S-ROM NOILES ROTATING HINGE KNEE SYSTEM
Transcript of S-ROM NOILES ROTATING HINGE KNEE SYSTEM
S-ROM®
NOILES™ ROTATING HINGE KNEE SYSTEM
SURGICAL TECHNIQUE
2 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
TABLE OF CONTENTS
SURGICAL TECHNIQUE Key Surgical Steps Summary 4
S-ROM NOILES Rotating Hinge Knee System 6
The S-ROM Hinge System Overview 7
Incision and Exposure 8
Intra-operative Evaluation 10
Initial Preparation of the Tibia 11
Preparation of the Metaphyseal Bone 13 – Tapered Reamer
Proximal Tibial Resection – Tapered Reamer 14
Preparation of the Metaphyseal Bone – Broach 16
Tibial Trial Assembly 18
Preparation of Femoral Diaphysis 19
Reaming the Medullary Canal 20
Preparation of the Metaphysis – Sleeve Use 22
Femoral Preparation – Distal Resection 26
Femoral A/P and Chamfer Cuts 29
Femoral Box Cuts 31
Final Preparation of the Tibia 32
Femoral Trial Insertion 33
Trial Reduction 36
Implant Assembly – Tibia 37
Tibial Implantation 38
Implant Assembly – Sleeve and Stem Use 39
Bearing and Hinge Pin Insertion 42
Initial Patellar Resection 43
Patella Reaming 45
Patella Drilling 46
Trial Reduction and Implantation 47
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 3
APPENDICES
ORDERING INFORMATION
Appendix 1: The Cemented Tibial Stem Extensions 48
Appendix 2: Step Wedge Preparation 51
Appendix 3: Thick Tray Preparation 54
Implant Listing 55
Compatibility Chart 59
Instruments 60
KEY SURGICAL STEPS SUMMARY
Patella PreparationFinal Trialing
Distal Femoral ResectionFemoral Medullary Canal Preparation
Tibial Medullary Canal PreparationIncision and Exposure
4 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Implantation
Femoral Preparation - A/P and Chamfer Cuts
Femoral Box Cuts
Tibial Resection Tibial Trial Assembly
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 5
S-ROM NOILES ROTATING HINGE KNEE SYSTEM
The S-ROM NOILES Rotating Hinge features:
· S-ROM Femoral Components available in three sizes
· 7 degree physiological valgus, fixed in the femoral component
· Deep femoral trochlear groove
· Modular porous sleeves to accommodate bone defects of the Engh Type 2 and Type 3 classification and allow possible bone ingrowth
· Available with both cemented and press-fit slotted stems for both femur and tibia
· Broad, congruent contact areas between femoral and tibial components to best distribute surface and sub-surface stresses in the polyethylene
· A rotating hinge that accommodates axial rotation, reducing stresses at the bone cement/implant interfaces
6 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
THE S-ROM HINGE SYSTEM OVERVIEW
The MBT Revision Knee System is comprised of the following components:
· Tibial Components are available in eight sizes, 1, 1.5, 2, 2.5, 3, 4, 5 and 6
· Tibial Metaphyseal Sleeves are available in 29 mm (cemented or porous), 37 mm, 45 mm, 53 mm and 61 mm sizes (M/L dimension)
· Tibial Wedge Augmentation Components: Step Wedge in 5, 10 and 15 mm thicknesses
· 75, 115 and 150 mm Fluted Universal Stem lengths in 10 to 24 mm diameters in 2 mm increments
· 30 and 60 mm Cemented Universal Stem lengths in 13 mm diameters. 90, 120, 150 Cemented Tapered Universal stem lengths in 13 mm diameters
· Thick Trays are available in three different sizes (2, 3 and 4) and two different thicknesses (+15 mm and +25 mm)
· Accepts Rotating Platform hinged insert from the LPS™ Limb Preservation System, which is compatible with the S-ROM NOILES Rotating Hinge Femoral Component and LPS Femoral Component
The S-ROM Hinge Knee System is comprised of the following components:
· Hinged Femoral Component is available in three sizes, X-Small, Small and Medium
· Femoral Metaphyseal Sleeves are available in 20 mm (cemented only), 31 mm, 34 mm, 40 mm and 46 mm sizes (M/L dimension), and can be used with or without a stem
· 5 and 10 mm Distal Femoral Augments
· 75 mm, 115 mm and 150 mm Fluted Universal Stem lengths in 10 mm to 24 mm diameters in 2 mm increments
· 30 mm and 60 mm Cemented Universal Stem Lengths in 13 mm and 15 mm diameters
· 90 mm, 120 mm, and 150 mm Tapered Cemented Universal Stem lengths in a 13 mm diameter
· 90 mm Tapered Cemented Universal Stem length in a 15 mm diameter (Must be used with a sleeve)
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 7
INCISION AND EXPOSURE
Initial IncisionWhen possible, follow the scar from the primary procedure (Figure 1). Where parallel incisions are present, the more lateral is usually preferred, as the blood supply to the extensor surface is medially dominant. Where a transverse patellectomy scar is present, the incision should transect it at 90 degrees.
Where there are multiple incision scars or substantial cutaneous damage (burn cases, skin grafting, etc.), one may wish to consult a plastic surgeon prior to surgery to design the incision, determine the efficacy of pre-operative soft tissue expansion and plan for appropriate soft tissue coverage at closure.
Capsular IncisionThe fascial incision extends from the rectus femoris proximal margin to the distal margin of the tibial tubercle following the patella’s medial border, maintaining a 3-4 mm cuff for reapproximation of the vastus medialis aponeurosis at closure (Figure 2). Where mobilization of the extensor mechanism and patella is problematic, extend the skin and capsular incisions proximally.
Figure 2
Figure 1
8 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
INCISION AND EXPOSURE
Occasionally an early retinacular release is indicated to assist with patellar eversion. Where eversion difficulties persist, a quadriceps snip, a proximal inverted quadriceps incision (modified V-Y) or a tibial-tubercle osteotomy may be indicated. Perform appropriate ligamentous release based upon pre-operative and intra-operative evaluation. Release fibrous adhesions to re-establish the suprapatellar pouch and medial and lateral gutters (Figure 3). In many revision cases, the posterior cruciate ligament will be absent or non-functional; when this is the situation, excise any residual portion. Exercise care when everting the patella. Frequently, subluxing the patella laterally is adequate. Doing so will help avoid patella tendon avulsion.
Implant Extraction from the Primary ProcedureTake care to preserve as much bone as possible. To this end, assemble a selection of tools, including thin Osteotomes, an Oscillating Saw, a Gigli Saw, a highspeed Burr and various extraction devices, but many cases will require only the thin Osteotome. Carefully disrupt the bone/cement or bone prosthesis interface before attempting extraction (Figure 4).
Disengage the implanted components and extract as gently as possible, in such manner as to avoid fracture and unnecessary sacrifice of bone stock. Where the entire prosthesis is to be replaced, it is advantageous to remove the femoral component first, as this will enhance access to the proximal tibia. Clear all residual methyl methacrylate with hand (chisels) or power tools.
Figure 3
Figure 4
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 9
INTRA-OPERATIVE EVALUATION
The surgeon should establish two anatomic conditions to facilitate revision arthroplasty: the level of the joint line and the disparity in the flexion and extension gaps.
Joint Line Evaluation
In an average knee in full extension, the true joint line can be approximated in reference to several landmarks.
· It lies 12–16 mm distal to the femoral PCL attachment
· It lies approximately 3 cm distal to the medial epicondyle and 2.5 cm distal to the lateral epicondyle
· It lies distal to the inferior pole of the patella (approximately one finger width)
· Level with the old meniscal scar, if available
Additional pre-operative joint line assessment tools include:
1) Review of original pre-operative roentgenogram of the Total Knee Arthroplasty (TKA)
2) Review of roentgenogram of contralateral knee if non-implanted
11 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Figure 5
INITIAL PREPARATION OF THE TIBIA
The Tibial Alignment SystemWhen pre-operative evaluations indicate that Fluted Stem Extensions, Metaphyseal Sleeves or Wedges are required, it is recommended that the proximal tibia be prepared with reference to the position of the I.M. Rod.
Note: Where a Cemented Stem Extension is indicated, see Appendix 1 (page 48).
Place the knee in maximal flexion with the patella laterally retracted and the tibia distracted anteriorly and stabilized. Release fibrosis around the tibial border or excise as required to ensure complete visualization of its periphery.
Approximate the location of the medullary canal with reference to pre-operative anterior/posterior (A/P) and lateral X-rays and to the medial third of the tibial tubercle.
Introduce a 9 mm drill into the canal to a depth of 2–4 cm. Avoid cortical contact (Figure 5).
4 cm
2 cm
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 11
Reaming the Medullary Canal Assemble the Straight Reamer to the T-Handle. If power reaming, it will be necessary to attach the modified Hudson Adapter to the Straight Reamer. The shaft of the reamer contains markings in 25.4 mm (1 in) increments. Each marking is numbered to use as a reference when reaming to the appropriate depth. Fluted stem lengths are available in 75, 115 and 150 mm. Determine the length and diameter of the prosthetic Stem Extension with Templates (Cat. No. 2178-30-100) applied to pre-operative X-rays.
Use the Reamer Depth Chart (Figure 6) to determine the appropriate mark on the reamer for canal reaming depth. Another option to determine reamer depth is to measure the trial assembly against the reamer and note the corresponding depth mark for reaming. Sequentially open the canal with progressively larger reamers until firm endosteal engagement is established (Figure 7).
Note: Simple cortical contact should not be construed as engagement.
The fixed relationship of the reamer to the cortices ensures the secure fit of the appropriate reamer and, subsequently, the corresponding fluted stem. It is equally important to not over-ream osteopenic bone. While reaming the proximal tibia, pay close attention to the reamer to assure that it is centrally located to the exposed proximal tibial surface. Eccentric reaming can occur, which could lead to undersizing of the tibial component.
The size of the final reamer indicates the diameter of the implant stem. The fluted stems are available in even sizes (10 through 24 mm). Perform final reaming with an even-sized reamer. The final implant will have a .4 mm press-fit versus the reamer and a .5 mm press- fit versus the Stem Trials.
Note: Refer to Appendix 1 (page 48) for cemented stem preparation.
T-Handle
Straight
Reamer
INITIAL PREPARATION OF THE TIBIA
Figure 6
Figure 7
MBT Revision Tray Reamer Line Depth
Press-Fit Stems
75 mm 2
115 mm 3
150 mm 4
Cemented Stems
30 mm 1
60 mm 2
90 mm 2.5
120 mm 3.5
150 mm 4
12 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Tibial Resection
Plane Notches on the Drill
PREPARATION OF THE METAPHYSEAL BONE – TAPERED REAMER
For Diaphyseal Engaging Stem andMetaphyseal Filling Sleeve
Attach the appropriately sized Stem Trial to the end of the MBT Revision Tapered Reamer.
Note: Assembly of the Stem Trial may be aided by the pre-attachment of the T-Handle to the MBT Revision Tapered Reamer.
Taper ream to the planned proximal tibial resection level (Figure 8). When finished reaming, the notches on the Drill should line up with the planned proximal tibial resection level.
Note: Use the “cemented” Tapered Reamer when requiring a cement mantle around the cone of the MBT Revision Tray or when utilizing a sleeve. Use the Press-Fit Tapered Reamer when line-to-line fit is desired around the cone of the MBT Revision Tray and a sleeve will not be utilized (Figure 10). Use End-Cutting Primary Reamer (Cat. No. 2178-63-199) when a stem or sleeve will not be used.
Note: To avoid Stem Trial disengagement, do not reverse ream.
At this point, intra-operatively determine if a Metaphyseal Sleeve will be used.
Note: Metaphyseal Sleeves are ideal to provide filling of Engh Type 2 or 3 defects in revision TKA. The steps of the Metaphyseal Sleeve also provide progressive loading of the bone with porous coating, which enhances fixation.
If a Metaphyseal Sleeve is selected, see page 16 in order to broach the metaphyseal bone.
If a Metaphyseal Sleeve will not be used, see the following page to prepare for the proximal tibial resection.
Figure 8
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 13
PROXIMAL TIBIAL RESECTION – TAPERED REAMER
Attach the 2 degree Tibial Cutting Block to the I.M. Tibial Referencing Device. Attach the I.M. Tibial Referencing Device to the shaft of the Tapered Reamer. Position the I.M. Tibial Referencing Device with the pre-attached 2 degree Cutting Block onto the shaft and allow it to descend to the proximal tibial surface. Since considerable bone stock may have been sacrificed in the primary TKA, minimize the amount resected: no more than 1-2 mm from the most prominent tibial aspect, managing residual defects of the contralateral tibial aspect with either prosthetic augment or bone graft.
Resection is based on tibial deficiency and the level of the joint line. Compensate deficiencies with sleeves, wedges and/or bone grafts. Advance the cutting block to the anterior tibial cortex and lock into position by tightening the knurled knob on the outrigger. Preliminary rotational alignment is based on the medial third of the tibial tubercle. Secure the alignment device to the reamer shaft with the lateral Setscrew (Figure 9).
Pin the Tibial Cutting Block so a minimal resection is made from the proximal tibia. Utilize the Stylus when necessary (Figure 9).
Note: There is a slotted and non-slotted end to the Stylus. The difference between the two is 5 mm.
Note: If a Metaphyseal Sleeve is to be used the tibial resection will be performed using the Tibial Sleeve Broach (see page 16, Figure 11).
Figure 9
Pins
14 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
PROXIMAL TIBIAL RESECTION – TAPERED REAMER
Remove the I.M. device while leaving the 2 degree Cutting Block in place. Remove the Tapered Reamer and resect the proximal tibia (Figure 10).
Note: At this point determine whether a step wedge is necessary on either the medial or lateral side to augment a defect, or both sides in order to restore the joint line. If a wedge is necessary on one side, it is recommended that the step wedge be prepared after rotational position of both the femoral and tibial components have been determined. For step wedge preparation see Appendix 2 (page 51).
Figure 10
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 15
PREPARATION OF THE METAPHYSEAL BONE – BROACH
Optional For Sleeve Utilization Only
Note: The MBT Revision Tibial Tray will accept either a tibial Metaphyseal Sleeve or a tibial Step Wedge. Only the 29 mm Sleeve is indicated for use with a Tibial Step Wedge.
Attach the MBT Revision Broach Handle to the smallest broach and then attach the appropriately sized Stem Trial. The broaches are asymmetrical, position the “ANT” engraving on the broach anteriorly. Impact the broach into the tibia until the top surface of the broach is at the desired proximal tibial resection level. When broaching the proximal metaphysis, take care to assure the appropriate rotation of the Broach.
Note: The corresponding tibial sleeve implant allows up to +/- 20 degrees of rotation from the centerline of the MBT Revision Tray.
Check for rotational stability of the broach. If the broach (not the handle) moves in the canal, it is not rotationally stable.
If the broach is unstable or the defect is unfilled, repeat with consecutively larger broaches until the desired fit is achieved (Figure 11). Remove the Broach Handle, leaving the last broach in place. Any defects remaining can be filled with allograft or autologous bone placed in intimate contact with the sleeve.
Two common tibial broaching techniques:
1) Chase the defect by rotating the broach to fill the defect until reaching rotational stability of the broach. If utilizing this technique the surgeon must be aware that the sleeves are allowed to rotate +/-20 degrees with respect to the MBT Revision Tibial Tray.
2) Align the broach with the medial third of the tibial tubercle and progressively broach until rotational stability of the broach is attained.
Figure 11
Tibial Resection
Plane
16 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
PREPARATION OF THE METAPHYSEAL BONE – BROACH
Resect the proximal tibia utilizing the top of the broach as a guide (Figure 12). The top of the broach has a 2 degree slope built in. The proximal cut should be parallel to the top of the broach.
Note: If a cutting guide is desired for resecting the proximal tibia with the tibial broach in place, assemble the SP2 0 degree Tibial Cutting Block (Cat. No. 96-6320) to the SP2 IM Tibial Guide and slide over the Broach Adapter Outrigger (Cat. No. 2178-01-108). Slide this assembly onto the boss of the seated tibial broach, pin the block, remove the outrigger, and resect through the slot of the cutting block (Figure 13).
Slide the tibial view plate which best covers the proximal tibial over the broach post. Note the view plate size as it will dictate the size of the MBT Revision Tibial Base Trial that will be used. The tibial view plate is transparent to help visualize tibial coverage (Figure 14). The template matches the implant to aid in orienting the tibial sleeve to the tibial base during assembly.
Figure 14
Figure 13
Figure 12
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 17
TIBIAL TRIAL ASSEMBLY
Assemble the tibial tray trial with the stem extension and sleeve trial, if applicable (Figure 15). Position the Tibial Trial construct into the prepared tibial canal (Figure 16). Assess proximal tibial coverage and rotation of tibial component. The base plate should be positioned to provide the best coverage of the tibial condylar surface.
Note: The MBT Revision Tibial Keel Punch with the Universal Handle may be utilized to assist with seating of tibial trial construct. Once the tibial trial construct is seated the Keel Punch must be removed in order to accommodate the use of Spacer Blocks.
Leave the trial in place and proceed to femoral preparation, final tibial preparation will occur after femoral preparation is complete.
Note: A 14 mm or smaller size stem implant can be pulled through the sleeve implant. If the stem is 16 mm or greater it will not pull through the sleeve.
Figure 15
Figure 16
18 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
PREPARATION OF FEMORAL DIAPHYSIS
Figure 17
Intramedullary Femoral Alignment SystemThis technique is designed to flow in a logical sequence, from reaming the diaphysis, to broaching the metaphysis and cutting the bone. The length and diameter of the stem extension is determined with templates applied to pre-operative roentgenograms.
Begin the procedure with the preparation of the medullary canal (Figures 17 and 18).
Enter the medullary canal with a 9 mm drill to a depth of 3-5 cm (Figure 19). Take care that the drill avoids the cortices. It is helpful to palpate the distal femoral shaft as the drill is advanced.
Where impedance of the intramedullary canal is anticipated, adjust the entry point accordingly.
Figure 19
Figure 18
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 19
REAMING THE MEDULLARY CANAL
Figure 20
Connect the Reamer Handle to a small diameter MBT Revision Reamer. If power reaming, it will be necessary to attach the modified Hudson Adapter to the Straight Reamer.
Note: The reamer shaft contains markings in 25.4 mm increments to accommodate the various Universal Stem/Sleeve length combinations (Figure 20).
Use the Reamer Depth Chart to determine reamer depth for each combination of components (Figure 21). Another option to determine reamer depth is to measure the trial assembly against the reamer and note the corresponding depth mark for reaming.
You may also determine the length and diameter of the prosthetic stem extension with templates (Cat. No. 2294-99-035: SIGMA® Femoral Adapter Sleeve and Stem Template) applied to pre-operative X-ray. The S-ROM Femoral Components can be found on the S-ROM Templates (XRT-115).
The S-ROM Femoral Component accepts the following stems, only with the use of a femoral sleeve: · Universal Fluted Stems of 75, 115 and 150 mm in
diameters of 10-24 mm in 2 mm increments
· Cemented Stems available in lengths of 30 and 60 mm lengths and a diameter 13 mm or 15 mm
· Cemented Tapered Stems available in lengths of 90 mm (13 mm and 15 mm diameter) and 120 mm and 150 mm (13 mm diameter only)
Note: The Stem is the same as is currently used with the MBT Revision Trays.
Example: Ream to the tick mark #4 when using S-ROM Femur with 31 mm Sleeve and 75 mm Press-Fit Stem
21 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
S-ROM Femur 20 mm31 mm34 mm
40 mm46 mm
Cemented Stems
30 mm 2 2
60 mm 3 3
90 mm 4 4
120 mm 5 6
150 mm 6 7
Press-Fit Stems
75 mm 4 4
115 mm 5 5
150 mm 6 7
REAMING THE MEDULLARY CANAL
Figure 21
In 1 mm diameter increments, sequentially open the medullary canal with MBT Revision Reamers of progressively greater size until firm endosteal engagement is established.
Take care to ream the canal in line with the femoral axis to avoid putting the implant in flexion.
Note: Do not reverse ream.
It is important that simple cortical contact of the tip not be construed as engagement.
Cemented Stem UseWhere a cemented stem extension is indicated, perform final reaming with a 15 mm diameter reamer for the 13 mm diameter stem extension; similarly a 17 mm diameter reamer is used to accommodate the 15 mm diameter stem extension.
This allows for creation of a cement mantle.
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 21
Figure 22 Figure 23
Broach
Reamer
After reaming the intramedullary canal, attach the Threaded Shaft to the Broach Reamer and then to the appropriate Stem Trial as determined by straight reaming (Figure 22).
Ream to the 20 mm, 31 mm, 34 mm etch mark on the Threaded Shaft (Figure 23).
PREPARATION OF THE METAPHYSIS – SLEEVE USE
22 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Medial SideLateral Side
Figure 25
Figure 24
PREPARATION OF THE METAPHYSIS – SLEEVE USE
When using the Broach Reamer, the next smaller diameter Stem Trial may be used to allow for easier reaming. The Broach Reamer will be necessary when utilizing a 20 mm sleeve and for the beginning of larger sequential broaching when using a 31 mm or larger sleeve. After broach reaming has been completed, attach the 31 mm broach to the Broach Handle (Figure 24). Attach the appropriate Stem Trial to the broach as determined by straight reaming. Give close attention to the medial orientation of the broach.
Note: The broach is asymmetrical; and the narrow side of the broach must point medially (Figure 25).
Note: When prepping for a 20 mm sleeve, leave the Broach Reamer and threaded shaft in the canal and perform the subsequent femoral cuts off the reamer.
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 23
PREPARATION OF THE METAPHYSIS – SLEEVE USE
Sequentially broach to the desired dimension of 31, 34, 40 or 46 mm (Figure 26). When the LCS/S-ROM etch mark on the Broach Handle is at the planned distal resection level, check the broach’s rotational stability. If the broach (not the handle) moves in the canal, it is not rotationally stable.
If the stability of the broach is unsatisfactory, move up to the next broach size. The last broach used will be the femoral sleeve size. The broach depth sets the extension gap/joint line.
In patients with a large degree of distal femoral bow, closely monitor the anterior progression of the broach during impaction. Excessive anterior placement of the broach may result in a loose flexion gap.
Figure 26
Planned Level of Distal Resection
24 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Figure 27
PREPARATION OF THE METAPHYSIS – SLEEVE USE
After broaching is complete, remove the Broach Handle from the broach. With the broach seated in the femur, attach the Threaded Shaft to the broach (Figure 27).
Distal, anterior, chamfer, and notch cuts will reference off the Threaded Shaft/Broach assembly.
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 25
FEMORAL PREPARATION – DISTAL RESECTION
Distal Femoral Resection Guide
Distal Femoral Connector
Revision Distal Cutting Block
Figure 28
Figure 30
Figure 29
Figure 31
Distal ResectionSet the valgus angle to 7 degrees and Left/Right on the Distal Femoral Alignment Guide by compressing the two triggers and lock in place by rotating the blue locking lever clockwise. Place the Femoral Alignment Guide on the Threaded Shaft and seat against the distal femur (Figure 28).
Rotate the knob on the Femoral Resection Guide counterclockwise until the arrow is pointing to the padlock symbol. Slide the Distal Femoral Connector into the Femoral Resection Guide. Rotate the knob on the Femoral Resection Guide clockwise. Every click moves the Revision Distal Cutting Block 1 mm proximal or distal. Turn the knob clockwise from 15 all the way down to 0 (which is the padlock symbol). This will set the block up for a 0 mm resection (Figure 29).
Slide the revision Distal Cutting Block onto the Distal Femoral Block attachment. The tang on the block connector will slide into the 0 mm cutting slot on the cutting block. The trigger should engage in the hole behind the 0 mm slot (Figure 30).
Note: An open resection will resect 4 mm less femur. When a 0 mm open resection is desired, the dial should be set to 4 mm.
Position the resection guide over the two legs of the Distal Femoral Alignment Guide until the Distal Cutting Block touches the anterior femur (Figure 31).
Note: The Revision Distal Block is equipped with 0, 4, and 8 mm saw slots. Please keep in mind that if the resection level is not at 0 (the padlock symbol) this will alter the resection. If the resection knob is set at 2, for instance, the saw slots will perform 2, 6, and 10 mm resections.
26 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
FEMORAL PREPARATION – DISTAL RESECTION
Rest the Femoral Alignment Guide against the most prominent distal condyle. If no distal augments are needed, proceed with pinning the distal resection block and making the distal cuts through the 0 mm resection slot (Figures 32 and 33).
If it is determined that a 5 mm or a 10 mm Distal Augment will be needed on only one condyle, perform a 0 mm clean-up cut first on the prominent condyle. Then turn the dial on the Femoral Resection Guide to 5 mm or 10 mm and make the cut on the other condyle through the 0 mm resection slot.
Note: Do not use the saw slots on the Distal Block to make augment cuts. The augment slots on this block are set up in 4 mm increments instead of the needed 5 mm increments for the S-ROM Knee System.
Note: When adding 5 or 10 mm Distal Augments to both sides of the femur, it may be necessary to re-evaluate the depth the sleeve was broached to, based upon the addition of augments. If the augments require the broach to be distalized, rebroaching should occur with a larger broach in order to distalize the sleeve in the canal without losing press-fit.
Figure 32
Figure 33
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 27
FEMORAL PREPARATION – DISTAL RESECTION
Figure 35
Figure 34
2. Remove femoral alignment guide towards the T-Handle
1. Slide femoral resection guide upwards
Release attachment
Once the pins are in place, unlock the Distal Cutting Block from the Distal Femoral Connector, using your thumb and index finger to release the attachment. Slide the Femoral Resection Guide upwards on the Alignment Guide legs until the block connector disengages from the Cutting Block and in one motion remove the Femoral Alignment Guide by pulling the instruments distally over the Threaded Shaft (Figure 34).
In many cases, little, if any, bone is removed from the distal femur as the joint line is effectively elevated with the removal of the primary femoral component. As the level of resection is based on the preservation of bone stock, each condyle is cut only to the level required to establish a viable surface, with augmentation employed to correct imbalance.
The resection is then performed through the slot appropriate for each condyle, using a standard 1.19 mm thick blade (Figure 35).
Note: If a ½ in. wide Standard Saw Blade is used it can complete both medial and lateral distal femoral cuts with the entire jig still in place.
An example of a medialresection at 0 mm
28 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
The Femoral Cutting Guide is size specific (two blocks – X-Small/Small and Medium). Determine the femoral component size by pre-operative templating and comparing the Femoral Component Trial to the size of the femur. Use the size which gives the best medial/lateral (M/L) coverage.
Slide the appropriately sized cutting guide over the Threaded Shaft. Use the corresponding hole for a left or a right knee (Figure 36).
Place the guide into neutral rotation by aligning the anterior cortex parallel with the anterior portion of the guide. The SIGMA Angel Wing (part number: 96-6530) may be helpful in this step. Also use the femoral epicondylar axis as the rotational reference.
Note: If distal augmentation will be used, use 5 or 10 mm Box Cut Guide Spacers on the appropriate condyle(s). Establish the proper rotation of the A/P block first, then pin through one each of the medial and lateral pin holes. Remove the block from the pins, then put the appropriate spacer(s) over the pins before replacing the A/P block. These spacers should rest between the cutting guide and the distal condyle(s) to fill these gaps appropriately (part numbers: 5 mm - 63-3305A and 10 mm - 63-3306A).
Achieve fixation of the cutting guide with 1/8 in. drill pins, introduced through the convergent holes on the side of the block. These Pins will need to be temporarily removed later to move to the notch guide (Figure 37).
Attach the Removable Handles to the cutting guide (optional).
FEMORAL A/P AND CHAMFER CUTS
Figure 36
Figure 37
PinPin
Box Cut Guide Spacer (5 mm)
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 29
FEMORAL A/P AND CHAMFER CUTS
Anterior Chamfer Cut
Remove Pins
Figure 38
Figure 39
Figure 40
Make the anterior cut (Figure 38) first. Proceed to make the anterior chamfer cut through the captured slot. If the block was previously pinned, temporarily remove one pin at a time while making the resection.
Make the posterior chamfer cut (Figure 39) by holding the saw blade flush with the cutting guide. If anterior pins are being used for fixation, remove the pin while resecting, then replace. Care should be taken to avoid damaging posterior soft tissue.
If not previously pinned, place at least one 1/8 in drill pin on each side of the guide. These will be used to position the Box Cut Guide. Next, remove the convergent pins.
Finally, remove the femoral cutting guide, leaving the 1/8 in. drill pins in place (Figure 40).
Note: It may be easier to remove the threaded shaft first, before trying to slide the blocks off the pins.
31 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
FEMORAL BOX CUTS
Use 5 or 10 mm Box Cut Guide Spacers if distal Augmentation Blocks will be used (Figure 41).
Slide the Hinge Femoral Box Cut Guide over the 1⁄8 in. drill pins placed in the previous step or align with the lines marked off the Femoral Cutting Guide. If Distal Augmentation Blocks will be used, slide 5 or 10 mm Box Cut Guide Spacers over the drill pins before positioning the Box Cut Guide.
Four additional 1⁄8 in. drill holes are provided on the anterior surface of the Box Cut Guide; 1⁄8 in. drill pins are recommended for additional stability.
Holding the saw blade flat against the inner surface of the Box Cut Guide, make the side cuts for the center box (Figure 42).
Use a narrow saw blade (12.7 mm or 0.5 in), placed on the sloped guide surface, to remove the bone block of the center box (Figure 43).
Figure 43
Figure 41
Box Cut Guide
Hinge Box Cut Guide
Narrow Saw Blade (1⁄2 in. wide)
Saw Blade
Box Cut Guide Spacer
1/8 in. Drill Pin
Figure 42
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 31
FINAL PREPARATION OF THE TIBIA
Figure 44
Assess proximal tibial coverage and rotation of tibial component. Impact the appropriate Keel Punch (utilize the cemented Keel Punch if a cement mantle is desired or the Press-Fit Keel Punch if line-to-line contact is desired) (Figure 44). The base plate should be positioned to provide the best coverage of the tibial condylar surface.
32 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
FEMORAL TRIAL INSERTION
Two femoral Augmentation Blocks are available for the S-ROM NOILES Rotating Hinge Total Knee System. They are 5 and 10 mm Distal Blocks. One size fits all, i.e. X-Small, Small and Medium hinge femoral components. If distal augmentation is required, attach the Augmentation Block Trial(s) with bone wax to the Femoral Component Trial (Figure 45).
Figure 45
Implant Cat. No.
Femoral Location
Use with S-ROM Noiles Rotating Hinge Femoral Size
Augment Thickness
Trial Cat. No.
623805 Distal All Sizes 5 mm 633785
623810 Distal All Sizes 10 mm 633790
Femoral Augment Block
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 33
Stem Trial
Trial Assembly inserted into Femoral Cavity
Femoral Component Trial
Femoral Sleeve Trial
Figure 46
Figure 47
Connect the Stem Trial into the appropriate Femoral Sleeve Trial. The diameter of the Stem Trial will be the same as the final Straight Reamer used; the size of the Femoral Sleeve Trial will be the same as the final Femoral Broach used (Figure 46).
Slide the Sleeve/Stem Trial assembly into the prepared cavity in the femoral canal to allow the assembly to self align with the broached surfaces (Figure 47).
Note: The narrow side of the Sleeve Trial points medially.
FEMORAL TRIAL INSERTION
34 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Universal Handle Femoral Driver
Femoral Component Trial
FEMORAL TRIAL INSERTION
Slide the Femoral Component Trial onto the resected femur, aligning the anterior cut with the posterior aspect of the patellar flange. After the Femoral Component Trial engages the Femoral Sleeve Trial, impact using the Femoral Driver on the Universal Handle. Check accuracy of the bone cuts. Revise or rebroach if necessary (Figure 48).
Note: If Distal Augmentation Blocks will be used, fix Distal Augment Block Trials to the Femoral Trial with bone wax before impacting the Trial onto the femur.
Figure 48
Push down on the button on the quick release handle and connect the femoral driver
Two ends engage
Femoral Sleeve Trial
Femoral Component Trial
Femoral Sleeve Trial
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 35
TRIAL REDUCTION
Without hinge pin,“lift off” during flexion indicates posterior impingement
Figure 49
Figure 51
Figure 50
Slide the condyles of the Femoral Trial into the Plateau Trial. If the insert trial lifts off the tibial baseplate during flexion, check the posterior area for soft tissue, osteophyte or bone impingement (Figure 49).
It is easier to insert the Hinge Pin Trial prior to placing the insert trial into the tibial baseplate. The Hinge Pin Trial can be inserted either medially or laterally (Figure 49).
With the leg in full extension, evaluate the mechanical axis. The center of the femoral head, knee and talus should all be in line (Figure 50).
The knee should be stable throughout the full range of motion (Figure 51).
Check ligament tension and leg length.
Revision of the tibial or femoral resection may be required if satisfactory stability cannot be achieved. Accommodate additional bone resection with rebroaching.
Remove the femoral trials and ensure that the rotational alignment of the assembly is preserved.This is used as a reference when assembling the modular implant.
Note: In patients with severe soft tissue loss, flexion of the knee beyond 90 degrees may cause distraction and subluxation of the tibial plateau out of the modular tibial base. In this instance, fit the patient with a post-operative brace, limiting flexion to 90 degrees and no more for at least three months. This helps soft tissue establishment of flexion tension. Consult package insert.
Hinge Pin Trial
36 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Figure 54
IMPLANT ASSEMBLY – TIBIA
Figure 52 Figure 53
Tibial Sleeve AssemblyNote: It is imperative to assemble the Sleeve prior to stem attachment.
Note: Sleeves and Step Wedges can only be used together if using a 29 mm Sleeve.
Remove trial component in one piece (use as guide for assembly of implants).
Place the MBT Revision Tray on a firm, stable, padded surface. Set the Tibial Sleeve in an orientation that matches the prepared canal. Matching the orientation of the Tray/Sleeve Trial is helpful in determining appropriate rotation of the final tibial tray/sleeve implant (Figure 52). The sleeve can rotate 20 degrees internally or externally.
Using the Sleeve Impactor and a Mallet, impact the sleeve onto the MBT Revision Tray. Deliver several strikes to engage the two components (Figure 53).
Stem Component AssemblyAttach the Stem Extension to the prosthetic tray using the two appropriate Wrenches to ensure full engagement (Figure 54).
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 37
Implanting the Tibial Component Thoroughly cleanse the site with pulsatile lavage. Perforate with small drill holes on the prepared tibial surface to facilitate penetration of methyl methacrylate (Figure 55). Pack residual small cavitory bone defects with cancellous autograft, if available, or allograft.
Apply methyl methacrylate cement to the proximal tibial surface (Figure 56) or directly to the underside of the tibial tray component.
When a Fluted Stem or a Fluted Stem with a Metaphyseal Sleeve is used, ensure the medullary canal remains free of cement. Clear all extruded cement with a Curette.
Seat the Tibial Implant construct into the prepared tibia by impacting the RP Tray Impactor and Universal Handle assembly (Figure 57).
Figure 57
Figure 55 Figure 56
TIBIAL IMPLANTATION
38 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Figure 58
Apply cement to this side of the distal augmentation block
Implant Assembly - Sleeve and Stem UseImplant assembly order (with Sleeve and Stem use):
· Add distal augments if necessary
· Attach stem to sleeve
· Attach sleeve construct to femoral construct
ImplantationAfter assembling the femoral components, prepare one package of bone cement according to instructions.
Apply cement to the augmentation block(s) on the side which contacts the femoral component, and to the corresponding surface(s) of the femoral component (Figure 58).
Attach the augmentation block(s) to the femoral component. Use an Augment Block Clamp to secure to the femoral component until the cement is fully cured.
Note: When distal augmentation blocks are used with the S-ROM Noiles Rotating Hinge femoral component, place the Augment Block Clamp into the distal condylar “pocket” of the femoral component (Figure 59).
The remainder of the mixed cement may be used to implant the patella and tibial component while the blocks are setting.
IMPLANT ASSEMBLY – SLEEVE AND STEM USE
Figure 59
S-ROM Noiles Rotating Hinge Femoral Component
Distal Augment
Apply cement only to this side of the distal augmentation block
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 39
To attach the Universal Stem to the Universal Femoral Sleeve, thread the stem onto the sleeve. Grasp the sleeve with the Tibial Sleeve Clamp and use the Stem Extension Wrench to grasp Universal Stem and tighten (Figure 60).
Apply sufficient force to both Wrenches to ensure that the Stem is secure.
Place the femoral component with the Femoral Adapter on a firm, stable surface. Place the appropriate sleeve and stem construct on top of the Femoral Adapter assembly (Figure 61). Use the sleeve and femoral construct trial to help set the final sleeve and femur implant rotation.
Figure 60
IMPLANT ASSEMBLY – SLEEVE AND STEM USE
Figure 61
41 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
IMPLANT ASSEMBLY – SLEEVE AND STEM USE
Slide the Femoral Stem/Sleeve Impactor on top of the stem and forcefully apply three strikes with a Mallet to engage the two component assemblies (Figure 62).
The definitive components are implanted in the following order:
· Tibial tray (with stem, sleeve or wedges)
· Femoral component (with stem, sleeve and augments)
· LPS Hinged Insert
Implant the femoral component using the Femoral Impactor (Figure 63).
Universal Handle
Figure 62
Figure 63
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 41
Figure 67
After the femoral component and tibial tray have been cemented into place, do one final check with the trial inserts. Once the proper thickness has been verified, introduce the actual implant into the sterile field.
Note: Take extreme care when opening the LPS Universal Insert to hold onto the bushings to ensure they do not fall out.
Put the condyles of the femoral component into the corresponding recesses in the tibial plateau.
Insert the Hinge Pin through the hole on the medial or lateral side of the femoral component. Orient the rectangular head of the Hinge Pin with the rectangular recess in the femoral component (Figure 64).
Squeeze the “clothespin” of the Hinge Pin together and insert the Hinge Pin into the femoral component. Make sure the Hinge Pin is securely locked in place (Figure 65).
Place the LPS Universal Insert post into the cone of the MBT Revision Implant (Figure 66).
Test the knee through full range of motion (Figure 67).
Figure 65
Hinge Pin
Needle Holder or Clamp
"Clothespin"
Figure 64
Figure 66
BEARING AND HINGE PIN INSERTION
42 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
INITIAL PATELLAR RESECTION
Measure and record the overall thickness of the patella using a Caliper (Figure 68).
Resect approximately 7 mm of bone from the posterior patella surface using an Oscillating Saw.
Measure and record the thickness of the resected/removed bone in order to properly duplicate the original thickness.
Figure 68
Caliper
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 43
INITIAL PATELLAR RESECTION
Figure 69
Figure 70
Patella Domes are available in four diameters (Figure 70). Select a patella trial with the diameter that best matches the patient’s patella (Figure 69).
Select the Patella Reamer Depth Adjuster that is the same diameter as the patella trial.
Insert the Patella Reamer Depth Adjuster into the Patella Restraining Instrument. Rotate the depth adjuster 120 degrees clockwise to lock into position (Figure 70).
38 mm
35 mm
32 mm
30 mm
Patella Reamer Depth Adjuster
Patella Restraining Instrument
44 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
PATELLA REAMING
Clamp the Patella Restraining Instrument assembly onto the patella. Lock into position by turning the Thumb Nut clockwise (Figure 71).
Insert the Patella Reamer Bushing into the appropriate set of slots on the Patella Reamer Depth Adjuster. Slots on the depth adjuster are marked 1, 2, 3 and 4, which indicate the reaming depth in millimeters. To determine the correct slot, use the formula shown in (Table 1) as a guide.
Select the Patella Reamer that matches the diameter of the patella component to be used and insert through the Patella Reamer Bushing into the Patella Reamer Depth Adjuster. Ensure that the Patella Reamer is making full contact with the bone prior to reaming. Ream until the Patella Reamer flange makes contact with the Patella Reamer Bushing.
Table 1
Figure 71
Turn Thumb Nut clockwise to lock into position
Patella Reamer
Flange
Patella Reamer Bushing
Patella Reamer Depth Adjuster
Patella Restraining Instrument
Thickness ofselected patella
component
Measuredthickness of
bone resectedand removed
Select Slotnumber
9 mm - =7 mm 2
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 45
PATELLA DRILLING
Figure 72
Figure 73
Remove the Patella Reamer and insert the Patella Drill Guide into the Patella Reamer Bushing. The Locating “Pin” on the Drill Guide will insert into the hole in the Patella Restraining Instrument (Figure 72).
Select the 3⁄16 in. Patella Shoulder Drill and prepare the three patella peg holes by drilling through the three larger holes in the Patella Drill Guide.
The depth of the holes drilled is correct for the length of the pegs on the selected Patella Button (Figure 73). Optional: Select the 1⁄8 in. Patella Shoulder Drill and drill through the four smaller holes to enhance the cement fixation to the patellar bone. Loosen the Thumb Nut on the Patella Restraining Instrument and remove the entire assembly from the patella bone.
Patella Shoulder Drill
Patella Drill Guide
Thumb Nut
Patella Reamer Bushing
Locating Pin
Patella Restraining Instrument
Patella Reamer Depth Adjuster
3⁄16 in. Drill Hole
Locating Pin
1⁄8 in. Drill Hole
Patella Drill Guide
46 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
TRIAL REDUCTION AND IMPLANTATION
Place the appropriate diameter Patella Trial into the prepared patella bone. Measure the overall thickness of the patella construct to ensure that it is the desired thickness, i.e. equal to or 1-2 mm less than the original patella thickness. A “no-thumbs” trial reduction and patella tracking evaluation can now be performed.
Note: If the reconstructed patella is too thick, repeat the reaming and drilling steps using the number 2, 3 or 4 slot on the Patella Reamer Depth Adjuster. If a greater thickness must be removed, take additional resection from the patella. The reaming and drilling steps must be repeated. (Take care to make sure the patella bone is not cut too thin. Maintain at least 10 mm of patella bone to prevent drill or peg penetration of the anterior cortex).
The appropriately sized Patella Dome may now be cemented into place. A Patella Cement Clamp is provided for this purpose (Figure 74).
Figure 74
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 47
APPENDIX 1: THE CEMENTED TIBIAL STEM EXTENSIONS
Figure 2 Figure 3
Cemented Drill Bushing
MBT Revision Drill Bushing
Figure 1
Cemented Stem ReamerAlign the Tibial Tray and secure with two Fixation Pins inserted through the holes designated (Figure 1). Seat the MBT Revision Drill bushing onto the tibia trial. Place in the posterior holes.
Place the Cemented Drill Bushing into the MBT Revision Drill Bushing (Figure 2).
Use the “cemented” reamer to ream to the predetermined selected depths for tray only or the Tray with a 30 or 60 mm cemented stem.
Remove the reamer and “cemented” bushing, leaving the tray trial and MBT Revision Drill Bushing in place (Figure 3).
Note: Only a 13 mm diameter cemented stem should be used in conjunction with the MBT Revision Tray to avoid a step off at the stem/tray junction.
48 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
APPENDIX 1: THE CEMENTED TIBIAL STEM EXTENSIONS
Tapered ReamerAssemble the Revision Reamer Adapter onto the Cemented Tapered Reamer.
Next, attach the modified Hudson Adapter to the Tapered Reamer, if power reaming.
Attach the appropriately sized Cemented Stem Trial
(13 x 30 mm or 13 x 60 mm) to the Tapered Reamer, if utilizing a cemented stem extension (Figure 4). Ream until the Revision Reamer Adapter is flush with the MBT Revision Drill Bushing (Figure 5).
Note: To avoid Stem Trial disengagement, do not reverse ream.
Figure 4
Figure 5
Modified Hudson Adapter
Revision Reamer Adapter
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 49
APPENDIX 1: THE CEMENTED TIBIAL STEM EXTENSIONS
Tapered Cemented StemsNote: Tapered Cemented Stem sizes 13 x 90/120/150 mm are compatible with MBT Revision Trays.
Ream the canal with a reamer two sizes larger than the stem. Ream the medullary canal with a 15 mm reamer to implant a 13 mm Tapered Cemented Stem, which allows for a 1 mm circumferential cement mantle at the proximal end of the stem. The cement mantle will be greater around the distal end of the Cemented Tapered Stem (3 mm per side).
This provides the following benefits:
· Thicker cement mantle distally helps assure that a circumferential mantle is present and reduces the possibility of thin or non-existent cement coverage of the stem distally
· Stresses are greatest at the tip of the stem. A larger cement mantle is advantageous in dissipating these stresses. Thinner cement mantles are more prone to breakdown when exposed to higher stresses
Tibial Keel PreparationPlace the knee in full extension and determine appropriate rotation of the Tibial Tray. Mark the appropriate rotation with electrocautery on the anterior tibial cortex at the center and sides of the Alignment Handle.
Assemble the appropriate Stem Trial to the MBT Revision Tray Trial and seat in the prepared bone bed. Impact the Cemented Keel Punch (Figure 6).
Disconnect the Universal Handle leaving the Keel Punch in place for trial reduction (if appropriate).
It is recommended that a Cement Restrictor be placed at the appropriate level prior to cementing the component. Use a Cement Gun to fill the canal with methyl methacrylate.
Figure 6
51 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Figure 1
Figure 2
APPENDIX 2: STEP WEDGE PREPARATION
Step Wedge Augmentation
Resection for supplementary tibial augmentation may be based on the established position of the Trial Tray. Remove the Femoral Trial to provide greater access. Confirm rotational alignment of the Tibial Tray Trial. Secure the Tray with two Fixation Pins.
Attach the Tray Trial Wedge cutting attachment with the Step Wedge Cutting Guide to the Trial Tray. The Step Wedge Cutting Block allows for a 5, 10 or 15 mm Step Wedge preparation as necessary. Slide the block forward to the anterior proximal tibia and secure in place with two Steinmann Pins through the holes marked with (Figure 1).
Unlock the block and slide the assembly out of the block. Disconnect the handle from the Trial Tray (Figure 2).
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 51
APPENDIX 2: STEP WEDGE PREPARATION
Trim the tibia accordingly with an Oscillating Saw so the cut does not extend beyond the central riser (Figure 3). Remove the Block and Pins.
Assemble the Trial Wedge to the appropriate Tibial Tray Trial (Figure 4) and introduce into the prepared site. Perform minimal correction with a Bone File where indicated to ensure maximal contact.
Figure 4
Step Wedge Cutting Block
Figure 3
52 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Confirm positioning, alignment and security of the tray assembly. If there is old cement or sclerotic bone, remove this first with a saw blade or burr prior to punching. Position the MBT Revision Tibial Keel Punch at the tray and cancellous bone interface and impact into the keel configuration. Leave the punch in place and perform a final trial reduction if necessary (Figure 5).
Note: Utilize the “cemented” Keel Punch when a cement mantle is desired.
Alternative Step Wedge Preparation This is a “free-hand” resection. Assemble the Wedge Trial and Stem Trial to the Tibial Tray Trial. Position the device slightly proximal to the planned resection level. Make a conservative “free-hand” wedge resection and then check cuts with the Trials (Figure 6).
Wedge Implant Assembly
Note: To aid wedge implant assembly, attach wedge prior to attaching the stem attachment.
Assemble the designated wedge to the tray and secure using the appropriate screw. Carefully tighten with the large T-Handle Torque Driver until an audible "click" is discerned, ensuring a full and permanent interlock (Figure 7).
Figure 7
APPENDIX 2: STEP WEDGE PREPARATION
Figure 5 Figure 6
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 53
APPENDIX 3: THICK TRAY PREPARATION
After impacting the cement or Press-Fit Keel Punch, remove the Keel Punch. Insert the MBT Thick Tray Trial Adapter (15 or 25 mm) onto the Tibial Tray Trial (Figures 1 and 2).
Note: The Tibial Tray Trial must be used with the Thick Tray Adapters as the two pieces equal the appropriate sizing – 15 or 25 mm.
Perform the final trial reductions utilizing the same technique as the standard MBT Revision Tray. Implant assembly and implantation is also the same as with the standard MBT Revision Tray. If utilizing a Wedge, refer to the Step Wedge Preparation in Appendix 2.
Note: A Tibial Wedge can be used with all Thick Tray sizes, except for size 2. Due to the taper, use size 2 Tibial Wedges with size 4 MBT Revision Thick Trays, and use size 1 Tibial Wedges with size 3 MBT Revision Thick Trays. Sleeves may be used with all Thick Trays.
Note: Due to the taper, trial with appropriate tray trial size. For example, a size 4 Thick Tray tapers down to a size 2. Use the size 2 Tray Trial with the size 4 Thick Tray Adapter. The size 3 Thick Tray tapers down to a size 1. And the size 2 Thick Tray tapers down to a size 0. The size 0 Tray Trial can be found in the MBT Thick Tray Instrument Set.
Figure 1
Figure 2
54 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
IMPLANT LISTING
S-ROM Femoral Components (hinge pin is packaged with the femur)
Universal Femoral Sleeves
Universal Press-Fit Stems
S-ROM Femoral Augments
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 55
62-3401L Medium Left S-ROM Femur
62-3401R Medium Right S-ROM Femur
62-3411L Small Left S-ROM Femur
62-3411R Small Right S-ROM Femur
62-3421L X-Small Left S-ROM Femur
62-3421R X-Small Right S-ROM Femur
1294-53-205 20 mm Cemented Femoral Sleeve
1294-53-215 31 mm Distally Porous Femoral Sleeve
1294-53-216 31 mm Fully Porous Femoral Sleeve
1294-53-225 34 mm Distally Porous Femoral Sleeve
1294-53-226 34 mm Fully Porous Femoral Sleeve
1294-53-235 40 mm Distally Porous Femoral Sleeve
1294-53-236 40 mm Fully Porous Femoral Sleeve
1294-53-245 46 mm Distally Porous Femoral Sleeve
1294-53-246 46 mm Fully Porous Femoral Sleeve
62-3805 Distal Augment, 5 mm (used on all femur sizes)
62-3810 Distal Augment, 10 mm (used on all femur sizes)
86-7410 75 mm x 10 mm Universal Fluted Stem
86-7412 75 mm x 12 mm Universal Fluted Stem
86-7414 75 mm x 14 mm Universal Fluted Stem
86-7416 75 mm x 16 mm Universal Fluted Stem
86-7418 75 mm x 18 mm Universal Fluted Stem
86-7419 75 mm x 20 mm Universal Fluted Stem
86-7420 75 mm x 22 mm Universal Fluted Stem
86-7421 75 mm x 24 mm Universal Fluted Stem
86-7424 115 mm x 10 mm Universal Fluted Stem
86-7426 115 mm x 12 mm Universal Fluted Stem
86-7428 115 mm x 14 mm Universal Fluted Stem
86-7430 115 mm x 16 mm Universal Fluted Stem
86-7432 115 mm x 18 mm Universal Fluted Stem
86-7433 115 mm x 20 mm Universal Fluted Stem
86-7434 115 mm x 22 mm Universal Fluted Stem
86-7435 115 mm x 24 mm Universal Fluted Stem
86-7438 150 mm x 10 mm Universal Fluted Stem
86-7440 150 mm x 12 mm Universal Fluted Stem
86-7442 150 mm x 14 mm Universal Fluted Stem
86-7444 150 mm x 16 mm Universal Fluted Stem
86-7446 150 mm x 18 mm Universal Fluted Stem
86-7447 150 mm x 20 mm Universal Fluted Stem
86-7448 150 mm x 22 mm Universal Fluted Stem
86-7449 150 mm x 24 mm Universal Fluted Stem
IMPLANT LISTING
56 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Universal Femoral Sleeves
S-ROM Patella
LPS Universal Inserts (compatible with S-ROM and LPS femurs, must match femur size-to-size)
86-6401 30 mm x 13 mm Cemented Stem
(used on MBT Revision Trays size 3 and smaller)
86-6402 60 mm x 13 mm Cemented Stem (used on MBT Revision Trays size 3 and smaller)
86-6403 30 mm x 15 mm Cemented Stem (used on MBT Revision Trays size 4 and larger)
86-6404 60 mm x 15 mm Cemented Stem
(used on MBT Revision Trays size 4 and larger)
86-6468 90 mm x 13 mm Cemented Tapered Stem
86-6469 90 mm x 15 mm Cemented Tapered Stem
86-6498 120 mm x 13 mm Cemented Tapered Stem
86-6499 150 mm x 13 mm Cemented Tapered Stem
62-1630 30 mm S-ROM Dome Patella
62-1632 32 mm S-ROM Dome Patella
62-1635 35 mm S-ROM Dome Patella
62-1638 38 mm S-ROM Dome Patella
1987-27-112 X-Small 12 mm LPS Universal Insert
1987-27-114 X-Small 14 mm LPS Universal Insert
1987-27-116 X-Small 16 mm LPS Universal Insert
1987-27-118 X-Small 18 mm LPS Universal Insert
1987-27-121 X-Small 21 mm LPS Universal Insert
1987-27-123 X-Small 23 mm LPS Universal Insert
1987-27-126 X-Small 26 mm LPS Universal Insert
1987-27-128 X-Small 28 mm LPS Universal Insert
1987-27-131 X-Small 31 mm LPS Universal Insert
1987-27-212 Small 12 mm LPS Universal Insert
1987-27-214 Small 14 mm LPS Universal Insert
1987-27-216 Small 16 mm LPS Universal Insert
1987-27-218 Small 18 mm LPS Universal Insert
1987-27-221 Small 21 mm LPS Universal Insert
1987-27-223 Small 23 mm LPS Universal Insert
1987-27-226 Small 26 mm LPS Universal Insert
1987-27-228 Small 28 mm LPS Universal Insert
1987-27-231 Small 31 mm LPS Universal Insert
1987-27-312 Medium 12 mm LPS Universal Insert
1987-27-314 Medium 14 mm LPS Universal Insert
1987-27-316 Medium 16 mm LPS Universal Insert
1987-27-318 Medium 18 mm LPS Universal Insert
1987-27-321 Medium 21 mm LPS Universal Insert
1987-27-323 Medium 23 mm LPS Universal Insert
1987-27-326 Medium 26 mm LPS Universal Insert
1987-27-328 Medium 28 mm LPS Universal Insert
1987-27-331 Medium 31 mm LPS Universal Insert
IMPLANT LISTING
Cat. No. Size (mm) A/P M/L Stem Length Tray Thickness
1294-35-110 1 39.0 59.2 61.8 4.8
1294-35-115 1.5 40.7 61.8 61.8 4.8
1294-35-120 2 42.6 64.6 61.8 4.8
1294-35-125 2.5 44.2 67.1 61.8 4.8
1294-35-130 3 45.8 69.6 61.8 4.8
1294-35-140 4 49.3 74.9 61.8 4.8
1294-35-150 5 53.1 80.6 61.8 4.8
1294-35-160 6 57.2 86.8 61.8 4.8
1294-35-215 2+15 42.6 64.6 61.8 15
1294-35-225 2+25 42.6 64.6 61.8 25
1294-35-315 3+15 45.8 69.6 61.8 15
1294-35-325 3+25 45.8 69.6 61.8 25
1294-35-415 4+15 49.3 74.9 61.8 15
1294-35-425 4+25 49.3 74.9 61.8 25
Cat. No. Size (mm) A/P M/L Height
1294-54-000 29 26 29 40
1294-54-140 (Cemented) 29 26 29 40
1294-54-100 37 27 37 40
1294-54-110 45 27 45 40
1294-35-120 53 31 53 40
1294-35-130 61 34 61 40
MBT Revision Tray
MBT Revision Sleeve
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 57
IMPLANT LISTING
Cat. No. Size (mm)
1294-56-110 1-5
1294-56-111 1-10
1294-56-112 1-15
1294-56-115 1.5-5
1294-56-116 1.5-10
1294-56-117 1.5-15
1294-56-120 2-5
1294-56-121 2-10
1294-56-122 2-15
1294-56-125 2.5-5
1294-56-126 2.5-10
1294-56-127 2.5-15
Cat. No. Size (mm)
1294-56-130 3-5
1294-56-131 3-10
1294-56-132 3-15
1294-56-135 4-5
1294-56-136 4-10
1294-56-137 4-15
1294-56-140 5-5
1294-56-141 5-10
1294-56-142 5-15
1294-56-145 6-5
1294-56-146 6-10
1294-56-147 6-15
MBT Revision Augments
Note: If a tibial sleeve is used, MBT Revision Augments are only compatible with the 29 mm sleeve.
58 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
COMPATIBILITY CHART
LPSXX-SmallFemoral
Component
LPS X-SmallFemoral
Component
and
S-ROMX-Small Femoral
Component
S-ROMSmall
FemoralComponent
S-ROMMediumFemoral
Component
Tray No. M/L 56.6 66.7 66.7 71.2
MBT Revision TraySize 1
1294-35-110 59.2
MBT Revision TraySize 1.5
1294-35-115 61.8
MBT Revision TraySize 2
1294-35-120 64.6
MBT Revision TraySize 2.5
1294-35-125 67.1
MBT Revision TraySize 3
1294-35-130 69.6
MBT Revision TraySize 4
1294-35-140 74.9
MBT Revision TraySize 5
1294-35-150 80.6
MBT Revision TraySize 6
1294-35-160 86.8
LPS Universal inserts must match S-ROM or LPS Femoral Component size-to-size. For example, XX-Small femoral component = XX-Small polyethylene; Small femoral component = Small polyethylene, etc.
3
3
3 3 3
3 3 3
3 3 3 3
3 3 3 3
3 3 3 3
3 3 3 3
signifies preferred match
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 59
MBT REVISION PREPARATION CASECAT. NO. 2178-64-100
Description Cat. No.
A MBT Revision Cemented Stem Reamer, 13 mm 2178-63-185
B Tibial Cutting Block, 2 Degree 2178-40-086
C MBT Revision Reamer Adapter 2178-63-128
D MBT Revision Press-Fit Tibial Punch 2178-63-118
E MBT Revision Cemented Tibial Punch 2178-63-120
F Pin Driver 2490-94-000
G MBT Revision Drill Bushing 2178-63-100
H Pin Puller 96-6515
I SP2 IM Rod, 400 mm 96-6120
J SP2 IM Rod Handle 99-2011
K MBT Revision Cemented Bushing, 13 mm 2178-63-196
L MBT Revision Tapered Press-Fit Reamer 2178-63-104
M MBT Revision Tapered Cemented Reamer 2178-63-106
N Steinman Pins (Package of 10) 86-9117
Top Tray
A
CD F
G
B
H
NM
L
I
E
J
K
61 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Description Cat. No.
A MBT Revision 2-Degree Tibial Broaches 29 mm 2178-63-109
MBT Revision 2-Degree Tibial Broaches 37 mm 2178-63-111
MBT Revision 2-Degree Tibial Broaches 45 mm 2178-63-113
MBT Revision 2-Degree Tibial Broaches 53 mm 2178-63-115
MBT Revision 2-Degree Tibial Broaches 61 mm 2178-63-117
B MBT Tray Sleeve Trials 29 mm 2294-54-000
MBT Tray Sleeve Trials 37 mm 2294-54-100
MBT Tray Sleeve Trials 45 mm 2294-54-110
MBT Tray Sleeve Trials 53 mm 2294-54-120
MBT Tray Sleeve Trials 61 mm 2294-54-130
C LCS Completion Tibial Stylus 2178-40-045
D MBT Revision Tibial Broach Handle 96-6521
E Revision Sleeve Impactor 2178-63-124
F Revision Femoral Sleeve/Stem Impactor 2178-63-126
G SP2 Universal Handle 96-6520
H SP2 IM Tibial Alignment Device 96-6315
I MBT Tibial Impactor 9505-01-558
J MBT Revision Tibial View Plate, Size 1 2178-65-110
MBT Revision Tibial View Plate, Size 1.5 2178-65-115
MBT Revision Tibial View Plate, Size 2 2178-65-120
MBT Revision Tibial View Plate, Size 2.5 2178-65-125
MBT Revision Tibial View Plate, Size 3 2178-65-130
MBT Revision Tibial View Plate, Size 4 2178-65-140
MBT Revision Tibial View Plate, Size 5 2178-65-150
MBT Revision Tibial View Plate, Size 6 2178-65-160
MBT REVISION PREPARATION CASECAT. NO. 2178-64-100
A B C
E
D H
F I J
G
Bottom Tray
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 61
MBT REVISION STEM TRIALS AND INSTRUMENTS CASECAT. NO. 2178-64-110
Top Tray
Description Cat. No.
A Revision Femoral/Tibial Sleeve Clamp 2178-63-134
B SIGMA Tibial Cemented Stem Trial, Sizes 2-3, 13 x 60 mm 86-6502
C SIGMA Tibial Cemented Stem Trial, Sizes 1.5-3, 13 x 30 mm 86-6501
D Stem Trial Extractor 86-5226
E Fluted Tibial Stem Trials – 75 mm
75 x 10 mm 86-6874
75 x 12 mm 86-6875
75 x 14 mm 86-6876
75 x 16 mm 86-6877
75 x 18 mm 86-6878
75 x 20 mm 86-6879
75 x 22 mm 86-6880
75 x 24 mm 86-6881
A
B F GE
C
D
62 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
MBT REVISION STEM TRIALS AND INSTRUMENTS CASECAT. NO. 2178-64-110
Top Tray
Description Cat. No.
F Fluted Tibial Stem Trials – 115 mm
115 x 10 mm 86-6882
115 x 12 mm 86-6883
115 x 14 mm 86-6884
115 x 16 mm 86-6885
115 x 18 mm 86-6886
115 x 20 mm 86-6887
115 x 22 mm 86-6888
115 x 24 mm 86-6889
G Fluted Tibial Stem Trials – 150 mm
150 x 10 mm 86-6890
150 x 12 mm 86-6891
150 x 14 mm 86-6892
150 x 16 mm 86-6893
150 x 18 mm 86-6894
150 x 20 mm 86-6895
150 x 22 mm 86-6896
150 x 24 mm 86-6897
A
B F GE
C
D
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 63
MBT REVISION STEM TRIALS AND INSTRUMENTS CASECAT. NO. 2178-64-110
Bottom Tray
Description Cat. No.
A Press-Fit Rod Wrench 86-5189
B SIGMA Tibial Cemented Stem Trial, Sizes 2-3, 13 x 60 mm 86-6502
C SIGMA Tibial Cemented Stem Trial, Sizes 1.5-3, 13 x 30 mm 86-6501
D Fluted Tibial Stem Trials – 75 mm
75 x 10 mm 86-6874
75 x 12 mm 86-6875
75 x 14 mm 86-6876
75 x 16 mm 86-6877
75 x 18 mm 86-6878
75 x 20 mm 86-6879
75 x 22 mm 86-6880
75 x 24 mm 86-6881
A
B D E F
C
64 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
MBT REVISION STEM TRIALS AND INSTRUMENTS CASECAT. NO. 2178-64-110
Bottom Tray
Description Cat. No.
E Fluted Tibial Stem Trials – 115 mm
115 x 10 mm 86-6882
115 x 12 mm 86-6883
115 x 14 mm 86-6884
115 x 16 mm 86-6885
115 x 18 mm 86-6886
115 x 20 mm 86-6887
115 x 22 mm 86-6888
115 x 24 mm 86-6889
F Fluted Tibial Stem Trials – 150 mm
150 x 10 mm 86-6890
150 x 12 mm 86-6891
150 x 14 mm 86-6892
150 x 16 mm 86-6893
150 x 18 mm 86-6894
150 x 20 mm 86-6895
150 x 22 mm 86-6896
150 x 24 mm 86-6897
A
B D E F
C
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 65
MBT REVISION REAMERS CASECAT. NO. 2178-64-105
Description Cat. No.
A Press-Fit Rod Wrench 86-5189
B I.M. Rod Sleeve Guide, 12 mm 2178-63-187
C I.M. Rod Sleeve Guide, 14 mm 2178-63-188
D LCS Reamer Depth Scale 2178-63-102
E Revision Femoral/Tibial/Sleeve Clamp 2178-63-134
F I.M. Reamer, 9 mm 2178-56-045
G MBT Revision Reamers
MBT Revision Reamer, 10 mm 2178-63-170
MBT Revision Reamer, 11 mm 2178-63-171
MBT Revision Reamer, 12 mm 2178-63-172
MBT Revision Reamer, 13 mm 2178-63-173
MBT Revision Reamer, 14 mm 2178-63-174
MBT Revision Reamer, 15 mm 2178-63-175
Top Tray
A
D
G
F E
B C
66 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
Description Cat. No.
A MBT Revision Reamers
MBT Revision Reamer, 16 mm 2178-63-176
MBT Revision Reamer, 17 mm 2178-63-177
MBT Revision Reamer, 18 mm 2178-63-178
MBT Revision Reamer, 19 mm 2178-63-179
MBT Revision Reamer, 20 mm 2178-63-180
MBT Revision Reamer, 21 mm 2178-63-181
MBT Revision Reamer, 22 mm 2178-63-182
MBT Revision Reamer, 23 mm 2178-63-183
MBT Revision Reamer, 24 mm 2178-63-184
B I.M. Rod Sleeve Guide, 16 mm 2178-63-189
C I.M. Rod Sleeve Guide, 18 mm 2178-63-190
D I.M. Rod Sleeve Guide, 20 mm 2178-63-191
E I.M. Rod Sleeve Guide, 22 mm 2178-63-192
F I.M. Rod Sleeve Guide, 24 mm 2178-63-193
G I.M. Rod Sleeve Guide, 26 mm 2178-63-194
H MBT Revision T-Handle 2178-63-137
I Modified Hudson Adapter 2178-63-136
MBT REVISION REAMERS CASECAT. NO. 2178-64-105
Bottom Tray
A
H
I
B
C
DF
E G
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 67
MBT REVISION WEDGE TRIALS AND INSTRUMENTSCAT. NO. 2178-64-115
Top Tray
Description Cat. No.
A Size 1, 5 mm 2294-56-110
Size 1, 10 mm 2294-56-111
Size 1, 15 mm 2294-56-112
B Size 1.5, 5 mm 2294-56-115
Size 1.5, 10 mm 2294-56-116
Size 1.5, 15 mm 2294-56-117
C Size 2, 5 mm 2294-56-120
Size 2, 10 mm 2294-56-121
Size 2, 15 mm 2294-56-122
D Size 2.5, 5 mm 2294-56-125
Size 2.5, 10 mm 2294-56-126
Size 2.5, 15 mm 2294-56-127
E Size 3, 5 mm 2294-56-130
Size 3, 10 mm 2294-56-131
Size 3, 15 mm 2294-56-132
F Size 4, 5 mm 2294-56-135
Size 4, 10 mm 2294-56-136
Size 4, 15 mm 2294-56-137
G Size 5, 5 mm 2294-56-140
Size 5, 10 mm 2294-56-141
Size 5, 15 mm 2294-56-142
H Size 6/7, 5 mm 2294-56-145
Size 6/7, 10 mm 2294-56-146
Size 6/7, 15 mm 2294-56-147
A B C D E F G H
Size 1
5 mm
10 mm
15 mm
5 mm
10 mm
15 mm
Size 1.5 Size 2 Size 2.5 Size 3 Size 4 Size 5 Size 6/8
68 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
MBT REVISION WEDGE TRIALS AND INSTRUMENTSCAT. NO. 2178-64-115
Bottom Tray
Description Cat. No.
A MBT Revision Tray Trial, Size 1 2294-36-110
MBT Revision Tray Trial, Size 1.5 2294-36-115
MBT Revision Tray Trial, Size 2 2294-36-120
MBT Revision Tray Trial, Size 2.5 2294-36-125
MBT Revision Tray Trial, Size 3 2294-36-130
MBT Revision Tray Trial, Size 4 2294-36-140
MBT Revision Tray Trial, Size 5 2294-36-150
MBT Revision Tray Trial, Size 6 2294-36-160
B MBT Revision Tray Trial with Stem, Size 1 2294-35-110
C MBT Revision Tray Trial with Stem, Size 1.5 2294-35-115
D MBT Revision Tray Trial with Stem, Size 2 2294-35-120
E MBT Revision Tray Trial with Stem, Size 2.5 2294-35-125
F MBT Revision Cutting Block 2178-63-122
G Tibial Wedge Trial Screwdriver 86-0277
H MBT Wedge Cutting Attachment 2178-63-130
I SP2 Alignment Rods 99-1016
J Tibial Trial Alignment Handle 96-6330
K MBT Revision Trial Post 2178-63-132
L Modular Plus Torque Driver 86-0284
M MBT Tray Trial Fixation Pins 2178-30-123
N MBT Revision Tray Trial with Stem, Size 3 2294-35-130
O MBT Revision Tray Trial with Stem, Size 4 2294-35-140
P MBT Revision Tray Trial with Stem, Size 5 2294-35-150
Q MBT Revision Tray Trial with Stem, Size 6 2294-35-160
A
B C
F
G
H IK
J
D E
M
QPON
L
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 69
Bottom Tray
SIGMA HP REVISION FEMORAL PREP INSTRUMENT CASECAT. NO. 2011-03-049
Description Cat. No.
A Visualization Wing 96-6530
B Completion Revision Femoral Tapered Reamer 2178-60-030
C Universal Revision Femoral Broach Reamer 96-1671
D Universal Revision Femoral Broach 31 mm 96-1683
E Universal Revision Femoral Broach 34 mm 96-1684
F Universal Revision Femoral Broach 40 mm 96-1685
G Universal Revision Femoral Broach 46 mm 96-1686
H SIGMA HP Revision Adapter Removable Shaft 2011-03-029
I SIGMA HP Distal Femoral Resection Guide 9505-01-235
J SIGMA HP Distal Femoral Connector 9505-01-238
K SIGMA HP Revision Distal Femoral Block 9505-01-239
L SIGMA HP Distal Femoral Alignment Guide 9505-01-234
M Universal Revision Femoral Broach Handle 96-1682
A
B C D
M
E F GI
K
J
L
H
71 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
SIGMA HP REVISION FEMORAL TRIAL INSTRUMENT CASECAT. NO. 2011-03-050
Top Tray
Description Cat. No.
A LCS/SIGMA Revision Femoral Sleeve Trial 20 mm 2294-53-100
B LCS/SIGMA Revision Femoral Sleeve Trial 31 mm 2294-53-110
C LCS/SIGMA Revision Femoral Sleeve Trial 34 mm 2294-53-120
D LCS/SIGMA Revision Femoral Sleeve Trial 40 mm 2294-53-130
E LCS/SIGMA Revision Femoral Sleeve Trial 46 mm 2294-53-140
F Modular Plus Torque Driver 86-0284
G SIGMA HP Revision TC3 Box Trial Size 2 2011-03-011
H SIGMA HP Revision TC3 Box Trial Size 2.5 2011-03-012
I SIGMA HP Revision TC3 Box Trial Size 3 2011-03-013
J SIGMA HP Revision TC3 Box Trial Size 4 2011-03-014
K SIGMA HP Revision TC3 Box Trial Size 5 2011-03-015
L SIGMA HP Revision Femoral Adapter Shift Tool 2011-03-057
M SIGMA Femoral Adapter Sleeve Bolt Trial Neutral 2011-03-052
N SIGMA Femoral Adapter Stem Bolt Trial Neutral 2011-03-051
O Femoral Augment Plug Puller 86-5151
P SP2 Torque Driver Handle Extension 96-6301
Q Large Fragment Screwdriver Shank 8242-19-000
R 2.0 Nm Torque-limiting Screwdriver 2141-18-001
S Femoral Adapter Torque Wrench 96-1673
T Femoral Adapter Holding Clamp 96-1674
A B C
O
QP
R
N
M K
L
S
T
ED
F G H I J
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 71
LPS HINGE INSERT TRIALS CAT. NO. 2987-99-039
Description Cat. No.
A Xsm Hinged Insert Trials
LPS Hinge Insert Trial 12 mm, Xsm 2987-22-112
LPS Hinge Insert Trial 14 mm, Xsm 2987-22-114
LPS Hinge Insert Trial 16 mm, Xsm 2987-22-116
LPS Hinge Insert Trial 18 mm, Xsm 2987-22-118
LPS Hinge Insert Trial 21 mm, Xsm 2987-22-121
LPS Hinge Insert Trial 23 mm, Xsm 2987-22-123
LPS Hinge Insert Trial 26 mm, Xsm 2987-22-126
LPS Hinge Insert Trial 28 mm, Xsm 2987-22-128
B Sm Hinged Insert Trials
LPS Hinge Insert Trial 12 mm, Sm 2987-22-212
LPS Hinge Insert Trial 14 mm, Sm 2987-22-214
LPS Hinge Insert Trial 16 mm, Sm 2987-22-216
LPS Hinge Insert Trial 18 mm, Sm 2987-22-218
LPS Hinge Insert Trial 21 mm, Sm 2987-22-221
LPS Hinge Insert Trial 23 mm, Sm 2987-22-223
LPS Hinge Insert Trial 26 mm, Sm 2987-22-226
LPS Hinge Insert Trial 28 mm, Sm 2987-22-228
C Med Hinged Insert Trials
LPS Hinge Insert Trial 12 mm, Med 2987-22-312
LPS Hinge Insert Trial 14 mm, Med 2987-22-314
LPS Hinge Insert Trial 16 mm, Med 2987-22-316
LPS Hinge Insert Trial 18 mm, Med 2987-22-318
LPS Hinge Insert Trial 21 mm, Med 2987-22-321
LPS Hinge Insert Trial 23 mm, Med 2987-22-323
LPS Hinge Insert Trial 26 mm, Med 2987-22-326
LPS Hinge Insert Trial 28 mm, Med 2987-22-328
A
B
C
72 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
LPS HINGE INSERT TRIALS CAT. NO. 2987-99-039
Description Cat. No.
A XXSm Hinged Insert Trials
LPS Hinge Insert Trial XX-Sm 12 mm 2987-22-012
LPS Hinge Insert Trial XX-Sm 14 mm 2987-22-014
LPS Hinge Insert Trial XX-Sm 16 mm 2987-22-016
LPS Hinge Insert Trial XX-Sm 18 mm 2987-22-018
LPS Hinge Insert Trial XX-Sm 21 mm 2987-22-021
LPS Hinge Insert Trial XX-Sm 23 mm 2987-22-023
B 31mm Hinged Insert Trials
LPS Hinge Insert Trial 31 mm, Xsm 2987-22-131
LPS Hinge Insert Trial 31 mm, Sm 2987-22-231
LPS Hinge Insert Trial 31 mm, Med 2987-22-331
C MBT Rev Trial Post 2178-63-132
D LPS Hinge Pin Trial 2987-15-000
D
A
B
C
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 73
S-ROM HINGE FEMORAL INSTRUMENTS CAT. NO. 2178-01-105
Description Cat. No.
A S-ROM Driver Femoral 633802
B S-ROM Knee Aug Blk Cem Clamp 633810
C S-ROM Separator,Femoral Sleeve 634182
D Sleeve/Stem Extractor 86-5226
E S-ROM NRH Dist Fem Aug Trl 5 mm 633785
F S-ROM NRH Dist Fem Aug Trl 10 mm 633790
G S-ROM NRH Hinge Pin Trial 634133
H S-ROM Universal Handle 633806A
I S-ROM NRH Fem Trial Right Med 634131R
J S-ROM NRH Fem Trial Left Med 634132L
K S-ROM NRH Fem Trial Right Small 634151R
L S-ROM NRH Fem Trial Left Small 634151L
M S-ROM NRH Fem Trial Right Xsmall 634161R
N S-ROM NRH Fem Trial Left Xsmall 634161L
A G
D
C
B E
F
H
I J K L
M N
74 DePuy Synthes Joint Reconstruction S-ROM® NOILES Rotating Hinge Surgical Technique
S-ROM HINGE FEMORAL INSTRUMENTS CAT. NO. 2178-01-105
Description Cat. No.
A S-ROM Guide Anterior Reference 633370A
B S-ROM Xsm/Sm A/P Block 2163-11-001
C S-ROM Med A/P Block 2163-11-004
D S-ROM NRH Fem Box Cut Guide 633385
E S-ROM Box Cut Guide Augment Space, 5 mm 633305A
F S-ROM Box Cut Guide Augment Space, 10 mm 633306A
G SP2 Removable Handles 96-6147
H Femoral Locating Device Outrigger 96-6112
I Femoral Locating Device 96-6110
J Distal Femoral Cutting Block 96-6115
C
B
F E
A G
H
I
D
J
Surgical Technique S-ROM® NOILES Rotating Hinge DePuy Synthes Joint Reconstruction 75
IMPORTANT:
This Essential Product Information sheet does not include all of the information necessary for selection and use of a device. Please see full labeling for all necessary information.
INDICATIONS AND USAGE:
The S-ROM NOILES Rotating Hinge Knee is indicated for use with PMMA bone cement in primary or revision cases in patients:
• who have reached skeletal maturity and
• for whom the surgeon has decided to resect both cruciate ligaments or whose cruciate ligaments are absent or incompetent and
• who exhibit insufficiency of lateral/collateral ligaments and other soft supporting tissue due to the following conditions:
• Rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, and arthritis secondary to a variety of diseases and anomalies
• Failure of a previous knee reconstruction procedure
• Trauma
CONTRAINDICATIONS:
1. Active infection or history of general infections or local infectious disease.
2. Vascular insufficiency, muscular atrophy or neuromuscular disease in the affected limb.
3. Advanced loss of osteochondral structure that would preclude proper fixation of the prosthesis.
4. Tumors of the supporting bone structure, systemic and metabolic disorders leading to progressive deterioration of solid bone support.
5. Drug or alcohol addiction, or limiting neuropathic disease.
6. Skeletal immaturity.
7. Obesity or very active lifestyle that can produce loads on the prosthesis that can lead to failure of the fixation of the device or device itself.
8. Allergic reaction to the implant materials.
9. Inadequate flexor and extensor mechanism necessary to achieve a functional prosthetic joint.
WARNINGS:
Improper prosthesis selection or alignment, inadequate fixation, use where contraindicated or in patients whose medical, physical, mental, or occupational conditions will likely result in extreme stresses to the implant, may result in premature failure due to loosening, fracture, or wear.
Post-operative care is extremely important. The patient should be instructed on the limitations of the device and should be cautioned regarding load bearing, ranges of motion, and activity levels permissible. Early motion and load bearing should be carefully controlled.
The S-ROM Tibial Base, Tibial Sleeve, Tibial Stem Extension, and Tibial Augmentation Blocks may not be used with the NOILES Posterior Stabilized Knee.
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
PRECAUTIONS:
The S-ROM NOILES Rotating Hinge Knee is designed to articulate from 6 degrees hyperextension to 110 degrees flexion. If, due to grossly inadequate soft tissue integrity, flexion beyond 90 degrees causes luxation of the plateau assembly out of the tibial base, the patient must have a knee brace posto-peratively to limit flexion to 90 degrees. In such cases, the surgeon should consider closing the wound with the knee in full extension.
The size of the tibial plateau assembly must correspond with the size of the femoral component.
The size of the tibial augmentation block must correspond to the size of the tibial base.
Femoral sleeves are required when using femoral stem extensions.
A femoral plug is required with the femoral sleeve when a femoral stem extension is not used.
A tibial cap is required with the tibial sleeve when a tibial stem extension is not used.
Tibial augmentation blocks cannot be used when tibial sleeves are being used.
An implant should never be reused. Any implant, once used, should be discarded. Even though it appears undamaged, it may have small defects and/or internal stress patterns that may lead to failure. Likewise, a new implant should be handled carefully to avoid damage that could compromise the integrity of the device and cause early failure or loosening.
The wear rate of prosthesis contact surfaces is greatly accelerated if loose fragments of bone cement become detached and act as an abrasive in the bearing surfaces. When using bone cement, care should be taken to remove all excess cement from the periphery of the implant.
*DePuy Synthes Joint Reconstruction Single Use devices have not been designed to undergo or withstand any form of alteration, such as disassembly, cleaning or re-sterilization, after a single patient use. Reuse can potentially compromise device performance and patient safety.
ADVERSE EFFECTS:
Fracture may occur due to improper preparation of the implant site or if excessive force is used during seating of the implant. Transient peroneal palsy has been reported following total knee arthroplasty, especially after correction of severe flexion or valgus deformities.
Patients have complained of persistent pain and stiffness following total knee arthroplasty. In addition, patellar tendon rupture, femoral-tibial subluxation or dislocation, and persistent ligamentary laxity have been reported with the use of total knee implants. Infection and loosening have been reported following total joint arthroplasty, as have wear and failure due to fracture of knee prosthesis components.
Histological reactions have been reported as an apparent response to exposure to a foreign material. The actual clinical significance of these reactions is unknown.
Serious adverse side effects may necessitate surgical intervention.
The third party trademarks used herein are trademarks of their respective owners.
S-ROM MODULAR TOTAL KNEE SYSTEM | NOILES ROTATING HINGE KNEE SYSTEM
DePuy Orthopaedics, Inc.700 Orthopaedic DriveWarsaw, IN 46582USAT. +1 (800) 366-8143
www.depuysynthes.com
DePuy (Ireland)Loughbeg Bingaskiddy Co. Cork Ireland Tel:+35 (321) 491-4278Fax: +35 (321) 491-4199
DePuy Internationl Ltd.St. Anthony's Road Leeds LS11 8DT England Tel:+44 (113) 270-0461Fax: +44 (113) 272-4101
*DePuy Synthes Joint Reconstruciton is a division of DePuy Orthopaedics, Inc.
© DePuy Synthes Joint Reconstruction, a division of DOI 2014. All rights reserved. DSUS/JRC/0414/0052
0086