REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the...

10
This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement between Intertek and its Client. Intertek's responsibility and liability are limited to the terms and conditions of the agreement. Intertek assumes no liability to any party, other than to the Client in accordance with the agreement, for any loss, expense or damage occasioned by the use of this report. Only the Client is authorized to copy or distribute this report and then only in its entirety. Any use of the Intertek name or one of its marks for the sale or advertisement of the tested material, product or service must first be approved in writing by Intertek. The observations and test results in this report are relevant only to the sample tested. This report by itself does not imply that the material, product, or service is or has ever been under an Intertek certification program. Measurement uncertainty budgets have been determined for applicable test methods and are available upon request. FOR THE SCOPE OF ACCREDITATION UNDER NVLAP LAB CODE 100402-0. REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100707553 Date: April 12, 2012 REPORT NO. 100707553CRT-002 TEST OF ONE LED HIGHBAY FIXTURE MODEL NO. HB2C4K RENDERED TO DIALIGHT CORPORATION 1501 ROUTE 34 SOUTH FARMINGDALE, NJ 08005 TEST : Electrical and Photometric tests as required to the IESNA test standard. LABORATORY NOTE : The laboratory that conducted the testing detailed in this report has been Qualified, Verified, and Recognized for LM-79 Testing for ENERGY STAR for SSL by US DOE's CALiPER program. STATEMENT OF LIMITATION : This report must not be used by the client to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the federal government. AUTHORIZATION : The testing performed was authorized by signed quote number 500371675. STANDARDS USED : The following American National Standards or Illuminating Engineering Society of North America Test Guides were used in part or totally to test each specimen: IESNA LM-79: 2008 Approved Method for Electrical and Photometric Measurements of Solid-State Lighting Products ANSI NEMA ANSLG C78.377: 2008 Specifications of the Chromaticity of Solid State Lighting Products DESCRIPTION OF SAMPLE : The client submitted two samples of model number HB2C4K. The samples were received by Intertek on April 3, 2012, in undamaged condition, and two samples were tested as received. The sample designations were D246152-1 and D246148-1. DATES OF TESTS: April 6, 2012.

Transcript of REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the...

Page 1: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement be tween Intertek and its Client. Intertek's responsib ility and liability are limited to the terms and conditions of the agreemen t. Intertek assumes no liability to any party, othe r than to the Client in accordance with the agreeme nt, for any loss, expense or damage occasioned by the use of this rep ort. Only the Client is authorized to copy or distr ibute this report and then only in its entirety. An y use of the Intertek name or one of its marks for the sale or advertisem ent of the tested material, product or service must first be approved in writing by Intertek. The obse rvations and test results in this report are relevant only to the sam ple tested. This report by itself does not imply th at the material, product, or service is or has ever been under an Intertek certification program. Measurement uncertainty bud gets have been determined for applicable test metho ds and are available upon request.

FOR THE SCOPE OF ACCREDITATION UNDER NVLAP LAB

CODE 100402-0.

REPORT

3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100707553 Date: April 12, 2012

REPORT NO. 100707553CRT-002

TEST OF ONE LED HIGHBAY

FIXTURE MODEL NO. HB2C4K

RENDERED TO

DIALIGHT CORPORATION 1501 ROUTE 34 SOUTH

FARMINGDALE, NJ 08005

TEST: Electrical and Photometric tests as required to the IESNA test standard. LABORATORY NOTE: The laboratory that conducted the testing detailed in this report has been Qualified,

Verified, and Recognized for LM-79 Testing for ENERGY STAR for SSL by US DOE's CALiPER program.

STATEMENT OF LIMITATION: This report must not be used by the client to claim product certification,

approval, or endorsement by NVLAP, NIST, or any agency of the federal government.

AUTHORIZATION: The testing performed was authorized by signed quote number 500371675. STANDARDS USED: The following American National Standards or Illuminating Engineering Society of

North America Test Guides were used in part or totally to test each specimen: IESNA LM-79: 2008 Approved Method for Electrical and Photometric Measurements of Solid-State Lighting Products ANSI NEMA ANSLG C78.377: 2008 Specifications of the Chromaticity of Solid State Lighting Products DESCRIPTION OF SAMPLE: The client submitted two samples of model number HB2C4K. The samples

were received by Intertek on April 3, 2012, in undamaged condition, and two samples were tested as received. The sample designations were

D246152-1 and D246148-1. DATES OF TESTS: April 6, 2012.

Page 2: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 2 of 10 Date: April 12, 2012

SUMMARY

Model No.: HB2C4K Description: LED Highbay

Criteria Result Total Lumen Output 10645 Lumens

Total Power 146.5 W Luminaire Efficacy 72.67

Power Factor (at 120 Vac) 0.996 Power Factor (at 277 Vac) 0.944 Current ATHD (at 120 Vac) 2.79% Current ATHD (at 277 Vac) 8.01%

Correlated Color Temperature (CCT) 5925 K Color Rendering Index (CRI) - Ra 76.8 Color Rendering Index (CRI) - R9 18.1

Duv 0.002 Chromaticity Coordinate (x) 0.324 Chromaticity Coordinate (y) 0.331 Chromaticity Coordinate (u’) 0.205 Chromaticity Coordinate (v’) 0.471

Maximum In-Situ Source Temperature Point 69.7°C

EQUIPMENT LIST

Equipment Used Model Number Control Number

Last Calibration

Date Calibration Due Date

Leeds & Northup Standard Resistor Manganin Y089 02/24/12 02/24/13 Data Precision Digital Voltmeter 3600 V124 02/24/12 02/24/13 Fluke Multimeter 45 M133 02/24/12 02/24/13 Kikusui DC Power Supply 35-10L E160 --- --- Sorenson DC Power Supply DLM150-20E -- --- ---

NIST Spectral Flux Standard Source RF1024 --- 09/18/10 100 hours of use

Yokogawa Power Analyzer WT1600 E462 06/22/11 06/22/12 ITS 3 Meter Sphere W/ CDS 1100 N307 w/use w/use EXTECH 445703 T1366 10/26/11 10/26/12 Elgar AC Power Supply CW1251 -- -- -- Yokogawa Power Meter WT210 E464 04/19/11 04/19/12 LSI High Speed Mirror Goniometer 6440 -- w/use w/use Cole Parmer Hygro Thermometer 445703 T1359 10/26/11 10/26/12

Page 3: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 3 of 10 Date: April 12, 2012

TEST METHODS Seasoning in Sample Orientation – LED Products No seasoning was performed in accordance with IESNA LM-79. Photometric and Electrical measurements – Distribution Method A LSI Type C High Speed Model 6440 Mirror Goniometer was used to measure the intensity (candelas) at each angle of distribution for each sample. Ambient temperature was measured equal to the height of the sample mounted on the Goniometer equipment. Each sample was operated at input rated voltage in its designated orientation. Each sample was allowed to stabilize for at least thirty minutes before measurements were made. Electrical measurements including voltage, current, and power were measured using the Xitron or Yokogawa Power Analyzer. Some graphics were created with Photometrics Plus software. Photometric and Electrical Measurements – Integrating Sphere Method A Labsphere Model CDS 1100 CCD Array Spectroradiometer and Two Meter or Ten Foot Sphere was used to measure correlated color temperature, chromaticity coordinates, and the color rendering index for each SSL unit. Ambient temperature was measured at a position inside the sphere. Each SSL unit was operated on the client provided driver at the rated input voltage in its designated orientation. Each SSL unit was allowed to stabilize for at least thirty minutes before measurements were made. Electrical measurements including voltage, current, and power were measured using the Xitron or Yokogawa Power Analyzer. The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of Standards and Technology. In-Situ Maximum Measured Power Supply Case and LED Source Point Temperature Power supply case and/or LED source operating temperature measurements were taken on one test sample per model with a thermocouple and Fluke 87 temperature meter. The SSL sample was allowed to reach thermal equilibrium for seven and a half hours before measurements were taken. Power supply or source temperature measurements were measured at the TMPPS or TS point as indicated by the included diagram in accordance with manufacturers declared hot spot location, or at a hot spot location found with a thermal camera when no diagram from the manufacturer is given. The maximum temperature was recorded for the sample. A simulated ceiling or other enclosure may be used in accordance to UL 1598 or UL 153 as applicable.

Page 4: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 4 of 10 Date: April 12, 2012

RESULTS OF TESTS Spectral Distribution over Visible Wavelengths nm mW/nm nm mW/nm nm mW/nm nm mW/nm

HB2C4K 350 1.514 460 123.066 570 141.057 680 28.146 355 1.532 465 87.749 575 139.197 685 24.841 360 1.326 470 62.857 580 136.804 690 21.774 365 1.528 475 47.058 585 133.744 695 19.186 370 1.346 480 39.389 590 129.945 700 16.815 375 1.470 485 36.985 595 126.235 705 14.721 380 1.475 490 39.428 600 121.734 710 12.874 385 1.475 495 47.576 605 116.515 715 11.267 390 1.750 500 60.369 610 110.685 720 9.861 395 2.051 505 75.661 615 104.907 725 8.546 400 2.939 510 91.483 620 98.234 730 7.466 405 4.867 515 105.981 625 91.386 735 6.486 410 9.323 520 118.067 630 84.540 740 5.678 415 18.386 525 127.418 635 77.517 745 4.959 420 35.893 530 134.417 640 70.712 750 4.367 425 63.365 535 138.418 645 63.918 755 3.921 430 99.288 540 141.208 650 57.389 760 3.407 435 143.899 545 143.310 655 51.855 765 0.000 440 202.741 550 144.080 660 46.003 770 2.679 445 261.151 555 144.554 665 40.969 775 2.338 450 254.998 560 143.937 670 36.282 780 2.107 455 183.807 565 142.933 675 31.951

DIALIGHT CORPORATION Sample No. D246152-1

Model No. HB2C4K Spectral Data Over Visible Wavelengths

Page 5: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 5 of 10 Date: April 12, 2012

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements at 25°C – I ntegrating Sphere Method

Intertek Sample

No.

Correlated Color

Temperature (K)

CRI -Ra

CRI -R9

DUV

CIE 31’ Chromaticity Coordinate

(x)

CIE 31’ Chromaticity Coordinate

(y)

CIE 76’ Chromaticity Coordinate

(u’)

CIE 76’ Chromaticity Coordinate

(v’) HB2C4K

D246152-1 5925 76.8 18.1 0.002 0.324 0.331 0.205 0.471

Intertek Sample No.

Base Orientation

Input Voltage (Vac)

Input Current

(mA)

Input Power (Watts)

HB2C4K D246152-1 UP 120.0 1222 146.4

Intertek Sample No.

Input Power Factor

120 Vac

Current ATHD

(%) 120 Vac

Input Power Factor

277 Vac

Current ATHD

(%) 277 Vac

HB2C4K D246152-1 0.996 2.79 0.944 8.01

Page 6: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 6 of 10 Date: April 12, 2012

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No.

Base Orientation

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy

(Lumens Per Watt)

HB2C4K D246152-1 UP 120.0 1220 146.5 0.997 10645 72.67

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 HB2C4K

0 4203 4203 4203 4203 4203 5 4112 4114 4102 4106 4107 10 3949 3972 3968 3982 3999 15 3796 3815 3821 3842 3870 20 3617 3649 3668 3692 3742 25 3464 3501 3525 3562 3650 30 3477 3495 3531 3609 3732 35 3503 3556 3595 3693 3825 40 3544 3588 3650 3741 3850 45 3411 3392 3454 3536 3436 50 2625 2626 2687 2698 2588 55 1909 1864 1845 1835 1712 60 1139 1164 1157 1093 1144 65 795 786 776 745 766 70 420 481 460 448 445 75 206 239 230 206 189 80 52 56 62 51 37 85 5 5 5 4 1 90 0 0 0 0 0

Page 7: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 7 of 10 Date: April 12, 2012

RESULTS OF TESTS (cont’d) Illumination Plots

Model No.: HB2C4K Mounting Height: 10 ft.

Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire HB2C4K

0-30 3185 29.9 0-40 5572 52.3 0-60 9699 91.1 60-90 946.1 8.9 0-90 10645 100.0

90-180 0.0 0.0 0-180 10645 100.0

Page 8: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 8 of 10 Date: April 12, 2012

RESULTS OF TESTS (cont’d) In-Situ Maximum Measured LED Source Temperature Manufacturer Supplied Documentation: LED identified as: CREE XT-E

Page 9: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 9 of 10 Date: April 12, 2012

RESULTS OF TESTS (cont’d) In-Situ Maximum Measured LED Source Temperature Maximum Junction Temperature from LED specification (Tj) = 150°C Thermal Resistance Formula from LED specification = 5°C/W Maximum Forward Voltage (Vf) from LED specification = 3.4 V Measured LED Current = 593 mA Calculated LED Wattage = Vf x Measured LED Current = 2.016 W Maximum Source Temperature (Ts) = Tj – (LED Wattage x Thermal Resistance) = 140°C Maximum Measured Manufacturer Designated Source Temperature

Sample No. Model No.

Maximum Measured Source Temperature (°C) Location

Maximum Rated Source

Temperature(°C) D246148-1 HB2C4K 69.7 Per diagram above 147

In-Situ Picture – Ts

In-Situ Picture – Ts location

Page 10: REPORT ACCREDITATION UNDER NVLAP LAB - LED Technology, LED …€¦ · The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of

Report No. 100707553CRT-002 page 10 of 10 Date: April 12, 2012

Picture (not to scale)

CONCLUSION The results tabulated in this report are representative of the actual test samples submitted for this report only. The data is provided to the client for further evaluation. Compliance to the referenced specification requirements was not determined in this report. In Charge Of Tests: Report Reviewed By:

Joseph Schledorn Jacki Swiernik Engineer Staff Engineer Lighting Division Lighting Division Attachment: None