PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical...

18
Add: No.1-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China 213022 Tel: 0086 519 85195556 Fax: 0086 519 85195551 Http://www.kanghui.com Add: F25, Shanghai International Pharmaceutical Trad & Exhibition Tower, No,1399, Jinqiao Road, Pudong District, Shanghai, China 200129 Tel: 0086 21 50319916 Fax: 0086 21 50312913 NeoGen Femoral Nail System Product Catalog & Surgical Technique

Transcript of PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical...

Page 1: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

Add: No.1-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China 213022Tel: 0086 519 85195556 Fax: 0086 519 85195551Http://www.kanghui.com

Add: F25, Shanghai International Pharmaceutical Trad & Exhibition Tower, No,1399, Jinqiao Road, Pudong District, Shanghai, China 200129Tel: 0086 21 50319916 Fax: 0086 21 50312913

NeoGen Femoral Nail SystemProduct Catalog & Surgical Technique

Page 2: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

Surgical Technique

PREFACE

INDICATIONS

PATIENT PREPARATION

ENTRY PORTAL

FEMORAL ANTEGRADE SURGICAL TECHNIQUE FOR STANDARD FEMORAL OR RECON LOCKING MODES

FRACTURE REDUCTION

CANAL PREPARATION

NAIL SELECTION

DRILL GUIDE ASSEMBLY

INTERLOCKING FOR FEMORAL MODE

RECON MODE

INTERLOCKING FOR RECON MODE

DISTAL LOCKING

OPTION-Estimating locking screw length using the Screw Depth Gauge

CLOSURE

INSTRUMENTS & IMPLANTS

1

1

2

3

4

6

7

8

9

10

12

13

16

23

23

24

Content

Page 3: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

NeoGen Femoral Nail System 2

Patient is placed supine with unaffected limb extended below the affected limb and trunk.

The affected limb is adducted. Flex the affected hip 15°. Apply traction through a skeletal pin

or the foot with the fracture table foot holder. Adjust the affected limb for length and rotation

by comparison with the unaffected limb. Rotation is further checked by rotating the arm to

align the femoral neck anteversion and then making the appropriate correction by foot, usually

in 0-15° of external rotation. Decubitus position may also be used with the fracture table, in

this situation because of the change of position of the femoral head, the leg is usually

internally rotated 10-15°. This is best checked by visualizing the femoral anteversion

proximally and matching it with correct rotation at the knee (Figure 1).

Make a small 2-3 cm incision, 2 cm proximal to the proximal tip of the greater trochanter.

Angle this incision posteriorly at its proximal end. Carry the incision through the fascia.

Palpate the greater trochanter (Figure 2).

The femoral nails are designed to accommodate a standard femoral locking mode or the

Recon locking mode in the same leg. Utilizing the same NeoGen instruments, simplify the

surgical approach by allowing placement of the nail through the tip of the greater trochanter.

The proximal section of the implant is 13 mm. Screw placement for the reconstruction mode is

at the standard 130°. The nail has a 5° anteversion and an anterior bow to match the femur.

PREFACE

INDICATIONS

1

Figure 1

Figure 2

PATIENT PREPARATION

The indications for intramedullary locking nail include:

Traumatic fractures

Pathological fractures

Re-fractures

Non-unions

Reconstructive surgery

Surgical Technique

Page 4: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

43

Figure 3

Figure 4

Surgical Technique

Place the Entry Tool with Honeycomb Insert through the incision to bone. Adjust to align

the Entry Tool with the axial line of the femoral shaft in the A/P and lateral image views. This

may require placing pressure on the Entry Tool to align the Tip Threaded Guide Wire (251110)

with the axial line of the femur. Insert the Guide Wire when the axial line and drill alignment is

acceptable. The position will usually located on the tip of greater trochanter.

A second 3.2 mm Tip Threaded Guide Wire can be used to further define the correct entry

portal. In this way, the position in the trochanter is maintained should the first pilot drill be

removed or repositioned. The Guide Wire will snap fit into the Power. Once proper placement

of the Guide Wire has been established, the “honeycomb” insert should be removed (Figure 4).

Assemble the Entry Tool with Honeycomb Insert (251100). The Entry Tool can be set for

the appropriate limb by pulling back on the suction port and allowing it to snap back into

place with the R (right) or L (left) showing in the circular window in the handle. This allows

blood flow to quickly exit the tool. Attach suction to the Entry Tool to assist in blood

evacuation and minimize aerosolisation of blood to operative team (Figure 3).

ENTRY PORTAL FEMORAL ANTEGRADE SURGICAL TECHNIQUE FOR STANDARD FEMORAL OR RECON LOCKING MODES

NeoGen Femoral Nail System

Page 5: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

6

Snap the T-Handle (030070) onto the Reducer (251150) (Figure 7). Place the Reducer

through the Entry Tool and 14 mm Channel Reamer to reduce the fracture (Figure 8). Once the

Reducer is in the medullary canal and has captured the distal fragment, the Ball-Tipped Guide

Rod (251160) is inserted through it with the use of the Gripper into the distal femur in the

region of the old postulus scar (Figure 9). The Gripper (251170) is useful in holding onto the

Guide Rod during insertion and can be used to steer the Guide Rod Tip to the center of the

canal (Figure 9 insert).

Tighten the Entry Reamer Connector (251140) onto the 14 mm Channel Reamer

(251120) and insert the 12.5 mm Entry Reamer (251130) until it “clicks” into the assembly

(Figure 5). Attach the 12.5 mm Entry Reamer to power to ream the proximal section of the

femur through the Entry Tool. Adjust the Entry Reamer assembly over the Guide Wire and

ream until the Entry Reamer Connector stops at the Entry Tool. This reamer assembly

enlarges the proximal femur 1.0 mm over the diameter of the head of the nail to 14 mm.

Remove the 12.5 mm Entry Reamer and Guide Wire, keeping the Entry Tool and 14 mm

Channel Reamer in place (Figure 6).

FRACTURE REDUCTION

5

Figure 7

Figure 8

Figure 9

Figure 5

Figure 6

2

1

3

Surgical TechniqueNeoGen Femoral Nail System

Page 6: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

87

Determine nail diameter from image intensifier or templating. Never insert a nail that has

a larger diameter than the last reamer used. Position the tip of the guide rod at the desired

level of the tip of the nail considering fracture patterns and locking screw positioning. Measure

the nail length by positioning the open end of the Nail Depth Gauge (251180) over the exposed

end of the guide rod pushing the end down to the level of bone through the 14 mm Channel

Reamer. Confirm the position on the image intensifier. The tip of the Nail Depth Gauge should

line up slightly below the tip of the Entry Tool for correct placement. Read the nail length from

the calibrations exposed at the other end of the Ruler. Leave the guide rod in place for

placement of the nail. Exchange of the ball-tipped guide rod is not necessary (Figure 11).

Canal preparation is dependent on surgical decision. If reaming is planned, use

progressive reamers through the Entry Tool. Unreamed nails are selected based on

preoperative planning, but should be of sufficient size to provide translational fill of the

intramedullary canal in the mid-diaphysis. Once the Guide Rod is in place, remove the

Reducer but leave the 14 mm Channel Reamer in place. Proceed to sequentially ream the

femoral shaft 0.5 to 1.0 mm or more above the chosen nail diameter through the 14 mm

Channel Reamer. For more curved femoral shafts, 0.5 to 1.0 mm of over reaming may be

beneficial.

NOTE: For reamers larger than 12.5 mm, the Channel Reamer must be removed before

reaming (Figure 10).

CANAL PREPARATION NAIL SELECTION

Figure 10 Figure 11

Surgical Technique

Proximal End Color

Pink

Darkgolden

Nail type

Left

Right

NeoGen Femoral Nail System

Page 7: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

109

Figure 12

Figure 14

Figure 15

Figure 13

Surgical Technique

5.0 mm screws are to be used with 11 mm,12 mm FAN Implants.

4.5 mm screws are to be used with 9 mm,10 mm FAN Implants.

Proximal Screw: To place screws at a 50° angle from the greater to lesser trochanter, the

following options are available (Figure 14):

A. PREDRILLING TECHNIQUE — Make a stab incision at the entry hole and push the

Outer Drill Sleeve (251260) through the Drill Guide hole until it is touching the lateral cortex.

Introduce the Inner Drill Sleeve (251270) through the Outer Drill Sleeve. Attach the 4.0 Drill Bit (251290) to power. Drill to, but not through the opposite cortex and measure for proper

length. The length measurements are taken from the calibrations off the drill in relation to the

end of the Inner Drill Sleeve (Figure 15).

FEMORAL MODE

Insert the Quick Bolt (251230) into the Drill Guide (251190) and use the Guide Bolt Wrench (251220) to secure the bolt to the nail. Then screw the Guick Bolt(251230) onto the

top of the Drill Guide. This assembly is used to drive the nail into the medullary canal (Figure

12). Insert the Skin Protector (251250) in the incision parallel to the Entry Reamer Tool.

Remove the Entry Reamer Tool and 14 mm Channel Reamer. The Skin Protector will assist in

maintaining control of the surrounding tissues and provide continued access to the bone.

Advance the nail over the guide rod and carefully past the fracture. Remove the guide rod

after the nail is inserted and before inserting the locking screws (Figure 13).

DRILL GUIDE ASSEMBLY INTERLOCKING FOR FEMORAL MODE

NeoGen Femoral Nail System

Page 8: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

1211

The appropriate length 5.0 mm screw is selected and attached to the Screwdriver for locking screws (251350). The Drill and Inner Drill Sleeve are removed and the screw is inserted through

the Outer Drill Sleeve. Rotate the Screwdriver Handle and place screw in bone. It is recommended

that final tightening of the 5.0 mm screw should always be under manual control using the

Screwdriver for locking screws(Figure 16).

B. SCREW LENGTH GAUGE— Predrill through both cortices. The surgeon should check that

the Outer Drill Sleeve is positioned so that it is touching the bone. The Screw Depth Gauge (251340) cover is then unscrewed and removed. The hooked end is inserted down the Screw Guide

and through the bone. It is then drawn back so that the hook engages the outer surface of the far

cortex. The correct length of screw can now be read at the top of the Outer Drill Sleeve (Figure

17). The appropriate length 5.0 mm screw is selected and attached to the Screwdriver. It is

recommended that final tightening of the 5.0 mm screw should always be under manual control

using the Screwdriver for locking screw (Figure 18).

NOTE: Once screw is seated, the Connect Rod (251360) in the canulate T-Handle turned counterclockwise and the

Screwdriver for locking screws releases the screw to remove

the T-Handle (Figure 19).

Figure 17Figure 18

Figure 16

Figure 19

Figure 20Figure 21

Surgical Technique

6.4 mm screws are to be used with our 10 mm, 11 mm and 12 mm FAN and Trochanteric

Implants for placing screws into the femoral head.

Insert the Guide Bolt (251210) into the Drill Guide (251190) and use the Guide Bolt

Wrench (251220) to secure the bolt to the nail. Connect the Proximal Aiming bar (Femoral) (256100) to the Drill Guide. The guide is keyed so that it will only fit one way.

Tighten the Proximal Guide Bolt (251200) by hand. Use the end of the Guide Bolt Wrench

(251220) to finish tightening the Proximal Aiming bar (Femoral) (256100) in place. Check

the alignment of the Aiming bar to the screw holes by passing the Trocar (251430) through the

Outer Drill Sleeve up into the holes of the nail. Screw the Impactor onto the top of the Drill

Guide (Figure 20). This assembly is used to drive the nail into the medullary canal. Insert the

Skin Protector (251250) in the incision parallel to the Entry Reamer Tool. Remove the Entry

Reamer Tool and Channel Reamer. The Skin Protector will assist in maintaining control of the

surrounding tissues and provide continued access to the bone. Advance the nail over the guide

rod and carefully past the fracture. Remove the guide rod after the nail is inserted and before

inserting the locking screws (Figure 21).

RECON MODE

NeoGen Femoral Nail System

Page 9: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

1413

6.4 MM SCREW PLACEMENT TECHNIQUE-Make an incision at the entry holes of the

proximal screw sleeves, and then connect the two puncture wounds for approximately a 3 cm

incision that will accommodate the insertion of both screws (Figure 24). Insert the Inner Drill

Sleeve (251270) into the Outer Drill Sleeve (251260) and push to bone. Insert the 4.0 Drill Bit

(251290) into the Inner Drill Sleeve and connect to power. Drill into the femoral neck and head

to the desired depth and position (Figure 25). Remove the Inner Sleeve and drill the femoral neck

with the 6.4 Drill Bit (251320) to slightly less than the depth desired. Check the alignment in A/P

and 15° lateral views again before removing the 6.4 mm drill. Use the 6.4 mm Tap (251330) through the Outer Drill Sleeve to prepare the bone for screw insertion. Measure the depth for the

screw length from the calibrations on the drill or tap with respect to the Outer Drill Sleeve

(Figure 26).

Proximal Screws: Two aspects of screw placement into the femoral head must be noted

before drilling into the femoral head: (1) Alignment of the anteversion; and (2) Depth of nail

insertion. To begin, rotate the C-Arm proximally until a true line of the hip is visualized, this

gives the correct axis of alignment for anteversion. Rotate the handle of the Proximal Aiming

bar (Femoral) (256100) until it bisects the femoral head in the lateral view (Figure 22). This

should assist in setting the correct anteversion position of the screws. Mark this position with

a skin marker on the leg parallel to the driving handle. Next, rotate the C-Arm into an A/P

view using the calibrated notches on the proximal attachment of the nail, which is visualized

radiographically, to determine from preoperative planning what depth of nail insertion will be

required to allow both screws to be centered in the head. As a rule, the inferior screw is

placed first, though in situations where the neck is large enough, the proximal screw can be

placed, again, approximately 4-5 mm from the superior cortical margin of the femoral neck.

These screws are angled at 130° in relation to the shaft. If both screws will not seat within the

femoral head, it is probable that too much varus positioning of the proximal fragment has

occurred, or the proximal nail entry portal is too lateral (Figure 23).

INTERLOCKING FOR RECON MODE

Figure 22

Figure 23

Figure 26

Figure 24 Figure 25

Surgical TechniqueNeoGen Femoral Nail System

Page 10: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

1615

There may be some bending of the nail, due to the pressure and weight of the soft tissues

and the bone. Medio-lateral bending of the nail will not affect the targeting significantly, since

this is the plane of screw insertion, but any bending antero-posteriorly will result in failure of

the locking. The stabilizing system is therefore designed to correct antero-posterior alignment

between the guide bar and the nail. The Distal Outrigger provides the mounting point for a

Positioning Rod (251390) which is inserted down to the nail through the anterior femoral

cortex, and the U-shaped Stabilizing Spacer correct the distance and lock the Stabilizing Rod

to the outrigger.

The stages of distal locking therefore are as follows:

Stabilize the Distal Aiming bar (Femoral) in the appropriate position to correct for any

bending of the nail.

Make the incision(s) for distal locking, insert the Drill Sleeve down to the bone, and

complete the procedure.

DISTAL LOCKING

Surgical Technique

Attach the appropriate length 6.4 mm screw to the Screwdriver for locking screws

(251350). Tighten the proximal screws when the traction is released to maximize

compression at the fracture site (Figure 28). Once an acceptable position is obtained, detach

the Screwdriver for locking screw from the screws by wrenching counterclockwise the

Connecting Rod (251360), proximal locking is complete (Figure 29).

Figure 28

Figure 29

Figure 27

NeoGen Femoral Nail System

Page 11: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

1817 Surgical Technique

An Inner Drill Sleeve (251270) is inserted through the hole in the outrigger down to the

skin anteriorly, and by palpation is centred over the middle of the femur. The point of contact

with the skin is noted. A 15 mm incision is made at this point, down to the deep fascia. The

muscle is then split longitudinally down to the bone.

The Trocar (251430) is inserted into the Inner Drill Sleeve, and the two pushed together

down to the bone. The Inner Drill Sleeve is centered over the middle of the femoral shaft, by

palpation, using gentle pressure on the Aining bar in the frontal plane .

The Trocar is withdrawn, the 4.0 DRILL BIT is inserted down to the bone, using gentle

pressure to keep the point in contact with the cortex. The anterior cortex only is then drilled.

Remove the 4.0 Drill bit. The Positioning Rod (251390) is inserted through the hole in the

Distal Targeter, and the hole in the anterior femoral cortex, down to the nail (Figure 32),

contact being confirmed by tapping its tip on to the nail.

Figure 32

Stabilization of the Distal Aiming bar(Femoral)

The Distal Aiming bar (Femoral) (256110) is attached to the Proximal Aiming bar

(Femoral) (256100) and the Distal Guide Bolt (251420) tighten firmly by hand (Figure 30).

There have two holes for Guide Bolt to fit in the Proximal Aiming bar (Femoral). Make sure to

use the correct one that will promise the curvature of the Aiming bar structure match the

curvature of the femur or the nail.

Figure 30

The Targeter (Femoral) (256130) is now attached on the anterior side of the Distal

Aiming bar, at the middle of the two distal locking holes. The Distal Guide Bolt (251420) is

tightened firmly by hand (Figure 31).

Figure 31

NeoGen Femoral Nail System

Page 12: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

2019 Surgical Technique

Distal Locking

Outer drill sleeve are now inserted through each of the holes in the Targeter (Femoral) (Figure

34). A single 4-5 cm incision is made over the points of contact with the skin, down through the

deep fascia. The incision is deepened by blunt dissection, splitting the ilio-tibial tract

longitudinally, down to the bone, taking care to keep the incision in line with the fibres of the

ilio-tibial band.

The more proximal Outer drill slleve is now inserted down to the bone, with the aid of the

Trocar (Figure 35).

Figure 34

Figure 35

The E Block (Femoral) (256150) is now attached so that: the upper fork, marked the Number

corresponding with the nail size, fits into the groove in the shaft of the Stabilizing rod. The two

other forks grip the screw guide and the Targeter (Femoral) (Figure 33).

The handle of the Stabilizing rod is now held so that its tip is in contact with the nail. The

surgeon maintains this contact throughout. If the handle is pushed too hard, it is sometimes

possible to push the tip of the stabilizing rod past the nail. This must be avoided, since it will result

in the drill bit passing posterior to the nail. Gentle contact is all that is required.

Figure 33

NeoGen Femoral Nail System

Page 13: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

2221 Surgical Technique

Figure 37

A locking screw of correct length is now inserted into the second Outer drill sleeve, rotated

through the bone with the Screwdriver for locking screws (Fig.38).

Now that Screw Guide alignment is maintained by this Replacement Rod, the surgeon may

release the handle of the stabilizing rod.

The distal Outer drill sleeve with the Inner Drill Sleeve inserted is now advanced down to the

bone, and the second locking hole drilled (Figure 37). The appropriate screw length can be read

from the mark on the Drill bit, which related with the top of the Outer drill sleeve.

The Replacement Rod is removed from the first screw guide, and the surgeon again maintains

the position of the Stabilizing rod by gripping its T-handle. The same technique is followed for

insertion of the remaining locking screw. A check is now carried out with the Image Intensifier or

by X-ray to confirm that both screws have passed through the nail and that reduction has been

maintained. The Stabilizing rod, Outer drill sleeve and Targeter are removed.

Figure 38

An Inner drill sleeve is inserted into the outer drill sleeve, the 4.0 Drill Bit is attached

to the Inner drill sleeve (Figure 36-A).

The surgeon now grips the T-handle of the Stabilizing Rod, to keep its tip against the

nail, and MAINTAINS THIS POSITION THROUGHOUT THE DRILLING

PROCEDURE. The first locking hole is now drilled as for proximal locking, and the Inner

drill sleeve is removed.

The Replacement Rod (251400) is now inserted into the Outer Drill Sleeve (Figure

36-B), so that it passes through the nail, and engages the far cortex. This Replacement rod

has now stabilized the position of the Aiming bar. Do not drill the second hole until the

angled Replacement rod is in position.

Figure 36-A

Figure 36-B

NeoGen Femoral Nail System

Page 14: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

2423 Surgical Technique

Product Code

251100251110251120251130251140251150030070251160251170050011050085050090050095050100251230251240050105050110050115050120050125050130050135050140251180251190251200

Product Code

251420251210251220251360251370251250251260251270251280251290251300251310251320251330251340251350251380251390251400251410251430251440256100256110256130256150

10737230

Product DescriptionEntry ToolTip Threaded Guide Wire14mm Channel Reamer12.5mm Entry ReamerAdaptorReducerT-Handle with Quick Coupling Ball Tip Guide RodGripperFlexible Reamer ShaftFlexible Reamer,(Φ8.5)Reamer Head,(Φ9.0)Reamer Head,(Φ9.5)Reamer Head,(Φ10.0)Quick Bolt Hammer Reamer Head,(Φ10.5)Reamer Head,(Φ11.0)Reamer Head,(Φ11.5)Reamer Head,(Φ12.0)Reamer Head,(Φ12.5)Reamer Head,(Φ13.0)Reamer Head,(Φ13.5)Reamer Head,(Φ14.0)Nail Depth GaugeDrill GuideProximal Guide Bolt

Product Description Distal Guide BoltGuide BoltGuide Bolt WrenchConnecting RodExtractorSkin ProtectorOuter Drill Sleeve Inner Drill Sleeve Small Drill Sleeve 4.0 Drill Bit5.0 Tapping4.5 Tapping6.4 Drill Bit6.4 TappingScrew Depth GaugeScrewDriver for locking screwsPositioning Rod DrillStabling RodReplacement RodHandle with Quick Coupling TrocarThread Pin SleeveProximal Aiming bar(Femoral)Distal Aiming bar(Femoral)Targeter(Femoral)E Block(Femoral)2.0 Threaded Pin

Quantity131111111111111111111111112

Quantity221211222111111211111111112

INSTRUMENTS

Figure 39

OPTION-Estimating locking screw length using the Screw Depth Gauge

If there is any doubt about the correct length of locking screw, either in respect of the

measurement recorded following drilling, or because the surgeon omitted this step, the Screw

Depth Gauge (251340) may be used as follows: the surgeon should first check that the screw guide

is positioned so that it is touching the bone. The Screw depth gauge cover is then unscrewed and

removed.

The hooked end is inserted down the Outer drill sleeve and through the bone. It is then drawn

back so that the hook engages the outer surface of the distal cortex. The correct length of screw

can now be read at the top of the Outer drill sleeve. This Screw depth gauge is only suitable for use

with NeoGen nails, since its accuracy depends on a fixed length of Outer drill sleeve.

CLOSURE

On completion of the procedure, unscrewed the Quick bolt and the Proximal Aiming bar

(Femoral) is removed, wounds are irrigated and closed in a standard fashion (Figure 39).

NeoGen Femoral Nail System

Page 15: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

2625 Surgical Technique

NeoGen Locking ScrewΦ5

Product Code33101232331012343310123433101238331012403310124233101244331022323310223433102234331022383310224033102242331022443310323233103234331032343310323833103240331032423310324433104232331042343310423433104238331042403310424233104244

Product Code33101132331011343310113433101138331011403310114233101144331021323310213433102134331021383310214033102142331021443310313233103134331031343310313833103140331031423310314433104132331041343310413433104138331041403310414233104144

RemarkCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CM

RemarkCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CMCM

RegularRegularRegularRegularRegular

CM

IMPLANTS

NeoGen Femoral NailsAngle: 130° Proximal Radian: 5°

Product Code

33112070

33112075

33112080

33112085

33112090

33112091

33112092

33112093

33112094

Products Description

NeoGen Locking Screws,5X70mm

NeoGen Locking Screws,5X75mm

NeoGen Locking Screws,5X80mm

NeoGen Locking Screws,5X85mm

NeoGen Locking Screws,5X90mm

NeoGen Locking Screws,5X95mm

NeoGen Locking Screws,5X100mm

NeoGen Locking Screws,5X105mm

NeoGen Locking Screws,5X110mm

Remark

Regular

Regular

Regular

Regular

CM

CM

CM

CM

CM

Product Code

33112025

33112030

33112035

33112040

33112045

33112050

33112055

33112060

33112065

Products Description

NeoGen Locking Screws,5X25mm

NeoGen Locking Screws,5X30mm

NeoGen Locking Screws,5X35mm

NeoGen Locking Screws,5X40mm

NeoGen Locking Screws,5X45mm

NeoGen Locking Screws,5X50mm

NeoGen Locking Screws,5X55mm

NeoGen Locking Screws,5X60mm

NeoGen Locking Screws,5X65mm

Remark

CM

Regular

Regular

Regular

Regular

Regular

Regular

Regular

Regular

NeoGen Locking ScrewΦ4.5Product Code

33111060

33111065

33111070

33111075

33111080

33111085

33111090

Product Code

33111025

33111030

33111035

33111040

33111045

33111050

33111055

Products Description

NeoGen Locking Screws,4.5X25mm

NeoGen Locking Screws,4.5X30mm

NeoGen Locking Screws,4.5X35mm

NeoGen Locking Screws,4.5X40mm

NeoGen Locking Screws,4.5X45mm

NeoGen Locking Screws,4.5X50mm

NeoGen Locking Screws,4.5X55mm

Products Description

NeoGen Locking Screws,4.5X60mm

NeoGen Locking Screws,4.5X65mm

NeoGen Locking Screws,4.5X70mm

NeoGen Locking Screws,4.5X75mm

NeoGen Locking Screws,4.5X80mm

NeoGen Locking Screws,4.5X85mm

NeoGen Locking Screws,4.5X90mm

Remark

CM

Regular

Regular

Regular

Regular

Regular

Regular

Remark

Regular

Regular

Regular

Regular

CM

CM

CM

NeoGen Caps (Femoral)Product Code

33110010

33110015

Product Code

33110000

33110005

Products Description

NeoGen Nails Cap,¢13

NeoGen Naisl Cap,¢13,+5

Products Description

NeoGen Nails Cap,¢13,+10

NeoGen Nails Cap,¢13,+15

Note: The Femoral nail could be used as Reconstruction nails with the RECON screws.

Remark

Regular

Regular

Remark

Regular

Regular

Products DescriptionNeoGen Femoral Nails,9X320mm,RightNeoGen Femoral Nails,9X340mm,RightNeoGen Femoral Nails,9X360mm,RightNeoGen Femoral Nails,9X380mm,RightNeoGen Femoral Nails,9X400mm,RightNeoGen Femoral Nails,9X420mm,RightNeoGen Femoral Nails,9X440mm,RightNeoGen Femoral Nails,10X320mm,RightNeoGen Femoral Nails,10X340mm,RightNeoGen Femoral Nails,10X360mm,RightNeoGen Femoral Nails,10X380mm,RightNeoGen Femoral Nails,10X400mm,RightNeoGen Femoral Nails,10X420mm,RightNeoGen Femoral Nails,10X440mm,RightNeoGen Femoral Nails,11X320mm,RightNeoGen Femoral Nails,11X340mm,RightNeoGen Femoral Nails,11X360mm,RightNeoGen Femoral Nails,11X380mm,RightNeoGen Femoral Nails,11X400mm,RightNeoGen Femoral Nails,11X420mm,RightNeoGen Femoral Nails,11X440mm,RightNeoGen Femoral Nails,12X320mm,RightNeoGen Femoral Nails,12X340mm,RightNeoGen Femoral Nails,12X360mm,RightNeoGen Femoral Nails,12X380mm,RightNeoGen Femoral Nails,12X400mm,RightNeoGen Femoral Nails,12X420mm,RightNeoGen Femoral Nails,12X440mm,Right

Products DescriptionNeoGen Femoral Nails,9X320mm,LeftNeoGen Femoral Nails,9X340mm,LeftNeoGen Femoral Nails,9X360mm,LeftNeoGen Femoral Nails,9X380mm,LeftNeoGen Femoral Nails,9X400mm,LeftNeoGen Femoral Nails,9X420mm,LeftNeoGen Femoral Nails,9X440mm,Left

NeoGen Femoral Nails,10X320mm,LeftNeoGen Femoral Nails,10X340mm,LeftNeoGen Femoral Nails,10X360mm,LeftNeoGen Femoral Nails,10X380mm,LeftNeoGen Femoral Nails,10X400mm,LeftNeoGen Femoral Nails,10X420mm,LeftNeoGen Femoral Nails,10X440mm,LeftNeoGen Femoral Nails,11X320mm,LeftNeoGen Femoral Nails,11X340mm,LeftNeoGen Femoral Nails,11X360mm,LeftNeoGen Femoral Nails,11X380mm,LeftNeoGen Femoral Nails,11X400mm,LeftNeoGen Femoral Nails,11X420mm,LeftNeoGen Femoral Nails,11X440mm,LeftNeoGen Femoral Nails,12X320mm,LeftNeoGen Femoral Nails,12X340mm,LeftNeoGen Femoral Nails,12X360mm,LeftNeoGen Femoral Nails,12X380mm,LeftNeoGen Femoral Nails,12X400mm,LeftNeoGen Femoral Nails,12X420mm,LeftNeoGen Femoral Nails,12X440mm,Left

NeoGen RECON ScrewΦ6.4Product Code

33113092

33113093

33113094

33113095

33113096

33113097

Product Code

33113065

33113070

33113075

33113080

33113085

33113090

33113091

Products Description

NeoGen RECON Screws,6.4X65mm

NeoGen RECON Screws,6.4X70mm

NeoGen RECON Screws,6.4X75mm

NeoGen RECON Screws,6.4X80mm

NeoGen RECON Screws,6.4X85mm

NeoGen RECON Screws,6.4X90mm

NeoGen RECON Screws,6.4X95mm

Products Description

NeoGen RECON Screws,6.4X100mm

NeoGen RECON Screws,6.4X105mm

NeoGen RECON Screws,6.4X110mm

NeoGen RECON Screws,6.4X115mm

NeoGen RECON Screws,6.4X120mm

NeoGen RECON Screws,6.4X125mm

Remark

CM

Regular

Regular

Regular

Regular

Regular

Regular

Remark

Regular

Regular

Regular

CM

CM

CM

RECON Screws

6.4 RECONScrews

6.4 RECON Screws

6.4 RECON Screws

6.4 RECON Screws

Proximal Locking Screws

5.0 Locking Screw

5.0 Locking Screw

5.0 Locking Screw

5.0 Locking Screw

Distal Locking Screws

4.5 Locking Screw

4.5 Locking Screw

5.0 Locking Screw

5.0 Locking Screw

Nails Type

Neogen Femoral Nails, ¢9mm

Neogen Femoral Nails, ¢10mm

Neogen Femoral Nails, ¢11mm

Neogen Femoral Nails, ¢12mm

Screw Selection

NeoGen Femoral Nail System

*CM: Customer Made

Page 16: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

27 28Surgical Technique

251110 Tip Threaded Guide Wire 251120 14mm Channel Reamer 251130 12.5mm Entry Reamer

251150 Reducer 251160 Ball Tip Guide Rod 251170 Gripper

251100 Entry Tool-1 251100 Entry Tool-2 251100 Entry Tool

030070 T-Handle with Quick Coupling 050090 Flexible Reamer,(Φ9.0)

INSTRUMENTS

050140 Flexible Reamer,(Φ14.0)

251270Inner Drill Sleeve 251280Small Drill Sleeve 251290 4.0 Drill Bit

251180 Nail Depth Gauge 251190 Drill Guide 251200 Proximal Guide Bolt

251210 Guide Bolt

251240 Hammer 251250 Skin Protector

251220 Guide Bolt Wrench 251230 Guick Bolt

251260 Outer Drill Sleeve

NeoGen Femoral Nail System

Page 17: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

3029 Surgical Technique

256110 Distal Aiming bar(Femoral)

251320 6.4 Drill Bit261130 Distal Aiming bar(Femoral)

256130 Targeter(Femoral) 256150 E Block(Femoral)

251350 ScrewDriver for locking screws

251380 Positioning Rod Drill

251430 Trocar 251440 Thread Pin Sleeve 256100 Proximal Aiming bar(Femoral)

251390 Positioning Rod 251400 Replacement Rod

251360 Connecting Rod 251370 Extractor

251310 4.5 Tapping 251340 Screw Depth Gauge251300 5.0 Tapping

NeoGen Femoral Nail System

Page 18: PT Jantra Reka Saksanamas - Product Catalog & …jantraindonesia.com/eng/documents/Surgical Technique...NeoGen Femoral Nail System 2 Patient is placed supine with unaffected limb extended

31

NOTE