Products Solutions Services Safety Instructions Proline Promass 300 Class I, Zone 1, AEx/Ex de ia IIC/IIB T6…T1 Class I, Zone 1, AEx/Ex d ia IIC/IIB T6…T1 Zone 21, AEx/Ex tb IIIC Txx °C C US Document: XA01472D Safety instructions for electrical apparatus for explosion- hazardous areas classified according to the National Electrical Code (NEC) and Canadian Electrical Code (CEC) 3 XA01472D/06/EN/01.16 71327890

Transcript of Proline Promass 300 - Endress+Hauser · 2016. 7. 29. · Proline Promass 300 XA01472D...

  • Products Solutions Services

    Safety InstructionsProline Promass 300Class I, Zone 1, AEx/Ex de ia IIC/IIB T6…T1Class I, Zone 1, AEx/Ex d ia IIC/IIB T6…T1Zone 21, AEx/Ex tb IIIC Txx °C

    C US

    Document: XA01472DSafety instructions for electrical apparatus for explosion-hazardous areas classified according to the National ElectricalCode (NEC) and Canadian Electrical Code (CEC) → 3

    XA01472D/06/EN/01.1671327890

  • Proline Promass 300 XA01472D

    Endress+Hauser 3

    Proline Promass 300

    Table of contents

    Associated documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    Manufacturer's certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Extended order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Safety instructions: General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    Safety instructions: Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    Safety instructions: Zone 21, Class II, Class III . . . . . . . . . . . . . . . . . . . . . . 10

    Temperature tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Connection data: Signal circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

  • XA01472D Proline Promass 300

    4 Endress+Hauser

    Associateddocumentation

    All documentation is available:• On the CD-ROM supplied (not included in the delivery for all device

    versions).• Available for all device versions via:

    – Internet: www.endress.com/deviceviewer– Smart phone/tablet: Endress+Hauser Operations App

    • In the Download Area of the Endress+Hauser web site:www.endress.com → Download

    This document is an integral part of the following OperatingInstructions:

    Measuringdevice

    Documentation

    HART FOUNDATIONFieldbus

    PROFIBUS PA ModbusRS485

    Promass A 300 BA01482D BA01515D BA01504D BA01493D

    Promass E 300 BA01484D BA01517D BA01506D BA01495D

    Promass F 300 BA01485D BA01518D BA01507D BA01496D

    Promass H 300 BA01486D BA01519D BA01508D BA01497D

    Promass I 300 BA01487D BA01520D BA01509D BA01498D

    Promass O 300 BA01488D BA01521D BA01510D BA01499D

    Promass P 300 BA01489D BA01522D BA01511D BA01500D

    Promass Q 300 BA01490D BA01523D BA01512D BA01501D

    Promass S 300 BA01491D BA01524D BA01513D BA01502D

    Promass X 300 BA01492D BA01525D BA01514D BA01503D

    Additional documentation:

    Document type Contents Documentation code

    Special documentation Remote display and operatingmodule DKX001

    SD01763D

    Safety instructions Remote display and operatingmodule DKX001:Class I, Zone 1, Zone 21

    XA01499D

    Control drawing As wanted on thenameplate.

    Please note the documentation associated with the device.

  • Proline Promass 300 XA01472D

    Endress+Hauser 5

    Manufacturer'scertificates

    Certificate numberCSA 16.70087388

    Notified bodyCSA: Canadian Standards Association

    Extended ordercode

    The extended order code is indicated on the nameplate, which is affixedto the device in such a way that it is clearly visible. Additionalinformation about the nameplate is provided in the associatedOperating Instructions.

    Structure of the extended order code

    * * * * * * _ * * * * * … * * * * * + A*B*C*D*E*F*G*…

    Device type Basic specifications Optional specifications

    * = Spaceholder: At this position, an option (number or letter) selected fromthe specification is displayed instead of the placeholders.

    • Device typeThe device and the device design is defined in the "Device type" section(Product root).

    • Basic specificationsThe features that are absolutely essential for the device (mandatoryfeatures) are specified in the basic specifications. The number ofpositions depends on the number of features available. The selectedoption of a feature can consist of several positions.

    • Optional specificationsThe optional specifications describe additional features for the device(optional features). The number of positions depends on the numberof features available. The features have a 2-digit structure to aididentification (e.g. JA). The first digit (ID) stands for the featuregroup and consists of a number or a letter (e.g. J = test, certificate).The second digit constitutes the value that stands for the featurewithin the group (e.g. A = 3.1 material (wetted parts), inspectioncertificate).

    More detailed information about the device is provided in the followingtables. These tables describe the individual positions and IDs in theextended order code which are relevant to hazardous locations.

  • XA01472D Proline Promass 300

    6 Endress+Hauser

    Device type

    Item Order code for Selected option Description

    1 Instrument family 8 Coriolis flowmeter

    2 Sensor A, E, F, H, I, O, P, Q, S, X Sensor type

    3 Transmitter 3 Transmitter type:4-wire, compact version

    4 Generation index B Platform generation

    5, 6 Nominal diameter DN 1 to 350DN 1: 01DN 2: 02toDN 350: 3E, 3F, 3R

    Nominal diameter of sensor

    Basic specifications

    Position 1, 2Order code for"Approval"Optionselected

    Position 4, 5Order code for"Output, input 1"Option selected

    Type of protection

    Transmitter Sensor

    C1 BA, BB, GA, LA, NA,RA, SA, MA

    Cl.I, Zone 1, AEx/Ex de ia IIB T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db

    Cl.I, Zone 1, AEx/Ex ia IIB T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIB

    T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db

    C2 BA, BB, GA, LA, NA,RA, SA, MA

    Cl.I, Zone 1, AEx/Ex de ia IIC T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db

    Cl.I, Zone 1, AEx/Ex ia IIC T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIC

    T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db

    C3 BA, BB, GA, LA, NA,RA, SA, MA

    Cl.I, Zone 1, AEx/Ex d ia IIB T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db

    Cl.I, Zone 1, AEx/Ex ia IIB T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIB

    T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db

    C4 BA, BB, GA, LA, NA,RA, SA, MA

    Cl.I, Zone 1, AEx/Ex d ia IIC T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db

    Cl.I, Zone 1, AEx/Ex ia IIC T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIC

    T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db

  • Proline Promass 300 XA01472D

    Endress+Hauser 7

    Position Order code for Option selected Description

    4, 5 Output, input 1 BA 4-20mA HART

    CA 4-20mA HART Ex-i

    GA PROFIBUS PA

    HA PROFIBUS PA Ex-i

    MA Modbus RS485

    SA FOUNDATION Fieldbus

    TA FOUNDATION Fieldbus Ex-i

    6 Output, input 2 A W/o;

    B 4-20mA

    C 4-20mA Ex-i

    D Configurable I/O initial setting off

    E Pulse/freq./switch output

    F Pulse output, phase-shifted

    G Pulse/freq./switch output Ex-i

    H Relay

    I 4-20mA input

    J Status input

    7 Output, input 3 A W/o;

    B 4-20mA

    C 4-20mA Ex-i

    D Configurable I/O initial setting off

    E Pulse/freq./switch output

    F Pulse output, phase-shifted

    G Pulse/freq./switch output Ex-i

    H Relay

    I 4-20mA input

    J Status input

    8 Display, Operation A W/o; via communication

    F 4-line, illuminated; touch control

    G 4-line, illuminated; touch control + WLAN

    M W/o; prepared for remote display DKX001 1)

  • XA01472D Proline Promass 300

    8 Endress+Hauser

    Position Order code for Option selected Description

    O Separate, with remote display DKX001, 4-line, illuminated; 10m/30ft cable; touch control

    17, 18 Device model A1 1

    1) DKX001 is approved according to CSA 160686-70030937.

    Optional specifications

    ID Order code for Option selected Description

    Jx Test, Certificate JP Ambient temperature of measuring device –50 °C

    Safetyinstructions:General

    • Staff must meet the following conditions for mounting, electricalinstallation, commissioning and maintenance of the device:– Be suitably qualified for their role and the tasks they perform.– Be trained in explosion protection.– Be familiar with national regulations (e.g. CEC or NEC).

    • Install the device according to the manufacturer's instructions andnational regulations.

    • Do not operate the device outside the specified electrical, thermal andmechanical parameters.

    • Only use the device in media to which the wetted materials havesufficient durability.

    • Refer to the temperature tables for the relationship between thepermitted ambient temperature for the sensor and/or transmitter,depending on the range of application, and the temperature classes.

    • Modifications to the device can affect the explosion protection andmust be carried out by staff authorized to perform such work byEndress+Hauser.

    • When using in hybrid mixtures (gas and dust occurringsimultaneously), additional measures should be taken. Please seeapproval body.

    • Open the device only if one of the following conditions is met:– An explosive atmosphere is not present.– A waiting time of 10 minutes is observed after switching off the

    power supply.• Observe all the technical data of the device (see nameplate).

    LWARNINGSubstitution of components is not permitted.‣ Substitution of components may impair intrinsic safety.

  • Proline Promass 300 XA01472D

    Endress+Hauser 9

    Safetyinstructions:Installation

    In the event of potentially explosive vapor/air mixtures, only operatethe device under atmospheric conditions.• Temperature: –20 to +60 °C• Pressure: 80 to 110 kPa (0.8 to 1.1 bar)• Air with normal oxygen content, usually 21 % (V/V)If no potentially explosive mixtures are present, or if additionalprotective measures have been taken, the device may also be operatedunder non-atmospheric conditions in accordance with themanufacturer's specifications.• Continuous service temperature of the connecting cable:

    –40 to +80 °C (–50 to +80 °C for optional specification, ID Jx (Test,Certificate) = JP); but at least according to the service temperaturerange of the application taking into account additional influences ofthe process conditions (Ta,min and Ta,max + 20 K).

    • Only use certified cable entries suitable for the application. Observeselection criteria as per CEC or NEC.

    • The following applies when connecting the transmitter with aconnection compartment in Ex d:Only use separately certified cable and wire entries (Ex d IIC) whichare suitable for operating temperatures up to 85 °C and for IP 67protection. If using conduit entries, the associated sealingmechanisms must be mounted directly on the housing.

    • The following applies when connecting the transmitter with aconnection compartment in Ex e:Only use separately certified cable and wire entries and sealing plugs(Ex e II) which are suitable for operating temperatures up to 85 °Cand for IP 67 protection. The cables must be routed such that they aresecurely seated, and sufficient strain relief must be ensured.

    • When the measuring device is connected, attention must be paid toexplosion protection at the transmitter.

    • In potentially explosive atmospheres:– Do not disconnect the electrical connection of the power supply

    circuit when energized.– Do not open the connection compartment cover when energized.

    • Install the transmitter circuit wiring according to Canadian ElectricalCode (CEC) respective National Electrical Code (NEC) using threadedconduit or other wiring methods in accordance with articles 500 to510.

    • When connecting through a conduit entry approved for this purpose,mount the associated sealing unit directly at the housing.

    • Seal unused entry glands with approved sealing plugs that correspondto the type of protection. The plastic transport sealing plug supplieddoes not meet this requirement and must therefore be replacedduring installation.

    • Only use certified sealing plugs. The metal sealing plugs suppliedmeet this requirement.

  • XA01472D Proline Promass 300

    10 Endress+Hauser

    Intrinsic safety• Observe the guidelines for interconnecting intrinsically safe circuits

    (e.g. , Proof of Intrinsic Safety).• When the intrinsically safe Ex ia circuits of the device are connected

    to certified intrinsically safe Ex ib circuits with explosion group IIC orIIB:The type of protection changes to Ex ib IIC or Ex ib IIB.

    • The device can be connected to the remote display and operatingmodule DKX001 with Ex ia explosion protection; refer to the SpecialDocumentation and Ex documentation.

    If the remote display and operating module DKX001 is used, theinternal display unit must be removed from transmitter!

    Potential equalization• Integrate the device into the local potential equalization .• If the ground connection has been established via the pipe as

    specified, it is also possible to integrate the sensor into the potentialequalization system via the pipe.

    Safetyinstructions: Zone21, Class II, ClassIII

    • To ensure dust-tightness, securely seal the transmitter housing, cableentries and sealing plugs.

    • Only open the transmitter housing briefly, ensuring that no dust ormoisture enters the housing.

    • Only use certified cable entries. The metal cable entries supplied meetthis requirement.

    • When connecting the transmitter to the remote display andoperating module DKX001: Connection values → 21

    Temperaturetables

    Ambient temperatureMinimum ambient temperature:• Ta = –40 °C• Optional specification, ID Jx (Test, Certificate) = JP

    Ta = –50 °CMaximum ambient temperature:Ta = +60 °C depending on the medium temperature and temperatureclass

  • Proline Promass 300 XA01472D

    Endress+Hauser 11

    Medium temperatureMinimum medium temperature:• Promass A, F, H, I, P, Q, S, X:

    Tm = –50 °C• Promass E, O: Tm = –40 °C• Promass F, Q with cryogenic temperature version (order code for

    "Measuring tube material", Option LA):Tm = –200 °C

    Maximum medium temperature:• Tm for T6…T1 depending on the maximum ambient temperature Ta• ( ) = The maximum permitted medium temperatures in brackets only

    apply if the sensor is installed in such a way that the transmitter isnot mounted above the sensor and free convection can occur on allsides.

    Compact version

    Without thermal insulation

    Promass A

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    1…4 50 205 50 95 130 150 205 205

    60 – 95 130 150 205 205

    Promass E

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8…50 50 150 50 100 130 130 150 150

    55 – 80 100 130 150 150

    60 – (80) (100) (130) (150) (150)

    80 50 150 50 75 110 150 150 150

    55 – 75 110 150 150 150

    60 – (75) (110) (150) (150) (150)

  • XA01472D Proline Promass 300

    12 Endress+Hauser

    Promass F

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    08…15 50 150 50 95 130 150 150 150

    60 – 95 130 150 150 150

    50 240 50 95 130 160 240 240

    60 – 95 130 160 (240) (240)

    50 150 2) 50 95 100 150 150 150

    60 – 95 100 150 (150) (150)

    25…50 50 150 50 95 130 150 150 150

    60 – 95 130 150 150 150

    50 240 50 95 130 160 240 240

    60 – 95 130 160 (240) (240)

    50 150 2) 50 95 100 150 150 150

    60 – 95 100 150 (150) (150)

    50 350 45 85 115 175 275 350

    60 – 85 115 175 275 350

    80 50 350 45 85 115 175 275 350

    60 – 85 115 175 275 350

    80…250 50 150 50 75 110 150 150 150

    60 – 75 110 150 150 150

    50 240 50 75 110 170 240 240

    60 – 75 110 170 (240) (240)

    50 150 2) 50 75 110 150 150 150

    60 – 75 110 150 150 150

    1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C

    Promass H

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 50 150 1) 50 65 100 150 150 150

    60 – 65 100 150 150 150

  • Proline Promass 300 XA01472D

    Endress+Hauser 13

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 50 205 2) 50 65 100 160 205 205

    60 – 65 100 160 205 205

    15…50 50 150 1) 50 75 115 150 150 150

    60 – 75 115 150 150 150

    15…50 50 205 2) 50 75 115 180 205 205

    60 – 75 115 180 205 205

    1) Measuring tube material tantalum2) Measuring tube material zirconium

    Promass I

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8, 15,15FB, 25

    50 150 50 95 130 150 150 150

    60 – 95 120 (150) (150) (150)

    25FB, 40,40FB, 50

    50 150 50 85 120 150 150 150

    60 – 85 120 (150) (150) (150)

    50FB, 80 50 150 50 85 120 150 150 150

    60 – 85 120 (150) (150) (150)

    FB = Full bore

    Promass O

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    80 … 250 50 205 50 75 110 170 205 205

    55 – 75 110 170 205 205

    60 – 75 110 170 (205) (205)

  • XA01472D Proline Promass 300

    14 Endress+Hauser

    Promass P

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 45 150 45 65 100 150 150 150

    60 – 65 100 150 150 150

    45 205 45 65 100 160 205 205

    60 – 65 100 160 205 205

    15…50 50 150 50 75 115 150 150 150

    60 – 75 115 150 150 150

    50 205 50 75 115 180 205 205

    60 – 75 115 180 205 205

    1) Maximum temperature range see nameplate

    Promass Q

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    25 … 100 50 205 50 75 110 160 205 205

    60 – 75 110 160 205 205

    25 … 100 50 150 2) 50 75 110 150 150 150

    60 – 75 110 150 150 150

    1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C

    Promass S

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 45 150 45 65 100 150 150 150

    60 – 65 100 150 150 150

    15…50 50 150 50 75 115 150 150 150

    60 – 75 115 150 150 150

  • Proline Promass 300 XA01472D

    Endress+Hauser 15

    Promass X

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    350 50 180 50 90 120 170 180 180

    55 – 90 120 170 180 180

    60 – (90) (120) (170) (180) (180)

    With thermal insulation according to Endress+Hauser specificationsFor information on the thermal insulation of the device, see the"Thermal insulation" section of the Operating Instructions

    Promass A

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    1…4 50 205 50 95 130 150 205 205

    55 – (95) (130) (150) (205) (205)

    Promass E

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8…50 50 150 50 100 130 130 150 150

    55 – (100) (130) (130) (150) (150)

    80 45 150 50 75 110 150 150 150

    50 – 75 110 150 150 150

    55 – (75) (110) (150) (150) (150)

    Promass F

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    08…15 50 150 50 95 130 150 150 150

    60 – 95 110 (150) (150) (150)

    50 240 50 95 130 160 240 240

  • XA01472D Proline Promass 300

    16 Endress+Hauser

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    55 – (95) (130) (160) (240) (240)

    60 – 95 110 110 110 110

    25…50 50 150 50 95 130 150 150 150

    60 – 95 110 (150) (150) (150)

    50 240 50 95 130 160 240 240

    55 – (95) (130) (160) (240) (240)

    60 – 95 110 110 110 110

    50 350 45 85 115 175 275 350

    60 – 85 115 175 275 350

    80 50 350 45 85 115 175 275 350

    60 – 85 115 175 275 350

    80…250 50 150 50 75 110 150 150 150

    60 – 75 110 (150) (150) (150)

    50 240 50 75 110 170 240 240

    55 – (75) (110) (170) (240) (240)

    60 – 75 110 110 110 110

    1) Maximum temperature range see nameplate

    Promass H

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 50 150 1) 50 65 100 150 150 150

    55 – (65) (100) (150) (150) (150)

    60 – 65 100 100 100 100

    8 50 205 2) 50 65 100 160 205 205

    55 – (65) (100) (160) (205) (205)

    60 – 65 100 100 100 100

    15…50 50 150 1) 50 75 115 150 150 150

    55 – (75) (115) (150) (150) (150)

    60 – 75 115 115 115 115

  • Proline Promass 300 XA01472D

    Endress+Hauser 17

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    15…50 50 205 2) 50 75 115 180 205 205

    55 – (75) (115) (180) (205) (205)

    60 – 75 115 115 115 115

    1) Measuring tube material tantalum2) Measuring tube material zirconium

    Promass I

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8, 15,15FB, 25

    50 150 50 95 130 150 150 150

    60 – 95 120 (150) (150) (150)

    25FB, 40,40FB, 50

    50 150 50 85 120 150 150 150

    60 – 85 120 (150) (150) (150)

    50FB, 80 50 150 50 85 120 150 150 150

    60 – 85 120 (150) (150) (150)

    FB = Full bore

    Promass O

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    80…250 50 205 50 75 110 170 205 205

    55 – (75) (110) (170) (205) (205)

    Promass P

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 45 150 45 65 100 150 150 150

    50 – 65 100 150 150 150

    60 – 65 100 125 (150) (150)

    45 205 45 65 100 160 205 205

  • XA01472D Proline Promass 300

    18 Endress+Hauser

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    50 – 65 100 160 205 205

    60 – 65 100 115 (205) (205)

    15…50 50 150 50 75 115 150 150 150

    60 – 75 115 125 (150) (150)

    50 205 50 75 115 180 205 205

    60 – 75 115 (150) (150) (150)

    1) Maximum temperature range see nameplate

    Promass Q

    DN Ta[°C]

    Tm, max 1)

    [°C]Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    25…100 50 205 50 75 110 160 205 205

    55 – (75) (110) (160) (205) (205)

    25…100 50 150 2) 50 75 110 150 150 150

    55 – (75) (110) (150) (150) (150)

    1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C

    Promass S

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    8 45 150 45 65 100 150 150 150

    50 – 65 100 150 150 150

    60 – 65 100 125 (150) (150)

    15…50 50 150 50 75 115 150 150 150

    60 – 75 115 125 (150) (150)

  • Proline Promass 300 XA01472D

    Endress+Hauser 19

    Promass X

    DN Ta[°C]

    Tm, max[°C]

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    350 50 180 50 90 120 170 180 180

    55 – (90) (120) (170) (180) (180)

    With thermal insulation without Endress+Hauser specifications

    1

    A B

    A0031198

    1 Position of the reference point for temperature measurement

    A Standard versionB Extended temperature version, cryogenic temperature version, high-

    temperature version1 Reference point

    Tm [°C]

    T6[85 °C]

    T5[95 °C]

    T4[130 °C]

    T3[195 °C]

    T2[290 °C]

    T1[440 °C]

    59 72 75 76 77 77

  • XA01472D Proline Promass 300

    20 Endress+Hauser

    Connection data:Signal circuits

    The following tables contain specifications which are dependent on thetransmitter type and its input and output assignment. Compare thefollowing specifications with those on the nameplate of the transmitter.

    Terminal assignment

    Transmitter: supply voltage, input/outputs

    HART

    Supply voltage Input/output 1 Input/output 2 Input/output 3

    1 (+) 2 (–) 26 (+) 27 (–) 24 (+) 25 (–) 22 (+) 23 (–)

    Device-specific terminal assignment: adhesive label in terminalcover.

    FOUNDATION Fieldbus

    Supply voltage Input/output 1 Input/output 2 Input/output 3

    1 (+) 2 (–) 26 (A) 27 (B) 24 (+) 25 (–) 22 (+) 23 (–)

    Device-specific terminal assignment: adhesive label in terminalcover.

    PROFIBUS PA

    Supply voltage Input/output 1 Input/output 2 Input/output 3

    1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–)

    Device-specific terminal assignment: adhesive label in terminalcover.

    Modbus RS485

    Supply voltage Input/output 1 Input/output 2 Input/output 3

    1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–)

    Device-specific terminal assignment: adhesive label in terminalcover.

  • Proline Promass 300 XA01472D

    Endress+Hauser 21

    Remote display and operating module DKX001

    1

    2

    3

    81

    Vcc

    82

    Gn

    d

    83

    A

    84

    B

    81

    Vcc

    82

    Gn

    d

    83

    A

    84

    B

    A0027518

    1 Remote display and operating module DKX0012 Connecting cable3 Measuring device

    Safety-related values

    Order code for"Output; input 1"

    Output type Safety-related values"Output; input 1"

    26 (+) 27 (–)

    Option BA Current output4 to 20 mA HART

    UN = 30 VUM = 250 V

    Option GA PROFIBUS PA UN = 32 VUM = 250 V

    Option MA Modbus RS485 UN = 30 VUM = 250 V

    Option SA FOUNDATIONFieldbus

    UN = 32 VUM = 250 V

    Order code for"Output; input 2";"Output; input 3"

    Output type Safety-related values

    Output; input 2 Output; input 3

    24 (+) 25 (–) 22 (+) 23 (–)

    Option B Current output4 to 20 mA

    UN = 30 VUM = 250 V

    Option D User configurableinput/output

    UN = 30 VUM = 250 V

  • XA01472D Proline Promass 300

    22 Endress+Hauser

    Order code for"Output; input 2";"Output; input 3"

    Output type Safety-related values

    Output; input 2 Output; input 3

    24 (+) 25 (–) 22 (+) 23 (–)

    Option E Pulse/frequency/switch output

    UN = 30 VUM = 250 V

    Option F Double pulse output UN = 30 VUM = 250 V

    Option H Relay output UN = 30 VIN = 100 mA DC/500 mA ACUM = 250 V

    Option I Current input4 to 20 mA

    UN = 30 VUM = 250 V

    Option J Status input UN = 30 VUM = 250 V

    Intrinsically safe values

    Order code for"Output; input 1"

    Output type Intrinsically safe values"Output; input 1"

    26 (+) 27 (–)

    Option CA Current output 4-20mA HART Ex i

    Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0

    Option HA PROFIBUS PA Ex i Ex iaUi = 30 Vli = 570 mAPi = 8.5 WLi = 10 µHCi = 5 nF

    Option TA FOUNDATIONFieldbus Ex i

    Ex iaUi = 30 Vli = 570 mAPi = 8.5 WLi = 10 µHCi = 5 nF

  • Proline Promass 300 XA01472D

    Endress+Hauser 23

    Order code for"Output; input 2";"Output; input 3"

    Output type Intrinsically safe values

    Output; input 2 Output; input 3

    24 (+) 25 (–) 22 (+) 23 (–)

    Option C Current output 4-20mA Ex i

    Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0

    Option G Pulse/frequency/switch output Ex i

    Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0

    Remote display DKX001

    Basic specification,position 1, 2Approval

    Terminalassignment

    Basic specification, position 8Display; OperationOption O

    Option C2, C4 81, 82, 83, 84

    A connecting cable with the value L/R ≤24 µH/Ω and Ccable ≤ 1 000 nF must be usedfor the version for connecting to the remotedisplay DKX001 or ODKX001.

  • www.addresses.endress.com