Qualification and Validation Basic Principles of GMP.

50
Qualification and Validation Basic Principles of GMP

Transcript of Qualification and Validation Basic Principles of GMP.

Page 1: Qualification and Validation Basic Principles of GMP.

Qualification and Validation

Basic Principles of GMP

Page 2: Qualification and Validation Basic Principles of GMP.

“Validation of an analytical procedure is the process by which it is established, by laboratory studies, that the performance characteristics of the procedure meet the requirements for the intended analyticalapplications.”

Definition of Validation: USP

Page 3: Qualification and Validation Basic Principles of GMP.

A written plan stating how validation will be conducted and defining acceptance criteria.

For example, the protocol for a manufacturing process identifies processing equipment, critical process parameters/operating ranges, product characteristics, sampling, test data to be collected, number of validation runs, and acceptable test results.

Validation Protocol:

Page 4: Qualification and Validation Basic Principles of GMP.

Action of proving and documenting that equipment or ancillary systems are properly installed, work correctly, and actually lead to the expected results. Qualification is part of validation, but the individual qualification steps alone do not constitute process validation.

Qualification

Page 5: Qualification and Validation Basic Principles of GMP.

The purpose of I.Q is to check the installation site/environment, confirms equipment specifications and verifies the condition of installed equipment.

Installation Qualification (I.Q):

Page 6: Qualification and Validation Basic Principles of GMP.

Confirmation of the specifications of the analytical equipment. Confirmation and maintenance of documents (Instruction manuals, qualification protocol and certificates). Confirmation of installation site and conditions Confirmation of delivered equipment. Confirmations of Software and Firmware i.e., verify that the equipment is consistent with actual versions displayed when power is turned on.

I.Q protocol shall include the following:

Page 7: Qualification and Validation Basic Principles of GMP.

O.Q includes procedures and documentation of O.Q of analytical instrument. When all procedures are executed and all items pass the inspection, it is verified that the system operates to satisfy the intended purpose.

Operational Qualification (O.Q):

Page 8: Qualification and Validation Basic Principles of GMP.

Operation check on each unit Due to modular nature of the system, the operation of eachunit is checked properly.Operation check on overall system. Confirm that the system controller and work station control each unit during analysis and that the analysis results meet the prescribed criteria. Software and Firmware check Here Firmware checking isconducted based on version display and Software certificate of Compliance. The Software and Firmware must be properly managed and change procedures must be properly clarified.

O.Q protocol shall include:

Page 9: Qualification and Validation Basic Principles of GMP.

The objective is to ensure that the instrument is performing within specified limits. Hence documented verification that the equipment and ancillary systems, as connected together, can perform effectively and reproducibly based on the approved process method and specifications.

Performance Qualification (PQ):

Page 10: Qualification and Validation Basic Principles of GMP.

The demonstration that a particular instrument or device produces results within specified limits by comparison with those produced by a reference or traceable standard over an appropriate range of measurements.

Definition of Calibration: ICH

Page 11: Qualification and Validation Basic Principles of GMP.

Various Calibrationparameters are: Flow rate accuracy Injector accuracy System Precision Wavelength accuracy Detector linearity Injector linearity Gradient Performance Check Column oven temperature accuracy

Calibration of HPLC:

Page 12: Qualification and Validation Basic Principles of GMP.

1. Prime all the solvent lines with Milli Q water.2. Set the flow rate to 0.500 ml/m.3. Wait for about 15 m to stabilize the system and ensure that the pressure is stable.4. Insert the outlet tubing into a 10 ml volumetric flask and start the stop watch simultaneously.5. Stop the stopwatch when the lower meniscus reaches the 10 ml mark on the flask.6. Record the elapsed time.7. Similarly check the flow for 1.0 ml/mand 2.0 ml/m.

Flow Rate Accuracy:

Acceptance criteria: The time taken tocollect the water should be with in ± 2.0% ofthe actual value.

Page 13: Qualification and Validation Basic Principles of GMP.

1. Connect the pump and detector inletwith union.2. Prepare mobile phase consisting of a mixture of water and Methanol (70:30 v/v)3. Set a flow rate of 0.5 ml/m and a runtime of 1 m.4. Set the column temperature at 25± 2°C.5. Fill a standard HPLC vial to 2/3rd with Milli-Q water. Seal the vial properly with a cap.6. Weigh the vial and record the weight as W1 grams.7. Place the vial in the chromatographic system and perform 6 injections of 50μl volume from this vial. 8. Weigh the vial again and note the weigh after the injections as W2 grams. Calculate the mean volume injected per injection as follows:Mean injected volume (μl) = (W1 – W2)×100/6

Injector Accuracy:

Acceptance criteria: The mean injectedvolume should be 50.0±1.0 μl.

Page 14: Qualification and Validation Basic Principles of GMP.

Standard Preparation: Accurately weigh and transfer about 60mg of Caffeine into a 100ml volumetric flask. Dissolve and dilute to the volume with mobile phase. Transfer 10ml of this solution into a 100ml volumetric flask and dilute to the volume with mobile phase.Procedure: Inject blank, followed by standard preparation in 6 replicates. Note down the areas and retention times. Now calculate the %RSD of retention time and peak areas for 6 replicates injections.Acceptance criteria: The %RSD of retention time & peak area should be <1.0%.

System Precision:

Page 15: Qualification and Validation Basic Principles of GMP.

Select 3D mode and set the wavelength range as 200-400nm.Inject 20 μl of standard preparation once into the chromatographic system. Extract and record the chromatograms at wavelengths of 202 to 208nm with an interval of 1nm and at 269 to 275 nm with an interval of 1nm.Note down the height or absorbance.Acceptance criteria: The maximum absorbance should be at 205±2nm and 272±2nm.

PDA Detector Accuracy:

Page 16: Qualification and Validation Basic Principles of GMP.

Standard Preparation: Accurately weigh and transfer about 60mg of Caffeine into a 100ml volumetric flask. Dissolve and dilute to the volume with mobile phase. Detector linearity solution 1(0.06mg/ml): Transfer 10ml of Standard Preparation into a 100ml volumetric flask and dilute to the volume with mobile phase Detector linearity solution 2(0.048mg/ml): Transfer 8ml of Standard Preparation into a 100ml volumetric flaskand dilute to the volume with mobile phase. Detector linearity solution 3(0.03mg/ml): Transfer 5ml of Standard Preparation into a 100ml volumetric flaskand dilute to the volume with mobile phase. Detector linearity solution 4(0.24mg/ml): Transfer 4ml of Standard Preparation into a 100ml volumetric flaskand dilute to the volume with mobile phase. Detector linearity solution 5(0.012mg/ml): Transfer 2ml of Standard Preparation into a 100ml volumetric flaskand dilute to the volume with mobile phase.Procedure: Inject blank, followed by Detector linearity solutions and record the peak responses of Caffeine standard plot between the concentration Vs the peak responses.Acceptance criteria: The plot should belinear and regression coefficient (R2) shouldnot be less than 0.99.

Detector Linearity:

Page 17: Qualification and Validation Basic Principles of GMP.

Standard Preparation: Accurately weigh and transfer about 60mg of Caffeine into a 100ml volumetric flask. Dissolve and dilute to the volume with mobile phase. Transfer 10ml of Standard Preparation into a 100ml volumetric flask and dilute to the volume with mobile phase.Procedure: Inject 5 μl of the mobile phase as blank injection. Inject 5 μl, 10 μl, 20 μl, 50 μl and 80 μl of the Standard Preparation and record the peak areas. Plot a curve for the volume injected Vs peak area.Acceptance criteria: The plot should be linear and regression coefficient (R2) should not be less than 0.99.

Injector Linearity:

Page 18: Qualification and Validation Basic Principles of GMP.

Add 5ml of acetone to 1000ml of methanol filter and degas. Connect the pump and detector inlet with union. Set the detectorwave length at 254 nm. Place Channels A and C in methanol and channel B and D in 0.5% acetone in methanol. Set binary gradient with a total flow rate of 2.0ml/m.Set gradient program as shown below for channels A, B and C, D individually.

Gradient performance check:

Page 19: Qualification and Validation Basic Principles of GMP.

Purge all the channels at a flow rate of 2ml/m for about 5 m. Set the flow rate at 2.0ml/m and wait until the base line is stable. Set the gradient profile for A and B and run the gradient profile by injecting “0.0”volume of methanol. Record the height of the peaks. Consider the height of the peak resulting from B at 100% concentration as 100 and calculate the percentage height of other peaks. Perform the gradient performance check similarly for channels C and D. Consider the height of the peak resulting from D at 100% concentration as 100 and calculate the percentage height of other peaks.

Acceptance criteria: The calculatedpercentage composition (Height (%) shouldbe within ±1.0% of the set composition.

Page 20: Qualification and Validation Basic Principles of GMP.

It is evaluated with a calibrated digital thermometer at 30°C and 60°C .Place the thermometer probe in the column oven and set the column oven temperature at 30°C.Wait till the temperature stabilizes. Record the temperature displayed on the thermometer. Similarly performs the column oven temperature accuracy test at 60°C.

Acceptance criteria: The resulting oven temperature from the thermometer display should be within ±2°C of the set temperature.

Column Oven Temperature Accuracy:

Page 21: Qualification and Validation Basic Principles of GMP.

Calibration of UV-Visible spectrophotometer

Spectral calibration: visible spectral region Ensure-the socket of the power cord of the instrument is inserted properly-cuvettes are clean Switch ON the instrument. Allow 15m to warm up. Keep dummy cuvette in position of sample holder. Set the λ to 485 nm and press %T button. Press 0%T in appropriate direction to adjust 0.00 reading on read out. Now remove dummy cuvette from sample holder. Close the lid. By adjusting coarse and fine control set a reading of around 80.0 on read out Now set the value of wavelengths in increments of 0.1 nm up to λ of 487 nm and read the value of %T at each increment of λ

Page 22: Qualification and Validation Basic Principles of GMP.

Draw a curve %T Vs λ. If the peak value of %T is occurring at a λ 486.1 ± 0.5 nm, the spectral calibration of the instrument in the visible spectral region is proper.

Page 23: Qualification and Validation Basic Principles of GMP.

U.V spectral region: Keep blank (distilled water) filled cuvette and sample (benzene vapor) filled cuvette. Set the λ to 253 nm and press absorbance button. Adjust blank to 0.000 on the read out by using coarse and fine adjustment Now place the sample into optical path, value of Absorbance of sample at the λ set appears on the read out. Again set the values of wavelengths increments of 0.1nm up to a λ of 255nm.Measure the A at each increment. If maximum A is obtained at λ 253.9 ± 0.5nm the “spectral calibration” of the instrument in U.V region is confirmed to be proper.

Spectral calibration:

Page 24: Qualification and Validation Basic Principles of GMP.

Visible region Place dummy cuvette in sample holder and set %T to “zero”. Now remove dummy cuvette, by using fine & coarse control set a reading exactly 40.0 on the read out. Press Absorbance push button. If the maximum absorbance obtained at λ of 485nm is 0.398 ± 0.002, the photometric calibration of instrument is confirmed to be proper.

Photometric Calibration:Absorbance:

Page 25: Qualification and Validation Basic Principles of GMP.

U.V region Place blank 0.1N H2SO4 cuvette and 60ppm K2Cr2O7 as sample Set λ exactly to 257 nm, if the value of Absorbance of sample at the set λ is 0.864±0.005, the instrument is measuring Absorbance properly.

Absorbance:

Page 26: Qualification and Validation Basic Principles of GMP.

a) Holmium filter: For routine calibrations, holmium filter is satisfactory. Record the absorption spectrum from 500 to 230 nm using slowest scan speed and narrowest slit setting. Identify 3 fused absorption bands centered on 452.2nm and single band at around 360.9nm.Instruments with accurately calibrated λ scales will show λ max at 453.2, 418.4, 360.9, 287.5, 279.4 and 241.5nm.

b) Holmium perchlorate solution: Prepare a solution of Holmium (III) perchlorate by dissolving 0.5g of holmium oxide in 2.4 ml perchloric acid (72% AR grade) by warming gently and diluting to 10ml with water. Record the absorption spectrum from 500- 230nm.The wavelengths of principal bands should be 485.8, 450.8, 416.3, 361.5, 287.1, 278.7, 241.1 nm.

Calibration of Wavelength:

Page 27: Qualification and Validation Basic Principles of GMP.
Page 28: Qualification and Validation Basic Principles of GMP.

Holmium(III) in perchloric acid: Consists of a 4% W/V solution of Holmium Oxide in 1.4M Perchloric Acid. λmax………………241.1 ± 0.4nm 278.0 ± 0.4nm 287.5 ± 0.4nm 361.2 ± 0.4nm 416.6 ± 0.4nm 451.3 ± 0.4nm 458.3 ± 0.4nm 537.5 ± 0.4nm 640.8 ± 0.4nm

Page 29: Qualification and Validation Basic Principles of GMP.

consistently produced

Page 30: Qualification and Validation Basic Principles of GMP.

Procedure

Page 31: Qualification and Validation Basic Principles of GMP.

Broader than SOP

Page 32: Qualification and Validation Basic Principles of GMP.

Whole plan and documents of validation

Page 33: Qualification and Validation Basic Principles of GMP.

Modification monitering

Page 34: Qualification and Validation Basic Principles of GMP.

Parts

Page 35: Qualification and Validation Basic Principles of GMP.

calibration

Page 36: Qualification and Validation Basic Principles of GMP.

Validation of Continuity

Page 37: Qualification and Validation Basic Principles of GMP.

E.g. cleaning validation and analytical method validation.

Page 38: Qualification and Validation Basic Principles of GMP.
Page 39: Qualification and Validation Basic Principles of GMP.

Repeatability (Intra-assay), intermediate precision (inter-assay including inter-day variability) and reproducibility (different labs): in order to trust a procedure the procedure should repeatable results for the same samples. For example testing a standard solution by HPLC method of analysis considered not repeatable if the results are 100%, 80%, 110%, 30% and 99%. However, if the data are far from the real values but still close then the procedure is considered precise. For example, 70%, 69%, 71%, 70% and 70%. Precision is evaluated statistically by relative standard deviation RSD (coefficient of variation CV).the analytical method is usually considered precise if RSD <3.0% according to USP or less than 2.0 according to BP.

Page 40: Qualification and Validation Basic Principles of GMP.

Accuracy: it means the proximity of results to the real value. For example if a solution of concentration of 2mg% gave rise the following HPLC results 2.1, 2.0, 2.0, 1.95 and 2.0 then the method is considered accurate and also it is precise.

Page 41: Qualification and Validation Basic Principles of GMP.

Accuracy is tested practically by preparing the formula by the analyst at a small scale. For example if tablet of Drug A is the candidate product the n the analyst will add API and the tablet scale with the same proportions of the tablet in a mortar then the mixed together to resemble grinded powder of the tablet. If 10mg of API used in the prepared formula then the HPLC method should recover this amount to be accurate. For example if the results of many samples gave 100%, 1010%, 99.9%, 100% and 101% of the claimed amount then the method is considered accurate.

Page 42: Qualification and Validation Basic Principles of GMP.

In order to test the accuracy at different concentration of API, analyst prepares three formulae one of them contains the same amount at the label of tablet (100%). The others are 50 and 200% of the labeled amount of the tablet. Sometimes, it is difficult to know the excipients of the dosage form or the technology of the preparation is not feasible to the analyst. The analyst can use recovery study which relies on a ready preparation. For example. If ointment contains 10 mg dexamethasone. The analyst might prepare the preparation by just add 10 mg of dexamethasone to the already prepared ointment. If the recovery was 20mg, 19.9mg, 21mg, 20.1 mg and 19.9mg then the method will considered accurate because it was able to recover the amount added by the analyst. The preparations should cover also 200 and 50% of the labeled amount as well.

Page 43: Qualification and Validation Basic Principles of GMP.

Linearity: it is important to check if the method is still valid when the analyst prepared a concentration more or less than the required one. The reason behind this is the behavior of detector with change of concentration for example UV detector is only linear at the range of absorbance = 0.1-1. Linearity study can be accomplished by analyzing several solutions of standard API at different concentrations. If the curve was linear then the method is considered valid for linearity point.

Page 44: Qualification and Validation Basic Principles of GMP.

mg% AUC

1 10000

2 19990

3 30100

4 40000

5 50600

6 60130

Example assume you get the following results for this group of concentrations

Page 45: Qualification and Validation Basic Principles of GMP.

Add these values to excel sheetSelect the table and choose insert form tool barSelect chartForm (standard types) choose XY (scatter)Choose: Next, next, next then finish.Left click on the mice over one of the data point at the curve to select themRight click and choose (add trendline) From options choose (display eq on chart) and (display R_squared value on chart)

Page 46: Qualification and Validation Basic Principles of GMP.

•You should get the following chart

Chart Title y = 10068x - 101.33

R2 = 0.9999

0

10000

20000

30000

40000

50000

60000

70000

0 1 2 3 4 5 6 7

AUC

Linear (AUC)

R2 (regression factor) is often accepted if it is above 0.999 (very close to one). This kind of regression (best-fit) in which the total of point deviation above t

he curve (residuals) equal those below the curve produce strong linearity.

Page 47: Qualification and Validation Basic Principles of GMP.

Selectivity or specificity: if an excipient interfere with the peak of API, the analyst can find this out by injection a solution of placebo. The method is considered not selective if one or more of placebo peaks are co-eluted with the HPLC peak of API. However, excipients are not the only interferant, the degradation products of API are also potential to deteriorate the method selectivity. To test that, API solutions are incubated at stress conditions with strong acidic , basic and oxidative conditions. The possibility of co-eluted degradants with same retention time of the API peak is test by Photo diode array (PDA).

Page 48: Qualification and Validation Basic Principles of GMP.

PDA is used for a test called selectivity, specificity or purity. If the chromatographic separation is not optimum then it is possible to have two analytes or more hidden into a peak of another compound. PDA can take the UV scan spectrum at different time points into the chromatographic peak if the spectra are alike then the purity of the peak is 100% and the peak is for a single compound. If not the method should be further optimized by choosing different mobile phases and columns until reaching nice separation for the analytes.

Page 49: Qualification and Validation Basic Principles of GMP.

LOD= when S/N is 2 or 3 LOQ= 10 LOD LOD=3 SD of blank/ slope in linearity LOQ= smallest conc linear with others and

with RSD less than 25% in biosamples.