Review CR & DIGITAL IMAGING (1)
2012 – RT 244 wk 15
References: Bushong Vol 9www.sprawls.org/resources/DIGRAD/classroom.htm
1
![Page 2: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/2.jpg)
ObjectivesObjectivesDigital imaging review
Review CR fundamentals
2
![Page 3: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/3.jpg)
Digital ImagingDigital ImagingImage acquisition that produces
an electronic image that can be viewed and manipulated on a computer.
Examples?
3
![Page 4: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/4.jpg)
Methods to Digitize an Image• 1. Film Digitizer
• 2. Video Camera (vidicon or plumbicon)
• 3. Computed Radiography
• 4. Direct Radiography– PACS– DICOM
4
![Page 5: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/5.jpg)
Digital
Radiography
DirectCapture
IndirectCapture
Direct-to-DigitalRadiography
(DDR)-Selenium
ComputedRadiography
(CR) - PSL
LaserScanningDigitizers
Direct-to-DigitalRadiographySilicon Scint.
DDR CR
5
![Page 6: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/6.jpg)
6
![Page 7: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/7.jpg)
Computer Language
• Computers operate on the Binary Number System
• It has only two digits, 0 and 1
• Computers function by converting all data into binary values.
7
![Page 8: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/8.jpg)
8
![Page 9: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/9.jpg)
ByteByteRepresents one character, digit, or
value. A bit describes the smallest unit of
measure 0 or 1 – computers ultimately understand only 0 or 1
Byte are 8 bitsA kilobyte represents 1024 bytes,
megabyte is 1 million bytes, gigabyte is approximately 1 billion bytes
9
![Page 10: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/10.jpg)
Alphabet in Binary
10
![Page 11: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/11.jpg)
What is a Pixel? 11
![Page 12: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/12.jpg)
Basics of Digital Images
• digital images are a (matrix) of pixel (picture element) values
12
![Page 13: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/13.jpg)
Pixel
13
![Page 14: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/14.jpg)
14
![Page 15: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/15.jpg)
Computed Radiography
Fundamentals of
Computerized Radiography
15
![Page 16: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/16.jpg)
CR SYSTEM CR SYSTEM COMPONENTSCOMPONENTS
What are the CR system components?
16
![Page 17: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/17.jpg)
CR SYSTEM CR SYSTEM COMPONENTSCOMPONENTS
CASSETTES (phosphor plates)
ID STATIONIMAGE PREVIEW (QC) STATION
DIGITIZERVIEWING STATION
17
![Page 18: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/18.jpg)
Imaging Plate (IP)Imaging Plate (IP)Contained in a cassette
Handled the same as S/F cassettes
Processed more like daylight processor with no chemicals
IP has lead backing to reduce scatter
18
![Page 19: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/19.jpg)
CR – PSP plate• photostimulable phosphor (PSP) plate
• Exit photons energizes the PSP plate
• The energy is stored in traps on plate (latent image)
• PLATE scanned in CR READER
19
![Page 20: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/20.jpg)
Imaging Plate Imaging Plate ConstructionConstructionA thin sheet of plastic
IP’s have several layers◦A protective layer. This is a very
thin, tough, clear plastic that protects the phosphor layer
◦A phosphor or active layer. This is a layer of photostimulable phosphor that “traps” electrons during exposure
20
![Page 21: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/21.jpg)
Active Layer - Crystals
• The materials that make up the PSP plate are from the barium fluorohalide family.
• Barium fluorohalide, chlorohalide, or bromohalide crystals. The most common crystal uses is barium fluorohalide with europium
21
![Page 22: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/22.jpg)
Acquiring the Image What is the correct order?
• Violet light is captured by PMT – is amplified and converted into a digital signal
• cassette is put into the reader, the imaging plate is extracted
• light is sent to the analog to digital converter (ADC). To convert light to binary.
• e- return to ground state, visible light is emitted
• remnant beam interacts with electrons in the barium fluorohalide crystals
• imaging plate is scanned with a helium laser beam or solid-state laser diodes 22
![Page 23: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/23.jpg)
Acquiring the Image• The remnant beam interacts with electrons in
the barium fluorohalide crystals. This interaction stimulates, or gives energy to, electrons in the crystals, allowing them to enter the conductive layer, where they are trapped in an area of the crystal known as the color or phosphor center.
• This trapped signal will remain for hours, even days, although deterioration begins almost immediately. IR should be processed as soon as possible.
• The trapped signal is never completely lost.
23
![Page 24: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/24.jpg)
Imaging Plate Imaging Plate ConstructionConstructionA reflective layer. This is a layer
that sends light in a forward direction when released in the cassette reader. This layer may be black to reduce the spread of stimulating light and the escape of emitted light. Some detail is lost in this process.
24
![Page 25: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/25.jpg)
IP Construction
25
![Page 26: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/26.jpg)
Cross section of a PSP Cross section of a PSP screenscreen
26
![Page 27: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/27.jpg)
Needle PSP increase the absorption of x-rays and limit the spread of light emission
27
![Page 28: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/28.jpg)
IP DesignIP DesignDesigned to optimize the
intensity of light release. (CE)
Enhance the absorption of x-rays (DQE)
Limit the spread of light emission for more detail.
28
![Page 29: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/29.jpg)
Photostimulable Luminescence• When the cassette is put into the reader,
the imaging plate is extracted and scanned with a helium laser beam or, in more recent systems, solid-state laser diodes. This beam, about 100μm wide with a wavelength of 633 nm (or 670 to 690 nm for solid state), scans the plate with red light in a raster pattern and gives energy to the trapped electrons.
29
![Page 30: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/30.jpg)
X-ray interaction with a PSP screen
1
2
X-ray interactions with the screen phosphors causes an e- to excited
When e- return to groundstate visible light is emitted
30
![Page 31: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/31.jpg)
CR Phosphor PlatesCR Phosphor Plates
ABSORPTION EMISSION
X-RAY
LIGHT
LASER STIMULATION
ELECTRONTRAP
ELECTRONTRAP
31
![Page 32: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/32.jpg)
CR Reader – PSP plateCR Reader – PSP plateStimulates the matrix of trapped E-
by a RED OR ULTRAVIOLET laser light
Trapped E- energy is released in a form of VIOLET/BLUE light
Violet light is captured by PMT – is amplified and converted into a digital signal
32
![Page 33: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/33.jpg)
Producing a PSL signalProducing a PSL signal50% of the excited e- return to
ground state immediately, resulting in light (VIOLET/BLUE) emission.
Slow scan = plate Fast scan = laser
33
![Page 34: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/34.jpg)
How CR worksHow CR worksReleased light is captured by a
PMT (photo multiplier tube). An ultrasensitive photomultiplier tube or CCD (charged couple device)
PSP light is amplified by the PMT or CCD
This light is sent to the analog to digital converter (ADC). To convert light to binary.
34
![Page 35: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/35.jpg)
Sequence of CR imaging
35
![Page 36: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/36.jpg)
Processing of digital images can be used to
change most image characteristics. • Three possibilities include
processing methods to:• Adjust and optimize the
image contrast characteristics
• LUT & Processing Algorithms
• Reduce image noise • Increase visibility of detail • Some type of digital
image processing is used with most of the medical imaging modalities.
36
![Page 37: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/37.jpg)
Brightness & Contrast
• Optimum kVp & mAs has changed for digital
• kVp changes for SUBJECT contrast – not image contrast
• mAs does not influence DENSITY the same as it did with F/S
• The image is POSTPROCESSED – with changing PROCESSING ALGORITHMS
37
![Page 38: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/38.jpg)
38
![Page 39: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/39.jpg)
39
![Page 40: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/40.jpg)
• digital processing methods that can be used to adjust the contrast characteristics of an image.
•Look Up Table (LUT) processing •Windowing •Are used in digital radiography as well as with many of the other imaging modalities.
40
![Page 41: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/41.jpg)
41
![Page 42: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/42.jpg)
DIGITAL “DENSITY”
42
![Page 43: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/43.jpg)
Dynamic Range
• The range of exposure values to which the
• image receptor will respond.
• The greater the range of values that a
• receptor will respond to the greater the
• dynamic range.
43
![Page 44: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/44.jpg)
Characteristic curveof radiographic film
44
![Page 45: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/45.jpg)
45
![Page 46: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/46.jpg)
46
![Page 47: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/47.jpg)
47
![Page 48: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/48.jpg)
Widow level & width
Same photons at the image receptor
Image is post processed – changing
Brightness and contrast of image appearance
48
![Page 49: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/49.jpg)
• The ability to window is a valuable feature of all digital images.
• Windowing is the process of selecting some segment of the total pixel value range
• and then displaying the pixel values within that segment over the full brightness (shades of gray) range from white to black.
49
![Page 50: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/50.jpg)
windowing
• Important point...Contrast will be visible only for the pixel values that are within the selected window.
• All pixel values that are either below or above the window will be all white or all black and display no contrast.
• The person controlling the display can adjust both the center and the width of the window. The combination of these two parameters determine the range of pixel values that will be displayed with contrast in the image.
50
![Page 51: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/51.jpg)
advantages of windowing?51
![Page 52: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/52.jpg)
52
![Page 53: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/53.jpg)
53
![Page 54: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/54.jpg)
54
![Page 55: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/55.jpg)
55
![Page 56: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/56.jpg)
Detective Quantum EfficiencyDQE
• An indicator of the potential “speed class” or dose level required to acquire an optimal image.
• The DQE performance is obtained by
comparing the image noise of a detector
with that expected for an “ideal” detector
having the same signal-response characteristics.
56
![Page 57: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/57.jpg)
57
![Page 58: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/58.jpg)
58
![Page 59: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/59.jpg)
59
![Page 60: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/60.jpg)
60
![Page 61: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/61.jpg)
Exposure Latitude
• It is the optimal exposure range
• relative to the “ideal” exposure
• that produces a quality image at
• an appropriate patient dose.
61
![Page 62: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/62.jpg)
Why do digital systems havesignificantly greater latitude?
• Linear response give the imaging plates greater latitude
• Area receiving little radiation can be enhanced by the computer
• Higher densities can be separated and brought down to the visible density ranges
62
![Page 63: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/63.jpg)
Exposure Latitude
• The analog receptor exposure latitude ranges from approximately
• 30% underexposed
• to 50% overexposed relative to
• the “ideal” exposure level.
63
![Page 64: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/64.jpg)
Exposure Latitude
The digital image receptor
• exposure latitude ranges from
• approximately
• 50% underexposed
• to 100% over exposure
• relative to the “ideal” exposure level.
64
![Page 65: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/65.jpg)
65
![Page 66: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/66.jpg)
Exposure Indicators
• Imaging plates get a signal from the exposure they receive
• The value of the signal is calculated from the region identified as the anatomy of interest
• The signal for the plate is an average of all signals given to the plate
![Page 67: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/67.jpg)
NoteIt is important to note that just because a• digital imaging system has the capacity to• produce an image from gross underexposure• or gross overexposure it does not equate to• greater exposure latitude. • The reason the system is capable of producing
an image when significant exposure errors occur is through a process called automatic rescaling.
67
![Page 68: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/68.jpg)
• In a digital system, underexposure of
• 50% or greater will result in a mottled
• image.
• In a digital system, overexposure
• greater than 200% of the ideal will result
• in loss of image contrast.
68
![Page 69: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/69.jpg)
Histogram showing pixel values in an image. The pixel values in gray are on the horizontal with the total number for each on the vertical.
DarkerLighter
![Page 70: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/70.jpg)
Screen / Film Imaging = self regulating
2 mAsUnder exposed
6 mAsCorrect exposure
24 mAsOver exposed
70
![Page 71: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/71.jpg)
CR imaging results in 10,000 shades of gray
Fixed kVp exposures
mAs = 0.5S = 357
mAs = 1.0 S = 175
mAs = 2.0S = 86
mAs = 5.0S = 35 71
![Page 72: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/72.jpg)
LUT
• Look Up Table (LUT)
• Each anatomic area has a LUT
• Used to adjust contrast and density
• Other terms that may be used for this– Contrast rescaling– Contrast processing– Gradation processing– Tone scaling
![Page 73: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/73.jpg)
LUT
• The image data from the histogram is rescaled for application of the LUT
• The LUT maps the adjusted data through a “S” curve that is similar to an H & D curve
• The result is an image that has the correct contrast and brightness (density)
![Page 74: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/74.jpg)
Characteristic curve & Characteristic curve & histogramhistogram
![Page 75: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/75.jpg)
UnderexposedUnderexposed
![Page 76: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/76.jpg)
OverexposedOverexposed
![Page 77: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/77.jpg)
Just right!Just right!
![Page 78: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/78.jpg)
LOOK UP TABLE (LUT)
Linear LUT
Black Shirt
Facial Tones
Black Saturation
White Saturation
* No Detail in Black Areas* High Contrast* Only Detail in White Areas can be seen
* No Detail in White Areas* Low Contrast* Only Detail in Black Areas can be seen
![Page 79: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/79.jpg)
Digital Images – Bit Depth
• Pixel values can be any bit depth (values from 0 to 1023)
• Bit depth = # or gray shades available for image display
• Image contrast can be manipulated to stretched or contracted to alter the displayed contrast.
• Typically use “window width” and “window level” to alter displayed contrast and brightness
79
![Page 80: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/80.jpg)
Display Bit Depth1 bit 6 bit 8 bit
2 shades 64 shades 256 shades
80
![Page 81: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/81.jpg)
BIT Depth
• The number of gray shades available for image display.
Number of gray shades is 2n .
• Where n is the number of bits available for each pixel
81
![Page 82: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/82.jpg)
Digital - GrayscaleBit depth.
Number of gray shades available for display
• 8 bit 256
• 10 bit 1024
• 12 bit 4096
• 14 bit 16384
82
![Page 83: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/83.jpg)
83
![Page 84: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/84.jpg)
84
![Page 85: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/85.jpg)
85
![Page 86: Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15 References: Bushong Vol 9 1.](https://reader035.fdocuments.us/reader035/viewer/2022062305/56649f2a5503460f94c440b7/html5/thumbnails/86.jpg)
CR Image QualityCR Image QualityPixels, Field of View, Image receptorSampling frequency, Quantization,
Nyquest frequencyNoiseMagnificationImage compression (lossless vs
lossee)
Questions?
86