Quantitative pharmacokinetic analysis of DCE-MRI data without...
Transcript of Quantitative pharmacokinetic analysis of DCE-MRI data without...
Magnetic Resonance Im
Original contributions
Quantitative pharmacokinetic analysis of DCE-MRI data without
an arterial input function: a reference region model
Thomas E. Yankeelova,b,T, Jeffrey J. Lucia,b, Martin Lepagea,b, Rui Lib,c,
Laura Debuskd, P. Charles Lind,e, Ronald R. Pricea,b, John C. Gorea,b
aInstitute of Imaging Science, Vanderbilt University, Nashville, TN 37232-2675, USAbDepartment of Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN 37232-2675, USA
cDepartment of Electrical Engineering and Computer Science, Vanderbilt University, Nashville, TN 37232-2675, USAdDepartment of Cancer Biology, Vanderbilt University, Nashville, TN 37232-2675, USA
eDepartment of Radiation Oncology, Vanderbilt University, Nashville, TN 37232-2675, USA
Received 17 February 2005; accepted 17 February 2005
Abstract
Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) can assess tumor perfusion, microvascular vessel wall permeability
and extravascular–extracellular volume fraction. Analysis of DCE-MRI data is usually based on indicator dilution theory that requires
knowledge of the concentration of the contrast agent in the blood plasma, the arterial input function (AIF). A method is presented that
compares the tissues of interest (TOI) curve shape to that of a reference region (RR), thereby eliminating the need for direct AIF measurement.
By assigning literature values for Ktrans (the blood perfusion-vessel permeability product) and ve (extravascular–extracellular volume fraction)
in a reference tissue, it is possible to extract the Ktrans and ve values for a TOI without knowledge of the AIF. The operational RR equation for
DCE-MRI analysis is derived, and its sensitivity to noise and incorrect assignment of the RR parameters is tested via simulations. The method
is robust at noise levels of 10%, returning accurate (F20% in the worst case) and precise (F15% in the worst case) values. Errors in the TOI
Ktrans and ve values scale approximately linearly with the errors in the assigned RR K trans and ve values. The methodology is then applied to a
Lewis Lung Carcinoma mouse tumor model. A slowly enhancing TOI yielded K trans=0.039F0.002 min�1 and ve=0.46F0.01, while a rapidly
enhancing region yielded Ktrans=0.35F0.05 min�1 and ve=0.31F0.01. Parametric K trans and ve mappings manifested a tumor periphery with
elevated Ktrans (N0.30 min�1) and ve (N0.30) values. The main advantage of the RR approach is that it allows for quantitative assessment of
tissue properties without having to obtain high temporal resolution images to characterize an AIF. This allows for acquiring images with higher
spatial resolution and/or SNR, and therefore, increased ability to probe tissue heterogeneity.
D 2005 Elsevier Inc. All rights reserved.
Keywords: DCE-MRI; Arterial input function; Pharmacokinetics; Reference region model
1. Introduction
Dynamic contrast-enhanced magnetic resonance imaging
(DCE-MRI) involves the serial acquisition ofMR images of a
tissue of interest (TOI) (e.g., a tumor locus) before, during
and after an intravenous injection of contrast agent (CA). As
the CA perfuses into the tissue under investigation, the T1 and
T2 values of tissue water decrease to an extent that is
determined by the concentration of the agent. By considering
0730-725X/$ – see front matter D 2005 Elsevier Inc. All rights reserved.
doi:10.1016/j.mri.2005.02.013
* Corresponding author. Institute of Imaging Science, Vanderbilt
University, Nashville, Tennessee 37232-2675, USA.
E-mail address: [email protected] (T.E. Yankeelov).
a set of images acquired before, during and after a CA
infusion, a region of interest (ROI) or individual voxel will
display a characteristic signal intensity time course that can
be related to CA concentration. This time course can be
analyzed with an appropriate mathematical pharmacokinetic
model. By fitting the DCE-MRI data to such model,
physiological parameters can be extracted that relate to, for
example, tissue perfusion, microvascular vessel wall perme-
ability and extracellular volume fraction [1]. It has been
shown that both healthy and pathologic tissues exhibit
characteristic signal intensity time courses as well as
pharmacokinetic parameter values (see, e.g., Ref. [2–4]).
Furthermore, since these parameter values are probes of
aging 23 (2005) 519–529
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529520
tissue status, theymay be used to differentiate malignant from
benign tumors [5], aid in tumor staging [6,7] and monitor
treatment response [6,8–10]. Thus, there is considerable and
continuing interest in developing new and improved methods
to obtain these parameter values accurately and precisely.
Analysis of DCE-MRI data is usually based on indicator
dilution theory [11] and requires knowledge of the concen-
tration of the CA in the blood plasma, Cp, the so-called
arterial input function (AIF). This is a notoriously difficult
problem and three main approaches have been developed to
estimate the AIF in DCE-MRI studies. One approach
involves introducing an arterial catheter into the subject and
sampling blood during the imaging process for later analysis
[12,13]. An advantage of this approach is that the Cp in each
sample can be determined accurately through standard
spectroscopic methods (e.g., inductive coupled plasma
emission spectroscopy), thereby allowing for characteriza-
tion of Cp as a function of time. However, the disadvantages
include its invasive nature, poor temporal resolution and
relative ambiguity concerning the actual time at which the
sample was drawn. In laboratory mice, which are used in
many DCE-MRI experiments, the total blood volume is very
small (~2 ml) so that very few samples (2–5; assuming a
volume of 50–100 Al per sample, the minimum amount
needed for standard spectroscopic measurements) can be
taken in total, and fewer still can be used to characterize the
uptake portion of the Cp curve.
A second method assumes that the AIF is similar for all
subjects. The AIF is first measured via blood samples in a
small cohort of subjects [14], and the resulting average AIF is
then assumed to be valid for subsequent studies [15,16]. A
major advantage of this approach is its simplicity in both data
acquisition and data analysis; no AIF measurement is
required for the experimental subjects, and the subsequent
curve fitting uses a common AIF on all data sets. The
disadvantages include the influences of both inter- and intra-
subject variations in AIF, which can introduce large errors in
both AIF characterization and subsequent pharmacokinetic
analysis [16]. Also, by measuring the AIF in one cohort of
subjects and applying it to another, changes in the AIF that
may be introduced by the pathology under investigation are
ignored, reducing the validity of the assumption in important
practical situations.
A third method obtains the AIF from the DCE-MRI data
sets themselves [11,17,18]. Methods have been developed
that simultaneously measure signal intensity changes (due to
CA passage) in both the blood and tissue. A calibration is then
employed to convert the blood signal intensity to the
intravascular concentration of CA. Such a method has the
potential advantage of measuring the AIF accurately on an
individual basis, and since it does not require any further
measurements, being completely noninvasive. However, it
requires the presence of a large vessel within the field of view
(FOV) [19]. Additionally, the images must be acquired such
that the lumen signal is devoid of partial volume or flow
effects. Specialized pulse sequences can be employed that
selectively saturate spins to avoid inflow effects, and this
allows acquisition of a set of slices containing a feeding
vessel without inflow effects. A recent elegant example of
such an approach was presented by McIntyre et al. [20].
However, this method still requires high temporal resolution
and restricts the choices of both which regions can be
characterized and the imaging slice orientation. In general,
accurate AIF measurements require (significantly) higher
temporal resolution (less than 10 s) than tissue measurements
(30–60 s), so the temporal resolution is dictated by the AIF
measurements process, and the spatial resolution and signal-
to-noise ratio are compromised. The former is potentially an
especially important drawback as a major use of DCE-MRI is
to assess heterogeneous tissues (such as tumors) that demand
high spatial resolution.
We present here a general method derived from the
positron emission tomography (PET) literature [21], which
allows for quantitative pharmacokinetic analysis of
DCE-MRI data without knowledge of the AIF. The PET
community refers to this formalism as the reference region
(RR) model because it relies on finding a well-characterized
RR from which to bcalibrateQ the signal intensity changes in aTOI. We have amended this model to allow for analysis of
T1-weighted DCE-MRI data. It should be noted that a similar
method has previously been proposed by Kovar et al. [22].
The two main differences between that approach and this
contribution are that Kovar et al employed the differential
form of the Kety equation [11] to estimate the AIF from an
RR, whereas our theory incorporates the integral form of the
Kety equation that allows for the development of an opera-
tional equation that is independent of the AIF and therefore
eliminates the requirement for AIF estimation entirely. This
allows for the application of a simple (one-step) curve-fitting
algorithm to obtain estimates on the pharmacokinetic
parameters. Additionally, employing the integral versions
of theKety theory offers the possibility for refined RRmodels
as presented in the Discussion.
In this report, we present the mathematical framework of
a model for DCE-MRI analysis that does not require
knowledge of an AIF. We then test the model’s accuracy,
precision and sensitivity to incorrect RR assumptions
through simulations and show the practical effectiveness
of the model. The potential ease of implementation of this
method promises to provide quantitative DCE-MRI analy-
ses in both experimental and clinical settings. Future efforts
will seek to validate this method.
2. Theory
In an effort to encourage the use of standardized nota-
tion, we use only those symbolic conventions described by
Tofts et al. [1].
Fig. 1 displays a simple two-compartment model in
which CA diffuses from the blood plasma into the
extravascular–extracellular spaces of the RR and the TOI.
Fig. 1. A cartoon depiction of the RR model. Contrast agent from the blood
plasma diffuses bidirectionally from the intravascular space to the
extravascular–extracellular space of the reference tissue and to the
extravascular–extracellular space of the tissue-of-interest. The equations
describing this simple two-compartment model are given as Eqs. (1) and (2).
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529 521
The linear, ordinary differential equations that describe this
system are given as Eqs. (1) and (2):
d
dtCRR tð Þ ¼ K trans;RR
bCp tð Þ � K trans;RR=ve;RR�bCRR tð Þ
�ð1Þ
d
dtCTOI tð Þ ¼ K trans;TOI
bCp tð Þ�ðK trans;TOI=ve;TOIÞbCTOI tð Þ ð2Þ
where CRR and CTOI are the tissue concentrations (expressed
in millimolar) of CA in the RR and the TOI, respectively;
K trans,RR and K trans,TOI are the CA extravasation rate
constants for the RR and TOI, respectively; and ve,RR and
ve,TOI are the extravascular–extracellular volume fractions
for the RR and TOI, respectively [1]. Eqs. (1) and (2) make
clear how the concentration of CA in the blood [Cp(t), i.e.,
the AIF] is incorporated into the analysis: if Cp and CTOI [as
in Eq. (2)] can be measured, then a curve-fitting routine can
return estimates on Ktrans and ve. We seek to remove the
dependence of Eqs. (1) and (2) on the AIF, and to do so, we
must eliminate all Cp terms from the formalism.
Solving Eq. (1) for Cp(t) yields Eq. (3):
Cp tð Þ ¼ ð1=K trans;RRÞ d
dtCRR tð Þ þ 1=ve;RR
�bCRR tð Þ:
�ð3Þ
Eq. (3) can then be substituted into Eq. (2) to yield Eq. (4):
d
dtCTOI tð Þ þ ðK trans;TOI=ve;TOIÞbCTOI tð Þ
¼ Rbd
dtCRR tð Þ þ K trans;TOI=ve;RR
�bCRR tð Þ;
�ð4Þ
where RuKtrans,TOI/Ktrans,RR. Note that Eq. (4) is indepen-
dent of Cp(t ). Next define the integrating factor
Iuexp[(Ktrans,TOI/ve,TOI)t] and note that (via the product
rule of calculus)
d
dtðCTOI tð ÞbIÞ ¼ Ib
� d
dtCTOI tð Þ
þ ðK trans;TOI=ve;TOIÞbCTOI tð Þ�; ð5Þ
which is identical to the left-hand side of Eq. (4) multiplied
by I. Then Eq. (4) may be rewritten as
d
dtðCTOI tð ÞbIÞ ¼ Rb
d
dtCRR tð ÞbI
þ ðK trans;TOI=ve;RRÞbCRR tð ÞbI ; ð6Þ
which is an exact equation and therefore can then be inte-
grated directly (by parts) to obtain a relationship between
CTOI and CRR that is independent of Cp (see, e.g., Ref. [23]):
CTOI Tð Þ ¼ RbCRR Tð Þ þ Rb½ðK trans;RR=ve;RRÞ
� ðK trans;TOI=ve;TOIÞ�bZT
0
CRR tð Þ
� expðð � K trans;TOI=ve;TOIÞb T � tð ÞÞdt: ð7ÞIt is important to note that CTOI and CRR are the bulk
concentrations in the tissue water and are related to the real
concentration of CA by the extravascular–extracellular
water volume fraction; that is, CTOI and CRR are only
proportional to the true concentrations (those in the volumes
in which the CA is actually dissolved) — and then only if
the compartment volumes do not change during the
experiment [24]. The transformations from tissue concen-
tration, Ct, to concentration in the extravascular–extracellu-
lar space is given by Eq. (8):
Ct ¼ vebCe; ð8Þ
where Ce denotes concentration of CA in the extravascular–
extracellular space [1]. Thus, the CTOI and CRR expressions
of Eq. (7) can be converted to the true concentrations via the
following transformations:
CTOI ¼ ve;TOIbCe;TOI ð9Þ
CRR ¼ ve;RRbCe;RR; ð10Þ
where Ce,TOI and Ce,RR denote the concentrations of CA in
the extravascular–extracellular (water) space of the TOI and
RR, respectively [24,28]. Furthermore, since concentration
of CA is not measured directly in a DCE-MRI experiment,
a calibration to the measured longitudinal relaxation rate
constant, R1 (u1/T1), is required [1]. In the fast exchange
limit approximation [1,24,25], the following relations
are assumed:
R1;TOI ¼ r1;TOIbve;TOIbCe;TOI þ R10;TOI ð11Þ
R1;RR ¼ r1;RRbve;RRbCe;RR þ R10;RR; ð12Þ
where r1,TOI and r1,RR are CA longitudinal TOI and RR
relaxivities (in mM�1 s�1), respectively, and R10,TOI and
2D Graph 4
time (min)0 10 20 30 40
R1(t
) (s
-1)
1.0
2.0
3.0
R1bR1,TOIR1,RR
C (
mM
)
0.0
0.4
0.8
1.2
1.6
2.0
CpCe,TOICe,RR
A
B
Fig. 2. Panel A depicts simulated Cp, Ce,TOI and Ce,RR time courses. Cp was
converted to Ce,TOI and Ce,RR via Eq. (15) with Ktrans,RR and ve,RR set to
0.1 min�1 and 0.1, respectively, and K trans,TOI and ve,TOI set to 0.1 min�1
and 0.1. Panel B represents the linear transformation for concentration of
CA time courses to R1 time courses as described by Eqs. (10) and (11).
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529522
R10,RR are the TOI and the RR R1’s before contrast
administration, respectively. Though a simplifying assump-
tion, it is common to assume that r1 is constant throughout the
FOV, and we do the same here [1,24,28]. Combining Eq. (9)
with Eq. (11) and Eq. (10) with Eq. (12) yields Eqs. (13) and
(14), respectively:
R1;TOI ¼ r1;TOIbCTOI þ R10;TOI ð13Þ
R1;RR ¼ r1;RRbCRR þ R10;RR: ð14Þ
Solving Eqs. (13) and (14) for CTOI and CRR gives
expressions that are readily substituted into Eq. (7), yielding
an operational equation that can be employed in a curve-
fitting routine to extract Ktrans,TOI and ve,TOI if R1,TOI,
R10,TOI, R1,RR and R10,RR can be measured:
ðR1;TOI Tð Þ � R10;TOIÞ ¼ RbðR1;RR Tð Þ � R10;RRÞ
þ R½ðK trans;RR=ve;RRÞ� ðK trans;TOI=ve;TOIÞ�
�ZT
0
ðR1;RR tð Þ � R10;RRÞbexp
�ðð�K trans;TOI=ve;TOIÞ� T� tð ÞÞdt;
ð15Þ
or, alternatively, as
R1;TOI Tð Þ ¼ RbðR1;RR Tð Þ � R10;RRÞ þ Rb½ðK trans;RR=ve;RRÞ
� ðK trans;TOI=ve;TOIÞ�bZT
0
ðR1;RR tð Þ�R10;RRÞ
� expðð � Ktrans;TOI=ve;TOIÞb T � tð ÞÞdt
þ R10;TOI: ð16Þ
3. Methods
3.1. Simulations
To test the method, we simulated AIF, RR and TOI CA
curves. The AIF curve (Cp time course) was generated using
Eq. (17):
Cp tð Þ ¼ Abtbexp � tbBð Þ
þ C 1� exp � tbDð Þ½ �bexp � tbEð Þ; ð17Þ
where A= 0.6 mM, B = 0.18 min�1, C = 0.45 mM,
D=0.5 min�1, E=0.013 min�1. This AIF is reasonable
and similar in form to that of Simpson et al. [16]. The
resulting AIF was discretized with 1-min temporal resolu-
tion and truncated after the first 40 min to produce the AIF
shown in Fig. 2A. This Cp AIF was then converted to the
Fig. 2A RR and TOI CA curves via the integral form of the
Kety-Schmidt equation [1,11], which is less sensitive to
noise than the differential form (Eqs. (1) and (2)):
Ct Tð Þ ¼ K trans
ZT
0
Cp tð Þexp � K trans=veÞ T � tð Þð Þdt;ð ð18Þ
with Ktrans=0.10 min�1 and ve=0.10 assigned for the Ce,RR
curve, whileKtrans=0.25min�1 and ve=0.4 were assigned for
the Ce,TOI curve. The RR values are reasonable for muscle
tissue [25–27], and the TOI values are reasonable for
enhancement kinetics seen in a variety of tumor types (e.g.,
Refs. [9,28]). Ct curves are then converted to R1 curves via
the fast exchange limit approximation [24,25], Eqs. (13) and
(14). The CA relaxivity was set at 3.6 mM�1 s�1
[appropriate for gadolinium diethylenetriamine pentaacetic
acid (Gd-DTPA) at 7.0 T], and R10 values of 0.7 s�1 and
0.55 s�1 were used for the RR and TOI, respectively. These
transformations from Ct(t) to R1(t) yield the curves depicted
in Fig. 2B. The Fig. 2B blood R1 time course, R1b,
was computed from Cp via R1b(t)=r1(1�h)Cp(t)+R1b0,
where h is the hematocrit (set to 0.5) and R1b0 is the pre-
CA blood R1.
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529 523
We then input the RR and TOI R1(t) curves with the RR
model (Eq. (18)) into a curve-fitting routine written in IDL
(Research Systems, Boulder, CO). The Ktrans,RR and ve,RRvalues were fixed at their assigned values (0.1 min�1 and
0.1, respectively), while the Ktrans,TOI and ve,TOI values were
allowed to vary. This tests the accuracy of the RR model. To
test the precision of the model, the curve-fitting procedure
was then repeated with n% random Gaussian noise [n%
noiseu(maximum R1 value)(n/100)] added to the RR and
TOI R1 curves. First 5% and 2% noise was added to the TOI
and RR curves, respectively. Then the entire process was
repeated with 10% and 4% noise added to the TOI and RR
curves, respectively. Assigning a higher noise values to the
TOI over the RR is justified since the RR curve is the
average of many voxels (below, we use 20), whereas the
TOIs will ultimately be single voxels. Additionally, the RR
curve was smoothed with a boxcar average of length 5
before fitting. The process of adding random noise was
repeated a hundred times (runs) to yield arrays of each
parameter (one value per run) from which means and
standard deviations were computed.
One of the potential drawbacks of the RR method is that
it requires assigning the K trans,RR and ve,RR values.
Potentially, even for a well-characterized tissuelike muscle,
these values could vary significantly from subject to subject,
and therefore, knowledge of Ktrans,RR and ve,RR might be
limited. To address this issue, we have run a series of
simulations aimed at elucidating the errors inherent in
assuming an incorrect Ktrans,RR and/or ve,RR value. We
constructed data using the Ktrans,RR and ve,RR values listed
previously (0.1 min�1 and 0.1, respectively) and then ran
the fitting routines with Ktrans,RR incorrectly assigned at a
value of xKtrans,RR, where x was first assigned the value of
0.70. This procedure was repeated with x varied from
0.70 to 1.30 incremented in units of 0.05. An analogous
procedure was performed for ve,RR. In this way, errors of
F30% in both Ktrans,RR and ve,RR were investigated. All
permutations of these incorrect parameter assignments and
their subsequent errors in the returned parameters were
assessed. The values returned by the fitting routines were
stored as two-dimensional arrays and rendered as three-
dimensional error plots. These simulations were performed
without noise so that systematic errors (rather than random
errors) could be assessed.
3.2. Experimental
A male C57/BL mouse (22 g) received a hind limb
subcutaneous injection of 2�105 Lewis Lung Carcinoma
cells. The mouse was fed a standard diet in a controlled
environment with a 12/12-h light/dark cycle. Just prior to
imaging, anesthesia was induced via a 5%/95% isoflurane/
oxygen mixture; anesthesia was maintained via a 2%/98%
isoflurane/oxygen mixture. A 26 G catheter (Abbott Labo-
ratories, Switzerland) was inserted into the tail vein. The
temperature of the animal was maintained via a flow of warm
air through the magnet bore. Heart and respiratory rate were
monitored throughout the experiment. All procedures
adhered to Vanderbilt University’s IACUC guidelines.
The mouse was imaged in a Varian 7.0-T scanner
equipped with a 38-mm quadrature birdcage coil on 16,
20, 24 and 33 days postinjection (we present here the results
from the day 33 study). With TR=400 ms, a variable flip
angle spoiled gradient echo (with flip angles of 158, 308, 458,608 and 758) approach was employed to produce a 1282 R10
map over a 30-mm2 FOV. This method of T1 measurement is
attractive as the total measurement time required (for large
volumes of interest) is drastically reduced compared to spin
echo or inversion recovery methods, while still maintaining
high the SNR of those methods [29]. The slice thickness was
1.5 mm, TE=4.1 ms and NEX=4. The DCE-MRI protocol
employed a standard T1-weighted, GRE sequence to obtain
50 serial images for each of 12 axial planes in 60 min of
imaging. The DCE-MRI parameters were TR=100 ms, flip
angle=308, with other imaging parameters the same as
above. Three images were acquired before a bolus of
0.3 mmol/kg Gd-DTPA was delivered within 30 s via the tail
vein catheter. The tumor TOI was manually selected to allow
both whole-tumor and single-voxel analyses. Twenty voxels
within the perivertebral muscle were selected as the RR.
All images were first coregistered to the first precontrast
image using a mutual information-based rigid registration
algorithm [30] and then analyzed on both a TOI and voxel-
by-voxel basis. Tumor volume was calculated by manually
outlining the visible tumor and multiplying the number of
voxels within the outline by the voxel volume (0.08 mm3);
repeating this procedure twice (by different investigators)
provided a standard deviation estimate for the tumor
volume. An R10 map was constructed by fitting the multiflip
angle spoiled-GRE data to Eq. (19):
S ¼ S0 sina 1�exp �TRbR1ð Þð Þ= 1� exp �TRbR1Þbcosað Þð �;½ð19Þ
where a is the flip angle, S0 is a constant describing the
scanner gain and proton density, and we have assumed that
TEbT2*. Voxels for which Eq. (19) could not fit the data
were set to 0 and colored white on the parametric R10 map.
The R10 map was then used to estimate R1 time courses for
each voxel within the FOV from the raw signal intensity
time courses in the manner of Landis et al. [24]:
R1 tð Þ ¼ � 1=TRð Þblnf½S0bsinabexpð�TE=T2TÞ
�S tð Þbcosa�= S0bsinabexpð�TE=T2TÞ�S tð Þ½ �g; ð20Þ
where S0 represents fully relaxed magnetization for a given
pixel and is computed via Eq. (21):
S0 ¼ S� 1� expð � TR=T1Þcosa½ �=f 1� expð � TR=T1Þ½ �
� sinabexpð � TE=T2TÞg; ð21Þ
where S- is the steady-state pixel-averaged intensity before
CA was injected. In the computation of the R1 time course
from Eqs. (20) and (21), we again took TEbT2*.
R1 (
s-1)
0.5
1.0
1.5
2.0
2.5
time (min)0 10 20 30 40
R1
(s-1
)
0.5
1.0
1.5
2.0
2.5
R1 (
s-1)
0.5
1.0
1.5
2.0
2.5
3.0
RRTOIfit
RRTOIfit
RRTOIfit
no noise
5% noise
10% noise
A
B
C
Fig. 3. Results of fitting the RR model (Eq.(13)) to simulated TOI curves
with 0% (A), 5% (B) and 10% (C) noise added to the bdataQ sets. See
Table 1 for the returned parameter values.
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529524
We then input the RR and TOI R1(t) curves with the RR
model (Eq. (16)) into the curve-fitting routine. Again, the
Ktrans,RR and ve,RR values were fixed at their assigned
muscle values (0.1 min�1 and 0.1, respectively), while the
Ktrans,TOI and ve,TOI values were allowed to vary. For the
ROI analysis, a cluster of nine (3�3) contiguous voxels was
selected, averaged and input into the curve-fitting routine.
To compute parameter uncertainties for this analysis, we
Table 1
Summary results of the RR fits to simulated data with 0%, 5% and 10% noise a
Parameter 0% Noise 5% No
Actual Returned Accuracy Precision Return
Ktrans (min�1) 0.25 0.25 – – 0.23
ve 0.40 0.40 – – 0.40
first find the average absolute deviation of the data points
from the best-fitted curve returned by Eq. (16); that is, we
compute du 1n
Ptnti¼t0
je tiÞjð , where e(ti)ufit(ti)�data(ti).
Then each point in the best-fitted curve is summed with a
random value from �d to +d, yielding a new bdataQ set.
This new curve is then fit with Eq. (16) to yield a new set of
parameters. Repeating this process 100 times yields
100 values each for Ktrans,TOI and ve,TOI from which means
and standard deviations are computed.
Voxel-by-voxel analysis allows for the production of
pharmacokinetic parameter maps to probe tumor heteroge-
neity. For the Ktrans parametric map, each voxel was then
assigned a color based on the Ktrans value returned from the
fit routine. An identical procedure was used to construct the
ve map. Voxels for which the model either did not converge
or converged to unphysical values (i.e., Ktransb0.0 min�1,
KtransN2.5 min�1; veb0.0, veN1.0) were displayed as black.
Each slice in which the tumor was visible was mapped
(slices 4–8).
4. Results
4.1. Simulations
As stated previously, the R1,TOI(t) and R1,RR(t) curves of
Fig. 2B (the two data sets that would actually be measured
in a DCE-MRI experiment) were discretized with 1-min
temporal resolution and input into a curve-fitting routine for
extraction of Ktrans,TOI and ve,TOI. The results of those
simulations are presented in Fig. 3A and Table 1. The fit is
good and the parameters returned (Ktrans,TOI=0.25 min�1,
ve,TOI=0.4) are identical to the values used to construct the
simulated curve. Next, we tested the model’s sensitivity to
noise as described previously. The data and the subsequent
curve fits are seen in Fig. 3B (5% noise) and C (10% noise),
while the parameters output by themodel are given in Table 1.
Again, the fits are good and the model returns
Ktrans,TOI=0.25F0.02 min�1, ve,TOI=0.4F0.02, for the 5%
case and Ktrans,TOI=0.25F0.04 min�1, ve,TOI=0.42F0.05,
for the 10% case. The v2 for the slowly and rapidly enhancingregions were 6.32�10�4 and 6.05�10�4, respectively. These
results indicate that the RR model is robust enough to
accommodate reasonable experimental noise levels. We next
investigate the systematic errors that can enter if an incorrect
assignment of Ktrans,RR and/or ve,RR is made.
Fig. 4 displays three-dimensional plots of the errors in
the returned Ktrans,TOI and ve,TOI values if incorrect RR
values are assigned. A horizontal plane at 0% on the ver-
tical axis indicates zero error in the returned parameter.
dded to the TOI (see Fig. 3)
ise 10% Noise
ed Accuracy Precision Returned Accuracy Precision
0.92 F0.03 0.20 0.80 F0.03
1.00 F0.01 0.40 1.00 F0.01
-30
-20
-10
0
10
20
30
-30-20
-100
1020
30
-30-20
-100
1020
Ktr
ans,
TO
I err
or
Ktrans,RR error
ve,RR error
Ktrans,TOI error
-30
-20
-10
0
10
20
30
-30-20
-100
1020
30
-30-20
-100
1020
v e,T
OI e
rro
r
ve,RR error
ve,TOI error
0
30
Error scale
KKtrans,RR error
-10 0
20
-30-20
10
Fig. 4. Three-dimensional renderings of systematic errors resulting from incorrect assignment of the RR parameters K trans,RR and ve,RR. The vertical axis
represents the degree to which the incorrect RR parameter assignments affects the Ktrans,TOI (left panel) and ve,TOI (right panel) output parameters.
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529 525
Considering the ve,TOI error plot, maximum ve,TOI errors
occur at the points where ve,RR is at its maximum incorrect
value. In particular, the ve,TOI errors scale linearly with error
in ve,RR; that is, a �30% error in ve,RR yields a �30% error
in ve,TOI, and a +30% error in ve,RR yields a +30% error in
ve,TOI. The ve,TOI error is almost completely independent of
the Ktrans,RR value; if we pick any ve,RR error threshold, the
error introduced into the ve,TOI value is essentially the same
as we move along the Ktrans,RR error axis. Thus, the ve,TOIparameter is extremely stable to incorrect Ktrans,RR assign-
ment. This is reasonable because Ktrans determines the initial
uptake portion of the enhancement curve [1,31] and has
very little effect on the washout slope, whereas vedetermines the washout slope (and to a lesser extent, the
peak height achieved by the enhancement curve). Thus, it is
reasonable that mischaracterizing the uptake portion of the
curve (through assigning an incorrect value to Ktrans,RR) will
have little effect on the ve value. Eq. (16) essentially
calibrates the TOI curve to the RR curve, so errors in
Fig. 5. Panel A displays the slice 4 t =0 MR image, while panel B displays the t =
the tumor periphery, and to a much lesser extent, the tumor core, from panel B. T
and slowly enhancing TOIs, respectively, depicted in Fig. 6. Panel C displays the T
T10 map.
Ktrans,RR will translate into the Ktrans,TOI parameter almost
exclusively with little effect on ve,TOI. Similarly, error in
ve,RR effects both ve,TOI and (to a lesser extent) Ktrans,TOI
since (as mentioned previously) ve determines the washout
slope and influences the ultimate height achieved by the
enhancement curve. Consequently, the plot of Ktrans,TOI
error demonstrates more structure. Maximum errors again
occur only at the points where ve,RR and Ktrans,RR are both
off by +30% of their true values, and when they are both off
by �30% of their true values. However, when ve,RR is off by
�30% and Ktrans,RR is off by +30%, the error in Ktrans,TOI
approximately vanishes. A similar pattern occurs when ve,RRand Ktrans,RR are off by +30% and �30%, respectively, with
K trans,TOI error dipping to ~17%. This increases the
robustness of the returned Ktrans,TOI parameters as there is
a smaller area within the Ktrans,TOI error plot for which
Ktrans,TOI is significantly affected by incorrect assignments
of the RR parameters. These results are encouraging as the
amount of systematic error inherent in the model is tractable
40 min image obtained from this study. Note the significant enhancement in
he red and blue circles in panels A and B represent the rapidly enhancing
10 map from this same slice. A grayscale is provided at the far right for the
Fig. 7. The results of RR parameter mappings for (representative) slices 4–6
(panels A, C and E, respectively) and the associated color scale on the far
right. The superior portion of the displays many red voxels on both the
K trans (the left panels) and ve map (right panels), indicating K trans values
z0.27 min�1 and 0.20bveb0.45. These voxels are most likely associated
with highly perfused and leaky vessels with increased extravascular–
extracellular space volume fractions over healthy tissue (vec0.10).
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529526
and directly related to the amount of error in the RR
parameters. Many literature values for muscle Ktrans and veare within the 0.1 min�1 F30% and 0.1F30% explored
here. Neither parameter returns values worse than F30% of
the true value even in those cases when both Ktrans,RR and
ve,RR are off by that amount. In an elegant series of
experiments, Simpson et al. [16] showed that incorrect
characterization of the AIF can lead to errors in perfusion
estimation by as much as 60%. Nevertheless, these errors
are still of concern and we therefore propose a variation on
Eq. (16) in the Discussion.
4.2. Experimental
Fig. 5 displays an axial slice through the tumor hind limb
at t=0 (panel A) and 40 min obtained from this study, as well
as the corresponding T10 map (panel C). The tumor volume
was calculated at 536.0F26.8 mm3. The signal-to-noise ratio
is approximately 52 and 35 for a 20-voxel TOI and an
arbitrary voxel, respectively — both above the range
explored in the above simulations. There is a significant
change in signal intensity from panels A to B in the tumor
periphery, and to a much lesser extent, in the tumor core. The
T10 map is reasonable (for 7.0 T) with most muscle voxels
between 1.45 and 1.85 s, while most tumor voxels are
between 1.8 and 2.2 s. The t=4-min frame indicates two
voxel groups used for TOI analysis (red circles denote a
rapidly enhancing region, while the blue circle indicate a
slowly enhancing region), as well as the RR (white circle)
that was used for both the ROI and voxel analysis. The Fig. 6
circles, triangles and squares indicate the RR, a slowly
enhancing TOI (blue circle in Fig. 5) and a rapidly enhancing
TOI (red circle in Fig. 5), respectively. The results of the
fitting routines are displayed as solid (slowly enhancing) and
dashed (rapidly enhancing). Again, the fits are good and the
parameters are well within the range of reported values for a
9 pixel TOIs22 pixel RR
time (min)
0 10 20 30 40
R1
(s-1
)
0.4
0.8
1.2
1.6
RR slowly enhancing TOIrapidly enhancing TOI
Fig. 6. The results of the RR model fits to data taken from the TOIs labeled
in Fig. 5. The model accommodates both the rapidly enhancing data (filled
squares) and the slowly enhancing data (filled triangles). The data curve
used as the RR for every slice in the study is depicted as the filled circles.
number of tumor types [3,4,9,32]: Ktrans=0.039F0.002
min�1 and ve=0.46F0.01 for the slowly enhancing region,
Ktrans=0.35F0.05 min�1 and ve=0.31F0.01 for the rapidly
enhancing region. We proceed to voxel-by-voxel analysis to
construct pharmacokinetic parameter maps of Ktrans and ve.
As noted previously, the tumors were manually outlined
and each voxel within the outline was fit with Eq. (16) and
characterized by a Ktrans and ve value. These parameter
values are then assigned a color. All voxels for which the
fitting routine did not converge or return physical values are
colored black. The results of these mappings for slices 4–6
(representative slices) and the associated color scale are
displayed in Fig. 7. (The bottom slice is that depicted in
Fig. 5). First, consider the Ktrans parametric map of panel A.
The superior portion of the tumor displays many red voxels
indicating Ktrans values z0.27 min�1. These voxels are
most likely associated with highly perfused and leaky
vessels where angiogenesis-mediated neovascularization is
occurring, as is commonly the case in the periphery of many
tumor types [33,34]. A similar pattern, though not as
pronounced, is also seen in the inferior portion of the tumor.
Both regions slowly fade into the central portion of the
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529 527
tumor, which is characterized by reduced perfusion and/or
permeability (Ktransb0.15 min�1) and many black voxels.
Since nearly all of these voxels occur within the tumor core,
we assume that these are located in dense necrotic regions that
are poorly vascularized as well as containing high intra-
tumoral pressure to prevent both active and passive delivery,
respectively, of the CA to those voxels. As mentioned in the
Discussion, this is a presumption that needs to be verified by
comparison with histology (which we are actively pursuing).
Moving from panel A to C to E, the number of red pixels
greatly diminishes in the superior portion of the tumor, while
the inferior portion of the tumor maintains high perfusion
and/or vessel permeability. All slices seem to display a central
necrotic zone as manifested by many black pixels for which
there was little or no enhancement.
The ve parameter map of slice 4 (panel B) displays, in the
superior and inferior portions of the tumor, increased
extravascular–extracellular space volume fractions
(0.20bveb0.45) over healthy tissue (ve~0.10). Again, the
values reach a maximum in the tumor periphery and begin
to decrease toward the tumor core. This pattern is seen in
both panels D and F as well, though the central necrotic
zone appears to expand just as in the Ktrans maps.
5. Discussion
We have presented a method by which DCE-MRI data
can be quantitatively analyzed on a voxel-by-voxel basis for
the extravasation transfer constant, Ktrans, and extravascular–
extracellular space, ve, without direct measurement of the
AIF. The assumptions inherent in the method are those
common to all compartmental models (Eq. (1)); principally,
that the subject’s body may be represented by one or more
pools, or bcompartments,Q into and out of which the CA
dynamically flows, and that each compartment is assumed
to be bwell mixedQ in the sense that CA entering the
compartment is immediately distributed uniformly through-
out the entire compartment. The method is fast [the results
presented here indicate that the method can analyze an entire
1282 DCE-MRI data set in less than 5 min on a Pentium P4
(Intel, Santa Clara, CA) running at 2.4 GHz], easily applied
and straightforward in implementation, thereby making it
useful in, for example, experiments to measure tumor
kinetics before and after treatment. The results are reason-
able and consistent with other methods that attempt to
measure the AIF directly. We now discuss some of the
assumptions inherent in the RR model.
It should be noted that an additional assumption inherent
in Eqs. (11) and (12) (and, indeed, nearly all of the DCE-
MRI literature), though simplifying, may not be accurate. It
has recently been shown that there frequently exists signi-
ficant water exchange effects between separate compart-
ments, and this can lead to errors in the analysis of dynamic
MR data. Significant transendothelial water exchange
effects have been seen in arterial spin labeling [35,36] and
DCE-MRI data [37,38], while significant transcytolemmal
water exchange effects have been seen in diffusion-
weighted [39] and DCE-MRI [24,25,28] data. We acknowl-
edge that the theory presented here does not account for
those complicating factors, and we are currently working to
modify Eq. (16) to account for water exchange affects. We
also note that though Fig. 1 implies that the RR and TOI are
in close spatial proximity, they are actually separated by tens
of millimeters as indicated by Fig. 5. Thus, water exchange
between the RR and TOI compartments should not be
incorporated into the model.
The simulations reveal that systematic errors in Ktrans,TOI
and ve,TOI may be caused by incorrect assignments of
Ktrans,RR and ve,RR. That is, there could be both intra- and
interanimal variation in RR parameter values, and by
assigning literature reported values (particularly to
Ktrans,RR), systematic error will be introduced as manifested
by the above simulations. Thus, both the assignment of RR
parameters (interanimal variation) and the selection of the
RR itself (intraanimal variation) can confound the results
obtained by the RR model. An alternate approach would be
to formulate Eq. (16) in terms of ratios of Ktrans,TOI to
Ktrans,RR and ve,TOI to ve,RR, thereby allowing for the
production of relative Ktrans and ve maps. This could
potentially reduce the model’s systematic error due to
incorrect RR parameter assignments since a model reporting
relative parameter values requires no assignments on the
RR. If we make the following assignments:
R1uRuK trans;TOI=K trans;RR ð22Þ
R2uK trans;RR=ve;RR ð23Þ
R3uK trans;TOI=ve;TOI ð24Þ
then Eq. (18) may be expressed as Eq. (25):
R1;TOI Tð Þ ¼ R1bðR1;RRðTÞ � R10;RRÞ
þ R1½R2 � R3�bZT
0
ðR1;RRðtÞ � R10;RRÞ
� expð � R3 T � tð ÞÞdt þ R10TOI: ð25Þ
By noting that (R1/R3)R2=ve,TOI/ve,RR, Eq. (25) can pro-
vide a three-parameter fit to a DCE-MRI time course to
extract rKtrans(R1), rve((R1/R3)R2) and rkep(R
3) values (kepis the rate constant describing the flow of CA from the
extravascular–extracellular space to the blood plasma [1]),
where brQ denotes relative. Such relative measurements have
been shown to be of clinical value, perhaps most noticeably
in investigations of cerebral perfusion. Our preliminary
results have shown that Eq. (25) is very robust to exper-
imental noise, returns accurate values and slightly increases
standard deviations over the Eq. (16) model — as one would
expect for a model with an additional degree of freedom.
Furthermore, we have made some progress on applying an
iterative approach in which Eq. (25) is used to first fit an
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529528
enhancement curve (to acquire relative parameter measure-
ments), and then Eq. (16) is used to conduct a full four
parameter fit (Ktrans,TOI, Ktrans,RR, ve,TOI and ve,RR) subject to
the results of the relative parameter values. If realized, such an
approach would eliminate the requirement for assumptions
on the RR and effectively eliminate the problems associated
with inter- and intraanimal variation discussed previously.
This represents another advantage of working with the
integral form of the Kety equation rather than the differential
version as in Ref. [22]; in Kovar et al., RR values must be
assumed to acquire the AIF. In the iterative approach just
described, one need not do that and it is a direct consequence
of formulating the RR method as an integral equation.
In conclusion, we have proposed a simple compartmental
model that allows for estimation of the Ktrans and vepharmacokinetic parameters on a voxel-by-voxel basis
without needing to characterize the AIF. The method needs
to be verified by correlation with accepted methods such as,
for example, standard histological analysis. To improve the
accuracy of the analysis, Eq. (16) needs to be amended to
account for transcytolemmal water exchange. It also needs
to be verified that the RR method can detect subtle
longitudinal changes in tumor vasculature so that the
approach could be applied in, for example, studies
evaluating the efficacy of novel anticancer therapies.
Acknowledgments
We thank Dr. Calum Avison for reviewing an early
version of this manuscript. We thank Drs. Adam Anderson,
Bruce Damon, Natasha Deane, Robert Kessler and Kenneth
Niermann for many stimulating and informative discussions.
Mr. Richard Baheza and Mr. George Holburn gave excellent
assistance in managing technical imaging and animal care
issues. We thank the National Institutes of Health for funding
through NCI 1R25 CA92043 and 5RO1 EB00461.
References
[1] Tofts PS, Brix G, Buckley DL, Evelhoch JL, Henderson E, Knopp
MV, et al. Estimating kinetic parameters from dynamic contrast-
enhanced T1-weighted MRI of a diffusible tracer: standardized
quantities and symbols. J Magn Reson Imaging 1999;10:223–32.
[2] Daniel BL, Yen YF, Glover GH, Ikeda DM, Birdwell RL, Sawyer-
Glover AM, et al. Breast disease: dynamic spiral MR imaging.
Radiology 1998;209:499–509.
[3] Robinson SP, McIntyre DJO, Checkley D, Tessier JJ, Howe FA,
Griffiths JR, et al. Tumour dose response to the antivascular agent
ZD6126 assessed by magnetic resonance imaging. Br J Cancer 2003;
88:1592–7.
[4] Checkley D, Tessier JJL, Wedge SR, Dukes M, Kendrew J, Curry B,
et al. Dynamic contrast-enhanced MRI of vascular changes induced
by the VEGF-signalling inhibitor ZD4190 in human tumour xeno-
grafts. Magn Reson Imaging 2003;21:475–82.
[5] Kelz F, Furman-Haran E, Grobgeld D, Degani H. Clinical testing of
high-spatial-resolution parametric contrast-enhanced MR imaging of
the breast. Am J Roentgenol 2002;179:1485–92.
[6] Daldrup-Link HE, Rydland J, Helbich TH, Bjornerud A, Turetschek
K, Kvistad KA, et al. Quantification of breast tumor microvascular
permeability with feruglose-enhanced MR imaging: initial phase II
multicenter trial. Radiology 2003;229:885–92.
[7] Brasch R, Turetschek K. MRI characterization and grading angio-
genesis using a macromolecular contrast media: status report. Eur J
Radiol 2000;23:148–55.
[8] Morakkabati N, Leutner CC, Schmiedel A, Schild HH, Kuhl CK.
Breast MR imaging during or soon after radiation therapy. Radiology
2003;229:893–901.
[9] Hayes C, Padhani AR, Leach MO. Assessing changes in tumor
vascular function using dynamic contrast-enhanced magnetic reso-
nance imaging. NMR Biomed 2002;15:154–63.
[10] Delille J-P, Slanetz PJ, Yeh ED, Halpern EF, Kopans DB, Garrido L.
Invasive ductal breast carcinoma response to neoadjuvant chemother-
apy: noninvasive monitoring with functional MR imaging-pilot study.
Radiology 2003;228:63–9.
[11] Kety SS. Peripheral blood flow measurement. Pharmacol Rev 1951;3:
1–41.
[12] Fritz-Hansen T, Rostrup E, Larsson HBW, Sondergaard L, Ring P,
Henrikson O. Measurement of the arterial concentration of Gd-DTPA
using MRI: a step toward quantitative perfusion imaging. Magn Reson
Med 1996;36:225–31.
[13] Larsson HBW, Stubgaard M, Frederiksen JL, Jensen M, Henriksen O,
Paulson OB. Quantitation of blood–brain barrier defect by magnetic
resonance imaging and gadolinium-DTPA in patients with multiple
sclerosis and brain tumors. Magn Reson Med 1990;16:117–31.
[14] Weinmann HJ, Laniado M, Mutzel W. Pharmacokinetics of GdDTPA/
dimeglumine after intravenous injection into healthy volunteers.
Physiol Chem Phys 1984;16:167–72.
[15] Degani H, Gusis V, Weinstein D, Fields S, Strano S. Mapping
pathophysiological features of breast tumors by MRI at high spatial
resolution. Nat Med 1997;3:780–2.
[16] Simpson NE, He Z, Evelhoch JL. Deuterium NMR tissue perfusion
measurements using the tracer uptake approach: I. Optimization of
methods. Magn Reson Med 1999;42:42–52.
[17] Port RE, Knopp MV, Hoffmann U, Milker-Zabel S, Brix G.
Multicompartment analysis of gadolinium chelate kinetics: blood–
tissue exchange in mammary tumors as monitored by dynamic MR
imaging. J Magn Reson Imaging 1999;10:233–41.
[18] van Osch MJO, Vonken E-JPA, Viergever MA, Grond J, Bakker CJG.
Measuring the arterial input function with gradient echo sequences.
Magn Reson Med 2003;49:1067–76.
[19] Kim YR, Rebro KJ, Schmainda KM.Water exchange and inflow affect
the accuracy of T1-GRE blood volume measurements: implications for
the evaluation of tumor angiogenesis. Magn Reson Med 2002;47:
1110–20.
[20] McIntyre DJO, Ludwig C, Pasan A, Griffiths JR. A method for
interleaved acquisition of a vascular input function for dynamic
contrast-enhanced MRI in experimental rat tumours. NMR Biomed
2004;17:132–43.
[21] Lammertsma AA, Bench CJ, Hume SP, Osman S, Gunn K, Brooks
DJ, et al. Comparison of methods for analysis of clinical [11C]Raclopr-
ide studies. J Cereb Blood Flow Metab 1995;16:42–52.
[22] Kovar DA, Lewis M, Karczmar GS. A new method for imaging
perfusion and contrast extraction fraction: input functions derived
from reference tissues. J Magn Reson Imaging 1998;8:1126–34.
[23] Rainville ED, Bedient PE. Elementary differential equations. New
York (NY)7 McMillan; 1989. p. 17–46.
[24] Landis CS, Li X, Telang FW, Coderre JA, Micca PL, Rooney WD,
et al. Determination of the MRI contrast agent concentration time
course in vivo following bolus injection: effect of equilibrium
transcytolemmal water exchange. Magn Reson Med 2000;44:563–74.
[25] Yankeelov TE, Rooney WD, Xin Li, Springer CS. Variation of the
relaxographic bshutter-speedQ for transcytolemmal water exchange
affects CR bolus-tracking curve shape. Magn Reson Med 2003;50:
1151–69.
[26] Padhani AR, Hayes C, Landau S, Leach MO. Reproducibility of
quantitative dynamic MRI of normal human tissues. NMR Biomed
2002;15:143–53.
T.E. Yankeelov et al. / Magnetic Resonance Imaging 23 (2005) 519–529 529
[27] Donahue KM, Weiskoff RM, Parmelee DJ, Callahan RJ, Wilkinson
RA, Mandeville JB, et al. Dynamic Gd-DTPA enhanced MRI
measurement of tissue cell volume fraction. Magn Reson Med 1995;
34:423–32.
[28] Landis CS, Li X, Telang FW, Molina PE, Palyka I, Vetek G, et al.
Equilibrium transcytolemmal water-exchange kinetics in skeletal
muscle in vivo. Magn Reson Med 1999;42:467–78.
[29] Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic
resonance imaging: physical principles and sequence design. New
York7 Wiley-Liss; 1999. p. 654.
[30] Li R. Automatic placement of regions of interest in medical
images using image registration. MSEE Thesis. Vanderbilt
University; 2001.
[31] Yankeelov TE. The effects of equilibrium transcytolemmal water
exchange on magnetic resonance imaging measurement of contrast
reagent pharmacokinetics. Ph.D. dissertation. Stony Brook (NY):
State University of New York; 2003. p. 176–280.
[32] Maxwell RJ, Wilson J, Prise VE, Vojnovic B, Rustin GJ, Lodge MA,
et al. Evaluation of the anti-vascular effects of combretastatin in
rodent. NMR Biomed 2002;15:89–98.
[33] Su M-Y, Cheung Y-C, Fruehauf JP, Yu H, Nalcioglu O, Mechetner E,
et al. Correlation of dynamic contrast enhancement MRI parameters
with microvessel density and VEGF for assessment of angiogenesis in
breast cancer. J Magn Reson Imaging 2003;18:465–77.
[34] Cha S, Johnson G, Wadghiri YZ, Jin O, Babb J, Zagzag D, et al.
Dynamic, contrast-enhanced perfusion MRI in mouse gliomas:
correlation with histopathology. Magn Reson Med 2003;49:848–55.
[35] Barbier EL, St. Lawrence KS, Grillon E, Koretsky AP, Decorps M. A
model of blood–brain barrier permeability to water: accounting for
blood inflow and longitudinal relaxation effects. Magn Reson Med
2002;47:1100–9.
[36] Parkes LM, Tofts PS. Improved accuracy of human cerebral blood
perfusion measurements using arterial spin labeling: accounting for
capillary water permeability. Magn Reson Med 2003;48:27–41.
[37] Schwarzbauer C, Morrissey SP, Deichmann R, Hillenbrand C, Syha J,
Adolf H, et al. Quantitative magnetic resonance imaging of capillary
water permeability and regional blood volume with an intravascular
MR contrast agent. Magn Reson Med 1997;37:769–77.
[38] Donahue KM, Weisskoff RM, Burstein D. Water exchange and
diffusion as they influence contrast enhancement. J Magn Reson
Imaging 1997;7:102–10.
[39] Lee J-H, Springer CS. Effects of equilibrium exchange on diffusion-
weighted NMR signals: the diffusigraphic bshutter-speedQ. Magn
Reson Med 2003;49:450–8.