Transcript of PT-V3-MKT-01-Phototherapy Brochure-2.00EN · PDF filePT3-1011 Eye Patch (L) PT3-1012 PT3-1013...
An LED phototherapy device for effective treatment of neonatal jaundice.
V3000
EN ISO 13485: 2012 Certi�ed Company
Distributor:
For more information or to place an order please contact us at:
[email protected]@[email protected]
+84 43 766 6521
http://mtts-asia.com
No 26, Lane 41 An Duong Vuong StreetPhu Thuong Ward, Tay Ho DistrictHanoi, Vietnam
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
182 cm
84 cm
22.8 kg
18.4 kg
Height
Width
Weight
Dimensions and Weight
Shipping Weight
- user selectable “standard” and “intensive” modes- used with premature and full term neonates- compatible with MTTS Warmer
Overview
- light spectrum set to optimized treatment of jaundice
1. LED Lights
- medical-grade power supply
2. Power Supply
- on/o� toggle switch- push button to toggle between normal and intensive modes
EN ISO 13485: 2012 Certi�ed Company
Subject to MTTS policy
PT-V3-MKT-01-Rev-2.00EN
4. Displays/Indicators
- bright , ‘across the room’ readable LED display
Requirements
Power Requirements
WARRANTY1 year
- no moving parts, no internal fan
- angle and placement to maximise skin surface area exposure
- total system time counter (lifetime)- total intensive mode system time counter (lifetime)- normal/intensive mode indicator lights
3. Controls
5. Stand
- adjustable height- adjustable angle of lights
MAIN UNIT
PT3-1000
Ordering Information and Servicing
ITEM ORDER NO
Eye Patch (S)
V3000 PT Unit Main Unit
ITEM ORDER NOACCESSORIES
PT3-1011
Eye Patch (M)
Eye Patch (L)
Curtains
PT3-1012
PT3-1013
PT3-1020
MTTS Lightmeter PT3-1030
Size: Small
Size: Medium
Size: Large
English
-
- treatment time counter
Agency Requirements
Voltage: 220V AC 10%
Frequency: 47 - 63 Hz
Wattage: 40WISO 13485: 2012
- 3 prong connection for safe electric grounding (Class I)- fuse for electrical protection- energy saving
Height Adjustment 34 cm
Type of LEDs
Phillips Luxeon Rebel Blue, High Power
Performance Speci�cations
42pcs
450 to 475 nm
458 nm 4nm
Standard: 20 µW.cm .nm
Light Spectrum
Light Intensity
Number of LEDs
Light Spectrum Peak + _
-2 -1
Intensive: 40 µW.cm .nm -2 -1
Lifetime 44,000 hours to 30% degradationSkin and Surface Area Coverage 75% of the body
EN IEC 60601-2-50: 2003
V3000 unique angled design feature ensures maximum skin coverage
A cross section of the V3000 shows the Philips LUXEON high-powered LEDs are posi- tioned so that the patient receives e�ective phototherapy on 75% of the skin surface.
User Manual PT3-1040
Standard Warranty
1 34
Medical Technology Transfer and Services