Rapid HIV Testing:2003 Update
Bernard M. Branson, M.D.Chief, Lab Determinants and Diagnostics Section
Centers for Disease Control and Prevention
![Page 2: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/2.jpg)
Why do we need rapid HIV tests?
180,000 – 280,000 persons don’t know they are infected
40,000 new HIV infections per year
27,000 – 30,000 HIV positive tests each year at publicly-funded testing sites
31% of those who test positive do not receive their results
![Page 3: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/3.jpg)
HIV Prevalence, 1989-2000 Publicly-funded Counseling &
Testing Sites
0
1
2
3
4
5
1989 1991 1993 1995 1997 1999 2000
HIV
Pre
va
len
ce,
%
![Page 4: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/4.jpg)
What if rapid HIV tests were used
in all public testing sites?
HIV C/T sitesSTD Clinics
Drug Treatment Family Planning Other testing sites
Rapid test
1.9%1.6%2.9%0.4%2.1%
HIV+82.1%67.8%73.6%76.9%73.2%
97.0%
HIV-84.3%48.1%70.8%63.0%64.6%
93.0%
Return for Results Site
Source: CDC Client Record Database, 1995
Prevalence
![Page 5: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/5.jpg)
040,000 0 2,000,000
1,442,847 786,9371,147,25126,46522,5141997 1,147,25122,5149,843
27,947 1,374,7091996 708,2201,442,84726,4658,023
What if rapid HIV tests were usedin all public testing sites?
Additional, Rapid Test
Receive HIV Positive Results
Standard testing Additional, Rapid Test
Standard testing
1998 1,442,847 856,8761,147,25126,46522,514 1,091,04518,1189,843
1,442,847 903,8391,147,25126,46522,514 995,95218,7968,4441999
1,442,847 913,0571,147,25126,46522,514 1,028,27418,5038,3132000
Receive HIV Negative Results
![Page 6: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/6.jpg)
1998: PHS Recommendation Changed
Provide preliminary positive HIV test results before confirmatory results are available in situations where tested persons benefit.
MMWR 47:211-15, 1998
![Page 7: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/7.jpg)
OraQuick Rapid HIV-1 Antibody Test
Easy to use: untrained first-time users report 98.5% of results correctly
Read test results in 20-60 minutes
Sensitivity = 99.6% / Specificity = 100%
CLIA-waived
![Page 8: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/8.jpg)
OraQuick: Oral fluid, serum, whole blood
![Page 9: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/9.jpg)
FDA-approved for use with whole blood; granted CLIA waiver.
![Page 10: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/10.jpg)
Obtain finger stick specimen…
![Page 11: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/11.jpg)
Or whole blood
![Page 12: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/12.jpg)
Loop collects 5 microliters of whole blood
![Page 13: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/13.jpg)
Insert loop into vial and stir
![Page 14: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/14.jpg)
Insert device; test develops in 20 minutes
![Page 15: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/15.jpg)
Read results
Positive Negative
Reactive Control
Positive HIV-1
TT
C C
![Page 16: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/16.jpg)
Sold only to “clinical laboratories”
To perform CLIA-waived tests, entities must:1) Enroll in CLIA program2) Obtain a Certificate of Waiver3) Pay a biennial fee4) Follow manufacturers’ instructions5) Meet state requirements
Requirements for OraQuick Testing
![Page 17: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/17.jpg)
Have an adequate quality assurance program
Assurance that operators will receive and use instructional materials
QA guidelines for OraQuick testing and sample forms:
www.cdc.gov/hiv/testing.htm
Requirements for OraQuick Testing
![Page 18: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/18.jpg)
Reveal HIV-1 Antibody Test
Requires serum or plasma, more dependent on laboratory
CLIA: moderate complexity Read results within minutes Sensitivity 99.8% Specificity: serum 99.1% plasma 98.6% Run controls with each test
![Page 19: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/19.jpg)
Reveal HIV-1 Rapid Antibody Test for serum, plasmaCLIA Category - Moderate Complexity
![Page 20: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/20.jpg)
Centrifuge to obtain serum or plasma
![Page 21: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/21.jpg)
Add 20 drops of buffer to reconstitute conjugate. (Refrigerate to store)
![Page 22: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/22.jpg)
Add 3 drops buffer to moisten membrane
![Page 23: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/23.jpg)
Add one drop of serum or plasma, followed by 3 drops of buffer.
![Page 24: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/24.jpg)
Add 4 drops of conjugate solution
![Page 25: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/25.jpg)
Add 3 drops of buffer to wash
![Page 26: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/26.jpg)
Read results immediately
Positive Negative
![Page 27: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/27.jpg)
Rapid Test Performance: Serum
Sensitivity Specificity OraQuick 100% 100% Reveal 99% 100% SUDS 97.9% 94.5% HIV 1-2 EIA - 95.1%
206 HIV+, 194 HIV- stored sera
![Page 28: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/28.jpg)
Point-of-Care Testing
To expand testing in non-clinical settings:
–Fingerstick or whole blood specimen–One-step–Easy to interpret– Internal control
![Page 29: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/29.jpg)
Example: Three possible OraQuick test results
Non-reactive Reactive Invalid
![Page 30: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/30.jpg)
Non-reactive
![Page 31: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/31.jpg)
Reactive
![Page 32: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/32.jpg)
Invalid
![Page 33: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/33.jpg)
Invalid
![Page 34: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/34.jpg)
Invalid
![Page 35: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/35.jpg)
Invalid
![Page 36: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/36.jpg)
The challenge: Weakly Reactive
![Page 37: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/37.jpg)
The Need for Training
Blood & body fluid precautions Obtaining the specimen (finger stick or blood
draw) Performing the test Providing test results and counseling Quality assurance OSHA requirements
![Page 38: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/38.jpg)
Remember the tradeoffs…
Good News: More HIV-positive people receive their test results.
Bad News: Some people will receive a false-positive result before confirmatory testing.
![Page 39: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/39.jpg)
Interpreting Rapid Test Results
For a laboratory test:Sensitivity: Probability test=positive if
patient=positiveSpecificity: Probability test=negative if
patient=negativePredictive value:
Probability patient=positive if test=positive Probability patient=negative if
test=negative
![Page 40: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/40.jpg)
Example: Test 1,000 persons
HIV prevalence = 10%
True positive:
False positive:
Positive predictive value:
100/104 = 96%
100 4
Test Specificity = 99.6%
(4/1000)
![Page 41: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/41.jpg)
Example: Test 1,000 persons Test Specificity = 99.6% (4/1000)
HIV prevalence = 10%True positive: 100 False positive: 4
Positive predictive value: 100/104 = 96%
HIV prevalence = 0.4%
True positive:
4 False positive:
4
Positive predictive value:
4/8 = 50%
![Page 42: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/42.jpg)
Positive Predictive Value of a Single Test
Depends on Specificity & Varies with Prevalence
Test Specificity
HIV Prevalence
Predictive Value, Positive Test
10% 99% 98% 92% 5% 98% 96% 85% 2% 95% 91% 69% 1% 91% 83% 53% 0.5% 83% 71% 36% 0.3% 75% 60% 25% 0.1% 50% 33% 10%
OraQuick
EIA Reveal
99.9% 99.8% 99.1%
![Page 43: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/43.jpg)
Reports from the 2003 HIV Prevention Conference
Promising news with rapid HIV tests for ––Routine screening in emergency rooms–Increasing receipt of results at CT sites–Screening in labor and delivery–Outreach testing
![Page 44: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/44.jpg)
Routine Screening of Emergency Department Patients Using OraQuick®
Rapid HIV-1 Antibody Test
Cook County Bureau of Health ServiceChicago, Illinois
![Page 45: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/45.jpg)
Number rapid tested 1664
Number received results 1624 (98%)
Number new HIV+ 39 (2.3%)
Number who entered care 28 (76%) median 18 days
Rapid Test Outcomes
![Page 46: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/46.jpg)
N=39
No previous test 22 (57%)
Risk Factors
MSM 12 (31%) IDU 5 (13%) Sex Partner IDU 3 ( 8%) No identified risk 19 (49%)
CharacteristicsRapid Test Positive Patients
![Page 47: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/47.jpg)
HIV Screening in Acute Care Settings
Cook County ED, Chicago 2.3% Grady ED, Atlanta 2.7% Johns Hopkins ED, Baltimore 3.2%
HIV testing sites 1.3%
New HIV+
![Page 48: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/48.jpg)
HIV Screening with OraQuick in Labor and Delivery: the MIRIAD
Study
Testing of pregnant women in labor for whom no HIV test results are available; 12 hospitals in 5 cities: Atlanta, Chicago, Miami, New Orleans, New York
To date 3178 women screened 27 new HIV infections identified 2 false positive OraQuick tests, no false negatives 7 false-positive EIAs
![Page 49: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/49.jpg)
The rapid test is done on this counter, extra supplies are stored below.
OB physicians and midwives share MIRIAD testing
Point-of-Care Testing Station
![Page 50: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/50.jpg)
Turnaround Times for Rapid Test Results,
Point-of-Care vs Lab Testing
Point-of-care testing: median 45 min– (range 30 min – 2.5 hours)
Testing in Laboratory: median 3.5 hours
– (range 94 min – 16 hours)
MMWR 52:36, Sept 16, 2003
![Page 51: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/51.jpg)
OraQuick Outreach to High Risk Persons of Color
Patrick Keenan MDUniversity of Minnesota Medical
SchoolDepartment of Family Practice and
Community Health
![Page 52: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/52.jpg)
OraQuick Outreach Study (7/02 – 6/03) N =1021
On-site group pretest counseling. Individual testing and post-test
counseling. Testing procedure:
Fingerstick OraQuick (results given) Fingerstick neg -> OraSure backup Fingerstick pos -> venous EIA/WB
![Page 53: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/53.jpg)
Outreach Testing Sites
Chemical Dependency Programs
Homeless shelters Sex worker support
program Drop-in center for gay
youth Teen clinic Gay bars
Sex offender groups “Johns” programs Half-way houses Health fairs Strip club workers African-born groups Drug court support
groups
![Page 54: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/54.jpg)
Results
99.7% of clients received their test results and post-test counseling.
The average time between fingerstick and learning test result was 28 minutes.
![Page 55: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/55.jpg)
OraQuick Fingerstick Results:N = 1021
Preliminary positive 5 (0.5%) True positives 4 (0.4%) False Positives 1 (0.1%) Sensitivity 4/4 (100%) Specificity 1016/1017
(99.9%) Positive Predictive Value 4/5 (80%)
![Page 56: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/56.jpg)
Client Survey Results I
“I have tested for HIV in the past and I prefer receiving my results the same day”
Strongly agree or agree = 98%Disagree or strongly disagree = 2%
![Page 57: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/57.jpg)
Client Survey Results II
“I would rather have my finger stuck than have blood drawn from my vein”
Agree or strongly agree = 95% Disagree or strongly disagree = 5%
![Page 58: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/58.jpg)
Client Survey Results: III
“I understand the results of my test.”
Agree or strongly agree = 99% Disagree or strongly disagree = 1%
![Page 59: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/59.jpg)
Confirmatory Testing
For Western blot: Venipuncture for whole blood Oral fluid specimen Dried blood spots on filter paper
![Page 60: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/60.jpg)
Requirements for OraQuick Testing
Register as a laboratory CLIA “Certificate of Waiver” or Limited Public Health Use Exception
Train staff Establish Quality Assurance Program For CDC testing programs:
Postmarketing surveillance
![Page 61: Rapid HIV Testing: 2003 Update Bernard M. Branson, M.D. Chief, Lab Determinants and Diagnostics Section Centers for Disease Control and Prevention.](https://reader034.fdocuments.us/reader034/viewer/2022051619/56649dc55503460f94ab8297/html5/thumbnails/61.jpg)
Additional Resources
General and technical information (updated frequently):
www.cdc.gov/hiv/testing.htm