RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC CHARACTERIZATION OF BLOOD COAGULATION CPOCT 25TH INTERNATIONAL SYMPOSIUM SEPTEMBER 2014 Reza Abdolvand Department of Elec. Eng. & Comp. Science University of Central Florida

Transcript of RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Page 1: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC CHARACTERIZATION OF BLOOD

COAGULATION

CPOCT 25TH INTERNATIONAL SYMPOSIUM

SEPTEMBER 2014

Reza AbdolvandDepartment of Elec. Eng. & Comp. ScienceUniversity of Central Florida

Page 2: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Blood Coagulation: Background2

Since Early 1900 blood clotting disorders have been detected by blood coagulation tests.

More common tests include: Prothrombin Time (PT) test: measures the tissue factor-

induced clotting time of plasma. Thrombin Time test (TT): measures the conversion rate of

fibrinogen to insoluble fibrin Activated Partial Thromboplastin Time (aPTT) Activated Clotting Time (ACT)

Page 3: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Coagulation Test: Earlier Techniques3

Initially, based on visual detection of clot formation

Later, optical and mechanical detections replaced the visual detection

Harris, Leanne F., et al. "Coagulation monitoring devices: Past, present, and future at the point of care." TrAC Trends in Analytical Chemistry 50 (2013): 85-95.

Page 4: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

PoC Coagulation Test4

The drive: Faster result Very small blood sample size (less painful) Potential for patient self-test (more data

points)

Mainly categorized in two classes based on the measured property: Mechanical: measuring change in viscosity or

flow (e.g. microINR, iLine Microsystems) Relatively complicated measurement technique

Electro-chemical: measuring change in impedance (e.g. INRatio, Alere) Prone to error resulted from interfering

properties

CoaguCheck XS from Roche

Page 5: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

MEMS for Coagulation Test5

Motivation: Directly measuring mechanical properties at very small scales and lower lost

Pioneered by Microvisk Company

Page 6: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Basics of Resonant Sensing6

Commonly liquid placed in direct contact with the resonant body

Resulting change is observed Wave damping Frequency

Excessive damping of acoustic waves: shear mode preferred

Measuring highly viscous liquids gives near zero output

Electrical interference

Is a contact-free method viable?

Quartz Shear mode liquid sensor*

Microbalance principle*** Sensor Review. Volume 28 · Number 1 · 2008 · 68–73

** http://en.wikipedia.org/wiki/File:QCM_principle.gif2

Page 7: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Thin Film-Piezoelectric-on-Substrate Resonators

Piezoelectric transduction Low loss substrate (e.g. Si) Lithographically-defined

frequency Advantages:

Higher Q relative to piezoelectric resonators

Lower Rm relative to capacitive resonator

Higher power handling relative to both

7

Page 8: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Fabrication

6 mask process Mo patterned by dry etching in SF6 & O2 plasma AlN is etched by wet and dry etching in Cl2 plasma Silicon is etched in DRIE (Topside and Backside) Devices are released in BOE

Top View Side View

Gold

Mo

AlNSi

SiO2

8

Page 9: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Anchor/Tether Loss9

Finite displacement at tether/substrate boundary Pressure waves are radiated into the substrate

Tether Displacement Example (top view)

Resonant Body Suspension Anchors

3D acoustic propagation simulation

Page 10: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

In-Plane Acoustic Reflectors A circular trench in the substrate serves as a very efficient

reflector Location of the trench is expected to be critical Finite Element Modeling predicts the enhancement of Q for a

device with properly designed reflector

10

Page 11: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Experimental Validation

Devices with properly designed reflector show >%400 improvement in Q

Multiples devices of the same design are measured and Q’s are averaged

Tether length (μm)No acoustic reflector

Reflector distance to tether (μm) a

18 ( λ/4) 36 ( λ/2) 54 (3λ/4)

9 ( λ/8) 3492 1475 15531 4676

18 ( λ/4) 3247 5023 3788 4613

27 (3λ/8) 1818 3326 12042 1969a Lengths expressed in acoustic wavelength are given in parentheses.

11

Page 12: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Sensor Design12

Sample collection site Capillary-based microfluidics Electrode pads for I/O measurement Transmission signal measurement

3D schematic of sensor

Page 13: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Preliminary Data Q is monitored as the liquid is injected in the channel Change in the Q is a measure of acoustic impedance of

the filling

RF probes

Micro-syringe

Test wafer

13

Page 14: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Ultrasonic Blood Analyzer

Using the same device the change in Q is monitored for different fluids

The goal is to characterize blood properties based on its acoustic signature

14

Dynamic output response with PR drying

48004900500051005200530054005500560057005800

0 100 200 300

Quality Factor (Q

)

Time (minutes)

Quality factor dependence on sample liquid (different device)

Initial Q = 6100 (in air)

σ =32, N=6

σ =34, N=10

σ =35,N=2

0

100

200

300

400

500

600

700

1813 Water Blood

Dec

reas

e in

Q

Liquid Filler

Device Response to Fillers

Page 15: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Transmitted Signal15

Transmission response is measured for liquid with variable glycol in water concentration

Transmission response increases with increased glycol concentrations

Q decreases (∆Q= 550) with increased glycol concentration

Transmission signal collected with filled channel

Transmission signal and Q for different glycol solutions

Page 16: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Simple Capping Techniques16

Polymer based capping of the channels and device is examined

Bonded PDMS films and deposited paryline films yield inexpensive covers

Polymer channel covering Polymer device cover

Page 17: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Final Test: PT monitoring17

IntegSense Inc. recently received funding to do a feasibility study on using this platform for PT monitoring

Results are due in a few month

Page 18: RESONANT MICRO-SENSING PLATFORM FOR ULTRASONIC ...

Acknowledgement18

I would like to acknowledge my former group members whose contribution was presented in this talk: Brandon Harington (now with Knowles)

Mohsen Shahmohammadi (now with Qualtre)

Jonathan Gonzales (now with IntegSense)

Thank you for your time!