Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and...
-
Upload
anubha-gupta -
Category
Documents
-
view
232 -
download
2
Transcript of Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and...
![Page 1: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/1.jpg)
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.1983
Quantitative determination of cetirizine enantiomers
in guinea pig plasma, brain tissue and microdialysis
samples using liquid chromatography/tandem
mass spectrometry
Anubha Gupta1*, Britt Jansson1, Pierre Chatelain2, Roy Massingham2
and Margareta Hammarlund-Udenaes1
1Division of Pharmacokinetics and Drug Therapy, Department of Pharmaceutical Biosciences, Uppsala University, Box 591,
SE-751 24 Uppsala, Sweden2UCB S.A. Pharma Sector, Braine-l’Alleud, Belgium
Received 6 April 2005; Revised 26 April 2005; Accepted 26 April 2005
Sensitive enantioselective liquid chromatographic assays using tandem mass spectrometric
detection were developed and validated for the determination of S-cetirizine (S-CZE) and R-
cetirizine (R-CZE) in guinea pig plasma, brain tissue, and microdialysis samples. Enantioselective
separation was achieved on an a1-acid glycoprotein column within 14 min for all methods. A cetir-
izine analog, ucb 20028, was used as internal standard. Cetirizine and the internal standard were
detected by multiple reaction monitoring using transitions m/z 389.1! 200.9 and 396.1! 276.1,
respectively. The samples were prepared using protein precipitation with acetonitrile. For guinea
pig plasma, the assay was linear over the range 0.25–5000 ng/mL for both S-CZE and R-CZE, with a
lower limit of quantification (LLOQ) of 0.25 ng/mL. For the brain tissue and microdialysis samples,
the assays were linear over the range 2.5–250 ng/g and 0.25–50 ng/mL, respectively, and the LLOQ
values were 2.5 ng/g and 0.25 ng/mL, respectively. The intra- and inter-day precision values were
�7.1% and �12.6%, respectively, and the intra- and inter-day accuracy varied by less than �8.0%
and �6.0% of the nominal value, respectively, for both enantiomers in all the matrices investigated.
Copyright # 2005 John Wiley & Sons, Ltd.
Cetirizine (CZE) is a non-sedative second-generation H1-
antihistamine widely prescribed for the treatment of allergic
disorders. It exists as a racemic mixture of levocetirizine
(R-CZE) and dextrocetirizine (S-CZE) (Fig. 1). The antihista-
minic properties of CZE are attributable to R-CZE.1 The
main side effects of the first-generation antihistamines are
sedation, drowsiness and impaired performance, which
are caused by their action on histamine H1-receptors in the
brain. The second-generation antihistamines have lower inci-
dence of these side effects than the first-generation drugs
because of lower brain penetration. It has been suggested
that the reduced brain penetration of CZE could be the result
of P-glycoprotein-mediated efflux at the blood-brain barrier
(BBB).2,3 Because of the probable involvement of an active
efflux process, the transport of CZE across the BBB could be
stereoselective and therefore is worthy of investigation.
Microdialysis is a valuable tool for studying the transport
of drugs across the BBB.4 It offers the advantage of the ability
to estimate unbound drug concentrations in both blood and
brain in one animal over time. Measuring the concentration of
unbound drug in the blood and the brain interstitial fluid
(ISF), and the total concentration of drug in the brain, allows
characterization of both transport into and distribution
within the brain. The guinea pig is preferable to the rat as a
model species because of the higher brain concentrations of
H1-receptors, and is often used in work on antihistamines.5
Investigation of the brain transport of CZE enantiomers
using microdialysis techniques requires an analytical method
capable of detecting the CZE enantiomers in plasma, whole
brain tissue, and in Ringer’s solution (perfusion fluid). CZE is
highly protein-bound, which will result in lower unbound
concentrations in the blood. It has been reported that only a
small proportion of the available CZE enters the brain,
resulting in even lower unbound concentrations of the two
enantiomers in brain ISF.6 Therefore, a highly sensitive
analytical method was a prerequisite for this study, and mass
spectrometric detection was an obvious choice. Bakhtiar et al.
have reviewed the use of mass spectrometry with enantio-
selective liquid chromatography.7 Other challenges with
such an analytical method involve the handling of the small
sample volumes (8–15 mL) obtained from the microdialysis
experiment. The Ringer’s solution used in the microdialysis
experiments contains high salt levels, which causes signal
depression in electrospray ionization (ESI) mass spectro-
metry. In addition, only a limited volume of blood can be
Copyright # 2005 John Wiley & Sons, Ltd.
*Correspondence to: A. Gupta, Division of Pharmacokinetics andDrug Therapy, Department of Pharmaceutical Biosciences,Uppsala University, Box 591, SE-751 24 Uppsala, Sweden.E-mail: [email protected]
![Page 2: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/2.jpg)
drawn from each guinea pig without affecting the home-
ostasis of the animal.
Several methods for the quantification of racemic CZE
have been described in the literature.8–14 However, to date,
only Choi et al. have described a stereoselective method,
using UV detection, for the determination of CZE in rat
plasma15 and human urine.16 A chiral a1-acid glycoprotein
(AGP) column was used by these workers to separate the
enantiomers. A volume of 100 mL was used for rat plasma,
and the lower limit of quantification (LLOQ) was 800 ng/mL.
For human urine samples, a 5 mL aliquot was used, and the
LLOQ was 400 ng/mL. The internal standard was roxatidine,
eluting at 32 min. This method, however, has poor sensitivity
and takes a relatively long time.
Four non-enantioselective methods which use mass spec-
trometric detection for the estimation of cetirizine concentra-
tions have also been described.8–10,13 Eriksen et al.9 used a
sample volume of 250 mL and validated the method with an
LLOQ of 5 ng/mL. Li et al.10 and Song et al.13 achieved an
LLOQ of 1 ng/mL, using a sample volume of 100mL. De Jager
et al.8 reported an LLOQ of 0.5 ng/mL, using a 100 mL plasma
sample.
In this paper we describe the development and validation
of three enantioselective liquid chromatographic assay
methods using tandem mass spectrometric detection (LC/
MS/MS) for the determination of S- and R-CZE in guinea pig
plasma (plasma method), guinea pig brain tissue (brain
method), and microdialysis samples from guinea pig blood
and brain ISF (Ringer method). Enantioselective separation
was achieved within 14 min for all three methods. Because of
the sensitivity and specificity of MS, it was possible to both
keep the sample volume small and retain a high level of
sensitivity. Preparation of the samples using protein pre-
cipitation was simple and rapid. The methods were used to
analyze guinea pig plasma and brain tissue samples and
samples from microdialysis of blood and brain ISF to in-
vestigate the possible stereoselective transport of CZE across
the BBB.
EXPERIMENTAL
Chemicals and reagentsRacemic CZE, S-CZE, and the internal standard (IS, ucb
20028), were provided by UCB Pharma (Braine-l’Alleud,
Belgium). Acetonitrile (ACN) and 2-propanol (LiChrosolve,
gradient grade), ammonia and ammonium acetate (analytical
grade) were obtained from Merck (Darmstadt, Germany).
Methanol (HPLC gradient grade) was purchased from
J.T. Baker (Deventer, The Netherlands). MQ water, which
is in-house deionized water, further purified in a Milli-Q
Academic system (Millipore, Bedford, MA, USA), was used.
Ringer’s solution (KCl 33 mg, CaCl2 H2O 330 mg, NaCl
8.6 g, sterilized water for injection Ph. Eur. ad 1000 mL,
pH ca. 6, osmolarity 290 mosm/kg H2O; Fresenius Kabi
Norge AS) was obtained from the Uppsala Hospital Phar-
macy, Uppsala, Sweden. Bovine serum albumin (BSA, initial
fractionation by heat shock) was provided by Sigma Chemi-
cal Co. (St. Louis, MO, USA). Blank guinea pig plasma and
brain tissue were obtained from male Dunkin Hartley guinea
pigs (Charles River, France).
InstrumentationChemical analyses were performed using an LC/MS/
MS system consisting of two Shimadzu LC-10AD pumps
(Shimadzu, Kyoto, Japan), a Triathlon 900 autosampler
(Spark Holland, The Netherlands) equipped with a 100 mL
loop and an extra six-port valve, a pre-filter (0.5mm;
Upchurch Scientific Inc, Hagersten, Sweden), an AGP col-
umn (150� 4.0 mm; ChromTech, Hagersten, Sweden), and a
Quattro Ultima triple-quadrupole mass spectrometer (Micro-
mass, Manchester, UK).
Liquid chromatographyFor the plasma and brain methods, the mobile phase con-
sisted of 6.5% ACN in 10 mM ammonium acetate, pH 7.0
(adjusted with 1 % ammonia). The column was maintained
at 308C in a water bath (type JB1, Grant Instruments,
Cambridge, UK).
For the Ringer method, the column was maintained at
room temperature and 5.5% ACN was used in the mobile
phase. A switch was used after the analytical column to avoid
salt overload in the mass spectrometer. Two pumps were
used; one was connected to the column through the auto-
sampler and the other was connected directly to the extra
six-port valve. During the time period 0–6.5 min after
sample injection, the flow of the mobile phase from the
column was discarded and the flow of mobile phase from
the other pump entered the mass spectrometer. During the
period 6.5–14 min, the flow from the column entered the
mass spectrometer.
The mobile phase was pumped at a flow rate of 0.9 mL/
min, generating a pressure of 100 bar. The flow was split
to 250–270 mL/min before entering the mass spectrometer.
The order of elution of the enantiomers was determined by
injection of pure S-CZE.
Mass spectrometeryThe triple-quadropole mass spectrometer was used in
the positive ESI mode. The measurements were made at
NO
O
O
O
ucb 20028
ONN
Cl
O
OH*
Cetirizine
Figure 1. Chemical structures of cetirizine and the internal
standard (ucb 20028). The asterisk represents the chiral
center.
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
1750 A. Gupta et al.
![Page 3: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/3.jpg)
4008C desolvation temperature, 1308C source temperature,
200 L/h cone gas, 1000 L/h desolvation gas and a collision
gas pressure of 3� 10�3 Torr. The capillary voltage was
1.5 kV and the cone voltage was 40 V. The collision energy
was set to 20 eV for CZE and 25 eV for the IS. The resolution
was set at 0.9 u at half height for Q1 and Q3. The dwell time
for all channels was set to 0.5 s and inter-channel delay was
0.1 s. To monitor the multiple reactions (MRM), the MS/MS
parameters were optimized to provide the highest sensitivity
by directly infusing (200 mL/min) working solutions con-
taining CZE and IS (100 ng/mL in the mobile phase) using a
Harvard 22 syringe pump (Harvard Apparatus Inc., Hollis-
ton, USA). The transitions that gave the most intense signals
were m/z 389.1! 200.9 for CZE and 396.1! 276.1 for the IS
(Fig. 2).
Preparation of standard and qualitycontrol (QC) samplesStock solutions of CZE and IS (1 mg/mL) were prepared in
MQ water and stored at 48C, protected from light.
For the plasma method, blank guinea pig plasma was
spiked with the stock CZE solution. The standard curve
consisted of ten samples containing each enantiomer in
the range of 0.25–5000 ng/mL. Three QC samples, low-
concentration (QCL, 0.75 ng/mL), medium-concentration
(QCM, 250 ng/mL), and high-concentration (QCH, 3710 ng/
mL), were prepared.
For the brain method, the blank brain sample was pre-
pared by homogenizing blank brain tissue with a 4-fold
volume (w/v) of saline. The standard samples were prepared
by spiking blank brain tissue samples with the stock
CZE solutions before homogenization; the brain was then
homogenized with a volume of saline equivalent to a 4-fold
volume (w/v) minus the volume of the added analyte
solution, and the homogenate was used for further proces-
sing. The standard curve for guinea pig brain composed of
seven samples containing each enantiomer in the range of
2.5–250 ng/g brain. Three QC samples (concentrations 12.5,
75 and 188 ng/g brain) were prepared.
For the Ringer method, BSA was soaked in Ringer’s
solution overnight and the volume was adjusted the next
day to give a blank of 0.5% BSA in Ringer’s solution. This
solution was then spiked with the stock CZE solutions to give
eight standard samples in the range of 0.25–50 ng/mL for
each enantiomer. Three QC samples (concentrations 0.38,
7.5 and 38 ng/mL) were prepared.
The standards and QC samples of guinea pig plasma,
guinea pig brain homogenate, and Ringer-BSA solution were
stored at �208C until analysis. The working solution for
the IS was prepared daily in ACN.
Study sample collectionThe guinea pigs were anaesthetized by inhalation of
Enflurane1 (2.5% balanced with 1.5 L/min oxygen and
1.5 L/min nitrous oxide) and 0.25 mL of Dormicum (midazo-
lam 5 mg/mL) intraperitoneally. FEP tubings were inserted
into the left jugular vein for drug administration and into
the left common carotid artery for blood sampling. The blood
probe (CMA/20, polycarbonate, 20 kDa cutoff, 10 mm) was
inserted into the right jugular vein. A CMA/12 guide cannula
with a dummy probe was implanted in the brain stereotaxi-
cally and fixed to the skull with a screw and dental cement.
When the cement had stiffened, the dummy probe was
replaced with the 3 mm (CMA/12) brain probe. The guinea
pig was placed in a CMA/120 system for freely moving
animals with free access to water and food, and the experi-
ment was performed approximately 24 h later. On the day
of the experiment, each guinea pig received a 60 min intrave-
nous (i.v.) infusion of cetirizine (2.7 mg/kg). The microdialy-
sis blood and brain ISF samples were obtained by collecting
dialysate from the microdialysis probes in fractions of 15 min.
The probes were continuously perfused with Ringer’s solu-
tion containing BSA from the beginning of the infusion
up to 360 min. The blood samples were collected at pre-
defined time points, and the plasma was separated by centri-
fugation for 5 min at 10 000 rpm. The brain tissue samples
were obtained by decapitating the guinea pig at the end
of the experiment. All the samples were stored at �208Cuntil analysis.
Sample preparation
Plasma methodPlasma (50mL) was precipitated with ACN (100mL) con-
taining IS at a concentration of 100 ng/mL in a 1.5 mL poly-
propylene Eppendorf tube (Brand, Wertheim, Germany).
The sample was vortexed for 5 s and centrifuged at
100 150 200 250 300 350 400m/z0
100
%
276
176167 202
396
338
50 100 150 200 250 300 350 400m/z0
100
%
201
201
50166
201
389
MRM 389.1 200.9
MRM 396.1 276.1
(a)
(b)
Figure 2. Product ion MS/MS spectra of [MþH]þ ions of (a)
IS and (b) CZE.
Cetirizine enantiomers in guinea pig plasma and brain ISF and homogenate 1751
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 4: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/4.jpg)
10 000 rpm for 5 min using a Force 7 centrifuge (Denver
Instrument Co., Denver, CO, USA). Then, 50 mL of the super-
natant were evaporated at 408C under N2 gas, and the residue
was dissolved in 500mL mobile phase by vortex mixing
and sonication. Finally, 60 mL of the reconstituted sample
were transferred to a polypropylene autosampler vial and a
volume of 50 mL was injected, using the partial loop fill tech-
nique, onto the LC/MS/MS system.
Brain methodA volume of 50mL of the brain homogenate was sampled and
processed as in the plasma method but with the following
modifications: the concentration of IS used was 5 ng/mL,
the residue was dissolved in 1000 mL mobile phase, and a
volume of 40 mL was injected.
Ringer methodSamples from the microdialysis experiment were collected
directly into polypropylene autosampler vials. The vials
were weighed before and after sample collection. The volume
collected varied from 6 to 8mL. A volume of 8mL of each
standard or QC sample was transferred into a polypropylene
autosampler vial. The microdialysis study, standard and QC
samples were precipitated with 100 mL ACN containing IS
(2 ng/mL), as described above. The supernatant (100mL)
was transferred to an Eppendorf tube and evaporated. The
Figure 3. MRM chromatograms of (a) a blank guinea pig plasma sample, (b) a spiked
guinea pig plasma sample containing 125 ng/mL of each of S- and R-CZE plus 100 ng/
mL of the IS, and (c) a plasma sample from a guinea pig dosed with CZE where the
measured concentrations of S- and R-CZE were 190 and 583ng/mL, respectively.
1752 A. Gupta et al.
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 5: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/5.jpg)
residue was reconstituted in 50 mL mobile phase and a
volume of 40 mL was injected.
For all three methods, the autosampler was thermostated
at 88C. A wash solution (50% ACN) of 1500mL was used
between the injections.
ValidationLinearity, precision, accuracy, specificity and lower limit
of quantification (LLOQ) were assessed. The linearity
was determined from peak area ratios (analyte/IS) as a
function of analyte concentration, using linear regression.
For the plasma method, the calibration curve was
forced through the origin. For both the brain and Ringer
methods a weighting factor of 1/y2 was used and the
calibration curve was not forced through the origin.
Peak areas were measured using Masslynx software,
version 4.0, individually checked and rectified manually if
required.
The precision was assessed as the coefficient of variation
(CV) of the back-calculated concentrations of the three
QC samples. The accuracy was calculated as the percent
deviation of the analyzed concentration from the nominal
value of the three QCs. For intra-day precision and accuracy,
one calibration curve and six independent sets of three QCs
were analyzed on one occasion. The inter-day precision
and accuracy were determined by two methods, either by
analyzing six independent sets of three QCs on two
additional occasions, or by pooling the two sets of three
QCs during six runs. During these runs, QC samples were
interspersed with the unknown study samples. The LLOQ
was set at the concentration that could be analyzed with a
precision of �20% and an accuracy of �20%.
(a)
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00Time0
100
%
0
100
%
CZE_111_74 MRM of 2 Channels ES+396.15 > 276.1
597
CZE_111_74 MRM of 2 Channels ES+389.1 > 200.9
1.20e3
(b)
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00Time0
100
%
0
100
%
CZE_111_52 MRM of 2 Channels ES+396.15 > 276.1
9.52e3
CZE_111_52 MRM of 2 Channels ES+389.1 > 200.9
1.58e4
(c)
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00Time0
100
%
0
100
%
CZE_111_61 MRM of 2 Channels ES+396.15 > 276.1
8.94e3
CZE_111_61 MRM of 2 Channels ES+389.1 > 200.9
1.61e4
S-CZE
S-CZE R-CZE
R-CZE
IS
IS
Figure 4. MRM chromatograms of (a) blank guinea pig brain tissue sample, (b) a spiked
standard sample containing 50 ng/g of each of S- and R-CZE plus 5 ng/mL of the IS, and (c)
a homogenized brain tissue sample from a guinea pig dosed with CZE where the measured
concentrations of S- and R-CZE were 91 and 56 ng/g, respectively.
Cetirizine enantiomers in guinea pig plasma and brain ISF and homogenate 1753
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 6: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/6.jpg)
Application of the analytical methodThe validated methods were applied to the analysis of
guinea pig plasma, brain and microdialysis blood and brain
ISF samples obtained from the experimental study of the
transport of CZE enantiomers across the BBB.
RESULTS AND DISCUSSION
Chromatographic conditionsTo optimize the chromatographic separation of CZE enantio-
mers, a mobile phase giving a resolution of 1.8 or higher and a
retention time less than 15 min was selected. The enantiose-
lective separation on the AGP column was affected by the
type (ACN, methanol or 2-propanol) and concentration of
organic modifier, the pH of the mobile phase, and the buffer
concentration. Higher concentrations of organic modifier
resulted in shorter retention times. Increases in the pH
improved the resolution of the peaks. The recommended
pH range for the AGP column is 4 to 7.17 Adequate separa-
tion of the enantiomers within a reasonable time was
achieved with a mobile phase of pH 7.0 containing 6.5%
ACN. A reduction in retention time after injecting ap-
proximately 3500 samples onto the AGP column indicated
that the number of theoretical plates was reduced. The
ACN content was therefore lowered to achieve the same
retention times.
Figure 5. MRM chromatograms of (a) blank Ringer’s solution with BSA, (b) a spiked
Ringer-BSA solution sample containing 25 ng/mL of each of S- and R-CZE plus 2 ng/mL of
the IS, and (c) a Ringer-BSA solution sample from the brain tissue of a guinea pig dosed
with CZE where the measured concentrations of S- and R-CZE were 0.80 and 0.55 ng/mL,
respectively.
1754 A. Gupta et al.
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 7: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/7.jpg)
In the previously reported enantioselective assay for CZE,
the mobile phase used was 5% ACN in 10 mM phosphate
buffer, at pH 7.15,16 In our study, ammonium acetate was
selected as the buffer because of its compatibility with LC/
MS systems.7 Buffer concentrations of 5–20 mM were
investigated. At pH 7, a buffer concentration of 10 mM
gave better resolution than 5 mM. Ammonium acetate
concentrations of 20 mM did not significantly influence the
resolution of the peaks, but the response of the mass
spectrometer was half that seen using 10 mM. On the basis
of these observations, the mobile phase composition was set
to 6.5% ACN in 10 mM ammonium acetate, at pH 7.0. A
column temperature of 308C resulted in decreased retention
times without compromising the peak resolution but, for the
Ringer method, the high salt concentration affected the signal
of S-CZE. By maintaining the column at room temperature
and using 5.5% ACN in the mobile phase for the Ringer
method, the retention times for the CZE enantiomers were
increased and the salts were eluted before the peaks.
Representative chromatograms for the plasma method, the
brain method and the Ringer method are shown in Figs. 3–5.
S-CZE, R-CZE and IS were eluted with retention times of
approximately 6.9, 8.7, and 7.9 min, respectively, for the
plasma samples; at 6.5, 8.2, and 7.7 min, respectively, for the
brain samples; and at 8.6, 11.4, and 11.1 min, respectively,
for the microdialysis samples.
Sample preparationVarious methods of sample preparation for the estimation
of CZE have been reported in the literature. These
include liquid-liquid extraction,11,14 solid-phase extraction
(SPE),9,10,13 and protein precipitation.8,12,15 The methods
screened in this study included SPE and protein precipita-
tion with acetonitrile. For a sensitive protein column an
SPE method would have been preferable, but CZE was
found to be adsorbed to the SPE assembly (C18 Sep-Pak1).
Protein precipitation with acetonitrile was a more rapid
and simple preparation approach, but this method can result
in high peak suppression. In the plasma method the sample
was diluted 30 times. The response of CZE and IS, obtained
from the extracted spiked plasma sample, was similar to
that from spiked mobile phase, indicating that suppression
effects were not significant. However, a depression in the
baseline was observed at 4.5–5 min in the chromatogram of
a blank plasma sample, shown in Fig. 3.
The method used for the brain homogenate samples was
that reported by Chen et al.2 The effect of the matrix was more
pronounced for the brain samples than the plasma samples.
When the residue was reconstituted in 500mL mobile phase,
the signal for S-CZE was depressed and the peak area for
S-CZE was approximately half that for R-CZE. When the
residue was dissolved in 1000mL of mobile phase, this
difference disappeared.
The samples obtained in microdialysis studies are usually
protein-free, allowing direct injection into the chromato-
graphic system. However, for CZE, 0.5% BSA was added to
the Ringer’s solution to prevent the adsorption of CZE onto
the materials used in the microdialysis experiments. For this
reason, the microdialysis samples needed to be prepared
before they were injected into the column.
ValidationThe standard curve obtained using the plasma method was lin-
ear over the range of 0.25–5000 ng/mL for both S- and R-CZE
(Fig. 6(a)). The three standard samples with the highest con-
centrations (5000, 4500 and 2500 ng/mL) and the QCH were
dissolved in 1000mL of mobile phase to avoid overloading
the mass spectrometer. All the unknown study samples that
showed a peak area larger than that of the standard sample
with a concentration of 2500 ng/mL were diluted twice before
re-injection. The standard curve obtained using the brain
method was linear over the range of 2.5–250 ng/mL and that
using the Ringer method was linear over 0.25–50 ng/mL. Both
these calibration curves were made with a weighting factor
of 1/y2, which resulted in an even residual distribution,
giving similar emphasis to all concentrations over the entire
standard curve range. The coefficient of determination was
greater than 0.996 for both enantiomers in all methods.
(a)
S-CZEy = 0.0051xR2 = 0.9996
R-CZEy = 0.0055xR2 = 0.9998
0
5
10
15
20
25
30
0 1000 2000 3000 4000 5000 6000
Concentrat ion (ng/mL)
Rat
io
(b)
S-CZEy = 0.0358x + -0.0049
Coefficient of Determinat ion = 0.9984
R-CZEy = 0.0363x + -0.0037
Coefficient of Determination = 0.9983
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
0 50 100 150 200 250 300
Concentrat ion (ng/mL)
Rat
io
(c)
S-CZEy = 0.0269x + - 0.0002
Coefficient of Determinat ion = 0.9975
R-CZEy = 0.0305x + - 0.0012
Coefficient of Determinat ion = 0.9962
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
0 10 20 30 40 50 60
Concentration (ng/mL)
Rat
io
Figure 6. Standard curves for (a) the plasma method, (b)
the brain method, and (c) the Ringer method. S-CZE and R-
CZE are represented by diamonds and squares, respectively.
The response is expressed as the ratio of the area of the
analyte to that of the internal standard.
Cetirizine enantiomers in guinea pig plasma and brain ISF and homogenate 1755
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 8: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/8.jpg)
The precision and accuracy of the plasma, brain and Ringer
methods are summarized in Tables 1–3. For the plasma
method, the intra-day precision was �7.1% and the accuracy
deviated by less than �3.1% of the nominal value. In this
method, the inter-day precision was�12.6% and the accuracy
deviated by less than �1.8% of the nominal value. For the
brain method, the intra- and the inter-day precisions were
�4.2% and �6.8%, respectively. The intra- and inter-day
accuracies deviated by less than �8.0% and �6% of the
nominal value, respectively. For the Ringer method, the intra-
day precision was�6.5% and the accuracy varied by less than
�5.2% of the nominal value. The inter-day precision was
�10% and the accuracy varied by less than �5.9% of the
nominal value.
The enantioselective assay of CZE was specific, as blank
plasma, brain and Ringer’s solution with 0.5% BSA samples
were free from any direct interferences in the MRM channels
at the retention times for IS, S-CZE and R-CZE (Figs. 3–5).
Table 2. Intra- and inter-day precision and accuracy of cetirizine enantiomers in guinea pig brain tissue. The enantiomers were
separated using an a1-acid glycoprotein (AGP) column prior to tandem mass spectrometry. Ucb 20028 was used as internal
standard
QCH (188 ng/mL) QCM (75.8 ng/mL) QCL (12.6 ng/mL) LLOQ (2.51 ng/mL)
S R S R S R S R
Intra-dayObserved conc.(ng/mL) 192 191 81.4 81.9 12.9 12.9 2.54 2.61CV (%) 2.9 2.9 2.6 2.4 3.4 4.2 10.2 7.6Accuracy (%) 1.8 1.7 7.3 8.0 1.8 1.8 1.3 4.0n 6 6 6 6 6 6 6 6Inter-dayObserved conc.(ng/mL) 190 190 80.2 80.4 13.0 13.1CV (%) 2.6 2.9 6.4 6.8 4.9 4.8Accuracy (%) 1.0 1.0 5.8 6.0 3.0 3.5n 18 18 18 18 18 18
Table 1. Intra- and inter-day precision and accuracy of cetirizine enantiomers in guinea pig plasma. The enantiomers were
separated using an a1-acid glycoprotein (AGP) column prior to tandem mass spectrometry. Ucb 20028 was used as internal
standard
QCH (3710 ng/mL) QCM (251 ng/mL) QCL (0.75 ng/mL) LLOQ (0.25 ng/mL)
S R S R S R S R
Intra-dayObserved conc.(ng/mL) 3690 3780 243 251 0.73 0.76 0.29 0.28CV (%) 3.4 2.3 2.7 0.7 7.1 3.7 10.8 12.7Accuracy (%) �0.6 1.9 �3.1 0.0 �3.0 0.7 15.0 12.4n 6 6 6 6 6 6 6 6Inter-dayObserved conc.(ng/mL) 3730 3760 248 245 0.76 0.77CV (%) 3.95 3.57 6.61 6.19 11.9 12.6Accuracy (%) 0.51 1.28 �0.82 �2.31 0.88 1.82n 12 12 12 12 12 12
Table 3. Intra- and inter-day precision and accuracy of cetirizine enantiomers in Ringer’s solution with bovine serum albumin
(BSA). The enantiomers were separated using an a1-acid glycoprotein (AGP) column prior to tandem mass spectrometry.
Ucb 20028 was used as internal standard
QCH (37.8 ng/mL) QCM (7.56 ng/mL) QCL (0.38 ng/mL) LLOQ (0.25 ng/mL)
S R S R S R S R
Intra-dayObserved conc.(ng/mL) 37.3 37.6 7.50 7.57 0.39 0.36 0.26 0.25CV (%) 1.4 1.0 2.3 1.7 6.5 6.0 13.3 10.0Accuracy (%) �1.4 �0.6 �0.8 0.1 3.6 �5.2 3.8 �1.5n 6 6 6 6 6 6 6 6Inter-dayObserved conc.(ng/mL) 38.8 39.0 7.78 7.66 0.40 0.40CV (%) 3.9 3.8 2.1 2.5 10.0 9.0Accuracy (%) 2.6 1.9 2.9 1.4 5.9 5.1n 12 12 12 12 12 12
1756 A. Gupta et al.
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757
![Page 9: Quantitative determination of cetirizine enantiomers in guinea pig plasma, brain tissue and microdialysis samples using liquid chromatography/tandem mass spectrometry](https://reader035.fdocuments.us/reader035/viewer/2022081122/5750256f1a28ab877eb3cc5f/html5/thumbnails/9.jpg)
The LLOQ for the CZE enantiomers for all the methods
was set to the lowest concentration of the calibration curve
ranges, since the precision and accuracy were sufficient to
fulfill the acceptance criteria (Tables 1–3). The CV, deter-
mined from six replicate samples, was below 13%, and the
accuracy varied by less than �15% of the nominal value for
both enantiomers in all matrices.
CONCLUSIONS
Chiral LC/MS/MS (ESI) methods to determine S- and R-CZE
in guinea pig plasma, brain tissue and microdialysis samples
were developed and validated. The LLOQ was 0.25 ng/mL
for the CZE enantiomers using only 50 mL plasma samples.
The LLOQ for the microdialysis samples was also 0.25 ng/
mL using a sample volume of approximately 8 mL. Because
of a high dilution factor, the LLOQ for the brain tissue sample
was 2.5 ng/g brain. The run time was no more than 14 min for
any of the methods. These validated methods were success-
fully applied to a preclinical study investigating the BBB
transport of CZE enantiomers in the guinea pig.
REFERENCES
1. Devalia JL, De Vos C, Hanotte F, Baltes E. Allergy 2001; 56:50. DOI:10.1034/j.1398-9995.2001.00726.x.
2. Chen C, Hanson E, Watson JW, Lee JS. Drug Metab. Dispos.2003; 31: 312.
3. Polli JW, Baughman TM, Humphreys JE, Jordan KH, MoteAL, Salisbury JA, Tippin TK, Serabjit-Singh CJ. J. Pharm. Sci.2003; 92: 2082. DOI: 10.1002/jps.10453.
4. Hammarlund-Udenaes M. Adv. Drug Deliv. Rev. 2000; 45:283. DOI:10.1016/S0169-409X(00)00109-5.
5. Hill SJ, Young JM. Br. J. Pharmacol. 1980; 68: 687.6. Pagliara A, Testa B, Carrupt PA, Jolliet P, Morin C, Morin D,
Urien S, Tillement JP, Rihoux JP. J. Med. Chem. 1998; 41: 853.DOI: 10.1021/jm9704311.
7. Bakhtiar R, Ramos L, Tse FL. Chirality 2001; 13: 63. DOI:10.1002/1520-636X(2001)13:2<63::AID-CHIR1000>3.0.CO;2-5.
8. de Jager AD, Hundt HK, Swart KJ, Hundt AF, Els J. J.Chromatogr. B Anal. Technol. Biomed. Life Sci. 2002; 773: 113.DOI:10.1016/S1570-0232(02)00134-4.
9. Eriksen H, Houghton R, Green R, Scarth J. Chromatographia2002; 55: S-145.
10. Li AC, Junga H, Shou WZ, Bryant MS, Jiang XY, NaidongW. Rapid Commun. Mass Spectrom. 2004; 18: 2343. DOI:10.1002/rcm.1623.
11. Macek J, Ptacek P, Klima J. J. Chromatogr. B Biomed. Sci. Appl.1999; 736: 231. DOI: 10.1016/S0378-4347(99)00464-8.
12. Moncrieff J. J. Chromatogr. 1992; 583: 128.13. Song Q, Junga H, Tang Y, Li AC, Addison T, McCort-Tipton
M, Beato B, Naidong W. J. Chromatogr. B Anal. Technol.Biomed. Life Sci. 2005; 814: 105. DOI:10.1016/j.jchromb.2004.10.009.
14. Zaater MF, Tahboub YR, Najib NM. J. Pharm. Biomed. Anal.2000; 22: 739. DOI: 10.1016/S0731-7085(99)00294-0.
15. Choi SO, Lee SH, Kong HS, Kim EJ, Choo HY. J. Chromatogr.B Biomed. Sci. Appl. 2000; 744: 201. DOI: 10.1016/S0378-4347(00)00229-2.
16. Choi SO, Lee SH, Kong HS, Kim EJ, Choo HY. Arch. Pharm.Res. 2000; 23: 178.
17. Instruction Manual Chiral-AGP, Chrom Tech AB.
Cetirizine enantiomers in guinea pig plasma and brain ISF and homogenate 1757
Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 1749–1757