RASNIK middle scale test at NIKHEF
description
Transcript of RASNIK middle scale test at NIKHEF
![Page 1: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/1.jpg)
1
RASNIK middle scale test at NIKHEF
Michał Jankowski
Krzysztof SyryczyńskiSoltan Institute for Nuclear Studies
2005/06/08CERN LHCb Week
![Page 2: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/2.jpg)
2
RASNIK
PVSS
![Page 3: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/3.jpg)
3
C-frame at NIKHEF - March ‘054000 mm
6050
mm
![Page 4: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/4.jpg)
4
Installation
![Page 5: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/5.jpg)
5
Mask placement
Final position – inside the C-frame
![Page 6: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/6.jpg)
6
Telescope = lens + CCD
![Page 7: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/7.jpg)
7
Software
![Page 8: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/8.jpg)
8
DAQ software
Frame grabber DT3162 driver
RASDIM server(containing
Analysis240.dll library)PVSS client
MySQL database (external data storage)
Graphicalpresentation
of the latest position
(IN FUTURE)
makes bit-map from video signal
plots in a new PVSS panel
controls everythingcalculates x, y, z
![Page 9: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/9.jpg)
9
Rasdim server by Robert Hart
• RASNIK multiplexing
• Image acquisition directly from driver
• Analysis in external dll library
![Page 10: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/10.jpg)
10Main Console
PVSS client by Robert Hart
Smallchangesare possible for LHCb OTto remove detectors we don’t have
![Page 11: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/11.jpg)
11Sequence selection
for choosing in which order channels are read out
PVSS client by Robert Hart (user layer)
![Page 12: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/12.jpg)
12
Channel parameters window – for choosing channel by multiplexers
and for defining distances, CCD and mask sizes etc.
PVSS client by Robert Hart (user layer)
![Page 13: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/13.jpg)
13Plotting channel positions in time
Java RASNIK data analysis program by Michał Jankowski
x, y – movement ofRASNIK mask inits coordinates
scale = magnification =
mask) ens,distance(l
CCD) ens,distance(l
![Page 14: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/14.jpg)
14
Results
![Page 15: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/15.jpg)
15
RASNIK lines on the C-frame0 1
23
4
• 5 lines were installed
• data is stored in MySQL database.
line focal
length
f [mm]
distance
lens-mask
[mm]
distance
lens-CCD
[mm]
0 - 3 70 140 140
4 104 388 142
XLAB
ZLAB
XLAB= +Y0 +X1 +X2 -X3 -Z4
YLAB = +X0 +Y1 +Y2 +Y3 -Y4
ZLAB = +Z0 +Z1 +Z2 +Z3 +X4
YLAB
from RASNIK to LAB coordinates
![Page 16: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/16.jpg)
16
Temperature & Length changes
Temperature measured on top and bottom of C-frame
Length of horizontal parts of C-frame
Tdown
Lup
Ldown
Tup
We observe thermal expansion of C-frame.
![Page 17: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/17.jpg)
17
movements of C-frame corners against steel pillars during 60 hours
• On upper lines 0 & 1 mainly thermal expansion in opposite directions• horizontal movement in the same direction on bottom lines 2 & 3.
0 1
3 2
![Page 18: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/18.jpg)
18
Stability of the module – RASNIK results
Movement in Z of middle of long module in respect of C-frame during 8 days.
• movement <200 µm of free module, confirm laser measurements: see note “Module flatness”.• bigger movements with polystyrene bar (it was during laser measurement of C-frame flatness)
free module
module connected by polystyrene barto other modules
module disconnected laser measurement
![Page 19: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/19.jpg)
19
RASNIK measurements for period of laser measurement 9th April 1230-
1830
Z of middle of 5m long module type F. On line 4 “Z” is perpendicular to optical axis, it is measured with precision of order of 1µm.
Hsi
de
Lup
Ldown
Hm
idd
le
Z
![Page 20: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/20.jpg)
20
RASNIK measurements - weekend 30th April – 1st May
Z of middle of 5m long module type F.On line 4 “Z” is perpendicular to optical axis, It is measured with precision of order of 1µm.
No correlation between module bending andchanges of length of C-frame sides “Hside”.Thus no correlation between module bendingand temperature.
Fast changes of C-frame height in the middle
Hsi
de
Lup
Ldown
Hm
idd
le
Z
![Page 21: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/21.jpg)
21
Z and Y of middle of 5m long module type F.
see on plots:2 * (-y4) Hside 2 * 2.7m * |Al-module| * T 5.5m * |Al-Fe|* T’y4 is up, so Al - module Al-Fe coefficient of linear expansion of moduleis close to normal steel.
75 µm
150 µm
Hsi
de
Hm
idd
le(y,z)
RASNIK measurements - weekend 30th April – 1st May
![Page 22: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/22.jpg)
22
Range of RASNIK
on test lines magnification = 1 range +- 7mm easy in future on same lines magnification = 1/ 2.5 range +- 3mm
This 3mm range should be used for measurement of movement of the C-frame. Holes in rails and pillars for mounting telescopes should be prepared with precision of about 1mm.Tuning sharpness – now more adjustments than needed.Movement of only CCD in Z is sufficient. Movement of lens has little use, so “telescopes” will be simplified.
CC
D4.
6 m
m
seen
by
CC
D11
.5 m
m
mas
k 20
mm
1mm in the shadow of mask mount(small distortion of the edge of image)
per
fect
mas
k 18
mm
minimal magnification on our lines 1/ 2.5
1 2.5
“half” range 3mm
RASNIK measures better when full surface of CCD sees mask.
minimum:full segment9*9 of340µm squares(if 18*18 fields seensegment is seen)minimum 6.1mm
![Page 23: RASNIK middle scale test at NIKHEF](https://reader035.fdocuments.us/reader035/viewer/2022062322/568151a2550346895dbfd14b/html5/thumbnails/23.jpg)
23
Conclusions
• Installation of 5 RASNIK lines completed successfully
• Full surfaces of all CCDs see mask on all 4 lines.
• Precision σx σy 1µm, σz 100µm
• Telescopes will be simplified.
• PVSS client is user friendly, should be adapted to OT needs
• Test results show that 4 RASNIK lines monitor thermal expansion of C-frame with sufficient precision