Proportional Assist Ventilation(PAV+) and WOB
L. Brochard, G Carteaux
![Page 2: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/2.jpg)
Conflicts of interest
• Our clinical research laboratory has received research grantsfor clinical trials from the following companies– Covidien (PAV+)
– Dräger (SmartCare)
– General Electric (FRC)
– Respironics (NIV)
– Fisher Paykel (Optiflow)
– Vygon (CPAP)
![Page 3: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/3.jpg)
ABC of ventilationPSV, PAV, NAVA
Trigger Control Cycling end
expiration
PSV Flow or
Pressure
Constant
Pressure
% Peak-
Flow
PEEP
NAVA patient
EMG
Pressure
proportion-
nal to
EMGdi
End of
EMGdi
PEEP and
EMGdi
PAV Flow or
Pressure
Pressure
proportion-
nal to flow
and volume
End of
Patient
Flow
PEEP
![Page 4: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/4.jpg)
![Page 5: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/5.jpg)
VENTILATEUR
Nerf phrénique
Système nerveux central
Stimulation et contractiondiaphragmatique
Variation de pression
Expansion de la paroi thoracique et des poumons
Variation de débit, de volume
NAVA
Technologie
idéale
Technologie
actuelle
PAV
![Page 6: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/6.jpg)
huge
Paw proportional
![Page 7: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/7.jpg)
RESPIRATORY SYSTEM MODEL
Compliance and Resistance = CONSTANT
Pressure, Volume, Flow= VARIABLES
Equation of Motion
muscle + ventilator
![Page 8: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/8.jpg)
Equation of motion
Pappl = (VxE) + (V'xR)
Pappl = Paw + Pmus
Pappl = (VxE) + (V'xR) = Pel + Pres
Gain
![Page 9: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/9.jpg)
G: gains for volume (VA) and flow (FA)VA is a fraction (percentage) of Ers, and FA is a fraction of Rrs
![Page 10: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/10.jpg)
Estimation of Elastance
300 ms end inspiratory pause manoeuvre at a random intervals of 4 to 10 breaths : PplatPAV
EPAV = (PplatPAV - PEEPtot) / Vt
Younes M, et al. AJRCCM 2001;164:50-60
![Page 11: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/11.jpg)
FLOW
Pes
Paw
GAIN = 40%
![Page 12: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/12.jpg)
Settings in PAV+
FiO2 PEEP Gain
How to adjust it?
![Page 13: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/13.jpg)
![Page 14: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/14.jpg)
understand « Gain »
K =1
1 - Gain
(P0 = 0) Ptot = K . |Pmus|For a Gain of 75% support
K =1
0.25
K =1
1 - 0.75
K = 4
Paw
25% 75%
100%
Ptot
Pmus
Carteaux et al CCM 2013
![Page 15: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/15.jpg)
Pmus,Peak = (Paw,Peak – PEEP) . 1 – Gain
Gain
Delta Paw (cm H20) = Paw,Peak - PEEP
1 2 3 4 5 6 7 8 9 10 12 15 17 20 25 30 35 40
% assist
20 4 8 12 16 20 24 28 32 36 40 48 60 68 80 100 120 140 160
25 3 6 9 12 15 18 21 24 27 30 36 45 51 60 75 90 105 120
30 2 5 7 9 12 14 16 19 21 23 28 35 40 47 58 70 82 93
35 2 4 6 7 9 11 13 15 17 19 22 28 32 37 46 56 65 74
40 2 3 5 6 8 9 11 12 14 15 18 23 26 30 38 45 53 60
45 1 2 4 5 6 7 9 10 11 12 15 18 21 24 31 37 43 49
50 1 2 3 4 5 6 7 8 9 10 12 15 17 20 25 30 35 40
55 1 2 2 3 4 5 6 7 7 8 10 12 14 16 20 25 29 33
60 1 1 2 3 3 4 5 5 6 7 8 10 11 13 17 20 23 27
65 1 1 2 2 3 3 4 4 5 5 6 8 9 11 13 16 19 22
70 0 1 1 2 2 3 3 3 4 4 5 6 7 9 11 13 15 17
75 0 1 1 1 2 2 2 3 3 3 4 5 6 7 8 10 12 13
80 0 1 1 1 1 2 2 2 2 3 3 4 4 5 6 8 9 10
85 0 0 1 1 1 1 1 1 2 2 2 3 3 4 4 5 6 7
90 0 0 0 0 1 1 1 1 1 1 1 2 2 2 3 3 4 4
Pmus,Peak
Gain
()
Carteaux et al CCM 2013
![Page 16: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/16.jpg)
Pmus
RR
Ti Te
Pmus,Peak
PTPmus
Pmus,Peak x Ti
2PTPmus = x RR
![Page 17: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/17.jpg)
CRITERIA FOR WEANING TRIAL
-Resolution of the pathologie requiringintubation-No sign of ARD-Fi02 50%-SaO2 ≥ 90%-Conscious patient-No vasopressor, haemodynamic stability
Adjust FiO2 and PEEP
According to:-Oxygenation (SpO2, PaO2)- Underlying respiratory disease
How to adjust the Gain 1
Standard settings
Assess Pmus
Pmus < 5 cmH2O 5 Pmus 10 cmH2O Pmus > 10 cmH2O
Decrease Gain in steps of 10%
Optimal WOB area.Decrease Gain in steps of 5% while Pmus remains in this area.
Increase Gain in steps of 10%
Initiating PAV+
INITIAL SETTINGS-Gain: 50%-Inspiratory trigger: 1 l/min-Expiratory trigger: 1%-FiO2 et PEEP: no specific rule
ALARMS-Paw max: 40 cmH2O-RR: 40/min-Vte max: 10 ml/kg-Vte min: 0 ml
If appears, while Pmus is in the optimal area
-Signs of ARD-Vte < 5 ml/kg-Respiratory acidosis
Increase Gain in steps of 10%
-Vte > 10 ml/kg-Respiratory alcalosis
Decrease Gain in steps of 10%
FiO2 & PEEP optimisation
2Aditionalsettings
Search another hypertotilation’s cause
Persistance Lack
Regardless of the Gain
![Page 18: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/18.jpg)
Adaptation of the Gain every 8h if needed
Clinical data collected daily
A computer was connected to the
ventilator for continuous recording
of ventilator’s data (sampling frequency =
1 min)
![Page 19: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/19.jpg)
![Page 20: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/20.jpg)
19 switch to ACV
52 PAV+
33 until Extubation
27 settings according to
protocol (82%)
4 needed additional
settings2 out of the
rules
1 Intolerance 18 Aggravation
Carteaux et al CCM 2013
![Page 21: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/21.jpg)
![Page 22: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/22.jpg)
![Page 23: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/23.jpg)
Each box-plot represents a patient
PTPmus median values all over PAV+ ventilation per patientP
TP
mu
s (c
mH
20.s
.min
-1)
Carteaux et al CCM 2013
![Page 24: Proportional Assist Ventilation (PAV+) and WOB · ventilator for continuous recording of ventilator’s data (sampling frequency = 1 min) 19 switch to ACV 52 PAV+ 33 until Extubation](https://reader030.fdocuments.us/reader030/viewer/2022040315/5e20630465db285f4a477048/html5/thumbnails/24.jpg)
Conclusions
• Many patients tolerate PAV+ ventilation over several days
• Setting PAV+ to target a « normal » level of effort seems feasible
• The target is reached in 85% of the cases
• Refinement of the Paw criteria may benecessary