Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center...

16
Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia

Transcript of Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center...

Page 1: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Results of IAEA supported TPS audit in

Europe

Eduard Gershkevitsh

North Estonia Medical Center

Tallinn, Estonia

Page 2: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

• TPS audit uses IAEA TECDOC 1583 methodology

• IAEA provides CIRS Thorax phantom on loan for 6 month to the Member State

• IAEA together with national audit coordinator (Medical Physics group, RT department nominated by the national authorities) organise one day workshop in the country

• National audit coordinator’s centre is audited by independent auditor

• National audit coordinator is performing audits through on site visits

Principles of operation

Page 3: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

IAEA TECDOC 1583 methodology

• Based on anthropomorphic phantom

• To verify that logistic chain: CT scanning

Treatment planning

Data transfer

Dose delivery

is operational and leads to desired results with sufficient accuracy

•Employs ionisation chamber measurements

Page 4: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

IAEA TECDOC 1583 methodology

• Eight test cases with 15 measurement points Single 10x10cm2 field at nominal SSD

Tangential field with wedge

Corner blocks

4-field “box”

Customised blocking

Oblique incidence with L-shape block

Half fields with wedges

Non-coplanar field arrangement

•Agreement criteria 2-5% depending on complexity

Page 5: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

• B. Petrovic – Institute of Oncology Vojvodina, Sremska Kamenica, Serbia

• C. Pesznyak – National Institute of Oncology, Budapest, Hungary

• K. Chelminski – M. Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw, Poland

• J. Grezdo – St. Elizabeth Institute of Oncology, Bratislava, Slovakia

• M. do Carmo Lopes – Portuguese Institute of Oncology, Coimbra, Portugal

• E. Gershkevitsh – North Estonia Medical Centre, Tallinn, Estonia

National TPS audit coordinators

Page 6: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Participants of the TPS Audit

• 8 countries

• 61 centres

• 195 datasets (combination of algorithms and beam quality)

Page 7: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

• CT to RED conversion curve required adjustment in 2/3 of centres (criteria used adopted from IAEA TRS 430 – for the same electron density the variation should not exceed ± 20HU for all materials except water (± 5HU)

CT to RED conversion

Page 8: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

• Discrepancies requiring intervention and not related to algorithm limitations were found in approximately 9% of datasets

Reasons for deviations

Calibration39%

Input beam data and model fitting

50%

Mech. problems 11%

Dosimetry problems

Page 9: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Calibration

• Use of chamber with outdated calibration factor

• Use of plastic phantom instead of water for calibration

• Incorrect value in TPS

Page 10: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Input beam data & model fitting

• Typographic errors

• Use of standard data

• Quality of measurement data

• Sub-optimal beam fitting

Page 11: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

0

5

10

15

20

25

30

35

40

1 2 3 4 5 6 7 8 9 10 11 12

Centres

% o

f m

ea

sure

me

nt

po

ints

exc

ee

din

g

ag

ree

me

nt

crite

ria

4-6MV Varian HE

10-18MV Varian HE

Problems with model beam fitting

Page 12: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Problems with treatment unit

• New couch top was installed and unaccounted. This lead to 8% underdose from posterior fields

• Loose mechanical wedge

Page 13: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Accuracy achievable

• With corrected data and advanced algorithms the majority of the measurements are within agreement criteria

Page 14: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Conclusion

• Input beam data and suboptimal beam modelling were the largest contributors to observed deviations

• CT to RED conversion is customised in minority of centres

• The majority of observed deviations have been corrected

• Contribution to better understanding of TPS performance and its limitations

Page 15: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Acknowledgement

• Stanislav Vatnitsky for drafting a proposal for TPS audit

• Joanna Izewska for support and implementation of the TPS audit at IAEA

• To national audit coordinators

• To medical physicist at audited RT departments

Page 16: Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia.

Thank you for attention!

More details: Gershkevitsh E, Pesznyak C, Petrovic B, Grezdo J, Chelminski K, do Carmo Lopes M, Izewska J, Van Dyk J. Dosimetric inter-institutional comparison in European radiotherapy centres: Results of IAEA supported treatment planning system audit. Acta Oncol. 2014 May;53(5):628-36.