PROFESSIONAL HANDPIECE MAINTENANCE

2
STEP 4 STERILISATION Place each handpiece in a separate sterilisation pouch and seal. Sterilise handpieces up to a max of 135°C (refer to your Autoclave user guide). AUTOMATICALLY CLEAN AND LUBRICATE HANDPIECES USING NSK iCARE EVERY TIME AFTER USE AND BEFORE STERILISATION NSK Oceania Pty Ltd www.nskoceania.com.au Product Information | Technical Support Freephone: 1300 44 33 21 (Australia) 0800 44 33 21 (New Zealand) Email: [email protected] KaVo ® and MULTIflex LUX ® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona ® is a registered trademark of Sirona Dental Systems GmbH, Germany. W&H ® and Roto Quick ® are registered trademarks of W&H Dentalwerk Burmoos GmbH, Austria. Bien-Air ® and Unifix ® are registered trademarks of Bien-Air Dental S.A., Switzerland. STEP 2 EXTERNAL CLEANING Manual: Clean the external surface of the handpiece with either a neutral detergent or enzymatic solution. A soft brush may be used if external debris are still present. Rinse with running water (<38°C, mineralised water is recommended). Do not spray directly on to the handpiece, use a wipe or lint free cloth if required. Dry handpieces using a dry mat, lint free cloth or compressed air. Automated : NSK handpieces with this symbol are thermodisinfector ready. When using a thermo-disinfector ensure the correct handpiece adaptors are used and the rinse and dry cycles activated. Solutions must not exceed PH9.5. STEP 5 STORAGE After sterilisation leave each handpiece sealed in the unopened autoclave pouch. Store in an hygienic area ready for use. STEP 1 PREPARATION Remove the bur and disconnect the handpiece from the dental unit. A. Select the desired purge programme Mode. By selecting the selection button (1, 2 or 3), until the mode light is illuminated. S: Short / M: Middle / L: Long S Mode for High Speed handpieces (air turbines) (Cycle time 28 seconds) S Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 50 secs) M (Middle) Mode for High Speed handpieces (air turbines) (Cycle time 30 seconds) M (Middle) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 64 secs) L (Long) Mode for High Speed handpieces (air turbines) (Cycle time 33 seconds) L (Long) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 84 secs) B.Chuck Cleaning insert chuck cleaning nozzle into the joint. hold handpiece securely by hand, insert chuck cleaning nozzle into bur insertion hole of the handpiece. when cleaning is completed, lubricate the handpiece following the instructions in the NSK iCare instructions manual. C. Straight handpieces Insert the supplied test bur into the handpiece and lock the chuck by turning the bur lock ring. You then connect this to the e-type handpiece joint until it locks. D. Surgical handpieces To avoid any foreign material such as blood remaining in the handpiece it is recommended that the handpiece is lubricated immediately after use (e.g. with NSK Pana Spray plus). E. Press the ‘Start’ button. The door should not be opened until the cycle is finished. E STEP 3 AUTOMATIC CLEANING & LUBRICATION USING NSK iCARE E-Type Handpiece For All Brands ISO E-Type and E-Type LUX Low Speed handpiece attachments – Z280-035 E-Type NSK Coupling For High Speed handpieces that connect to NSK couplings – Z257010 NSK Borden 2 hole For Borden 2 hole High Speed handpieces – T904 Borden ® KaVo Coupling For High Speed handpieces that connect to KaVo ® Multiflex LUX® couplings – Z257020 Kavo ® Sirona Coupling For High Speed handpieces that connect to Sirona ® Quick couplings – Z257023 Sirona ® W&H Coupling For High Speed handpieces that connect to W&H ® Roto Quick ® couplings – Z257021 W&H® Bien Air Coupling For High Speed handpieces that connect to Bien-Air ® Unifix ® couplings – Z280022 Bien-Air ® F-Type Head For ENDO-MATE DT, TC / TASKAL Wizard Heads – Z280-042 F-Type Morita For High Speed handpieces that connect to Morita couplings – Z257-048 Morita iCare Adaptor Guide Use the correct iCare adaptor if your handpiece coupling is one of the following: PROFESSIONAL HANDPIECE MAINTENANCE A

Transcript of PROFESSIONAL HANDPIECE MAINTENANCE

Page 1: PROFESSIONAL HANDPIECE MAINTENANCE

STEP 4STERILISATIONPlace each handpiece in a separate sterilisation pouch and seal. Sterilise handpieces up to a max of 135°C (refer to your Autoclave user guide).

AUTOMATICALLY CLEAN AND LUBRICATE HANDPIECES USING NSK iCARE EVERY TIME AFTER USE AND BEFORE STERILISATION

NSK Oceania Pty Ltd www.nskoceania.com.auProduct Information | Technical Support

Freephone: 1300 44 33 21 (Australia) 0800 44 33 21 (New Zealand) Email: [email protected]

KaVo® and MULTIflex LUX® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany. W&H® and Roto Quick® are registered trademarks of W&H Dentalwerk Burmoos GmbH, Austria. Bien-Air® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.

STEP 2 EXTERNAL CLEANING

Manual: Clean the external surface of the handpiece with either a neutral detergent or enzymatic solution. A soft brush may be used if external debris are still present. Rinse with running water (<38°C, mineralised water is recommended). Do not spray directly on to the handpiece, use a wipe or lint free cloth if required. Dry handpieces using a dry mat, lint free cloth or compressed air.

Automated : NSK handpieces with this symbol are thermodisinfector ready. When using a thermo-disinfector ensure the correct handpiece adaptors are used and the rinse and dry cycles activated. Solutions must not exceed PH9.5.

STEP 5STORAGEAfter sterilisation leave each handpiece sealed in the unopened autoclave pouch. Store in an hygienic area ready for use.

STEP 1 PREPARATIONRemove the bur and disconnect the handpiece from the dental unit.

A. Select the desired purge programme Mode.

By selecting the selection button (1, 2 or 3), until the mode light is illuminated.S: Short / M: Middle / L: Long

• S Mode for High Speed handpieces (air turbines) (Cycle time 28 seconds)

• S Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 50 secs)

• M (Middle) Mode for High Speed handpieces (air turbines) (Cycle time 30 seconds)

• M (Middle) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 64 secs)

• L (Long) Mode for High Speed handpieces (air turbines) (Cycle time 33 seconds)

• L (Long) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 84 secs)

B.Chuck Cleaning

– insert chuck cleaning nozzle into the joint.

– hold handpiece securely by hand, insert chuck cleaning nozzle into bur insertion hole of the handpiece.

– when cleaning is completed, lubricate the handpiece following the instructions in the NSK iCare instructions manual.

C. Straight handpieces

Insert the supplied test bur into the handpiece and lock the chuck by turning the bur lock ring. You then connect this to the e-type handpiece joint until it locks.

D. Surgical handpieces

To avoid any foreign material such as blood remaining in the handpiece it is recommended that the handpiece is lubricated immediately after use (e.g. with NSK Pana Spray plus).

E. Press the ‘Start’ button.

The door should not be opened until the cycle is finished.

E

STEP 3 AUTOMATIC CLEANING & LUBRICATION USING NSK iCARE

E-Type HandpieceFor All Brands ISO E-Type and E-Type LUX Low Speed handpiece attachments – Z280-035 E-Type

NSK CouplingFor High Speed handpieces that connect to NSK couplings – Z257010 NSK

Borden 2 holeFor Borden 2 hole High Speed handpieces – T904 Borden®

KaVo CouplingFor High Speed handpieces that connect to KaVo® Multiflex LUX® couplings – Z257020 Kavo®

Sirona CouplingFor High Speed handpieces that connect to Sirona® Quick couplings – Z257023 Sirona®

W&H CouplingFor High Speed handpieces that connect to W&H® Roto Quick® couplings – Z257021 W&H®

Bien Air CouplingFor High Speed handpieces that connect to Bien-Air® Unifix® couplings – Z280022 Bien-Air®

F-Type HeadFor ENDO-MATE DT, TC / TASKAL Wizard Heads – Z280-042 F-Type

MoritaFor High Speed handpieces that connect to Morita couplings – Z257-048 Morita

iCare Adaptor GuideUse the correct iCare adaptor if your handpiece coupling is one of the following:

PROFESSIONAL HANDPIECE MAINTENANCE

A

Page 2: PROFESSIONAL HANDPIECE MAINTENANCE

STEP 1 PREPARATIONRemove the bur and disconnect the handpiece from the dental unit.

Low Speeds and Surgical Optic and Non Optic handpieces and heads:

E Type (all brands) Nozzle Z019-090

ISO E Type (all brands) and NSK Surgical Heads SGM-I, SGMS-I & SGQ-H Nozzle Z114-090

NSK HEADS used with Endo-Mate DT, TC, TC2 and TASKAL Wizard Nozzle Z256-090

NSK Phatelus Airmotor Heads Nozzle M070-090 Nozzle P183-090

Pana Spray Plus Nozzle GuideHigh Speed Optic and Non Optic handpieces that connect direct to:

NSK coupling Nozzle P183-090

KaVo® coupling Nozzle Z090-051

W&H® coupling Nozzle Z090-054

Sirona® coupling Nozzle Z090-052

Bien-Air® coupling Nozzle Z090-053

NSK Pana Spray Plus ORDER CODE: Z182100G

Placing the handpieces upright on the NSK multi-stand is an easy way to remove any excess oil from the handpieces after lubrication. ORDER CODE: Z036

STEP 3SPRAY CAN CLEANING & LUBRICATION WITH NSK PANA SPRAY PLUSNSK Pana Spray Plus is a specially formulated, highly effective, spray cleaner/lubricant formulated by NSK for the correct internal maintenance of all dental handpieces, airmotors and air scalers. Regular application of NSK Pana Spray Plus will extend long term product performance, reliability and durability.

It is preferable to purge any excess solution by briefly operating the handpiece. When purging cover the handpiece head with a tissue. Screw the nozzle onto the valve at the top of the NSK Pana Spray Plus can. Position the nozzle into the handpiece or head as illustrated.Apply NSK Pana Spray Plus as follows:– High Speed handpieces 1 to 2 seconds.– Low Speed handpieces and heads 2 seconds.– Always use the correct nozzle for the handpiece being lubricated.

NSK Oceania Pty Ltd www.nskoceania.com.auProduct Information | Technical Support

Freephone: 1300 44 33 21 (Australia) 0800 44 33 21 (New Zealand) Email: [email protected]

KaVo® and MULTIflex LUX® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany. W&H® and Roto Quick® are registered trademarks of W&H Dentalwerk Burmoos GmbH, Austria. Bien-Air® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.

MANUALLY CLEAN AND LUBRICATE HANDPIECES USING NSK PANA SPRAY PLUS EVERY TIME AFTER USE AND BEFORE STERILISATION

PROFESSIONAL HANDPIECE MAINTENANCE

STEP 4STERILISATIONPlace each handpiece in a separate sterilisation pouch and seal. Sterilise handpieces up to a max of 135°C (refer to your Autoclave user guide).

STEP 5STORAGEAfter sterilisation leave each handpiece sealed in the unopened autoclave pouch. Store in an hygienic area ready for use.

STEP 2 EXTERNAL CLEANING

Manual: Clean the external surface of the handpiece with either a neutral detergent or enzymatic solution. A soft brush may be used if external debris are still present. Rinse with running water (<38°C, mineralised water is recommended). Do not spray directly on to the handpiece, use a wipe or lint free cloth if required. Dry handpieces using a dry mat, lint free cloth or compressed air.

Automated : NSK handpieces with this symbol are thermodisinfector ready. When using a thermo-disinfector ensure the correct handpiece adaptors are used and the rinse and dry cycles activated. Solutions must not exceed PH9.5.