Radiography and Mammography Viewing Systems - · PDF fileFor assistance contact your...

22
For assistance contact your authorized RADX Dealer, or call S&S Technology at 800-231-1747 E-mail [email protected] www.ssxray.com Radiography and Mammography Viewing Systems

Transcript of Radiography and Mammography Viewing Systems - · PDF fileFor assistance contact your...

For assistance contact your authorizedRADX Dealer, or call S&S Technology at 800-231-1747

E-mail [email protected]

Radiography and MammographyViewing Systems

Administrator
Rectangle

MammographyDigital ViewingStation

Mammography-Digital Viewing Station (MDVS),

consisting of the MS-400DVS as shown on top of

the DVS-2-MS

60.5"

The Mammography Digital Viewing Station (MDVS) is designed for departments converting todigital mammography systems, as an efficient workstation to compare the new digital studies with theprevious "hard copy" films.

Mammography films are mounted on a motorized viewer directly above the location of the monitors,minimizing space between images. The viewing station incorporates a single motorized film belt onthe upper tier and the capability of accommodating up to four CRT/Flat Panel Monitors in the lowersection.

The single motorized belt has a capacity of 400 (24 x 30 cm) films or 600 (18 x 24 cm) films. It isextremely bright, with luminance output exceeding ACR recommendations for Mammography.

It features external horizontal shutters and automated vertical shades for efficient film masking.Convenient controls provide user-friendly operation of the belt and vertical shades. The optional Auto-Select feature allows for quick film retrieval.

The large work counter surface minimizes clutter, and two pullout-writing shelves allow the userabundant work space.

Four convenient power receptacles are present for a dictaphone, laptop, or other auxiliary powerrequirements. The DVS was designed to have abundant leg room for a comfortable work areaoptimized for up-close, detail work.

DS023/0406June 2004

10625 Telge Road • Houston, TX 77095800-231-1747 • Fax 281-815-1494

www.ssxray.com • [email protected]

Mammography DigitalViewing Station

SPECIFICATIONS

MS400DVSDimensions (inches)

Height......................................................20.25Width.......................................................60.50Depth.......................................................24.125

CapacityFilms......................................................... 400 (24 x 30 cm)

600 (18 x 24 cm)Approx Ship Weight (lbs).................. 320

DVS2-MSDimensions (inches)

Height......................................................58.50Width.......................................................66Depth.......................................................46.40 w/shelf

*Height is variable depending on the size of the monitors used.

MS400DVSPower Requirements120 VAC / 60 Hz, 6 AmpHeat Output - 2000 BTU / Hr Max.

24.12"

20

.25

"

VistaViewerFor Mammography

10625 Telge Road • Houston, TX 77095800-231-1747 • Fax 281-815-1444 • www .ssxray.com • [email protected]

DS018/0405© S&S TechnologyMay 2004

Vertical answer to mammography reading. The VistaViewer for mammography features a high-capacity verticalbelt and automated shades to support reading centers concerned with efficient through-put. The viewer offers highluminance for reading dense mammograms. The VistaViewer holds up to 1,200 18 x 24 cm films with no strings todistract or obscure the images; morever, the belt can be factory customized for displaying 14 x 17 inch radiographs.The frame is designed to allow the radiologists to comfortably stretch their legs while reading films.NOTE: The 14 x 17 format will decrease belt capacity.

Improved radiologist efficiency. During film placement, the technologist can program the VistaViewer to rememberthe positions for the four independent curtains for each frame. Thus, when the radiologist accesses a particularframe, the shades will automatically adjust to the appropriate setting for the displayed films. The VistaViewer alsohas an optional variable intensity bright spot with joystick control for focusing on extra-dense areas of film. Whenactivated, the spotlight automatically finds the non-shaded area.

At your fingertips for security and ease of use. The standard touchpad controls put random frame selection, aswell as lighting and shade management at the radiologist’s fingertips. Next and Previous buttons support quickmovement during sequential film reading; the standard Foot Switch also can be used for this purpose. To conserve,energy, the VistaViewer includes an auto-shutoff feature that triggers after 30 minutes of inactivity. To ensurepatient data confidentiality, the viewer has a security code feature that limits access to films. Two convenienceoutlets installed in the front of the viewer allow for easy installation of such accessories as a spot light and voicerecorder.

SpecificationsProduct Number ........................ 08-01002Luminance ................................. Exceeds ACR recommendations

for mammographyDimming ................................. 5 positions

Lamps ........................................ 10 fluorescent lamps; 40 Wattseach

Viewing format ......................... 4-over-4 (24 x 30 cm); 6-over-6(18 x 24 cm);3 across (14 x 17 inch) 1

Viewing area (WxH) ................. 102 x 64 cm (40 x 25 inches)Films

Transport mechanism ........... 1 vertical belt; 200 framesCapacity, total ....................... 800–24 x 30 cm; 1,200–18 x 24

cmMasking

Vertical .................................. Two motorized shutters;programmable frame memory

Horizontal .............................. Two motorized shutters;programmable frame memory 1Belt may be custom-configured, at no extra charge, to support 14 x 17 inch radiographs

Frame accessNext/Prior ............................... Standard buttonsSpecific Frame Access ........ Numeric keypad integrated with

control panelDimensions, in inches

Operating (WxHxD) .............. 61.25 x 72 x 44.5, includingcasters

Shipping (WxHxD) ................ 61.25 x 72 x 30, including castersDesk (WxHxD) ..................... 57 x 28 x 14.5

Weight ....................................... 840 lbs.Electrical .................................... 120 VAC, 60 Hz, 15 Amps;

Two Convenient 3 Amp (MAX)Accessory outlets

Color Temperature .................... 6,500 KMaximum Heat Output ............... 3,600 BTU/hr

For Mammography

����������������� �������������������� �

������ �������������������� �

�������� ���������� �!""�#�

$��%�&�%�"'"���(�$�%$��%�'''����)��(���)*�������������+���()*�

,-��".����/-0-��*��� �%12��3������

The perfect choice for multi-film reading. The VersaViewer supports both 14 x 17 inch and mammography-sized films on a vertical belt, allowing the radiology department to mix or match film sizes with equal security of filmretention. The combination of a large viewing area and a customized vertical belt allows the radiologist to see threeyears of mammography studies (18 x 24 cm films) concurrently for the ultimate in film comparison efficiency! Theframe is designed to allow the radiologists to comfortably stretch their legs while reading films.

For fast, efficient reading. Moreover, during film placement, the technologist can program the VersaViewer tostore the positions for the four independent curtains for each frame. Frame-specific shading improves efficiency byreducing the radiologist’s time spent on adjusting the viewer for proper reading conditions.

The viewer also has an optional variable intensity bright spot with joystick control for highlighting and focusing in onextra dense areas of film. When activated, the spotlight automatically finds the non-shaded area.

At your fingertips. The standard touchpad controls put frame selection, as well as lighting and shade managementat the radiologist’s fingertips. Next and Previous buttons support quick movement during sequential film reading; anoptional Foot Switch also can be used for this purpose. To conserve, energy, the VersaViewer includes an auto-shutoff feature that triggers after 30 minutes of inactivity. To ensure patient data confidentiality, the viewers have asecurity code feature that limits access to films. Four convenience outlets installed in the front of the viewer allow foreasy installation of such accessories as a spot light and voice recorder.

SpecificationsProduct Number ........................ 08-01005Luminance ................................. Exceeds ACR recommendations

for mammographyDimming ................................. 5 positions

Lamps ........................................ 14 fluorescent lamps;50 Watts each

Viewing format ......................... 4-over-4 (14 x 17 inch);6-over-6 (24 x 30 cm);8-over-8-over-8 (18 x 24 cm)

Viewing area (WxH) ................. 145 x 89 cm (57 x 35 inches)Films

Transport mechanism ........... 1 vertical belt (customized);100 frames

Capacity, total ....................... 400–14 x 17 inch;800–24 x 30 cm;1,200–18 x 24 cm

Frame accessNext/Prior ............................... Standard buttonsSpecific Frame Access ........ Numeric keypad integrated with

control panel

MaskingVertical .................................. Two motorized shutters;

programmable frame memoryHorizontal .............................. Two motorized shutters;

programmable frame memoryDimensions, in inches

Operating (WxHxD) .............. 73.5 x 79 x 44.5, including castersShipping (WxHxD) ................ 73.5 x 79 x 30, including castersDesk (WxHxD) ..................... 69 x 30 x 14.5

Weight ....................................... 840 lbs.Electrical .................................... 118 VAC, 60 Hz, 10 Amps;

four convenience outletsColor Temperature .................... 6,500 KMaximum Heat Output ............... 3,600 BTU/hr

MammographyWorkstationFor Comparing Films with Electronic Images

��������������� ��!�"�#� ��$��%�&&�'�

(��)�*�)�&+&��!����,��(�)(��)�+++��!�����-��,��-�����!��������#���� .�� ,-���

/���(0�����1232�����������)�4�5��6�������

Elegant combination of digital and film display. The Mammography Workstation is the perfect answerfor those departments converting to new digital formats while comparing these images to prior film-basedstudies. It combines into one compact frame both a motorized film belt and an area for a PACS or CADworkstation, resulting in an elegant mammography reading station.

Improving the radiologist’s efficiency. The Mammography Workstation places both the films and digitalimages in front of the radiologist, thus eliminating awkward reading positions that occur when a computerworkstation is adjacent to a motorized viewer.

The motorized belt holds up 600 films (18 x 24 cm), and exceeds ACR recommendations for mammographyreading. Its external horizontal shutters and automated vertical shade provide efficient film masking.

At your fingertips for convenience and ease of use. To maximize the radiologist’s workspace, the deskextends the width of the viewer. The standard touchpad controls for the motorized belt put random frameselection, as well as lighting and shade management at the radiologist’s fingertips. Next and Previous buttonssupport quick movement during sequential film reading; moreover, an optional Foot Switch also can be usedfor hands-free belt activation. For convenient computer hardware connections, the Mammography Workstationincludes a six-outlet, hospital-grade power strip with its own power switch.

SpecificationsProduct Number .................... 08-01001Luminance ............................. Exceeds ACR recommendations

for mammographyDimming ............................. 5 positions

Lamps .................................... 4 fluorescent lamps;40 Watts each

Viewing format ..................... 4-across (24 x 30 cm);6-across (18 x 24 cm)

Viewing area (WxH) ............. 102 x 30 cm (40 x 12 inches)Films

Transport mechanism ....... 1 horizontal belt; 100 framesCapacity, total ................... 400–24 x 30 cm;

600–18 x 24 cmMasking

Vertical .............................. One continuous motorizedshade; 2 programmablepositions

Horizontal .......................... Two manual shutters (left andright)

Frame accessNext/Prior ........................... Standard buttonsSpecific Frame Access .... Numeric keypad integrated with

control panelDimensions, in inches

Operating (WxHxD) .......... 62 x 62.5 x 44.75, includingcasters

Shipping (WxHxD) ............ 62 x 62.5 x 30, includingcasters

Desk (WxHxD) ................. 61.75 x 29 x 15.5, with cut-outfor close reading

Weight ................................... 600 lbs.Electrical ................................ 118 VAC, 60 Hz, 10 Amps;

6-outlet, hospital-grade powerstrip included with separatepower switch

Color Temperature ................ 6,500 KMaximum Heat Output ........... 3,000 BTU/hr

CAD and/or PACS system not included.

FlexiViewerFor Mammography andGeneral Radiology

���������������� ����������� ������

��������������� ���������������� �!�

"��#�$�#� % �������&��"�#"��#�%%%�����'''(��&���()�����������������*�� &()��

+����,�$�

For when you screen both mammograms and 14 x 17-inch films. The film belts on the FlexiViewersupport a wide variety of films, from 14 x 17-inch to 18 x 24 cm. Moreover, films of various sizes can beintermixed throughout the dual, horizontal belts for more convenient reading.

The viewer’s independent belt system allows one belt to move while the other remains stationary. Thus,the radiologist can scroll through an unlimited number of the patient’s earlier films while current filmsremain constantly in view. The vertical shades span the width of each light box, thus maskingmammograms no matter where they’re placed on the wide viewing area.

Easy to learn and control. The FlexiViewer’s digital control panel has been arranged for effortless use.It includes belt action controls, Previous/Next buttons, buttons for adjusting luminance and the verticalshades and a Liquid Crystal Display. The integrated keypad provides a quick, efficient method forreading films that are not placed sequentially on the viewer. The LCD allows at-a-glance status of theluminance level and belt position. The standard Foot Switch provides a hands-free method of framecontrol.

Specifications

Product Number ..................................... 08-00976Luminance ..............................................Exceeds ACR recommendations for reading mammography

Dimming .........................................5 positionsLamps .....................................................6 – 60-inch daylight fluorescent lamps per light panel; 50 W atts eachViewing format ........................................ 4-over-4 (14 x 17 inch); 6-over-6 (24 x 30 cm); 8-over-8 (18 x 24 cm)Viewing area (WxH) ................................145 x 43 cm (57 x 17 inches) per beltFilms

Transport mechanism ....................2 single-tier, horizontal belt; 50 frames eachCapacity, total .................................400–14 x 17 inch; 600–24 x 30 cm; 800–18 x 24 cm

MaskingVertical ............................................One motorized shade (per tier); two programmable preset positionsHorizontal .......................................Two manual shutters (left and right) per tier

Frame accessNext/Prior ........................................Standard buttonsSpecific Frame Access ..................Numeric keypad integrated with control panel

Dimensions, in inchesShipping (WxHxD) ..........................69 x 74 x 30, including castersOperating (WxHxD) ........................69 x 74 x 42, including castersDesk (WxHxD) ............................... 57 x 29 x 14Depth of leg space ......................... 21

Weight .....................................................740 lbs.Electrical ................................................. 117 VAC, 60 Hz, 10 Amps

........................................................Convenient 1 Amp (MAX) Accessory outletColor Temperature Rating ...................... 6,500 KMaximum Heat Output ...........................3,070 BTU/hr

-�./.���)������

0�������$

CrystalViewerFor Mammography

�������������

Ideal system for mammography evaluations. The CrystalViewer enhances productivity by allowingthe radiologist to adjust luminance to the intensity needed for each image’s density. The motorizedvertical shade spans the width of each light box, eliminating the problem of light leakage that occursbetween divided shade systems. Each vertical shade has two programmable settings.

The CrystalViewer’s dual independent belt system allows one belt to move while the other remainsstationary. Thus, the radiologist can scroll through an unlimited number of the patient’s earlier filmswhile current mammograms remain constantly in view. The two 100-frame belts combine for up to 800films (24 x 30 cm).

Easy to learn and control. The CrystalViewer’s digital control panel has been arranged for intuitiveuse. It includes belt action controls, Previous/Next buttons, buttons for adjusting luminance and thevertical shades, and a Liquid Crystal Display. The integrated keypad provides a quick, efficient methodfor reading films that are not placed sequentially on the viewer. The LCD shows at-a-glance status of theluminance setting and belt position. The standard Foot Switch provides a hands-free method of framecontrol.

SpecificationsProduct Number .....................................08-00972Luminance ..............................................Exceeds ACR recommendations for reading mammography

Dimming .........................................5 positionsLamps .....................................................4 – 48-inch daylight fluorescent lamps per light panel; 40 W atts eachViewing format .......................................4-over-4 (24 x 30 cm); 6-over-6 (18 x 24 cm)Viewing area (WxH) ...............................102 x 30 cm (40 x 12 inches) per beltFilms

Transport mechanisms ..................2 single-tier, horizontal belts; 100 frames per beltCapacity, total ................................800–24 x 30 cm; 1,200–18 x 24 cm

MaskingVertical ............................................One motorized shade (per tier); programmable preset positionsHorizontal .......................................Two manual shutters (left and right) per tier

Frame accessNext/Prior .......................................Standard buttonsSpecific Frame Access ..................Numeric keypad integrated with control panel

Dimensions, in inchesShipping (WxHxD) .........................62 x 62.5 x 30, including castersOperating (WxHxD) .......................62 x 62.5 x 42, including castersDesk (WxHxD) ..............................40 x 29 x 12Depth of leg space .........................21

Weight ....................................................615 lbs.Electrical .................................................117 VAC, 60 Hz, 10 Amps

........................................................Convenient 1 Amp (MAX) Accessory outletColor Temperature Rating ......................6,500 KMaximum Heat Output ...........................2,870 BTU/hr

� �������������������������������� ��

!�"�!��#��������$�� �! ��!�###�����%%%&��$���&'�����������������(���$&'��

)� ��* " �

+�,-,� ��'������

.���� � "

A -1

8 0 0 /347 -XRAY

www. s s x r a y . c om

MO

TO

RIZ

ED

VIE

WE

RS

MAMMOGRAPHIC VERTICAL VIEWERMV800

Base: 2.5” CastersFinish: Attractive durable baked on polyester, in light grey with blue sides. Shelf is finished in matching blue laminate with Oak bull nose trim.Electrical: 118 volts, 60HZ AC, 1 phase, total amperage draw with all lights on and motor running-10 amps. For 220 volts, 50 HZ or other vol t-ages, consult factory for details.Approx. Weight: 840 Lbs.

28.37"44.5"

30.25"

57"

71.187"

LED READOUT(FRAME NUMBER IDENTIFICATION)

AUTOMATIC SHUTOFF TIMER

POSITIONING JOY STICK

CURTAIN MEMORY CURTAIN AUTO RETRACT

SPOT LIGHT INTEN-SITY

CONTROL

PANEL SELECTORS PREVIOUS/NEXT BUTTON

LIGHT INTENSITY CONTROLCURTAIN PRESETS

CURTAIN POSITIONING

Here is our top of the line mammographymotorized viewer, which features intuitiveuse, no obstructing strings, and exceedsall ACR requirements for mammography!Four independent curtains can easily be indi-vidually positioned by the technologist when thefilms are placed on the viewer (using the manu-al curtain controls or one of four programmablecurtain presets), and their position stored intothe frame-specific memory with the push of abutton. When the particular “frame” is accessedby the radiologist, the shades will automatical-ly go to the position previously set, providingoptimal frame-specific shading for all films onthe viewer. The unit is designed with the comfortof the radiologist in mind, with extra leg room, alarge work desk and easily accessible utility out-lets.

Variable intensity illumination is standard,ranging from 100% down to 60% in incrementsof 10%. This allows the radiologist to select theproper level of illumination for a specific type offilm, change the intensity for other films, andeasily and precisely return to this original level.High frequency energy efficient electronic bal-lasts provide for flicker-free, reduced eyestrainviewing.

An optional variable intensity bright spot per-mits highlighting and focusing in on extra denseareas of the film. It can be moved to any locationon the lower tier . The spotlight has been pro-grammed to automatically find the non-shadedarea when turned on. The positioning joystickcan be used to move the spotlight to the preciselocation desired.

The vertical transport belt features exceptionally quiet and smooth operation. The MV800 has 200 frames and can hold 80020x30cm films (or 1200 18x24cm films). There are no strings to distract or obscure the viewing of films.

This viewer is ideal for sequential interpretation of film, and automatically centers the next set of films.The standard auto-index-ing feature rapidly accesses a selected film (moving at a speed greater than one frame per second). Any film can be viewed inthe upper or lower tier position. Previous/Next buttons or the standard dual foot switch can be used to move the transport beltand quickly center the adjacent films. Angulated viewing of 50 cuts glare.

To conserve energy, the units feature auto-shutoff when there is inactivity for 30 minutes. There is also the ability to program secu-rity codes into the machine to limit access to films loaded into the viewer .

MEMBRANE TOUCH PAD CONTROLS

PROGRAMMABLE SHADES WITH INDIVIDUAL PANEL SPECIFIC MEMORY

Dimensions

ULC UL

S & S X-RAY PRODUCTS INC

.

Administrator
Rectangle

MS880Mammography Viewer

MS880Mammography Viewer

Three years

of films

at once.

At your fingertips.

Accessories and

Options.

The MS880A single-belt viewer is wide enough to present twelve 24 x 30 cm mammography films concurrently. Thus, the radiologist can comfortably view three years of film without ever changing belt frames.

The control panel includes Next and Prior buttons to ensure the radiologist can efficiently screen films.

Auto Select, Belt Advance Foot Switch, Pivot or Fixed Instrument Mounting Arm, and Film Jacket File.

SpecificationsProduct Number 08-00987Peak Luminance 7,000 NitColor Temperature 6,500 KLamps 6 Daylight fluorescent lamps

per light panel (2 panels)Maximum Heat Output 3,600 BTU/hrFluorescent Dimming 5 PositionsFilm Capacity (Standard) 780-24 x 30 cmFilm Belt 1 Dual-tier, horizontal; 65 framesVertical Masking 2 Motorized ShadesHorizontal Masking 2 Manual ShuttersOperating Dimensions (WxHxD) 69 x 62.5 x 42.5 inches, including castersShipping Dimensions (WxHxD) 69 x 62.5 x 30 inches, including castersWeight 750 lbs.Maximum Current 10 Amps

S&S Technology10625 Telge RoadHouston, Texas 77095281-815-1300Fax: 281-815-1444© S&S Technology

February 2002

For assistance contact your authorized S&S RADX Dealer or call S&S Technology at 1-800-231-1747email: [email protected]

DR014/0202

MS824Mammography Viewer

MS824Mammography Viewer

Perfect for

high-volume

screening centers.

At your fingertips.

Accessories and

Options.

The MS824A mammography viewer provides a cost-effective solution for high-volume centers. The single-belt system supports and 800-film capacity in a 4-over-4 format, thus ensuring that radiologists won't spend extra time waiting for film to be loaded or unloaded.

The control panel includes Next and Prior buttons to ensure the radiologist can efficiently screen films. Four motorized vertical and two manual horizontal shades provide quick film masking.

Auto Select, Belt Advance Foot Switch, Pivot or Fixed Instrument Mounting Arm, and Film Jacket File.

SpecificationsProduct Number 08-00984Peak Luminance 6,000 NitColor Temperature 6,500 KLamps 4 Daylight fluorescent lamps

per light panel (2 panels)Maximum Heat Output 3,600 BTU/hrFluorescent Dimming 5 PositionsFilm Capacity (Standard) 800-24 x 30 cmFilm Belt 1 Dual-tier, horizontal; 100 framesVertical Masking 2 Motorized ShadesHorizontal Masking 2 Manual ShuttersOperating Dimensions (WxHxD) 62 x 62.5 x 42.5 inches, including castersShipping Dimensions (WxHxD) 62 x 62.5 x 30 inches, including castersWeight 615 lbs.Maximum Current 10 Amps

© S&S TechnologyFebruary 2002

S&S Technology10625 Telge RoadHouston, Texas 77095281-815-1300Fax: 281-815-1444

For assistance contact your authorized S&S RADX Dealer or call S&S Technology at 1-800-231-1747email: [email protected]

DR015/0202

This versatile vertical belt automatic viewer is ideal for sequential interpretation of films. The viewers feature advanced microprocessor con-trol technology, smooth and quiet operation, and film viewing without obstructing strings! The standard auto-indexing feature rapidly accesses aselected frame (moving at a speed greater than one frame per second) and automatically centers the film in the proper viewing p osition.

Variable intensity illumination, ranging from 100% down to 60% inincrements of 10%, allows the radiologist to select the proper level of illu-mination for a specific type of film, change the intensity for other films,

and easily and precisely return to the originallevel. High frequency energy ef ficient electronicballasts provide flicker free reduced eyestrainviewing.

The combination of a microprocessor and acoffee resistant membrane touch pad con-trol insures years of trouble free use. The ver-tical transport belt features exceptionallyquiet operation, and is designed to prevent filmloss or damage while in the viewer. Angulatedviewing of 50 cuts glare. The belt can accom-

modate any size film up to 14”x17”. Mix or match your film sizeswith equal security of film retention. Films slide easily andsecurely into narrow plastic grips on each frame. There are nostrings to distract or obscure the viewing of films. The durable plas-tic film grips minimize the space between the top and bottom tier forease of comparison.

Eight 14”x17” films can be viewed simultaneously. The units hold 220or 400 14”x17” films (greater quantity with smaller films).

The C and CS models have full manual and programmable shut-ter capability allowing for proper collimation to any film format.When films are loaded into the viewer , the shade locations can beeasily set manually (or through the use of one of four programmablepresets) and programmed into the frame-specific memory with thetouch of a button. The curtain memory can be easily individually setfor each frame on the machine, so that when the radiologist calls upa specific frame, the curtains move automatically to their presetlocation for that particular set of films (Manual movement of the cur-tains can easily be used to fine tune the curtain locations). Optional

customized belts are available and recommended (for C and CS models only) for mixing film formats, and in particular for mammog raphic films.The customized belt enables the simultaneous display and full shading of 3 years of screening 18x24cm mammographic studies!

A joystick-controlled variable intensity moving bright spot can be quickly positioned anywhere on the lower tier to add the extra brightness needed toread dense film. When the bright spot is activated it rapidly locates the non-shaded region on the viewing surface, then turns on. Care was taken in thedesign of the control pad and in the programming of the unit so that all functions are intuitive to the user , and easily and qu ickly learned and assimilated.

To conserve energy, the units feature auto-shutoff when there is inactivity for 30 minutes. There is also the ability to program security codes intothe machine to limit access to films loaded into the viewer .

The C and CS models with fully programmable motorized shades (with frame specific memory in addition to programmable memory pre-sets) easily exceed the proposed ACR requirements for mammography.

MO

TO

RIZ

ED

VIE

WE

RS

IDEAL FOR THE FULL SPECTRUM OF GENERAL RADIOLOGY

A - 2

OPTIONS: MV220-2 FOOT SWITCH - Hands free selection of frames (scan frames either forward or back).

MOTORIZED CURTAINS: FULL SHUTTER CAPABILITY Four curtains adjust independently for any opening size or loca-tion. Curtain memory can be set individually for automatic collimation of each frame.

CUSTOMIZED BELTS: (C &CS models only) Design your belt to fit your needs - consult factory .

ROVING SPOTLIGHT: (Bottom Tier Only) Spotlight can be positioned anywhere on the lower tier with the touch of the joy-stick. Roving spotlight is programmed to coordinate with automatic collimation of shutters and will automatically move to theopen area when activated. To allow the bright spot to move, the standard internal dividers are removed.

Base: 2.5” Swiveling CastersFinish: Attractive durable baked on polyester, in light grey with blue sides. Shelf is finished in blue laminatewith Oak bullnose trim.Electrical: 118 volts, 60HZ AC, 1 phase, total amperage draw with all lights on and motor running-10 amps.Approx. Weight: MV220 - 1140 Lbs., MV400 - 1190 Lbs.

MV220 or MV400 - VERTICAL VIEWERMV220C or MV400C - VIEWER W/ MOTORIZED CURTAINS

MV220S or MV400S - VIEWER W/ ROVING SPOTLIGHTMV220CS or MV400CS - VIEWER W/CURTAINS AND SPOTLIGHT

VERTICAL VIEWER MV220 / MV400

AUTOMATIC SHUT OFF TIMER

POSITIONING JOY STICK

SPOT LIGHT INTENSITY CON-TROL

LIGHT INTENSITY CONTROL

MEMBRANE TOUCH PAD CONTROLS

PANEL SELECTORSPREVIOUS/NEXT BUTTONS

LED READOUT (FRAME NUMBER IDENTIFICATION)

27.5"43.5"

30"

69.125"

78"

Lighting: MV220, MV220S, MV400, MV400S: 4 - 15 W att Daylight fluorescent vertical bulbs per section. MV220C, MV220CS, MV400C, MV400CS 14 - 60Watt Daylight fluorescent horizontal bulbs per viewer .

Dimensions

(Shown for the MV200CS/MV400CS)

CURTAIN POSITIONING

CURTAIN PRESETSCURTAIN MEMORY

CURTAIN AUTO RETRACT

ULC UL

8 0 0 /347 -XRAY

www. s s x r a y . c om

S & S X-RAY PRODUCTS INC

.

Administrator
Rectangle

Foot SwitchModel Number 20-0007The Foot Switch module enables hands-free operation of the film transport system on RADX Motorized Viewers. The momentary action of the switch drives the film belt at low speed (like the manual travel switch), allowing the film belt to be stopped in any position for optimal viewing flexibility.

Instrument Mounting ArmsWhen mounted to the frame of a motorized viewer, the Instrument Mounting Arm reduces workspace clutter and positions the needed device at a convenient, adjustable location. Standard versions of these arms support such ancillary devices as the RADX CoolBrite Illuminator or a dictaphone.

Pivot ArmModel Number 08-00718In addition to height, tilt, and rotation adjustments, a joint at the base of the Pivot Arm allows it to pivot at the point of attachment to the viewer, giving the user maximum control of the viewing position and distance.

Fixed ArmModel Number 08-00717The Fixed Arm provides adjustments for height, tilt, and rotation of the attached device.

5.1

AutoSelectModel Number 08-00059The AutoSelect is an optional handheld control system for any RADX Motorized Viewer. Its numeric keypad and display system allow quick, easy access to patient films. The AutoSelect provides an added level of efficiency and productivity to motorized film viewing, whether the radiologist prefers numeric entry, single-frame indexing keys (Next and Prior), momentary jog keys, or a Store/Recall feature.

RADX Products Viewer Accessories January 2001

Shown with CoolBrite(see page 5.3)

5.2

Panoscope Image Magnifier

2+ Diopter 08-009073.5+ Diopter 08-00936

Especially valuable for mammography viewing, the Panoscope magnifies image detail while eliminating extraneous light. Available with 2+ or 3.5+ Diopter lenses.

Belt ChangerModel Number 08-00904By supporting the weight of the film belt and providing the necessary "spooling" action, the Belt Changer greatly simplifies the process of installing a new film belt on a motorized viewer. The risk of belt damage during installation is practically eliminated and the detachable hand crank makes quick work of removing the old belt in a neat and orderly manner.

NitMeterIlluminance Meter

Model Number 08-00908

The RADX NitMeter provides a method of measuring both viewer luminance (nit) and room ambient illuminance (lux) in a single instrument. The calibrated scale provides simple operation and assures radiologists that their viewing systems are in conformance with these 1999 American College of Radiology (A.C.R.) recommendations*:

1. "Viewboxes used for interpreting mammograms and clinical quality review by the technologist should be capable of at least 3,000 candela per square meter (cd/m2)."

2. "The illumination levels should be 50 lux, or preferably less."

* 1999 A.C.R. mammography quality control manual

RADX Products Viewer Accessories January 2001

Film Jacket FileLower Mount File (MV-3 and MV-4 models only)

Model Number 08-00719 Field Installed08-00719 Factory Installed

RADX MV-Series motorized viewers (except the MV200DT) are available with the internally-mounted lower film jacket file that has 25 slots for film jackets. Due to the height of the MV-Series viewers, many users find the height of the lower file more convenient and accessible than top-mounted files. To maximize leg room and close-proximity film viewing, lower film jacket files are not available on Mammoscope models.

Top Mount File (MV and MS models)

Model Number 08-00934

Model Number 08-00713 Field Installed08-00713 Factory Installed

Model Number 08-00726 Field Installed08-00726 Factory Installed

Excellent as an addition or alternative to the lower file, the top-mounted film jacket file, which has 25 slots for film jackets, can be utilized for increased storage of film jackets, reports, films, documentation, etc. The upper file is more visible than the lower file and is accessible without bending or reaching below the desktop of the viewer.

5.4