Date: October 24, 2006 1 of 17 Programming data for Spectrophotometer and Spectroquant ® Test Kits measured with a Reference Spectrophotometer Date: October 24, 2006 Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against No. range time min length Acid capacity CT to pH 4.3 (Total Alkalinity) 1.01762. 0.20-8.00 10-400 mmol/l OH mg/l CaCO 3 0 + 0 16 mm 10 mm 16 mm 10 mm 620 nm 4) dist. water dist. water dist. water dist. water Alcohol CT deleted, see Reflectoquant Test 1.16130.0001. Additionally RQflex required. Alkalinity see Acid capacity to pH 4.3 Aluminium Test 1.14825.0001 0.10-1.20 0.020-0.200 mg/l Al 2 + 0 10 mm 50 mm 550 nm 4) dist. water dist. water Aluminium CT 1.00594.0001 0.02 – 0.50 mg/l Al 5 + 0 16 mm 10 mm 545 nm 4) dist. water dist. water Ammonium CT 1.14739.0001 0.010-2.000 0.01-2.58 mg/l NH 4 -N mg/l NH 4 + 15 + 0 16 mm 10 mm 16 mm 10 mm 690 nm 0.877 1.18 1.13 1.53 cell with blank own blank cell with blank own blank Ammonium Test 1.14752.0001 0.05-3.00 0.03-1.50 0.010-0.500 0.06-3.86 0.04-1.93 0.013-0.644 mg/l NH 4 -N mg/l NH 4 + 5 + 5 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm 690 nm 1.23 0.615 0.246 1.58 0.792 0.316 own blank own blank own blank own blank own blank own blank Ammonium CT 1.14558.0001 0.20-8.00 0.26-10.30 mg/l NH 4 -N mg/l NH 4 + 15 + 0 16 mm 10 mm 16 mm 10 mm 690 nm 4.26 5.75 5.49 7.41 cell with blank own blank cell with blank own blank Ammonium CT 1.14544.0001 0.5-16.0 0.6-20.6 mg/l NH 4 -N mg/l NH 4 + 15 + 0 16 mm 10 mm 16 mm 10 mm 690 nm 7.84 10.6 10.1 13.6 cell with blank own blank cell with blank own blank Ammonium Test 1.00683.0001 2.0-75.0 5-150 2.6-96.6 6-193 mg/l NH 4 -N mg/l NH 4 + 15 + 0 10 mm 10 mm 10 mm 10 mm 690 nm 27.8 55.6 35.8 71.5 own blank own blank own blank own blank Ammonium CT 1.14559.0001 4.0-80.0 5.2-103.0 mg/l NH 4 -N mg/l NH 4 + 15 + 0 16 mm 10 mm 16 mm 10 mm 690 nm 36.4 49.1 46.8 63.3 cell with blank own blank cell with blank own blank AOX 1.00675.0001 0.05-2.50 mg/l AOX 15 + 0 16 mm 10 mm 445 nm 4.00 5.40 own blank own blank Antimony 2) 0.10-8.00 mg/l Sb 3 + 2 10 mm 620 nm 3.45 own blank Arsenic Test 1.01747.0001 0.005-0.100 0.001-0.020 mg/l As 120 + 0 10 mm 20 mm 525 nm 0.0910 0.0455 dist. water dist. water BOD CT 1.00687.0001 0.5-3000 mg/l BOD 0 + 0 16 mm 10 mm 500 nm 13.5 18.2 dist. water dist. water Boron Test 1.14839.0001 0.050-0.800 mg/l B 12 + 2 10 mm 565 nm 0.328 own blank Boron CT 1.00826.0001 0.05-2.00 mg/l B 60 + 0 16 mm 10 mm 405 nm 1.38 1.86 own blank own blank 2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on the Internet homepage http://photometry.merck.de . 4) Please see programming data under non-linear calibration or absorption-concentration- table. Remark: “own blank” means distilled water and all reagents in the same quantity as used for the measuring sample preparation. Separately prepared in standard test tubes and poured into the type of cell used. “cell with blank” means the white capped ready-to-use blank sample in cell tests for use with round cell measurement. The instrument must be equipped with a round cell holder. “dist. water” means distilled water without reagents and poured into the type of cell used. “clear sample without reag.” means sample must be colourless and turbid samples must be filtered.

Transcript of Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference...

Page 1: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 1 of 17

Programming data for Spectrophotometer and Spectroquant® Test Kits

measured with a Reference Spectrophotometer

Date: October 24, 2006

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against No. range time min length

Acid capacity CT to pH 4.3

(Total Alkalinity)

1.01762. 0.20-8.00

10-400

mmol/l OH

mg/l CaCO3

0 + 0 16 mm 10 mm 16 mm 10 mm

620 nm 4) dist. water dist. water dist. water dist. water

Alcohol CT deleted, see Reflectoquant Test 1.16130.0001. Additionally RQflex required. Alkalinity see Acid capacity to pH 4.3

Aluminium Test 1.14825.0001 0.10-1.20 0.020-0.200

mg/l Al 2 + 0 10 mm 50 mm

550 nm 4) dist. water dist. water

Aluminium CT 1.00594.0001 0.02 – 0.50 mg/l Al 5 + 0 16 mm 10 mm

545 nm 4) dist. water dist. water

Ammonium CT 1.14739.0001 0.010-2.000

0.01-2.58

mg/l NH4-N

mg/l NH4+

15 + 0 16 mm 10 mm 16 mm 10 mm

690 nm 0.877 1.18 1.13 1.53

cell with blank own blank

cell with blank own blank

Ammonium Test 1.14752.0001 0.05-3.00 0.03-1.50

0.010-0.500 0.06-3.86 0.04-1.93

0.013-0.644

mg/l NH4-N

mg/l NH4+

5 + 5 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

690 nm 1.23 0.615 0.246 1.58

0.792 0.316

own blank own blank own blank own blank own blank own blank

Ammonium CT 1.14558.0001 0.20-8.00

0.26-10.30

mg/l NH4-N

mg/l NH4+

15 + 0 16 mm 10 mm 16 mm 10 mm

690 nm 4.26 5.75 5.49 7.41

cell with blank own blank

cell with blank own blank

Ammonium CT 1.14544.0001 0.5-16.0

0.6-20.6

mg/l NH4-N

mg/l NH4+

15 + 0 16 mm 10 mm 16 mm 10 mm

690 nm 7.84 10.6 10.1 13.6

cell with blank own blank

cell with blank own blank

Ammonium Test 1.00683.0001 2.0-75.0 5-150

2.6-96.6 6-193

mg/l NH4-N

mg/l NH4+

15 + 0 10 mm 10 mm 10 mm 10 mm

690 nm 27.8 55.6 35.8 71.5

own blank own blank own blank own blank

Ammonium CT 1.14559.0001 4.0-80.0

5.2-103.0

mg/l NH4-N

mg/l NH4+

15 + 0 16 mm 10 mm 16 mm 10 mm

690 nm 36.4 49.1 46.8 63.3

cell with blank own blank

cell with blank own blank

AOX 1.00675.0001 0.05-2.50 mg/l AOX 15 + 0 16 mm 10 mm

445 nm 4.00 5.40

own blank own blank

Antimony 2) 0.10-8.00 mg/l Sb 3 + 2 10 mm 620 nm 3.45 own blank Arsenic Test 1.01747.0001 0.005-0.100

0.001-0.020 mg/l As 120 + 0 10 mm

20 mm 525 nm 0.0910

0.0455 dist. water dist. water

BOD CT 1.00687.0001 0.5-3000 mg/l BOD 0 + 0 16 mm 10 mm

500 nm 13.5 18.2

dist. water dist. water

Boron Test 1.14839.0001 0.050-0.800 mg/l B 12 + 2 10 mm 565 nm 0.328 own blank Boron CT 1.00826.0001 0.05-2.00 mg/l B 60 + 0 16 mm

10 mm 405 nm 1.38

1.86 own blank own blank

2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

the Internet homepage http://photometry.merck.de. 4) Please see programming data under non-linear calibration or absorption-concentration- table. Remark: “own blank” means distilled water and all reagents in the same quantity as used for the measuring sample preparation. Separately prepared in standard test tubes and poured into the type of cell used. “cell with blank” means the white capped ready-to-use blank sample in cell tests for use with round cell measurement. The instrument must be equipped with a round cell holder. “dist. water” means distilled water without reagents and poured into the type of cell used. “clear sample without reag.” means sample must be colourless and turbid samples must be filtered.

Page 2: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 2 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Bromine Test 1.00605.0001 0.10-10.00

0.05-5.00 0.020-2.000

mg/l Br2 3 + 0 10 mm 20 mm 50 mm

550 nm 7.52 3.76 1.50

dist. water dist. water dist. water

Cadmium Test 1.01745.0001 0.010-0.500 0.005-0.250 0.002-0.100

mg/l Cd 2 + 0 10 mm 20 mm 50 mm

525 nm 0.680 0.340 0.136

own blank own blank own blank

Cadmium CT 1.14834.0001 0.025-1.000 mg/l Cd 2 + 0 16 mm 10 mm

550 nm 0.595 0.804

own blank own blank

Calcium Test 1.14815.0001 10-160 25-400 14-224

5-80 12-200 7-112

1.0-15.0 1)

2.5-37.5 1)

1.4-21.0 1)

mg/l Ca mg/l CaCO3mg/l CaO mg/l Ca

mg/l CaCO3mg/l CaO mg/l Ca

mg/l CaCO3mg/l CaO

8 + 0 10 mm 10 mm 10 mm 20 mm 20 mm 20 mm 10 mm 10 mm 10 mm

550 nm 137 342 192 68.5 171 95.9 29.4 73.5 41.2

own blank own blank own blank own blank own blank own blank own blank own blank own blank

Calcium CT 1.00858.0001 10-250 25-625 14-350 10-250 25-625 14-350

mg/l Ca mg/l CaCO3mg/l CaO mg/l Ca

mg/l CaCO3mg/l CaO

3 + 0 16 mm 16 mm 16 mm 10 mm 10 mm 10 mm

565 nm 4) dist. water dist. water dist. water dist. water dist. water dist. water

Chloride CT 1.14730.0001 5-125 mg/l Cl- 0 + 0 16 mm 10 mm

525 nm 114 153

own blank own blank

Chloride Test 1.14897.0001 10-250 2.5-25.0

mg/l Cl- 1 + 0 10 mm 10 mm

500 nm 445 nm

107 28.2

own blank own blank

Chlorine Test (free chlorine)

1.00598. 0.05-6.00 0.02-3.00

0.010-1.000

mg/l Cl2 3 + 0 10 mm 20 mm 50 mm

550 nm 3.33 1.67

0.667

dist. water dist. water dist. water

Chlorine CT (free chlorine)

1.00595.0001 0.03-6.00 mg/l Cl2 3 + 0 16 mm 550 nm 2.47 dist. water

Chlorine Test (total Chlorine)

1.00602. 0.05-6.00 0.02-3.00

0.010-1.000

mg/l Cl2 3 + 0 10 mm 20 mm 50 mm

550 nm 3.33 1.67

0.667

dist. water dist. water dist. water

Chlorine Test (free+total Chlorine)

1.00599.0001 0.05-6.00 0.02-3.00

0.010-1.000

mg/l Cl2 3 + 0 10 mm 20 mm 50 mm

550 nm 3.33 1.67

0.667

dist. water dist. water dist. water

Chlorine CT (free+total chlorine)

1.00597.0001 0.03-6.00 mg/l Cl2 3 + 0 16 mm 550 nm 2.47 dist. water

Chlorine (liquid) (free+total Chlorine)

1.00086.00011.00087.00011.00088.00011.00089.0001

0.03-6.00 0.010-1.000

mg/l Cl2 3 + 0 16 mm 50 mm

550 nm 2.47 0.667

dist. water dist. water

Chlorine Test 1.14828.0001 replaced by 100599 or 100598 and 100602 Chlorine dioxide

Test 1.00608.0001 0.10-10.00

0.05-5.00 0.020-2.000

mg/l ClO2 3 + 0 10 mm 20 mm 50 mm

550 nm 6.33 3.16 1.27

dist. water dist. water dist. water

Chlorine dioxide Test

1.14732.0001 replaced by 100608

Chromate Test 1.14758.0001 0.05-3.00 0.03-1.50

0.010-0.600 0.11-6.69 0.07-3.35 0.02-1.34

mg/l Cr

mg/l CrO4

0 + 0 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

550 nm 1.30 0.650 0.260 2.90 1.45

0.579

dist. water dist. water dist. water dist. water dist. water dist. water

Chromate CT 1.14552.0001 0.05-2.00

0.11-4.46

mg/l Cr

mg/l CrO4

0 + 0 16 mm 10 mm 16 mm 10 mm

550 nm 0.971 1.31 2.17 2.92

cell with blank dist. water

cell with blank dist. water

Chromium bath 2) 20-400 10-200

4.0-80.0

g/l CrO3 0 + 0 10 mm 20 mm 50 mm

445 nm 556 278 111

dist. water dist. water dist. water

1) Handling details: see Appendix 2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

the Internet homepage http://photometry.merck.de. 4) Please see programming data under non-linear calibration or absorption-concentration- table.

Page 3: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 3 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length COD CT 1.14560.0001 4.0-40.0 mg/l COD 0 + 0 16 mm

10 mm 340 nm -41.7

-56.3 own blank own blank

COD CT 1.14540.0001 10-150 mg/l COD 0 + 0 16 mm 10 mm

445 nm -217 -295

own blank own blank

COD CT 1.14895.0001 15-300 mg/l COD 0 + 0 16 mm 10 mm

445 nm -222 -300

own blank own blank

COD CT 1.14690.0001 50-500 mg/l COD 0 + 0 16 mm 10 mm

445 nm -397 -536

own blank own blank

COD CT 1.14541.0001 25-1500 mg/l COD 0 + 0 16 mm 10 mm

605 nm 1667 2249

own blank own blank

COD CT 1.14691.0001 300-3500 mg/l COD 0 + 0 16 mm 10 mm

605 nm 3226 4355

own blank own blank

COD CT 1.14555.0001 500-10000 mg/l COD 0 + 0 16 mm 10 mm

605 nm 4608 6221

own blank own blank

COD CT (Hg free)

1.09772.0001 10-150 mg/l COD 0 + 0 16 mm 10 mm

445 nm -238 -321

own blank own blank

COD CT (Hg free)

1.09773.0001 100-1500 mg/l COD 0 + 0 16 mm 10 mm

605 nm 1923 2596

own blank own blank

Colour 0.5-50.0 m-1 0 + 0 50 mm 445 mm525 nm 620 nm

20.0 20.0 20.0

dist. water dist. water dist. water

Colour measurement

see Hazen Colour number Pt/Co or HZ

Copper Test 1.14767.0001 0.10-6.00 0.05-3.00 0.02-1.20

mg/l Cu 5 + 0 10 mm 20 mm 50 mm

605 nm 4.44 2.22

0.889

dist. water dist. water dist. water

Copper CT 1.14553.0001 0.05-8.00 mg/l Cu 5 + 0 16 mm 10 mm

605 nm 3.60 4.86

cell with blank dist. water

Copper bath 2) 10.0-80.0 5.0-40.0 2.0-16.0

g/l Cu 0 + 0 10 mm 20 mm 50 mm

820 nm 43.1 21.6 8.62

dist. water dist. water dist. water

Cyanide Test 1.09701.0001 0.010-0.500 0.005-0.250 0.002-0.100

mg/l CN- 10 + 0 10 mm 20 mm 50 mm

605 nm 0.221 0.110

0.0442

dist. water dist. water dist. water

Cyanide Test 1.14800.0001 0.010-0.500 0.005-0.250 0.002-0.100

mg/l CN- 5 + 0 10 mm 20 mm 50 mm

585 nm 0.208 0.104

0.0416

dist. water dist. water dist. water

Cyanide CT 1.14561.0001 0.010-0.500 mg/l CN- 10 + 0 16 mm 10 mm

605 nm 0.161 0.218

dist. water dist. water

Detergents see Surfactants Dissolved Oxygen see Oxygen

Ethanol see Alcohol Fluoride CT 1.14557.0001 0.10-1.50

0.025-0.500 1)

mg/l F- 5 + 0 16 mm 10 mm 50 mm

620 nm 1.33 1.80

0.323

own blank own blank own blank

Fluoride Test 1.14598.0001 0.10-2.00 1.0-20.0

mg/l F- 5 + 0 10 mm 10 mm

620 nm 2.02 21.5

own blank own blank

Formaldehyde Test

1.14678.0001 0.10-8.00 0.05-4.00 0.02-1.50

mg/l HCHO 10 + 0 10 mm 20 mm 50 mm

565 nm 4.22 2.11

0.844

dist. water dist. water dist. water

Formaldehyde CT

1.14500.0001 0.10-8.00 mg/l HCHO 5 + 0 16 mm 10 mm

565 nm 3.23 4.36

dist. water dist. water

Gold Test 1.14821.0001 0.5-12.0 mg/l Au 0 + 0 10 mm 550 nm 6.25 own blank Hardness see Residual hardness or Total hardness

Hazen colour number

0-500 0-250 0-100

HZ or Pt/Co 0 + 0 10 mm 20 mm 50 mm

340 nm 370 185 74

dist. water dist. water dist. water

Hazen colour number

0-1000 HZ or Pt/Co 0 + 0 50 mm 445 nm 769 dist. water

Hazen colour number

0-1000 HZ or Pt/Co 0 + 0 50 mm 455 nm 725 dist. water

Hazen colour number

0-1000 HZ or Pt/Co 0 + 0 50 mm 465 nm 741 dist. water

1) Handling details: see Appendix 2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

the Internet homepage http://photometry.merck.de.

Page 4: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 4 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Hydrazine Test 1.09711.0001 0.02-2.00

0.01-1.00 0.005-0.400

mg/l N2H4 5 + 0 10 mm 20 mm 50 mm

445 nm 0.870 0.435 0.174

dist. water dist. water dist. water

Hydrazine Test 1.14797.0001 replaced by 1.09711.0001 Hydrogen peroxide

CT 1.14731.0001 2.0-20.0

0.25-5.00

mg/l H2O2 2 + 0 16 mm 10 mm 50 mm

410 nm 38.5 52.3 10.5

cell with blank dist. water dist. water

Hydrogen peroxide Test

1.18789.0001 0.03 – 6.00 0.015 – 3.000

mg/l H2O2 10 + 0 10 mm 20 mm

445 nm 2.67 1.34

own blank own blank

Hydrogen Sulfide see Sulfide Iodine Test 1.00606.0001 0.20-10.00

0.10-5.00 0.050-2.000

mg/l I2 3 + 0 10 mm 20 mm 50 mm

550 nm 11.9 5.95 2.38

dist. water dist. water dist. water

Iodine colour number

0.05-3.00 0.03-1.50

0.010-0.600

IFZ 0 + 0 10 mm 20 mm 50 mm

340 nm 1.27 0.637 0.255

dist. water dist. water dist. water

Iodine colour number

1.0-50.0 0.5-25.0 0.2-10.0

IFZ 0 + 0 10 mm 20 mm 50 mm

445 nm 17.1 8.55 3.42

dist. water dist. water dist. water

Iron Test 1.14761. 0.05-5.00 0.03-2.50

0.005-1.000

mg/l Fe 3 + 0 10 mm 20 mm 50 mm

565 nm 2.08 1.04

0.416

dist. water dist. water dist. water

Iron Test 1.00796.0001 0.10-5.00 0.05-2.50

0.010-1.000

mg/l Fe 5 + 10 10 mm 20 mm 50 mm

500 nm 5.56 2.78 1.11

dist. water dist. water dist. water

Iron CT 1.14549.0001 0.05-4.00 0.05–5.00

mg/l Fe 3 + 0 16 mm 10 mm

565 nm 1.64 2.21

cell with blank dist. water

Iron CT 1.14896.0001 1.0-50.0 mg/l Fe 5 + 0 16 mm 10 mm

525 nm 28.3 38.2

cell with blank dist. water

Lead Test 1.09717.0001 0.10-5.00 0.05-2.50

0.010-1.000

mg/l Pb 0 + 0 10 mm 20 mm 50 mm

525 nm 6.25 3.13 1.25

own blank own blank own blank

Lead CT 1.14833.0001 0.10-5.00 mg/l Pb 0 + 0 16 mm 10 mm

525 nm 4.39 5.93

cell with blank own blank

Mercury 2) 0.025-1.000 mg/l Hg 5 + 0 50 mm 565 nm 0.563 own blank Magnesium CT 1.00815.0001 5.0-75.0 mg/l Mg 3 + 0 16 mm

10 mm 565 nm 4) dist. water

dist. water Manganese

Test 1.01739.0001 0.05-2.00

0.03-1.00 0.005-0.400

mg/l Mn 10 + 0 10 mm 20 mm 50 mm

565 nm 1.72 0.862 0.345

own blank own blank own blank

Manganese Test

1.14770.0001 0.50-10.00 0.25-5.00 0.01-2.00

mg/l Mn 2 + 2 10 mm 20 mm 50 mm

445 nm 5.62 2.81 1.12

dist. water dist. water dist. water

Manganese CT 1.00816.0001 0.10-5.00 mg/l Mn 2 + 5 16 mm 10 mm

445 nm 4.13 5.58

cell with blank dist. water

Molybdenum CT 1.00860.0001 0.02-1.00 mg/l Mo 2 + 0 16 mm 10 mm

620 nm 1.10 1.49

own blank own blank

Monochloramine Test

1.01632.0001 0.25-10.00 0.13-5.00

0.050-2.000 0.18-7.25 0.09-3.63

0.036-1.450 0.05-1.96 0.03-0.98

0.010-0.392

mg/l Cl2

mg/l NH2Cl

mg/l NH2Cl-N

5 + 5 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

690 nm 5.00 2.50 1.00 3.67 1.83

0.733 0.988 0.494 0.198

own blank own blank own blank own blank own blank own blank own blank own blank own blank

Nickel Test 1.14785.0001 0.10-5.00 0.05-2.50 0.02-1.00

mg/l Ni 1 + 2 10 mm 20 mm 50 mm

445 nm 4.76 2.38

0.952

dist. water dist. water dist. water

Nickel CT 1.14554.0001 0.10-6.00 mg/l Ni 1 + 2 16 mm 10 mm

445 nm 3.82 5.16

cell with blank own blank

Nickel bath 2) 10-120 5.0-60.0 2.0-24.0

g/l Ni 0 + 0 10 mm 20 mm 50 mm

690 nm 60.6 30.3 12.1

dist. water dist. water dist. water

2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

the Internet homepage http://photometry.merck.de. 4) Please see programming data under non-linear calibration or absorption-concentration- table.

Page 5: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 5 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Nitrate Test 1.09713.0001 1.0-25.0

0.5-12.5 0.10-5.00 4.4-110.7 2.2-55.4 0.4-22.1

mg/l NO3-N

mg/l NO3-

10 + 0 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

340 nm 19.6 9.80 3.92 86.8 43.4 17.4

own blank own blank own blank own blank own blank own blank

Nitrate Test 1.14773.0001 0.5-20.0 0.2-10.0 2.2-88.5 0.9-44.3

mg/l NO3-N

mg/l NO3-

10 + 0 10 mm 20 mm 10 mm 20 mm

525 nm 9.62 4.81 42.6 21.3

own blank own blank own blank own blank

Nitrate CT 1.14542.0001 0.5-18.0

2.2-79.7

mg/l NO3-N

mg/l NO3-

10 + 0 16 mm 10 mm 16 mm 10 mm

525 nm 7.14 9.64 31.6 42.7

cell with blank own blank

cell with blank own blank

Nitrate CT 1.14563.0001 0.5-25.0

2.2-110.7

mg/l NO3-N

mg/l NO3-

10 + 0 16 mm 10 mm 16 mm 10 mm

340 nm 14.7 19.9 65.1 87.9

cell with blank own blank

cell with blank own blank

Nitrate CT 1.14764.0001 1.0-50.0

4-221

mg/l NO3-N

mg/l NO3-

10 + 0 16 mm 10 mm 16 mm 10 mm

340 nm 27.9 37.7 124 167

cell with blank own blank

cell with blank own blank

Nitrate CT 1.00614.0001 23-225

102-996

mg/l NO3-N

mg/l NO3

5 + 0 16 mm 10 mm 16 mm 10 mm

340 nm 132 180 586 797

cell with blank own blank

cell with blank own blank

Nitrate CT in seawater

1.14556.0001 0.10-3.00 0.10-2.50 0.4-13.3 0.4-11.1

mg/l NO3-N

mg/l NO3-

30 + 0 16 mm 10 mm 16 mm 10 mm

500 nm 4) dist. water dist. water dist. water dist. water

Nitrate Test in seawater

1.14942.0001 0.2-17.0 0.9-75.3

mg/l NO3-N mg/l NO3

-15 + 60 10 mm

10 mm 500 nm 4) dist. water

dist. water Nitrite Test 1.14776. 0.02-1.00

0.010-0.500 0.002-0.200

0.07-3.28 0.03-1.64

0.007-0.657

mg/l NO2-N

mg/l NO2-

10 + 0 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

525 nm 0.370 0.185

0.0741 1.22

0.608 0.243

dist. water dist. water dist. water dist. water dist. water dist. water

Nitrite CT 1.14547.0001 0.010-0.700

0.03-2.30

mg/l NO2-N

mg/l NO2-

10 + 0 16 mm 10 mm 16 mm 10 mm

525 nm 0.274 0.370 0.900 1.22

cell with blank dist. water

cell with blank dist. water

Nitrite CT 1.00609.0001 1.0-90.0

3.3-295.2

mg/l NO2-N

mg/l NO2-

20 + 0 16 mm 10mm 16 mm 10mm

585 nm 80.2 109.9 263.4 360.9

dist. water dist. water dist. water dist. water

Nitrogen, total CT 1.00613.0001 0.5-15.0 mg/l N 10 + 0 16 mm 10 mm

340 nm 15.3 20.6

cell with blank own blank

Nitrogen, total CT 1.14537.0001 0.5-15.0 mg/l N 10 + 0 16 mm 10 mm

525 nm 7.81 10.5

own blank own blank

Nitrogen, total CT 1.14763.0001 10-150 mg/l N 10 + 0 16 mm 10 mm

340 nm 154 208

cell with blank own blank

Nitrogen, Ammonium

see Ammonium

Nitrogen, Nitrate see Nitrate Nitrogen, Nitrite see Nitrite Organic Acids,

Volatile see Volatile organic acids

Organic Carbon, Total

see TOC

Oxygen, dissolved CT

1.14694.0001 0.5-12.0 mg/l O2 0 + 0 16 mm 10 mm

500 nm 13.5 18.2

dist. water dist. water

Oxygen Demand, Biological

see BOD

Oxygen Demand, Chemical

see COD

4) Please see programming data under non-linear calibration or absorption-concentration- table.

Page 6: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 6 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Ozone Test 1.00607. 0.05-4.00

0.02-2.00 0.010-0.800

mg/l O3 3 + 0 10 mm 20 mm 50 mm

550 nm 2.25 1.13

0.450

dist. water dist. water dist. water

Ozone Test 1.14732.0001 replaced by 100607 Palladium 2) 0.05-1.25 mg/l Pd 1 + 0 10 mm 525 nm 0.588 own blank Peroxide see Hydrogen Peroxide pH CT 1.01744.0001 6.4-8.8 pH units 0 + 0 16 mm

10 mm 550 nm 4) dist. water

dist. water Phenol Test 1.00856.0001 0.002-0.100

0.10-5.00 0.013-2.500 0.025-1.000

mg/l Phenol 30 + 10 10 + 0

20 mm 10 mm 20 mm 50 mm

445 nm 495 nm

0.114 8.33 4.17 1.67

own blank own blank own blank own blank

Phenol CT 1.14551.0001 0.10-2.50 mg/l Phenol 1 + 0 16 mm 10 mm

500 nm 3.39 4.58

cell with blank own blank

Phosphate Test (PMB)

1.14848.0001 0.05-5.00 0.03-2.50

0.010-1.000 0.2-15.3

0.09-7.67 0.03-3.07

0.11-11.46 0.07-5.73 0.02-2.29

mg/l PO4-P

mg/l PO43-

mg/l P2O5

5 + 0 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

690 nm 2.12 1.06

0.424 6.50 3.25 1.30 4.85 2.43

0.971

own blank own blank own blank own blank own blank own blank own blank own blank own blank

Phosphate CT (PMB)

1.14543.0001 0.05-5.00

0.2-15.3

0.11-11.46

mg/l PO4-P

mg/l PO43-

mg/l P2O5

5 + 0 16 mm 10 mm 16 mm 10 mm 16 mm 10 mm

690 nm 1.61 2.18 4.95 6.68 3.70 4.99

cell with blank own blank

cell with blank own blank

cell with blank own blank

Phosphate CT (PMB)

1.14729.0001 0.5-25.0

1.5-76.7

1.1-57.3

mg/l PO4-P

mg/l PO43-

mg/l P2O5

5 + 0 16 mm 10 mm 16 mm 10 mm 16 mm 10 mm

690 nm 8.06 10.9 24.7 33.4 18.5 24.9

cell with blank own blank

cell with blank own blank

cell with blank own blank

Phosphate CT (VM)

1.14546.0001 0.5-25.0 1.0-33.0 1.5-75.0 3.0-90.0 1.1-57.3 1.5-77.4

mg/l PO4-P

mg/l PO43-

mg/l P2O5

0 + 0 16 mm 10 mm 16 mm 10 mm 16 mm 10 mm

410 nm 13.4 18.2 41.2 55.6 30.8 41.5

cell with blank own blank

cell with blank own blank

cell with blank own blank

Phosphate Test (VM)

1.14842.0001 1.0-30.0 0.5-15.0 3.1-92.0 1.5-46.0 2.3-68.7 1.1-34.4

mg/l PO4-P

mg/l PO43-

mg/l P2O5

0 + 0 10 mm 20 mm 10 mm 20 mm 10 mm 20 mm

410 nm 18.0 9.00 55.2 27.6 41.3 20.6

own blank own blank own blank own blank own blank own blank

Phosphate Test (PMB)

1.00798.0001 1.0-100.0 3-307 2-229

mg/l PO4-P mg/l PO4

3- mg/l P2O5

5 + 0 10 mm 10 mm 10 mm

690 nm 35.1 108 80.4

own blank own blank own blank

Phosphate CT (PMB)

1.00616.0001 3.0-100.0

9-307

7-229

mg/l PO4-P

mg/l PO43-

mg/l P2O5

5 + 0 16 mm 10 mm 16 mm 10 mm 16 mm 10 mm

690 nm 39.2 52.9 120 162 89.9 121

cell with blank own blank

cell with blank own blank

cell with blank own blank

Platinum 2) 0.10-1.25 mg/l Pt 5 + 1 10 mm 690 nm 2.38 own blank Platinium-Cobalt

Standard see Hazen Colour number Pt/Co or HZ

2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

Internet homepage http://photometry.merck.de analysis. 3) This methods needs to be instrument-specific calibrated (see appendix). 4) Please see programming data under non-linear calibration or absorption-concentration- table.

Page 7: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 7 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Potassium CT 1.14562.0001 5.0-50.0 mg/l K 5 + 0 16 mm

10 mm 690 nm turbidity

3)dist. water dist. water

Potassium CT 1.00615.0001 30-300 mg/l K 5 + 0 16 mm 10 mm

690 nm turbidity 3)

dist. water dist. water

Residual Hardness CT

1.14683.0001 0.50-5.00 1.25-12.50 0.70-7.00

0.070-0.700 0.088-0.875 0.125-1.250

0.50-5.00 1.25-12.50 0.70-7.00

0.070-0.700 0.088-0.875 0.125-1.250

mg/l Ca mg/l CaCO3mg/l CaO

°d °e °f

mg/l Ca mg/l CaCO3mg/l CaO

°d °e °f

10 + 0 16 mm 16 mm 16 mm 16 mm 16 mm 16 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm

565 nm 4) dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water

Silicate Test 1.14794.0001 0.10-5.00 0.05-2.50

0.005-0.750 0.21-10.70 0.11-5.35 0.01-1.60

mg/l Si

mg/l SiO2

3 + 10 10 mm 20 mm 50 mm 10 mm 20 mm 50 mm

665 nm

820 nm 665 nm

820 nm

3.73 1.87

0.286 7.98 3.99

0.611

dist. water dist. water dist. water dist. water dist. water dist. water

Silicate Test 1.00857.0001 0.5-50 5-500

1.1-107.0 11-1070

mg/l Si

mg/l SiO2

2 + 2 10 mm 10 mm 10 mm 10 mm

410 nm 28.2 279 60.3 598

dist. water dist. water dist. water dist. water

Silver Test 1.14831.0001 0.50-3.00 0.25-1.50

mg/l Ag 5 + 0 10 mm 20 mm

550 nm 2.22 1.11

own blank own blank

Sodium CT in nutrient solutions for

fertilization

1.00885.0001 10-300 mg/l Na 1 + 0 16 mm 10 mm

550 nm 263 355

own blank own blank

Sulfate CT 1.14548.0001 5-250 mg/l SO42- 2 + 0 16 mm

10 mm 525 nm turbidity

3)Clear sample without reag. Clear sample without reag.

Sulfate Test 1.14791.0001 25-300 mg/l SO42- 5 + 7 10 mm 525 nm 256 own blank

Sulfate CT 1.00617.0001 50-500 mg/l SO42- 2 + 0 16 mm

10 mm 525 nm turbidity

3)Clear sample without reag. Clear sample without reag.

Sulfate CT 1.14564.0001 100-1000 mg/l SO42- 2 + 0 16 mm

10 mm 820 nm turbidity

3)Clear sample without reag. Clear sample without reag.

Sulfide Test 1.14779.0001 0.10-1.50 0.020-0.500

mg/l S2- 0 + 0 10 mm 50 mm

665 nm 4)

0.244 dist. water dist. water

Sulfite CT 1.14394.0001 1.0-20.0

0.05-3.00 1)

mg/l SO32- 2 + 0 16 mm

10 mm 50 mm

410 nm 8.77 11.8 1.69

cell with blank own blank own blank

Sulfite Test 1.01746.0001 1.0-60.0 0.8-48.0

mg/l SO32-

mg/l SO2

2 + 0 10 mm 10 mm

410 nm 29.4 23.5

own blank own blank

Surfactants (anionic) CT

1.14697.0001 0.05-2.00 mg/l MBAS 10 + 0 16 mm 650 nm 1.80 cell with blank

Surfactants (cationic) CT

1.01764.0001 0.05-1.50 mg/l CTAB 5 + 0 16 mm 620 nm 1.92 cell with blank

Surfactants (nonionic) CT

1.01787.0001 0.10-7.50 mg/l Triton X-100

2 + 0 16 mm 605 nm 6.06 own blank

Suspended Solids 2) 25-750 mg/l SuS 0 + 0 20 mm 820 nm turbidity 3)

dist. water

Tensides see Surfactants Tin CT 1.14622.0001 0.10-2.50 mg/l Sn 15 + 0 16 mm

10 mm 665 nm 4) dist. water

dist. water TOC CT 1.14878.0001 5.0-80.0 mg/l TOC 0 + 0 16 mm 605 nm -64.5 own blank TOC CT 1.14879.0001 50-800 mg/l TOC 0 + 0 16 mm 605 nm -645 own blank

Total Alkalinity see Acid capacity to pH 4.3

2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

Internet homepage http://photometry.merck.de analysis. 3) This methods needs to be instrument-specific calibrated (see appendix). 4) Please see programming data under non-linear calibration or absorption-concentration- table.

Page 8: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 8 of 17

Parameter Catalogue Measuring Unit Reaction Cells Wave- Factor Measuring against

No. range time min length Total Hardness

CT 1.00961.0001 5-215

13-538 7-301

0.7-30.1 0.9-37.6 1.3-53.8

5-215 13-538 7-301

0.7-30.1 0.9-37.6 1.3-53.8

mg/l Ca mg/l CaCO3mg/l CaO

°d °e °f

mg/l Ca mg/l CaCO3mg/l CaO

°d °e °f

3 + 0 16 mm 16 mm 16 mm 16 mm 16 mm 16 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm

565 nm 4) dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water dist. water

Total Nitrogen CT see Nitrogen, total Turbidity 2) 1-100 FAU 0 + 0 50 mm 550 nm turbidity

3)dist. water

Volatile Organic Acids CT

1.01763.0001 50-3000 mg/l HOAc 3 + 10 16 mm 10 mm

495 nm 1430 1931

own blank own blank

Water Hardness see Residual hardness or Total hardness Zinc CT 1.00861.0001 0.025-1.000 mg/l Zn 15 + 0 16 mm

10 mm 500 nm 1.36

1.84 own blank own blank

Zinc Test 1.14832.0001 0.05-2.50 mg/l Zn 3 + 2 10 mm 565 nm 1.08 dist. water Zinc CT 1.14566.0001 0.20-5.00 mg/l Zn 15 + 0 16 mm

10 mm 500 nm 4.88

6.59 cell with blank

own blank 2) These parameters are not ready-to-use test kits. You will find all detailed information in our application list on

Internet homepage http://photometry.merck.de analysis. 3) This methods needs to be instrument-specific calibrated (see appendix). 4) Please see programming data under non-linear calibration or absorption-concentration- table. Appendix: Measuring range with appendix 1)

Each test kit contains a package insert with detailed information regarding the standard procedure of the measurement. In some cases, however, special points have to be considered which are not mentioned in the package inserts. In this appendix you will find specific applications and modified methods for analyses for sensitive measurements indicated with a 1).

Calcium: with Spectroquant® Calcium Test, Cat. No. 1.14815.0001 Measuring range: 1.0 – 15.0 mg/l Ca 2.5 - 37.5 mg/l CaCO3 1.4 - 21.0 mg/l CaO Procedure: Add 10 ml of reagent Ca-1 and 8 drops each of reagent Ca-2 and Ca-3 to 1 ml of the sample solution and mix. Leave to stand for 8 min for colour development. After filling into 10 mm cells measure in the photometer at 550 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 1 ml of distilled water instead of the sample solution. Fluoride: with Spectroquant® Fluoride Cell Test, Cat. No. 1.14557.0001 Measuring range: 0.025 – 0.500 mg/l F-

Procedure: Pipette 10 ml of sample solution into the reaction cell, mix and add 1 blue microspoon of reagent F-1K. Close the cell tightly and dissolve the solid by shaking. Leave to stand for 5 min for colour development. After filling into 50 mm cells measure in the photometer at 620 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 10 ml of distilled water instead of the sample solution.

Page 9: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 9 of 17

Sulfite: with Spectroquant® Sulfite Cell Test, Cat. No. 1.14394.0001 Measuring range: 0.005 – 3.00 mg/l SO3

2-

Procedure: Place 1 microspoon of reagent SO3-1K into the cell, close tightly and shake until the complete dissolution of the solid. Then pipette 7 ml of the sample solution into the cell and mix. After the reaction time of 2 min fill into 50 cells and measure in the photometer at 410 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 7 ml of distilled water instead of the sample solution. Alternative methods 2)

There are no test kits available for this parameter. The methods used there are reported in analytical literature. Applications and instructions how to perform the methods are available on the Internet homepage of Merck KGaA http://photometry.merck.de. Turbidity measurement 3)

The parameters Potassium Cat. No. 1.14562.0001 Potassium Cat. No. 1.00615.0001 Sulfate Cat. No. 1.14548.0001 Sulfate Cat. No. 1.00617.0001 Sulfate Cat. No. 1.14564.0001 Suspended Solids (physical measurement - application) Turbidity (physical measurement - application) are methods where the turbidity is measured. As the optical system of every spectrophotometer is different, we convince you to determine the factor for your spectrophotometer by your own and program it in your photometer. Non-linear calibration or absorption-concentration- table. 4)

Calibration curves for the methods are non-linear. 1. Use the below mentioned concentration-absorption-table 2. Beside using the concentration-absorption-table you can enter a non-linear curve by using the coefficients of

the following formula. C = a + bx + cx2 + dx3 Check your instrument and enter the coefficients according the given list. The calibration is always done against distilled water.

3. We recommend to take a standard and check if the calibration is done correctly because also spectrophotometers have small differences from instrument to instrument.

Page 10: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 10 of 17

Absorption-Concentration-Table for Acid capacity Cell Test, Cat. No. 1.01762.0001, calibration against distilled water. Absorption 16 mm

round cell mmol/l OH mg/l CaCO3

- Absorption 10 mm cell

mmol/l OH mg/l CaCO3-

0.544 0 0 0.400 0 0 0.571 0.20 10 0.420 0.20 10 0.680 1.00 50 0.500 1.00 50 0.816 1.90 95 0.600 1.90 95 0.979 2.80 140 0.720 2.80 140 1.129 3.70 185 0.830 3.70 185 1.292 4.50 225 0.950 4.50 225 1.469 5.40 270 1.080 5.40 270 1.632 6.30 315 1.200 6.30 315 1.782 7.20 360 1.310 7.20 360 1.904 8.00 400 1.400 8.00 400

Non-linear curve for Acid capacity Cell Test, Cat. No. 1.01762.0001, calibration against distilled water.

Non-linear 16 mm round cell

16 mm round cell

10 mm cell

10 mm cell

calibration 0.20 – 8.00 mmol/l OH

10 – 400 mg/l CaCO3

-0.20 – 8.00 mmol/l OH

10 – 400 mg/l CaCO3

-

coefficient a: -5.582 -279.1 -5.581 -279.0 coefficient b: 13.170 658.5 17.899 895.0 coefficient c: -6.232 -311.6 -11.509 -575,4 coefficient d: 1.608 80.4 4.038 201.9

Absorption-Concentration- Table for Aluminium Cell Test, Cat. No. 1.00594.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Al Absorption 10 mm

cell mg/l Al

0.336 0 0.245 0 0.397 0.05 0.290 0.05 0.466 0.10 0.340 0.10 0.569 0.15 0.415 0.15 0.671 0.20 0.490 0.20 0.781 0.25 0.570 0.25 0.897 0.30 0.655 0.30 1.032 0.35 0.753 0.35 1.164 0.40 0.850 0.40 1.288 0.45 0.940 0.45 1.411 0.50 1.030 0.50

Non-linear curve for Aluminium Cell Test, Cat. No. 1.00594.0001, calibration against distilled water.

Nichtlineare 16 mm round cell

10 mm cell

Kalibrierung 0.02 – 0.50 mg/l Al

0.02 – 0.50 mg/l Al

Koeffizient a: -0.308 -0.308 Koeffizient b: 1.157 1.585 Koeffizient c: -0.748 -1.403 Koeffizient d: 0.237 0.609

Page 11: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 11 of 17

Absorption-Concentration- Table for Aluminium Test, Cat. No. 1.14825.0001, calibration against distilled water. Absorption 10 mm

cell mg/l Al Absorption 50 mm

cell mg/l Al

0.200 0 1.050 0 0.250 0.05 1.100 0.020 0.290 0.10 1.180 0.040 0.360 0.15 1.300 0.060 0.430 0.20 1.420 0.080 0.590 0.30 1.550 0.100 0.750 0.40 1.670 0.120 1.165 0.60 1.800 0.140 1.560 0.80 1.930 0.160 1.935 1.00 2.060 0.180 2.285 1.20 2.200 0.200

Non-linear curve for Aluminium Test, Cat. No. 1.14825.0001, calibration against distilled water.

Non-linear 10 mm cell 50 mm cell calibration 0.10 – 1.20

mg/l Al 0.020 – 0.240

mg/l Al coefficient a: -0.167 -0.493 coefficient b: 0.981 0.769 coefficient c: -0.378 -0.348 coefficient d: 0.092 0.065

Page 12: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 12 of 17

Absorption-Concentration-Table for Calcium Cell Test, Cat. No. 1.00858.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Ca mg/l CaCO3 mg/l CaO

0.075 0 0 0 0.100 10 25 14 0.170 35 88 49 0.245 60 150 84 0.330 85 213 119 0.450 115 288 161 0.575 145 363 203 0.705 175 438 245 0.835 200 500 280 0.950 225 563 315 1.075 250 625 350

Absorption 10 mm

cell mg/l Ca mg/l CaCO3 mg/l CaO

0.056 0 0 0 0.074 10 25 14 0.126 35 88 49 0.181 60 150 84 0.244 85 213 119 0.333 115 288 161 0.426 145 363 203 0.522 175 438 245 0.619 200 500 280 0.704 225 563 315 0.796 250 625 350

Non-linear curve for Calcium Cell Test, Cat. No. 1.00858.0001, calibration against distilled water.

Non-linear 16 mm round cell

16 mm round cell

16 mm round cell

calibration 10 – 250 mg/l Ca

25 – 625 mg/l CaCO3

14 – 350 mg/l CaO

coefficient a: -29.293 -73,418 -41.011 coefficient b: 416.85 1045 583.59 coefficient c: -252.27 -634.44 -353.18 coefficient d: 99.158 249.09 138.82

Non-linear 10 mm cell 10 mm cell 10 mm cell calibration 10 – 250

mg/l Ca 25 – 625

mg/l CaCO3

14 – 350 mg/l CaO

coefficient a: -29.666 -74.353 -41.533 coefficient b: 567.59 1422.9 794.63 coefficient c: -473.01 -1189.5 -662.22 coefficient d: 253.87 637.66 355.42

Page 13: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 13 of 17

Absorption-Concentration-Table for Magnesium Cell Test, Cat. No. 1.00815.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Mg Absorption 10 mm

cell mg/l Mg

0.070 0 0.052 0 0.085 5.0 0.063 5.0 0.120 13.0 0.089 13.0 0.150 20.0 0.111 20.0 0.190 28.0 0.147 28.0 0.235 36.0 0.174 36.0 0.285 44.0 0.211 44.0 0.335 52.0 0.248 52.0 0.385 59.0 0.285 59.0 0.435 67.0 0.322 67.0 0.490 75.0 0.363 75.0

Non-linear curve for Magnesium Cell Test, Cat. No. 1.00815.0001, calibration against distilled water.

Non-linear 16 mm round cell

10 mm cell

calibration 5.0 – 75.0 mg/l Mg

5.0 – 75.0 mg/l Mg

coefficient a: -20.79 -19.51 coefficient b: 337.41 425.65 coefficient c: -513.33 -778.18 coefficient d: 458.21 892.38

Absorption-Concentration-Table for Nitrate Cell Test in sea water, Cat. No. 1.14556.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l NO3-N mg/l NO3

- Absorption 10 mmcell

mg/l NO3-N mg/l NO3-

0.000 0 0 0.000 0 0 0.085 0.10 0.4 0.063 0.10 0.4 0.260 0.40 1.8 0.193 0.40 1.8 0.480 0.70 3.1 0.356 0.70 3.1 0.725 1.00 4.4 0.537 1.00 4.4 1.000 1.30 5.8 0.741 1.30 5.8 1.275 1.60 7.1 0.944 1.60 7.1 1.600 1.90 8.4 1.185 1.90 8.4 1.900 2.20 9.7 1.407 2.20 9.7 2.200 2.50 11.1 1.630 2.50 11.1 2.500 2.80 12.4 1.852 2.80 12.4 2.700 3.00 13.3 2.000 3.00 13.3

Non-linear curve for Nitrate Cell Test in seawater, Cat. No. 1.14556.0001, calibration against distilled water.

Non-linear 16 mm round cell

16 mm round cell

10 mm cell

10 mm cell

calibration 0.10 – 3.00 mg/l NO3-N

0.4 – 13.3 mg/l NO3

-0.10 – 3.00 mg/l NO3-N

0.4 – 13.3 mg/l NO3

-

coefficient a: -0.01 -0.0442 -0.0102 -0.0454 coefficient b: 1.6224 7.182 2.1895 9.6924 coefficient c: -0.3722 -1.6478 -0.6766 -2.9951 coefficient d: 0.0687 0.3041 0.1683 0.7449

Page 14: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 14 of 17

Absorption-Concentration-Table for Nitrate Test in sea water, Cat. No. 1.14942.0001, calibration against distilled water. Absorption 10 mm

cell mg/l NO3-N mg/l NO3

-

0.000 0.0 0 0.050 1.0 4.43 0.160 2.0 8.85 0.300 3.0 13.28 0.580 5.0 22.13 0.860 7.0 30.99 1.140 9.0 39.84 1.410 11.0 48.69 1.700 13.0 57.55 1.980 15.0 66.40 2.350 17.0 75.26

Non-linear curve for Nitrate Test in sea water, Cat. No. 1.14942.0001, calibration against distilled water.

Non-linear 10 mm cell 10 mm cell calibration 0.2 – 17.0

mg/l NO3-N 0.9 – 75.3 mg/l NO3

-

coefficient a: 0.421 -1.865 coefficient b: 8.251 36.524 coefficient c: -0.5679 -2.515 coefficient d: 0.0335 0.149

Absorption-Concentration-Table for pH Cell Test, Cat. No. 1.01744.0001, calibration against dist. water. Absorption 16 mm

round cell pH Absorption 10 mm

cell pH

0.000 6.2 0.000 6.2 0.100 6.4 0.074 6.4 0.175 6.7 0.130 6.7 0.330 7.0 0.244 7.0 0.500 7.2 0.370 7.2 0.600 7.4 0.444 7.4 0.800 7.6 0.593 7.6 1.020 7.8 0.741 7.8 1.325 8.0 0.981 8.0 1.700 8.4 1.259 8.4 1.900 8.8 1.407 8.8

Non-linear curve for pH Cell Test, Cat. No. 1.01744.0001, calibration against dist. water.

Non-linear 16 mm round cell

10 mm cell

calibration pH 6.4 – 8.8 pH 6.4 – 8.8 coefficient a: 6.144 6.141 coefficient b: 3.218 4.359 coefficient c: -2.318 -4.231 coefficient d: 0.714 1.755

Page 15: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 15 of 17

Absorption-Concentration-Table for Residual Hardness Cell Test, Cat. No. 1.14683.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Ca mg/l CaCO3 mg/l CaO °d °e °f

0.100 0 0 0 0 0 0 0.220 0.50 1.25 0.70 0.070 0.088 0.125 0.350 1.00 2.50 1.40 0.140 0.175 0.250 0.470 1.50 3.75 2.10 0.210 0.263 0.375 0.600 2.00 5.00 2.80 0.280 0.350 0.500 0.750 2.50 6.25 3.50 0.350 0.438 0.625 0.900 3.00 7.50 4.20 0.420 0.525 0.750 1.200 4.00 10.00 5.60 0.560 0.700 1.000 1.500 5.00 12.50 7.00 0.700 0.875 1.250

Absorption 10 mm

cell mg/l Ca mg/l CaCO3 mg/l CaO °d °e °f

0.074 0 0 0 0 0 0 0.163 0.50 1.25 0.70 0.070 0.088 0.125 0.259 1.00 2.50 1.40 0.140 0.175 0.250 0.348 1.50 3.75 2.10 0.210 0.263 0.375 0.444 2.00 5.00 2.80 0.280 0.350 0.500 0.556 2.50 6.25 3.50 0.350 0.438 0.625 0.667 3.00 7.50 4.20 0.420 0.525 0.750 0.889 4.00 10.00 5.60 0.560 0.700 1.000 1.111 5.00 12.50 7.00 0.700 0.875 1.250

Non-linear curve for Residual Hardness Cell Test, Cat. No. 1.14683.0001, calibration against distilled water.

Non-linear 16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

calibration 0.50 – 5.00 mg/l Ca

1.25 – 12.50mg/l CaCO3

0.70 – 7.00 mg/l CaO

0.070 – 0.700°d

0.088 – 0.875 °e

0.125 – 1.250°f

coefficient a: -0.46153 -1.1538 -0.64615 -0.06461 -0.08077 -0.11538 coefficient b: 4.62293 11.557 6.4721 0.64721 0.80901 1.15573 coefficient c: -1.14084 -2.8521 -1.59717 -0.15972 -0.19965 -0.28521 coefficient d: 0.32358 0.8089 0.45301 0.0453 0.05663 0.08089

Non-linear 10 mm cell 10 mm cell 10 mm cell 10 mm cell 10 mm cell 10 mm cell calibration 0.50 – 5.00

mg/l Ca 1.25 – 12.50mg/l CaCO3

0.70 – 7.00 mg/l CaO

0.070 – 0.700°d

0.088 – 0.875 °e

0.125 – 1.250°f

coefficient a: -0.46099 -1.15246 -0.64538 -0.06454 -0.08067 -0.11525 coefficient b: 6.23776 15.5944 8.73286 0.87329 1.09161 1.55944 coefficient c: -2.0739 -5.18475 -2.90346 -2.9035 -0.36293 -0.51847 coefficient d: 0.79354 1.98386 1.11096 0.1111 0.13887 0.19839

Page 16: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 16 of 17

Absorption-Concentration-Table for Sulfide Test, Cat. No. 1.14779.0001, calibration against distilled water. Absorption 10 mm

cell mg/l S

0.000 0 0.080 0.10 0.220 0.25 0.350 0.40 0.490 0.55 0.600 0.70 0.770 0.90 0.890 1.05 1.080 1.35 1.200 1.50

Non-linear curve for Sulfide Test, Cat. No. 1.14779.0001, calibration against distilled water.

Non-linear 10 mm cell calibration 0.10 – 1.50

mg/l S coefficient a: 0.0052 coefficient b: 1.1079 coefficient c: -0.0055 coefficient d: 0.1062

Absorption-Concentration-Table for Tin Cell Test, Cat. No. 1.14622.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Sn Absorption 10 mm

cell mg/l Sn

0.000 0 0.000 0 0.040 0.10 0.030 0.10 0.230 0.40 0.170 0.40 0.500 0.70 0.370 0.70 0.780 1.00 0.578 1.00 0.980 1.20 0.726 1.20 1.190 1.40 0.881 1.40 1.370 1.60 1.015 1.60 1.650 1.90 1.222 1.90 1.900 2.20 1.407 2.20 2.150 2.50 1.593 2.50

Non-linear curve for Tin Cell Test, Cat. No. 1.14622.0001, calibration against distilled water.

Non-linear 16 mm round cell

10 mm cell

calibration 0.10 – 2.50 mg/l Sn

0.10 – 2.50 mg/l Sn

coefficient a: 0.03 0.03 coefficient b: 1.5544 2.0985 coefficient c: -0.5016 -0.9141 coefficient d: 0.147 0.3617

Page 17: Programming Data for Spectrophotometer and Spectroquant Test Kits Measured With a Reference Spectrophotometer

Date: October 24, 2006 17 of 17

Absorption-Concentration-Table for Total Hardness Cell Test, Cat. No. 1.00961.0001, calibration against distilled water. Absorption 16 mm

round cell mg/l Ca mg/l CaCO3 mg/l CaO °d °e °f

0.068 0 0 0 0 0 0 0.088 5 13 7 0.7 0.9 1.3 0.150 25 63 35 3.5 4.4 6.3 0.240 50 125 70 7.0 8.8 12.5 0.345 75 188 105 10.5 13.1 18.8 0.465 100 250 140 14.0 17.5 25.0 0.570 120 300 168 16.8 21.0 30.0 0.710 145 363 203 20.3 25.4 36.3 0.855 170 425 238 23.8 29.8 42.5 1.015 195 488 273 27.3 34.1 48.8 1.145 215 538 301 30.1 37.6 53.8

Absorption 10 mm

cell mg/l Ca mg/l CaCO3 mg/l CaO °d °e °f

0.050 0 0 0 0 0 0 0.065 5 13 7 0.7 0.9 1.3 0.111 25 63 35 3.5 4.4 6.3 0.178 50 125 70 7.0 8.8 12.5 0.256 75 188 105 10.5 13.1 18.8 0.344 100 250 140 14.0 17.5 25.0 0.422 120 300 168 16.8 21.0 30.0 0.526 145 363 203 20.3 25.4 36.3 0.633 170 425 238 23.8 29.8 42.5 0.752 195 488 273 27.3 34.1 48.8 0.848 215 538 301 30.1 37.6 53.8

Non-linear curve for Total Hardness Cell Test, Cat. No. 1.00961.0001, calibration against distilled water.

Non-linear 16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

16 mm round cell

calibration 5 – 215 mg/l Ca

13 – 538 mg/l CaCO3

7 – 301 mg/l CaO

0.7 – 30.1 °d

0.9 – 37.6 °e

1.3 – 53.8 °f

coefficient a: -22.354 -55.594 -31.296 -3.1296 -3.8748 -5.5594 coefficient b: 339.98 849.94 475.97 47.597 59.261 84.994 coefficient c: -195.26 -489.08 -273.36 -27.336 -33.641 -48.908 coefficient d: 69.671 175.17 97.54 9.754 11.864 17.517

Non-linear 10 mm cell 10 mm cell 10 mm cell 10 mm cell 10 mm cell 10 mm cell calibration 5 - 215

mg/l Ca 13 – 538

mg/l CaCO3

7 - 301 mg/l CaO

0.7 – 30.1 °d

0.9 – 37.6 °e

1.3 – 53.8 °f

coefficient a: -22.177 -55.151 -31.048 -3.1048 -3.844 -5.5151 coefficient b: 457.29 1143.2 640.2 64.02 79.71 114.32 coefficient c: -351.63 -880.8 -492.29 -49.229 -60.575 -88.08 coefficient d: 168.46 423.59 235.85 23.585 28.674 42.359

Please contact our local Merck companies or Merck agents for any questions. Alternatively contact: [email protected]