Page 1: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3
Page 2: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 2 / 38

TABLE OF CONTENTS

1. GENERAL DESCRIPTION OF EUT 3

2. GENERAL INFORMATION OF TEST 4

3.1 CONDUCTED EMISSION TEST 7

3.2 RADIATED EMISSION TEST 11

3.3 ELECTROSTATIC DISCHARGE IMMUNITY TEST 14

3.4 RF ELECTROMAGNETIC FIELD IMMUNITY TEST 17

3.5 FAST TRANSIENTS IMMUNITY TEST 20

3.6 SURGES IMMUNITY TEST 22

3.7 RF COMMON MODE IMMUNITY TEST 24

3.8 VOLTAGE DIPS AND INTERRUPTIONS IMMUNITY TEST 26

3.9 POWER FREQUENCY MAGNETIC FIELD 28

3.10 POWER FREQUENCY HARMONICS AND FLICKER EMISSION TESTS 30

4. EUT PHOTOGRAPHS 34

5. DESCRIPTION OF EUT MODIFICATION 37

Page 3: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 3 / 38

1.1 Applicant

Company Name : HANMED Co., Ltd.

Address : #112 Gimhae Bio Medical Center, 155-1, Nogso-ri, Juchon-myeon, Gimhae-si, Gyeongsangnam-do, Korea

1.2 Manufacture

Company Name : HANMED Co., Ltd.

Address : #112 Gimhae Bio Medical Center, 155-1, Nogso-ri, Juchon-myeon, Gimhae-si, Gyeongsangnam-do, Korea

1.3 Basic Description of EUT

1. General Description of EUT

Product Name : Orthopedics appliance

Model Name : KNX-7000

Serial Number : Prototype

Input Rating : AC 100 ~ 240 V , 50/60 Hz

Page 4: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 4 / 38

2. General Information of Test 2.1 Test Facility

This test was carried out by SGS Tesco Korea. Test Site Location : 413-15, Gomae-dong, Giheung-gu, Yongin-si, Gyeonggi-do, Korea TEL : 82-31-8005-6020, FAX : 82-31-8005-6025

2.2 Standard for Methods of Measurement

Basic Standard Description Test Result

EN 55011:2007 CISPR 11:2003+A1:2004

Conducted Emission Pass Fail

Radiated Emission Pass Fail

EN61000-4-2 / IEC61000-4-2 Electrostatic Discharge Pass Fail

EN61000-4-3 / IEC61000-4-3 RF Electromagnetic field Pass Fail

EN61000-4-4 / IEC61000-4-4 Fast Transient/burst Pass Fail

EN61000-4-5 / IEC61000-4-5 Surge Pass Fail

EN61000-4-6 / IEC61000-4-6 RF common mode Pass Fail

EN61000-4-11 / IEC61000-4-11 Voltage Dips & Interruptions Pass Fail

EN61000-4-8 / IEC61000-4-8 Power-frequency Magnetic Field Pass Fail

EN 61000-3-2:2006 * Harmonic Current Emission Pass Fail

EN 61000-3-3:1995 +A1:2001+A2:2005 * Voltage Fluctuations and Flicker Pass Fail

* This test was performed by the LTA Co., Ltd., which is our sub-contracted Laboratory

Page 5: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 5 / 38

2.3 Description of EUT modification

Used ferrite core and noise filter (Refer to photograph of page 37 ~ 38)

2.4 Variations covered by this report N/A

2.5 Additional information related to Testing

Test results apply only to the particular sample tested and functionality described in this test report. This report may be reproduced in full. Partial reproduction may only be made with the written permission of the SGS Tesco Korea.

2.6 Test Conditions EUT Operating Mode EUT was tested according to the following operation modes provided by the specifications given by the manufacturer and reported the worst emissions.

Operation Modes Worst Case Mode

The EUT set up normal operating by program.

2.7 Immunity Method

Observation of display & operating condition.

Test System layout on EUT and peripherals

Power cable, Signal cable

AC 230V

EUT

EUT PC SYSTEM

Page 6: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 6 / 38

2.8 Description of Test System

Type of Peripheral Equipment Used:

Description Model Name Serial No. Manufacturer

PC N/A N/A N/A

Monitor MY19PS N960HVCQA12596F SAMSUNG ELECTRONICS

Keyboard TGK-7500 165856846000969 Goldland Electronics.,Ltd.

Mouse AN3050 N/A AIONTECH CO., LTD.

Type of Cables Used:

Device from Device to Type of Cable Length (m) Type of shield

EUT AC Power Power 1.8 Unshield

EUT EUT Signal 2.0 Unshield

Criteria description The following degradations associated with essential performance and safety shall not be allowed: - Component failures - Changes in programmable parameters - reset to factory defaults (manufacturer’s presets) - Change of operating mode - Fuse alarms - Cessation or interruption of any intended operation, even if accompanied by an alarm - Initiation of any unintended operation, including unintended or uncontrolled motion, even if accompanied by an alarm

- Error of a displayed numerical value sufficiently large to affect diagnosis or treatment - Noise on waveform in which in which the artefact would interfere with diagnosis, treatment or monitoring

- Failure of automatic diagnosis or treatment equipment and systems to diagnose or treat, even if accompanied by an alarm

Page 7: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 7 / 38

3.1 Conducted Emission Test Disturbance voltages at the mains terminal were measured from 150 kHz to 30 MHz with a bandwidth of 9 kHz on the 230 V AC power input terminal. The EUT was placed on a non-metallic stand in a shielded room 0.8 meters above the ground plane as shown in photograph of test setup. The interface cables and equipment positioning were varied within limits of reasonable applications to determine the position producing maximum conducted emissions.

3.1.1 Test Condition

Frequency Range of Test : 150 kHz to 30 MHz

Test Standard : EN 60601-1-2:2007

Test method : EN 55011:2007 Group 1, Class A

Test Date : August 10, 2009

Temperature/Humidity : (21 ± 1) °C / (43 ± 2) % R.H.

Input Voltage : AC 230 V / 50 Hz 3.1.2 Test Standard.

Frequency Range (MHz)

Limit dB() Quasi-Peak Average-Peak

0.15 ~ 0.5 79.0 66.0

0.5 ~ 30 73.0 60.0

3.1.3 Test Equipment List.

Equipment Type Model Manufacture Serial No Cal Due Date Use

2-LINE V-NETWORK ENV216 R&S 100415 2010.06.22

EMI TEST RECEIVER ESU8 R&S 100128 2010.01.19

PULSE LIMITER ESH3-Z2 R&S 100850 2009.09.14

Page 8: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 8 / 38

3.1.4 Test Result of Conducted Emission

Test Results: PASS Test data sheets follow.

Frequency C.F. Phase Quasi-Peak Mode Average Mode

Limit Result Margin Limit Result Margin

[MHz] [dB] [N/L1] dB[uV] dB[uV] dB[uV] dB[uV] dB[uV] dB[uV]

5.310 19.5 L1 73.0 59.0 14.0 60.0 36.0 24.0

5.574 19.6 N 73.0 58.1 14.9 60.0 36.3 23.7

6.378 19.6 L1 73.0 60.9 12.1 60.0 36.6 23.4

6.634 19.6 N 73.0 62.0 11.0 60.0 35.8 24.2

7.534 19.6 L1 73.0 61.5 11.5 60.0 38.4 21.6

Notes : 1. C.F = LISN Factor + Cable Loss + Pulse Limiter Loss / Result = Reading Level + C.F 2. All modes of operation were investigated and the worst-case emissions are reported. 3. Measurement uncertainty estimated at ± 3.76 dB. The measurement uncertainty is given with a confidence of 95.45 % with the coverage factor, k=2.

4. See next page for measurement graph.

Tested by Seung-bum Cho

Page 9: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 9 / 38

0

10

20

30

40

50

60

70

80

150k 300 400 500 800 1M 2M 3M 4M 5M 6 8 10M 20M 30M

Leve

l in

dB킮

Frequency in Hz

CISPR11 CLASS A Gruop 1 QP

CISPR11 CLASS A Gruop 1 AV

<N LINE>

0

10

20

30

40

50

60

70

80

150k 300 400 500 800 1M 2M 3M 4M 5M 6 8 10M 20M 30M

Leve

l in

dB킮

Frequency in Hz

CISPR11 CLASS A Gruop 1 QP

CISPR11 CLASS A Gruop 1 AV

<L1 LINE>

Page 10: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 10 / 38

3.1.5 Photographs of Conducted Emission Test Configuration

Page 11: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 11 / 38

3.2 Radiated Emission Test

Radiated disturbance from 30 MHz to 1 GHz were measured with a bandwidth of 120 kHz for both vertical and horizontal polarization of the measuring antenna according to the methods defines in EN 60601-1-2:2007. The EUT was placed on a non-metallic table in the semi-anechoic chamber, 0.8 meter above the ground plane, and the measurement distance from the measuring antenna to the center of the EUT was set to 10m distance as shown in Test setup photograph. The measuring antenna was adjusted in height of 1m to 4m range for the maximum emission. The interface cables and equipment positions were varied within limits of reasonable applications to determine the positions producing maximum radiated emissions.

3.2.1 Test Condition

Frequency Range of Test : 30 MHz to 1 GHz Test Standard : EN 60601-1-2:2007

Test Method EN 55011:2007 Group 1, Class A Test Date : August 10, 2009

Temperature/Humidity : (29 ± 1) °C / (52 ± 2) % R.H. 3.2.2 Test Standard

Frequency Range (MHz)

Limit Quasi-Peak dB(/m)

30 ~ 230 40

230 ~ 1 000 47 3.2.3 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

EMI TEST RECEIVER ESU8 R&S 100128 2010.01.19

Antenna Mast ANTENNA MAST AM4.0 MATURO GMBH N/A N/A

Antenna Turntable Controller

Turntable TT1.2 SI-0 MATURO GMBH N/A N/A

BILOG ANT CBL6112D TESEQ 25233 2010.06.02

Preamplifier AM-1431 MITEQ 1336160 2010.07.08

Page 12: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 12 / 38

3.2.4 Test Result of Radiated Emission

Test Results : PASS Test data sheets follow.

Frequency (MHz)

Quasi-PeakdB(uV/m)

Antenna height (cm)

Polarity Turntable position

(deg)

Corr. (dB)

Margin (dB)

Limit dB(uV/m)

42.12 33.7 180 V 200 -24.4 6.3 40.0

67.83 32.1 120 V 230 -29.1 7.9 40.0

163.37 35.4 100 H 60 -24.2 4.6 40.0

232.73 33.9 150 V 170 -21.5 13.1 47.0

450.98 39.7 130 H 300 -15.1 7.3 47.0

471.83 41.1 220 H 255 -14.6 5.9 47.0

511.60 41.2 250 H 310 -13.8 5.8 47.0 Notes 1. H : Horizontal polarization , V : Vertical polarization 2. QuasiPeak Level = Reading + Corr.(Antenna factor + Cable loss – PreAmplifier) 3. Margin value = Limit – QuasiPeak Level 4. Measurement uncertainty estimated at ± 4.86 dB.

The measurement uncertainty is given with a confidence of 95.45 % with the coverage factor, k = 2. 5. The result was passed but within measurement uncertainty

Antenna : Horizontal / Vertical

Distance : 10 m

Tested by Seung-bum Cho

Page 13: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 13 / 38

3.2.5 Photographs of Radiated Emission Test Configuration

Page 14: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 14 / 38

3.3 Electrostatic Discharge Immunity Test Enclosure Port

The EUT was set-up on insulators 0.5 mm above the Horizontal Coupling Plane and tested in accordance with EN 61000-4-2.

Air Discharge

A potential of 2 kV was initially applied to each applicable test point in both polarities, then verity level was increased to 4 kV and then to 8 kV. The full discharge level was applied a total of 20 times to each test point, 10 positive discharges and 10 negative discharges. The interval between discharges was a minimum of 1 second.

Contact Discharge

A potential of 2 kV was initially applied to all user accessible conductive surfaces. The level was increased to 6 kV. The full discharge level of 6 kV was then applied to each test point, 25 positive discharges and 25 negative discharges. Each vertical side of the EUT was subjected to Vertical Coupling Plane (VCP) discharges of 6 kV, 25 positive discharges and 25 negative discharges. The base of EUT was subjected to Horizontal Coupling Plane (HCP) discharges of 6 kV, 25 positive discharges and 25 negative discharges. Correct functioning of the EUT was checked throughout this test.

3.3.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN61000-4-2:1995+A1:1998+A2:2001

Test Date : August 11, 2009

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (24 ± 1) °C / (41 ± 2) % R.H./ 100.2 kPa 3.3.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

ESD SIMULATOR ESS-2000 NOISEKEN ESS0422543 2010.06.22

3.3.3 Test Result of Electrostatic Discharge Immunity Test Electrostatic Discharge Immunity Test : PASS

The equipment meets the requirements.

Page 15: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 15 / 38

Test data sheet follows.

Test Point Discharge

Type CD=Contact

AD=Air

Test Voltage(kV) Tested No Observation Result

VCP CD ±2,4,6 kV Each 25 times Normal PASS Screw CD ±2,4,6 kV Each 25 times Normal PASS

Enclosure CD ±2,4,6 kV Each 25 times Normal PASS Seat AD ±2,4,8 kV Each 10 times Normal PASS

Cable AD ±2,4,8 kV Each 10 times Normal PASS LED AD ±2,4,8 kV Each 10 times Normal PASS

[Test Point] AD / CD

Remarks The primary functions as described below were fully functional during and after test. 1. There is no deterioration on the EUT

Tested by Seung-bum Cho

Page 16: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 16 / 38

3.3.4 Photographs of Electrostatic Discharge Immunity Test Configuration

Page 17: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 17 / 38

3.4 RF Electromagnetic Field Immunity Test Tests were conducted in accordance with EN 60601-1-2:2007 over the frequency range of 80 MHz to 2.5 GHz. The transmitting antenna was located 3 meters(or 1.8 meter) from the EUT at a height of 1 meter above the floor. Sides of the EUT were exposed to a uniform field of 3 V/m using both horizontal and vertical antenna polarizations. 3.4.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN 61000-4-3: 2002+A1:2002

Test Date : August 12, 2009

Test field strength (V/m) 3 V/m

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (20 ± 1) °C / (46 ± 2) % R.H./ 100.3 kPa 3.4.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

AMPLIFIER 250W1000A AR 0327278 N/A

AMPLIFIER 60S1G3 AR 0327239 N/A

POWER METER NRVD R&S 101297 2009.11.13

SIGNAL GENERATOR SMR40 R&S 100540 2010.01.19

HIGH GAIN LOG-PERIODIC ANTENNA HL046E R&S 100043 N/A

10V INSERTION UNIT URV5-Z2 R&S 100346 2009.11.13

10V INSERTION UNIT URV5-Z2 R&S 100347 2009.11.13

FIELD PROBE HI-6005 ETS-LINDGREN 0086968 2010.01.31

3.4.3 Test Result of RF Electromagnetic Field immunity test Electromagnetic field immunity test : PASS

The equipment meets the requirements.

Page 18: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 18 / 38

Test data sheet follows.

Frequency Range (MHz)

Position (Angle)

Antenna Polarity

Field Strength

(V/m) Modulation Result

80~2 500 Front Vertical 3 80 % AM(1 kHz) PASS

80~2 500 Rear Vertical 3 80 % AM(1 kHz) PASS

80~2 500 Right Vertical 3 80 % AM(1 kHz) PASS

80~2 500 Left Vertical 3 80 % AM(1 kHz) PASS

80~2 500 Front Horizontal 3 80 % AM(1 kHz) PASS

80~2 500 Rear Horizontal 3 80 % AM(1 kHz) PASS

80~2 500 Right Horizontal 3 80 % AM(1 kHz) PASS

80~2 500 Left Horizontal 3 80 % AM(1 kHz) PASS

Remarks The primary functions as described below were fully functional during and after test. 1. There is no deterioration on the EUT

Tested by Seung-bum Cho

Page 19: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 19 / 38

3.4.4 Photographs of Electrostatic Discharge Immunity Test Configuration

Page 20: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 20 / 38

3.5 Fast Transients Immunity Test Tests were conducted in accordance with EN 60601-1-2:2007. Tests were performed to 2 kV to AC Power lines. 3.5.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN 61000-4-4:2004

Test Date : August 11, 2009

Test Voltage : ± 2 kV

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (24 ± 1) °C / (41 ± 2) % R.H./ 100.2 kPa 3.5.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

Combinded test system ECOMPACT 4 Haefely 170328 2010.06.18

3.5.3 Test Result of Electrical fast transient/burst immunity test

Electrical fast transient/burst immunity test : PASS

The equipment meets the requirements.

Page 21: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 21 / 38

Test data sheet follows.

Line Voltage Coupling Inject Time (sec) Result

N ± 2 kV Direct 60 PASS

L1 ± 2 kV Direct 60 PASS

PE ± 2 kV Direct 60 PASS

N,L1 ± 2 kV Direct 60 PASS

L1,PE ± 2 kV Direct 60 PASS

N,PE ± 2 kV Direct 60 PASS

L1,N,PE ± 2 kV Direct 60 PASS

Remarks The primary functions as described below were fully functional during and after test.

1. There is no deterioration on the EUT

3.5.4 Photographs of Fast Transient Immunity Test Configuration

Tested by Seung-bum Cho

Page 22: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 22 / 38

3.6 Surges Immunity Test Tests were conducted in accordance with EN 60601-1-2:2007. Tests were performed to 1 kV to AC Line to Line, 2 kV to AC Line to PE. 3.6.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN61000-4-5:2006

Test Date : August 11, 2009

Test Voltage : ±1 kV(Line to Line), ± 2 kV(Line to PE)

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (24 ± 1) °C / (41 ± 2) % R.H./ 100.2 kPa 3.6.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

Combinded test system ECOMPACT 4 Haefely 170328 2010.06.18

3.6.3 Test Result of Surge immunity test

Surge immunity test : PASS

The equipment meets the requirements.

Page 23: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 23 / 38

Test data sheet follows.

Line Voltage Coupling Phase (Degree)

Inject (count - min) Result

L1-N ± 1 kV 2 Ω, 18 0,90,180,270 5 times - 5 min Pass

L1-PE ± 2 kV 12 Ω, 9 0,90,180,270 5 times - 5 min Pass

N-PE ± 2 kV 12 Ω, 9 0,90,180,270 5 times - 5 min Pass

Remarks The primary functions as described below were fully functional during and after test.

1. There is no deterioration on the EUT 3.6.4 Photographs of Surges Immunity Test Configuration

Tested by Seung-bum Cho

Page 24: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 24 / 38

3.7 RF Common Mode Immunity Test Tests were conducted in accordance with EN 60601-1-2:2007. Over the frequency range of 150 kHz to 80 MHz. Tests were performed at 3 V on power line. 3.7.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN 61000-4-6:1996+A1:2001

Test Date : August 11, 2009

Test Voltage : 3 V

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (24 ± 1) °C / (41 ± 2) % R.H./ 100.2 kPa

3.7.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

CONTINUOUS WAVE SIMULATOR CWS 500 C S1 EM TEST 0744103074 2010.04.22

CDN M2/M3 EM TEST 1107-59 2010.05.14

Page 25: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 25 / 38

3.7.3 Test Result of RF Common Mode immunity test Tes data sheet follows.

Frequency (MHz) Coupling Test

voltage Dwell time Modulation Step Size Result

0.15 ~ 80 CDN(M3) 3 V 3 s 80 % AM @ 1 kHz 1.0 % Pass

Remarks The primary functions as described below were fully functional during and after test.

1. There is no deterioration on the EUT 3.7.4 Photographs of RF Common Mode immunity test Configuration

Conducted immunity test : PASS The equipment meets the requirements.

Tested by Seung-bum Cho

Page 26: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 26 / 38

3.8 Voltage Dips and Interruptions Immunity Test Voltage Dips and Interruptions tests were conducted in accordance with EN 60601-1-2:2007. 3.8.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN 61000-4-11:2004

Test Date : August 11, 2009

Test Voltage : AC 120 V , AC 230 V

Voltage Dip and Voltage interruptions : 100 % (0.5 cycle), 60 % (5 cycle), 30 % (25 cycle)

100 % (250 cycle)

Performance criterion :The EUT remains safe, experiences no component failures ant is restorable to the pre-test state with operator intervention.

Temperature/Humidity/Pressure : (23 ± 1) °C / (40 ± 2) % R.H./ 100.2 kPa

3.8.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

Combinded test system ECOMPACT 4 Haefely 170328 2010.06.18

3.8.3 Test Result of Voltage Dips and Voltage interruptions immunity test

Voltage Dips and Voltage interruptions immunity test : PASS

The equipment meets the requirements.

Page 27: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 27 / 38

Test data sheet follows.

Residual Voltage (%) Cycle Angle ( ° ) Result

0 0.5 0,180 Pass

40 5 0,180 Pass

70 25 0,180 Pass

0 250 0,180 Pass(Note 1)

Remarks The primary functions as described below were fully functional during and after test. Note 1. The EUT turn on/off during test but after test there have been no deterioration of EUT.

3.8.4 Photographs of Voltage Dips and Interruptions Immunity Test Configuration

Tested by Seung-bum Cho

Page 28: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 28 / 38

3.9 Power Frequency Magnetic Field Tests were conducted in accordance with EN 61000-4-8. 3.9.1 Test Condition

Test Standard : EN 60601-1-2:2007

Test Method : EN 61000-4-8:1993+A1:2001

Test Date : August 11, 2009

Test Voltage : 3 A/m

Performance criterion : Refer to Criteria description of page 6

Temperature/Humidity/Pressure : (23 ± 1) °C / (40 ± 2) % R.H./ 100.2 kPa

3.9.2 Test Equipment List

Equipment Type Model Manufacture Serial No Cal Due Date Use

Magnetic Field Coil MFS100 HAEFELY 1684 2009.09.09

3.9.3 Test Result of Power Frequency Magnetic Field immunity test

Conducted immunity test : PASS The equipment meets the requirements.

Page 29: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 29 / 38

Test data sheet follows.

Test Current Frequency Axis Test Time Result

3 A/m 50Hz X, Y, Z 60 s Pass

3 A/m 60Hz X, Y, Z 60 s Pass

Remarks The primary functions as described below were fully functional during and after test.

1. There is no deterioration on the EUT

3.9.4 Photographs of Power Frequency Magnetic Field Configuration

Tested by Seung-bum Cho

Page 30: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 30 / 38

3.10 Power Frequency Harmonics and Flicker Emission Tests 3.10.1 Test Procedure

Flicker Emission Tests : The total impedance of the test circuit, excluding the appliance under test, but including the internal impedance of the supply source, shall be equal to the reference impedance. The stability and tolerance of the reference impedance shall be adequate to ensure that the overall accuracy of ±8% is achieved during the whole assessment procedure.

3.10.2 Test Condition

Test Standard : EN61000-3-2:2006 EN61000-3-3:1995+A1:2001+A2:2005

Test Date : August 13, 2009

Temperature/Humidity : (23 ± 1) °C / (42 ± 2) % R.H. 3.10.3 Test Result of Harmonics and Flicker Emission Tests

Power Frequency Harmonics : PASS

Flicker Emission Tests : PASS

Page 31: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 31 / 38

Page 32: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 32 / 38

Page 33: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 33 / 38

Page 34: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 34 / 38

4. EUT PHOTOGRAPHS

Page 35: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 35 / 38

Page 36: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 36 / 38

Page 37: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 37 / 38

5. Description of EUT modification

- Ferrite Core(Manufacture : TDK / Model No.:ZCAT3035) - Ferrite Core x 2(Manufacture : E-TECH / Model No.:CU1330B)

- Ferrite Core(Manufacture : TDK / Model No.:ZCAT3035)

Page 38: Report No.: TE-09-0705-EC0082kgbmed.ru/docs/60601-1-2_knx-7000_elet_ispytaniya.pdf · report no.: te-09-0705-ec0082 page : 2 / 38 table of contents 1. general description of eut 3

TES-D05-02/2 SGS TESCO Korea Co., Ltd. GSBC, 2F, 906-5, Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea. t +82 (0)31 259 6500 f +82 (0)31 259 6511 www.kr.sgs.com www.tesco21.com

Member of SGS Group (Société Générale de Surveillance)

Report No.: TE-09-0705-EC0082 Page : 38 / 38

- Noise Filter(Manufacture : FINE SUNTRONIX / Model No.:SN-M6H-CM)