Exosomes: Emerging Therapy Delivery Tools and Biomarkers ...
R-3Oxygen Therapy and Oximetry in the Delivery Room
-
Upload
redina-tuya-ramirez -
Category
Documents
-
view
214 -
download
0
Transcript of R-3Oxygen Therapy and Oximetry in the Delivery Room
-
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
1/6
Review
Oxygen therapy and oximetry in the delivery room
Yacov Rabi a,b,*, Jennifer A. Dawson c,d,e
a Division of Neonatology, Department of Paediatrics, University of Calgary, Calgary, Alberta, CanadabAlberta Childrens Hospital Research Institute, Calgary, Alberta, Canadac Newborn Research Centre, The Royal Womens Hospital Melbourne, Melbourne, Australiad The Murdoch Childrens Research Institute, Melbourne, Australiae University of Melbourne, Melbourne, Australia
Keywords:
Delivery room
Heart rate
Oxygen saturation
Pulse oximetry
s u m m a r y
Pulse oximetry is increasingly being used in the delivery room. Expert recommendations state that
oxygen therapy during newborn resuscitation should be guided by pulse oximetry. Obtaining accurate
and stable oxygen saturation and heart rate information from a pulse oximeter in the delivery room can
be challenging. Understanding the properties of this device is important in overcoming these challenges.
This article describes several aspects of pulse oximetry use in the delivery room ranging from technical
issues with the device itself to clinical applications of the technology.
Crown Copyright 2013 Published by Elsevier Ltd. All rights reserved.
1. Introduction
The practice of transmitting light through tissue to measure a
patients oxygenation is nearly 80 years old. Early iterations of thistechnology could not differentiate oxygen saturations of arterial
blood from venous blood and tissue. In the early 1970s a Japanese
biomedical engineer, Takuo Aoyagi, discovered that the pulsatile
changes in light transmission through tissue could be used to
measure oxygen saturations[1] and pulse oximetry was born. In
1978, only 6 years after its discovery, the Minolta Company was the
rst to broadly market this new piece of technology. It was a major
breakthrough and quickly adopted into clinical settings.
Pulse oximetry use in the delivery room has been investigated
for more than 30 years. The earliest studies were published in 1986.
Sendak et al. [2] reported on pulse oximetry measurements in a
case series of four newborns at birth. That year, Harris et al. [3]
published a study describing continuous oxygen saturations for
the rst seven minutes after birth in 76 term newborns. A similarstudy published the following year concluded that pulse oximetry
use in the delivery room was very useful in objectively judging the
adequacy of resuscitative efforts[4].
More recently, we have seen renewed interest in this technology
for use in the delivery room. This was largely prompted by the
realization of the dangers of hyperoxia during resuscitation and the
subsequent International Liaison Committee on Resuscitations
(ILCOR) recommendation that oxygen concentration should be
titrated during resuscitation [5]. Clinicians now have to consider
what methods are best for guiding oxygen use in the delivery room.In this article, we review how pulse oximetry works as a prelude to
understanding its strengths and limitations in the delivery room
setting.
2. How does pulse oximetry work?
Oxygen saturation is the percentage of hemoglobin that is bound
with oxygen, also called oxyhemoglobin. Oxygenated and deoxy-
genated hemoglobin absorb light differently, with oxygenated he-
moglobin maximally absorbing light in the infrared spectrum at
660 nm compared to deoxyhemoglobin which maximally absorbs
light in the red spectrum at 910e940 nm. Pulse oximeters calculate
the ratio of tissue absorptions of red and infrared light to derive the
oxygen saturation of hemoglobin (SpO2). Two wavelengths of light,660 nm and 910e940 nm, are directed intoa tissuebed and a sensor
placedon theopposite side of the light source measures the amount
of light absorbed at both wavelengths. The absorption ratio is then
compared to reference values obtained from healthy volunteers
who were made hypoxic with saturations as low as 60e70% under
controlled conditions. This is why manufacturers only quote the
accuracy of their devices for oxygen saturations above 60e70%.
Pulse oximeters also display pulse rate. In a well-perfused per-
son, each heartbeat results in a pulsatile increase in the blood
volume of tissue beds. With larger tissue blood volumes, more light
from the pulse oximeter probe is absorbed causing less light to
* Corresponding author. Address: Foothills Medical Center, 1403 29 Street NW,
Room C211, Calgary, Alberta, Canada T2N 2T9. Tel.: 1 403 944 1087; fax: 1 403
944 4892.
E-mail address:[email protected](Y. Rabi).
Contents lists available at ScienceDirect
Seminars in Fetal & Neonatal Medicine
j o u r n a l h o m e p a g e : w w w . e l s e v i e r . co m / l o c a t e / s i n y
1744-165X/$e see front matter Crown Copyright 2013 Published by Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.siny.2013.08.007
Seminars in Fetal & Neonatal Medicine 18 (2013) 330e335
mailto:[email protected]://www.sciencedirect.com/science/journal/1744165Xhttp://www.elsevier.com/locate/sinyhttp://dx.doi.org/10.1016/j.siny.2013.08.007http://dx.doi.org/10.1016/j.siny.2013.08.007http://dx.doi.org/10.1016/j.siny.2013.08.007http://dx.doi.org/10.1016/j.siny.2013.08.007http://dx.doi.org/10.1016/j.siny.2013.08.007http://dx.doi.org/10.1016/j.siny.2013.08.007http://www.elsevier.com/locate/sinyhttp://www.sciencedirect.com/science/journal/1744165Xhttp://crossmark.crossref.org/dialog/?doi=10.1016/j.siny.2013.08.007&domain=pdfmailto:[email protected] -
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
2/6
reach the sensor. The pulse oximeter measures heart rate by
detecting this pulsatile change in light absorption.
The newest generation of pulse oximeters uses multiple wave-
lengths of light to measure concentrations of other types of he-
moglobin, including methemoglobin and carboxyhemoglobin.
Whereas these are of limited clinical signicance for the newborn
at birth, fetal haemoglobin may be important. Fetal hemoglobin
binds oxygen more tightly than adult hemoglobin and constitutes
close to 100% of the hemoglobin in very lowbirth weight babies [6].
Currently, pulse oximeters cannot detect fetal hemoglobin. The
practical implications are discussed later in this article.
Pulse oximeters are calibrated by their manufacturers and
cannot be adjusted by clinicians. Internal calibration occurs auto-
matically within the oximeter and, unlike other devices, does not
require an external reference standard.
3. Evidence for best practices in using pulse oximetry in the
delivery room
The transitional physiology of the newborn at birth differs
greatly from that encountered in the neonatal intensive care unit
and provides unique challenges to using pulse oximetry in the de-
livery room. Poor perfusion, motion and high ambient light can allbe present during resuscitation. Choosing the appropriate settings
can aid in rapidly obtaining accurate and stable measurements in
this environment. In addition to choosing the optimal settings on
the pulse oximeter, the user should consider how and where to
apply the sensor. Below, we provide evidence-based recommen-
dations for proper use in the delivery room.
3.1. .Pulse oximeter
In an effort to provide stable readings, pulse oximeters display a
moving average of their measurements over a user-specied time-
period. Whereas the options vary by manufacturer, the shortest
available selectable averaging time is typically 2 s. Some pulse
oximeters do not allow the user to specify the averaging time; thisis set by the device and may vary according to signal quality [7].
Leone and Finer[8] recommended that oximeters used during
neonatal resuscitation should be set to the minimal averaging time
for the SpO2 valuesto allow rapid detection of changes in oxygen
saturation. The detection of brief and severe periods of desaturation
is improved with a shorter averaging time (2 s) compared to longer
averaging times (16 s)[9]. At present, there are no outcome studies
examining the effects of averaging time. However, given the current
state of knowledge, we recommend selecting the shortest averaging
time available when the pulse oximeter is used in thedelivery room.
The sensitivity setting on a pulse oximeter controls the devices
ability to detect pulsations. A higher sensitivity will increase
detection of pulsations and therefore is more likely to provide SpO2
measurements during periods of low perfusion. Studies of pulseoximetry in the delivery room routinely used the highest sensitivity
setting available [9e13]. Again,in the absence of outcome studies, it
is reasonable to use a high sensitivity setting in situations where
the patient is poorly perfused.
3.2. Application of pulse oximeter probe
Care must be taken not to apply the probe too tightly. Bucher
et al. [14]showed that SpO2 measurements became increasingly
inaccurate with increasing pressure from the pulse oximeter probe
on the underlying tissue. The authors concluded that this was likely
due to venous congestion.
Studies of time to acquisition of a reliable signal with different
pulse oximeters have con
rmed manufacturer recommendations
for the optimal technique for probe application[15e17]. When an
oximeter is turned on with the probe already connected, it will
immediately begin taking measurements. If the probeis notapplied
to the baby, the oximeter measures environmental signals such as
ambient light or infrared light from a radiant warmer. This can
cause a delay in obtaining accurate measurements once the probe is
placed on the patient. To avoid this, the pulse oximeter should be
turned on rst with the oximeter cable connected. Next, the probe
is applied to the baby. Once the probe is well-positioned on the
baby, it is then connected to the oximeter cable (Fig. 1).
Is the location of probe application clinically important? The
majority of studies using pulse oximetry in the delivery room have
placed the probe on the right wrist as this preductal location re-
ects the saturation of blood that is perfusing the brain. Further-
more, nomograms of oxygen saturation at birth were created using
preductal oxygen saturations. If we are to use such nomograms to
guide the titration of oxygen, as recommended by ILCOR, it is
important that we also use a preductal probe location. There is a
large body of evidence documenting that pre- and postductal ox-
ygen saturations are signicantly different for several minutes after
birth due to right-to-left shunting of blood.
Mariani et al. [12] reported that postductal SpO2 was 5e8%
lower than preductal SpO2 but that this difference decreased after5 min. However, others have shown that this difference persists for
up to 20 min and ranges from w14% at 3 min to 5% at 17 min[18e
20]. Furthermore, pulse wave signals aredetectable sooner with the
probe applied to the hand versus the foot: 1.24 min from the hand
versus 3.06 min from the foot [18]. For these reasons, we recom-
mend placing the probe on the right wrist.
4. Interpreting the quality of signal
Delivery room resuscitation provides one of the most technically
challenging environments for obtaining an accurate and stable
signal from the pulse oximeter. Each model has its own algorithm
for identifying and displaying signal quality. The user should befamiliar with the indicators of signal quality fortheir model of pulse
oximeter to determine their condence in the displayed values.
For example, the user can be condent of a good signal when
using the Nellcor (OxiMan N600x; Tyco Healthcare, Pleasanton, CA,
USA) pulse oximeter if they observe a regular plethysmograph with
pulse search and interference indicators not lit continuously.
Similarly, a good signal on the Radical pulse oximeter (Masimo,
Figure 1. Nellcor OxiMax N-600x pulse oximeter (A) with a patient cable (B) to which
the sensor (C) is connected.
Y. Rabi, J.A. Dawson / Seminars in Fetal & Neonatal Medicine 18 (2013) 330 e335 331
-
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
3/6
Irvine, CA, USA) is indicated by a regular plethysmograph, a high
signal IQ which represents the condence of the pulse oximeter in
the oxygen saturation and heart rate displayed with each arterial
pulse, and a good perfusion index reecting the relationship of
pulsatile to non-pulsatile blood ow at the sensor site. Other
manufacturers of pulse oximeters have different methods for
identifying a good signal.
It is common practice to manually calculate the heart rate and
compare it to the heart rate displayed on the pulse oximeter. If the
heart rate values are similar, the clinician is usually condent in the
quality of the signal. This seems to be a reasonable practice but the
reader should be aware that this approach has not been validated.
5. Challenges in using pulse oximetry in the delivery room
Pulse oximeters have been tested for validity and reliability by a
number of researchers who compared SpO2measurements against
arterial blood under a variety of conditions [21,22]. These studies
were generally conducted with rst or second generation oximeters.
In a study of preterm newborns, SpO2measurements were accurate
at arterial oxygen saturations of >93% [23]. However, at arterial
saturations of 80% fewer than half of SpO2 measurements were
within 3% of the arterial oxygen saturation. These lower oxygensaturation values are normal in therst few minutes after birth.
Other major challenges to obtaining accurate pulse oximetry
measurements in the delivery room are motion, low perfusion and
ambient light. Pulse oximeters identify arterial blood by detecting
pulsations. Motion artefact can occur with patient movement
resulting in irregular venous blood ow that can be interpreted as
pulsatile ow. This can confuse the pulse oximeter and lead to
inaccurate measurements. Different manufacturers have developed
various strategies for dealing with this problem and have met with
some success. In a review article examining the performance of
new-generation motion-tolerant pulse oximeters, Giulinao and
Higgins[24]concluded that newer pulse oximeters have superior
performance.
Newborns often have poorly perfused extremities immediatelyafter birth. As the pulse oximeter is dependent upon identifying a
pulse to perform its calculations, this can present a challenge. If the
option is available, we recommend selecting a high sensitivity
setting for delivery room applications.
Fluorescent lights, bright lights common to surgical suites and
even light produced by radiant warmers can adversely affect
measurements [25e27]. If ambient light is excessive, the pulse
oximeter probe should be shielded with an opaque wrap [28,29].
At 26 weeks, the fetus has nearly 100% fetal hemoglobin [6].
Even at term, fetal hemoglobin represents about 70% of total he-
moglobin. At present, pulseoximeters do not measure or correct for
fetal hemoglobin which differs slightly in its light absorption
characteristics from adult hemoglobin[30,31]. Some investigators
have demonstrated that by not correcting for fetal hemoglobinconcentration, pulse oximetry overestimates the true oxygen
saturation of arterial blood by 3e6% [32e34]. There is another,
equally compelling, body of evidence from different investigators
showing that fetal hemoglobin has a trivial effect on the accuracy of
pulse oximetry SpO2 measurements [22,35e37]. Whereas studies
appear to agree that fetal hemoglobin does affect the accuracy of
pulse oximetry SpO2 measurements, there is disagreement over
whether the effect is clinically signicant.
6. Does pulse oximetry provide useful information in the
delivery room?
For pulse oximetry to be useful in the delivery room, it must
provide reliable information soon after delivery. Pulse oximeter
performance for SpO2and HR measurements in the delivery room
has been evaluated in several studies.
6.1. Oxygen saturation
Before the introduction of pulse oximeters intothe delivery room,
clinicians relied upon assessment of colour to determine the babys
oxygenation status. We now understand that clinical assessment of
colour is an inaccurate and imprecise sign of oxygenation[38].
Many studies have shown that pulse oximetry can be used to
measure oxygen saturations in the delivery room for both term and
preterm babies. The majority of these studies focused on doc-
umenting normal oxygen saturations in stable babies over the rst
minutes after birth[10,11,13,19,20,39].
How soon after birth can pulse oximetry provide reliable SpO2measurements? In experienced hands, a probe can be applied to a
baby within 15e20 s of birth[11,15,17]. Kamlin et al.[11]reported a
53% success rate in obtaining stable SpO2measurements by 1 min.
In another study, the median time to achieve stable SpO2 mea-
surements was 82 s in healthy newborns 35 weeks of gestation
[13]. In a recent study of preterm babies
-
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
4/6
the delivery room. This allows clinicians to focus on other aspects of
resuscitation without having to stop and auscultate or palpate the
heart rate. Though evidenceof any inuence on importantoutcomes
is lacking, many experts recommend continuous oxygen saturation
monitoring [5,43,51,52] and targeting normal oxygen saturation
values[53].
8. Can we target oxygen saturation ranges in the delivery
room?
Neonatal intensive care units generally have specic SpO2target
ranges for babies receiving supplemental oxygen. However, re-
searchers have shown that it is difcult for staff to comply with
such target ranges[54e56]. In the delivery room it is likely to be
even more difcult to keep SpO2 measurement within a target
range when SpO2rises rapidly in the rst minutes.
We now understand that in the rst minutes after birth it is
normal for babies to have a low SpO2. There is the potential for
oxygen to be administered needlessly if clinicians are not aware
that babies are normally blue in the rst minutes after birth and it
may take several minutes for SpO2to reach the normalpostnatal
range [53]. When oxygen is administered, its use should betitrated to achieve normal time-based oxygen saturation values.
ILCOR recommends targeting the middle quartile of oxygen
saturation values observed in healthy babies transitioned in room
air [5]. Although there is no evidence that this approach affects
outcomes, given the current state of knowledge, it seems
reasonable.
Prospective studies of preterm resuscitation in the delivery
room that targeted an oxygen saturation value or range met with
modest success in achieving these targets. Escrig et al. [57] tar-
geted an SpO2 of 85% in conjunction with heart rate >100 bpm to
guide oxygen titration. They were able to use oxygen saturation to
guide the titration of oxygen though they did not report on the
amount of time on the SpO2target, perhaps because it was a single
value. The ROAR study (Room air vs Oxygen Administration forResuscitation of preterm infants) targeted a static SpO2 target
range of 85e92%. The primary outcome was the proportion of the
total resuscitation time the babys SpO2 remained in the target
range. They compared three groups. The high oxygen group
received a static concentration of 100% oxygen while the moderate
and low oxygen groups had the oxygen concentration titrated
starting from either 100% or 21%, respectively. The moderate ox-
ygen group spent 21% of the total resuscitation time in the target
range compared to 11% in the high oxygen group and 16% in the
low oxygen group[58].
Goos et al. [59]measured the extent to which observed SpO 2levels matched the European Resuscitation Council (ERC) oxygen
saturation targets during resuscitation of babies 30 weeks of
gestation. During the initial 10 min after birth, SpO2 values wereabove the ERC target for 44% (IQR: 12e66) of the time and below
the ERC target for 51% (IQR: 27e82) of the time. Gandhi et al.[60]
have described a system intended to facilitate maintaining SpO2values between the 10th and 50th percentiles of the normal target
range. Their Transitional Oxygen Targeting System (TOTS) plots
real-time SpO2 values in relation to the 10th and 50th percentile
SpO2 curves from a nomogram. This provides clinicians with a
visual representation of a babys SpO2 pattern in relation to the
SpO2 target range. Babies managed with this system spent a
greater proportion of total resuscitation time (52%) within the
target range compared to a control group (37%) [60]. Further
studies are needed to determine whether targeting normal oxygen
saturations during resuscitation affects clinically meaningful long-
term outcomes.
9. Does pulse oximetry use in the delivery room affect
outcomes?
The question of whether pulse oximetry affects outcomes has
long been debated in the adult, pediatric and neonatal intensive
care settings. Concerns over frequent alarms and low condence in
measurements, especially with earlier models, led some to
disparagingly refer to pulse oximeters as random number gener-
ators. Fortunately, the performance of newer generation pulse
oximeters has successfully addressed these issues to a large extent.
Research focus has shifted from validating the technology to
determining how the technology can be put to best use to improve
outcomes.
A few recent studies quantied the amount of oxygen given to
babies during resuscitation where continuous pulse oximetry was
used. Preterm babies initially resuscitated with 30% oxygen were
estimated to have received 465.6 ml/kg of pure oxygen in com-
parison to babies initially resuscitated with 90% oxygen who
received 864 ml/kg of pure oxygen [57]. Another study reported
that preterm babies resuscitated with an oxygen titration protocol
that started with room air received w37% less pure oxygen by
volume than babies resuscitated with a staticconcentration of 100%
oxygen[58]. Using pulse oximetry to guide oxygen titration mayhelp reduce the oxidative stress we impose on sick newborns at
birth, though this still awaits conrmation.
Pulse oximetry is simply a measurement tool, hence its use will
not directly inuence outcomes. Rather, its potential lies in
providing information about the safety and effectiveness of in-
terventions and guiding decision-making. As an example, three
randomized controlled studies all using pulse oximetry showed
that oropharyngeal suctioning has a negative effect on oxygenation
[61e63]. ODonnell et al. [64] measured the effects of attempted
endotracheal intubation on SpO2in the delivery room and showed
that SpO2 frequently fell during intubation attempts. During in-
terventions such as oropharyngeal suction or endotracheal intu-
bation, pulse oximetry is able to show clinicians the effect of the
intervention on a babys SpO2in real time.Kopotic and Lindner[65]studied babies at risk for respiratory
failure. Half the babies were managed without pulse oximetry
and compared with the other half managed with pulse oximetry.
Babies managed with oximetry were less likely to be admitted to
the nursery (32% vs 52%). In another study by Deckardt et al.,
respiratory care was based on the baby s clinical state and SpO2measurements [66]. Oxygen was started at 100% and adjusted to
achieve an SpO2 between 80% and 92%. Using pulse oximetry
they were able to reduce the fraction of inspired oxygen (FiO 2)
from 1.0 to 0.40. The Deckardt et al.[66]and Kopotic and Lindner
[65] studies, although unmasked and non-randomized, suggest
that pulse oximetry can be effective in guiding oxygen admin-
istration to improve short-term outcomes, such as admission to
nursery and the use of oxygen or continuous positive airwaypressure. We are not aware of any studies assessing the effect of
using SpO2 measurements to guide treatment decisions imme-
diately after birth on long-term outcomes.
There may be lasting effects from short exposure to oxygen in
the delivery room. There is evidence to support concerns
regarding exposure to oxygen in the delivery room and an asso-
ciation with an increased risk for the development of childhood
cancers. Spector et al. reviewed data from a large prospective
cohort of>50 000 children enrolled in the National Collaborative
Perinatal Project (NCPP) and described a higher risk of cancer in
children exposed to more than 3 min of oxygen in the delivery
room than in children without oxygen exposure (hazard ratio:
2.87) [67]. The authors speculate that reactive oxygen species
could damage DNA and increase susceptibility to developing
Y. Rabi, J.A. Dawson / Seminars in Fetal & Neonatal Medicine 18 (2013) 330 e335 333
-
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
5/6
cancer. In a similar study, the odds ratio for developing childhood
cancer following exposure to oxygen at delivery was 2.57 and
increased to 3.54 if manual ventilation lasted more than 3 min
[68]. Both studies point to a potential dose response with higher
risks of cancer associated with longer exposure to oxygen [67,68].
Whereas there is a clear biologic relationship between oxygen
exposure and the formation of cancerous cells in vitro, it remains
unclear whether it was the oxygen exposure and/or the babiesunderlying conditions that contributed to the increased risk in
these studies.
10. Conclusions
Pulse oximetry is a useful adjunct to clinical assessment in the
delivery room but clinicians attending newly born infants should
have a good understanding of normal fetal-to-neonatal transition.
Different models of pulse oximeter have differences in calibration
and algorithms for detecting and managing artefact, and demon-
strating signal quality. Therefore, it is important for clinicians
managing babies in the delivery room to understand the charac-
teristics of the pulse oximeter that they use and to be aware of
indicators that measurements might be unreliable. Titrating oxygen
administration to SpO2measurements is reasonable at the currenttime. However, there are no randomized trials measuring the
impact of using different SpO2 targets in the delivery room on long-
term outcomes.
Funding sources
J.A. Dawson is a recipient of a National Health and Medical
Research Council (NHMRC) Post Doctoral Fellowship and is sup-
ported by the Victorian Governments Operational Infrastructure
Support Program.
Conict of interest statement
Y. Rabi has a patent in Japan and patent pending applications in
the USA and European Union, for technology to help guide oxygen
titration in the delivery room. He has a revenue sharing agreement
with Masimo Corp. for distribution of this technology.
References
[1] Aoyagi T, Miyasaka K. Pulse oximetry: its invention, contribution to medicine,and future tasks. Anesth Analg 2002;94(1 Suppl):S1e3.
[2] Sendak MJ, Harris AP, Donham RT. Use of pulse oximetry to assess arterialoxygen saturation during newborn resuscitation. Crit Care Med 1986;14:739e40.
[3] Harris AP, Sendak MJ, Donham RT. Changes in arterial oxygen saturationimmediately after birth in the human neonate. J Pediatr 1986;109:117e9.
[4] House JT, Schultetus RR, Gravenstein N. Continuous neonatal evaluation in thedelivery room by pulse oximetry. J Clin Monit 1987;3:96e100.
[5] Perlman JM, Wyllie J, Kattwinkel J, Atkins DL, Chameides L, Goldsmith JP, et al.Part 11: Neonatal resuscitation: 2010 International Consensus on Cardiopul-monary Resuscitation and Emergency Cardiovascular Care Science WithTreatment Recommendations. Circulation 2010 Oct 19;122(16 Suppl 2):S516e38.
[6] Bard H. The postnatal decline of hemoglobin F synthesis in normal full-terminfants. J Clin Invest 1975;55:395e8.
[7] Nellcor Puritan Bennett, Inc.. OxiMaxN-600signal processing. In: http://www.nellcor.com/_Catalog/PDF/Product/N600SignalProcessing.pdf; 2005.
[8] Leone TA, Finer NN. Neonatal resuscitation: beyond the basics. NeoReviews2005;6:e177e83.
[9] Ahmed SJ, Rich W, Finer NN. The effect of averaging time on oximetry valuesin the premature infant. Pediatrics 2010;125:e115e21.
[10] Altuncu E, Ozek E, Bilgen H, Topuzoglu A, Kavuncuoglu S. Percentiles of ox-ygen saturations in healthy term newborns in the rst minutes of life. Eur JPediatr 2008;167:687e8.
[11] Kamlin CO, ODonnell CP, Davis PG, Morley CJ. Oxygen saturation in healthyinfants immediately after birth. J Pediatr 2006;148:585e9.
[12] Mariani G, Dik PB, Ezquer A, Aguirre A, Esteban ML, Perez C, et al. Pre-ductaland post-ductal O2 saturation in healthy term neonates after birth. J Pediatr2007;150:418e21.
[13] Rabi Y, Yee W, Chen SY, Singhal N. Oxygen saturation trends immediatelyafter birth. J Pediatr 2006;148:590e4.
[14] Bucher HU, Keel M, Wolf M, von Siebenthal K, Duc G. Artifactual pulse-oximetry estimation in neonates. Lancet 1994;343(8906):1135e6.
[15] ODonnell CPF, Kamlin COF, Davis PG, Morley CJ. Feasibility of and delay in
obtaining pulse oximetry during neonatal resuscitation. J Pediatr 2005;147:698e9.
[16] ODonnell CPF, Kamlin COF, Davis PG, Morley CJ. Obtaining pulse oximetrydata in neonates: a randomised crossover study of sensor application tech-niques. Archs Dis Childh Fetal Neonat Ed 2005;90:F84e5.
[17] Saraswat A, Simionato L, Dawson J, Thio M, Kamlin C, Owen L, et al. Deter-mining the best method of nellcor pulse oximeter sensor application in ne-onates. Acta Paediatr 2012;101:484e7.
[18] Meier-Stauss P, Bucher HU, Hrlimann R, Knig V, Huch R. Pulse oximetryused for documenting oxygen saturation and right-to-left shunting immedi-ately after birth. Eur J Pediatr 1990;149:851e5.
[19] Toth B, Becker A, Seelbach-Gobel B. Oxygen saturation in healthy newborninfants immediately after birth measured by pulse oximetry. Arch GynecolObstet 2002;266:105e7.
[20] Zubarioglu U, Uslu S, Can E, Bulbul A, Nuhoglu A. Oxygen saturation levelsduring the rst minutes of life in healthy term neonates. Tohoku J Exp Med2011;224:273e9.
[21] Barrington KJ, Finer NN, Ryan CA. Evaluation of pulse oximetry as a contin-
uous monitoring technique in the neonatal intensive care unit. Crit Care Med1988;16:147e53.
[22] Ramanathan R, Durand M, Larrazabal C. Pulse oximetry in very low birthweight infants with acute and chronic lung disease. Pediatrics 1987;79:612e7.
[23] Rosychuk RJ, Hudson-Mason A, Eklund D, Lacaze-Masmonteil T. Discrepanciesbetween arterial oxygen saturation and functional oxygen saturationmeasured with pulse oximetry in very preterm infants. Neonatology2012;101:14e9.
[24] Giuliano KK, Higgins TL. New-generation pulse oximetry in the care of criti-cally ill patients. Am J Crit Care 2005;14:26e37.
[25] Amar D, Neidzwski J, Wald A, Finck AD. Fluorescent light interferes with pulseoximetry. J Clin Monit 1989;5:135e6.
[26] Block Jr FE. Interference in a pulse oximeter from a beroptic light source.J Clin Monit 1987;3:210e1.
[27] Costarino AT, Davis DA, Keon TP. Falsely normal saturation reading with thepulse oximeter. Anesthesiology 1987;67:830e1.
[28] Fouzas S, Priftis KN, Anthracopoulos MB. Pulse oximetry in pediatric practice.
Pediatrics 2011;128:740e
52.
Practice points
Pulse oximetry can accurately measure oxygen satura-tions and heart rate during delivery room resuscitation.
Pulse oximeters have user-modifiable settings thataffect the ability to obtain stable measurements in thedelivery room.
Acquiring stable readings of oxygen saturation andheart rate within 2 min of birth is feasible in clinicalpractice.
Pulse oximetry provides real-time information impor-tant for decision-making (e.g. oxygen titration) duringresuscitation.
Research directions
Creation of new oxygen saturation nomograms forbabies who have had delayed cord clampingperformed.
Determine whether targeting normal oxygen satura-tions during resuscitation affects clinically importantoutcomes.
Determine the optimal approach to titrating oxygenduring resuscitation (i.e. starting FiO2, frequency of FiO2adjustments, magnitude of FiO2 adjustments).
Y. Rabi, J.A. Dawson / Seminars in Fetal & Neonatal Medicine 18 (2013) 330e335334
http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://www.nellcor.com/_Catalog/PDF/Product/N600SignalProcessing.pdfhttp://www.nellcor.com/_Catalog/PDF/Product/N600SignalProcessing.pdfhttp://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref28http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref27http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref26http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref25http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref24http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref23http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref22http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref21http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref20http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref19http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref18http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref17http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref16http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref15http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref14http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref13http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref12http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref11http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref10http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref9http://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://refhub.elsevier.com/S1744-165X(13)00070-X/sref8http://www.nellcor.com/_Catalog/PDF/Product/N600SignalProcessing.pdfhttp://www.nellcor.com/_Catalog/PDF/Product/N600SignalProcessing.pdfhttp://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://refhub.elsevier.com/S1744-165X(13)00070-X/sref6http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref5http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref4http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref3http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref2http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1http://refhub.elsevier.com/S1744-165X(13)00070-X/sref1 -
7/25/2019 R-3Oxygen Therapy and Oximetry in the Delivery Room
6/6
[29] Ralston AC, Webb RK, Runciman WB. Potential errors in pulse oximetry. III:Effects of interferences, dyes, dyshaemoglobins and other pigments. Anaes-thesia 1991;46:291e5.
[30] Fogh-Anderson N, Siggaard-Anderson O, Lundsgaard FC, Wimberley PD.Spectrophotometric determination of hemoglobin pigments in neonatalblood. Clin Chim Acta 1987;166:291e6.
[31] Pologe JA, Raley DM. Effects of fetal hemoglobin on pulse oximetry. J Perinatol1987;7:324e6.
[32] Jennis MS, Peabody JL. Pulse oximetry: an alternative method for theassessment of oxygenation in newborn infants. Pediatrics 1987;79:524e8.
[33] Shiao SY, Ou CN. Validation of oxygen saturation monitoring in neonates. Am JCrit Care 2007;16:168e78.
[34] Whyte RK, Jangaard KA, Dooley KC. From oxygen content to pulse oximetry:completing the picture in the newborn. Acta Anaesthesiol Scand Suppl1995;107:95e100.
[35] Anderson JV. The accuracy of pulse oximetry in neonates: effects of fetalhemoglobin and bilirubin. J Perinatol 1987;7:323.
[36] Durand M, Ramanathan R. Pulse oximetry for continuous oxygen monitoringin sick newborn infants. J Pediatr 1986;109:1052e6.
[37] Rajadurai VS, Walker AM, Yu VY, Oates A. Effect of fetal haemoglobin on theaccuracy of pulse oximetry in preterm infants. J Paediatr Child Health1992;28:43e6.
[38] ODonnell CP, Kamlin CO, Davis PG, Carlin JB, Morley CJ. Clinical assessment ofinfant colour at delivery. Arch Dis Child Fetal Neonatal Ed 2007;92:F465e7.
[39] Hulsoore R, Shrivastav J, Dwivedi R. Normal oxygen saturation trend inhealthy term newborns within 30 minutes of birth. Indian J Pediatr 2011;78:817e20.
[40] Ghandi B, Rich W, Finer N. Time to achieve stable pulse oximetry values inVLBW infants in the delivery room. Resuscitation 2013;84:970e3.
[41] Kratky E, Pichler G, Rehak T, Avian A, Pocivalnik M, Mller W, et al. Regionalcerebral oxygen saturation in newborn infants in the rst 15 min of life aftervaginal delivery. Physiological Measurement 2012:33.
[42] Dawson JA, Morley CJ. Monitoring oxygen saturation and heart rate in theearly neonatal period. Semin Fetal Neonatal Med 2010;15:203e7.
[43] Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, et al. EuropeanResuscitation Council Guidelines for Resuscitation 2010. Section 1. Executivesummary. Resuscitation 2010;81:1219e76.
[44] Perlman JM, Wyllie J, Kattwinkel J, Atkins DL, Chameides L, Goldsmith JP, et al.Neonatal Resuscitation: 2010 International Consensus on CardiopulmonaryResuscitation and Emergency Cardiovascular Care Science With TreatmentRecommendations. Pediatrics 2010;126:e1319e44.
[45] Kamlin CO, Dawson JA, ODonnell CP, Morley CJ, Donath SM, Sekhon J, et al.Accuracy of pulse oximetry measurement of heart rate of newborn infants inthe delivery room. J Pediatr 2008;152:756e60.
[46] Dawson JA, Kamlin CO, Wong C, Te Pas AB, Vento M, Cole TJ, et al. Changes inheart rate in the rst minutes after birth. Arch Dis Child Fetal Neonatal Ed2010 May;95:F177e81.
[47] Katheria A, Rich W, Finer N. Obtaining a continuous heart rate during neonatalresuscitation: EKG vs pulse oximetry. Baltimore: Pediatric Academic Societies;2012 [E-Pas2012:2855.4].
[48] Mizumoto H, Tomotaki S, Shibata H, et al. Electrocardiogram shows reliableheart rates much earlier than pulse oximetry during neonatal resuscitation.Pediatr Int 2012;54:205e7.
[49] Urlesberger B, Kratky E, Rehak T, Pocivalnik M, Avian A, Czihak J, et al.Regional oxygen saturation of the brain during birth transition of term
infants: comparison between elective cesarean and vaginal deliveries. JPediatr 2011;159:404e8.
[50] Wang CL, Anderson C, Leone TA, Rich W, Govindaswami B, Finer NN. Resus-citation of preterm neonates by using room air or 100% oxygen. Pediatrics2008;121:1083e9.
[51] Australian Resuscitation Council. Neonatal guidelines 13.1e13.10. AustralianResuscitation Council; 2010.http://www.resus.org.au/.
[52] Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, et al.Neonatal resuscitation: 2010 American Heart Association guidelines for car-diopulmonary resuscitation and emergency cardiovascular care. Pediatrics
2010;126:e1400e
13.[53] Dawson JA, Kamlin CO, Vento M, Wong C, Cole TJ, Donath SM, et al. Dening
the reference range for oxygen saturation for infants after birth. Pediatrics2010;125:e1340e7.
[54] Clucas L, Doyle LW, Dawson J, Donath S, Davis PG. Compliance with alarmlimits for pulse oximetry in very preterm infants. Pediatrics 2007;119:1056e60.
[55] Hagadorn JI, Furey AM, Nghiem TH, Schmid CH, Phelps DL, Pillers DA, et al.Achieved versus intended pulse oximeter saturation in infants born less than28 weeks gestation: the AVIOx study. Pediatrics 2006;118:1574e82.
[56] Sink DW, Hope SA, Hagadorn JI. Nurse:patient ratio and achievement of ox-ygen saturation goals in premature infants. Arch Dis Child Fetal Neonatal Ed2011;96:F93e8.
[57] Escrig R, Luis A, Izquierdo I, Villar G, Gimeno A, Senz P, et al. Achievementof target oxygen saturation in extremely low gestational neonates resus-citated with different oxygen concentrations: a prospective randomizedclinical trial. Pedatrics 2008;121:875e81.
[58] Rabi Y, Singhal N, Nettel-Aguirre A. Room-air versus oxygen administrationduring resuscitation of preterm infants: the ROAR Study. Pedatrics 2011;128:E374
e81.
[59] Goos TG, Rook D, van der Eijk AC, Kroon AA, Pichler G, Urlesberger B, et al.Observing the resuscitation of very preterm infants: are we able to follow theoxygen saturation targets? Resuscitation 2013;84:1108e13.
[60] Gandhi B, Rich W, Finer N. Achieving targeted pulse oximetry values in pre-term infants in the delivery room. J Pediatr 2013;163:412e5.
[61] Carrasco M, Martell M, Estol PC. Oronasopharyngeal suction at birth: effectson arterial oxygen saturation. J Pediatr 1997;130:832e4.
[62] Gungor S, Teksoz E, Ceyhan T, Kurt E, Goktolga U, Baser I. Oronasopharyngealsuction versus no suction in normal, term and vaginally born infants: a pro-spective randomised controlled trial. Aust NZ J Obstet Gynaecol 2005;45:453e6.
[63] Waltman PA, Brewer JM, Rogers BP, May WL. Building evidence for practice: apilot study of newborn bulb suctioning at birth. J Midwifery Womens Health2004;49:32e8.
[64] ODonnell CPF, Kamlin CO, Davis PG, Morley CJ. Endotracheal intubation at-tempts during neonatal resuscitation: success rates, duration, and adverseeffects. Pediatrics 2006;117:e16e21.
[65] Kopotic RJ, Lindner W. Assessing high-risk infants in the delivery room withpulse oximetry. Anaesth Analg 2002;94:S31e6.[66] Deckardt R, Schneider KT, Graeff H. Monitoring arterial oxygen saturation in
the neonate. J Perinat Med 1987;15:357e60.[67] Spector LG, Klebanoff MA, Feusner JH, Georgieff MK, Ross JA. Childhood cancer
following neonatal oxygen supplementation. J Pediatr 2005;147:27e31.[68] Naumburg E,BelloccoR, Cnattingius S,JonzonA,EkbomA. Supplementaryoxygen
and risk of childhood lymphatic leukaemia. Acta Paediatr 2002;91:1328e33.
Y. Rabi, J.A. Dawson / Seminars in Fetal & Neonatal Medicine 18 (2013) 330 e335 335
http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref35http://refhub.elsevier.com/S1744-165X(13)00070-X/sref35http://refhub.elsevier.com/S1744-165X(13)00070-X/sref35http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://www.resus.org.au/http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref68http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref67http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref66http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref65http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref64http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref63http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref62http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref61http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref60http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref59http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref58http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref57http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref56http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref55http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref54http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref53http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://refhub.elsevier.com/S1744-165X(13)00070-X/sref52http://www.resus.org.au/http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref50http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref49http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref48http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref47http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref46http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref45http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref44http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref43http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref42http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref41http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref40http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref39http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref38http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref37http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref36http://refhub.elsevier.com/S1744-165X(13)00070-X/sref35http://refhub.elsevier.com/S1744-165X(13)00070-X/sref35http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref34http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref33http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref32http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref31http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref30http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29http://refhub.elsevier.com/S1744-165X(13)00070-X/sref29