RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC...

53
RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE T 6 TOWNS ET AL. AUG 84 UNCLASSIFIED SARRI-EN-84-ii F/G 7/4 NL Ehhmmmhmhhlo mohmhhEEmhmhEE mh||h|hhE|hE

Transcript of RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC...

Page 1: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/iI ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL ILI ENGINEERING DIRECTORATE T 6 TOWNS ET AL. AUG 84

UNCLASSIFIED SARRI-EN-84-ii F/G 7/4 NL

EhhmmmhmhhlomohmhhEEmhmhEEmh||h|hhE|hE

Page 2: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

.12.

I1I 11111220

iI ii L.-' m

11111.25 l.4z 1

MICROCOPY RESOLUTION TEST CHARTNATIONAL BUREAU OF STANDAPDS 1963 A

rI'.1

Page 3: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

EN-84-1 1

00

, 0

* TRACE METALS ANALYSIS OF WASTEWATER BYQ ATOMIC ABSORPTION SPECTROSCOPY

BY

I THEODORE G. TOWNS, Ph.D.and

KAREN L. FRANICH

AUGUST 1984

0

rSTRIION STATEMENT ARY

*1 -vv*dI public releoseg JAN 14 1985C2, Dlstrbufton Unimited

B

ENGINEERING DIRECTORATEROCK ISLAND ARSENAL

ROCK ISLAND, ILLINOIS 61299-5000

*85 01 08 00

- •

Page 4: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

* DISPOSITION INSTRUCTIONS:

Destroy this report when it is no longer needed. Do not return to theoriginator.

DISCLAIMER:

The findings of this report are not to be construed as an official -

Department of the Army position unless so designated by Dther authorized-0documents.

7S

Page 5: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

UnclassifiedSECURITY CLASSIFICATION O0 TIS PAGE (When Data EnterocO

REPOT DCUMNTATON AGEREAD INSTRUCTIONSREPOT DCUMNTATON AGEBEFORE COMPLETING FORM1REPORT NUMBER 2. GOVT EACCESSION NO. 3. RECIPIENT'S CATALOG NUMBER

* ~EN'-84-11a /_____________

*4. TITLE (and Subti) S. TYPE OF REPORT & PERIOD COVERED

TRACE METALS ANALYSIS OF WASTEWATER BY ATOMICABSORPTION SPECTROSCOPY Technical Report

6. PERFORMING ORG. REPORT NUM1BER

*7. AUTHOR(e) 8. CONTRACT OR GRANT NUMBER(a)

Theodore G. Towns, Ph.D. and Karen L. Franich

9. PERFORMING ORGANIZATION NAME AND ADDRESS 10. PROGRAM ELEMENT. PROJECT. TASK

Commander, Rock Island Arsenal AREA A WORK UNIT NUMBERS

ATTN: S11CRI-ENM1-ERock Island, IL 61299-5000 _____________

11. CONTROLLING OFFICE NAME AND ADDRESS 12. REPORT DATE

*Commander, Rock Island Arsenal August 19844

ATTN: SMCRI-EN13NUBROPAERock Island, IL 61299-5000 35 UMEOFAE14- MONITORING AGENCY NAME & AODRESS(If different from Controllind Office) 1S. SECURITY CLASS. (of Chia report)

UnclassifiedS. DECLASSI FI CATION/ DOWNGRADING

SCHEDULE

16. DISTRIBUTION STATEMENT (of this Report)

Approved for public release, distribution unlimited.

17. DISTRIBUTION STATEMENT (of the abstract entered In Block 20, It different ftem Report)

I1. SUPPLEMENTARY NOTES

*IS. K EY WORDS (Continue on reverse aide If necessary and Identifyr by block anumber)

*1. Atomic Absorption Spectroscopy 6. Calibration Curves2. Effluent Water 7. Linear Regression3. Wastewater 8. Mulianalyte Reference Solutions4. Trace Metals 9. Spiking of Wastewater5. Quality Control 10. EPA Quality Control Samples020. ABSTACT (Coname tw roere sele N nosoema sai Ide.ify by block mmb~w)

The primary project objectives of upgrading effluent trace metals analysis byatomic absorption spectroscopy (AA), providing advanced AA instruction, andmodeling a schedule for quality control data in the analyses have been success-hItlly completed. The upgraded and automated system will meet the needs of

fwastcwater analysis In an accurate and efficient manner that yields a high* level of confidence.,I

DO1AN3 72 DTOWFNvSIOSLT Unclassified

iSECURITY CLASSIFICATION OF THIS PAGE (When Does Enteredt)

Page 6: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

* I:

FOREWORD

This report was prepared by Dr. Theodore C. Towns, Howard University,Washington, D.C. and Ms. Karen L. Franich, Rock Island Arsenal, RockTsland, Illinois. The work ws accomplished during the Summer of 1I',Rwhile Dr. Towns 's employed under the Historically Rlak College Programfor professors.

The project ws carried out in the Analytical Chemistry l.ahoratory,Engineering Directorate, Rock Island Arsenal. The authors would like to ""thank Mr. George Powers for his assistance in the experimentation. . -

S I

* i-

-.7

\

Accession For-.4NTIS GRA&I

DTIC TAB

ELECTE ousoif [] t.

JAN 1 4 1985 -" ""

SDist ribut ion/

B Availability ('odos

IDist SpLuo 1-11

0Iiii *'

* •I,

* . .,. .

. . . . . . . . . . . . . . . . . . . . . . . . . . ..

i

Page 7: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

1.0 TNTROP1Cr"TON . ... a-1.1 Problem Sratp ,Pnt. . . ... ... .... .. ....

1.,2 Obiective .. . . .......

1.3 Miscellaneou . .. .?.. . . . . . . . . . .

2.0 T)TSruSS ION . . . . . . . . . . . . . . . . . . . . . . . . -2.1 EPA Ouality Control .Samples for Trace petals. ..... .- 1,"

2.9 The Calibration Curve. ....... ................. -

2.3 Preparation of Reerence Solutions .............. -?i

2.4 Multianalyte Re'ference Solutions . ...... ..... . -. ..

?.5 Absorbance versus Concentration. ... ............ ... 2-r,

T. The Instrument ............................. -q

?.7 'aintenance. ........... .............. - 0

?.8 Instrument Alignment ....... .................. 2-11

P.0) Recorder Setup ........ ..................... -1

3.0 PF.SUL"5. . . . ..

3.1 Spiked Samples ....... ..................... .

I.P Sampling Sequence ....... .................... .

)i.0 CONCL11STOS..s..... ................................. -

9 .0 n,F, RAL R~rOnWFDACTON' . .. .. .. .. .. ..... . ..

r.0 0ONTIMMEr RCOMMEMNDArTn, . ".

7.0 RFERFNCrF.. .......... ........................ 7-1

APPFN Y"" '

iv4_

~-.* '2 - -- ** - -

* .. -- -.-

Page 8: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

S

'4 0

0

"I.

0

0

0

0

0

0

4 0

4 0

. . . - . . . .

* . . . . . . . . . . . . *'.~*v. *-*. \****.

Page 9: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

LIST OF FIGUIES'

Fig. No. Title

1 Distribution Curves for Ouality Assessment . . . . 2-?

2(A) Generalized Calibration Curve (dati for C.' . . P-7

(P) Cadmium Standards (low ppb level) ......... ...

I(A)(R) Chromium Standards (low ppm level) . ....... .

*I(A),B) Chromium Standards (for data in Table I) . .-....

5(A) Iron 7tandards (low ppb level) ...........

(B) Iron Standards (low ppm level) .... ............ it_ -

Lead Standards (low ppm level ........... it_-,s

Lach standards curve has inset the result of linear regression performedon the data; A or M slope, B : intercept, R correlation coefficient.

v0

'" " 0

• .

-o

.• t

Page 10: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

]

J0

0

p0

II. S

S

I p

A

*

9 S

*

............. . . . . . . .

Page 11: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

LIST OF TABLES

Table No. Title Pg

I ~Example F~dta from Chiromium AA Analysi . . -

IT Statistical Evaluation for Table I Data. ...... J

III (urrent Limits for RIA Effluent . 5 .... 4-

I S

Page 12: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

1.0 T'DOrYIrTTON

.

1.1 Problem ',tatei.ent .

Frratic or inconsist.ent innlvtical r.sults in atom;P ah- ortin r.-asalys.s can usually be attrihuted to the occurrene ono or more or th'.-ol lowing:

(a) Poorly prepared -ererenoe mvteria!s.

(h) Tnadequate instrment optim 5 ,ion.

(c) Meglect oP routine instrument imaintenance.

"he scope o" wnk conce.rns the remiation oP' thee ociirrences.Point. (a) generally results r-om pither inadequate star-inq materials orinadequate laboratory technique in the preparption; the Ponus o" this wn-!"has been on the laboratory technique. nf points (b anid (o), tho lant.e"is perhaps the more serious since it inevitablv results i n hoth anelevated minimum detectable limit., i.e., lost sensitivity, and adegradation ol analytical accuracy, while the "ormer la's p-incinallv tolost sensitivitv. Tnstrument optimization is addressed in Tho form n" qstandard operating procedure with notes detailing the interplav andadiustment or the operatin, parameters of the Varian model AA- 7 7£ uised at •Rock Island Arsenal (RTA).

NOT

This operatin, prnocedure mst. in no wa, be conr-f'-d

to supersede the nanfaturer's "Operations "aniinl."Rather, it represents a synopsis of procedures r!.rman# tothe routine use of the instrument in the PTA %nrineerinr'Directorate and it provides el.aboration for point,briefly mentioned in the AA manual. [1]

1.? Objectives.

The ohiectives ofr this report are to de.cribe:

(a) The upgrado in capability of' the Varian model A-77r A~trn: "

Ahsorption .pectrophotometlr, with particular reanrd for t'#- appic'tiontn routine e-ffluent trace metals analysis. 0

(b) The provision of' advanced instruction on the use o' AtomicAbsorption Spectroscopv (AA.) in wastewater trace metals analysis.

4"o

1-1

.... .= -- - - .. .,- , -, . • , .., .:, , ,. i . . .-.. .. ..- ) : . ... i0

Page 13: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

A ',odel "ohediile In-' th'-n inai nri "v''s-0nenxiv ty cotrn' (4atn st etn o mee-t or Pxrope4 pvt-an+. ?rlw-rlnrecguti'1ios a:ndI T' A"-"'r "nvirnrmnt.al TT\,f-irip A17onolCv 'ATr"A) I)mli-f in o

rf) vh rvew *m.' 'ahorrat orv prnod 1r- "r' 4ho~ n i rrno',''n'.o i nr t hose -sterns knF qppr'c ia Ie uneo- ~it i 4es-r .ir 7 r, ImIvz n'rr 4i r-l~

* ~ .v pprsonnel.

The IIVaoe&,4 itilru- t ion on the uno or AA ani- os1-.;t~s ~'* t~~'ull or' seotion m).C, M-ous.~ssion. Tn snection r.l nA Omi'' to

!7arm's r'or Trnv'e ?W t mls, arse notes on 4-he "nij-r'onrwnt.a'1 Potention T'genev(FPA) iullitv e-ont-rrol sa'p.sand theiv' li we; no's on Th"orrtorrv

*tochniqiios riecessar- 'or the rolitino detectfion of -innlvtpr- in thernocentration. ran(ge helow one part per Mn'lior. (ppm); rosesq on n-ar, it,o)f a c,.alihration curve iIts usne; nnI notes, on the t,'il;nterpret.ition 0" fhr oxruerimcflpl danta, chieo'anprorririle onorafinTParameters, and -vainftenanre. Notes on lahoratorv nrocenrr are P .~otop~c a" "ections 1.? throupgh 1.0.

'2Fni~canf- enhancr-mrnt. in fhn 'atistical -irnurnov ;irA- nicr'iqion il'th~conce ntration meaiwer-ont. can I- achieved wi~th 4hp at'a1chment or bt

an autoramnling access-orv aniI remotp outpult deOvies! such -1recorder an4 a d~ipital p-'intpr. fRnhnncement in tho user o" the- -tt.e-n'a.-ntpersonnel's time mav ,Also accompany t ', s upgrade or' the ?"'Icinstrument. Details on "-le use or' selected accessories wre "ounrl insecotion ?KThp Tn7strumpn-..

The qualityv control schedule is caiven inrH oxernplificd in scin*.lPen I fts. T he diata is ~anf'romi the previln' effluent nn.alv!-': loar,luring 41ho course o' th's Thd.0

Tn addlition, the term "lahsov'hne' asi7 e in ~ "the rinto!urvfen rcr thlq report -)CtuUlv 'ef'er.c to t-he rel'tvp aThsorhancQ

detcmne f'rom the r)m :VI-!hts on '-hp chart r(ecorler mreasurQ& inmillimeters. The poi-- hr-Tthts are U rnctlv proportiona1i tor 1'oo metwl.

obsrbrtenn,tc~r, f-ive n valir and uq're'ul calhatio O fVC"rthe? letermnition o,' sav'1plo7 concentrntion.

.2.

Page 14: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

_7% I - -*

P.0l D I SCUSS ION

2. V.PA Qualitv Control Samples for Trace Meta1. I..The EPA quality control (EPA QC) samples3, ivai 1 , a.- mutt ir, nalyto,

oonrett rate-, are re ference so lutions which have bee. pren1,,ed ard stnr1ander exacting laboratory conditions. The mdividui analIvteconcentration is riven both in terms of , "true" value definirf the numterof significant figures associated with its preparation, and in terms of

st;,tistical value. (mean, X; and standard deviation,, ) derived fromthe amount detected by a number of independent analytical laboratcries.T"he quality control program is based on a 95 percent confidence interval,and thus, results for a OC samplm are considered acceptahle if they fn' Iwithirn X + 2 (3. The estimate of a OC analyte concentration derived fror awork inp standard--, ourve shou d be accompanied by its own standarddeviation reflecting the var -ine in its determination. Lahoratoryprocedures must be sufficient'v exacting to give a detected analyteconcentration, X', includinj, its .-tandard deviation, ', within the Qr--ample limits. The areceptable range given by EPA takes this into account..-he analytical sitiat ion is represented in Figure 1. Note that, for, implicity, the figure has been couched in terms of a distribution with. ; other equivalent representations are achieved by shifting theorigin along the ah.sissa by an amount equal to the non-zero mean valuewith no change in the shapes of or relationship between the twod Istributions.

,PA OC samples may be put to several uses: a control for thedigest ion efficiency; a check on the quality of prepared referencesolutions; And a guniide to sample analysis accuracy. As a check on thereference solutions,, the EPA QC samples may be analyzed unadulterated(neat.) and this should be done at least as often as "fresh" referenceroIut ions are pr epared. As ai control on either the digestion or thrii raly.'i , the !'A Q' sample or one of the prepared reference solutionn i-

.ritrodice, ,-o a ","pike" in one of the unknown samples, either before orafter the digest ion process. When the digestion efficiency is not inquest ion, and the chemical interferences in the sample matrix are known tobe nor!nal or absent, the neat EPA QC sample is an adequate control.However, fo;r -;amples with unknown history or for those with knowninterferring matrices, the EPA QC sample or one of the prepared referencesolutions must be rmn as a spike. This use of a spike is preferred andrecommended ideally in all cases, but it may be deferred as appropriate tnthe neat EPA QC sample in the interest of time and material savings.

'7 T"he Calibration Curve.0|

If the CPA OV sample is considered the "heart" of a quality controlprogrim, the (alibration, or standards, curve must be taten as the "mind;",h! two must work together to provide the desired quality assurance. The

. -.

,.

Page 15: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

4 S1

A.0

C

InXU

'-4

I.L)

0'

0

F--

2-24

Page 16: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

etrfieay of tho -1andards curve will be determined in par. hy t!.ta'<en in it!, preparaition and in part by the though' l 1n05';s of it, ue.",' r'A OC sample concerns the laboratory technique exerci!-,ed In t.h,.v iltimetric prepar-tion of solutior.n, the calibrntiorn erv," corncerr:- horelationship between absorbance and concentration for i 7-iven analytl.

9.: >'eparation of Reference Solutions.

The primary trace metal stock solutions, typically avaii'Ke',,rlrc!ally as MOO ppm + 1 percent oonoentrates, must eithor he .Ir,',ot i fjCd standards or be traceable In such standards. 'he roomt.-mr,.riture shelf life of the one pint (100 mL) volumes as supp'ird is a'. ..-

!ea:;t one year, and prohahly long,,r. Secondary stocks and seriali i lu, ions (hereafter referred to as "work ing standards") derived from theprimary stock will have a considerably reduced shelf life which, however,c:in 1- maximized by the procedures given here. Degradat ion of work i rr•st*andards is generally attributed to two phenomena: thermal instabilit.ios(mini-,ized by refrigerated storace), and surface adsorption (minimized hyprior "acid soaking" of the storage containers). 'he importance of thisacid soaking procedurr is often overlooked.

Iefore its '15c, the volumetric glassware (class A, horosilicateglass) including rtoppers and any other containers destined for ntora£pduty, must be soaked in one to one strong acid, , nitric, phosphoric,or perchloric, for at least four hours. After soaking, the one to oneacid can he returned to its own container for reuse over a period ofapproximately six months.

NOTr

This presumes the glassware has already been scrupulouslycleaned with a good quality laboratory detergent, e.g.Alconox, and rinsed with tap and then deionized water.The acid sxoaking solution must not be used as a cleaningsolution for "dirty" glassware.

After emptying the acid solution, the vessels should be rinsed threetimes with deionized water, the excess shaken out, and dried. Drying maybe omitted when the vessels are for immediate use, however, a recommendedpractice is to then make a final rinse with a small portion of the solventor matrix used. Drying can be effected in several ways: evaporation atroom temperature (air drying), rinsing with reagent grade acetone andshaken dry, or blown dry with a compressed air source; if this latter isu-ed, care should he taken to remove the traces of lubricants which oftencontaminate air from such sources. When the empty vessels are. to beshelved for later uise, their openings should be either stoppered, using ashort, 1/R to 1/h inch wide strip of lens paper between the ground glass -.;urfaces, or covered with an inert material such as "Parafilm".Vnlimetric flasks thus pretreated can then be used for convenient,rreliable preparation and storage or working standards.

1

4 "0

Page 17: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

Experience has shown that under the procedures g;iven above:

(a) Secondary stocks of the order of 100 ppm are usahie for, at-most six weeks.

(h) Serial dilutions in the range ().I to 10 ,rm are usahlc for,at nost two week.".

(c) Dilutions to below 0.1 ppm should not be kept for longerthan five days.

As mentioned Parlier, the working standards sho-ld he prepared as oneor more serial dilutions of the secondary stock. Serial dilution affordsa proven method for minimizing the uncertainties passed through thevolumetric work., and provides a tractable design for assessing anomalousoccurrences attributable to the standards. By the same token, serialdilution also demands that the same care be given the preparation of eachin the series. The quality of the dilution will depend on two criticalfactors: temperature and the handling of the volumetric ware.

For very precise work, the glassware and reagents must all he at theambient temperature of the volumetric vessel's calibration, typicallysoecified as 20 deg C on the apparatus label (TD 2N deg r means to deliverat ?0 deg (). Since N) deg r is not always the room temperature of thelaboratory, having the reagents and apparatus involved all at the sametemperature, which is near that of the calibration, is sufficient.Considering the roefficient of thermal expansion of water in theneighborhood of' room wmpersture (approimnately 0.(l00-./deg C), there in' amargin of abtuot 5 deg C over which temperature variation will notintroduce significant loss in accuracy or precision; that is, all elsebeing, equal, laboratory room temperature of, say, deg C will introducean error of less than 1 percent when using glassware calibrated at ?0 der,* The effect of temperature on the volumetric ware itself is also aconsideration, but for good quality borosilicate glassware the effect oftemperature is negligible.

if pipets are not dry before use, always rinse them to a point abovethe fiducial mark, with the solution to be transferred. A recommendedpractice is to wash one's hands before beginning the work, and then avoidhandling any part of a pipet that will contact the reagents. Pipetsshould have a clean, dry surface on which to rest when not in hand. Thefiducial mark must !e at eye level when filling and the miniscus should"sit" on thin mark. Volumetric pipets are calibrated either "to contain"(TI) or "to deliver" (TD); the two letter designation may be found on theglassware in the vi-inity of Its capacity marking. To deliver means that

Safter gravity has withdrawn all but the last drop, the pipet tip istouched to the si !o ,f' the receiver vessel to remove both the drop andSnart of what remains in the tip. There will still be a small amount ofI quid remaining inside the pipet tip, it is there by design, leave it.

A 'P pipet will be labeled as such. A TC pipet may or may not

0P

2-il

. .'. ° ..

Page 18: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

hav- an imprint; the absence of such a marking near the capacity la-el canhP taken an to contain, which means that the last portion inside the tipshould be blown out, after gravity has finished its work.

The amount of analyte transferred for each solution in the seriesshould be as large ds practicable; i.e., smaller vohi mes at highconcentration, larger volumes at lower concentrations. After discharrinpthe contents into a volumetric flask, the matrix is added to brinp, theliquid level to a point below the neck and the vessel is swirled a fewt.ime.3 to insure mixing. The volume is then brought to the mark, the lasthalf milliliter, ml., or so introduced dropwise taking care not to rundrop.s; down the side of the neck because this can lead to a false reading.After filling, the flask is stoppered and inverted at least twice. Thesolution is now ready for use.

Since the ahsorbance versus concentration relation is not oftenlinear over more than an order of magnitude (decade) change inconcentration, adequate characterization of the relationship requires at.least five concentrations per decade span. For example, to cover the spann.1 to 1.0 ppm, a representative schedule of standards is (0.1, 0.3, 0.9,O.P, and 1.0 ppm). Serial dilution implies that each decade of

4 concentration is prepared from the preceding decade. A representativeschedule is:

(a) Blegin with secondary stock of 100 ppm and pipet 10.0, 8.0,5.0, and 3.0 ml, aliquots per 100 mL to prepare the decade 10.0, 8.0, 5.0,and 3.O ppm.

(hW Use 1M ml. aliquots of each from step (a) per 100 mL toprepare 1.0, 0.1, 0.5, and 0.3 ppm.

(c) And so on until the desired limit is reached.

.A 'Multianalyte Reference Solutions. S

This is a practical means of minimizing both the amount of glasswarerequired for routine analysis over a wide concentration range and theamount of time devoted to standards preparation. The foremost concernshere are "chemical interferences" and "chemical (metathesis) reactions."

4 A recommended practice, though not essential, is to compose standards 9solutions as free as possible from both of these Influences. "Chemicalinterferences" is the phrase used to describe the suppression of ananalyte's ahsorbance caused by the presence of another species in thematrix. For instance, chromium is known to suffer this interference fromthe presence of cobalt, iron, nickel, and some of the Group IT-A elements;

4 lead sufferr no cationic interferences but is subject to anionicinterference from concentrations of phosphate, carbonate, acetate,fluoride, ard i"i-ide in excess of ten times the concentration of lead.r'or. on the other hand, while an interference for chromium, is itself

typically suhject to none. The kinds of "chemical interferences" for mostof the common elements analyzed by the atomic absorption are well

* S

p* .

Page 19: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

documented and available. (361"Chemical reactions" are probable whenthe form in which the components are available as primary stock are

* unstable when mixed. Caution is in order only when a reaction removesanalyte from solution, vzprecipitation. For example, while neitherlead nor chromium exhibit any mutual "chemical interference,"~ their formin our primary stock is as the nitrate and dichrom~ate salts, respectively,and their mixture, at the 100 ppm level, leads within seconds to theformation of a flocculent, pale-orange precipitate (a thoroughinvestigation of the nature of this precipitate w~as not feasible, but byits color and cursory solubility studies one of the lead oxides is aprobable candidate; for all of them the solubility product constant iswell below 100 ppm.)

7.5 Absorbance versus Concentration.

Te atomic absorption method is essentially a single wa~velengthultraviolet or visible (tJV/VIS) experiment. The quantitative use is basedon the well known Beer-Lamb~ert Law, A =a-b-c, where a =absorptivity (anintensive property of the absorber which is wavelength dependent), b=optical path length, and c =concentration. The law defines a linearrelationship which passes through the origin between absorbance A (anextensive property) and concentration. However, inherent in thederivation of the law are two principal assumptions, the second of whichhas significant bearing on the preparation and use of the calibrationcurve: that the absorptivity at a given wavelength Is constant; and thatthe absorbing species are non-interacting. While the absorptivity is

* appreciably dependent on the absorber's environment, in the presentapplication (flame photometry) the first assumption is appropriate. Thesecond assumption clearly presents no conflict at very low concentrations(parts per billion (ppb) or smaller). But just as clearly, every elementor absorber must exhibit a concentration level at which this assumptionbecomes entirely untenable. Indeed, positive and negative deviations fromthe Beer-Lambert Law expectation over a wide concentration range are quitecommion; the interactions gain sufficient magnitude to significantly affectthe "constancy" of the absorptivity. Thus a realistic expectation for theshape of the calibr-ation (or standards) curve is not a straight line fromthe origin out to the highest concentration of interest, thereby requiringbut a few (two or three) reference solutions spanning perhaps three orfour orders of magnitude in concentration. Rather, it is a linear portionin the dilute region and a concentration level at which decided curvature

* sets In, thus requiring groups of several (ideally five) referencesolutions where each group spans only one order of magnitude change inconcentration. An empirical standards curve for cadmium is representative 1of the situation and is given In Figure 2(A). The feature of note is thatthere is a quite good linear correlation up~ to about one ppm, butthereafter, the relation is indeed a standards "curve."

What is essential for the analysis is that the unknown sampleconcentration be br-acketed by reference solutions. The remarks of the

* preceding paragraph are intended to provide insight into the number and* value of' reference concentrutions needed to achieve an analytically useful

2-6

Page 20: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

FIGURE 2

4 Generalized Calibration Curve (data for Cd).0

+ = CRDMIUM STRNDRRDS; 7/15/83El = . 1 -< (CD) ( 1.0 PPM

LINERR REIRESSION:e . NRe. 999-, R=MO. 0iS!

* 7.31

cc-- . 111. - (A)

4.11-

* U 0

3.9n11 t..

RISDRBRNCE (MM)0

. Cadmium Standards (low ppb level)

+ = CRDMIUM STRNDRRDS; 7/15/B3

0 " W.C. RECOVERY; 7/15/B3F4. 08-, -(X-E85 VS Y-TRUE)

- LU~~CRRELATICIN1 'CUFF= .9e0A= -C.001(B

B=. e0.08

AB. 04.1

6 ,PFEII RE3 -I DUFiL' u267. 100 0.010 0. 011 U - .0Sic

211 -: 0.020 C-. 020 - 0. 00 c1 . 00.030 0.030 -0. 000

S .00 -401 0.040 -0. ONCO.lUk .050 0.049 l. 00 1

U. 14. 500 0.080 0 . 0 7 0. 00 1* 1-0.000 1.183 0. 11 -1.. 1

7S[ In. 1, NO

R851R3RNCE (MM)

2-7 S

Page 21: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

calibration curve. For example, when one is working in a region of theah.nrbance function which as been found to be inherently linear, i.f .,less than one ppm for the metals of current interest, the number of -bracketing standards may be taken as a minimum of three, spanning anapproximate order of magnitude change In concentration. When working in anon-linear region, concentration intervals must be made sufficiently smailto achieve practical linearity or else a non-linear curve-fit must beused.

?.6 The Instrument.

The Varian model AA-775 is a thoroughly modern atomic absorptionspectrometer offering several capabilities that facilitate routineanalyses for trace metals. There are input and output ports forsimultaneous operation of an autosampler and chart recorder and digital 0printer. Benefits from these automating accessories accrue both to theprecision of the analytical method and to the flexibility of theoperator's time. An internal microprocessor provides, among otherfeatures, an automatic calibration curve based on up to five referencesolutions and subsequent direct output of sample concentration.

The automatic calibration curve can, for highly controlled samples,offer substantial time savings in the conversion of absorbance toconcentration. However, its use requires a judicious choice of referencesolutions since there is no "automatic" correction for any inherentcurvature over the range of standards used. Consequently, its range mustbe limited and its practical utility as the sole output device for 0unknowns of highly variable composition is questionable.

The sample Introduction process is an important part of the AAanalysis and can be conveniently described in terms of a "sampling time,"the period over which a sample is aspirated, and a "settling time," theperiod between successive samplings. Adequate sample time allows: the Sestablishment in the flame of a steady-state condition reflecting the trueanalyte concentration; and sufficient time for an integrated response tobe made by the instrument. Adequate settle time allows: the previoussample to be cleared from the nebulizer capillary; and the instrumentresponse to return to its baseline. Efficient use of time requiresminimization of these two parameters; however, they should not be made tooshort and the same set of values for samples and standards alike Isdesirable.

The Varian model 51 Autosampler provides a highly reproducible meansof regimenting the sample introduction; and the sampling and settlingtimes can be varied to optimize the sensitivity for a given analyte andconcentration range. The sample time should be three to four times theintegration period and the settle time should be equal to or greater thanthe sample time. We have found that a sampling time of about eightseconds (with integration periods up to two seconds) provides adequateresoloit.ion of concentrations down to about 0.1 ppm for the metals ofinterest here. At lower concentrations sample times as great as 1? to 1 4seconds have been necessary.

L I:.

Page 22: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

The actual time savings that results from the use of an autosam'perdepends on the number and nature of samples (unknowns, standards, andquality controls) and is difficult to quantify. However, we can becertain that an automatic device will perform the sample introduction morereproducibly and no slower than a person, so that whatever the timerequired here, it does not involve personnel attendance; hat. Is,personnel may engage in other analytical activity (e.g., tandardspreparation, paperwork, etc.) while sampling proceeds unattended. Thefull (50 samples) model 51 carousel may be 3et up in about 10 to 15minutes and will cycle in about 111 minutes at a combined sample and settletime of 16 seconds, or in about 24 minutes at a comb ined sample and settletime of ?8 seconds. The individual sample cup volume (six t) issufficient for three to five metal analyses, depending on the sample time,so that reduced manual handling of the aspiration tube minimizes theopportunity for sample contamination.

There are several advantages to using a chart recorder as an outputdevice. Principal among them are: such output affords a more accurateestimate of the noise level and hence the true signal strength; any drift

rn- .hange in the baseline (zero calibration) is more readily evident and*,i('countable; the shape of the absorption peaks provides a useful guideIboth to the adequacy of the selected operating parameters (specifically,the sample and settle times and the integration time) and to the need formaintenance activity; and, the full scale Recorder Span (mV) and theabsorbanee assigned to it can be selected to allow collection at one timeof a wider range of data than could be effectively used by theautocalibration feature alone.

For highly controlled samples, that is, for unknowns whoseconcentrations all fall within one linear region of the calibration curve,uie of the internal autocalibration feature with a digital printer and

-, chart recorder can realize appreciable timesavings for data processing. pF)r example with cadmium (see Figure 2A), over the range 0.1 to 5.0 ppmthe calibration curve is non-linear, but this interval nay be divided Intotwo sufficiently linear ranges (0.1 to 1.0 ppm and 1.0 to 5.0 ppm), onlyone of which at a time could bc. used for the autocalibation curveoittputing to the digital printer. On the other hand, the dynamic range ofthe chart recorder at REC RANGE z 0.5 A and Reconder Span = 500 mV willallow calibration data to be collected over the whole range (0.1 to 5.0ppm). The internal calibration curve may be used to process unknowns in,sty, the lowest concentration range; the digital printer will then outputv°lues for all unknowns in that range directly. The chart output can thenhc used to 'manually' determine unknowns that fall outside the automaticcalibration curve. Obviously, for those cases when all unknowns are within1he range selected for the Internal processor, there is no 'residual'processing to be done by the operator.

rhe full scale recorder deflection in absorbance units (REC RANGE),tr:'n recorder sensitivity in mV (Recorder Span), the recorder time constant'F<'C FA3T, REC NORM, REC Ek I D), and the integration period (TIME SEC)

2-)

Page 23: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

are -l1 chosen in consort to achieve adequate Mignal to noise at th,-desirec sensitivity. Signal to noise (S/N) ratios of' less than tthree ofour are of marginal quantitative use.

!he treatment (-.f the recorder output is straightforward. Rasel i-ripand peak height. lines should be drawn parallvl to th, me axis of the.cart and nThould be nl-ced In the "middle" rf their , t cive noiselevels, that. is, to 4ha.t equal excursions ff magnit'. e ind lurntion x:curbot - above arid tbcow the line. When the baseline Is very stable, oi;c linemay be drawn from the beginning to the end of the sequence, otherwise it 0should he drawn hetween the average noise levels on either side of eachpeak. When the baseline must be drawn for each peak, it will likely notbe parallel to the time axis, and the height must be measured from themidpoint of the peak base width.

The noise level (sensitiv ity) can be affectec, by excessive turbulencein the burner compartmient. in this connection, the exhaust system draftn.y he too vigorous. The draft. is simply controlled by closing the stackcamper in, say, Ml to 1r degree increments until an optimum draft isohtained. Thi.- 1o-ocd V. a one time adjustment for the current exhaustsystem configurat ion.

When or- encounters unexpectedly high unknown concentrations whichare off-scale at the current recorder setting, one may choose: analternate, less sensitive analytical wavelength; reduced sensitivitysettings (i.e., increase RF:( RANGE or the Recorder Span); or a rotation ofthe burner head to yield an absorbence reading of less than 1.0 in themost concentrated sample. In any case the change should be made at theenn of the initially programmed sample sequence, and standards must be rununtil the uknowns are bracketed by at least three references forming alinear segmert of the calibration curve.

2.7 Maintenance.

The cleanlin.. ,f the burner head and nebulizer assembly hasprourourced influe,e or, the noise level and thus the sensitivity. 'hepresence of (carbon) 'eposit: in the burner slot is evident byirregularities in the Peak-line of the inner flame profile. Smalldeposits are easi ly remn)ved, after turning off the flame, with an unwantedbusiness card or a soft metal (e.g., copper or brass) strip. A good Spractice at the end of a sample run, before shutting down the instrument,is to aspirate 25 to 50 mL of deionized water as a flush for the nebulizerand spray chamber.

The burner head should be dismantled periodically for a thoroughinternal .crubbing, with soap and water. A plastic domestic dish scrubberor onp of the copper mesh (not steel) variety is appropriate for removingencrustation inside the burner. Hard metal implements (e.g. screwdrivers,raor t -ivies, vte,. ) must not he used for cleaning; also, acids or other!ir.t cheiical clean 3ers or abrasives should not he used. Take carewhile cleaning not to immerse the interlock socket. The spray chamber

106

S

Page 24: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

should he removed periodically and cleaned with soap and water. Theaspiration capillary hose should be replaced whenever it becomes cloudy orotherwi:;e visibly encrusted on its interior walls. When fitting the freshtubing one must take care that the (sharp) metal edge of the nebulizerentrance tube does not gouge the wall of the capillary hose and therebycreate blockage.

There are only thr.,e optical surfaces that require regularinspection, and cleaning as necessary: the hollow cathode lamp w!ridow andthe entrance and exit windows to the burner compartment. '"hese surfacesmust be free of fingerprints and dust or other smudges and debris. Theymust be cleaned only with quality lense paper and optical grade solvent,e.g., spectrograde ethanol, methanol, or acetone.

Gas supply hoses must be inspected regularly for both physicalintogrity (major cracking or deterioration) and secure fit. Whenacetylene supply tank pressure gets below 100 psi, a fresh supply shouldbe fitted. Below this pressure the acetone used as stabilizer in the tanl<gains sufficient concentration in the gas supply to pose both a serioussafety hazard and an increase in flame Instability.

The autosampler sample cups must not be allowed to dry out when theycontain sample; they should be rin.3d several times with deionized water,shaken out and allowed to dry secure fmm contact with metal containingapparatus and particulate fallout.

Gas supply filters (whether for fuel or oxidant, particularly air)should be changed according to the manufacturer's instructions or whensubstantial deviations from usual S/N levels are noticed. The replacementdate of such filters should be noted for future reference. This presumesthat the reduced S/N cannot be traced to some other alignment or matrixproblem.

".R Instrument Alignment.

Instrument alignment response and analytical performance are asintimately tied to the care in setting up the instrument as they are tothe quality of the reference solutions used. Diligence and patience arethe keywords here; an instrument may be stable at less than optimumperformance. While manufacturers make every effort to make operation asuser friendly" as possible, one must remember that the ultimatefriendliness is only achieved through familiarity and practice with theexercise of the key operator qualities. (This Is particularly importantto the adjustment of the burner position and flame composition.) To seti p thro instrument, proceed as follows:

(a) 'heck that gas supplies and lamp current switches are OFF.

b) Power O'I; check that the following conditions have beeni-Oomatically selected:

2-11

• o. .. , . ___-____ ____ ___, __

Page 25: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

ARSorbance operating mode;INTegrate REPeat readout >de;Background Correction OFFRECorder NORMalRECorder RANGE = 1.0 (absorban,, unit)TIME SEConds = 1.0 second)

NOTE

C'AL .T1:10 Ii~ht wi ll be flashing; both display and rwterindications are meaningless.

(c) Fit burner; connect iriterlock; exhaust ON; gas supplies ON.

(d) Fit lamp; adjust current to specification.

NOTE

The current give, on the ipecification card accompanyingeach lamp i: 'recommended' (not absolute); somewhat highercurrents rrry improve the signal to noise ratio and therebyafford grpater sensitivity. For example, operation of theiron lamp a! 11) mA showed signficant improvement in thesignal to nni.'e ratio for the analysis of this element. Thecurrent selected must not exceed the maximum specified.Once the d,.esired current is selected, the lamp should be

allowed to warm up for a period of 15 to 20 minutes, in theinterest oF rnximum stability and minimum noise at thechos4n setting.

(e) Adjust the burner vertical and horizontal positions so thatthe lamp light spot, intercepted at the midpoint of the burner slot, is

approximately tmo to four mm above the slot and centered. Ad iust theangular position so that the light spot is centered when alternatelyintercepted at the ends of the burner slot. These adjustments can beconveniently made with the use of a business carr' for example.

(M) Set the monochrometer wavelength and slit at the valuesrecommended for the concentration range of interest; select Single Beam

(S9) and adjust the gain so that the meter reading is about half scale.Using the gain adjustment to keep the meter reading on scale, max imize thesignal by fine adjustment of a) the wavelength selector, and b) the lampfocus, with the two knurled thumbscrews on the sector base..

(g) Select. Double Beam (DR); light the flame.

At this point tne coarse alignment of the instrument has beeneffected; fine tuning is achieved (in DR) as follows:

(a) Aspirate the matrix or analytical blank; select CAl. '7"1

2-1?

Page 26: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

(b) Aspirate a solution known to give A : O.14, whileadjusting:

(1) Horizontal, vertical, and angular burner position.

(?) Fuel and oxidant flow.

(3) (1]ass head position (nebulizer).

(4) Burner vertical position again.

NOTE

A RSAnalyte ppm giving A1S0.1 (best observed at given ppm)

Cd 2.0 0.5 1qCr 10.0 0.504Fe 7 0.480

Ph10.0 0.400NOTE0.

Steps (1),(P), and (3) above, are all cruciallyinterdependent; here, perhaps more than any other step inthe procedure, diligence and patience must prevail. Theflame configuration, physical position, gas volume ratio,and nebulization rate will 'make or break" the AA analysis.

Step (?) represents the optimization of the "flamestoichiometry" or composition, which will be determined bythe nature of both the analyte and the matrix. For thedigestion process prescribed by the US EPA, i.e., for agiven matrix, the identity of the analyte will be thedeciding factor. Flame composition for best performance isgiven for each element as either oxidizing (lean), reducing(rich), or stoichiometric (intermediate). Oxidizing flamesare the hottest and appear non-luminous white with a slightbluish hue in the body of the flame profile. Reducingflames are the coolest and appear luminous with red to pinkhue in the flame profile. The exact composition of theflame must be found during aspiration of the appropriatestandard solution; a previously achieved set of indexnumbers (as shown on the AA-775 control panel.) shouldhe taken only as a guide.

Properly installed, the glass bead will be about one mm fromthe nehulizer orifice and its adjustment should be small,less than one tull turn of its adjusting screw. Thea11justment should be made for maximum absorbance andstability. The effect seen from this adjustment can be

2-13

Page 27: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

quite significant particularly If the setting has been mov,,Igrossly cxit of range. The best sensitivity his been foundfor, sample uptake rates of three to six mL/minute. [6]

Since the adjustments of steps (') and (3) will have .i major effect or! thevertical geometry of the flame profile, step (4) is P:;sential as the last,step in optimizing the "flame configuration".

(c) Aspirate the ratrix or analytical blank; select kCAL ZERO;aspirate the optimization standard to check performance.

(d) Select an ASorbance EXPansion factor (1.0 usua]ly) and an

integration period (TIME SEConds 1.0, usually).

NOTE

The Pxpansion factor will only affect the digital display;it is not presented tr) input-output ports. The factoraffects both the sigrl anil the noise; to be practicallyuseful factors greater than one must be accompanied bysubstantially increased integration times and consequentlygreater s1pmring times. For this reason a preferred scale 0expansion technique is outlined in the final note underRecorder Setup.

An integralion period of one second will provide arelatively noise-free signal for metal concentrations downto about 0. ppm for all the metals of interest here. SLonger periods many be used to smooth out the noiseinherently encountered when working at the tens of ppblevel. The only caution here is that the sampling timeshould be on the order of three to four times theintegration period.

-. q Recorder Setup.

Power/ON; Select full scale (f/s) = 1 V; Select standby. Allow towarm up, for examplp, wile setting up the AA-775.

.elect INT pryP, T11:: SEC = "x" (x greater than or equal to 3).

,elect BAP GRAPH; when pen comes to rest, you have "x" seconds to se -the baseline with the ZERO POS potentiometer (10 turn). If you miss,simplv- select bar graph again.

.Select RE HAT;:'; wi Ii deflect to or near f/s; you have as muchtime as you please to ,it the pen to exactly f/s with the RECORDER ADJUST.potentiometer (one turn). When completed, go on to the next step.

2-1)1

- S

Page 28: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

. -

NOTE

The number in the digital display after selecting REC RANGEwill be the f/s deflection in absorbence units and at thispoint has no bearing on the zero and f/s positioning of thepen.

Select REC NORH, K' 'AST or REC IDMPFD.

NOTE

After selecting, for example, REC NORM, the pen will come torest at a position determined by the status of the CAL ZEROfunction. That is:

(a) If r AL ZEO .:- assigned (LED on) and the AA-775 isaspirating the assignent., then the pen will rest at zeroposition established above.

(b) If CAI, ZERO is assigned and the AA-775 isaspirating anything other than the assignment, the pen will •rest at a position up- or down-scale from the originalselection according to the difference in absorbance betweenthe assignment and what is being aspirated.

(c) If CAI, ZERO is not assigned (LED flashing), thepen will behave as based on the last previous assignment.

The assignment should be whichever (matrix or air) has thelower absorbance, most likely air for wastewater analysis.When the difference in absorbance between matrix and air issmall, the baseline position is not critical. When thisdifference is substantial, as for lead, the baselineposition established must be able to accomodate the penexcursion due to the air aspirated between samples. For Pb,if the assignment was air, pen moves up-scale whenaspirating matrix; if the assignment was matrix, pen movesdown-scale when aspirating air.

Select the desired recorder SPAN; a decrease from f/s = 1 V may,quire a small final adjustment of the baseline position with the ZERO'OS control, unless the recordgr functions are perfectly,a lanced/alibrated.

NOTE

The choice of necorder SPAN in concert with the choice ofRF( RANGE determines the magnitude of the pen displacementf'or a given analyte absorbance value. For example, if thepen response to a given absorbance value is, say, 14 mm at

2--]0

Page 29: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

PPC ANGE = 1 A arid recorder 'PAN 1 \V, then nominally t:esame (14 [m) response can be achieved with r' RANG A. Aand SPAN = c.lo mV. By the ,same token, an effective two-foldamplification of the "readability" of the recorded 7ip,nal(i.e., 28 m) is achieved either by halving the ! "AN at theoriginal REC ANGF, or by the converse; if both i:tialparameters otc: clividea by two, the enhancemeni i, four-foli(i.e., c~r, is). ~temember that such enhancement:. apnlv toboth the signail level and noise level; however, we navefound emirio'lcly with the present, system that inerease'; inrecorder nerunitivity (decreased .2PAN) generally tend toproduce somewhat higher noise levels than the corresponwingchange in the REFC RANGE. The relevance of these comments tothe above note is that at analyte concentrations above oneppm settings of PEC RANGE = 1 A and SPAN = V are quiteadequate and render nominal the difference in absorbancebetween matrix and air', thereby mitigating the choice ofbaseline position. Also, at concentration levels below oneppm wiere more sensitive settings are required, a judiciouschoice of hoth parameters may eliminate the concern withbaseline position.

2-1'

... 2*-

0

.0. , ..

Page 30: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

'0 RFSILT iThe data and analyses given in this section are based on the

techniques set forth in section 2.r, Discussion. During the summer weekswhen the daytime, ambient laboratory temperature was consistently above o)deg F (3P' deg C), all secondary stocks and glassware were allowed toequilibrate at the temperature of the instrument room (approximately '6

deg C), and volumetric preparations were performed therein. Flame methodswere used without simultaneous use of the background correction(deuterium) lamp and ahsorbance was outputted to the Varian Model 0176chart recorder. Peak heights were measured to the nearest 0.1 millimeterand a linear regression was performed (using a Hewlett-Packard 34C handheld calculator) as concentration versus absorbance on typically four orfive points spanning not more than an order of magnitude chanrre inconcentration. When necessary the end-points were varied to achieve alinear correlation coefficient of not less than 0.09. Additional programcode was entered in the calculator to allow a determination of thestandard deviation; for the regression slope, regression interceapt, andthe interpolated unknown concentration. (The formulae on which theadditional program code is based may be found in the appendix.)

The data in Table T is a representative, abridged listing of an

original series of unknowns analyzed for chromium using an oxidizing airand acetylene flame. The certified concentration of chromium in "QC-'6" is:true value, 21 0 pph; reported 242 + V, ppb. There are two sets of data inTable 1: the left three columns result, from instrumental conditionsnosted at the table foot; the right three columns result from samplingunder identical conditions except that REC RANGE = O.r and Recorder Span1 'J (i.e., lower sensitivity). Table II is a listing of statistical datao)-.tained by linear regression analysis of the data in Table I.

A.,; for the QC sample results, the values when derived from standardsn, mbered one through seven are within the acceptable range, and theprecision of their measurement, in terms of the relative estimate ofvariance (s , l*qs/X), is below the maximum value (eight percent)e-tablishei for chromium by AEHA. [2] That the amount detected is, inact, i the low part of the acceptable range may be attributed in largene,,-tre to m atrix mismatch between standards and sample; particularly the

relatively high concentration of iron (000 ppb in the QC sample, along4ith other metals). That the QC results, when derived from standards

inhbered six through twelve, are outside their expected range and low isattributed to extrapolation error and to the negative curvature in thecalibration curve (c.f. Figure 4). S

For samples whose concentration is less than RO ppb, the results are6-iven as such since standards below RO pph were not used during the sample

"-1

*. " .S'-'" ; ; " " ' "'i"; . . .- ; - ; ', i . . . " .

Page 31: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

oG

1 '.0 'a0+1 +1 +1

- ' 0

.4) 0

0- 0 N O cu %0 UN &A Go s-

0 q4j

v- - A - % C t-'0 I I -- 00 U

00 - N9 %o I '0z0 0 ' -E4*u

w m CA4 , v V

'- -4

*6*

8 +1 +10 9 +0c

m I -.0 .o0 .94 4. 05. 0 0 0 0 0 0

16 0

00

N0 N 0

NA U; a LA 'C t-

00

Page 32: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

0 .. aa ~ NI 0% en 'd

06J

0 a r cm ao -4)

~ N UN

-4

9-1~~ 0c 9-N

r-4 .4 >1

1- c-en.o g cm . c I n . - 6)C

12 AT In Go8 NV

4) 9-4

0 4)0

%D %V .4 r t % % oI

0 00

49~~~~ 0-N*0004

060it OD r LD

-~ VY KY C

IS x c

3-3

Page 33: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

TABLE II

Statistical Evaluation for Table I Data

.

RtAta Pairs (Conc. Range) 7(0.08 to 1.0 ppm) to 5.0 ppm)

Slope of Conc. vs Abs. 0.016004

Variance on Slope 0.0003 0.0000

Intercept -0.039 -0.235 :

Variance on Intercept 0.012 0.06h1..-

Correlation Coefficient 0.999 0.900

.40

3-11

Page 34: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

= - * - - - -=---. - -- -u .- - - ~ ----_ . - • co- , , - .- - .S * • ,o ... -° .- .,

run. However, there is sufficient sensitivity or readability under theconditions listed to detect approximately 20 ppb before the analyte peal<becomes lost in the matrix or background signal.

For samples whose concentration is not bracketed Iv standards onethrough seven, confident determination of the concent at.ion requiresbracketing standards seven through eleven, even though the extrapolatedresult is only about. 10 percent different. While one can expect thisdifference will become less as higher concentration levels are approached,extrapolation to lower levels leads to much greater uncertainties anderrors, and must be avoided.

3.1 Spiked Samples.

As mentioned in paragraph 2 of section 2.1, EPA Quality Control 0

Samples for Trace Metals, quality control should be effected by the use ofspiked samples. The spike can be a convenient aliquot of one of thereference solutions added to a duplicate of one of the unknowns; an equalaliquot of matrix should be added to the same volume of unspiked sample sothat the base or background concentrations will be the same, therebysimplifying the data reduction. The calculation is based on the

following:

where, Y [-]X = observed concentration of the analyte in unknown sample duplicate;Y = observed concentration of the analyte in the spiked unknown sample;S = true concentration of spike as milligrams added per total volume of

spiked sample; andC = corrected concentration of unknown sample.

The quotient in brackets above represents an enhancement or

suppression factor which will be applicable to correct otherdeterminations of the same metal under the same conditions at the sametime (i.e., during that sample run). This factor is determined ad hoc andis transferable neither from day to day nor from analyte to analyte.Acceptable values for this factor are in the range 0.5 to 1.5 (i.e., 501fenhancement or 50% suppression); factors outside this range indicate aserious matrix mismatch which must be corrected. The correction need notbe applied to the unknown sample determinations if the factor is in therange 0.9 to 1.1.

?.P Sampling Sequence.

A recommended sampling sequence is: three to five standards, twoneat quality control solutions (EPA certified), duplicate unknown, spikedunknown, five to six samples, standard number one, five to six samples,standard number two, and so on, alternating five to six samples with oneof the initial standards until the calibration curve has been reproduced.

3K ..... .

Page 35: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

* - . ...........- U a YTVWV.'a.

0

* End the sequence with a known (QC, spike or standard). If alA of the* subsequent responses for the standards are within approximately l')% or* their initial value, there is rv problem.

0

9

S

I ~0

6

S

4 S

* 0

* S

3-6

* 5

* - . . * . . . - -. *. * * . * -'."-a

a - * - . a * * * * ......... * * . -j.*... a.. . * . *j. .. , )1 . - *..~ - -. **,- **'- ~"- :. *.

Page 36: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

4.0 CONCLUSIONS

Establishing a limit of detection, while straight-forward, is notquite as simple, procedurally, as extending one's calibration curve to the"lowest concentration level for which signal response- can be distinguishedfrom noise. Statistically meaningful quantitation cannot be made at thelimit of detection [7 As a functional rule of thumb, detectabilityshould be at a level at. least 50% below the desired level of quantitation.For example, if one desires to determine an element at the 0.1 ppm level,detection should be evident at the 0.05 ppm level. Figures P(R) through6 are presented simply as empirical evidence that under the proceduresoutlined herein routine detectability with high confidence can be acheivedfor the metals of current interest. A definitive determination of theLimit of Detection must await the results of subsequent experiments 5designed expressly for such purpose.

There is currently sufficient capability, and facility, to detect therequired maximum allowable level of cadmium, chromium, iron,and lead inRIA effluent by flame atomic absorption methods (c.f. Table III).

(a) Without any special procedures, the detection limits forcadmium, chromium (both tri- and hexavalent) and iron are about an orderof magnitude lower than are needed to meet the present effluentspecification.

(b) With special procedures, the detection limit for lead in 5samples can be brought down to the current required specification (100 ppbmaximum permitted). Special procedures could be the concentration of thesample during the acid digestion procedure and flame AA analysis, the useof electrothermal atomization (graphite furnace), solvent extractiontechniques, or combinations of these.

NOTE

Maximum care in the preparation of referencesolutions is essential.

(c) Reference solutions (working standards and certified '.PAquality control samples) must be prepared in acid soaked glassware and

stored under refrigeration.

(d) Lack of adequately climatized (thermally stable) environmentfor reference solution preparation will introduce significant error andnon-reproducibility in analytical results, particularly at theconcentration levels needed for effluent analysis.

NOTE

Maximum care in the construction and use ofcalibration curves is essential.

4-1

-ji " . ' .• _l " , - , .. - . "i. , - . .• ,

Page 37: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

I . *1(' S

. Chroaium Standards (low ppm level).

+ C CHROMIUM STRNDRRDS; 7/25/838" N1G.C. RECOVRY

CX-CJ9S. V5 Y-TRUC)i. ll~m0. 0-8'

. (A)

-. I 1.21 E - 8C:C RRE L. AT 10I .i1" I::IEFF= 0 ,.

0. 0 14 -7 ,e in, '',PRED RES; I HiLAL4. ;0 0- OC5 0 0.C5 l.c

9.0 .CIO 0.1112c -0. 01215. 100 0. 200 0 19 0 .00422,. .,,:330 -0. 000

LU I. Il 21.1

RB5RBRNCE (MM)

+ & CHROMIUM STRNDRRDS; 7/25/B3.- =.C. RECOVERY

(X-095. VS Y-TRUE)7. 400 0.078

(B)

iERRELFITI01 C :-EFF= . :9:,._35C3 M.ggA 0.0 16

O / U iPREr FD E':1U1AL

4.0.05 0 ', 4:'"" 0. 007C2 9.21.00.0 0 1101O, 111 -.0.0 1 1

15.100 0.200 0. 20 0 -. 004.. 900 0.400 0. 405 -0. 0 0 5

" . " 0 . " , -- ":'. _I ,,- -0.500 H. 4:5 0. 015

5 .20 0 0 .8 0 0 0 . '-0 - . r 1 '7I*. E.i . 200 0 1. 000 1 H.0 --0. 08

RISORENCE(MM)

4-2

*P "I.-:--i .-;> ::ii.-:-:~~~~~-...--.-! -- .i --. --:iL - '-- - --...

Page 38: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

FIGURE 4

* Chromium Standards (for data in Table 1).

+ - CHROMIUM STRNDRRDS; B/1/63(DATR FROM TRBLE I/ LEFT)

F3 U .C. RECO\ERY(-05E5. V5 Y-TFRLE)15.4

a:CORLTO C O_.FF= 0. 99887

*.I.sI.= 0.016 (A) RR;L

= == B=-o~o,_,9.-oil.~~~0 0:39 E.5 R

6.0 0 .080 0.07 0t-.010

2.7.00 0.400 0.404 -0.0043200 0.500 0.473 0.027

51.800 0.800 0.789 0.01166.000 1.000 1.016 '0.016

RBSDRBRNCE (mm)

+ - CHROMIUM STRNDRRDS; /1/83(DRTR FROM TRBLE It RIEHT)

0 = LINERR REGRESSION0.Q = I.C. RECCVERY p

(X-0BS. VS Y-TRUE)

SCORRELATION COEFF= 0.99931R= 0. 044

4.= B= -0. 235

SY y YPRED RESIDUAL21.900 0.800 0.729 0.07128.200 1.000 1.006 -0.00651.80E 2.000 2.046 -0.04675.200 3.000 3.076 -0.076117.600 5.000 4.-943 0.0 57

mm l~ II IxI 171.IRISDRRNCE CMM)

4-3* S

Page 39: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

FIGJE 5

Iran Standards (low ppb level). .

+ = IRON 5TRNDRRD5; 7/ZS/ 53

Q = .C. R LV r-v ¥ -

to(X-LBS. VS Y-TRUFj)

6.500 80.000Z (A)

CORRELATION COEFF= 0.98354+ A= 14.260

B= -26.073Y =AX +B

x YPRED RESIDUAL

+2.800 10.000 13.854 -3.8543.000 20.000 16.706 3.2944.100 30.000 32. 392 -2. 3924.200 40.000 33.818 El.1825.700 50.000 5,5. 207 -5.'2077.300 88.0880 78.023 1.977

3.n 4.1 i.10 5.B 7..

fRSSnREfNCE CNN)

Iron Standards (low ppm level).

+ -I RON STRNDRRDS; 7/IS/B3

El O.C. RECOVERY(X-095. VS Y-TRUE)

Si.4 1 .3808070. 00.. 88

(B)CORRELATION COEFF= 0.99993

X Y YPRED RESIDUAL12.080 0.100 0.102 -8.08022:3.00 0..201 -0.00133.680 0.380 0.296 0.00444.800 0.400 0.396 0.00456.500 0.500 0.502 -0.00279.300 0.708 0.706 -0.806111.708 1.000 0.997 0.003

* .1 Sl.3 531 N.1 I l1.1

RESDRBRNCE (NM)

4-4

S..

Page 40: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

LL) I

IL

Ii j T. D- Ji, IT jjM.

td -71, ti- a r - -, .F

''4 r LL . '. . ..

'IL

M EN

cc LL

zzw cr

Lfl L-

D-3n

I- La X

(Wnd a011UINDa4-

Page 41: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

TABLE III

Current Limits for RIA Effluent (ppm)

Element (1) C Max. (2) LOD (3) LODA

Cd 0.15 0.0005 0.010

Cr(III) 1.0 0.005 0.05Cr(VI) 0.3 0.005

Fe (Total) 2.0 0.003 0.01Fe (Dissolved) 0.5 0.003

Pb 0.1 00.01 0.2

(1) City of Rock Island maximum allowable.

(2) Limit of Detection, from literature (c.f. reference 10).(3) Limit of Detection attempted here.

S

4 40

4-

4-6

. .. ''" . .

Page 42: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

(e) The determination of unknown concentrations must be* bracketed by the determination of known concentrations.

(f) The actual mgnitude of the reference concentrations and theintervals between them will be determined by the behavior of thecalibration curve in the region of interest.

NOTE

Quality assurance procedures must be applied toboth reference solutions and analysis of unknowns.

(g) A neat EPA certified quality control sample, prepared tospecif ications, is a hona fide control for reference solutions, and inspecial cases can be also an adequate control for the analysis ofunknowns.

(h) A spiked unknown, with unknown duplicate, is in all cases abona fide control for the analysis of unknowns.

Automated sampling and output accessories, when used routinely,facilitate more reliable or reproducible data and better use of personneltime.

* IP

4-7

* I- .-, . ., , ..- . . , .,

* , . . °

.-.* "- I- "

Page 43: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

5.0 GENERAL FRCOWENDATIONS0

A certified EPA quality control sample, one for each concentrationrange of interest, should be analyzed with every set of unknown samples.

(a) Results should be maintained only for as long as a given setof secondary stock solutions is in use.

S(b) When secondary stock is replenished, the mean recovery and

variance may be computed and subjected to control charting procedures ifdesired.

A spiked sample with duplicate unknown should be included in everyset of samples analyzed. The spike result need not be retained beyond the Sdata reduction stage for the sample set of which it was a part.

Analytes should be determined at a level of 50% below the requirementand results reported to two significant figures or simply as less than therequired limit.

The Hewlett-Packard Model 82905-B Digital Printer should be acquiredand installed as an additional output device for the Varian model AA-775.

Volumetric preparation of reference solutions should be consistentlyperformed in an environment where the temperature is consistently as near20 deg C as practicable. S

5-1

* p61

Page 44: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

4

.0

I

U *1

S

I

I

JS

* 0

4 0

4 0

I S

I S

- . . . ., ..

Page 45: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

.0 CONTINGENT RECOMMNDATIONS

If the current analysis load is subject to increarp, in term., oreither the number of samples per week or the nunber of analyses per sampleor hoth, consideration should be given to the permanent placement of' atleast a one-half time laboratory assistant for maintaining glassware.

If the results f, m daily effluent sampling sites are to he

correlated with the operation of any production function (e.g., for thesake of optimizing production procedures), that is, if daily unknownsample results are to be subjected to control charting procedures, thensteps should be taken to establish computer collection, processing andtransmittal (in digital or hard copy format) of AA data.

0I

6-I

4 p

"* ' " 'S .. : :

F -" ...: : .,-" k - :.i ":

Page 46: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

a - - -- -. '-j -' - - .. .-. *7. -. - -. -. -' ~-., ~ -.-..;-~ X~L~<-r -~--V-~-ZS--w--~ -WV.--,----j------- - - .... -~ -~- -,-----r----z-*-w - rrrw -.-----~- w~

4 0

-0

0

S

*

0

S

I* 9

r *1*

* S

C S

* - - . ii

Page 47: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

7.0 REFERENCES

[11 AA-775 Series Atomic Absorption Spectrophotometers: OperationsManual; Varian Techtron Publication No. 85 - 100101 - 00.

[21 US Army Environmental Hygiene Agency, REG 7O-!; November, 11,9?,Append ix-H.

131 Analytical Methods for Flame Spectroscopy, June, 1078; VarianTechtron Publication No. 85 - 100009 - 00.

[4] Methods for the Chemical Analysis of Water and Wastes, March, 1170;EPA - 600/4 - 7 - 020.

[5] Slavin, Morris, Chemical Analysis, Vol. 25: Atomic AbsorptionSpectroscopy, John Wiley & Sons, New York 1978.

[ Price, W.J., Spectrochemical Analysis by Atomic Absorption, Heyden &Son, Philadelphia, IQ79.

[71 Parsons, M.L.; Major, S; and Forster, A.R; Appl. Spectrosc., 7(1983), 411-I18.

(9] US Army, AMCP 706-110.

[n] Ewing, Galen W., Instrumental Methods of Chemical Analysis, 4th Ed.,McGraw-Hill, New York, 1975.

[101 Kolthoff, I.M. et al., Quantitative Chemical Analysis, Pith Ed.,Macmillan, New York, 199q; p. 397 ff.

S i

7-1

Page 48: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

- 2*.. ': ~.-. ~ -. -~ - -. -. '~'~'~"7' -~- ~ -......... IiI S

S

.1S

S

'II S_

4 S

* S

.1

Page 49: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

APPENDIX

The variance associated with the slope and the intercept derived fromlinear regression equations are useful variahles for defining theconfidence limits for the linear relationship. The v.riance associatedwith the value interpolated from regression parameters is also a usefulquantity for defining the confidence limit on the interpolation. Thesevariances can all be computed from the same set of factors used to computethe slope, intercept and correlation coefficient (particularly regardinghand-held calculators).

The equations given here are excerpted from reference 3, chapter five.

Definit ions:

given: n, Zy, Ey 2, Zx, Zx2 , and Zxy

.Sxx) [Zx - (Ex) 2 /n] 2(yy) : [Ey2- (Ey) 2 /n]

Sq(xy) :[Exy - (Ex) - (Ly)/n]S():[y) -(xy) 2/, (xx)]/n-?)

the estimated variance of the slope,

s2 : S2(y)/.,(xx)

the estimated variance of the intercept,

2 = 2 (y) [(1/n) + (R) 2 /S(xx)] [52(y)/n 2 ] In + (Fx) 2 /'(xx)]

the estimated variance of a value, Y', interpolated from x

s- (X') .2 (y) [(l/n) + (X'- )2 /Srxx)]y

f.4

*__A

. .- . .. ". .

.. .. 0.". .. .

Page 50: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

- --N- - -

DISTRIRUTION

romwnder

Ballistic Research LaboratoriesA".]: A!4SMC-BLAberdeen Proving Ground, MT) 21005-506r)

CommanderRock Island ArsenalA" 'N: SMCRT-FY 1

qMCRI-FNm,MfRI-FNM-E 1

SMCRT-AP 1Rock Island, TL 'Il?90-5000

rommanderUS Army Armament, Munitions and Chemical rommandA.".: AMSMC-OA (R)/Mr. FerrRock T.sland, If. A1QnO-AO00

CommanderUS Army Chemical Research and Development CenterA!"N: SMICR-CLD (A)Aberdeen Proving Ground, MD ?1010

Commander*IS Army Foreign Science and Technology CenterATTN: AMXSr-SD1??0 Seventh St., NErharlottesvil le, "A ?201-5396

rommanderUJ5 Army Malteriel CommandATTN: AMCOA-E 1

AMCOA-P 1*SMmCR-rLB 1

AMGM "501 Eisenhower AvenueAlexandria, VA 2?3 3

fommanderre TUS Army Matick Research and Development renter

A7TN: .TRNr-E1Natick, MA 0170-900

. . . .* * . ... .* . . . ° .• ..t - . • . " . . . " . . * *" -" . ."* - *. . . ° " . - . ... * . .

Page 51: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

• I

Defense Technical Information Center Jo

Cameron Station, Puilding 5010 Duke StreetAlexandria, VA 2"3111

Director

ITS Army Industrial Base EngineeringATTN: AMXIB-MT/mr. BrimRock Island, IL 61299-r000

Dr. Theodore Towns2217 Flagler Place NWWasington, DC 20001

Dr. Joseph B. Morris, ChairmanChemistry DepartmentHoward UniversityWashington, DC 20059

0I

re • ,

*. : . .. : , -I . . : . - . . ., . : . - . . . , . : _ . . . , , . : . . .: : : - .. - .

Page 52: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

0 z 0

c I40 - 4-

-C I4 I - .

* -v -2.2AC C, Cl Eg

a0 LA Ao - A C 0 a c ~ 0

.4~ ~~ ~ ~~~ C 4 1 - c C 4 I A OA C- a - a

a a4 3 c 0 C-c A 4- . 4 I- - J S A 44 .

- M -N 1 4 C C 0% 0- N N 0 - C 0

v 0 I00n. tM V EM C V -- c

A~~~- CA C -CA A S -44 - "At

-0 A . uC4 4 4 - A .C a A -

* 0 Z44. CCA C A 0 .4 CCA CAo 4- -o 2,4 0OM 420 41 C0 44 'L 4 2 4l - - CC4

vi Li A . C 4 A J L A - SC~ A- C Lw-t A C A- C 4 n. 4 Ao a C - at- - v - C 0 c> a - - v . 0w

v nz.i a-440 0 - nA 0-4- - AU- '4L A0> - - AA J-A -t

A ~~~~~~ ~a o-. n C l 410A n 3 C

Li= wLA >4 '14.C 0i 3LA > - A

AZ-A - w12 Z4 A IC m. r C-

64> 0 4) L - ~ c ~ c4 0 4>)- C - 400

A 0 34.4 0% !t- A t- 61 14 A ,4 4 % - 4 -0 0- m 4-LC- c 0 0- -nI --

I 0 0 m-4 C-4 C -I 0 0 -- , O C

444 0i 0. MS

-4-2.4 ZLA - C CA C; 4-2. Zi A C 1A

4-444-~~~~ 40 -A>A C A 6 --- 4 - C A 4

2I cc C .2 0 UAM L 4- C m 4C 4A

0 A C0-

M. 0- a A- VAm-U i Cox --.

C0 6. C- 44

0~ ~~~~~ CE 1 0 0> C 4 44444~ ~ ~ ~~~~~ .- - 4 4 - 4 4 4

-~ ~~~~~~~~ 44> 4-C A A C0 0 - > .

-- I4 4- .4 61

Vc 0C wV Ac -- Ca-f - . 4 - 5 A5~>0 - I C L - Ci CCCL A- A A A - c L4 A O n.4 %-- A 4- A A 1- A a.041- A - A A - A 0. 0

C Ca AC C CA AAw 0C~ ~ ~~~~ 4 - -A v- 4 4 4 C l A -A ~ ~ ~ ~ < CA C - 0C - S -

Page 53: RD-R149 181 TRACE METALS ANALYSIS OF ...RD-R149 181 TRACE METALS ANALYSIS OF WASTEWATER BY ATOMIC i/i I ABSORPTION SPECTROSCOPV(U) ROCK ISLAND ARSENAL IL I ENGINEERING DIRECTORATE

FILMED

2-85

* DTIC