill need different fitting files or CAMSIZER Meta-Fitting®

Page 54: Sampling, Sieving, Imaging, Correlating the Data Between ... · If sieve analysis is used for quality control. within the context of DIN EN ISO 9000:2000. then both the . sieve shaker

Thank you for yourattention !