1 R&D status report of LST technology (2) Changguo Lu Princeton University BaBar IFR Workshop SLAC, December 8, 2002

description

R&D status report of LST technology (2). Changguo Lu Princeton University BaBar IFR Workshop SLAC, December 8, 2002. Outlines. LST cell size study; Aging test results; LST rate capability; QA at US assembling sites. LST cell size study. - PowerPoint PPT Presentation

Transcript of R&D status report of LST technology (2)

Page 1: R&D status report of LST technology (2)

1

R&D status report of LST technology (2)

Changguo Lu

Princeton University

BaBar IFR Workshop

SLAC, December 8, 2002

Page 2: R&D status report of LST technology (2)

2

Outlines

• LST cell size study;

• Aging test results;

• LST rate capability;

• QA at US assembling sites.

Page 3: R&D status report of LST technology (2)

3

LST cell size study

Comparison of three different cell geometries:

• Standard cell size — 9 9 mm2, ST type;

• Small size, anode wire is located at the center of the cell — 9 8 mm2, SC type;

• Small size, anode wire is equally distant from three conductive walls — 9 8 mm2, SE type.

Criterion for the comparison:

Relative variation of the E-field on the anode wire surface along the circumferential direction.

Page 4: R&D status report of LST technology (2)

4

E-field variation on anode wire surface

Coverless LST chamber has one distinguish feature: its E-field on the anode wire surface varies along the circumferential direction.

9 x 9 mm2 cell with conductive cover

V contour plot E-field lines

Page 5: R&D status report of LST technology (2)

5

Coverless cell with normal cell size

V contour plot E-field lines

Page 6: R&D status report of LST technology (2)

6

E-field distribution on the anode wire surface

9 x 9 mm2 cell size, 100m anode wire, Vanode = 4700V

For all-covered cell E-field on the anode wire surface is constant, but for coverless cell E-field varies with .

The criterion for comparison:

(Emax – Emin)/Emax

Page 7: R&D status report of LST technology (2)

7

… E-field distribution

ST type(standard cell)

SC type(anode at center)

SE type(anode equally distant to conductive walls)

Emax (V/cm) 199,137 199,704 196,126

Emin (V/cm) 197,938 198,002 194,352

(Emax – Emin) / Emax

0.6% 0.85% 0.9%

V(V)* 28.3 40.0 42.5

* V = 4700* (Emax – Emin)/Emax

The differences among three geometries are small, 9 8 mm2 cell size won’t be a problem.

Page 8: R&D status report of LST technology (2)

8

Plateau curve of different gas mixture

Standard binary gas mixture.

Safe gas mixture (ZEUS)

Both of them can get good single’s rate plateau curve.

Page 9: R&D status report of LST technology (2)

9

LST for ZEUS, SLD, Phenix and our proposed one

ZEUS SLD PHENIXProposed BaBar barrel IFR upgrade

Building material

Noryl PVC PVC Noryl

coating resistivity

Low resistivity

High resistivity High resistivity (?)

High resistivity

readout anode readout only

strip and pad readout

anode readout only

strip readout only

Gas mixture Ar/C4H10/CO2

(3/8/89)Ar/C4H10/CO2

(2.5/9.5/88)C4H10/CO2

(9/91)Ar/C4H10/CO2

(3/8/89)

Working mode

Limited streamer

Limited streamer

Avalanche Limited streamer

No one system is a duplication of the other, we have to conduct our own aging test.

Page 10: R&D status report of LST technology (2)

10

Update on aging test

So far the accumulated charge dose has reached 170 mC/cm, no aging sign has shown up yet.

The test chamber used in this test was made of PVC, anode is 100m Au-coated Cu-Be wire, a ceremic end board used to distribute high voltage to eight wires, PVC profile was painted with standard graphite paint. The chamber was heat sealed and the wires were heat staked onto plastic supports. No glue was used.

We are proposing to use Noryl instead of PVC as the profile and envelop building material. According to ZEUS aging test results Noryl certainly will be much better than PVC. SMC experience shows that their PVC LST aging was well beyond 1C/cm, so we believe that the Noryl LST should be even better than 1C/cm.

Page 11: R&D status report of LST technology (2)

11

… Update on aging test

Page 12: R&D status report of LST technology (2)

12

Effect of the gas pressure on the streamer signal size

LST signal size is sensitive to temperature and gas pressure.

Aging test chamber with Sr-90 source. During this period the room temperature was quite stable, ~ 240.5C.

Page 13: R&D status report of LST technology (2)

13

Is LST suitable for BaBar endcap IFR?

Two key performances need to be checked (based on Jerry’s estimation):

• Rate capability - 100Hz/cm;

• Aging performance – 2C/cm.

Qrpc = 1000pC Q = 20C/cm2

Qlst = 100pC Q = 2C/cm2

Page 14: R&D status report of LST technology (2)

14

Rate capability of LST

The spectra of 61Hz/cm and 111Hz/cm are generated by Sr-90 source with different absorbers.

Very similar to cosmic ray background LST can handle the rate of BaBar IFR endcap layer 18.

Page 15: R&D status report of LST technology (2)

15

… rate capability

Further increase the source intensity we can see the charge spectrum shifting towards left:

Aging performance up to 2C/cm2 for Noryl LST is still unknown. But so far no any negative evidence. We are going to perform this test as long as we get final version of LST from PolHiTech.

Page 16: R&D status report of LST technology (2)

16

Proposed QA procedure at Princeton assembling site

Never overestimate the importance of QA for a successful large LST system.

PolHiTech will continue to use the QA procedure for Phenix LST mass production. They will continue to monitor the anode wire current for one month. The final selection of good chamber will be based on the current history of the chamber. After receiving the LST chambers at US assembling sites (Princeton and Ohio State), before making the final assembly unit, second round of QA will take place.

The proposed QA procedure at assembling site is based on successful OPAL LST system. That procedure had been used to test more than 28000 LST tubes (3609 8-cell chambers of length 3.0 m to 7.3 m) at CERN. Similar QA program was used at University of Maryland during that time.

Page 17: R&D status report of LST technology (2)

17

QA procedure

1. Stability of the anode current — at the maximum voltage continuously monitor and record the current drawn by each chamber for one month. Criterion for acceptance: high current activity above 100nA < 5% of the time in the first half of the history, no activity in the second half of the time;

2. Pulse-form test — decent pulse shape should be seen on the scope: 30 ~ 50 mV, 40 ns. Criterion fro rejection: anomalously shaped pulses or no pulse at all, indication of construction faults;

3. Cathode scanning with a radioactive source — using radioactive source to produce exceptionally high rate of charge exchange on the cathode to detect defect on the graphite coated cathode. Garfield calculation shows at 4700V the total cathode charge ~ 55nC/m, if one anode wire current is 55nA, the source moving speed is 10cm/s, then within 10 seconds we can scan 1m long chamber, and make 10 times cathode charge exchange.

Page 18: R&D status report of LST technology (2)

18

Cathode scan

Due to graphite coating non-uniformity, some isolated islands may exist on the cathode surface, under the normal background rate those bad spots may not cause trouble, but under high rate condition it would trigger self-sustaining discharge, and reveal this type of defect.

Even w/o source the high current is still remaining.

F. Fabbri et al. CERN/EP 87-134, 29 July, 1987)

Page 19: R&D status report of LST technology (2)

19

Conclusion

• LST E-field study by Garfield for three different cell’s geometries shows that the small cell size (9x8mm2) has very little affect on the E-field distribution, will not deteriorate the performance. A real test will be conducted on the new prototypes;

• Aging test on an old PVC LST chamber shows that up to 170mC/cm charge dose, no aging symptom can be seen, which indicates even with PVC LST should have no problem for barrel IFR;

• Rate test shows that up to 110Hz/cm the charge spectrum shows no difference from cosmic ray background, which indicates that LST should be able to handle the rate for BaBar IFR endcap layer 18 @ 4x1034/cm2sec PEP II luminosity.