Raafat Abdel-Azim 1 Bellows assembly Distensible bellows Rigid housing Control Unit.
-
Upload
joleen-hodges -
Category
Documents
-
view
224 -
download
3
Transcript of Raafat Abdel-Azim 1 Bellows assembly Distensible bellows Rigid housing Control Unit.
Raafat Abdel-Azim1
Bellows assembly
Distensible bellows
Rigid housing
Control Unit
Raafat Abdel-Azim
Control Unit
Bellows assembly
Distensible bellows
Rigid housing
2
Raafat Abdel-AzimBS3Driving Gas Flow
Raafat Abdel-Azim
Spill
Exhaust
Safety relief
Driving Gas Flow BS
Insp
Raafat Abdel-AzimBS
Spill
Exhaust
Safety relief
Flow5
Raafat Abdel-AzimBS6
Raafat Abdel-AzimBS
To atmosphere
Exp
7
Raafat Abdel-Azim
Spill
Scavenging
Overpressure
8
Raafat Abdel-Azim
Spill Valve
To scavenging system
9
Spill Valve
I & beginning E End E (2-4 cmH2O)
Driving gas
Patient’s gas
Raafat Abdel-Azim10
Raafat Abdel-Azim11
Raafat Abdel-Azim
Traditional circle system
12
Raafat Abdel-Azim
Traditional circle system
13
Raafat Abdel-Azim
Components of the Ventilator
• Driving Gas Supply• Injector (some)• Controls (F, V, T, P)• Alarms• Safety-Relief Valve• Bellows Assembly• Exhaust Valve• Spill Valve• Ventilator Hose Connection
preset 65-80 cmH2O adjustable
14
Raafat Abdel-Azim
Injector
Driving gas
Air
Air
Some ventilators use a device called an injector (Venturi mechanism) to increase the flow of driving gas. As the gas flow meets restriction, its lateral pressure drops (Bernoulli principle). Air will be entrained when the lateral pressure drops below atmospheric. The end result is an increase in the total gas flow leaving the outlet of the injector but no increase in consumption of driving gas
15
Raafat Abdel-Azim
Control of Parameters of Ventilation
• VT
• VM
• f (frequency= rate)
• I:E ratio
• IFR (Insp. Flow Rate)
• Maximum Working Pressure
16
Raafat Abdel-Azim17
Raafat Abdel-Azim18
Raafat Abdel-Azim
Narkomed - Drager
19
Raafat Abdel-Azim
Fabius - Drager
20
Raafat Abdel-Azim
Cisero - Drager
21
Raafat Abdel-Azim22
Dameca
IFR (Inspiratory Flow Rate)VT (ml) f
(b/min)Cycle
time (s)IFR TI (s) TE (s) I:E
L/min L/s ml/s
500 20 60/20 = 360 1 1000 0.5 2.5 1:5
30 0.5 500 1 2 1:2
Raafat Abdel-AzimTime (sec)
0.500
100
200
300
400
500
600
1.5 2.5 3.5 4.5 5.51 2 3 4 5 6
60 L/min30 L/min
23
Inspiratory Waveforms
Raafat Abdel-Azim
30
30
60
60
0
Constant Decelerating Accelerating Sinusoidal
TI TI TI TI
24
Raafat Abdel-Azim
The Ventilation Cycle
25
Raafat Abdel-Azim
Paw
0 t
PmaxPplat
IF E
ZEEP
IPPV
20
I EP
ause26
Raafat Abdel-Azim
• Duration of ventilation cycle (sec)
• f (60/duration)
• I phase (IF period, IP period)
• E phase
• VT, VM
• TI, TE, I:E
F V P T27
Raafat Abdel-Azim
Inspiratory Phase
During the IF period:Paw depends on:
• The airway resistance (R)
• The total thoracic compliance (C) (V/P)
R C28
Raafat Abdel-Azim
PP PP
P
P
P
PP
PP
P
P
P
Resistance
Flow Rate: FRP29
Raafat Abdel-Azim
Paw
0 t
20
N R F R F
Resistance
30
Raafat Abdel-Azim
Compliance
C
C
P PVolume
31
Raafat Abdel-Azim
Compliance
PVolume: VP
P
32
Raafat Abdel-Azim
During I pause
• No gas F into or out of the lungs Paw depends only on VI & CT
• Gas redistributes among alveoli
• This improves gas distribution in the lungs of patients with small AWD (BA, smokers).
Pause
33
Raafat Abdel-Azim
Paw
0 t
20
C R
Secretions
Bronchospasm
Kinked ETT
EB intubation
CW rigidity
Pulmonary edema
34
Compliance = V/P
• Dynamic compliance:
= VT/(PIP-PEEP) L/cmH2O
• Static compliance:
= VT/(Pplat-PEEP) L/cmH2O
Raafat Abdel-Azim35
Raafat Abdel-Azim
During the expiratory phase
VT
FRC
FRC
EI
FRCIA contents•Pulmonary edema•ARDS
PEEPPaO2 PaO2
36
Raafat Abdel-Azim
Goals of Pulmonary Ventilation
37
Raafat Abdel-Azim
• To provide adequate minute alveolar ventilation
• and to side effects
necessary to maintain the desired PaCO2
PPV ITP
38
Raafat Abdel-Azim
Adequate Ventilation
• PaCO2 of 40 mmHg = 5.3% of 760 mmHg40/760 = 0.053
• Normal resting VCO2= 200 ml/min= 0.2 L/min
• This requires VM of 3.8 L0.2/ ? = 0.0530.2/ 0.053 =
• Add dead space
Goals
3.8 L/min
39
Raafat Abdel-Azim
• VD phys: ventilated but not perfused• N = 1 ml/pound 150 ml in a 70 Kg (154 pound) adult• = 0.15 x 10 (f) =
• Required VM = 3.8 + 1.5 =
• A larger VM is required for patients who have VD or VCO2
1.5 L/min
5.3 L
Goals, Adequate Ventilation
For VCO2 For VD
40
Raafat Abdel-Azim
• When VCO2 & VD are stable:
VM 1/ PaCO2
VM x PaCO2 = constant
• e.g., PaCO2 = 50 mmHg with VM = 5 L/min
VM to 7 L/min PaCO2 to 36 mmHg
Goals, Adequate Ventilation
V1 x P1aCO2 = V2 x P2aCO2
41
Raafat Abdel-Azim
Side effects
Goals
ITP
VR
EDV
CO
PVR RVAPPV
42
Raafat Abdel-Azim
Goals, Side Effects
PPV
ITP PA > PAP VVD
43
Raafat Abdel-Azim
All these effects mean Paw
Therefore, a goal of PPV is to mean Paw while maintaining adequate ventilation and oxygenation
Goals
44
Raafat Abdel-Azim
Modes of Pulmonary Ventilation
45
Raafat Abdel-Azim
Paw
0 t
SB
-ve
+ve
Anesthesia
ICU: demand valve WOB CPAP
46
Raafat Abdel-Azim
Paw
0 t
CPAP
47
Raafat Abdel-Azim
Paw
0 t
IPPVAnesthesia (or sedation) + MRFew days muscle atrophy
ZEEP
48
Raafat Abdel-Azim
Paw
0 tPEEP
IPPV + PEEP
49
Raafat Abdel-Azim
Assisted Ventilation (A)Paw
0
A
t
S
Patient f and timingHazard: hypoventilation
If or No S
No A
50
Raafat Abdel-Azim
Assisted Ventilation (A)Paw
0
CPAP
A
t
Patient f and timingHazard: hypoventilation
If or No S
No A
51
Raafat Abdel-Azim
Paw
0
A+CProvides a minimum f below which C
Assist-Control (AC)
52
Raafat Abdel-Azim
• Conscious patients are more comfortable on A & AC > C
• Muscle atrophy is still a problem
• Most AV don’t provide A or AC
AC
53
Raafat Abdel-Azim
IMV
0
Paw
St
Preset VT and f
SB is allowedMuscle atrophy is less likely
IMV PaCO2 < apneic threshold no SB IMV = IPPV
54
Raafat Abdel-Azim
Paw
0 tTriggering
window
Mandatory
S
Synch. Mandatory
NO
Preset VT and fSIMV
55
Raafat Abdel-Azim
Paw
t
Pressure-limited ventilation (PLV) (PCV)
Pmax
Pplat
0
56
Raafat Abdel-Azim
Ventilator Classification
• Pressure versus Volume Ventilators
• Single-Circuit versus Double-Circuit Ventilators
• Bellows Assembly
57
Raafat Abdel-Azim
P vs V Ventilators
P Ventilators V Ventilators
Changes in R & C change
VT P
Commonly used in
Neonates Barotrauma Compressible V
ICU
Most AV
Other names P generators
P-limited
P-stable
Flow generators
V-stable
58
Raafat Abdel-Azim
ICU
Anesthesia
59
Raafat Abdel-Azim
Bellows assembly
Distensible bellows
Rigid housing
Driving gas BS to patient
E
I
60
Raafat Abdel-Azim
Drager (End-insp.)
Ohmeda (End-exp.)
VT
VT
61
Raafat Abdel-Azim
NA Drager bellows Ohmeda bellows
E Expands partially Fully expands
I Empties completely Partially empties
Spill valve External (visible) Hidden in housing
62
Raafat Abdel-Azim
Ascending bellows Descending bellows
E
AttachmentMovement
EEPVT adjustment
Disconnection or leak
63
Raafat Abdel-Azim
Interaction between the Ventilator and Breathing Circuits
64
Raafat Abdel-Azim
Preset / delivered VT difference due to:
1. Effect of circuit compliance
for all V ventilators: both single circuit and double circuit
2. Effect of FGFR
only for double circuit ventilators
Interaction
65
Raafat Abdel-Azim
Effect of Circuit ComplianceInteraction
PIP Delivered VT
Circuit compliance = Compressible volume (gas compression, tubing distension)
CC= the V of gas that must be injected into a closed circuit to cause a unit in the circuit P
It is related to the volume of the circuit PV
Ventilator tubing V C
Anesthesia 6 L 9 ml/cmH2O
ICU (1.5 ml/cmH2O)66
Raafat Abdel-Azim
• Preset – Del VT difference CC and PIP
Interaction, Circuit Compliance
Del VT = Preset VT – [(PIP - PEEP) x CC](i.e, VT loss = P x CC)
Ventilator tubing
CC
(ml/cmH2O)
PIP
(cmH2O)
V Loss
(ml)
Anesthesia 9 20
30
180
270
ICU 1.5 20
30
30
45
Example:
67
Raafat Abdel-Azim
Volume loss is specially important in:
• Patients with PIP
ICU patients
• Patients with small VT
Neonates PLV
Interaction, Circuit Compliance
68
Raafat Abdel-Azim
Effect of Fresh Gas Flow Rate(from anesthesia machine)
Interaction
FGFR Delivered VT
V
FGF from AM
69
Raafat Abdel-Azim
Volume added = TI x FGFR
TI depends on
I:E I fraction of the respiratory cycle f
TI = 60/f x I fraction
Example: I:E = 1:2, f = 20, FGFR = 6 L/min
TI = 60/20 x 1/3 = 1 s
FGFR = 6000 ml/min = 100 ml/s
V added = 1 x 100 = 100 ml
At a certain I:E, f TI V added
Interaction, FGFR
I:E I fr
1:1 ½
1:2 1/3
1:3 ¼
70
Raafat Abdel-Azim
Interaction, FGFR
71
Raafat Abdel-Azim
Volume added is specially important:
When making large changes in FGFR
O2 flush 55 L/min (don’t use during MV)
When delivering small VT
For a neonate, changing FGFR from 1 to 5 L/min double VT
Interaction, FGFR
72
Raafat Abdel-Azim
Delivered VT = Preset VT - V lost (compression) + V added (An. machine)
Interaction
73
Raafat Abdel-Azim
Fresh Gas Decoupling (FGD)
74
Circle system with piston ventilator and fresh gas decoupling (FGD)(Dräger Narkomed 6000)
Fresh Gas Decoupling (FGD)
During the I phase the FG coming from the AWS through the fresh gas inlet is diverted into a separate reservoir by a “decoupling valve” (located between the fresh gas source & the circle BS). In the case of the NM 6000 series, the reservoir bag serves as an accumulator for FG storage until the E phase begins.
75 Raafat Abdel-Azim
Raafat Abdel-Azim
During the E phase, the decoupling valve opens, allowing the accumulated FG in the reservoir bag to be drawn into the circle system to refill the piston ventilator chamber. Because the ventilator exhaust valve also opens during the E phase, excess FG & exhaled patient gases are vented into the scavenging system
NPR valve = negative pressure relief valve
76
Raafat Abdel-Azim
• Most FGD systems are designed with either piston-type or descending bellows-type ventilators.
• Because the ventilator piston unit or bellows in either of these type systems refills under slight negative pressure, it allows the accumulated fresh gas from the reservoir to be drawn into the breathing circuit for delivery to the patient during the next ventilator cycle.
• As a result of this design requirement, it is not possible for FGD to be used with conventional ascending bellows ventilators, which refill under slight +ve pressure
77
Raafat Abdel-Azim
Advantage of FGD
risk of barotrauma & volutrauma
With a traditional circle system, in FGF from the flow meters, or from inappropriate activation of the oxygen flush valve, may contribute directly to VT excessive FG pneumothorax, pneumomediastinum, other serious patient injury, or even death.
Because systems with FGD isolate the patient from FG coming into the system while the ventilator is in I phase, delivered VT tends to remain stable over a broad range of FGF rates, and the potential risk of barotrauma to the patient is greatly reduced.
78
Raafat Abdel-Azim
Disadvantages of FGD
1. The possibility of entraining room air into the patient gas circuit
• In a FGD system, the ventilator piston unit (or descending bellows) refills under slight -ve pressure. If the total volume of gas contained in the reservoir bag + that returning as exhaled gas from the patient’s lungs is inadequate to refill the ventilator piston unit, -ve patient Paw could develop.
• To prevent this, a negative pressure relief valve is placed in the FGD BS. If BS P < a preset threshold value, then the relief valve opens and ambient air is allowed to enter into the patient gas circuit.
• If this goes undetected, the entrained atmospheric gases could lead to dilution of either or both the inhaled anesthetic agent(s) or an enriched oxygen mixture (lowering an enriched oxygen concentration toward 21%).
• If allowed to continue, this could lead to intraoperative awareness and/or hypoxia.
• High-priority alarms with both audible and visual alerts should notify the user that fresh gas flow is inadequate and room air is being entrained.
79
Raafat Abdel-Azim
2. If the reservoir bag is removed during mechanical ventilation, or if it develops a significant leak (from poor fit on the bag mount or a bag perforation), room air may enter the breathing circuit as the ventilator piston unit refills during the expiratory phase.
This could also result in dilution of either the inhaled anesthetic agent(s) concentration and/or the enriched oxygen mixture. This type of a disruption could lead to significant pollution of the operating room with anesthetic escaping into the atmosphere.
80
Raafat Abdel-Azim
Hazards of Ventilators
81
Raafat Abdel-Azim
• Apnea– Erroneous setting of ventilator selector
switch– Ventilator switched off– Disconnection– Obstruction of inspiratory gas pathway– Ventilator failure
Hazards
V B
82
Raafat Abdel-Azim
• Hypoventilation– Circuit leak– Incompetent spill valve– Open APL valve in circuit– Driving gas leak– Improper settings– Ventilator limitation at high airway
pressure
Hazards
PL
S
83
Raafat Abdel-Azim84
Raafat Abdel-Azim
Hyperventilation & dilution of anesthetic gasesBellows leak
85
Raafat Abdel-Azim
• High airway pressure
Exhaust obstruction
Obstruction of expiratory gas pathway
Ventilator cycling failure
Improper settings
Oxygen flush during inspiration
Bellows leak
Hazards
86
Summary of important pointsSummary of important points
Raafat Abdel-Azim
•Driving gas and patient gas don’t mix•The ventilation cycle is controlled by F, V, P and T•During the I phase, Paw depends on R and CT
•Goals of MV are to provide adequate VM (necessary to maintain
the desired PaCO2) and to side effects•Modern AV provide a range of ventilatory modes•The difference between the delivered VT and the preset VT depends on CC and FGF•FGD decreases the risk of barotrauma and volutrauma•Hazards of MV can be avoided
87
Examples of questions to assess the ILOsExamples of questions to assess the ILOs
Raafat Abdel-Azim
•Describe causes and ventilatory management of high airway pressure during inspiratory phase of mechanical ventilation•Discuss goals of pulmonary ventilation under anesthesia•With IPPV: if f (rate) = 10/min, I: E= 1:1 and FGF= 3 LPMWhat will be the volume added to the preset tidal volume?
88
Raafat Abdel-Azim
Thank you
http://telemed.shams.edu.eg/moodle
89