RP HPLC,Efficiency

34
RP-HPLC, EFFICIENCY PARAMETERS, RESOLUTION, COMPARISON OF SENSITIVITY, SELECTIVITY AND FIELD OF APPLICATIONS OF DETECTORS Mr. Shrishail Pattadakal

Transcript of RP HPLC,Efficiency

Page 1: RP HPLC,Efficiency

RP-HPLC, EFFICIENCY PARAMETERS, RESOLUTION,

COMPARISON OF SENSITIVITY, SELECTIVITY

AND FIELD OF APPLICATIONS OF

DETECTORS

Mr. Shrishail Pattadakal

Page 2: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 2

Contents

Introduction

Types of HPLC

RP-HPLC

HPLC Parameters

Resolution

Conclusion

HPLC Detectors

Page 3: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 3

Introduction

H PLC

= High

= Pressure

= Liquid

= Chromatography

(Performance)

Page 4: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 4

What is HPLC ?

HPLC is a form of column chromatography used frequently in biochemistry and analytical chemistry. To separate, identify and quantify compounds.

Separation technique involving mass-transfer between stationary and mobile phase.

Page 5: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 5

Types of HPLC

HPLC

NP-HPLC RP-HPLC Aq. NP-HPLC

Page 6: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 6

Normal Phase HPLC

Using normal phase solvents :

Most retained compounds are polar

Least retained compounds are less polar

More non-polar solvents make compounds retain longer

Page 7: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 7

Aqueous NP Chromatography

Aqueous normal-phase chromatography (ANP) is a chromatographic technique which encompasses the mobile phase region between reversed-phase chromatography (RP) and organic normal phase chromatography (ONP). This technique is used to achieve unique selectivity for hydrophilic compounds, showing normal phase elution using reverse-phase solvents.

Page 8: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 8

Reverse Phase HPLC

Solvents using :

Most retained compounds are least polarLeast retained compounds are most polarMore polar solvents make compounds retain longer.

Page 9: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 9

Principle• Nonpolar (nonspecific) interactions of analyte with hydrophobic

adsorbent surface (-C18, C8, Phenyl, C4)

• Difference in analyte sorption affinities results in their separation

• More polar analytes retained less

• Analytes with larger hydrophobic part are retained longer

• Almost no separation of structural isomers

Page 10: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 10

Parameters Used In HPLC

1) Retention parameter

2) Column efficiency parameter

3) Peak symmetric parameter

Page 11: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 1111

Retention Parameters

Page 12: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 1212

Efficiency Parameters

Page 13: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 13

Peak symmetry

S : Symmetry factor ( T : Tailing factor )

S = 2f

W0.05

fW0.05

h x 0.05h

S = 1 : The peak is completely symmetric.S > 1 : TailingS < 1 : Leading

Page 14: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 1414

Resolution

Page 15: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 15

Selectivity

• Selectivity (a) is equivalent to the relative retention of the solute peaks and, unlike efficiency, depends strongly on the chemical properties of the chromatography medium.

• The selectivity, a, for two peaks is given by α = k2´ /k1´ = V2 - V0/V1 – V0 = V2/V1

where V1 and V2 are the retention volumes, and k2´ /k1´ are the capacity factors, for peaks 1 and 2 respectively, and V0 is the void volume of the column.

Page 16: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 1616

Eluent Composition Effect on Selectivity

0

2

4

6

8

10

12

14

16

18

0 1 2 3 4 5 6 7

0

5

10

15

20

25

0 1 2 3 4 5 6 7

0

2

4

6

8

10

12

14

16

0 1 2 3 4 5 6 7 8 9

0

2

4

6

8

10

12

0 2 4 6 8 10 12

90% MeCN

80% MeCN

70% MeCN

60% MeCN

Page 17: RP HPLC,Efficiency

Sensitivity sensitivity was considered to define how small a mass or

concentration of a solute could be unambiguously identified by a detector. This concept of sensitivity still persists, although it has been replaced in some cases by minimum detectable mass (MDM) or, alternatively, minimum detectable concentration (MDC).

MDM or MDC are classically defined as the mass or concentration of solute passing through the detector that will give a signal equivalent to twice that of the noise. Originally the signal given by a detector for unit mass or unit concentration change was termed the detector response.

Monday, April 10, 2023 Mr. Shrishail Pattadkal 17

Page 18: RP HPLC,Efficiency

HPLC Detectors

Page 19: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 19

Requirements for an HPLC detector

• Good sensitivity (high signal, low noise)

• No interference from mobile phase

• Must be able to work in a liquid phase environment

Page 20: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 20

Types of Detectors

• Refractive Index• UV-Visible• Fluorescence• Conductivity (for ion chromatography)• Mass Spectrometry

Page 21: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 21

Types of HPLC Detectors

Solute Property

UV-Vis

Electrochemical

Fluorescence

FixedVariable

Photodiode ArrayAmperometry

Pulse Amperometry

Voltammetry

Coulometry

Bulk Property

Refractive Index

ConductivitySuppressed

Non-suppressed

Deflectance Type

Reflectance Type

Page 22: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 22

UV-Vis Absorbance Detector

UV Detectors measure the ability of a sample to absorb light. This can be accomplished at one or several wavelength.

Fixed wavelength measures at one wavelength, usually 254 nm.

Variable wavelength measures at one wavelength at a time, but can detect over a range of wavelengths.

Diode Array measures a spectrum of wavelengths simultaneously.

Page 23: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 23

UV-Vis Absorbance Detector

Page 24: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 24

Fluorescent Detectors

• Fluorescent detectors measures the ability of a compound to absorb then re-emit light at given wavelength.

• Each compound has a characteristic fluorescence.

• The excitation source passes through the flow-cell to a photo detector while a monochromator measures the emission wavelengths.

• Has sensitivity limit of 10-9 to 10-11 gm/ml.

Page 25: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 25

Page 26: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 26

Fluorosence

Page 27: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 27

Electrochemical Detector• It measure compounds that undergo oxidation

or reduction reactions.

• Usually accomplished by measuring gain or loss of electrons from migrating samples as they pass between electrodes at a given difference in electrical potential.

• Has sensitivity of 10-12 to 10-13 gm/ml.

Page 28: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 28

Optical Detectors

There are mainly 4 types of detectors,

1) Circular Dichroism Chiral Detector.

2) Chiral Detector.

3) Chemiluminiscence Detector.

4) Refractive Index Detector.

Page 29: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 29

Circular Dichroism Chiral Detector

• Simultaneous acquisition of UV factor signals.

• Direct determination of optical isomer separation & purity.

• Spectral scanning capability.

Page 30: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 30

Chiral Detectors• These are specially designed for optically active

compounds.

Chemiluminescence Detector

• Detection limits as demand increases for detection of trace amounts of lipids, nucleotides, nitrogen oxides & catachol amines.

Page 31: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 31

Refractive Index

• Monitors are the closest to being universal HPLC detectors, as nearly all dissolved solutes alter the refractive index of the mobile phase.

• They are differential detectors, generating a signal that depends on the difference between the RI of the pure mobile phase and the modified value caused by the dissolved solute.

Page 32: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 32

Instrumentation of Refractive Index

Page 33: RP HPLC,Efficiency

ConclusionHPLC is probably the most universal type of analytical procedure;

its application areas include, quality control, process control, forensic analysis, environmental monitoring and clinical testing. In addition HPLC also ranks as one of the most sensitive analytical procedures and is unique in that it easily copes with multi-component mixtures. It has achieved this position as a result of the constant evolution of the equipment used in LC to provide higher and higher efficiencies at faster and faster analysis times with a constant incorporation of new highly selective column packings.

Monday, April 10, 2023 Mr. Shrishail Pattadkal 33

Page 34: RP HPLC,Efficiency

Monday, April 10, 2023 Mr. Shrishail Pattadkal 34

Bibliography

Separation Chemistry – R.P.Budhiraja, New age international(p) Limited.

Analytical Chromatography – Gurudeep R. Chatwal Edited By Madhu Arora, Himalaya Publishing House