QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE GOLD NANOPARTICLES AND CHROMIUM 51 SEED IMPLANTS A Dissertation presented to the Faculty of the Graduate School at the University of Missouri In Partial Fulfillment Of the Requirements for the Degree Doctor of Philosophy By RAJESH R KULKARNI Dr Kattesh V. Katti and Dr Jimmy C. Lattimer, Dissertation Supervisors DECEMBER 2012

Transcript of QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

Page 1: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

QUANTIFICATION OF ABSORBED DOSES FROM

RADIOACTIVE GOLD NANOPARTICLES AND

CHROMIUM 51 SEED IMPLANTS

A Dissertation presented to the Faculty of the Graduate School at the University of Missouri

In Partial Fulfillment

Of the Requirements for the Degree

Doctor of Philosophy

By

RAJESH R KULKARNI

Dr Kattesh V. Katti and Dr Jimmy C. Lattimer,

Dissertation Supervisors

DECEMBER 2012

Page 2: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

© Copyright by Rajesh R Kulkarni 2012

All Rights Reserved

Page 3: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

The undersigned, appointed by the dean of the Graduate School, have examined the DISSERTATION entitled

QUANTIFICATION OF ABSORBED DOSES OF RADIOACTIVE GOLD NANOPARTICLES AND CHROMIUM 51 SEED IMPLANTS

Presented by Rajesh R Kulkarni

A candidate for the degree of DOCTOR OF PHILOSOPHY

And hereby certify that, in their opinion, it is worthy of acceptance.

Dr Kattesh V Katti

Dr Jimmy C Lattimer

Dr Sudarshan K Loyalka

Dr William H Miller

Dr Tushar K Ghosh

Dr Steven J Westgate

Page 4: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

To my wife Indira and precious daughters, Bhakti and Annika……………

Page 5: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

ii

ACKNOWLEDGEMENTS

I would like to take this opportunity to thank and appreciate all of those who have been

a part of my PhD research work, without their support it would be impossible for me to

complete my work.

I would like to start with Dr Kattesh Katti, who has been supportive of my graduate

studies all times at MU. I would like to express my gratitude for his patience, endurance

and trust that has always kept me motivated in my academic research. I would like to

thank Dr Jimmy Lattimer, who has been supportive and a good mentor in radiation

science applications. Working with Dr Lattimer, has laid a solid foundation in my medical

physics career. I am thankful to Dr. Sudarshan Loyalka for his support and advice. I

would like to thank Dr William Miller (Bill Miller), who has been constantly helping me

with my dosimetry calculations. I would like to appreciate his patience in replying my

long trail of emails. He has been a good teacher and a good mentor throughout my

graduate program. I would like this opportunity to thank Dr Tushar Ghosh, who has

been advising me on academic course work for both my Masters and PhD degrees. I

have worked with Dr Steven Crooks in the year of 2008, who is board certified medical

physicist in radiation oncology physics. His advises helped me outline my research plan

in radiation therapy and identify Cr51 as a potential high dose brachytherapy agent. I

would like to thank Dr Steven Westgate, for his acceptance to serve in my research

committee. Further, I wish to thank Dr Evan Boote for his advices on medical physics

Page 6: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

iii

diagnostics measures. I would like to thank Dr Cathy Cutler for sharing information on

neutron activation of gold to Au198, or Au199 isotopes. I must not forget to thank for

the invaluable technical help from Jan Kunkel, Jeff March and James Avenille of

Veterinary Medicine Teaching Hospital. Also I would like to thank all my colleagues and

co workers, Kavita katti, Dr. Kattumuri, Dr. Kannan, Dr. Upendran, Dr. Chanda, and Dr.

Shukla, from the department of Radiology in the Alton Building who have shared

information in nanoparticles synthesis and characterization techniques. I would like to

thank Dr Mark Milanick who mentored me in ‘Clinical Biodetectives’ project grant

whose specific objective involved identifying technical difficulties and to provide

solutions to clinical treatment procedures in day to day life. I believe that quality of life

can be significantly improved with better understanding of the technology and there by

its continuous improvements.

Last but not the least, I would like to thank my wife Indira, my precious daughters Bhakti

and Annika who kept faith in me, and endured patiently in achieving this important

milestone of PhD degree in my academic career. Also I would like to thank my parents

and my parents in law who kept their patience throughout my graduate studies.

Page 7: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

iv

TABLE OF CONTENTS

ACKNOWLEDGEMENTS ……………………………………………………………………………………………..…ii

TABLE OF CONTENTS…………………………………………………………………………………....................iv

LIST OF FIGURES……………………………………………………………………………………..…………………….vi

LIST OF TABLES……………………………………………………………………………………………………………..ix

NOMENCLATURE……………………………………………………………………………………………………………x

ABSTRACT……………………………………………………………………………………………….…………………….xi

CHAPTERS

CHAPTER 1. INTRODUCTION ............................................................................................. 1

CHAPTER 2. BACKGROUND ............................................................................................. 20

CHAPTER 3. MATERIALS AND METHOD .......................................................................... 44

CHAPTER 4. QUANTIFICATION OF RADIOACTIVE GOLD NANOPARTICLES DOSE ........... 63

CHAPTER 5. QUANTIFICATION OF CR-51 DOSE .............................................................. 90

APPENDIX…………………………………………………………………………………………………………………….93

BIBLIOGRAPHY .................................................................................................................. 97

Page 8: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

v

VITA ................................................................................................................................. 103

PUBLICATIONS……………………………………………………………………………….…………………………..104

Page 9: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

vi

LIST OF FIGURES

Figure 1 – Transmission Electron Microscope image 20,000 times magnified in JEOL 1400 microscope. Tea leaves synthesized gold nanoparticles are internalized in PC3 cells.

Figure 2 – Therapeutic efficacy studies of Au-198 nanoparticles in prostate tumor bearing SCID mice. SCID mice bearing human prostate cancer lines were treated with a single dose of 30µl of GA-198AuNP (408 µCi) into the tumor. Tumor volume in cm3 was measured every 3 day for 30 days.

Figure 3 – Gel dosimetry schematics for radioactive gold nanoparticles of Au-198 and Au-199.

Figure 4 - Coordinate system used for brachytherapy dosimetry calculations.

Figure 5 – Radiation Treatment Plan of gel dosimeter on CT acquired image with XiO software (Ver.4.6). Dosimeter was located at 100 mm SAD in an acrylic water tank of size 1 cubic foot.

Figure 6 - The dosimeters treated with Siemens 8X Linac.

Figure 7 - Phantom set of acrylic construction of beagle dog’s head representing pelvic bone and Play-Doh ball representing prostate gland wrapped with adhesive tape. Left picture-Play Doh ball closed inside the acrylic construction. Right picture

Figure 8 -The MCNP model for radioactive nanoparticles.

– Inside view with one side of the acrylic construction removed.

Figure 9- The MCNP model for Cr-51 brachytherapy seed. The dark blue region or cell 4 in MCNP5 is Chromium material. The violet regions or cells 1, 2, and 3 are representing material for encapsulation of stainless steel. The cells 5, 6, 7, etc. are for different tallies that follows the AAPM TG43 protocol.

Figure 10 The beta energy spectrum of Au-198 isotope. The maximum beta energy is 1372.5 keV.

Figure 11 The beta energy spectrum of Au-199 isotope. The maximum beta energy is 452.3 keV.

Figure 12 Range of beta particles energy in inches within various media.

Figure 13 – Gel dosimeter consistency response. Dosimeters made in separate batches were scanned in GE 1.5T MR scanner with T2 Transverse pulse sequence at various TE,

Page 10: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

vii

echo times of 15, 30, 45 and 60 milliseconds. T2, spin-spin relaxation time in milliseconds was the response collected against the all the TE’s.

Figure 14 - (Top) this is an image of 3D localizer, wherein MR scanner by default will scan briefly sample in Transverse/Sagittal orientations, which further helps to locate sample and choose the area of interest for full scan. (Bottom) MR Images obtained with 30 Gy dose at TE=14.5 ms, TE=33.9 ms, TE=53.3 and TE=63 ms, TR=5000 ms, slice thickness = 5mm, Transverse T2 weighted imaging. Two large regions around the dosimeter were of saline bags, which were used for homogeneous field response.

Figure 15 – (Top) this is an image of 3D localizer, wherein MR scanner by default will scan briefly sample in Transverse/Sagittal orientations, which further helps to locate sample and choose the area of interest for full scan. (Bottom) MR Images obtained with 70 Gy dose at TE=14.5 ms, TE=33.9 ms, TE=53.3 and TE=63 ms, TR=5000 ms, slice thk= 5mm, Transverse T2 weighted imaging.

Figure 16 – Normalized T2 relaxation time as function of echo time. The dosimeters treated with lower doses in the range, 0-40 Gy show linear response. With increase of dose >40 Gy, the curve shows exponential responses. The results are shown in mean T2 relaxation time ± standard error. The standard deviation is more to the dosimeters treated with doses >70 Gy.

Figure 17 – Dose Calibration Curve.

Figure 18 – Radioactive nanoparticles injected in gel dosimeter. The red spot in the center is nanoparticles spread volume, was approximately about 0.5 cm in diameter.

Figure 19 – MR Image in Sagittal section with T2 weighted imaging with FRFSE pulse sequence. The dosimeter, cylindrical shaped object with apex was held between saline bags for better field response. The region marked inside the circle is the dose response from radioactive nanoparticles (Au-198). The dose response in region R is due to particle backflow during injection.

Figure 20- T2 relaxation time plots of dosimeters at different echo times of 0 Gy, 70 Gy and unknown dose (Au-198 nanoparticles). From these plots it can be deduced that the radiation absorbed dose from Au-198 nanoparticles is >70 Gy. However it was difficult to measure that dose precisely beyond the threshold limit ~ 70 Gy of gel dosimeter.

Figure 21- SPECT image of Au198 in the gel dosimeter.

Figure 22 – Sinogram for Au-198 nanoparticles. (Left) The Sinogram for the full data set. (Right) 16 sets of Sinogram, each representing set of 4 projections.

Figure 23 – Dose Rate of Au198 photon transport that includes other secondary particles born with photon interactions.

Page 11: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

viii

Figure 24 – Dose Rate of Au198 electron transport. The homogeneous volumetric source of 0.316 cm radius was used. The dose rates shown in the plot are measured from the center of the source sphere formed with nanoparticles diffusion.

Figure 25– Photograph taken after 1 day of radioactive nanoparticles of Au-199 injected in gel dosimeter.

Figure 26 – MR Image in Sagittal section with T2 weighted imaging with FRFSE pulse sequence. The dosimeter, cylindrical shaped object with apex was held between saline bags for better field response. The region marked inside the circle is the dose response from radioactive nanoparticles (Au-199). The dose response in region R is due to particle backflow during injection.

Figure 27 - T2 relaxation time plots of dosimeters at different echo times of 0 Gy, 70 Gy and unknown dose (Au-199 nanoparticles). From these plots it can be deduced that the radiation absorbed dose from Au-199 nanoparticles is >70 Gy. However it was difficult to measure that dose precisely beyond the threshold limit ~ 70 Gy of gel dosimeter.

Figure 28 – SPECT image of Au-199 in the gel dosimeter.

Figure 29 – Sinogram for Au-199 nanoparticles.

Figure 30 – Dose rate profile of Au199 photon transport. The homogeneous volumetric source of 10.7 mm radius was used. The dose rates shown in the plot are measured from the center of that source in regular intervals through the periphery of the source.

Figure 31 – Dose rate profile of Au199 electron transport. The homogeneous volumetric source of 10.7 mm radius was used. The dose rates shown in the plot are measured from the center of that source in regular intervals through the periphery of the source.

Figure 32 – Dose rate profile of Cr-51 photon transport.

Figure 33–CPS schematics of operating principles.

Figure 34 – Overlay of particulate distributions by CPS disc.

Page 12: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

ix

LIST OF TABLES

TABLE I – THE UNSHIELDED GAMMA RAY DOSE EQUIVALENT RATE AT 1 METER FROM

POINT SOURCE FOR CHROMIUM AND IRIDIUM NUCLIDES.

TABLE II - THE UNSHIELDED GAMMA RAY DOSE EQUIVALENT RATE AT 1 METER FROM

POINT SOURCE FOR GOLD NUCLIDES

TABLE III – GUIDELINES FOR INTERPRETING THE RELATIVE ERROR R

TABLE IV A&B – COMPOSITION AND MASS FRACTIONS OF GEL DOSIMETER USED IN

MCNP MODELING OF Au-198 AND Au-199 SOURCES.

TABLE V – DECAY PROPERTIES OF Au-198

TABLE VI - DECAY PROPERTIES OF Au-199

TABLE-VII, COMPARISON OF CPS DERIVED PARTICLE AVERAGE SIZES TO TEM DERIVED

PARTICLE AVERAGE SIZES OF DIFFERENT TYPES OF THE GOLD NANOPARTICLES IN

NANOMETERS

Page 13: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

x

NOMENCLATURE

AuNP – Gold nanoparticles

Ci– Curie

CT– Computer Tomography

DNA – Deoxyribonucleic acid

FSE – Fast Spin Echo

GA-AuNP – Gold Nanoparticles suspended in Gum Arabic matrix

HDR – High Dose Rate

LDR – Low Dose Rate

LEAP – Low Energy All Purpose

LINAC – Linear Accelerator

MCNP – Monte Carlo Neutral Particle

MCNP5 – Monte Carlo Neutral Particle version 5

MR – Magnetic Resonance

MU – University of Missouri

MURR – Missouri University Research Reactor

RNA – Ribonucleic acid

SCID – Severe combined immunodeficiency

SPECT– Single Photon Emission Computer Tomography

THPAL – Phosphine Functionalized trimeric alanine (a chemical compound)

VMTH – Veterinary Medicine Teaching Hospital

XiO – Radiation Treatment Planning System

Page 14: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

xi

ABSTRACT

Purpose: The purpose of this work is 2 fold, firstly to estimate that gold nanoparticles of

Au-198 and Au-199 which emit both gammas and betas can be used in brachytherapy,

and secondly the chromium isotope of Cr-51 as brachy seed can be used in

brachytherapy. This research work elucidates the absorbed doses and radioactivity

distribution by Au-198 and Au-199 nanoparticles experimentally and estimates the

absorbed doses and dose rates by Monte Carlo simulations. Whereas, Cr-51 with

complete beta shielded is estimated for dose rates and corresponding doses by Monte

Carlo Simulations.

I. Radioactive gold nanoparticles: A polymer gel dosimeter was prepared using about

12% gelatin (Sigma Aldrich, 300 bloom type A), 9% methacrylic acid(Sigma Aldrich),

10mM THPC (Sigma Aldrich), 18mM hydroquinone (Sigma Aldrich) in deionized

water. These dosimeters were calibrated using 8X Siemens Liner Accelerator and GE

1.5T MR scanner. To calibrate for radioactivity of Au-198 and Au-199 SPECT imaging

and the balls of Play-Doh[

Materials and Methods:

1] made by Hasbro Toys[2] as phantoms were used. The

radioactive gold nanoparticles were synthesized at the University of Missouri

nuclear reactor. To simulate photons and electrons transport MCNP5 (Monte Carlo

Neutral Particle) code from Los Alamos National Lab was used.

Page 15: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

xii

II. Novel Cr-51 radionuclide: This nuclide has potential benefits to be considered in

brachytherapy applications for both temporary and permanent dose implants.

MCNP simulation in photon transport for Cr-51 was used to estimate the dose rates.

I. Radioactive gold nanoparticles: Radioactive gold nanoparticles with 800 µCi (micro

Curies) of Au-198 and 500 µCi of Au-199 were considered in separate studies. T2

weighted imaging with FRFSE pulse on 4 different intervals of echo times

demonstrates the absorbed dose response. The Gamma camera peaked at 411keV

photon with a high energy (Iodine collimator) demonstrates activity distribution of

Au-198 nanoparticles. Similarly, at 158 keV energy photons were peaked using LEAP

(Low energy all purpose) collimator to demonstrate activity distribution of Au-199

nanoparticles.

Results:

II. Cr-51 radionuclide: Through MCNP5 (MCNP version) simulation it is demonstrated

that 4 mCi of activity per seed is required to deliver at least 5 Rads/hr in water

sphere of radius 1.0 cm and mass of 4.2 g.

Page 16: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

1

CHAPTER 1

INTRODUCTION

American Cancer Society estimates that about 1,638,910 new cases of cancer are

expected to be diagnosed in 2012. About 577,190 cancer deaths were projected to

occur in the U.S., in 2012. The most predominant cancer deaths occur in respiratory,

digestive, and genital systems. There were about 164,770 deaths projected for the year

of 2012. However, since 1990 the reduction in overall cancer deaths across most of the

ethnic groups was due to cancer control knowledge and improved techniques. Four

major areas that contributed towards this decline are lung, colorectal, breast and

prostate[3]. Surgery and radiation therapy are the available treatment methods. The

research work presented here relates to physical measurements in radiation therapy for

cancer.

Radiation therapy deals with treatment procedures that involve high energy

photons that upon interactions with matter can ionize particles or scatter within

interacting media e.g. normal tissue or cancer tissue. Radiation therapy is a complex

phenomenon that may be applied with radiating energy that penetrates deeply into

tissue. Radiation therapy is applied based on the type of radiation, its source strength,

and its biological effectiveness. Radiation interactions can ionize the genetic structures

Page 17: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

2

of the living organism. This is designed to be lethal to cancer cells but can also cause

carcinogenesis, acute and late effects to normal tissue.

Cancer can be treated with radiation by two methods. The first method EBRT (External

Beam Radiation Therapy) involves the use of Linac (Linear Accelerator). The second

method is using permanently or temporary interstitial implants of radionuclides

encapsulated in inert metals, is called as brachytherapy. Radionuclides of short life lives

are generally considered as permanent implants (seeds left inside the body). Common

sources currently include Pd-103 and I-125. Radionuclides of long half lives usually Cs-

137 or Ir-192 are considered as temporary implants.

1.1 Brachytherapy

Brachytherapy is a Greek acronym for ‘therapy at short distance’. Brachytherapy can be

described as radiation therapy treatment using radioactive sources in proximity of the

tissue. Brachytherapy, also called internal radiation therapy, allows clinicians to use

particular dose of radiation to treat a small volume in a calculated time intervals.[4] In

brachytherapy, the radiation travels a short distance into the tissues, whereas with an

external beam source the radiation usually travels a long distance to reach its target

tissue. Depending on the position of the source with respect to the tissue,

brachytherapy is further classified into interstitial implant (in the tissue), intracavitary

implant (in a cavity), intraluminal implant (in a lumen), intraoperative implant (during

Page 18: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

3

surgery) and surface implant (over the tissue).[4, 5] Over nearly three decades,

brachytherapy has been used as treatment for metastatic prostate, breast, liver and

malignant brain tumors in both prospective clinical trials and as part of standard

therapy.

Sources are further categorized with respect to their dose rates. Clinically, there is a

range of reference of dose rates measured in Gray per min (Gy/min). Sources that

deliver radiation doses from 0.4 to 2 Gy/min are low dose rate (LDR). Dose rates

between 2 and 12 Gy/min are considered as medium dose rates (MDR). Sources that

deliver doses more than 12 Gy/min are high dose rates (HDR). Further there is a

specialized treatment of pulsed dose rate[6] (PDR), which is usually treated between 1

and 3 Gy/min for every 1 hour or until the prescribed dose is completely delivered to

the tissue[7].

Brachytherapy implants may be either temporary or permanent. In temporary

brachytherapy; the radioactive material is placed inside or near a tumor for a specific

amount of time and then withdrawn. Permanent brachytherapy involves placing

radioactive seeds or pellets (about the size of a grain of rice) in or near the tumor and

leaving them there permanently. After several weeks or months, the radioactivity level

of the implants eventually curtails to nothing. The seeds then remain in the body, with

no lasting effect on the patient. For permanent implants, the radioactive material

(which is enclosed within small seeds or pellets) is placed directly in the site of the

Page 19: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

4

tumor using a specialized delivery device. For temporary implants, needles, plastic

catheters or specialized applicators are placed in the treatment site. [8]

A radionuclide’s half life and its initial dose rate determine whether it can be used as

permanent implant or being used as a temporary implant. Some types of radiation

sources (isotopes) used in brachytherapy are: iodine-125, palladium-103, strontium-90,

cesium-137 and iridium-192. In all cases of brachytherapy radionuclides, the source of

radiation is encapsulated by an inert metal capsule. This encapsulation serves as a shield

for beta radiation. Mostly the radiation dose deposition is due to gamma radiation

alone. However, therapeutic doses are significantly deposited by photons (Gammas, X-

rays) and their secondary particles born on their surfaces due to the photonic

interactions.

Brachytherapy is a vast subject by itself and has a history of one hundred years.

However, brachytherapy is only a curative option to treat cancer[9]. Currently, cervix,

breast cancer and prostate cancer have shown good success rate.[10, 11] Besides, there

are few incidences where brachytherapy has not been successful in cervical carcinoma,

and also some of the prostate cancer cases. This indicates that more treatment specific

approaches are necessary depending upon the cancer location. The research work

presented in this thesis pertains to brachytherapy. In one section of the dissertation

there is an attempt to demonstrate the novel ways of treating cancer using radioactive

nanoparticles. Other section demonstrates about Cr-51 radionuclide as a brachytherapy

seed implant. Both of these sections do not have any inter dependency except their

Page 20: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

5

physical forms of source and their radiation effects for comparison. Radioactive

nanoparticles are of gold in Au-198 and Au-199 isotopes considered separately have

both effects of betas and gamma radiation on the medium of interest. Whereas, Cr-51

isotope is considered in a contemporary design for only gamma radiation to the medium

of interest shielding beta particles. However, the comparison with Cr-51 seed will

demonstrate the effects of beta radiation from unshielded gold nanoparticles in the

medium of interest. This work exhibits techniques for dose quantifications that have

references mostly prior to 20 years.[12-14].

1.2 Introduction to gold nanoparticles

Nanoparticles are particulates of material in nanometer scale. There is a very interesting

phenomenon that nanoparticles behave differently in their physical and chemical

properties as compared to their bulk size properties[15]. Today it is new emerging field

of science and with titles such as nano-science, nanotechnology or nano engineering so

on so forth. Nanoparticulates that are synthesized in aqueous solutions show significant

promise in medicinal applications. This research field is categorically known as

Nanomedicine. Nanoparticulates could be made of metallic or non metallic forms.

Advantages of using nanoparticles are,

a. Higher surface to volume ratio.

b. Internalization into cells and their structures.

Page 21: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

6

c. Biological clearance.

d. Charge effects.

e. Higher atomic number if they are metallic.

Gold has been the most preferred element of choices due to its non toxic

properties.[16] Syntheses of gold nanoparticles involve gold complexes or gold salts that

are commercially available. Further these gold complexes are to be chemically reduced

to gold nano particles by suitable chemical reactions. These types of chemical reactions

are usually performed in aqueous solution for particle suspension. The best solvent for

such type of chemical reactions is to use deionized water. In principle, all the metallic

elements are reduced due to charge effects and start to coagulate instantaneously. The

controlled coagulation process yields larger particles of different sizes. To prevent rapid

coagulation or slowdown particles coagulating process stabilizing agents like

carbohydrates or polyphenols are used. Gold nanoparticles are required to be

characterized for their physical appearances and sizes. Some of the characterizing

techniques available include a) UV-visible absorption spectroscopy, b) Transmission

Electron Microscopy, c) Particle Sedimentation Technique, d) Atomic Force Microscopy,

and e) Dynamic light scattering.

Well stabilized with good size distribution, the nanoparticles become useful in various

types of applications. The nanoparticles bear charges [17]and can conjugate or attach

with other biochemical ligand and become target specific to the histological sites of

Page 22: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

7

interest within the living organism. Two methods to synthesize gold nanoparticles

available were:

a. Green nanotechnology

Scheme: gold complex + phytochemicals + water

b. THPAL (amino acid analog) reduction method

Scheme: gold complex + THPAL + carbohydrates + water

Green nanotechnology involves synthesizing nanoparticles with natural phytochemicals

from edible materials. On one hand, this method of synthesis can be a faster and may

involve fewer steps in the drug development pathways as there are no toxic side effects

[18-22]. On the other hand, these nanoparticles have certain limitations with increase of

gold concentrations, yield of uniform size distributions, etc. Similarly, THPAL reduction

method can synthesize gold nanoparticles with good stability and uniform size

distributions. This method was preferred a stable method to increase gold

concentration. These two properties help in target specificity and diagnostic abilities

when used in treatments. THPAL is a synthetic peptide, with amino acid functionalities;

this peptide has potential to synthesize gold nanoparticles in controlled size

distribution[23] and also it aids to increase the gold concentration[24] or content of gold

in gold nanoparticulate solutions.

Page 23: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

8

1.3 Motivation

The search for better therapeutic ratio will always be there until the last possible normal

cell can be spared when eradiating cancer tissue. Gold nanoparticles, particularly

radioactive gold nanoparticles, have shown promising responses when applied in

therapeutics [25-27]. Radioactive gold nanoparticles of Au198 and Au199 could be

considered individually for brachytherapy applications.

Ir-192 is extensively used radionuclide in brachytherapy. It is usually activated to about

12 Ci/g and contained in the radiation protected containers in the machine used for

brachytherapy treatment. This machine is called remote after loader. This is computer

controlled machine and the source (of Ir-192) is briefly introduced into the patients

through catheters and suitable applicators. Ir-192 has half life of 74 days. It emits both

gammas with average energy of 380 keV in 100% photon abundance and also betas up

to the maximum energy of 675 keV. One isotope of chromium element, Cr-51 has half

life of 27.7 days. Cr-51 emits both gammas with average energy of 320 keV in 10%

photon abundance and also betas up to maximum energy of 753 keV. Fundamentally Ir-

192 and Cr-51 are comparative for brachytherapy type of applications. Unlike Ir-192, it

can be used with any desired dose rates. However, it may be suitable as a permanent

LDR implant. Also Cr-51 seed could be cost effective as it has lesser radiation safety

issues in transporting to remote areas as compared to Ir-192. Few characteristics of

interest for both Ir-192 and Cr-51 with reference to radiation protection are shown in

Table 1. As first step of proposal for use of Cr-51 in the contemporary brachy seed

Page 24: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

9

designs[28] Monte Carlo Simulations to calculate the dose rates have been presented in

this dissertation report.

TABLE I – THE UNSHIELDED GAMMA RAY DOSE EQUIVALENT RATE AT 1 METER FROM

POINT SOURCE FOR CHROMIUM AND IRIDIUM NUCLIDES [29]

Nuclide Energy

keV

Prob.

%

Half Life

days

Γ

(mSv/h)/MBq

t

cm

µ

cm-1

Ir-192

296

308

316

468

29

30

82.8

48

74 1.597×10-4 1.284 2.334

Cr-51 320 10 27.7 6.32×10-6 0.782 3.833

Γ- Specific gamma ray dose constant, which is unshielded gamma ray dose equivalent

rate at 1 meter from a point source.

t= thickness of lead to attenuate gamma rays by 95%.

µ=linear attenuation coefficient in lead to stop gamma rays by 95%.

Page 25: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

10

1.3.1 Gold nanoparticles

Gold nanoparticles due to their sizes can pass through the cellular membrane or other

pathways. Cellular internalization studies of gold nanoparticles provide insights with

cellular uptake and this may enhance the scope of gold nanoparticles applications in

biomedicine. Gold nanoparticles are currently investigated for their potential

applications as diagnostic/therapeutic agents, therapeutic delivery vectors, and

intracellular imaging agents.[30, 31] Selective cell and its nuclear matrix targeting by

gold nanoparticles will provide new pathways for the site specific delivery of gold

nanoparticles as diagnostic/therapeutic agents. A number of studies have demonstrated

that phytochemicals in tea have the ability to penetrate the cell membrane and

internalize within the cellular matrix.[32, 33] Cancer cells are highly metabolic and

porous in nature and are known to internalize solutes rapidly compared to normal

cells.[34, 35] Therefore, it is hypothesized that gold nanoparticles, would show

internalization within cancer cells. Cellular interaction and their uptake via incubation of

aliquots of tea reduced gold nanoparticles with prostate (PC-3) cells in vitro studies were

performed. As a result, significant internalization of gold nanoparticles was observed in

PC-3 cells through electron microscopy (Figure 1). The internalization of nanoparticles

within cellular structures is probabilistic and it could have been via phagocytosis, fluid-

phase endocytosis, and receptor-mediated endocytosis. Gold nanoparticles presence in

cellular structure in closer vicinity to genetic material such as DNA, RNA poses a greater

advantage. In principle, if the nanoparticles are energized or made radioactive with

Page 26: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

11

average energy of 200 keV at least in beta particle emission then the cell death is

certain.

Figure 1 – Transmission Electron Microscope image 20,000 times magnified in JEOL 1400 microscope. Tea leaves synthesized gold nanoparticles are internalized in PC3 cells.

1.3.2 Radioactive gold nanoparticles

The motivation to this work comes from in vivo pre-clinical studies performed on mice

at VMTH of the University of Missouri. The therapeutic efficacy of radioactive gold

nanoparticles (198AuNP) was investigated by intratumoral administration of GA-198AuNP

( Gum Arabic suspended radioactive gold nanoparticles) in prostate tumor bearing mice

Page 27: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

12

models.[36] Figure 2 shows the results from the single dose radiotherapy study of GA-

198AuNP in human prostate cancer bearing SCID mice.[37]

Figure 2 – Therapeutic efficacy studies of Au-198 nanoparticles in prostate tumor bearing SCID mice. SCID mice bearing human prostate cancer lines were treated with a single dose of 30µl of GA-198AuNP (408 µCi) into the tumor. Tumor volume in cm3 was measured every 3 day for 30 days.

Unilateral solid tumors were allowed to grow three weeks, and animals were

randomized (denoted Day 0) into control and treatment groups (n = 7) with no

significant differences in tumor volume. On Day 8, 30 μL of GA-198AuNP (408 μCi) was

injected directly into the tumor to deliver an estimated dose of 70Gy. These dose

estimations were based on few rudimentary assumptions like amount of human

prostate cancer radioactivity injected in Curies of activity and confirmatory test of

Page 28: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

13

radioactivity presence was recorded using SPECT camera. However, there was no tool to

quantify the radiation doses. The volume of the injected material contained the mixed

species of radioactive and non radioactive gold nanoparticles. This mixture complicated

to pursue quantification of radiation doses from gold nano particulates and its dose

distribution within the tissue. To overcome these difficulties special techniques of

dosimetry were necessary [38, 39] with radioactive gold nanoparticles. At this stage, it

was imperative to find some tool that would elevate the dose quantification method to

another level. In that perspective, the radiation absorbed dose and its distribution from

radioactive gold nanoparticles gel dosimetry was found as the best available choice. This

method also has ability to quantify dose volumetrically.

1.3.3 Chromium 51 seed

Cr-51 isotope behaves in penetration and dose characteristics similar to Ir-192 isotope.

In this perspective, Cr-51 could be used as a LDR implant. Cr-51 isotope can be produced

in the nuclear reactor with (n, γ) reaction. MURR has neutron flux of 8×1013

neutrons/cm2/sec and is accessible to synthesize Cr-51 isotope. In principle, Cr-50 would

be the target isotope and it has 3.3 barns of neutron absorption cross section. It would

need approximately about 64 hrs of irradiation time in the reflector region of the

reactor to get activity of 4 mCi with about 18 mg of Cr-50. Cr-51 can be produced with

low cost materials such as chromium wire, chromium containing alloys such as nickel

chrome, and chrome coated titanium wires. On contrary, iridium targets are rare and it

Page 29: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

14

involves expensive enrichment processes. Another note to remember is expensive costs

of handling and shipping of Ir-192 isotope.

1.4 Rationale

The objectives of this research work are to provide dosimetric measures to new

methods of treatments proposed in brachytherapy. There are 3 nuclides presented

here. To avoid confusion, the readers should make a note that there is no

interdependence between radioactive gold nanoparticles and the chromium seed.

They are categorically separated as radioactive gold nanoparticles and chromium seed

implants. Radioactive nanoparticles of gold have never been used in the past and there

are no defined mechanisms to understand the dosimetry of these nanoparticles for

therapy applications. On one hand, attempt has been made to understand and

estimate the dose deposition and dose distributions of the radioactive gold

nanoparticles. The radioactive nanoparticles of gold (Au-198 or Au-199), if delivered in

the target site, may have uniform dose distribution across the site and potentially

overcome the anisotropy factors measured in conventional brachytherapy seeds. On

the other hand, Cr-51 nuclide is considered as brachytherapy seed in conventional

brachy seed form, a radioactive material encapsulated within inert metal e.g., stainless

steel, in dimensions of 1.31mm diameter and 1cm length. The brachy seed is in

cylindrical form and its dose distribution can be quantified based on several dose

Page 30: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

15

factors. Dose distribution can be laid out or mapped depending upon the dose rate.

AAPM (American Association of Physicists in Medicine) recommends a formalism that is

based on computer simulated values and it is routinely used by all the manufacturers in

developing the brachytherapy implants for both temporary and permanent.

1.5. Radiation dose quantifications

Ion chambers, TLD’s and X-rays films are commonly used for dose measurements in

clinical setting. However, these are limited to one or two dimensional dose

measurement and in retrospect measurements are estimated. Such measurements have

uncertainties regarding the accuracy of measurements. Whereas, the three dimensional

dosimetry provides more reliable data that help significantly in relative dosimetry

analysis. The demand for a tool that can quantify doses in three dimensions has been

increasing as more advanced radiotherapy techniques continue to grow. For case of

radioactive nanoparticulates, polymer gel dosimeters will be used to quantify absorbed

doses of radiation. These dosimeters are gaining preferences for dose distributions from

a three dimensional perspective. In this approach polymer gel dosimetry[40, 41] has

characteristic response to the NMR (Nuclear Magnetic Resonance) properties of spin-

spin relaxation time, also called as T2 relaxation time. This type of dosimeters can retain

therapeutic dose levels and also they have tissue equivalent properties in radiation

interactions. In principles, within gel dosimeters, radiation induced free radicals initiate

Page 31: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

16

polymerization processes from aqueous monomers.[42] This polymerization bears

directly certain degree of dependency on the energy of the radiation and also on type of

particle interactions. Chemically induced polymerization has some relationship to the

absorbed doses in the dosimeter. This response can be easily measured through spin-

spin relaxation time (T2 relaxation) using magnetic resonance imaging scanner.

The dosimeter consists of monomers suspended in gelatin matrix (semi solid form) as

shown in Figure 3, which resembles a glue stick, in physical appearance. Any radiation

exposure to these dosimeters induces chemical reaction, causing monomers to fuse into

polymers. This polymerization process due to the radiation interactions directly relates

to the amount of dose deposited in that region.

The spin-spin relaxation time (T2 relaxation time) of the nuclear magnetic momentum of

the dosimeter materials shifts in a way that directly relates to the amount of absorbed

doses in that specific region. This response or time shift can be measured in MR scanner

with appropriate pulse sequence for e.g. FSE (Fast Spin Echo) pulse sequence.

Page 32: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

17

Figure 3 – Gel dosimetry schematics for radioactive gold nanoparticles of Au-198 and Au-199.

The dose quantifying schema used to quantify the absorbed doses deposited by

radioactive gold nanoparticles is shown in Figure 3. Initially, the gel dosimeter was

calibrated using some known doses. After determining the dose response though T2

weighted relaxation times, the radioactive nanoparticles were injected in known

amounts (in Curies) of activity. After certain interval of time, the dosimeters were

scanned in MR scanner to record T2 time shift response in the dosimeter due to the

interactions of radiation.

Page 33: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

18

1.6. Organization of thesis

The research work presented here is about the radiation therapy physics, specifically

applied in brachytherapy. There are 3 radioactive sources with 2 different modus

operandi. Two isotopes of gold (Au-198 and Au199) in the nanoparticulates form and

one isotope of chromium (Cr-51) in the contemporary seed form are considered. Gold

and chromium studies are presented in separate sections because they are intended for

different applications.

This dissertation is organized further in separate chapters as follows:

Chapter 2 - Background, provides the information about current medical procedures in

brachytherapy and also presents the reviews about gel dosimetry preferences and its

applications as suggested by other researchers.

Chapter 3 – Materials and Methods provides the information about the materials used

in this research work and also defines about the methods applied to quantify radiation

dose and its distributions.

Chapter 4 – Quantification of radioactive gold nanoparticles dose provides the details

about the results measured obtained with radioactive gold nanoparticles. Both Au198

and Au199 are discussed in separate sections and there is not intend of comparison

between them. The results obtained with experimental methods along with that

obtained through Monte Carlo Simulations are discussed in detail for gammas and betas

of both Au198 and Au199.

Page 34: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

19

Chapter 5 –Quantification of Cr-51 dose provides the details about the results of

chromium-51 isotope measured with Monte Carlo Simulations. Cr-51 isotope brachy

seed is considered encapsulated within a different material that shields beta emission

but high energy gammas. As a first step process only the MCNP calculated values are

discussed in this section.

Page 35: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

20

CHAPTER 2

BACKGROUND

2.1 Current medical procedures in brachytherapy

In this section some of the tools used in the current treatment procedures in

brachytherapy are discussed. The purpose of this information is to highlight the

diversified range in techniques applied in the brachytherapy treatments. There are at

least about 18 different anatomic sites treated with brachytherapy using various

approaches however it is gaining preferences to treat cancers of cervix, prostate, and

breast. The choice of applying brachytherapy is generally decided with reference to the

anatomic site and ease of accessibility by the ability of the site of interest to

accommodate the radioactive source. The source placement could either be temporary

or permanent. One of the main accessories used to transport or place the radioactive

sources is called an applicator. Applicators are chosen depending upon anatomic site, by

type of implant (permanent, temporary) etc. The details of some applicators used for

permanent and temporary implants are discussed in the following sections.

2.1.1 Permanent implants

Permanent implants are small size pellets or seeds of cylindrical shape of sizes of 0.5

mm in diameter and 5 mm length with rounded ends.[4] The size is compatible to

Page 36: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

21

administer with implant needles in the selective regions of body tissues[43]. Most of

such procedures are categorically placed as surgical and can be painful to the patients.

Using radioactive nanoparticles could be much easier and less painful compared to the

contemporary procedures. In permanent seed brachytherapy, clinicians implant these

seeds at a short distance to cancerous tissue within the organ of interest (for e.g.

prostate gland). Few applicators used in clinics are mentioned in the following

paragraphs to highlight their treatment specificity:

Stainless steel needle: One of the types of applicators that is available in 20 cm length;

also in sizes of 17 or 18 gauge diameters. This applicator is used with pre-loaded seeds

and spacers in prostate implants.[44]

Quick Seeder: This applicator holds 18 gauge sized needles and it accepts pre-loaded

cartridges with seeds and spacers in a polysulfone core. This type of applicator is used in

prostate implant similar to the stainless steel needle mentioned earlier.[44]

Seed guns: This type of applicator also called as seed gun holds the preloaded magazine

cartridge that accepts multiple seeds. After loading, the cartridge is attached to the gun

applicators. Some of the specific designs can eject up to 15 seeds per magazine one at a

time. These applicators are widely used in intraoperative lung implants, prostate, and

other clinically qualified sites.[5]

Absorbable suture: This type of applicator holds seeds unlike other methods mentioned

earlier. Seeds are actually spaced in suture material which later gets absorbed or

Page 37: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

22

dissolves leaving seeds in desired spatial pattern. This is biocompatible method that aids

in brachytherapy but has other limitations like seed migration or misalignments etc.

2.1.2 Temporary implants

Temporary implants are briefly placed in the target locations either manually or with the

aid of loading devices that are computer controlled. Broadly temporary implants are

categorized as interstitial applicators and intracavitary applicators.

2.1.2.1 Interstitial applicators

Stainless steel needle: These needles are usually available in 10, 15 and 20 cm lengths;

the needle sizes are approximately from 17 gauges to 14 gauges for dispensing the

brachytherapy seeds. The choice for the size used is relative to the external diameter of

the brachy seed or pellet. These types of needle applicators are used to treat for breast,

base of tongue and rectal implants.

2.1.2.2 Intracavitary applicators

Intracavitary radioactive brachytherapy means placement of radioactive materials using

an applicator into a body cavity. The radiation is exposed briefly through an enclosure to

the pre defined location for a calculated amount of time (called as dwell time). Only

clinically accepted radioactive material (Ir-192 or Sr-90 or Cs-137) is used in the

treatment procedures. The dwell time depends directly upon the on the dose rate and

the air kerma rate of the isotope applied in the treatments. Some of the applicators

used in this type of application are known by generic names as follows:

Page 38: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

23

Heyman capsule: Heyman capsules have long thin hollow plastic cylinders which have a

braided capsule [45, 46]. The lumen of the plastic tube is 1 mm in diameter and extends

into the capsule. A thin removable non-radioactive steel wire in the plastic tubing and

capsule provides structural rigidity for easy insertion of the capsule into the body cavity

for e.g. uterus. The stiffness of the inert steel wire within the plastic tube makes it

possible to apply enough force on the capsule to distend the uterine cavity. Radium-226

isotope was the most preferred radioactive source in these types of classic applicators.

Simon Capsule: These capsules are modified form of Heyman capsule[47]. The

biocompatible plastic is used for capsules instead of braided wire. Currently, long thin

flexible tubes are available with different capsule heads (e.g. 4/6/8 mm diameter). By

individual packing with such capsules, the application can be adapted to the pathologic

anatomy of the uterine cavity independently[48]. These are compatible with the

computer controlled loading devices as well. Small radiation sources of Cs-137 or Ir-192

are used in these types of capsules.

Fletcher suit tandem: This brachy applicator has ovoid and tandem. These are two main

parts in these types of applicators. This applicator is used in the gynecological sites.

Tandem is inserted into the uterus; ovoid is inserted into the vaginal vault. Typically,

ovoids are 2 cm diameter by 3 cm height with immovable tungsten shields. They are less

bulky and lighter for easier placement in human prostate cancer with remote controlled

procedure. Sources considered with this applicator are mostly isotopes of Cs-137 and Ir-

192.[49, 50]

Page 39: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

24

2.1.3 Nanoparticulates in nuclear medicine

Nanoparticles can be synthesized with the aid of natural products (edible) that have no

toxic side effects; particularly gold nanoparticulates have shown tremendous scope to

wide variety of medicinal applications. Nanoparticles bear large surface to volume ratios

and can chemically attach (for e.g. chloroquine,[51]) to target specific vector conjugates

based on the particles chemical affinity[52]. Gum Arabic is one such type of natural

product that is used in synthesizing gold nanoparticles as a stabilizing matrix. This is also

used in synthesis of radioactive gold nanoparticles. Gum Arabic suspended gold

nanoparticles have shown high affinity to tumor vasculatures.[36]

2.2 Polymer gel dosimetry

Many chemical compounds undergo structural changes after radiation exposure; this

phenomenon can be adopted to quantify radiation effects. Since, chemical dosimetry

directly assesses radiation effects based on ionization; it can be taken as a direct

measure of dose however there are some limitations to apply for absolute dosimetry.

Some chemical dosimeters display change of color associated due to the transition of a

metal ion in a different oxidation state. This allows the evaluation of the change

sphectrophotometrically. The most commonly used chemical dosimeter is based on the

Page 40: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

25

work of Fricke and Morse.[53] In these types of dosimeters radiation interaction oxidizes

Fe2+ in a ferrous sulfate solution to Fe3+, ferric iron. The basic reaction can be written as,

Fe2++OH+radiation→Fe3++OH-

Gelatin is used to contain the ferrous ions suspended uniformly in the container (gel

dosimeter) to validate radiation interactions accuracy. This was one of the important

enhancements for Fricke dosimetry. However these dosimeters had certain limitations

like working in controlled environments which lead to other discoveries. With constant

evolution in research and developments different materials were developed. Presently,

gel dosimetry is broadly classified into 2 categories as Fricke based gels and the polymer

gels. All types, of gel dosimeters exhibit physical density in equivalence to living tissue.

Polymer gels rely on radiation induced polymerization and cross linking of acrylic

monomers. The radiation products are part of the gel matrix and cannot diffuse within

the gel. Improved formulation exhibit a linear dose response relationship and is

independent of radiation energy and dose rate.

2.2.1 Gel dosimeter concept

Lin et. al. have used polymer gel dosimeters concept and named them MAGAT gels[40].

Their investigations suggested that Fricke gel dosimeters have low sensitivity of dose

response and the diffusion of ferrous and ferric ions after irradiation degraded the

Page 41: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

26

spatial resolution of the dosimeter. MAGAT gels were made using 6% gelatin, 9%

methacrylic acid, and THPC in aqueous solutions. Earlier researchers reported oxidation

problems with polymer gels and concluded that gel dosimeters were sensitive in the

normal atmosphere. To overcome this difficulty in preparation of polymer gels under

normal atmosphere, normoxic polymer gels were developed that incorporate an anti

oxidant (e.g. hydroquinone) to scavenge any free oxygen radicals generated within gel

from inhibiting radiation induced polymerization. Hydroquinone, the most effective anti

oxidant, was used in smaller proportions with methacrylic acid to get rid of oxygen

radicals. The accuracy of dose estimation was interpreted as the difference of planned

and measured dose maps. In conclusion, Lin et al. reported that gel dosimeters have

good dose response in the range of 0-12 Gy with R-square value of 0.99.

2.2.2 Synthesis and characterization of gelatin based dosimeters

The potential advantages of MRI gel dosimetry were recognized several years ago[54]

(1990), but early experience with the Fricke reaction identified some important

limitations such as blurring of dose distributions by diffusion of the products of radiation

interactions. To overcome such difficulties, alternatives to the Fricke effect were

investigated. Schulz and Fong report the changes in proton NMR relaxation in the gels

containing acrylic monomers that were exposed to radiation [41, 55]. Later monomer

based gel dosimeters gained preferences and many researchers accepted this

Page 42: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

27

phenomenon to estimate the radiation doses for MRI-based dosimetry. Polymer gel

dosimeters need one or more acrylic monomers suspended in gelatin (e.g. bovine

gelatin) for dosimetric studies. These acrylic gels respond to radiation through free

radical initiated polymerization that is rapidly quenched by oxygen. The loss in

sensitivity occurs because oxygen scavenges free radical initiators (like OH* and H*) that

are produced during radiolysis. To avoid oxygen quenching the gels were manufactured

in hypoxic conditions which involved exposure of toxic gases to personnel and thus

involved high costs to provide safer environment. To store it in oxygen free zones, gels

were stored in glass vials, and thus were incompatible for dosimetric use. For all such

reasons, newer techniques were constantly been in development. Fong et. al. termed

their polymer gel dosimeters as MAGIC gels.[41] These polymer gels are similar in

construction to MAGAT gels as explained earlier but use few chemicals in different

combinations. MAGIC gels are formulated in normal atmosphere and it is preferred

method to earlier hypoxic polymer gels. However, in conclusion MAGAT gels are

suggested to be more dose sensitive as compared to MAGIC gels. MAGIC gel dosimeters

were synthesized using gelatin (300 bloom strength), ascorbic acid, copper sulphate and

hydroquinone.

Irradiation: This section in the synthesis of gel dosimeters demonstrates about

characterizing gel dosimeters prepared. MAGIC gels in glass tubes were irradiated using

a 6000 Ci 137Cs gamma ray irradiator, which emits gammas at 662 keV. The samples

were irradiated at the calibrated dose rate of 1.83 Gymin−1. After exposing to irradiation

Page 43: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

28

the transverse NMR relaxation times (T2) were recorded at two different frequencies of

20 and 85 MHz. Particularly T2 measurement at 85 MHz was acquired through imaging

of transaxial sections through the tubes using a 31 cm bore 2 Tesla Bruker Avance

spectrometer. A 120 mm diameter birdcage coil was used for RF transmission and signal

reception. T2 measurements were made using single slice (10 mm) axial images

acquired using a RATE pulse sequence[56], which combines a train of RF refocusing

pulses with segmented echo planar acquisition for rapid T2 imaging measurements.

During each of the 32 spin-echo image acquisition periods, 8 independently phase-

encoded gradient echoes were collected. An 80 mm field-of-view (FOV) was encoded

with 128 samples in the read direction and 128 lines in the phase direction. The echo

time (TE) of 25 milliseconds and repetition time of 3 seconds (TR) were used, resulting in

total acquisition time of 1.5 min approximately. The recorded data were fit in at least

squares fashion to a single exponential. It was also reported that the gel dosimeters at

room temperature had physical density equivalent to that of tissue at 1060 kg.m-3.

In conclusion it is major concern that varied compositions of methacrylic acids in gel

dosimeters at different frequencies of magnetic resonance show unique dose

responses. Hence while synthesizing the gel dosimeters, it is important to maintain the

concentration of methacrylic acid and the tuning frequency of the MR scanner to be

maintained consistent to avoid dose sensitivity.[57]

Page 44: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

29

2.2.3 Importance of gelatin concentration

Shin-ichiro et.al. have reported the importance of gelatin and its role with different

concentrations of gelatin.[58] To achieve that direct temperature monitoring to the

changes in gel characteristics due to radiation was adopted as the mode of operandi.

Polymer gel dosimeters are made with gelatin, the main role of gelatin is to suspend

the synthetic monomers and also prevent the diffusion of the polymer formed after

interactions with radiation. Polymerization is an exothermal reaction, direct

temperature monitoring of gel formation after each irradiation process was adapted

to study the gels behavior. MAGAT gels were considered for measurements with

varied gelatin concentrations. The sample with concentration of 120 g/L of gelatin

gelatinized at room temperature. The correlation of temperature changes with

before and after irradiation process was reported with different concentrations of

gelatin. There was a rapid rise in the temperature due to radiation induced

polymerization at the beginning of the irradiation. However it was shown that

temperature effects have dependency over change of radiation doses except one set

of samples without gelatin. This means that presence of gelatin initiates the

polymerization process. Alternatively, few other amino acids like glycine, alanine and

proline were also tried with success to see similar temperatures effects and

polymerization. Although the viscosity of the solutions did increase after irradiation

but change in opacity was not appreciable. Also acetic acid and propionic acid, other

possible agents without amino groups, showed no responses to radiation

Page 45: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

30

interactions. In retrospect, it can be concluded that polymerization does occur with

an amino acid or gelatin. Hypothetically –NH-CH-CO- skeleton, which is common

structure in gelatin and/or amino acids, may be contributing to the polymerization

reaction.

2.2.4 THPC in polymer gel dosimeters

The radiation-induced polymerization is very sensitive to oxygen. Until a few years

ago, the only way to avoid the negative effects of oxygen on the polymerization

process was to remove all oxygen during the preparation process. The preparation

procedure had to take place in a nitrogen flushed glove box including extensive

bubbling of the polymer gel with nitrogen. The gel manufacturing process needed

special equipment and appeared quite elaborative. In 2002, De Deene et. al.[59]

suggested an alternative solution to use THPC (tetrakis-hydoxy-methyl-phosphonium

chloride). The THPC functions as an effective oxygen scavenger. THPC was found to

have the highest scavenging rate and it improved the dose sensitivity of the polymer

gel dosimeters.

Bayreder et.al. also have reported [60] about the performance of THPC as an

oxygen scavenger. The polymer gels made with THPC have better reproducibility,

accuracy and have good stability. Also it is reported that gel dosimeters made with

THPC have good dose rate dependence on the basis of MR scanning. However

accuracy of the gel dosimeters strongly depended on the calibration curve.

Page 46: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

31

2.4 Aqueous radioactive gold nanoparticles in brachytherapy

Nanoparticles are gaining attention due to unique change in characteristics as compared

their macro or bulk state. In wider perspective, nanoparticles are types of probe or drug

carrier agents used in research areas of pharmacology. Also nanoparticles can be tagged

with many functional moieties for target specificity [61]. Gold nanoparticles in specific

are being investigated for wide variety of applications in cancer diagnostics and cancer

therapy. Gold nano particulate approach may become the most preferred pathway in

the future, for many medical applications. Aqueous gold nanoparticles can be

synthesized by various methods. Similarly most of the metallic nanoparticles have been

characterized, to identify their unique physical and chemical properties. Nanoparticles

use in medicine is being considered as a novel approach, and this research presented in

this literature is related to applications in cancer therapy. Gold in particular is chosen,

because of its unique and beneficial properties in the nano size. Gold is found to be non

toxic, when used with conjugation to target specific ligands.[62]

There are many ways to synthesize gold nanoparticles, of various size and shape. Size

and shape significantly helps in characterizing gold nanoparticles to make them target

specific. For radiation applications gold can be neutron activated in the nuclear reactor

and made radioactive. Radioactive gold nanoparticles of Au-198 and Au-199 are short

lived isotopes (half life ~ 3 days). These isotopes emit both gammas and beta particles at

different energy levels. For comparison gamma energy characteristics for both radio

nuclides is shown in Table 2 to identify that Au198 have longer penetrability as to

Page 47: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

32

Au199. Both isotopes can be used to treat cancer lesions with their radiation energies

(betas and gammas) depending upon their suitability and specificity.

TABLE II – THE UNSHIELDED GAMMA RAY DOSE EQUIVALENT RATE AT 1 METER FROM POINT SOURCE FOR GOLD NUCLIDES

Nuclide Average Gamma Energy

keV

Half Life

days

Γ

(mSv/h)/MBq

t

cm

µ

cm-1

Au198

415 2.7 7.88 ×10-5 1.29 2.32

Au199 166 3.1 1.866 × 10-5 0.186 16.12

Γ- Specific gamma ray dose constant, which is unshielded gamma ray dose equivalent

rate at 1 meter from a point source.

t= thickness of lead to attenuate gamma rays by 95%.

µ=linear attenuation coefficient in lead to stop gamma rays by 95%.

2.4 AAPM TG43 recommendations

There are about 6 radionuclides currently used as sealed sources in both temporary and

permanent implants. They are cobalt 60, cesium 137, iridium 192, iodine 125, palladium

Page 48: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

33

103, and strontium-90/yttrium 90[4]. Total dose and the dose rate of these

radionuclides with respect to the region of interest, are the key parameters that

differentiate their treatment capabilities and preferences. The physical parameters such

as half life, source strength, exposure rate or air kerma rate are necessary to evaluate

total dose or the dose rates. Moreover, the AAPM Task Group 43 (TG-43) protocol

recommends the absorbed dose formalism, which is the collective effort of all the

medical physicists in United States through their clinical experience and clinical

research. This formalism was introduced in 1990, and it describes, two dimensional (r, θ)

dose distribution for the clinically used radionuclides as shown in Figure 4. The symbolic

details like t, T and all the angles of geometry are explained in detail within TG-43[63].

Earlier this formalism was used for iridium 192 and iodine 125 and in 1995 it was

expanded to palladium 103 seed sources. Essentially TG-43 with its wide acceptance

shall be used in this brachytherapy dosimetry study. Briefly, the formalism defines the

dose rate as,

D(r, θ) = 𝑆𝑘𝛬 �𝐺(𝑟,θ)𝐺(𝑟0,𝜃0)� 𝑔(𝑟)𝐹(𝑟, θ)

Where,

Sk →Air Kerma Strength

Λ →Dose Rate Constant

G (r, θ) →Geometry function

Page 49: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

34

G (1,𝜋2) →Reference position

g (r) →Radial dose function

F (r, θ) →2D anisotropic function

Figure 4 - Coordinate system used for brachytherapy dosimetry calculations.

Similarly, one dimensional formalism is defined in the TG-43 protocol for point sources,

which is simpler to the two dimensional formalism. Precise measurement of these

parameters independently determines the dose rate. The brief details about the

importance of these parameters are worthwhile to explain as this simulation is applied

to the chromium isotope in this report.

Air Kerma Strength: Measured and provided by NIST (National Institute of Standards and

Technology) as primary standard, is based on wide angle free air chamber.

Page 50: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

35

Dose Rate Constant: Calculated based on Air Kerma Strength and reference dose. The

reference dose is calculated by experimental studies and confirmed by Monte Carlo

Simulations.

Geometry function: Provides inverse law square correction. The purpose of a geometry

function is to improve the accuracy with which dose rates can be estimated by

interpolation from data tabulated at discrete points.

Radial dose function: This accounts for dose fall off on the transverse plane due to

photon scattering and attenuation. It’s a fifth order polynomial fit that may produce

some erroneous results.

2D anisotropic function: Describes the variation in dose as a function of polar angle

relative to the transverse plane.

Before clinical acceptance to any new brachy sources, TG-43 protocol would be a very

useful reference for dosimetry evaluations. In broader perspective radiation dosimetry

depends on mathematical computations, assumptions, and appropriate interpolations.

However, most of the experimental and computational investigations may still include

uncertainties. As per the AAPM recommendation[63], it is required to perform

dosimetry on every brachytherapy agent with an experiment supported by one Monte

Carlo simulation.

Page 51: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

36

2.5 Monte Carlo Simulation Technique

2.5.1 Introduction

Monte Carlo simulation deals with statistical sampling of particle histories. It works on

the principles of random sampling or gambling and derives its name of Monte Carlo

from the one of the oldest legalized casinos in Europe. Monte Carlo N-Particle Transport

Code (MCNP) is a software package for simulating nuclear processes. It is developed by

Los Alamos National Laboratory and is distributed within the United States by the

Radiation Safety Information Computational Center in Oak Ridge, TN. It is used primarily

for the simulation of nuclear processes, such as fission, but has the capability to

simulate particle interactions involving neutrons, photons, and electrons. Specific areas

of application where it is extensively used, but are not limited to, radiation protection

and dosimetry, radiation shielding, radiography, medical physics, nuclear criticality

safety, detector design and analysis, nuclear oil well logging, accelerator target design,

fission and fusion reactor design, decontamination and decommissioning. Monte Carlo

methods are very different from deterministic transport methods. Deterministic

methods, the most common of which is the discrete ordinates method[64], solve the

transport equation for the average particle behavior. By contrast, Monte Carlo obtains

answers by simulating individual particles and recording some aspects (called as tallies)

of their average behavior. The average behavior of particles in the physical system is

then inferred (using the central limit theorem)[65, 66] from the average behavior of the

Page 52: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

37

simulated particles. Deterministic methods typically give fairly complete information

throughout the phase space of the problem. Monte Carlo supplies information only

about specific tallies requested by the user.[67]

MCNP is a general-purpose, continuous-energy, generalized-geometry, time-dependent,

coupled or neutron or photon or electron Monte Carlo transport code. It can be used in

several transport modes: neutron only, photon only, electron only,

neutron/photon/electron, photon/electron, or electron/photon. In MCNP use, the

neutron energy regime is from 10 MeV to 20 MeV for all isotopes and up to 150 MeV for

some isotopes, the photon energy regime is from 1 keV to 100 GeV, and the electron

energy regime is from 1 KeV to 1 GeV.

2.5.2 MCNP features

2.5.2.1 Nuclear data and reactions

MCNP uses continuous-energy nuclear and atomic data libraries. The primary sources of

nuclear data are evaluations from the Evaluated Nuclear Data File (ENDF)[68] system,

Advanced Computational Technology Initiative (ACTI)[69], the Evaluated Nuclear Data

Library (ENDL)[70], Evaluated Photon Data Library (EPDL)[71], the Activation Library

(ACTL)[72] compilations from Livermore, and evaluations from the Nuclear Physics (T–

16) Group[73-75] at Los Alamos. Evaluated data are processed into a format appropriate

for MCNP by code such as NJOY[76, 77]. Photon interaction tables exist for all elements

Page 53: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

38

from Z = 1 through Z = 100. The data in the photon interaction tables allow MCNP to

account for coherent and incoherent scattering, photoelectric absorption with the

possibility of fluorescent emission, and pair production.

2.5.2.2 Source specification

MCNP has capability that allows users to specify a wide variety of source conditions

without having to make a code modification. Independent probability distributions may

be specified for the source variables of energy, time, position, and direction, and for

other parameters such as starting cell(s) or surface(s). Information about the

geometrical extent of the source can also be given. The user can bias all input

distributions. In addition to input probability distributions for source variables, certain

built-in functions are also available. These include various analytic functions for fission

and fusion energy spectra such as Watt, Maxwellian, and Gaussian spectra; Gaussian for

time; and isotropic, cosine, and mono directional for direction.

2.5.2.3 Tallies in MCNP

MCNP accepts instructions to make various tallies related to particle current, particle

flux, and energy deposition. MCNP tallies are normalized to be per starting particle

except for a few special cases with nuclear reaction criticality sources. Currents can be

tallied as a function of direction across any set of surfaces, surface segments, or sum of

surfaces in the problem. Charge can be tallied for electrons and positrons. Fluxes across

any set of surfaces, surface segments, sum of surfaces, and in cells, cell segments, or

Page 54: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

39

sum of cells are also available. Similarly, the fluxes at designated detectors (points or

rings) are standard tallies, as well as radiography detector tallies. A dose deposition tally

provides the energy deposited in specific region to its unit mass. A pulse height tally

provides the energy distribution of pulses created in a detector by radiation. In addition,

particles may be flagged when they cross specified surfaces or enter designated cells,

and the contributions of these flagged particles to the tallies are listed separately. Tallies

such as the number of fissions, the number of absorptions, the total helium production,

or any product of the flux times the approximately 100 standard ENDF reactions plus

several nonstandard ones may be calculated with any of the MCNP tallies. All tallies are

functions of time and energy as specified by the user and are normalized to be per

starting particle.

2.5.2.4 MCNP Output

The output file contains tables of standard summary information to give the user an

idea of program compilation. This information can give insight into the physics of the

problem and the adequacy of the Monte Carlo simulation. If errors occur during the

running of a problem, detailed diagnostic prints for debugging are given. Printed with

each tally is also its statistical relative error corresponding to one standard deviation.

Following the tally is a detailed analysis to aid in determining confidence in the results.

Ten pass/no-pass checks are made for the user-selectable tally fluctuation chart (TFC)

bin of each tally. The quality of the confidence interval still cannot be guaranteed

because portions of the problem phase space possibly still have not been sampled. Tally

Page 55: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

40

fluctuation charts, are also automatically printed to show how a tally mean, error,

variance of the variance, and slope of the largest history scores fluctuate as a function of

the number of histories run.

2.5.2.5 Estimation of Monte Carlo Errors

MCNP tallies are normalized to be per starting particle and are printed in the output

accompanied by a second number R, which is the estimated relative error defined to be

one estimated standard deviation of the mean divided by the estimated mean . In

MCNP, the quantities required for this error estimate−−the tally and its second

moment−−are computed after each complete Monte Carlo history, which accounts for

the fact that the various contributions to a tally from the same history are correlated.

For a well-behaved tally, R will be proportional to where N is the number of histories.

Thus, to halve R, we must increase the total number of histories fourfold. For a poorly

behaved tally, R may increase as the number of histories increases. The estimated

relative error can be used to form confidence intervals about the estimated mean,

allowing one to make a statement about what the true result is. The Central Limit

Theorem states that as N approaches infinity there is a 68% chance that the true result

will be in the range and a 95% chance in the range. It is extremely important to note that

these confidence statements refer only to the precision of the Monte Carlo calculation

itself and not to the accuracy of the result compared to the true physical value. A

statement regarding accuracy requires a detailed analysis of the uncertainties in the

physical data, modeling, sampling techniques, and approximations, etc., used in a

Page 56: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

41

calculation. The guidelines for interpreting the quality of the confidence interval for

various values of R are listed in Table iii. Depending upon the output values of R, as

shown in the Table III the data can be termed as trustworthy or not.

TABLE III – GUIDELINES FOR INTERPRETING THE RELATIVE ERROR R

Range of R Quality of the Tally 0.5 to 1.0 Not meaningful 0.2 to 0.5 Factor of a few 0.1 to 0.2 Questionable

< 0.10 Generally reliable < 0.05 Generally reliable for point detectors

Despite one’s best effort, an important path may not be sampled often enough, causing

confidence interval statements to be incorrect. To try to inform the user about this

behavior, MCNP calculates a figure of merit (FOM) for one tally bin of each tally as a

function of the number of histories and prints the results in the tally fluctuation charts

at the end of the output. The FOM is defined as,

F.O.M. = 1𝑅2𝑇

where T is the computer time in minutes. The more efficient a Monte Carlo calculation

is, the larger the FOM will be because less computer time is required to reach a given

value of R. The FOM should be approximately constant as N increases because R2 is

proportional to 1/N and T is proportional to N. Examining the tally fluctuation charts is

Page 57: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

42

good practice to be sure that the tally appears well behaved, as evidenced by a fairly

constant FOM. A sharp decrease in the FOM indicates that a seldom-sampled particle

path has significantly affected the tally result and relative error estimate.

2.5.2.6 Variance Reduction

As noted in the previous section, R (the estimated relative error) is inversely

proportional to N, where N is the number of histories. For a given MCNP run, the

computer time T consumed is proportional to N. Thus,

𝑅 =𝐶√𝑇

where C is a positive constant. There are two ways to reduce R: (1) increase T and/or (2)

decrease C. Computer budgets often limit the utility of the first approach. For example,

if it has taken 2 hours to obtain R = 0.10, then 200 hours will be required to obtain R =

0.01. For this reason MCNP has special variance reduction techniques for decreasing C.

(Variance is the square of the standard deviation.) The constant C depends on the tally

choice and/or the sampling choices.

Page 58: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

43

2.5.3 MCNP structure

MCNP is written in ANSI-Standard Fortran 90[78]. Global data is shared via FORTRAN

modules. The general internal structure of MCNP is as follows:

A. Initiation (IMCN):

• Read input file (INP) to get dimensions;

• Set up variable dimensions or dynamically allocated storage;

• Re-read input files (INP) to load input;

• Process source;

• Process tallies;

• Process materials specifications including masses without loading the data files;

• Calculate cell volumes and surface areas.

B. MCRUN sets up multitasking and multiprocessing, runs histories, and returns to

print, write.

C. RUNTPE dumps, or process another cycle.

Page 59: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

44

CHAPTER 3

MATERIALS AND METHOD

3.1 Radioactive gold nanoparticles synthesizing methods

198Au - Gold foil 5-30 mgs were irradiated at a neutron flux of 8 ×1013 𝑛𝑐𝑚2.𝑠

. The

radioactive foil was subsequently dissolved with aqua regia, dried down and

reconstituted in 0.05 –1 ml of 0.05 N HCl to form HAuCl4. The radioactive gold

solution is added to aqueous solutions of gum Arabic (GA), followed by solution of

THPAL (P (CH2NHCH (CH3) COOH) 3), a modified amino acid acts as reducing agent,

for formation of radioactive gold nanoparticles of well-defined particle sizes (15-20

nm). Then the final solution was diluted and prepared to the concentration of

0.033𝑚𝑔𝑚𝑙

.

199Au, will be synthesized in the similar sequence as 198Au. Platinum foil is irradiated

at a flux of 8 ×1013 𝑛𝑐𝑚2.𝑠

. The radioactive foil is subsequently dissolved with aqua

regia, dried down and reconstituted in 0.05 –1 ml of 0.05 N HCl to form HAuCl4. The

radioactive gold solution is added to aqueous solutions of gum Arabic (GA),

followed by THPAL for formation of radioactive gold nanoparticles of well-defined

particle sizes (15-20 nm). Then the final solution was diluted and prepared to the

concentration of 0.033𝑚𝑔𝑚𝑙

.

Page 60: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

45

3.2 Gel Dosimeter Synthesis

Polymer gels have been synthesized as per the procedure mentioned below.

Polymer gel formation procedure

Qty.: 60ml

Composition: bovine gel (~300 bloom strength), THPC, Methacrylic acid, water.

1) Add 51.6 ml of DI water in a clean 80ml beaker.

2) Add 7.2 g of gelatin in the water, and allow it to swell after soaking for half an

hour.

3) Heat the contents in microwave for approximately 15 seconds, until the solution

turns to clear solution.

4) Add a clean magnetic stir bar to the mixture and stir the solution continuously

until the solution reaches to 500C. This confirms that gelatin has dissolved

completely.

5) Allow the solution to cool down up to 400C.

6) Make a hydroquinone solution in a separate 20ml vial.

0.12 g of hydroquinone + 3 ml of DI water

7) Add the hydroquinone solution quickly in to the molten gel.

Note: As the temperature of the contents goes down, gelatinization begins to form.

8) Rapidly add 154.5µl of THPC, in to the mixture.

Page 61: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

46

9) Rapidly add 5.4g of methacrylic acid in to the solution.

10) Solution is stirred for few seconds for the contents to mix homogeneously.

11) Set the gel immediately, in a transparent plastic tube (50ml centrifuge tube) for

gelatinization.

Note: Gelatinization completes in few hours.

3.3 Dosimetry (Characterization)

The dosimeters were prepared in batches of 8 to get set of 8 readings. Individual

dosimeters were treated in 0.5, 1, 4, 8, 16, 32, and 64 Gray of dose with 8X Siemens

linear accelerator according to the radiation treatment plan. One gel dosimeter was

the untreated control. The radiation treatment planning was done on a CT image of

the dosimeter located in a water tank at the isocenter of 100 mm SAD (Source to

Axis Distance) with XiO 4.60, application software used for teletherapy treatment

planning procedures. Acrylic tank of 12×12×12 inches was filled with water and up

to the rim and the dosimeter was held at the center of the tank at 100 cm SAD

(Source to Axis Distance). Four beam angles at gantry angles of 90o, 1800, 2700 and

dorsal (00) on 8X Siemens Linear Accelerator were used to deliver uniform doses to

the gel dosimeter. Field size of 8cm ×8cm was used for all the beam angles. The

dose distributions planned in XiO treatment planning system is shown in the Figure

5. The dosimeters were treated in the steps of 0, 250, 500, 1000, 2000, 3000, 4000,

5000, 6000 and 7000 cGy’s.

Page 62: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

47

Figure 5 – Radiation Treatment Plan of gel dosimeter on CT acquired image with XiO software (Ver.4.6). Dosimeter was located at 100 mm SAD in an acrylic water tank of size 1 cubic foot.

After the incremental treatment on the dosimeters there was a change in physical

appearances. All of the dosimeters which were treated with more than 16 Gy of

dose had change in opacity as shown in Figure 6. This foggy appearance within the

gel dosimeters is the indication of polymerization due to interaction of radiation.

Page 63: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

48

Figure 6 - The dosimeters treated with Siemens 8X Linac

3.4 Sample identification

Samples were identified with ID# to identify the dosimeter batches. There are 4 digits in the ID #. Detail with one e.g.

e.g. 1, 2, 3…

First digit: is for batch numbers, to keep track between each batch made.

For e.g. if it is 6th batch and the dose to be delivered is 37.5 Gy, Then its ID# would be 6375.

Next three digits: gives information about doses

Page 64: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

49

3.5 MR Scanning

The dosimeters were then scanned in 1.5T GE MRI scanner with GP Flex® coil at Veterinary Medicine Teaching Hospital of University of Missouri on these polymer gels using the following MR protocol.

Pulse Sequence: T2 Transverse

TR: 5000 ms

TE: 15, 30, 45, 60, 75 ms

Slice thk: 5mm

Gap: 0.3 mm

Slices: 9

NEX: 3

FOV: 16 X 16

RF Coil: GP Flex®

Bandwidth: 31.25

Echo Train length: 16

Plane: Oblique

Mode: 2D

Acquisition timing: Freq x Phase 256 x 192

Page 65: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

50

3.7 SPECT imaging

There is no dosimetry tool to evaluate radioactive nanoparticles. The first dosimetry

evaluation that was used to evaluate the radioactive nanoparticles of gold is described

here. A previously constructed acrylic phantom of a dog’s head was used in this study

(Figure 7).This phantom is a realistic- reconstruction of the shape and size of a 50-70

pound dog. The phantom is divided down the middle and there is a cavity in the central

portion of the phantom. This cavity is representative of the nasal passages and sinuses.

An approximately 1 inch diameter (8-10ml) ball of Play Doh® was injected with

approximately 200 micro Curies of 198AuCl in solution. 198AuCl in solution used is form of

gold complex in aqueous medium and not in nanoparticulate form. The Play-Doh® ball

was then placed into the cavity in the phantom and the two sides of the phantom put

Figure 7 - Phantom set of acrylic construction of beagle dog’s head representing pelvic bone and Play-Doh® ball representing prostate gland wrapped with adhesive tape. Left picture-Play Doh® ball closed inside the acrylic construction. Right picture

– Inside view with one side of the acrylic construction removed.

Page 66: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

51

back together and held together with tape. The phantom was placed on the SPECT

imaging table near the isocenter of the imaging plane. A high energy collimator was

used along with the detection window of the gamma camera set for the 411 keV

photon. The radius of rotation was 11 inches, with data acquisition in regular intervals

with a fixed time of 15 seconds at every interval. The data collection was performed in

the array of matrix of 128×128.

3.4 MCNP Simulation

Method

Multiple radiation calculations were performed using MCNP version 5 (MCNP5)

obtained from Los Alamos National Laboratory (LANL). Full physics modeling utilizing

photon and electron transport was performed to calculate parameters as accurately as

possible. For photon interactions, the detailed physics treatment includes coherent

(Thomson) scattering and accounts for fluorescent photons after photoelectric

absorption. Form factors and Compton profiles are used to account for electron binding

effects. Analog capture is always used. For photon transport models thick –target

bremsstrahlung model is used. This model generates electrons and assumes that they

are locally slowed to rest. For electron transport all photon collisions except coherent

scatter can create electrons that are banked for later transport. The cross section data

libraries used for photon transport are termed as MCPLIB04. Cross section data within

these libraries for incident photon energies are in the range of 1 keV to 100 GeV.

MCNP5 uses the data libraries el and el03 for electron interactions.

Page 67: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

52

The radioactive sources used in this experiments are both gamma and beta

emitters. Hence, random sampling of both photons and electrons are necessary in

modeling such cases. For 198Au and 199Au sources both photons and electrons have to be

considered since they are applied in to the medium without any shielding. To achieve

higher efficiency transport models for photons and electrons are simulated in MCNP

separately for all cases. However, for 51Cr source only photons will be considered

because the source is packed in denser material that can absorb beta radiation. For e.g.

AISI Type 316L, stainless steel. In this work the dose rate factors of the radioactive

sources will be calculated using MCNP5 code for all of the three sources independently.

The geometry of the interacting medium will be a 50ml polypropylene centrifuge tube

as shown in Figure 6. The composition of the gel dosimeter will be as shown in Table IV

A). Furthermore, the material fractions shown in Table IV B are necessary for MCNP

models are derived with reference to Table IV A.

Page 68: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

53

TABLE IV A&B – COMPOSITION AND MASS FRACTIONS OF GEL DOSIMETER USED IN

MCNP MODELING OF Au-198 AND Au-199 SOURCES.

A) COMPOSITION OF GEL DOSIMETER

Chemicals in gel dosimeter

Empirical formula

Mol. Wt.,

gm/mol

Wt. in sample,

g

C H N O P Cl

12 1 14 16 31 35.45

Methacrylic acid C4H6O2 86.09 5.4 4 6 2

Gelatin C37H60N12O12 864 7.2 37 60 12 12

THPC C4H12O4PCl 190.56 0.1656 4 12 4 1 1

Water H2O 18 54.641 2 1

Hydroquinone C6H6O2 110.11 0.12 6 6 2

B) MASS FRACTIONS OF CONTENTS OF GEL DOSIMETER

C H N O P Cl

10.12 10.31 2.07 77.41 0.04 0.05

The absorbed dose was calculated in average Rads/photon using F6 tally for photon

transport. Similarly, the absorbed dose was calculated by taking the ratio of energy

absorbed to the mass of the volume element (voxel) in question using MCNP *F8 tally

for electron transport. Voxel computational cells were placed in selected positions

necessary to determine relevant dosimetry parameters. The total number of photons

Page 69: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

54

transported for every simulation run was 109 in order to arrive at relative errors < 0.1%.

The number of electron transported per calculation was 106 in order to arrive at relative

errors < 0.1.

Source definition in this modeling was the key specifically for the radioactive gold

nanoparticles of Au-198 and Au-199. The source was assumed to be of spherical volume

of radius 0.25 cm. This assumption is based on the nanoparticulate spread by its flow

within the gel dosimeter after the injection of desired quantity. The source distribution

and probabilities for all possible radiation that is more than 50 keV and at least with

0.1% of abundance was weighted appropriately for all the MCNP runs and particle

histories will be run until estimated errors are well within the acceptable limits (< 0.10).

3.6.2 The MCNP geometry models

3.6.2.1 Model for radioactive nanoparticles

The MCNP geometry model for the radioactive gold nanoparticles of Au-198 and Au-

199 is shown in the Figure 8. This geometry shown is not drawn to scale and is only

conceptual. The gel dosimeter is held in cylindrical shaped container that has

dimensions of 2.66 cm diameter and 11 cm length. The imaginary spheres indicate

different regions of interest or cells in MCNP5 to evaluate tallies. The central core

Page 70: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

55

part represents the source in spherical volume. This region that represents source is

assumed and approximated based on the injected solution spread in that medium.

Figure 8 -The MCNP model for radioactive nanoparticles.

3.6.2.2 Model for Cr-51 seed

The MCNP geometry model for the proposed Cr-51 brachytherapy seed is shown in

Figure 9. This model is not drawn to scale but is conceptual representation. The

dimensions of the seed are 1.31 mm in diameter and 1 cm in length. This

assumption was based on the available raw material. This is fairly good

Page 71: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

56

assumption that resembles other clinically used brachy seed designs. The seed

is placed in the water medium. Regions of interest or cells in MCNP5 were

created in such a way that conveniently follows the AAPM TG43 protocol

recommendations.

Figure 9- The MCNP model for Cr-51 brachytherapy seed. The dark blue region or cell 4 in MCNP5 is Chromium material. The violet regions or cells 1, 2, and 3 are representing material for encapsulation of stainless steel. The cells 5, 6, 7, etc. are for different tallies that follows the AAPM TG43 protocol.

Page 72: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

57

3.1.3. Physical Source description

3.1.3.1. Gold nanoparticles

The MCNP models for both Au-198 and Au-199 are experimented separately. The

photon energy spectrum of both X-rays and gammas which are >50 keV and with

abundance of at least 0.1% as shown in Table V were used for Au-198 to simulate

emissions from Au-198. The beta emissions shown in Table V are the most applicable

values of energy. However, the beta spectrum[79] of Au-198 shown in Figure 10 was

used for MCNP5 electron transport model of Au-198.

TABLE V – DECAY PROPERTIES OF Au-198

Type of emission E (keV) Abundance (%) Decay Mode

X-rays

68.9 0.8 Hg Kα2

70.8 1.3 Hg Kα1

79.8 0.1 Hg Kβ3

80.2 0.3 Hg Kβ1

82.5 0.1 Hg Kβ2

Gammas

411.8 96 β-

675.9 0.8 β-

1087.7 0.2 β-

Betas

284.8 1 β-

960.7 98.99 β-

1372.5 0.025 β-

Page 73: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

58

Figure 10 The beta energy spectrum of Au-198 isotope[79]. The maximum beta energy is 1372.5 keV.

The photon energy spectrum of both X-rays and gammas which are >50 keV and with

abundance of at least 0.1% as shown in Table VI were used for Au-199 to simulate

emissions from Au-199. The beta emissions shown in Table VI are the most applicable

values of energy. However, the beta spectrum[79] of Au-199 shown in Figure 11 was

used for MCNP5 electron transport model of Au-199.

0.00E+00

2.00E-02

4.00E-02

6.00E-02

8.00E-02

1.00E-01

1.20E-01

1.40E-01

0.000 0.200 0.400 0.600 0.800 1.000 1.200 1.400

#/nt

Energy (MeV)

Au-198

Page 74: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

59

TABLE VI - DECAY PROPERTIES OF Au-199

Type of emission

E (keV) Abundance (%) Decay Mode

X-rays

68.9 5 Hg Kα2

70.8 8.6 Hg Kα1 79.8 1 Hg Kβ3 80.2 2 Hg Kβ1 82.5 0.7 Hg Kβ2 82.7 0.1 Hg Kβ4

Gammas 49.8 0.36 β-

158.4 40 β- 208.2 8.7 β-

Betas

244.1 21.5 β- 293.9 72 β- 452.3 6.5 β-

Figure 11 The beta energy spectrum of Au-199 isotope[79]. The maximum beta energy is 452.3 keV.

0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 1.00E-01 1.20E-01 1.40E-01 1.60E-01 1.80E-01

0.000 0.050 0.100 0.150 0.200 0.250 0.300 0.350 0.400 0.450

#/nt

Energy (MeV)

Au-199

Page 75: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

60

Both Au-198 and Au-199 have significant beta energies. The absorbed dose or the

radiation dose deposited by these nuclides is from both high energy photons and also

the electrons. To compute the absorbed dose, the dose rate of the radionuclide in

question is determined first. The dose rate is summation of dose rates from photons

and electrons. For simplification of the MCNP5 model the photon and electron transport

models are run separately for both Au-198 and Au-199 nuclides. Mode card, a data

section within MCNP5 coding determines the type of transport models. In MCNP5

photons only transport model is used as Mode P E. This tallies all the photons that were

born randomly from the source cell. Also the secondary electrons generated due to

photonic interactions were also tallied in this mode.

In MCNP5 electrons only transport model is used as Mode E. This tallies all the electrons

that were born randomly from the source cell until they were terminated in the process.

Also the secondary photons generated due to electron collisions were banked and

tallied. Both Au-198 and Au-199 nuclides had 2 models each. For Au198 MCNP model

and Au199 MCNP model spherical geometry radius 3.16 mm and 10.7 mm were applied.

This approximate assumption is based upon the mimic of the spread of nanoparticles

within gel dosimeters after injection. Also the MCNP5 transport models of Au198/Au199

have another assumption of uniform distribution of photons throughout the volume.

Page 76: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

61

3.1.3.2. Chromium 51

The MCNP model for Cr-51 has cylindrical geometry of 0.51 mm and 1 cm. The material

used for source is enriched Cr-50. The source was encapsulated with AISI Type 316L

stainless steel that has density of 8.02 g/cc to stop all the beta radiation. The thickness

of steel was derived from the standard reference of beta range spectrum of ICRP

(International Commission on Radiological Protection) shown in Figure 12. The

maximum beta energy of Cr-51 is 752.8 keV and about 0.4 mm thick stainless steel

should be able to absorb almost all of the beta energies. In MCNP5 code for photon only

transport encapsulation of stainless steel of at least 0.4 mm is added on the surface of

Cr-50 material. Also the electrons transport model for CR-51 is considered to compute

the dose rate, if by bremsstrahlung radiation. Source biasing is used as a variance

reduction technique to mitigate the self absorption factor.

Page 77: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

62

Figure 12 Range of beta particles energy in inches in some various media.

Page 78: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

63

CHAPTER 4

QUANTIFICATION OF RADIOACTIVE GOLD NANOPARTICLES DOSE

The order of data presentation

4.1 Gel dosimeter calibration

4.1.3 Gel dosimeter consistency

4.1.3 Treatment plan

4.1.3 MR Imaging

4.2 Au198 nanoparticles

4.2.1 MR Imaging

4.2.2 SPECT Imaging

4.2.3 MCNP models

4.2.3.1 Photon transport

4.2.3.2 Electron Transport

4.3 Au199 nanoparticles

4.3.1 MR Imaging

4.3.2 SPECT Imaging

4.3.3 MCNP models

4.3.3.1 Photon transport

4.3.3.2 Electron transport

Page 79: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

64

4.1 Gel dosimeter calibration

4.1.1 Gel consistency

To check the consistency of gel dosimeters 3 separate batches of gel dosimeters were

synthesized as described in Chapter 3. Three batches, made in separate batches were

chosen for statistical significance. Each dosimeter was scanned on MR scanner to check

the consistency across the three batches. For each dosimeter T2 Transverse pulse

sequence was applied on GE 1.5T MR scanner. Each dosimeter was scanned 4 times

with echo times of 15, 30, 45 and 60 ms. With reference of SNR=100 % (signal to noise

ratio), field of view (FOV=16 × 16 cm2) and thickness (=5mm) of the slices were used.

Data was acquired across frequency and phase of 256 × 192 for about 3.8 minutes.

Averaging 3 ROI (regions of interest) per slice the T2 relaxation was measured for all the

three batches. The relation between T2 relaxation times in milliseconds against the TE in

milliseconds as shown in the Figure 13 plotted.

Page 80: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

65

Figure 13 – Gel dosimeter consistency response. Dosimeters made in separate batches were scanned in GE 1.5T MR scanner with T2 Transverse pulse sequence at various TE, echo times of 15, 30, 45 and 60 milliseconds. T2, spin-spin relaxation time in milliseconds was the response collected against the all the TE’s.

4.1.2 Treatment plan

CT (Computer Tomography) imaging of tank and phantom are calibrated. CT used to

model the dose determination. 4 field plans was placed at equal intervals of 900 to

maintain the dose homogeneity. Effective dose weights were applied depending on the

percentage depth dose (PDD) at every field. The sample holder at the center of the tank

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

0 20 40 60 80

T2 (S

pin

spin

rel

axat

ion

tim

e) ,m

s

TE (Echo time),ms

T2 relaxation curves

Batch 1

Batch 2

Batch 3

Page 81: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

66

was in one fixed position. The distance from the surface of the water to the central axis

of the sample at 0, 90, 180, and 270 were 13, 15.3, 13.1, and 14.6 cm respectively. The

PDD at the axis of the sample for the fields at 0, 90, 180, and 270 were 0.764, 0.706,

0.761, and 0.725 respectively. Field size of 8 cm × 8 cm was used for all the beam angles.

One dosimeter per dose was used. The dosimeters were treated at 0, 2.5, 5, 10, 20, 30,

40, 50, 60 and 70 Gray. Each dose was delivered equally from all the beam angles to

maintain homogeneity.

4.1.3 MR Imaging

8 dosimeters were used for calibration study. Every dosimeter was scanned on 1.5T GE

MR Scanner with Fast Spin multi spin echo T2 Transverse pulse sequence. The samples

were scanned for 4 different echo times for e.g. 15, 30, 45, and 60ms with long TR

(Repetition time) of 5000 ms. Saline bags were used around the dosimeter to

homogenize field response. The random circular ROI (region of interest) of 69.03mm2

was considered. The average of T2 relaxation times of that ROI size was recorded from 5

different slices of 5 mm thickness for statistical significance. Then it was repeated for at

least 4 different echo times in incremental order. Two sets of MR images for 30 and 70

Gy dose are shown in Figure 14 & Figure 15 in sequence to demonstrate the

polymerization due to interaction of radiation. Change in physical appearance for both

of the dosimeters was there but not distinct. However, the MR response to these

changes was clearly distinct between these two dosimeters. This change of contrast

Page 82: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

67

across the images is the polymerization response that is required to evaluate to co-

relate it to the absorbed doses.

Figure 14 – (Top) this is an image of 3D localizer, wherein MR scanner by default will scan briefly sample in Transverse/Sagittal orientations, which further helps to locate sample and choose the area of interest for full scan. (Bottom) MR Images obtained with 30 Gy dose at TE=14.5 ms, TE=33.9 ms, TE=53.3 and TE=63 ms, TR=5000 ms, slice thickness = 5mm, Transverse T2 weighted imaging. Two large regions around the dosimeter were of saline bags, which were used for homogeneous field response.

Page 83: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

68

Figure 15 – (Top) this is an image of 3D localizer, wherein MR scanner by default will scan briefly sample in Transverse/Sagittal orientations, which further helps to locate sample and choose the area of interest for full scan. (Bottom) MR Images obtained with 70 Gy dose at TE=14.5 ms, TE=33.9 ms, TE=53.3 and TE=63 ms, TR=5000 ms, slice thk= 5mm, Transverse T2 weighted imaging.

Page 84: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

69

In the next step, T2 relaxation times and echo times were plotted for every

dosimeter as shown in Figure 16. The change in T2 responses across the dosimeters

is distinct and represents decay of the signal with increase in echo times. The

responses from each dose curve distinctly show characteristic slopes. Each curve

represents the dose response of the dosimeters treated with 20X Linac (Siemens

Mevatron).

(A)

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0 10 20 30 40 50 60 70

Nor

mal

ized

T2

rela

xed

valu

e

Echo Time, ms

Normalized T2 relaxation curves

0 Gy

2.5 Gy

5 Gy

10 Gy

20 Gy

30 Gy

40 Gy

50 Gy

60 Gy

70 Gy

Page 85: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

70

(B)

(C)

Figure 16 A, B, C, – Normalized T2 relaxation time as function of echo time. The dosimeters treated with lower doses in the range, 0-40 Gy show linear response. With increase of dose >40 Gy, the curve shows exponential responses. The results are shown in mean T2 relaxation time ± standard error. The standard deviation is more to the dosimeters treated with doses >70 Gy.

0.83

0.88

0.93

0.98

1.03

0 10 20 30 40 50 60 70

Nor

mal

ized

T2

rela

xed

valu

e

Echo Time, ms

Normalized T2 relaxation curves

0 Gy

2.5 Gy

5 Gy

10 Gy

20 Gy

30 Gy

40 Gy

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0 10 20 30 40 50 60 70

Nor

mal

ized

T2

rela

xed

valu

e

Echo Time, ms

Normalized T2 relaxation curves

0 Gy

50 Gy

60 Gy

70 Gy

Page 86: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

71

Further slopes of the curves for each dosimeter across the set of echo time intervals

were derived from the chart equation calculated in MS® Excel. Dose response curve is

plotted with a convenient function R2 (slope of the normalized T2 relaxation curve, 1𝑠𝑒𝑐

)

is used for demonstration as shown in Figure 17. This will be an important curve

referred as ‘Calibration Curve’ that determines the unknown doses from the radioactive

nanoparticles of Au198 and Au199.

Figure 17 – Dose Calibration Curve.

-0.0050

0.0000

0.0050

0.0100

0.0150

0.0200

-10 0 10 20 30 40 50 60 70 80

R2, 1

/S

Dose, Gy

Calibration Curve

Page 87: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

72

It can be inferred from this calibration curve that gel dosimeters used in this study

showed meager to no dose response of T2 relaxation time up to 40 Gy. However, the gel

dosimeters showed relatively linear dose response changed between 40 Gy to 70 Gy.

Beyond 70 Gy the response of the gel dosimeters seems untrustworthy with reference

to the standard deviation as shown in Figure 16C.

4.2 Au-198 nanoparticles

Radioactive gold nanoparticles of Au-198 with 800 µCi in

volume of 200 µl were injected with 25 gauge needles in the

gel dosimeters in the controlled facility. The red spot at the

middle of the dosimeter indicate localized nanoparticles as

shown in Figure 18.

Figure 18 – Radioactive nanoparticles injected in gel dosimeter. The red spot in the center is nanoparticles spread volume, was approximately about 0.5 cm in diameter.

Page 88: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

73

4.2.1 MR Imaging

The dosimeters after injecting were quarantined for at least 2-3 half lives to ensure

some dose deposition for quantification. The dosimeter was then scanned on GE 1.5T

MR scanner with GP Flex® coil. A FRFSE (Fast Relaxation Fast Spin Echo) pulse sequence

was applied with the MR protocol as explained earlier. T2 relaxation time was acquired

in steps of 15, 30, 45 and 60 ms echo times. The dark contrast region as seen in Figure

19 is the dose distribution of radioactive gold nanoparticles of Au-198 from 800 µCi of

activity. This is significant information for dose mapping, from the complexity involved

with Au-198 nanoparticles in dose quantification.

Region R shown in figure 19 is one another important information about the material

backflow effect. It was observed that radioactive nanoparticles at the time of injection,

pose high probability of backflow into the port of injection. This can be avoided with

application of gentle pressure on the syringe. There was an attempt made to quantify

the radiation absorbed dose due to Au-198 nanoparticles. It was observed that the gel

dosimeters had threshold dose response limits around 70 Gy. The mean T2 relaxation

time of the gel dosimeter measured with different intervals of echo times (15, 30, 45,

and 60 ms) in the region of interest (circled in Figure 19) from Au-198 nanoparticles is

plotted with 0 Gy, 70 Gy responses are shown in Figure 20.

Page 89: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

74

Figure 19 – MR Image in Sagittal section with T2 weighted imaging with FRFSE pulse sequence. The dosimeter, cylindrical shaped object with apex was held between saline bags for better field response. The region marked inside the circle is the dose response from radioactive nanoparticles (Au-198). The dose response in region R is due to particle backflow during injection.

Page 90: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

75

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 15 ms

0 Gy

70 Gy

Au-198

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 re

laxa

tion

tim

e, m

s

Echo time TE = 30 ms

0 Gy

70 Gy

Au-198

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 60 ms

0 Gy

70 Gy

Au-198

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 45 ms

0 Gy

70 Gy

Au-198

Figure 20 – T2 relaxation time plots of dosimeters at different echo times of 0 Gy, 70 Gy and unknown dose (Au-198 nanoparticles). From these plots it can be deduced that the radiation absorbed dose from Au-198 nanoparticles is >70 Gy. However it was difficult to measure that dose precisely beyond the threshold limit ~ 70 Gy of gel dosimeter.

This comparison reveals that dose deposited in the gel dosimeter across all of the echo

times is >70 Gy. Dosimeters that have higher threshold limits as shown in Figure 20

could be better choice for such quantifications.

Page 91: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

76

4.2.2 SPECT Imaging

Au-198 nanoparticles in liquid form, was prepared in dilution with non radioactive gold

nanoparticles of the same size. This was necessary to validate the required minimum

volume of injection. After injection in to the gel dosimeters, Au-198 nanoparticles

diffused in the gel dosimeter along with non radioactive nanoparticles, which were

added to match the required number of gold concentration (~33µg/ml). Attempt was

made to acquire dataset with SPECT images from high resolution gamma camera as

shown in Figure 21. The SPECT image of Au-198 in dosimeter was acquired peaking

gamma energy at 64 different projections

around 360⁰ to the gel dosimeter . The

radioactive hot spots were observed and

recorded.

Figure 21- SPECT image of Au198 in the gel dosimeter.

Page 92: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

77

Figure 22 – Sinogram for Au-198 nanoparticles. (Left) The Sinogram for the full data set. (Right) 16 sets of Sinogram, each representing set of 4 projections.

One commonly used way to display a full set of projection data is in the form of a 2

dimensional matrix formed by r (location in the projection) and θ (rotation angle of the

projection). The Sinogram shown in the Figure 22 provides a convenient way to

represent the full set of data acquired during a scan as shown. The artifacts during data

acquisition of Au-198 are distinct in set of Sinograms.

In retrospect, SPECT image is significant information about radioactivity distribution as it

discriminates between the cold and hot spots of radioactivity in the gel dosimeters.

Further, with the aid of image registration technique, it could be mapped along with MR

images to signify dose quantification.

4.2.3 MCNP Simulations

Au-198 isotope is both gamma and beta emitter. For MCNP model that mimics

Page 93: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

78

brachytherapy of Au-198 nanoparticles interaction of both gammas and betas with

tissue are considered. To keep simpler and efficient, MCNP code for Au198 was

simulated for photon transport and electron transport separately. Volumetric source

considered were measured from the MR images for both Au198 and Au199 models.

MCNP model of Au198 representing as volumetric source was estimated to a sphere of

radius 3.16 mm. Similarly, MCNP model of Au199 representing as volumetric source was

estimated to a sphere of radius 10.7 mm. Dose profiles of both betas and gammas for

both of the MCNP models were created to estimate the radiation interactions.

Individual cells in concentric spheres in steps of 0.05 cm radial increments were used for

purpose of MCNP simulation tally.

4.2.3.1 Photon transport

In photon transport, the source is isotropic and volumetric geometry. Both photons and

photon generated electrons were transported using MODE P E card. F6 Tally, tallies the

amount of energy deposited in 𝑀𝑒𝑉𝑔

was defined in the MCNP input deck (code

sequence) to calculate dose deposited per photon in all cells in question. With this

information average dose rate can be computed for any Curies of radioactivity that

emits gamma radiation. A dose rate profile as shown in Figure 23 demonstrates dose

rate in cGy/hr from center of the source towards the outer surface of the dosimeter (~

1.3 cm). Dose falls off sharply towards the end of diffused length (marked region in

Fig.19).

Activity of Au-198 injected = 800 µCi

Page 94: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

79

= 1.0229 ×1010 photons.hr-1 @ gamma rays abundance = 1.0

Figure 23 – Dose Rate of Au198 photon transport that includes other secondary particles born with photon interactions.

4.2.3.2 Electron transport

In electron transport, the source considered is isotropic and spherical volume in

geometry. Electrons are transported with MODE e card in the MCNP input deck. *F8

Tally card is called, that tallies the total amount of energy deposited by electrons in

MeV. Small detectors of radius 100 microns were placed at 0, 0.4, 0.8, 1.2, 1.6, 2, 2.75,

3.25 and 3.75 mm to calculate the amount of energy deposited by electrons in the cell

volumes along one of the axis starting from the center of the source through the

0.000

1.000

2.000

3.000

4.000

5.000

6.000

0 0.1 0.2 0.3 0.4

Dos

e ra

te, c

Gy/

hr

Distance, cm

Dose rate profile of Au198 gammas

Page 95: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

80

periphery of the source. The source strength of 800 micro Curies determined the dose

rate of electrons at those intervals. A dose rate profile representing all of the point

detectors is shown in Figure 24.

Activity of Au-198 injected = 800 µCi

= 3.5129 ×1010 electrons.hr-1

*F8 tally results obtained from MCNP simulation was the amount of energy deposited in

the gel dosimeters from the surface of the source.

Figure 24 – Dose Rate of Au198 electron transport. The homogeneous volumetric source of 0.316 cm radius was used. The dose rates shown in the plot are measured from the center of the source sphere formed with nanoparticles diffusion.

0

1

2

3

4

5

6

7

0 0.1 0.2 0.3 0.4

Dos

e ra

te, c

Gy/

hr

Distance, cm

Dose rate profile of Au198 betas

Page 96: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

81

4.3 Au-199 nanoparticles

Radioactive gold nanoparticles of Au-199 with 500 µCi in volume of 200 µl were injected

with 25 gauge needles in the gel dosimeters in the controlled facility. Physical

appearance of red spot (similar to Figure18) at the middle of the dosimeter transformed

the reactants (methacrylic acid) to dense polymeric material as shown in Figure 25. This

was chosen to demonstrate the effect of radiation from Au-199 nanoparticles on the gel

dosimeter. The bright white material (polymerization in process) at the middle of the

dosimeter shown in Figure 25 indicates radiation interactions of Au-199 nanoparticles

with methacrylic acid molecules. This transformation was pictured after 24 hrs of

exposure from the day of injection.

Figure 25 – Photograph taken after 1 day of radioactive nanoparticles of Au-199 injected in gel dosimeter.

Page 97: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

82

4.3.1 MR Imaging

The dosimeters after injecting were quarantined for at least 2 half lives. This was to

allow radiation to deposit some dose in the gel dosimeter and also from the radiation

safety perspective. The dosimeter was then scanned on GE 1.5T MR scanner using GP

Flex® coil for T2 weighted imaging with FRFSE pulse sequence. T2 relaxation time was

acquired in steps of 15, 30, 45 and 60 ms echo times. The dark contrast region as shown

in Figure 26 is the dose distribution of radioactive gold nanoparticles of Au-199 from

500 µCi of activity. This is significant information for dose mapping, from the complexity

involved with Au-199 nanoparticles in dose quantification.

The large circular region inside the dosimeter as shown in Figure 26 is the region of

interest where radioactive materials were injected. The black spot at the center is due

to air bubble. Air bubble was deliberately introduced in the dosimeter as a marker.

However, the bright spot surrounding the marker could be due to oxygen scavenging by

THPC or it could be due to localization of non radioactive nanoparticles in that region

which are proportionally more in percentage (1000:1) to radioactive nanoparticles.

Region R shown in Figure 26 is one another important information about the material

backflow effect as explained earlier with Au-198 nanoparticles. Unlike, Au-198 there

was sharp drop in T2 relaxation times with predefined echo times. Due to heavy

interactions of radiation the gel dosimeters may past their threshold dose response

limits (~ 70 Gy) and could not be validated for dose measurements. However the region

Page 98: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

83

of dose deposits provides important information that can be used to develop dose

maps. The mean T2 relaxation time of the gel dosimeter measured with different

intervals of echo times (15, 30, 45, and 60 ms) in the region of interest (circled in Figure

26) from Au-199 nanoparticles is plotted along with 0 Gy, 70 Gy responses are shown in

Figure 27.

Page 99: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

84

Figure 26 – MR Image in Sagittal section with T2 weighted imaging with FRFSE pulse sequence. The dosimeter, cylindrical shaped object with apex was held between saline bags for better field response. The region marked inside the circle is the dose response from radioactive nanoparticles (Au-199). The dose response in region R is due to particle backflow during injection.

Page 100: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

85

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 15 ms

0 Gy

70 Gy

Au-199

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 re

laxa

tion

tim

e, m

s

Echo time TE = 30 ms

0 Gy

70 Gy

Au-199

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 45 ms

0 Gy

70 Gy

Au-199

0.00

200.00

400.00

600.00

800.00

1000.00

1200.00

1400.00

1600.00

T2 r

elax

atio

n ti

me,

ms

Echo time TE = 60 ms

0 Gy

70 Gy

Au-199

Figure 27 – T2 relaxation time plots of dosimeters at different echo times of 0 Gy, 70 Gy and unknown dose (Au-199 nanoparticles). From these plots it can be deduced that the radiation absorbed dose from Au-199 nanoparticles is >70 Gy. However it was difficult to measure that dose precisely beyond the threshold limit ~ 70 Gy of gel dosimeter.

This comparison reveals that dose deposited in the gel dosimeter across all of the echo

times was found to be >70 Gy. However, dosimeters that have higher threshold limits

could be of better choice for such quantifications.

Page 101: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

86

4.3.2 SPECT imaging

Au-199 has 150 keV photons with 40% abundance. The activity locations of Au-199

nanoparticles within the gel dosimeters can be peaked with LEAP (Low Energy All

Purpose) collimator. However, mixed species of Au-199 nanoparticles with Au-198

nanoparticles was observed after neutron activation. Separating these species is not

part of this research but evaluations were. SPECT imaging peaked with high energy

collimator for Au-199 nanoparticles estimates the proportions of Au-198 mix, through

the number of counts received by the detector. It was estimated that Au-198 to Au-199

was in the ratio of 1:8. This measuring step with SPECT also serves in the quality

assurance of Au-199 nanoparticles applications. SPECT image shown in Figure 28

estimates the activity locations of Au-199 in the gel dosimeter.

Figure 28 - SPECT image of Au-199 in the gel dosimeter.

Page 102: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

87

Figure 29 – Sinogram for Au-199 nanoparticles.

The Sinogram shown in the figure 29 provides a convenient way to represent the full set

of data acquired during a scanning process. The artifacts during data acquisition of

Au199 can be observed in 16 different Sinogram sets. Each set represents data from 4

projection locations.

4.3.3 MCNP models

4.3.3.1 Photon transport

In photon transport of Au199, the source is assumed isotropic and is spherical volume in

geometry. The source radius of 10.7 mm was derived from the MR image as marked in

Figure 26. Both photons and photon generated electrons were transported using MODE

P E card. F6 Tally card use in the MCNP code sequence calculates the dose deposited by

average photons at points of interest within the defined geometry. The source strength

of 500 micro Curies determined the dose rate of photons at 0, 0.4, 0.8, 2, 5, 10, 11.4,

Page 103: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

88

11.7 and 12.5 mm intervals with point detectors of radius 100 microns. F6 Tally results

of average Rads/photon from Au199 photon spectra > 50 keV is converted to cGy/hr. A

dose rate profile (Figure.30) that represents all of those point detectors placed in one

direction (x-axis) from the center of the source through the periphery of the source.

Activity of Au-199 injected = 500 µCi = 1.3320 ×1010 photons.hr-1

Figure 30 – Dose rate profile of Au199 photon transport. The homogeneous volumetric source of 10.7 mm radius was used. The dose rates shown in the plot are measured from the center of that source in regular intervals through the periphery of the source.

4.3.3.2 Electron transport

In electron transport of Au199, the source considered is isotropic and spherical volume

in geometry. Electrons are transported with MODE e card in the MCNP input deck. *F8

Tally is called, that tallies the total amount of energy deposited by electrons in MeV.

0

0.05

0.1

0.15

0.2

0.25

0 0.2 0.4 0.6 0.8 1 1.2 1.4

Dos

e ra

te, c

Gy/

hr

Distance, cm

Dose rate profile of Au199 photons

Page 104: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

89

Small detectors of radius 100 microns were placed at 0, 0.4, 0.8, 2, 5, 10, 11.4, 11.7 and

12.5 mm to calculate the amount of energy deposited by electrons in the cell volumes in

one direction starting from the center of the source through the periphery of the

source. The source strength of 500 micro Curies determined the dose rate of electrons

at those intervals. A dose rate profile representing all of the point detectors is shown in

Figure 31.

Activity of Au-199 injected = 500 µCi = 1.1988 ×1010 electrons.hr-1

*F8 tally results obtained from MCNP simulation was the amount of energy

deposited in the gel dosimeters from the surface of the source.

Figure 31 – Dose rate profile of Au199 electron transport. The homogeneous volumetric source of 10.7 mm radius was used. The dose rates shown in the plot are measured from the center of that source in regular intervals through the periphery of the source.

0

20

40

60

80

100

120

0 0.5 1 1.5

Dos

e ra

te, c

Gy/

hr

Distance, cm

Dose rate profile of Au199 betas

Page 105: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

90

CHAPTER 5

QUANTIFICATION OF CHROMIUM-51 DOSE

2.3.1 Cr-51 isotope

51Cr is radionuclide, which has a half life of 27.7 days. This is a synthetic isotope and it

can be synthesized only in the nuclear reactor. 51Cr has gamma energy of 320 keV with

abundance of 10%. This isotope may be considered as both high dose rates (HDR) and

low dose rates (LDR) temporary or permanent implants in brachytherapy treatments.

Permanent implants also called ‘brachy seeds’ and are usually packed with radioactive

material in the cylindrical shells made of stainless steel with the dimensions of 0.8mm in

diameter and 5mm long.[80] There exist a long list of various seed designs but most of

them are in close match to the dimensions specified. The chromium filled seeds can be

made with a simple technique and they can be neutron activated in the nuclear reactor.

To obtain chromium targets either enriched or chromium alloys can be considered. The

alloys that qualify for activation in the nuclear reactor should strictly have contents that

if get activated will not yield long lived isotopes. Basically, for activation 50Cr is the

required as a target isotope. Natural chromium has ~4.345% of 50Cr, which can be

considered for activation. The absorbing cross section for 50Cr is 3.1 barns. The other

composites in the natural chromium are 84% of Cr-52 and 9.5% of Cr-53 which are

stable and there is no stray radiation contamination from them. These seeds could be

activated at Missouri University Research Reactor (MURR). The MU reactor has neutron

Page 106: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

91

flux of 8×1013 𝑛𝑐𝑚2.𝑠

. Neutron irradiation process shall be very brief to get enough

radioactivities that are within safe limits. The radioactivity should be approximately

about 40mCi, it is about 0.4 mCi is standard value currently used for iodine-125(I-125)

brachytherapy seeds that are used for LDR implants[81].

MCNP Simulations→ Photon transport

The dose profile shown in the following graph (Figure 32) is plotted in cGy/hr against the

distance from the source in centimeters, demonstrates the dose rate characteristics.

The radioactivity selected depends upon the dose rate at least 50 cGy/hr at the distance

of 0.5 cm to achieve the effective radiobiological response. Evaluations suggest that at

least 100 mCi of radioactivity per seed would deliver the radiation therapeutic doses

that have radiobiological effective dose significance. Both photons and photon

generated electrons due to secondary photonic interaction in the medium are tallied in

the regions at intervals of 0.72, 1, 2, 3, 4, 5 cm to determine the dose rate.

Page 107: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

92

Figure 32 - Dose rate profile of Cr-51 photon transport.

59.37

36.58

11.01

5.18 2.99

1.53 0.00

10.00

20.00

30.00

40.00

50.00

60.00

70.00

0 1 2 3 4 5 6

Dos

e Ra

te, c

Gy/

hr

Distance, cm

Dose Rate Profile of Cr51 gammas

Page 108: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

93

APPENDIX

Particle Sedimentation Technique

Introduction:

The CPS Disc Centrifuge is a particle size analyzer for measuring

particles in the range of 10 nanometers to 40 microns. The analyzer measures particle

size distributions using centrifugal sedimentation within an optically clear spinning disc

that is filled with fluid. (Figure 33) Sedimentation is stabilized by a density gradient

within the fluid, and accuracy of the measured sizes is insured through the use of a

known size calibration standard before each test. The concentration of particles at each

size is determined by continuously measuring the turbidity of the fluid near the outside

edge of the rotating disc. The turbidity measurements are converted to a weight

distribution using Mie Theory light scattering calculations. The weight distribution is

converted to a surface area or number distribution if required.

Figure 33-CPS schematics of operating principles.

Principle of operation: [82] [82] [82]

The rate of sedimentation inside the rotating disc is controlled by four factors: the size

of the particles, the density difference between the particles and the fluid through

Page 109: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

94

which they are passing, the viscosity of the fluid, and the strength of the centrifugal field

(rotational speed). The sedimentation of particles in a gravitational field was first

systematically investigated by Sir G. G. Stokes (Mathematical and Physical Papers, 11).

Stokes built upon earlier work by Newton (Principia, Lib 11, Loc. Cit. 21) and Rayleigh

(Scientific Papers, 6, Art. 392) which described how drag force on a spherical particle

moving through a fluid depends upon the diameter of the particle, the viscosity of the

fluid, and the velocity of movement. Stokes showed that when particles settle in a

gravitational field under a certain set of conditions, the forces acting on the particle are

in perfect balance, and the particle moves at a constant velocity (which can be

predicted) after a very brief period of initial acceleration. The required conditions are:

1. The particle must be smooth, spherical, and rigid enough to not deform due to the forces acting on it.

2. The particle must be small compared to the container of fluid: the fluid must be essentially infinite in size compared to the size of the particle.

3. The particles which make up the fluid (that is, the molecules) must be much smaller than the settling particle, so that the fluid is essentially homogeneous in how it acts on the particle. (Brownian motion of small particles is the first evidence of non-homogeneous interaction between the fluid and the particle.)

4. The settling speed must be slow enough that all viscous forces come from smooth (non-turbulent) flow. These conditions are normally satisfied for most samples run on the CPS Disc Centrifuge using normal operating parameters. There are two fairly common situations where these conditions are not completely satisfied.

Results: The distribution can be displayed as a weight distribution (weight of

particles plotted against diameter), surface area distribution (surface area of the

particles plotted against diameter), number distribution (number of particles

Page 110: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

95

plotted against diameter), and absorption (light absorbed/scattered plotted

against diameter). The absorption distribution is the raw data that comes from

the instrument as shown in Figure 34.

Figure 34 - Overlay of particulate distributions by CPS disc.

Particle sizes obtained from conjugated nano particles are shown in Table-VII, below.

The most preferred and universally accepted standard method for the particle size of

nanoparticles is by TEM (Transmission Electron Microscopy). The particle size

distribution determined through the particle sedimentation method is in agreement

Page 111: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

96

with the electron microscopy. However, values do not agree Bomesin-gold. This is

because; the sizes determined from the particle sedimentation technique do not show

core sizes. The larger size is due to surface coating of gold by Bombesin peptide.

TABLE-VII, COMPARISON OF CPS DERIVED PARTICLE AVERAGE SIZES TO TEM DERIVED PARTICLE AVERAGE SIZES OF DIFFERENT TYPES OF THE GOLD NANOPARTICLES IN NANOMETERS

Sample TEM, nm CPS, nm

Soy-gold 15±5 17±1

Tea-gold 30±5 25±1

Cumin-gold 13±5 12±1

Cinnamon-gold 13±5 32±1

Bombesin-gold 16±5 125±1

Page 112: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

97

BIBLIOGRAPHY

1. Ross, K. and S. Vargo, Play-Doh animal fun. 2002, Brookfield, Conn.: Millbrook Press. 48 p.

2. Inc., H., The manufacturer of Play Doh clays for kids. www.hasbro.com, 2010. 3. Siegel, R., D. Naishadham, and A. Jemal, Cancer statistics, 2012. CA Cancer J Clin,

2012. 62(1): p. 10-29. 4. Van Dyk, J., The Modern Technology of Radiation Oncology. 1 ed. 1999: Medical

Physics Publishing. 1. 5. Dyk, J.V., The Modern Technology of Radiation Oncology. 6. Pieters, B.R., et al., Treatment results of PDR brachytherapy combined with

external beam radiotherapy in 106 patients with intermediate- to high-risk prostate cancer. Int J Radiat Oncol Biol Phys, 2011. 79(4): p. 1037-42.

7. Polo, A., Pulsed dose rate brachytherapy. Clin Transl Oncol, 2008. 10(6): p. 324-33.

8. McDermott, M.W., P.K. Sneed, and P.H. Gutin, Interstitial brachytherapy for malignant brain tumors. Semin Surg Oncol, 1998. 14(1): p. 79-87.

9. Obasi, P.C., G.H. Barnett, and J.H. Suh, Brachytherapy for intracranial meningioma using a permanently implanted iodine-125 seed. Stereotact Funct Neurosurg, 2002. 79(1): p. 33-43.

10. Trnkova, P., et al., New inverse planning technology for image-guided cervical cancer brachytherapy: description and evaluation within a clinical frame. Radiother Oncol, 2009. 93(2): p. 331-40.

11. Newman, L.A. and T.A. Washington, New trends in breast conservation therapy. Surg Clin North Am, 2003. 83(4): p. 841-83.

12. Barus, C., Penetrating Radiation Associated with the X-Rays. Science, 1905. 21(537): p. 561-6.

13. Strong, W.W., The Penetrating Radiation. Science, 1908. 27(687): p. 351-3. 14. Schaeberle, J.M., The Law of Radiation. Science, 1909. 29(736): p. 227-8. 15. Proenca, M.P., et al., Size and surface effects on the magnetic properties of NiO

nanoparticles. Phys Chem Chem Phys, 2011. 13(20): p. 9561-7. 16. Thakor, A.S., et al., Gold nanoparticles: a revival in precious metal administration

to patients. Nano Lett, 2011. 11(10): p. 4029-36. 17. Sun, Y., et al., Rapid and quantitative quality control of microarrays using cationic

nanoparticles. Biotechnol Bioeng, 2009. 102(3): p. 960-4. 18. Shukla, R., et al., Soybeans as a phytochemical reservoir for the production and

stabilization of biocompatible gold nanoparticles. Small, 2008. 4(9): p. 1425-36. 19. Nune, S.K., et al., Green Nanotechnology from Tea: Phytochemicals in Tea as

Building Blocks for Production of Biocompatible Gold Nanoparticles. J Mater Chem, 2009. 19(19): p. 2912-2920.

Page 113: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

98

20. Katti, K., et al., Green Nanotechnology from Cumin Phytochemicals: Generation of Biocompatible Gold Nanoparticles. Int J Green Nanotechnol Biomed, 2009. 1(1): p. B39-B52.

21. Chanda, N., et al., Bombesin functionalized gold nanoparticles show in vitro and in vivo cancer receptor specificity. Proc Natl Acad Sci U S A, 2010. 107(19): p. 8760-5.

22. Chanda, N., et al., An effective strategy for the synthesis of biocompatible gold nanoparticles using cinnamon phytochemicals for phantom CT imaging and photoacoustic detection of cancerous cells. Pharm Res, 2011. 28(2): p. 279-91.

23. Katti, K.K., et al., Facile and General Method for Synthesis of Sugar Coated Gold Nanoparticles. Int J Green Nanotechnol Biomed, 2009. 1(1): p. B53-B59.

24. Graham, J.S., Y. Miron, and M. Grandbois, Assembly of collagen fibril meshes using gold nanoparticles functionalized with tris(hydroxymethyl)phosphine-alanine as multivalent cross-linking agents. J Mol Recognit, 2011. 24(3): p. 477-82.

25. Aziz, M.S., et al., Gold nanoparticle trapping and delivery for therapeutic applications. Int J Nanomedicine, 2012. 7: p. 11-7.

26. Park, J., et al., Intra-organ Biodistribution of Gold Nanoparticles Using Intrinsic Two-photon Induced Photoluminescence. Lasers Surg Med, 2010. 42(7): p. 630-639.

27. Elbialy, N., M. Abdelhamid, and T. Youssef, Low power argon laser-induced thermal therapy for subcutaneous Ehrlich carcinoma in mice using spherical gold nanoparticles. J Biomed Nanotechnol, 2010. 6(6): p. 687-93.

28. Mobit, P. and I. Badragan, An evaluation of the AAPM-TG43 dosimetry protocol for I-125 brachytherapy seed. Phys Med Biol, 2004. 49(14): p. 3161-70.

29. Laurie M. Unger*, D.K.T., OAK RIDGE NATIONAL LABORATORY, 1982. 30. Chithrani, B.D. and W.C. Chan, Elucidating the mechanism of cellular uptake and

removal of protein-coated gold nanoparticles of different sizes and shapes. Nano Lett, 2007. 7(6): p. 1542-50.

31. Yang, P.H., et al., Transferrin-mediated gold nanoparticle cellular uptake. Bioconjug Chem, 2005. 16(3): p. 494-6.

32. Mizuno, H., et al., Theaflavin-3, 3'-digallate induces epidermal growth factor receptor downregulation. Mol Carcinog, 2006. 45(3): p. 204-12.

33. Na, H.K. and Y.J. Surh, Intracellular signaling network as a prime chemopreventive target of (-)-epigallocatechin gallate. Mol Nutr Food Res, 2006. 50(2): p. 152-9.

34. Liebman, M.A., et al., Generation and preliminary characterization of an antibody library with preferential reactivity to human colorectal cancer cells as compared to normal human blood cells. Immunol Lett, 2004. 91(2-3): p. 179-88.

35. Benhar, M., et al., Enhanced ROS production in oncogenically transformed cells potentiates c-Jun N-terminal kinase and p38 mitogen-activated protein kinase activation and sensitization to genotoxic stress. Mol Cell Biol, 2001. 21(20): p. 6913-26.

Page 114: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

99

36. Kannan, R., et al., Functionalized radioactive gold nanoparticles in tumor therapy. Wiley Interdiscip Rev Nanomed Nanobiotechnol, 2012. 4(1): p. 42-51.

37. Afrasiabi, Z., et al., Nanoscale sensor design via in situ labeling of gold nanoparticles onto protein scaffolds. J Nanosci Nanotechnol, 2010. 10(2): p. 719-25.

38. Bondarkov, M.D., et al., Method for simultaneous 90Sr and 137Cs in-vivo measurements of small animals and other environmental media developed for the conditions of the Chernobyl exclusion zone. Health Phys, 2011. 101(4): p. 383-92.

39. Baechler, S., N. Stritt, and F.O. Bochud, Individual monitoring of internal exposure for nuclear medicine workers in Switzerland. Radiat Prot Dosimetry, 2011. 144(1-4): p. 464-7.

40. Lin, M.H., et al., Three-dimensional dosimetry in brachytherapy: A MAGAT study. Appl Radiat Isot, 2009. 67(7-8): p. 1432-7.

41. Fong, P.M., et al., Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere. Phys Med Biol, 2001. 46(12): p. 3105-13.

42. Lepage, M., et al., Modelling of post-irradiation events in polymer gel dosimeters. Phys Med Biol, 2001. 46(11): p. 2827-39.

43. Slessinger, E.D., Practical considerations for prostate HDR brachytherapy. Brachytherapy, 2010. 9(3): p. 282-7.

44. Dattoli, M. and K. Waller, A simple method to stabilize the prostate during transperineal prostate brachytherapy. Int J Radiat Oncol Biol Phys, 1997. 38(2): p. 341-2.

45. Weitmann, H.D., et al., Pilot study in the treatment of endometrial carcinoma with 3D image-based high-dose-rate brachytherapy using modified Heyman packing: clinical experience and dose-volume histogram analysis. Int J Radiat Oncol Biol Phys, 2005. 62(2): p. 468-78.

46. Simon, N., S.M. Silverstone, and L.C. Roach, Afterloading Heyman applicators. Acta Radiol Ther Phys Biol, 1971. 10(2): p. 231-8.

47. Bond, M.G., et al., Dosimetric considerations in the treatment of inoperable endometrial carcinoma by a high dose rate afterloading packing technique. Clin Oncol (R Coll Radiol), 1997. 9(1): p. 41-7.

48. Chun, M., et al., Uterine thickness determination using real-time ultrasonography: a guide for intracavitary brachytherapy in the treatment of endometrial carcinoma. Gynecol Oncol, 1990. 36(2): p. 176-80.

49. Saylor, W.L. and M. Dillard, Dosimetry of 137Cs sources with the Fletcher-Suit gynecological applicator. Med Phys, 1976. 3(2): p. 117-9.

50. Decker, W.E., et al., Comparison of traditional low-dose-rate to optimized and nonoptimized high-dose-rate tandem and ovoid dosimetry. Int J Radiat Oncol Biol Phys, 2001. 50(2): p. 561-7.

51. Joshi, P., et al., The anticancer activity of chloroquine-gold nanoparticles against MCF-7 breast cancer cells. Colloids Surf B Biointerfaces, 2012. 95: p. 195-200.

Page 115: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

100

52. Shao, X., et al., 125I-labeled gold nanorods for targeted imaging of inflammation. ACS Nano, 2011. 5(11): p. 8967-73.

53. Fricke, H., S. Morse., The chemical action of roentgen rays on dilute ferrous sulfate solutions as a measure of dose. Am. J. Roent. Rad. Ther., 1927. 18: p. 430-432.

54. Day, M.J., Radiation dosimetry using nuclear magnetic resonance: an introductory review. Phys Med Biol, 1990. 35(12): p. 1605-9.

55. Schulz, R.J., et al., Dose-response curves for Fricke-infused agarose gels as obtained by nuclear magnetic resonance. Phys Med Biol, 1990. 35(12): p. 1611-22.

56. Does, M.D. and J.C. Gore, Rapid acquisition transverse relaxometric imaging. J Magn Reson, 2000. 147(1): p. 116-20.

57. Gore, J.C., M.J. Maryanski, and R.J. Schulz, Test objects for MRI quality assurance based on polymer gels. Med Phys, 1997. 24(9): p. 1405-8.

58. Shin-ichiro Hayashi, M.Y., Shuji Usui, Kiyofumi Haneda and Takahiro Tominaga, The role of gelatin in a methacrylic acid based gel dosimeter. Journal of Physics: Conference Series 164, 2009.

59. De Deene, Y., et al., A basic study of some normoxic polymer gel dosimeters. Phys Med Biol, 2002. 47(19): p. 3441-63.

60. Bayreder, C., et al., Basic investigations on the performance of a normoxic polymer gel with tetrakis-hydroxy-methyl-phosphonium chloride as an oxygen scavenger: reproducibility, accuracy, stability, and dose rate dependence. Med Phys, 2006. 33(7): p. 2506-18.

61. Chanda, N., et al., Bombesin functionalized gold nanoparticles show in vitro and in vivo cancer receptor specificity. Proc Natl Acad Sci U S A. 107(19): p. 8760-5.

62. Agasti, S.S., et al., Photoregulated release of caged anticancer drugs from gold nanoparticles. J Am Chem Soc, 2009. 131(16): p. 5728-9.

63. Williamson, J., et al., Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sources. Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee. Ad Hoc Subcommittee of the Radiation Therapy Committee. Med Phys, 1998. 25(12): p. 2269-70.

64. Clancy, B.E., Slabbo- a discrete ordinate neutron transport code in plane geometry with anisotropic scattering. Aaec/Tm 505. 1969, Lucas Heights,: Australian Atomic Energy Commission Research Establishment. 8 p.

65. Berger, P.D., 101 special practice problems in probability and statistics. 4th ed. 2008, Lombard, IL: Marsh Publications. xvi, 318 p.

66. Adams, W.J., The life and times of the central limit theorem. 2nd ed. History of mathematics. 2009, Providence, R.I. London: American Mathematical Society ; London Mathematical Society. xii, 195 p. 67. Team, X.-M.C., MCNP — A General Monte Carlo N-Particle Transport Code, Version 5. Los Alamos National Laboratory LA-UR-03-1987, 2005. I & II.

Page 116: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

101

68. Rose, P.F., Compiler and Editor, ENDF-201, ENDF/B-VI Summary Documentation. Brookhaven National Laboratory, October 1991. BNL-NCS-17541.

69. S. C. Frankle, R.C.R., and P. G. Young,, ACTI An MCNP Data Library for Prompt Gamma-ray Spectroscopy. 12th Biennial Radiation Protection and Shielding Topical Meeting, Santa Fe, NM,, April 15-19, 2002.

70. R. J. Howerton, D.E.C., R. C. Haight, M. H. MacGregor, S. T. Perkins, and E. F. Plechaty,, The LLL Evaluated Nuclear Data Library (ENDL): Evaluation Techniques, Reaction Index, and Descriptions of Individual Reactions Lawrence Livermore National Laboratory report UCRL-50400 1975. Vol. 15, Part A.

71. D. E. Cullen, M.H.C., J. H. Hubbell, S. T. Perkins, E. F. Plechaty, J. A. Rathkopf, and J. H. Scofield, Tables and Graphs of Photon Interaction Cross Sections from 10 eV to 100GeV Derived from the LLNL Evaluated Photon Data Library (EPDL) Lawrence Livermore National Laboratory report UCRL-50400, 1989. Volume 6, Rev. 4, Part A: Z = 1 to 50 and Part B: Z = 51 to 100.

72. Howerton, M.A.G.a.R.J., ACTL: Evaluated Neutron Activation Cross–Section Library-Evaluation Techniques and Reaction Index Lawrence Livermore National Laboratory report UCRL-50400, 1978. 18.

73. Young, E.D.A.a.P.G., Evaluated Neutron-Induced Cross Sections for 54,56Fe to 40MeV. Los Alamos Scientific Laboratory report LA-8626-MS (ENDF-304), 1980.

74. D. G. Foster, J.a.E.D.A., Average Neutronic Properties of “Prompt” Fission Products. Los Alamos National Laboratory report LA-9168-MS, Feb 1982.

75. E. D. Arthur, P.G.Y., A. B. Smith, and C. A. Philis, New Tungsten Isotope Evaluations for Neutron Energies Between 0.1 and 20 MeV. Trans. Am. Nucl. Soc., 1981. 39(793).

76. Muir, R.E.M.a.D.W., The NJOY Nuclear Data Processing System Version 91. Los Alamos National Laboratory report LA-12740-M, Oct 1994.

77. R. E. MacFarlane, D.W.M., and R. M. Boicourt, The NJOY Nuclear Data Processing System, Volume I: User’s Manual. Los Alamos National Laboratory report LA-9303-M, May 1982. Vol. I (ENDF-324).

78. American National Standard for Programming Language - Fortran - Extended. American National Standards Institute, New York, NY, 1992. ANSI X3.198-1992.

79. K. F. Eckerman, R.J.W., J. C. Ryman, and M. Cristy, Availability of Nuclear Decay Data in Electronic Form Including Beta Spectra not Previously Published. Health Phys. , 1994. 67(4): p. 338-345.

80. Rivard, M.J., et al., Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med Phys, 2004. 31(3): p. 633-74.

81. Masucci, G.L., et al., Comparison between high and low source activity seeds for I-125 permanent seed prostate brachytherapy. Int J Radiat Oncol Biol Phys. 78(3): p. 781-6.

82. Day, T.W., Braxton Hicks vs. preterm labor. Minn Med, 1990. 73(8): p. 21-2.

Page 117: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

102

Acknowledgements

Work presented in thesis about radioactive gold nanoparticles is supported by the

following grants:

• National Cancer Institute grant on Cancer Nanotechnology Platform No.: 1RO1CA119412-01.

• Intramural grants from the University of Missouri and Phase II NIH SBIR contract #: HHSN261201000100C; PHS 2008-1(RFA).

Page 118: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

103

VITA

Rajesh R Kulkarni was born in Dharwad in Karnataka state of India, on 14th September

1972. After graduating from Sarvodaya Vidyalaya high school in Khanapur, Karnataka,

then he was educated at the Pre-University Course at the state education board of

Karnataka in India. He was further educated at the Karnataka University in Dharwad, in

Karnataka state of India, where he received a Bachelor of Engineering degree in

Mechanical Engineering in January of 1995. Thereafter, he was employed between 1995

and 2004 with different types of engineering industries that were involved in electrical

and electronics manufacturing, and also one of them in remodeling of marine

combustible engines. 8 years being with Siemens Ltd, he has experiences in materials

supply chain, manufacturing, purchase, inventory control, process planning and quality

management systems. He received his Masters in Nuclear Engineering in July 2008 from

the University of Missouri in Columbia in the state of Missouri from United States of

America. In 2009 he became board eligible candidate in Therapeutic Radiation Physics

with ABR (American Board of Radiology). After receiving a Doctor of Philosophy, in

Nuclear Engineering at the University of Missouri in Columbia in July 2012, he will be

seeking for career opportunities in Therapeutic Radiation Physics or relevant fields in

medical physics. With diversified experiences in industry and academic research he

could be an easier fit in any type of organization.

Page 119: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

104

PUBLICATIONS

1. Ravi Shukla, Nripen Chanda, Ajit Zambre, Anandhi Upendran, Kavita Katti, Rajesh

R. Kulkarni, Satish Kumar Nune, Stan W. Casteel, Charles Jeffrey Smith, Jatin

Vimal, Evan Boote, J. David Robertson,Para Kan, Hendrik Engelbrecht, Lisa D.

Watkinson, Terry L. Carmack, John R. Lever, Cathy S. Cutler, Charles Caldwell,

Raghuraman Kannan and Kattesh V. Katti. Laminin receptor specific therapeutic

gold nanoparticles (198AuNP-EGCg) show efficacy in treating prostate cancer.

PNAS June 25, 2012

2. Raghuraman Kannan, Ajit Zambre, Nripen Chanda, Rajesh Kulkarni, Ravi Shukla,

Kavita Katti, Anandhi Upendran, Cathy Cutler, Evan Boote and Kattesh V. Katti.

Functionalized radioactive gold nanoparticles in tumor therapy. Wires

Nanomedicine and Nanobiotechnology January 2012

3. Nripen Chanda, Ravi Shukla, Ajit Zambre, Swapna Mekapothula, Rajesh R.

Kulkarni, Kavita Katti, Kiran Bhattacharya, Genevieve M. Fent, Stan W. Casteel, ,

Evan J. Boote, John A. Viator, Anandhi Upendran, Raghuraman Kannan, Kattesh

V. Katti. An Effective Strategy for the Synthesis of Biocompatible Gold

Nanoparticles Using Cinnamon Phytochemicals for Phantom CT imaging and

Photoacoustic Detection of Cancerous Cells. Pharmaceutical Research Sept. 13,

2010

4. Chanda N, Kattumuri V, Shukla R, Zambre A, Katti K, Upendran A, Kulkarni RR,

Kan P, Fent GM, Casteel SW, Smith CJ, Boote E, Robertson JD, Cutler C, Lever JR,

Katti KV, Kannan R. Bombesin functionalized gold nanoparticles show in vitro

and in vivo cancer receptor specificity. PNAS May 11, 2010 vol. 107 no. 19 8760-

8765.

Page 120: QUANTIFICATION OF ABSORBED DOSES FROM RADIOACTIVE …

105

5. Katti K, Chanda N, Shukla R, Zambre A, Kulkarni RR, Kannan R, Katti KV. Green

nanotechnology from cumin phytochemicals: generation of biocompatible gold

nanoparticles. Int J Green Nanotechnol. Biomedicine, 2009, 1: B39-B52.

6. Nune SK, Chanda N, Katti K, Mekapothula S, Kulkarni RR, Welshons WV, Kannan

R, Katti KV. : Green Nanotechnology from Tea: Phytochemicals in Tea as Building

Blocks for production of Biocompatible Gold Nanoparticles. Journal of Materials

Chemistry 2009, 19, 2912–2920.

7. Ravi Shukla, Satish K. Nune, Nripen Chanda, Kavita Katti, Swapna

Mekapothula, Rajesh R. Kulkarni, Wade V. Welshons, Raghuraman Kannan,

Kattesh V. Katti, Soybeans as a Phytochemical Reservoir for the Production and

Stabilization of Biocompatible Gold Nanoparticles, Small 2008, p 1425-1436.