RA-Micro 007
Desktop Rotating Anode
X-Ray Generator
Service Manual
Document No. 4156A101
RA-Micro 007
Desktop Rotating Anode
X-Ray Generator
Service Manual
Document No. 4156A101
August 2002
Rigaku/MSC, Inc. shall not be liable for technical or editorial
omissions or mistakes in this manual. Nor shall it be liable for
incidental or consequential damages resulting from the use of the
information contained herein.
Reproduction, adaptation, or translation of any portion of this
publication
without prior written consent from Rigaku is prohibited, except as
allowed under copyright laws.
© Copyright 2002 by Rigaku/MSC, Inc., 9009 New Trails Drive, The
Woodlands, TX 77381.
All rights reserved.
Technical writer: M. Liliana Riahi
RA-Micro 007 Service Manual v
Contents
Chapter 1. Equipment Outline and Specifications ................................................................... 2
Specifications.............................................................................................................. 2
Utilities ....................................................................................................................... 2
Manual................................................................... Error! Bookmark not defined.
Construction ............................................................................................................... 3
Power Supply Requirements...................................................................................... 4
Voltage ............................................................................................................................................... 4
Frequency.......................................................................................................................................... 4
Installation ................................................................................................................. 5
Turning ON the System ............................................................................................. 9
Stability of Focus Position........................................................................................ 10
Focus Size ................................................................................................................. 10
Determining the Focus Size ..................................................................................... 12
RA-Micro 007 vs. Ultrax-18 ..................................................................................... 13
Chapter 2. Individual Units ......................................................................................................... 15
Power Supply............................................................................................................ 15
Power Section ........................................................................................................... 16
Subpanel ................................................................................................................... 18
Target Driver ............................................................................................................ 22
Check Terminal.................................................................................................. 22
LED Display...................................................................................................... 22
TMP Driver ............................................................................................................... 23
Shutter Circuit ......................................................................................................... 23
CPU Circuit .............................................................................................................. 25
Operation Panel........................................................................................................ 26
vi RA-Micro 007 Service Manual
Secret Command ............................................................................................... 26
Rotation of Target .............................................................................................. 26
Aging Mode......................................................................................................... 26
Check Flag Mode................................................................................................ 27
Dimming............................................................................................................. 27
X-Ray LED/Shutter Open LED ......................................................................... 27
Shutter FS Circuit.................................................................................................... 28
Shutter Drive Unit ................................................................................................... 28
Ready Conditions...................................................................................................... 29
Chapter 3. Characteristics ........................................................................................................... 31
Consumption of Electricity....................................................................................... 31
Vacuum ..................................................................................................................... 33
Filament.................................................................................................................... 34
Focus size .................................................................................................................. 35
Definition............................................................................................................ 36
Focus Size Measurement................................................................................... 36
BIAS Voltage ............................................................................................................ 38
BIAS Setting–Filament Current.............................................................................. 39
TMP Driver ............................................................................................................... 40
Aging ......................................................................................................................... 41
Check Terminal and LED of High Voltage Transformer........................................ 43
X-Ray Intensity Monitoring ..................................................................................... 46
X-Ray Intensity Changing........................................................................................ 47
FC and Focus Size .................................................................................................... 48
Chapter 4. Alarm and Countermeasures................................................................................... 51
Chapter 5. Drawing ....................................................................................................................... 55
2 RA-Micro 007 Service Manual
Chapter 1
Equipment Outline and
Specifications
Specifications
Refer to the publication titled RA-Micro 007 Desktop Rotating
Anode X-Ray Generator: Specifications Manual (Document No.
DBG-S001).
Utilities
Refer to the publication titled RA-Micro 007 Desktop Rotating
Anode X-Ray Generator: Utilities Manual (Document No. DBG-
U001).
Service
Please refer to this publication titled RA-Micro 007 Desktop
Rotating Anode X-Ray Generator: Service Manual (Document No.
4156A101).
RA-Micro 007 Service Manual 3
Construction
Name
Name of Components
Cat. No.
Drawing No.
Micro 007 Electrical components 4541N201 A792-001-3-12
Micro 007 Tube tower 4540H701 A792-001-3-13
Micro 007 Power supply 4541N101 A792-0026
RA-Micro 007 Right 4601L711 A116-0053
RA-Micro 007 Left 4601M111 A116-0054
R2HV cable 4259B2 4437-0049
Top table BS 2807F312 A340-6-1D
RA-Micro 007
4156A101
A792-001-3-11
Filament APBS D2H64T 4547-0072
Cathode
4479A118
5726-0069
DPTA2
4364F401
A347-001-1
4 RA-Micro 007 Service Manual
Power Supply Requirements
Power
Voltage
Frequency
Note
3 -phase AC 200V ± 10% 50/60 Hz ± 1% Standard specification.
3-phase AC 220V ± 10% 60 Hz ± 1% Standard specification.
3-phase AC 220V ± 10% 50 Hz ± 1% Roughing pump needs
to be changed.
Current and voltage
Three-phase AC 200V 3.5 kVA.
NOTE
More than 15 A must be used at the main breaker.
RA-Micro 007 Service Manual 5
Installation
X-ray generator unit
This unit provides a power supply for the RA-Micro 007. For the unit
to be operational, the ventilation cap must be set to operate rather
than to transfer.
Connection at power supply terminal
TB 401
Display
Work (See A792-0040H)
1 L1/U Connect wire at L1/U from main breaker.
2 L2/V Connect wire at L2/V from main breaker.
3 L3/W Connect wire at L3/W from main breaker.
4 PE/G Connect grounding wire from main breaker.
6 RA-Micro 007 Service Manual
TB 402
Display
Work (See A792-0040H)
1 L1/U Connect wire at L1/U of X-ray generator
power unit.
2 L2/V Connect wire at L2/V of X-ray generator
power unit.
3 L3/W Connect wire at L3/W of X-ray generator
power unit.
4 PE/G
XG-PS
Connect grounding wire to X-ray generator
power unit.
RA-Micro 007 Service Manual 7
TB 403
Display
Work (See A792-0040H)
1 COM
2 200V Jumper Pin 1 (COM) and Pin 2 (COM) when using AC
200 V and AC 208 V.
3 220V Jumper Pin 1 (COM) and Pin 3 (COM) when using AC
220 V.
4 240V
IG
TRANS
TAP
SELECT
Do not use.
5 Do not use.
6 Do not use.
7
8
100V
Radiation
Enclosure
AC 100 V power supply output when the power is ON
at subpanel.
MAX 0.4(A)
9
10
100V
CW
AC 100 V power supply output when the power is ON
at the subpanel. It is for the remote signal of the heat
exchanger.
MAX 0.1(A)
11 L1/U Connect the U line wire of the roughing pump.
12 L2/V Connect the V line wire of the roughing pump.
13 L3/W Connect the W line wire of the roughing pump.
14 PE/G
RP
Connect the ground wire of the roughing pump.
8 RA-Micro 007 Service Manual
Installing the Roughing Pump
To install the roughing pump, connect the power and the ground
wire at TB403 and connect the vacuum tube between the
turbomolecular pump (TMP) and the roughing pump.
Cooling the water line
Connect the water hose between the RA-Micro 007 IN, OUT and the
heat exchanger.
RA-Micro 007 Service Manual 9
Turning ON the System
After all the parts are assembled and connected, follow the steps below
to turn on the system:
1. Turn ON the main breaker on the wall. Be sure that the line
lamp in the power section lights up.
2. Turn ON the breaker on the system. Be sure that the line
lamp in the subpanel lights up.
3. Turn ON the power switch on the subpanel. There will be the
buzzing sound of cooling water.
4. Turn ON the heat exchanger. Be sure that the buzzing sound
disappears.
5. Be sure that Power Off Now is lit up in the display, then
push the Start button of the vacuum.
6. Be sure that the Ready light is on in the display when pushing
the Power ON button
7. The X-ray generator is now ON and running.
10 RA-Micro 007 Service Manual
Stability of Focus Position
The system needs 1.5 hours to maintain its focus position. Therefore,
after 1.5 hours, it is necessary to take measurements and to align the
CMF. To do this,
1. Set 600 dev at BIAS volume.
2. Execute a short in the aging mode.
NOTE
The best conditions for this step will occur automatically
after 1.5 hours.
Focus Size
To set the correct (official) focus size, follow the steps below:
1. Mount the front cover of the system, and then start the
vacuum.
2. Set the BIAS setting at 600 dev.
3. Turn the X-ray generator ON and execute aging in short time.
4. Connect the voltage meter (DC) to the FC check terminal.
RA-Micro 007 Service Manual 11
NOTE
It is necessary to wait for 1.5 hours.
Be sure that the X-ray generator is set at
40 kV/20 mA.
FC
5. After the filament change, the FC voltage must be checked.
At the FC check terminal, set the voltage to *.* V through the
BIAS setting.
NOTE
The BIAS knob may need to be adjusted to set the voltage
to *.* V.
CAUTION!!!
Gloves must be worn when working with a vacuum.
Do not apply tension to the filament.
+ –
12 RA-Micro 007 Service Manual
1.40
1.35
1.30
1.25
1.20
1.15
z ax
is [
mm]
1.851.801.751.701.65y axis [mm]
Determining the Focus Size
The tables below show the specifications for the focus size, as
measured with a CCD camera.
FWHM
FWHM Focus area (µmz)
Z(µ m) Y(µ m) Maximum Minimum
75 ± 15%
(64 ~ 86)
65 ± 15%
(55 ~ 75)
4900
(F 79)
3320
(F 65)
FWHM
FWHM Focus area (µ?)
Z(µ m) Y(µ m) Maximum Minimum
BIAS
setting
FC
(V)
Vaccum
(mV)
Focus picture
RA-Micro 007 Service Manual 13
RA-Micro 007 vs. Ultrax-18
There is a number of advantages to the RA-Micro 007 over the
Ultrax-18. Some of these advantages are as follows:
1. The focus size is only 70 µ m.
2. The capacity of the X-ray generator is 800 W.
3. The CE is specified.
4. There is only one copper target.
5. The tube tower is designed directly for the RA-Micro 007.
6. The HV bushing is designed directly for the RA-Micro 007, and
it is small. It is not a gas insulation.
7. The rotary shutter is designed directly for the RA-Micro 007.
8. The turbomolecular pump (TMP) is manufactured by the
Varian company.
9. The electronic component is stored in the subcabinet.
10. The cooling water flow rate is only 4 liters/min.
11. The operating power is AC 200 V 10 A.
12. It is not possible to increase the mA before reaching 40 kV.
13. It is not possible to decrease the voltage to < 40kV before
reaching 10 mA.
14. An error prompt may appear when the mA increase
command is issued at <39 kV.
14 RA-Micro 007 Service Manual
15. The aging mode is designed only for the RA-Micro 007.
During aging, if the UP or DOWN key is depressed, aging may
be canceled and the X-ray generator may be made to run
continually.
16. The target driver is of a new type (SGPXD-4). The target is a
DPTA2.
17. The turbomolecular pump (TMP) driver is of a circuit board
type manufactured by Varian.
18. The X-ray warning lamp does not have a power connector.
The RA-Micro 007 has 6 pins. The Ultrax-18 has 10 pins.
RA-Micro 007 Service Manual 15
Chapter 2
Individual Units
Power Supply
The power supply for the X-ray generator
• is named RMN1,
• is of 2 kW, and
• is made by RMS.
Drawing: A792-26-1D
A845-001-3-2.
NOTE
For more detailed information, see the RA-Micro 007
Desktop Rotating Anode X-Ray Generator:
Maintenance Manual.
16 RA-Micro 007 Service Manual
Power Section
The power section supplies power to each component of the RA-Micro
007, with the terminal being located in the back of the unit, in the
right lower part:
1. Terminal TB 401
Connect the power supply wires at TB 401 from the main
breaker on the wall.
2. Line lamp
The line lamp lights up when it receives AC 200 V power.
3. Circuit protector 1 (CP 1)
Circuit protector 1 trips when too much power flows to the
line lamp.
4. Main breaker (ELB 1)
The current at the main breaker is 15 A, and the electrical
leakage is 30 mA.
5. Circuit protector 2 (CP 2)
Circuit protector 2 trips when the Varista jumper lights up or
there is an abnormal input impulse.
6. Terminal TB 402
The TB 402 terminal provides a power supply for the X-ray
generator.
RA-Micro 007 Service Manual 17
7. Terminal TB 403
The TB 403 provides a power supply for each of the
components of the unit.
NOTE
See the TB 403 table in the section titled Connection at
Power Supply Terminal in Chapter 1 on p. 6.
8. Circuit protector 3 (CP 3)
Circuit protector 3 trips when the current flows more than
0.5 A at TB 402 7–8, 9–10.
9. Electric contactor RY 2
The electric contactor RY 2 is for the roughing pump. The
thermal relay should be set to 2.6 A.
18 RA-Micro 007 Service Manual
Subpanel
The subpanel is located under the OP panel. This subpanel displays
the hour meter and the status of each component. It also manages
the ON/OFF power control and the turbomolecular pump (TMP).
1. Main power switch
This switch controls the ON/OFF power and has a lamp that
lights up when the switch is turned ON. Inside the cabinet,
there is also a fan, which rotates when the main power switch
is turned ON.
NOTE
The main power switch cannot be turned on if the fan is
not rotating.
2. Hour meter
There is a reset switch in the hour meter under the black line.
IMPORTANT!!!
It appears to be best not to tell the customer about the
reset switch.
3. Focus selection switch
This switch controls the selection of the focus size. This size
must be set at 0.07. It is the only setting on the RA-Micro
007.
RA-Micro 007 Service Manual 19
4. Target selection switch
This switch is used to select the target material and must be
set at Cu. (Other targets are optional.)
5. BIAS setting knob
The BIAS must be set at 0.07. The BIAS setting knob will set
the RA-Micro 007 to *.* V at the FC terminal.
6. Turbomolecular pump (TMP)unit display
The TMP unit will display the following:
• a lighted power lamp when the main power is supplied by
the TMP driver;
• a lighted acceleration lamp when the TMP accelerates;
• a lighted run lamp when the final rotation speed is reached;
• a lighted alarm lamp when an abnormal condition occurs in
the rotation or the temperature of the unit.
7. TMP check terminal
The actual rotation speed output is 0–10 V between COM and
speed. The output is 10 V when the run lamp is lit (run = 75
krpm). The TMP load output is between COM and power.
1 V is about 10 W.
NOTE
See the section on the TMP driver below for driver
specifications.
20 RA-Micro 007 Service Manual
8. Shutter mode selection
The shutter mode switch has the following selections:
• L, for left;
• R, for right;
• XG, for operating the shutter with the OP panel and the
computer;
• Hose, for operating the shutter with the R-Axis unit or the
RCD.
9. Buzzer switch
This switch controls whether the buzzer is to be ON or OFF.
10. Auxiliary terminal
This terminal offers the following selections:
• Power ON, which yields a maximum AC 120 V–2 A;
• X-ray ON, which yields a maximum AC 120 V–2 A.
11. Vacuum output terminal
The output range of this terminal is 0–500 mV at 1 x 10ˆ – 1,
1 x 10ˆ – 5.
12. FC check terminal
The output of this terminal is the primary filament current.
NOTE
See the section on Filament specifications in Chapter 3
on p. 34.
RA-Micro 007 Service Manual 21
13. Sequencer
This item manages the vacuum system.
14. TMP driver
The TMP driver is of the circuit board type.
15. BIAS reference voltage
The BIAS reference voltage can be measured. It should be
measured when it is abnormal.
Setting: +10 (V) (+9.995 V ~ +10.005 V).
22 RA-Micro 007 Service Manual
Target Driver
The model number of the target driver is SGPXD-40. This target
driver is a newer version of the S30T. The SGPXD-40 provides
• economy of electricity,
• less noise, and
• fewer vibrations.
Check Terminal
Check
Point
Name
Sub
Name
Note
TP1 Sensor test point HX ±13 V ~ ±14 V
TP2 Monitor test point pul ±13 V ~ ±14 V
GND Signal common GND
LED Display
Name
LED Display
RDY Green color when in Ready mode.
Red color when in Alarm mode.
SRV Green color when in Run mode.
Red color when in Alarm mode.
RA-Micro 007 Service Manual 23
NOTE
When the alarm LED is lit, push the reset switch or the
ON switch .
TMP Driver
The specifications of the TMP driver are as indicated in the table.
TMP Driver
Model number 969-9509
Manufacturer Varian
Input voltage
Frequency
Power
90 V ~ 265 V
50/60 Hz
150 VA maximum
Output voltage
Frequency
Power
54 Vac ±10%
1250 Hz
70 W
Revolutions 75,000 r/min
Shutter Circuit
This circuit monitors the shutter status. The circuit will shut off the
X-ray when an abnormal condition occurs. The monitor will close the
sensor in the shutter.
24 RA-Micro 007 Service Manual
Card Rack
At the lower right in the cabinet, there is a card rack that holds the
switching regulators.
Specifications for the switching regulators.
P/N and
Manufacturer
Input
Output
Note
PS1 LWT30H-5FF/ Densei-Lambda
85~ 265VAC 0.7A
5V/5A? +15V/1.2A? -15V/0.6A
For control circuit.
PS2 RWS 30A-24/ Densei-Lambda
85~ 265VAC 0.7A
24V/1.3A For sequencer and warning lamp.
NOTE
Press keys F1 and F2 to activate the 3.15-A circuit
protector whenever an abnormal current flow is tripped
within the main power panel.
RA-Micro 007 Service Manual 25
CPU Circuit
The CPU circuit controls the X-ray generator.
ROM: Version 4.10 (12/13/1001)
IC3 in A171-14-ID
Rotary switch 2
(kV maximum)
6 (0 ~ F)
Rotary switch 3
(mA maximum)
D (0 ~ F)
D.W. switch 5 1 – > ON
2 – > OFF
3 – > OFF
4 – > OFF
5 – > OFF
6 – > OFF
7 – > ON
8 – > OFF
26 RA-Micro 007 Service Manual
Operation Panel
The operation panel controls the operation of and information in the
X-ray generator.
ROM version 3.10 (13/12/2001)
IC10 in the A171-22-ID
D.W. S2 1 – > OFF
2 – > OFF
3 – > OFF
D.W. S4 1 – > ON
2 – > ON
3 – > OFF
4 – > OFF
Secret Command (Password?)
The RA-Micro 007 requires no password.
Rotation of Target
The target may be rotated only in the Manual mode.
Aging Mode
There are four aging modes in the RA-Micro 007.
NOTE
The aging process will stop and the kV and mA settings
will be kept when the UP and DOWN keys are pressed
during aging.
RA-Micro 007 Service Manual 27
1. Short time mode: When the system is running, it is generally
in the short time mode. One hour of running time is needed
to reach the 40 kV and 20 mA level.
2. Medium time mode: When the filament is replaced. The
system is in the medium time mode. 12 hours of running time
are needed to reach the 40 kV and 20 mA level.
3. Long time mode: This mode is used for the aging of the new
tube tower. The factory uses this mode when running the first
aging. In this mode, 48 hours of running time are needed to
reach the 40 kV and 20 mA level.
4. Manual mode: This mode may be used for the aging
procedure.
Check Flag Mode
This mode is used for monitoring all the flag statuses in the CPU
circuit. When all the flags are normal, the check flag mode is at
ready.
• Input port: The input port displays the port condition.
• Echo back: The echo back displays communication
between the computer and the RA-Micro 007.
Dimming
This feature adjusts the brightness in the operation panel. The
standard brightness is 60%.
X-Ray LED/Shutter Open LED
This feature monitors the jumper and the open mode in the LED. The
LED must be replaced when all the LEDs are lit up.
28 RA-Micro 007 Service Manual
Shutter FS Circuit
This circuit controls two of the rotary shutters and the FS unit. To
operate the shutter, the shutter driver circuit must be installed
(A105-003, which is included in the R-Axis controller).
NOTE
When the R-Axis is not used, the shutter driver circuit
must be ordered.
Shutter Drive Unit
The RA-Micro 007 contains two rotary shutters. The A171-0035
(circuit board control) is included in the RA-Micro 007. However,
there is no shutter drive circuit board in the RA-Micro 007 because it
is included in the R-Axis and in the RCD.
D. S. S1 1 – > OFF
2 – > ON
3 – > OFF
4 – > OFF
5 – > ON
6 – > OFF
D. S. S2 1 – > ON
2 – > ON
3 – > OFF
4 – > OFF
RA-Micro 007 Service Manual 29
If the R-Axis and the RDC are not ordered, then the shutter driver
unit must be installed in the RA-Micro 007.
Ready Conditions
The ready conditions are as follows:
1. No alarm is active.
2. Shutters are closed (by the close sensor).
3. All circuit boards are installed in the RA-Micro 007.
4. There is more than 10ˆ4 Pa (250 mV less) in the vacuum.
5. The radiation enclosure doors are closed.
The power ON conditions are as follows:
• The shutter mounting sensor is not visible.
• The water leaking sensor is detected.
30 RA-Micro 007 Service Manual
This page intentionally left blank.
RA-Micro 007 Service Manual 31
Chapter 3
Characteristics
Consumption of Electricity
Condition A B C D Voltage (V)
Current (A)
Current (A)
C urrent (A)
Current (A)
180 9.95 11.42 11.34 11.26200 8.25 10.10 9.71 9.92208 7.94 9.53 9.49 9.45220 7.65 9.07 8.77 8.56242 7.37 8.42 8.41 8.04
180 9.09 10.98 10.60 10.66200 7.95 9.63 9.33 9.17208 7.57 9.01 8.83 8.69220 7.26 8.53 8.38 8.14242 6.86 8.17 7.84 7.61
Load 0.8 1.2 1.2 1.2
Frequency (Hz)
50
60
Difference between 50 Hz and 60
6.0 7.0 8.0 9.0
10.0 11.0 12.0
170 180 190 200 210 220 230 240 250 (V)
(A)
50 Hz 60 Hz
32 RA-Micro 007 Service Manual
40 kV, 20 mA, BIAS 156 V (950) 40 kV, 30 mA, BIAS 0 V (000)
Condition A B
Frequency
( Hz)
Voltage
( V)
Effective
electricity
(W)
Invalid
electricity
(Var)
Superficial
electricity
(VA)
Rate
(%)
Effective
electricity
(W)
Invalid
electricity
(Var)
Superficial
electricity
(VA)
Rate
(%)
50 200 2302 508 2366 0.97 2845 590 2913 0.98
200 2442 179 2453 1.00 2966 194 2978 1.00
208 2427 187 2436 1.00 2936 215 2947 1.00 60
220 2485 253 2505 0.99 2986 269 3009 0.99
Load (kW) 0.8 1.2
0.8(kW)
2200
2300
2400
2500
2600
2700
2800
2900
3000
3100
195 205 215 225 Volt (V)
(W)?
) (VA
)
Effe 50Hz Supe 50Hz
Effe 60Hz Supe 60Hz
RA-Micro 007 Service Manual 33
Vacuum
1.0E-06
1.0E-05
1.0E-04
1.0E-03
1.0E-02
1.0E-01
1.0E+00
0 50 100 150 200 250 300 350 400 450 500
IG? ? ? ? ? ? ( mV)
??
??
(Pa
)
IG terminal output (mV) Actual vacuum and Output of IG terminal
Vacu
um
(Pa)
34 RA-Micro 007 Service Manual
Filament
4.000
5.000
6.000
7.000
8.000
9.000
10.000
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00
FC? ? ? ? ? ? ( FCMF) ( V)
??
??
??
??
(A
)
FC terminal (FCMF) (V) Filament current and FC monitor output (FCMF)
FCL: A (at 40kV, 20mA, BIAS950) +10% Detect at 7.7 A Must measure at FCMF after 1.5 hours 40 kV-20 mA
(because of the drifting FCMF temperature).
Filam
ent cu
rrent (A
)
RA-Micro 007 Service Manual 35
Focus size
0.0
0.1
0.2
0.3
0.4
0.5
0 100 200 300 400 500 600 700 800 900 1000
? ? ? ? ? ?
??
??
?
(mm
)
FW10%M_X
FW10%M_Y
FWHM_X
FWHM_Y
BIAS deviation
This is a reference example chart between focus size and BIAS. BIAS setting is at about 866 = 156 V .
Focu
s size (mm
)
36 RA-Micro 007 Service Manual
1.40
1.35
1.30
1.25
1.20
1.15
z ax
is [
mm]
1.851.801.751.701.65y axis [mm]
Definition
The focus size of the RA-Micro 007 must be less than 80 µm in the X
and Y directions.
The X-ray focus size is the same as that using FWHM (X–Y).
Focus Size Measurement
The expansion rate should be 3–4 times.
Glancing angle must be 6 deg.
40 kV 20 mA BIAS 950 (148.2 V)
RA-Micro 007 Service Manual 37
100
80
60
40
20In
tens
ity
[arb
.uni
t]
-0.10 0.00 0.10y-axis [mm]
120
100
80
60
40
20
0Intensity [arb.unit]
0.15-0.15Z-axis [mm]
38 RA-Micro 007 Service Manual
BIAS???? BIAS? ? ? ? ? ?
0.0
50.0
100.0
150.0
200.0
250.0
300.0
350.0
400.0
450.0
0 100 200 300 400 500 600 700 800 900 1000BIAS???
BIAS??? ?
(V) Focus 0.1 40kV
Focus 0.1 50kVFocus 0.1 60kVFocus 0.5 40kVFocus 0.5 50,60kV
BIAS Voltage
RA-Micro 007 = Focus 0.1
BVMF---BIAS? ? ? ?y = 103.6x + 0.1479
0
50
100
150
200
250
300
350
400
450
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
BVMF(V)
BIAS
? ?(V
)
40kV 45kV 50kV Linear (50kV)
RA-Micro 007 Service Manual 39
BIAS Setting–Filament Current
40kV, 20mA
(Reference example)
FCMF must be measured after 1.5 hours because of the drifting temperature.
BIAS 866 = about 156 V.
BIAS? ? -FC (FC M F)? ?
5.00
5.50
6.00
6.50
7.00
7.50
0 100 200 300 400 500 600 700 800 900 1000BIAS? ?
FC(FCMF)(V)
40 RA-Micro 007 Service Manual
TMP Driver
Acceleration time around 1 min
Power at RUN About 13 W (1.3 V)
At RUN 75 k(r/min)(10V)
Stop without leak
RPM
Power
Power during acceleration time 86 W (8.6 V)
RA-Micro 007 Service Manual 41
Aging
The RA-Micro 007 has four aging modes.
Aging will stop and keep the present kV and mA when the UP or
DOWN keys are pressed.
AGING MODE: Short
0
10
20
30
40
50
60
0 60 120 180 240 300 360 420 480 540 600 660 720 780 840AGING TIME (min)
kV,mA
kV mA
After 1 hour 40 kV, 20 mA
After 12 hours X-ray OFF
42 RA-Micro 007 Service Manual
0
10
20
30
40
50
60
0 120 240 360 480 600 720 840 960 1080 1200 1320 1440
Aging Time (min)
kV, m
A
kV mAAGING MODE: Middle
After 24 hours X-ray OFF
After 12 hours 40 kV, 20 mA
RA-Micro 007 Service Manual 43
AGING MODE: Long
0
10
20
30
40
50
60
70
0 720 1440 2160 2880 3600 4320 5040AGING TIME (min)
kV,
mA
kV mA
After 48 hours 40 kV, 20 mV
After 72 hours X-ray OFF
44 RA-Micro 007 Service Manual
Check Terminal and LED of High Voltage Transformer
Display Action or Name Voltage LED or color
Power When provided power in the transformer.
Green LED
KVMR KV reference 6 V/–60kV Red color KVMF KV feed back 6 V/–60kV Red color MAMR MA reference 3 V/30 mA Red color MAMF MA feed back 3 V/30 mA Red color FCMR Filament current reference 8.23 V Red color FCMF Filament current feedback See p. 15:
Graph FCMF and filament current
Red color
BVMR Bias reference 4 V/– 400 V Red color BVMF Bias feedback 4 V/– 400V Red color + 5 V + 5V + 5 V Red color + 15 V + 15 V + 15 V Red color + 15 V + 15 V + 15 V Yellow
color +15 V(A) + 15 V + 15 V Red color COM Common Black color ARC ALARM
Arcing detection select ON/OFF switch.
Switch
XON Lit when X-ray ON. Green LED kV min OK
Lit when kV is more than 17 kV.
Green LED
MA min OK
Lit when mA is more than 1.8 mA.
Green LED
PSFLT Fan abnormal rotation, Thermal sensor, Fuseburned,
Arc detected, Power supply voltage
abnormal, Resister between BIAS and filament less than 30 KO when lit one above alarm.
Red color
HVL Lit when kV is more than 63 kV.
Red color
LVL Lit when kV is less than 17 kV.
Red color
TCL Lit when mA is more than 66 mA .
Red color
OLL Lit when passed + 5% OLL setting (0.8 kW).
Red color
RA-Micro 007 Service Manual 45
Display Action or Name Voltage LED or color
LCL Lit when flowed 150 A in IGBT.
Red color
FUSE Lit when burned out fuse. Red color THERM Lit when passed 60 deg in
inverter and converter heat sink.
Red color
FAN Lit when detected fan as abnormal.
Red color
ARC Lit when detected arcing. Red color UVL Lit when power supply
voltage is less than 170 V. Red color
BFS Lit when make short between filament and cathode or resister is less than 30KO.
Red color
46 RA-Micro 007 Service Manual
X-Ray Intensity Monitoring
150000
100000
50000
0
Inte
nsity
[cps
]
2.42.22.01.81.61.41.21.00.80.60.40.20.0time [hour]
warmup time 100min
X-ray intensity with Cu0.2foil
X-ray intensity monitoring over time.
RA-Micro 007 Service Manual 47
X-Ray Intensity Changing
X-ray intensity changing over time.
180000
175000
170000
165000
160000
155000
150000
Inte
nsity
[cps
]
80006000400020000time [sec]
26
24
22
20
18
tube
tem
p [C
°]
5000cps
1000cps
5000cps
1000cps
intensity water temp
48 RA-Micro 007 Service Manual
FC and Focus Size
Below is an FC and a focus size characteristic chart. The correct
(official) focus size may be obtained by setting the FC voltage at *.* V
as in *.* V (FCMF) = 6.2 V + FC Offset at 40 kV–20 mA.
6.2 V = Constant voltage for the RA-Micro 007.
FC offset = Every RA-Micro 007 has a typical offset value.
Focus characteristic in 5 of RA-Micro 007.
FC-FOCUS(FWHM Z) 5FCMF-Offset = 6.2 V
50556065707580859095100105110115120125
5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4
FCMF-Offset (V)
Z?FWHM??µm)
02-0.19
02-0.13
02-0.19(2)
02-0.145
02-0.08
02-0.115
02-0.08(2)
02-0.15
01-0.18
01-0.17(A)
01-0.15
01-0.17(B)
01-0.12
01-0.17
01-0.19
Pass? MAX
Pass MIN
01-0.15± 0.05
03-0.15± 0.05
04-0.17
05-0.15± 0.05
04-0.17-10.95
04-0.17-11.05
04-0.17-11.20
RA-Micro 007 Service Manual 49
Focus characteristics in 5 of RA-Micro 007.
FC-FOCUS (FWHM Y) 5FCMF-Offset = 6.2 V
50
55
60
65
70
75
80
85
90
95
100
105
5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40
FCMF-Offset (V)
Y FW
HM (
µm)
02-0.19
02-0.13
02-0.19(2)
02-0.145
02-0.08
02-0.115
02-0.08(2)
02-0.15
01-0.18
01-0.17(A)
01-0.15
01-0.17(B)
01-0.12
01-0.17
01-0.19
Pass MAX
Pass MIN
02-0.15± 0.05
03-0.15± 0.05
04-0.17
05-0.15± 0.05
04-0.17-10.95
04-0.17-11.05
04-0.17-11.20
50 RA-Micro 007 Service Manual
This page intentionally left blank.
RA-Micro 007 Service Manual 51
Chapter 4
Alarm and Countermeasures
The operation panel has 15 alarms. These alarms are as follows:
1. High voltage limit alarm: Goes off when the voltage
exceeds –63 kV.
Countermeasures
HVL alarm is triggered mostly by arcing in the tube tower.
Find arcing mark and remove it with sand paper.
Check arcing mark on the target surface.
2. Low voltage limit alarm (HVL): Occurs when the voltage
exceeds –17 kV.
Countermeasures
Check the register between L and Ground C and Ground S
after removing the high voltage cable from the transformer.
Check the voltage at the kVMR.
3. Over load alarm (OLL): Occurs when too much current has
flowed in kV and mA.
Countermeasures
The OLL alarm occurs in the case of arcing in the tube tower
and on the target surface. This is the same as the HVL alarm.
52 RA-Micro 007 Service Manual
4. Filament current limit alarm: Occurs when 7.7 A of current
has flowed into the filament or when an open circuit occurs in
the filament.
Countermeasures
Check the resistance between L and C after removing the high
voltage cable at the transformer. The display should read
open.
5. Tube current limit alarm: Occurs when the current in the
tube exceeds 66 mA.
Countermeasures
This may cause arcing in the tube. Check the arcing mark in
the tube tower.
6. PSFLT alarm: The PSFLT alarm will be visible whenever the
following conditions occur:
• The fan rotation is abnormal (FAN).
• The inverter heat sink is over 60 deg (THERM).
• The fuse is burned out (FUSE).
• Arcing is detected when the power supply voltage is
below 170 V (UVL).
• The current flow exceeds 150 A in the inverter (LCL).
• A jumper occurs between the BIAS and the filament (BFS).
• Insulation below 30 kO registers between the BIAS and the
filament (BFS).
Countermeasures
Be sure about which alarm occurred in the high voltage
transformer.
RA-Micro 007 Service Manual 53
7. Target rotation limit alarm: Occurs when a rotation speed
of < 5700 rpm is reached.
Countermeasures
Check the T1 frequency. It must be 400 Hz.
The target rotation limit is 95% of the rotation speed.
SD circuit board (A 287-008).
TP-25 = input.
TP-24 = output of LM2917.
TP-9 = target rotation limit set point (95% of TP-24).
8. Cooling water flow alarm: Occurs when the water flow rate
is <3.5 liters/min.
Countermeasures
Check the cooling water flow line.
9. Cooling water pressure alarm: Occurs when the water
pressure is >0.34 Mpa.
Countermeasures
Check the cooling water flow line.
10. Cooling water temperature alarm: Occurs when the
temperatures in the water line is over the set point.
Countermeasures
Check the cooling water flow line.
54 RA-Micro 007 Service Manual
11. Vacuum alarm: Occurs when the pressure in the tube tower
exceeds 1 x 10 –2 Pa and the X-ray shuts off.
Countermeasures
Check the filament in the RIG and check for leaking from the
tube tower.
12. TMP overload alarm: When an overload occurs in the TMP
driver or an abnormal temperature occurs in the TMP.
Countermeasures
• Check the TMP rotation by hand.
• Check for leaking in the tube tower.
• Listen for a “strange” sound.
13. X-ray warning lamp alarm: Occurs when a short circuit
takes place in the X-ray warning lamp.
Countermeasures
Check the connection at the connector. This alarm is indicated
by an LED display.
14. Operation panel X-ray lamp alarm: Occurs when the
operation panel is open or when a short circuit takes place in
the X-ray lamp of the panel.
Countermeasures
Check whether the X-ray lamp is ON. The X-ray ON is
indicated by an LED display.
15. Cooling water leakage alarm: Occurs when a water leak is
detected. This is an optional function not included in the
standard options.
RA-Micro 007 Service Manual 55
Chapter 5
Drawing
Tube tower
56 RA-Micro 007 Service Manual
This page intentionally left blank.
RA-Micro 007 Service Manual 57
Index A
acceleration lamp................................... 19
aging mode .............................. 9, 13, 26,41
alarm .................................... 51, 52, 53, 54
alarm lamp............................................. 19
alarm mode ............................................ 22
auxiliary terminal.................................. 20
B
BIAS deviation....................................... 35
BIAS reference....................................... 21
BIAS setting............... 9, 10, 11, 19, 35, 39
BIAS voltage .......................................... 38
BIAS volume ............................................ 9
buzzer switch ......................................... 20
C
card rack................................................. 24
check flag mode...................................... 27
circuit board ........................................... 13
circuit protector.......................... 16, 17, 24
consumption of electricity...................... 31
CPU circuit............................................. 25
current...................................................... 3
D
dimming ................................................. 27
E
electric contactor .................................... 17
error prompt........................................... 12
F
FC check terminal.................................... 9
filament .................................................. 34
filament change...................................... 10
filament current..................................... 39
focus position............................................ 9
focus size .......................... 9, 11, 18, 36, 48
G
ground wire .............................................. 6
H
heat exchanger................................. 6, 7, 8
hour meter.............................................. 18
L
line lamp............................................. 8, 16
long time mode....................................... 27
M
main breaker.............................. 3, 4, 8, 16
Maintenance Manual ............................. 15
manual mode.......................................... 27
medium time mode ................................ 27
58 RA-Micro 007 Service Manual
O
operation panel ...................................... 26
P
power supply ................................ 3, 15, 17
R
ready conditions..................................... 29
ready mode............................................. 22
reset switch ............................................ 18
rotary shutter................................... 12, 28
roughing pump................................... 7, 17
run lamp................................................. 19
run mode ................................................ 22
S
Service Manual ........................................ 1
short circuit............................................ 54
short time mode ..................................... 27
shutter circuit ........................................ 23
shutter driver......................................... 29
shutter mode .......................................... 20
shutter mounting................................... 29
Specifications Manual ............................. 1
T
target driver ..................................... 13, 22
target selection....................................... 19
TMP driver........................... 19, 21, 23, 40
TMP rotation.......................................... 54
tube tower ........................................ 12, 55
turbomolecular pump (TMP) 7, 12, 13, 18,
19
U
U line wire................................................ 6
Utilities Manual....................................... 1
V
V line wire ................................................ 6
vacuum................................................... 33
vacuum output ....................................... 20
vacuum system....................................... 21
vacuum tube............................................. 7
ventilation cap.......................................... 4
voltage ...................................................... 3
voltage meter ........................................... 9
W
W line wire ............................................... 6
water flow line........................................ 53
water flow rate ................................. 12, 53
water hose ................................................ 7
water line ................................................. 7
X
X-ray generator 4, 5, 8, 9, 10, 12, 13, 15,
16, 25, 26
X-ray intensity ................................. 46, 47
Rigaku/MSC, Inc.
9009 New Trails Drive
The Woodlands, TX 77381
U.S.A.
Telephone: (281) 363-1033
Facsimile: (281) 364-3628
E-mail: [email protected]
http://www.RigakuMSC.com
Rigaku/MSC, Inc.
(European Branch)
Unit D2, Chaucer Business Park
Watery Lane, Kemsing
Sevenoaks, Kent
TN15 6YU England
U.K.
Telephone: (44) 1732 763367
Facsimile: (44) 1732 763757
E-mail: [email protected]
Rigaku/MSC, Inc.
9009 New Trails Drive
The Woodlands, TX 77381
Telephone: (281) 363-1033 or 1-800-543-2379
Printed in the U.S.A.