INFUSION PUMPS
Fluid container
Drip chamber
Roller Clamp
Basic Infusion System
• Flow by gravity
• Flow controlled by roller clamp
• Difficult to set and control infusion rate
Infusion Pumps
What are they?• Usually electrically powered infusion devices
What do they do?• Use pumping action to infuse fluids, medication or
nutrients into patient• Suitable for intravenous, subcutaneous, enteral and
epidural infusions
Infusion Pumps
Why are they used?• To provide accurate and controllable flow over a prescribed
period or on demand
What are they used for?Wide range of drugs and therapies including• Chemotherapy• Pain management• Total parental nutrition• Anaesthesia/sedation• Etc. etc.
Infusion Pumps
TWO BASIC TYPES
• Syringe Pumps
• Volumetric Pumps
Syringe Pump
Syringe Pumps
• Generally used for low volume, low flow rate infusions
• Good short term accuracy• Long start up time at low flow rates
– Prime and purge line before connecting to patient
• Alarms: End/near end of infusion; drive disengaged, occlusion, battery low
• Specialised syringe pumps for ambulatory use, PCA, sedation, insulin etc
Volumetric Pumps
Latch
Cam followers (fingers)
Pressure sensor
Air in line detector
Volumetric Pumps
• Preferred for medium and high flow rates and large volumes
• Generally not suitable for rates < 5ml/h• Variable short term accuracy• Alarms: Latch/door open, set out,
occlusion, battery low, air-in-line• Specialised volumetric pumps for
ambulatory use, epidural infusions etc.
Infusion Pump Incidents
700 incidents/year reported to MHRA, including 10 deaths
20% Device related (e.g. design, failures etc)27% User error53% Not established (majority user error)
Many incidents not reported e.g. 6 Trusts, 321 incidents
Reporting Incidents
• All incidents should be reported on a Clinical Adverse Patient Incident Form
• Aim is to reduce risk in future, not to apportion blame
• Where an infusion pump is involved, the pump and its disposables must be retained, and Clinical Physics informed.
What Goes Wrong?
Medication ErrorsMedication Errors
• Prescription
• Preparation of infusion solution
• Calculation of rate of infusion
What Goes Wrong?
Medication ErrorsMedication Errors
• Prescription
• Preparation of infusion solution
• Calculation of rate of infusion
• Setting up infusion pump/unfamiliarity
If you accidentally use a BLUE 1hr pump instead of a GREEN 24 hour pump, you will deliver the drug at 24 times the intended rate.
What Goes Wrong?
Medication ErrorsMedication Errors
• Prescription• Preparation of infusion solution• Calculation of rate of infusion• Setting up infusion pump/unfamiliarity
Do not use a model you have not been trained and are deemed competent to use
What goes wrong?
Free flow by gravity/siphoningFree flow by gravity/siphoning
- What is it: Uncontrolled fluid flow by gravity from syringe or bag.
- Has resulted in a significant number of fatalities, none yet in North Glasgow.
Free Flow in Volumetric Pumps
If fluid container is a few inches above heart level, free flow by gravity can occur if:
• Pump latch/door open. Always close roller clamp before removing set from pump.
Latch closed
Clamp open
Latch open
Clamp closed
Free Flow in Volumetric Pumps
If fluid container is a few inches above heart level, free flow by gravity can occur if:
• Pump latch/door opened. Always close roller clamp before removing set from pump.
• Infusion set not correctly loaded
Free Flow in Volumetric Pumps
If fluid container is a few inches above heart level, free flow by gravity can occur if:
• Pump latch/door opened. Always close roller clamp before removing set from pump.
• Infusion set not correctly loaded• Damage to set resulting in an air leak
Free Flow in Syringe Pumps
If pump is a few inches above heart level, free flow by gravity can occur if:
1. Syringe not correctly located and secured to pump. Check syringe barrel clamp, barrel flange and plunger located correctly and secured.
Free Flow in Syringe Pumps
If pump is a few inches above heart level, free flow by gravity can occur if:
1. Syringe not correctly located and secured to pump. Check syringe barrel clamp, barrel flange and plunger located correctly and secured
2. Syringe removed from pump. Always close clamp first.
Free Flow in Syringe Pumps
If pump is a few inches above heart level, free flow by gravity can occur if:
1. Syringe not correctly located and secured to pump. Check syringe barrel clamp, barrel flange and plunger located correctly and secured
2. Syringe removed from pump. Always close clamp first.
3. Air leak caused by crack in syringe, plunger seal leak, loose luer connection, distortion of barrel/plunger.
To prevent free flow
• Never remove syringe or set from pump whilst connected to patient, without closing the clamp first (or checking it is closed)
• Always use a set with an anti free flow device (not available for Alaris/IVAC 59 series)
• Check set or syringe is correctly loaded• Check drip chamber on volumetric pump for
unexpected flow after set loading and during infusion
• Keep syringe pump near to or below infusion site
What Goes Wrong?
Occlusion alarm (all pumps)Occlusion alarm (all pumps)
- Occurs when pump is unable to sustain set flow rate and pressure in line increases
- Caused by partial or complete blockage in delivery tubing (kinked tube, clamp or tap closed) or cannula (clotted off, position changed)
Occlusion Alarm
Time to alarm• Dependent on occlusion pressure level
(usually variable) and flow rate– Low pressure, high flow rate 45 seconds– High pressure, low flow rate 45 minutes
• To reduce time to alarm and bolus size– Use highest flow rate clinically acceptable– Use lowest occlusion pressure setting possible
without causing nuisance alarms– User smaller syringes
Occlusion Alarm
Hazards
• Interruption to therapy– Problem with critical, fast acting drugs e.g.
inotropes
• Post occlusion bolus
Tissuing (Extravasation)Extravasation occurs when fluid that should be delivered intravenously is inadvertently delivered into a tissue space.
Tissuing
• Cannot be detected by infusion pumps– Usually little or no increase in pressure
• Secure and dress the catheter for stability
• Check IV site frequently for tenderness, skin tightening, cooling and blanching
What Goes Wrong?
Air-in-lineAir-in-line
• Volumetric pumps have a risk of air being delivered due to poor priming of set, upstream leak or pumping action drawing air out of solution
• Volumetric pumps have either a mechanism for preventing pumping of air or an air-in-line detector & alarm
Air-in-line Alarm
Hazards
• Nuisance alarms
• Interruption to therapy– Problem with critical, fast acting drugs e.g.
inotropes– Recent fatality in North Glasgow
What Goes Wrong?
Tampering by patients/visitors/carersTampering by patients/visitors/carers
1. Keylock
2. Lock box
Ambulatory (portable)
What Goes Wrong?
Equipment FaultsEquipment Faults• Often occur as a result of damage due to fluid
ingress or being dropped/knocked• Always return damaged pumps to Clinical
Physics – never use or attempt to repair• Infusion devices very reliable, faults rarely
result in adverse incidents
Training
• This presentation and demonstrations to follow are a general introduction ONLY
• Before using any infusion device you MUST have received specific training for that model and be signed off as competent – over 50 models in North Glasgow
• Otherwise DO NOT USE