Requirements on clinical data in Europe - TÜV SÜD · Requirements on clinical data ... Depending...
Transcript of Requirements on clinical data in Europe - TÜV SÜD · Requirements on clinical data ... Depending...
Requirements on clinical data in Europe Dr. Bassil Akra Director Global Clinical Affairs
TÜV SÜD Product Service
Current Medical Device Directives
TÜV SÜD Product Service Slide 2
Applicable Directives • Active Implantable Medical Devices AIMDD 90/385/EEC accepted: 20.06.90; adopted: 01.01.93; end of transition: 31.12.94 • Medical Devices MDD 93/42/EEC accepted: 14.06.93; adopted: 01.01.95; end of transition: 13.06.98 • In Vitro Diagnostics IVDMDD 98/79/EC accepted: 27.10.98; published: 07.12.98; adopted: 07.06.2000; end of transition: 07.12.2003
Where are we heading to ?
MDD AIMD IVDD
MEDDEV 2.7.1
2007/47EG
Rec. 2013/473
920/2013
TÜV SÜD Product Service Slide 3
Current Requirements for Clinical Data
TÜV SÜD Product Service Slide 4
Dir. 2007/47/EC Which medical devices require a clinical evaluation? ALL Medical Devices regardless of Classification Annex I: Demonstration of conformity with the essential requirements must include a
clinical evaluation in accordance with Annex X
One focus of the new regulations will be clinical evaluation
Clinical Evaluation according to current MEDDEV 2.7.1 Rev. 3
TÜV SÜD Product Service Slide 5
Clinical Evaluation The assessment and analysis of clinical data pertaining to a medical device to verify the clinical safety and performance of the device when used as intended by the manufacturer.
EU Regulations Proposal – Clinical Evaluation
TÜV SÜD Product Service Slide 6
Article 4: Placing on the market and putting into service A device shall meet the general safety and performance requirements which
apply to it, taking into account its intended purpose.
Demonstration of conformity with the general safety and performance requirements shall include a clinical evaluation in accordance with Artilce 49.
Current Requirements for Clinical Data
TÜV SÜD Product Service Slide 7
Dir. 2007/47/EC - CLINICAL DATA The safety and/or performance information that is generated from the use of a device
SOURCE OF CLINICAL DATA
Studies reported in scientific literature of a similar device (EQUIVALENCY)
Published and/or unpublished reports
Clinical investigation(s) of the device concerned
EU Regulations Proposal – Clinical Evaluation
TÜV SÜD Product Service Slide 8
Article 49: Clinical Evaluation Manufacturers shall conduct a clinical evaluation in accordance with the principles
set out in this Article and Part A of Annex XIII.
A clinical evaluation shall follow a defined and methodologically sound procedure based on either of the following:
A critical evaluation of the relevant scientific literature A critical evaluation of the results of all clinical investigations A critical evaluation of the combined clinical data
Equivalence Approach based on current available literature
TÜV SÜD Product Service Slide 9
Same intended use
(Clinical condition/Disease, Severity, Application Site, Patient Population, Critical Performance Parameter) +
Technical and biological equivalence +
No clinically significant difference regarding safety and performance
Clinical Evaluation – Example 1
TÜV SÜD Product Service http://www.orthopaedicsurgeon.com.sg/total-hip-replacement/ Slide 10
Which route should you follow for the device under consideration?
Example of available competitors New Device
Clinical Evaluation – Example 2
TÜV SÜD Product Service Slide 11
Which route should you follow for the device under consideration?
Example of available competitors New Device
A
B http://www.exac.com/patients-caregivers/joint-replacement-surgery/hip-replacement/components-hip-replacement http://evertsmith.com/innovations/
Clinical Evaluation – RISK MANAGEMENT PROCESS
TÜV SÜD Product Service Slide 12
Which route should you follow for the device under consideration?
IDEA
RISK IDENTIFICATION METHODS
RISK MITIGATION METHODS
RESIDUAL RISKS ACCEPTABILITY
Clinical Evaluation according to MEDDEV 2.7.1 Rev. 3
TÜV SÜD Product Service Slide 13
Clinical Evaluation Report
Stage 3Analysis of
relevant data
Stage 2Appraisal of
individual data sets
Stage 1Identification of
Clinical Data
Sufficient?
Generateadditional clinical
Data
NO YES
Determination ofScope
Clinical Investigation Route
TÜV SÜD Product Service Slide 14
MEDDEV 2.7/4 – When should a clinical investigation be undertaken? Decision Tree General Classification Rules • Class III
• Implants, Most Cardiovascular Devices • Class IIb
• Energy/Substance Delivering, Ionizing Radiation Emitting, Long Term Surgically Invasive Devices
• Class IIa • Most Active & Short Term Surgically Invasive
• Class I • Wheelchairs, Surgical Instruments
Current Clinical Investigation Route Requirements
TÜV SÜD Product Service Slide 15
MEDDEV 2.7/4 – When should a clinical investigation be undertaken? The Conformity Assessment process for active implantable medical devices as well as
for class III and implantable medical devices requires that a clinical investigation is undertaken unless it is duly justified to rely on existing data. Section 1.2 of Annex 7 of directive 90/385/EEC Section I.1a of Annex X of directive 93/42/EEC
Current Clinical Investigation Route Requirements
TÜV SÜD Product Service Slide 16
MEDDEV 2.7/4 – When should a clinical investigation be undertaken? Depending on clinical claims, risk management outcome and on the results of the
clinical evaluation, clinical investigations may also have to be performed for non-implantable medical devices of classes I, IIa and IIb.
Additional clinical investigations may be feasible to corroborate the existing clinical evidence with regard to aspects of clinical performance, safety, benefit/risk-ratio or to determine relative effectiveness and safety with suitable comparators.
EU Regulations Proposal – Clinical Investigations Requirements
TÜV SÜD Product Service Slide 17
Preamble (46) in the proposed regulation Clinical data of Class III medical devices and
Implantable medical devices should be, as a general rule, based on clinical investigations to be carried out under the responsibility of a sponsor.
Preamble (47) in the proposed regulation Clinical investigations in line with major
international guidance in this field, such as the international standard ISO 14155:2011.
Clinical Investigation Route
TÜV SÜD Product Service Slide 18
Clinical Investigation
Sponsor
Investigation site
Qualified investigators
CIB Annex B
normative
CIP Annex A
normative
Informed consent form
Insurance Favorable opinion of
ethics committee
Approval of Competent Authority
EU Regulations Proposal - Clinical Investigations Requirements
TÜV SÜD Product Service Slide 19
Preamble (48) in the proposed regulation An electronic system should be set up at Union level to ensure that every clinical investigation is registered in a publicly accessible database.
No personal data of subjects participating in a clinical investigation should be recorded in the electronic system.
Preamble (51) in the proposed regulation This Regulation should only cover clinical investigations which pursue regulatory purposes laid down in this Regulation.
Already there
TÜV SÜD Product Service Slide 20
Three Annexes: I. Product Assessment II. Quality System Assessment III. Unannounced Audit
Annex I – Product assessment
TÜV SÜD Product Service Slide 21
Notified Bodies should: Review All relevant preclinical data Clinical evaluation Post-market clinical follow-up undertaken or planned
Verify the clinical evaluation is up-to-date Assess the need for and appropriateness of a post- market clinical follow-up plan
April 2014
EU Regulations Proposal - Resource Requirements
TÜV SÜD Product Service Slide 22
Manufacturers have to perform qualified clinical evaluations
Designating authorities have to employ qualified staff
Notified Bodies have to employ qualified staff
EU Regulations Proposal - Resource Requirements
TÜV SÜD Product Service Slide 23
At all times and for each conformity assessment procedure and each kind or category of products in relation to which it has been notified, a notified body shall have within its organization the necessary administrative, technical and scientific personnel with technical knowledge and sufficient appropriate experience relating to medical devices and the corresponding technologies to perform the conformity assessment tasks, including the assessment of clinical data.
EU Regulations Proposal - Resource Requirements
TÜV SÜD Product Service Slide 24 TÜV SÜD Product Service Slide 24
Radiology Cardiology
Orthopedic Surgery
Neurology
Add value by
know-how
General Medicine
Computer Science
Electrical Engineering
Chemical Engineering
Biological Science
Mechanical Engineering
Medical Science
Environmental Science
EU Regulations Proposal - Resource Requirements
TÜV SÜD Product Service Slide 25 New Clinical Data Requirements in the EU
Notified bodies shall have available personnel with clinical expertise. This personnel shall be integrated in the notified body’s decision-making process in a steady way in order to: Identify when specialist input is required Appropriately train external clinical experts Be able to discuss and challenge the clinical data contained in the CER Be able to ascertain the comparability and consistency of the clinical Assessments conducted by clinical experts Be able to make an objective judgment
April 2014
PMCF Requirements according to current MEDDEV 2.12/2 Rev. 2
TÜV SÜD Product Service Slide 26
Why PMCF?
Rare complications or problems become apparent after wide-spread or long term use of the device
An appropriate PMS Plan is key to identifying and investigating residual risks associated with the use of medical devices placed on the market
To confirm the safety and performance throughout the expected lifetime of the device, the continued acceptability of identified risks and to detect emerging risks on the basis of factual evidence
PMCF Requirements according to current MEDDEV 2.12/2 Rev. 2
TÜV SÜD Product Service Slide 27
PMCF is mandatory per example, for: Innovative Products Significant changes High product related risks High risk anatomical locations/target populations Emergence of new information on safety or performance Where CE marking was based on equivalence Unanswered questions of long-term safety and performance
EU Regulations Proposal – PMCF Requirements
TÜV SÜD Product Service Slide 28
Why PMCF? To confirm the safety and performance throughout the expected lifetime of the device, the continued acceptability of identified risks and to detect emerging risks on the basis of factual evidence
Systematic registries for Class IIb and Class III devices
EU Regulations Proposal – Scrutiny Procedure
Proposal Commission
TÜV SÜD Product Service Slide 29
Proposed Scrutiny Procedure (Article 44, Page 66) Medical Devices Directives revision proposal
Manufacturer Submission
Notified Body
Notified Body Review
Notified Body Pre-report
Commission
Medical Devices Coordination Group MDCG T= 28 Days
Scrutiny? Scrutiny Start
MDCG Scrutiny
Notified Body Approval Board
T= 60 Days
MDCG final comment
CE
Consider
EU Regulations Proposal – Scrutiny Procedure
Proposal Parliament
TÜV SÜD Product Service Slide 30
Manufacturer Submission SNB SNB Evaluation
Commission
MDCG T= 20 Days
Scrutiny?
SNB Approval Board CE
If scrutiny
Evaluation of CER, PMCF
and other relevant doc
T= 30 Days*
*Clock Stop System
Opi
nion
°
°Appeal procedure
Follow
Summary: what to expect in the future?
TÜV SÜD Product Service Slide 31 New Clinical Data Requirements in the EU
Notified bodies
will
Involve external
clinical experts if deemed necessary
Discuss and challenge the clinical data contained in
the CER
Challenge the comparability
and consistency of
the clinical data
TÜV SÜD Product Service Slide 32 Lighting Services
Dr. Bassil Akra Director - Global Clinical Affairs For enquiries, email me at: [email protected]
Questions?
� TÜV SÜD TÜV SÜD Product Service
Global website: www.tuv-sud.com/medicaldevice Stay informed and updated with our Healthcare & Medical Device newsletter: www.tuv-sud.com/essentials
TÜV SÜD Product Service