[Prof] Visualimpact 0710.680imn1 - 7.1
-
Upload
frank-scialla -
Category
Documents
-
view
47 -
download
0
description
Transcript of [Prof] Visualimpact 0710.680imn1 - 7.1
-
Visual IMPACT
Code 0710.680.1 Code 0710.680.2 Code 0710.680.3 Code 0710.680.4
INSTRUCTION MANUAL 0710.680.MN1r Ed. 7 Rev. 1
The subject of this document is the exclusive property of ITW Test and Measurement Italia S.r.l. - INSTRON CEAST Divi-sion. This document or any part thereof must not be reproduced in any form, nor information contained therein dis-closed to third parties nor methods, procedures or test described therein performed, without the written permission of ITW Test and Measurement Italia S.r.l. - INSTRON CEAST Division.
CEAST Data Acquisition System Series
-
SUMMARY
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 3
SUMMARY
1 GENERAL INFORMATION ........................................ 11
1.1 CEASTVIEW SOFTWARE 12
1.2 VISUALIMPACT SOFTWARE 13
1.2.1 MAIN MODULE FUNCTIONS 13
1.2.2 OPTIONAL MODULES 13
1.3 WINDOWS VIEW 15
1.3.1 MENU WINDOW 15
1.3.2 WINDOW SIZING 15
2 INSTALLATION AND CONFIGURATION ......................... 17
2.1 MINIMUM HARDWARE AND SOFTWARE REQUIREMENTS 18
2.1.1 FOR UNINSTRUMENTED VISUALIMPACT 18
2.1.2 FOR INSTRUMENTED VISUALIMPACT 18
2.2 SOFTWARE INSTALLATION 20
2.3 UNINSTALL SOFTWARE 21
2.4 RUN CEASTVIEW 22
2.5 EXIT SOFTWARE 23
2.6 SYSTEM CONFIGURATION 24
2.6.1 COMPANY IDENTIFICATION FOLDER 24
2.6.2 INSTRUMENT FOLDER 28
2.6.3 INSTRUMENT CONFIGURATION 29
2.6.4 UNITS OF MEASURE FOLDER 51
2.6.5 MISCELLANEOUS FOLDER 53
2.6.6 LANGUAGE FOLDER 55
2.7 TUP/HAMMER/VICE CONFIGURATION 56
2.7.1 UNINSTRUMENTED TUP/HAMMER 57
2.7.2 INSTRUMENTED TUP/HAMMER/VICE 57
2.8 IMPORT FUNCTION 61
-
SUMMARY
4 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
3 TEST PARAMETERS .............................................. 63
3.1 TEMPLATES MANAGEMENT 64
3.2 OPERATIONS ON TEMPLATES 68
3.2.1 SET TEMPLATE AS DEFAULT 68
3.2.2 TOGGLE A TEMPLATE AS REFERENCE 68
3.2.3 SELECT NOT USED TEMPLATE SETS 69
3.2.4 DELETE TEMPLATE SETS 69
3.2.5 COPY TEMPLATE SETS 70
3.2.6 RENAME TEMPLATE SETS 70
3.2.7 VIEW OR MODIFY AN EXISTING TEMPLATE SET 70
3.2.8 ADD NEW TEMPLATE SETS 71
3.3 EDIT TEMPLATE SETS 72
3.3.1 GENERAL FOLDER 72
3.3.2 STANDARD FOLDER 74
3.3.3 SPECIMEN FOLDER 82
3.3.4 DATA ACQUISITION SYSTEM FOLDER 85
3.3.5 MATERIAL FOLDER 94
3.3.6 TEMPERATURE FOLDER 95
3.3.7 OPTIONS FOLDER 97
3.3.8 POINTS FOLDER 102
3.4 IMPACT APPEARANCE LIST MANAGEMENT 110
4 TEST METHOD .................................................. 115
4.1 LOADERS MANAGER WINDOW 116
4.2 SPECIMENS LOT 117
4.3 LOADER 120
4.3.1 CREATE A NEW LOADER 120
4.3.2 LOADER EDITING 121
4.3.3 LOADER WITH LOTS 122
4.4 BOX 124
4.4.1 CREATE A BOX 124
4.4.2 BOX EDITING 125
4.4.3 BOX WITH LOADERS AND LOTS 126
4.5 OBJECTS MANAGEMENT 127
5 TEST EXECUTION ............................................... 129
5.1 TEST EXECUTION 130
5.1.1 RUN IMPACT TEST 130
5.2 PENDULUM TEST PROCEDURE 134
5.2.1 TEST START 134
5.2.2 SPECIMEN DIMENSION CUSTOMIZATION 135
5.2.3 IMPACT EXECUTION 135
5.2.4 TEST END 135
-
SUMMARY
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 5
5.2.5 TEST SUSPENSION 136
5.2.6 TEST BREAK 136
5.3 CEAST 9340, CEAST 9350 AND FRACTOVIS PLUS TEST PROCEDURE 137
5.3.1 TEST START 137
5.3.2 SPECIMEN DIMENSION CUSTOMIZATION 137
5.3.3 IMPACT EXECUTION 138
5.3.4 TEST END 138
5.3.5 TEST SUSPENSION 139
5.3.6 TEST BREAK 139
5.4 RESILVIS TEST PROCEDURE 140
5.4.1 TEST START 140
5.4.2 SPECIMEN DIMENSION CUSTOMIZATION 140
5.4.3 IMPACT EXECUTION 141
5.4.4 TEST END 141
5.4.5 TEST SUSPENSION 141
5.4.6 TEST BREAK 142
5.4.7 COMPONENT DISABLING 142
5.5 REAL TIME RESULTS TABLE 144
5.5.1 RESULTS ADJUSTMENT 145
5.6 SYNOPTIC WINDOW 146
5.6.1 RESILVIS WINDOW 146
5.6.2 PENDULUM WINDOW 148
5.6.3 CEAST 9340 WINDOW 151
5.6.4 CEAST 9350 AND FRACTOVIS PLUS WINDOW 156
5.6.5 CEAST 9050/IMPACTOR II WINDOW 162
5.6.6 DAS WINDOW 167
5.6.7 MEASURING DEVICE WINDOW 168
5.6.8 DOUBLE MEASURING DEVICE WINDOW 170
5.6.9 WIN9 WINDOW 171
5.6.10 RESILVIS MOVEMENT WINDOW 174
5.6.11 THERMOSTATIC CELL SYNOPTIC WINDOW 179
5.7 SYNOPTIC WINDOW SIZING 181
5.8 REAL TIME GRAPHIC 182
5.8.1 UNINSTRUMENTED REAL TIME GRAPH 182
5.8.2 INSTRUMENTED REAL TIME GRAPH 182
5.9 TEST ANALYSIS DURING EXECUTION 183
6 TEST OPENING AND DATA ANALYSIS ....................... 185
6.1 OPEN TESTS FROM CURRENT DATABASE 186
6.1.1 OPEN LAST EXECUTED TEST 186
6.1.2 OPEN EXECUTED TESTS 186
6.1.3 TEST FILE OPTIONS 188
6.2 DATA ANALYSIS WINDOW 193
6.2.1 UNINSTRUMENTED ANALYSIS WINDOW 193
6.3 INSTRUMENTED ANALYSIS WINDOW 200
-
SUMMARY
6 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
6.3.1 GRAPH PANEL 200
6.3.2 ANALYSIS AREA 201
6.3.3 RESULTS AREA 209
6.4 RESULTS MANAGEMENT 222
6.4.1 RESULTS FILTERING ON BREAK TYPE 222
6.4.2 RESULTS DISCARD 222
6.4.3 ENABLE OR DISABLE MAIN POINTS 223
6.5 CHANGE SETTINGS DIALOG 224
6.5.1 CHANGE THE BREAK TYPE 225
6.5.2 CHANGE THE APPEARANCE BREAK IMAGE 225
6.5.3 CHANGE THE MAIN POINTS SETTINGS 226
6.5.4 SMOOTH TEST CURVE DATA 228
6.5.5 RESULTS TABLE AND MAIN POINTS CUSTOMIZATION 230
6.5.6 LOAD AND SAVE CUSTOM SETTINGS 232
6.5.7 PERCENTAGE POINT CUSTOMIZATION 233
6.5.8 MANUAL POINT CUSTOMIZATION 234
6.5.9 CHANGE SPECIMEN DIMENSIONS 234
6.5.10 BRITTLE DUCTILE LIMIT CUSTOMIZATION 235
6.5.11 VIEW PARAMETERS 235
6.5.12 CHANGE COMMENT 235
6.5.13 AUTOSTART 236
6.6 YIELD POINT WIZARD 237
7 GRAPH MANAGEMENT ......................................... 241
7.1 AXES CUSTOMIZATION 242
7.1.1 UNINSTRUMENTED TEST GRAPH 242
7.1.2 INSTRUMENTED TEST GRAPH 243
7.2 GRAPHIC FUNCTIONS 246
7.2.1 SET MAIN POINTS ON CURVE 246
7.2.2 SHIFTING CURVE 246
7.2.3 OFFSETTING CURVE 246
7.2.4 GRAPHIC MENU 248
7.2.5 SHOW/HIDE GRIDS 249
7.2.6 FULL SCREEN GRAPH VIEW 249
7.2.7 ZOOM MODE 249
7.2.8 GRAPH DATA VIEW 251
7.2.9 SAVE GRAPHIC 251
8 DATA PRINTOUT ................................................ 253
8.1 PRINTOUT CUSTOMIZING 254
8.1.1 PRINT CONFIGURATION 254
8.1.2 REPORT SELECTION 258
8.1.3 REPORT LIST CONFIGURATION 258
8.2 REPORT WIZARD 261
-
SUMMARY
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 7
8.2.1 REPORT OVERVIEW 261
8.2.2 REPORT WIZARD EXECUTION 262
8.3 PRINTING REPORT 270
8.3.1 MANUAL PRINT 270
8.3.2 AUTOMATIC PRINT 271
9 DATA EXPORT .................................................. 273
9.1 EXPORT CONFIGURATION 274
9.1.1 EXPORT FORM SETTINGS 275
9.1.2 GENERAL SETTINGS 276
9.1.3 PARAMETER AND DATA SETTINGS 277
9.1.4 SAVE FORM SETTINGS 279
9.2 MANUAL EXPORT 280
9.2.1 EXPORTING TEST 280
9.2.2 EXPORTING SPECIMEN 280
9.2.3 CLIPBOARD EXPORT 281
9.3 AUTOMATIC EXPORT 283
10 OPERATING PROFILES ......................................... 285
10.1 LOG ON FUNCTION 286
10.2 LOG OFF FUNCTION 287
10.3 EDIT FUNCTION 288
10.3.1 ADD PROFILES 288
10.3.2 EDIT PROFILES 288
10.3.3 DELETE PROFILES 289
11 TTA ANALYSIS ................................................. 291
11.1 THE TRANSITION TEMPERATURE ANALYSIS MODULE 292
11.1.1 TTA MODULE ACTIVATION 292
11.1.2 OPERATION ON VIRTUAL TESTS 293
11.1.3 ADD A VIRTUAL TEST 294
11.1.4 ABOUT VIRTUAL TEST SPECIMEN PARAMETERS 296
11.1.5 OTHER OPERATIONS ON VIRTUAL TEST SPECIMENS 296
11.1.6 PREVIEW GRAPHIC MADE BY SELECTED SPECIMENS 297
11.1.7 MODIFY A VIRTUAL TEST 297
11.2 DATA ANALYSIS WINDOW 298
11.2.1 TEMPERATURE VS AVERAGE RESILIENCE ANALYSIS 299
11.2.2 TEMPERATURE VS %B ANALYSIS 299
11.2.3 DATA ANALYSIS OPTIONS 300
11.2.4 HOMOGENEOUS TEMPERATURE SET COUNTING OUT 301
11.2.5 SINGLE SPECIMEN COUNTING OUT 301
11.2.6 DIT VARIATION 302
-
SUMMARY
8 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
11.3 REPORTING TTA 303
12 SQC IMPACT ANALYSIS ....................................... 305
12.1 THE SQC IMPACT ANALYSIS MODULE 306
12.1.1 SQC MODULE ACTIVATION 306
12.1.2 OPERATIONS ON VIRTUAL TESTS 307
12.1.3 ADD A VIRTUAL TEST 307
12.1.4 OTHER OPERATIONS ON VIRTUAL TEST SPECIMENS 309
12.1.5 PREVIEW SQC GRAPH 310
12.1.6 MODIFY A VIRTUAL TEST 310
12.2 DATA ANALYSIS WINDOW 311
-
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 9
MANUAL EDITION: #7#
MANUAL REVISION: #1#
DATE PRINTED (ISO 8601): 2010-01-29
OBJECT OF THE REVISION:
CEAST 9050 added.
VALIDITY OF THE MANUAL STARTING FROM LOT NO.
-
1 GENERAL INFORMATION
CHAPTER 1
GENERAL
INFORMATION
-
1 - GENERAL INFORMATION
12 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
1.1 CEASTVIEW SOFTWARE
The CeastVIEW software platform is based on a Mi-
crosoft Windows environment and is fully compatible
with the following operating system:
Windows 2000/SP2
Windows XP
Windows VISTA
Much of the software developed for the management
and use of CEAST measure instruments can operate on
this platform.
CeastVIEW software offers the following features:
Remote data processing. Allows the user to process
test results without being connected to the instrument.
Multi instrument management. Ability to perform
tests on many instruments simultaneously from the
same PC.
Customizable test reports with graphical data re-
presentation.
Advanced features:
- Storage of all acquired raw data
- Point-by-point management of graphs with sliding
cursor and advanced graph management with Zoom,
Pan and shift functions
- Units of measure customizing
- Customizable data export
Automatic adjustment to operator screen resolution.
Friendly MS Windows-based interface (e.g. easy
resizing of windows and columns, multi-document
opening, test graphic preview before opening, ).
MS Access database file. Can easily share data with
other MS compatible software.
-
1 - GENERAL INFORMATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 13
1.2 VISUALIMPACT SOFTWARE
The new VisualIMPACT software has been designed to
setup, manage, store and process test data executed
with CEAST pendulums and drop tower instruments.
VisualIMPACT software consists of a main module and
few optional modules.
1.2.1 MAIN MODULE FUNCTIONS
Parameters setup. Specific folders guide the opera-
tor through tup and specimen choices, filtered accord-
ing to selected test type and international standard
system.
Test execution. Acquired data, main settings and test
parameters can be displayed during test.
Data analysis and results management. Statistical
data from impacted specimens (min, max, average,
standard deviation and variance) are automatically cal-
culated. Data can be filtered by break type (complete
break, partial break, ..). Graphical representation of
absorbed energy and resilience during impact. Manual
discard of specimens to analyze statistical variation.
Graph management.
Print out of parameters and results
Data export in .TXT or Excel compatible format. Cus-
tomizable fields selection.
Other features: Possibility to connect up to two mi-
crometers for direct acquisition of specimen dimen-
sions. Full management of Data Acquisition System
(CEAST DAS 8000 or DAS 16000) for instrumented
tests, with drivers for USB connection to PC.
1.2.2 OPTIONAL MODULES
Operator: user can choose parameter set, perform
tests and open executed tests (See DATA EXPORT chap-
ter).
Transfer: (See GRAPH MANAGEMENT chapter).
Multi n (n = number of instruments connectable to the
same PC). Allows parallel management of many CEAST
instruments with same characteristics, from parame-
ters selection, to test execution, real-time synoptic and
data acquisition.
-
1 - GENERAL INFORMATION
14 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
MultiMIX: like "Multi n" module, but manages differ-
ent models (e.g.: Modular Melt Flow + Rheologic 1000
+ HDT/Vicat +)
TTA (Transition Temperature Analysis): (See TTA ANAL-
YSIS chapter).
SQC (Statistical Quality Control) Impact Analysis:
(See SQC IMPACT ANALYSIS chapter).
-
1 - GENERAL INFORMATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 15
1.3 WINDOWS VIEW
Using Menu window and windows sizing commands
are common operations in CeastVIEW software
1.3.1 MENU WINDOW
Under the Window menu the following options allow
the user to manage opened windows:
Get Focus: click on a window name to make that win-
dows current. Note that the current windows has a tick
on the left of its name.
Cascade: arranges windows on-screen in a cascade.
Tile: squeezes windows to show them all.
1.3.2 WINDOW SIZING
Visual IMPACT software automatically sizes window
dimensions to screen resolution. Suggested minimum
screen resolution is 1024 x 768 pixels - 65,536 colors
(16 bit).
To customize window and column size:
Position the cursor over the windows edge or column
header until the arrow shape becomes as shown ( ).
With the left mouse button drag the windows edge or
the column header to the desired dimension.
Release the left mouse button when done.
-
1 - GENERAL INFORMATION
16 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: To go back to the default view, press the Default
Dimensions icon from the tool bar.
-
2 INSTALLATION AND CONFIGURATION
CHAPTER 2
INSTALLATION
AND
CONFIGURATION
-
2 - INSTALLATION AND CONFIGURATION
18 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.1 MINIMUM HARDWARE AND SOFTWARE REQUIREMENTS
Following two paragraphs show you minimum hard-
ware and software requirements in the different fore-
seen measure instruments configurations.
2.1.1 FOR UNINSTRUMENTED VISUALIMPACT
Operative system: Windows 2000, XP, VISTA (Windows
XP recommended)
RAM memory:
- With Windows 2000: at least 64 MB, 128 MB recom-
mended.
- With Windows XP: at least 128 MB, 256 MB recom-
mended.
- With Windows VISTA: at least 512 MB, 1024 MB rec-
ommended.
Hard disk size: at least 200 MB; 1 GB recommended.
Video resolution: at least 800 X 600; 1024 x 768 rec-
ommended
Serial ports or USB1: one for each connected instru-
ment.
USB ports: one for the dongle.
Ethernet port: one TCP/IP 10/100 MiBit for remote as-
sistance if requested.
2.1.2 FOR INSTRUMENTED VISUALIMPACT
Operative system: Windows 2000, XP, VISTA (Windows
XP recommended)
RAM memory:
- With Windows 2000: at least 64 MB, 128 MB recom-
mended.
- With Windows XP: at least 128 MB, 256 MB recom-
mended.
- With Windows VISTA: at least 512 MB, 1024 MB rec-
ommended.
Hard disk size: at least 1 GB.
Video resolution: at least 1024 x 768 px
1 Please use only USB/serial converter Cod. 0700.165 for USB connections.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 19
Serial ports or USB2: one for each connected instru-
ment.
USB ports: one for the dongle and one for the DAS de-
vice.
Ethernet port: one TCP/IP 10base100 MiBit for remote
assistance if requested.
2 Please use only USB/serial converter Cod. 0700.165 for USB connections.
-
2 - INSTALLATION AND CONFIGURATION
20 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.2 SOFTWARE INSTALLATION
Before to install the software, read carefully the instal-
lation instructions given in the CeastVIEW Software
Installation CDROM.
If on your PC the CDROM drive is associated to the D:\
letter, then you can find the installation help files in:
D:\Readme - Quick Installation.txt
file, and in the:
D:\CeastVIEW Manual\Software Installation Pro-
cedure manual 0710.000MN1xx.pdf
file.
If you cannot read the PDF (Portable Document File)
file format, please install before the free Acrobat Read-
er program you can find in:
D:\CeastVIEW Manual\Adobe Reader 6.0 Install\
folder.
The program can be installed in a selected folder or in
the default path
C:\CeastVIEW x.xx yy3
in which x.xx yy is the version of the CeastVIEW soft-
ware installed.
A shortcut icon will be automatically created on
the Desktop at the end of installation.
NOTE: The system must be rebooted after software in-stallation.
3 The name %PROGRAMFILES% specifies the location of the Program Files folder on your
system. The %PROGRAMFILES% name varies depending on the language of your operative system.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 21
2.3 UNINSTALL SOFTWARE
NOTE: Before removing the CeastVIEW software, select File Configure Miscellaneous then press the
button to execute a copy in a safe place of the data stored by the Ceast-VIEW software.
NOTE: If you do not execute a copy of the CeastVIEW software database all configuration and test data will be lost.
To uninstall the CeastVIEW software:
From the Windows Control Panel double-click on the
Add/Remove programs icon.
Select the CeastVIEW program and click on the
Change/Remove button.
Follow the wizard uninstall procedure.
NOTE: When CWSRVCOM.exe file removing is requested, click on the Yes button.
Reboot the PC before installing a new software version.
NOTE: The CeastVIEW software directory must be de-leted manually after it has been uninstalled.
-
2 - INSTALLATION AND CONFIGURATION
22 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.4 RUN CEASTVIEW
To run the CeastVIEW software double-click on the
shortcut icon from the desktop.
If the Operator Module code 0710.651 is enabled, Log
On will be requested.
The default log on is:
User Name: Ceast
Password: Ceast
To customize the User Name and Password, see OPER-
ATING PROFILES chapter.
When dongle is not recognized a pop-up window with a
warning message will appear and the software will run
in DEMO mode.
NOTE: To solve problems with dongle recognition, see troubleshooting section in the Software Installation Procedure manual.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 23
2.5 EXIT SOFTWARE
To exit the software:
Click on File Exit from the menu bar.
The user will be prompted to save the open test not yet
saved, if any.
If the database has been modified, the system asks the
user to make a backup copy.
The backup files will be saved according to the Active
Backup Directory setting.
User can set the Active Backup Directory option us-
ing the setting:
File Configuration Miscellaneous Database
Configuration Active Backup Directory button.
NOTE: It is suggested to close all CeastVIEW opened windows before quitting the CeastVIEW program.
-
2 - INSTALLATION AND CONFIGURATION
24 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6 SYSTEM CONFIGURATION
To open the System Configuration window:
Click on File Configure from the menu bar,
or
Click on the System Configuration icon from the
tool bar.
A pop-up window for system configuration will appear
with the following folders:
Company Identification
Instruments
Units of Measure
Miscellaneous
Language
NOTE: To enable the software functions it is necessary to configure at least one instrument type in the System Configuration window.
2.6.1 COMPANY IDENTIFICATION FOLDER
In this folder you can specify your company identifica-
tion data and send system information to the CEAST
Technical Assistance in order to get help.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 25
2.6.1.1 COMPANY IDENTIFICATION
Fill in the Name and Address fields with the company
information you wish to print on your Test Reports.
To customize the Company Logo field:
Click on the Edit button.
Open your Company Logo file with Windows Paint
program.
Save the logo as LogoCust.bmp in the
C:\..\CeastVIEW\Data directory.
Close the Windows Paint program.
NOTE: Alternatively to the previous procedure, save a copy of your Company Logo LogoCust.bmp in the C:\\CeastVIEW\Data directory using Windows Ex-plorer.
Click on the Refresh button to view the updated im-
age.
NOTE: Logo suggested dimensions are 80x80 pixels.
2.6.1.2 TECHNICAL HELP
The Send e-mail button creates an e-mail message for
CEAST Technical Assistance, in order to notify them of
your software configuration and any technical prob-
lems in software installation and use.
Before sending notice of software configuration, re-
member to attach the CeastVIEWConfig.txt file, from
the CeastVIEW folder.
The View Config button opens the software configura-
tion file CeastVIEWConfig.txt with:
-
2 - INSTALLATION AND CONFIGURATION
26 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
The Company information strings pulled into the Name
and Address fields from the Company Identification
folder.
Software ID Key code.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 27
CeastVIEW information about installation and use.
Main settings configured from the Miscellaneous
folder.
Options:
0: option disabled
1: option enabled
Instrument configuration list: instruments configured
by the operator from the Instruments folder.
Main information is shown for each instrument.
Software modules configuration list: software
modules enabled from the hardware dongle.
Number and type of instrument enabled from the
hardware dongle.
System information: operating system information.
NOTE: CeastVIEWConfig.TXT file has to be sent to CEAST for software upgrade or for enabling new mod-ules.
Installation Date and ID Key Code fields are auto-
matically set by the program and cannot be altered.
-
2 - INSTALLATION AND CONFIGURATION
28 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.2 INSTRUMENT FOLDER
2.6.2.1 ADD INSTRUMENT
To choose the CEAST instruments to be configured in
your CeastVIEW software:
Click on the Add button
A pop-up window allows the user to choose from a list
of CEAST instruments supported by the CeastVIEW
software.
The windows shows a graphical area on the left and a
list of selectable instruments on the right in which:
Line is the name of the CEAST instrument family.
Instrument code is the set of instrument part num-
bers that can be configured by selecting that row in the
list. When the list of codes available for that row is too
long, an en dash is shown instead.
Description is a brief description of the instrument
type.
Type is an identifier by which the instrument is named
by software.
NOTE: The image shown in the graphical area of the window always shows the most representative instrument among the available Instrument Codes currently selected.
Select one instrument at a time and click on the OK
button.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 29
NOTE: For a description of the CEAST impact instru-ments see following paragraphs.
2.6.2.2 MODIFY INSTRUMENT CONFIGURATION
To modify the instrument configuration after it has
been added, choose the instrument from the list and
click on the Modify button.
The instrument configuration window will appear.
NOTE: For details on CEAST instrument configurations see specific instrument paragraphs.
2.6.2.3 REMOVE INSTRUMENT
To remove one instrument from the list of those confi-
gured select it and click on the Remove button.
2.6.3 INSTRUMENT CONFIGURATION
To add CEAST Impact Instruments to the configura-
tion window:
From the Instruments folder, click on the Add button.
Select from the list an instrument belonging to the
Visual Impact line. The impact instrument type avail-
able are:
Not instrumented pendulum. (Pendu-
lum + Measuring device)
Instrumented Pendulum (Pendulum +
Measuring device + DAS)
Not Instrumented CEAST
9050/Impactor II (CEAST 9050/Impactor II + Measur-
ing device + Double Measuring device)
Instrumented CEAST 9050/Impactor II
(CEAST 9050/Impactor II + Measuring device +
Double Measuring device + DAS)
Digital Micrometer Instrument
Double Digital Micrometer Instru-
ment
Ceast 9350 (Drop Tower + Win9 Weight
Device + Measuring device + DAS)
-
2 - INSTALLATION AND CONFIGURATION
30 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
Fractovis Plus (Drop Tower + Win9
Weight Device + Measuring device + DAS)
Ceast 9340 (Ceast 9340+ Measuring
device + DAS)
Resilvis system (Pendulum + Move-
ment System + DAS + W Measuring device + H Mea-
suring device + Thermal Cell Type 1 + Vice Thermo-
meter)
Generic Drop Tower (DAS + Measuring
device)
Generic Pendulum (DAS + Measuring
device)
Click on the OK button to confirm your choice.
The instrument configuration window appears, allow-
ing the user to:
Insert a customized System Name available for the
printout on the test report.
Select/Unselect System components.
Specify the System Version value as reported on the
Certificate of Quality and Origin shipped with the
device.
Customize the System Polling Time to set the refresh
time rate of complete system state (default value is
suggested).
Select Resilvis Type if a Resilvis instrument was se-
lected.
NOTE: To select/unselect components, click on the checkbox near the components name. When the box is checked the component is configured.
When an asterisk appears to the left of the component
checkbox, the component can be disabled but not unse-
lected.
To configure component parameters, double-click on
the component (e.g. Pendulum) to open the configura-
tion window.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 31
2.6.3.1 PENDULUM CONFIGURATION
To configure the Pendulum parameters, double-click on
the Pendulum label.
Set the following parameters according to the custom-
ers Pendulum characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the pen-
dulum has been connected.
NOTE: A warning message appears if two or more in-struments are configured to the same port, allow-ing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and pendulum (default value is sug-
gested).
Name: insert the customized name for the pendulum.
The Enable checkbox allows the user to enable/disable
the use of the Pendulum. Check to enable use.
NOTE: Pendulum enabling can also be set in Test Execu-tion.. The enabling state set in configuration is used as a default opening Test Execution window.
Move cursor over the pendulum image and select the
instrument in use from a list of Visual Impact-
compatible CEAST pendulums.
The Pendulum code is available from the instruments
CE label.
-
2 - INSTALLATION AND CONFIGURATION
32 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: For the Resilvis pendulum, only the Resil Impac-tor 25J code 6956.000 is available.
According to the pendulum type, the software shows
related characteristics:
Manual brake.
Automatic brake: pneumatic brake on the hammer
Automatic Hammer Recovery: motorized system to
recover the hammer in the rest position
Instrumented IZOD Vice: Izod vice equipped with
load cell for clamping force measurement
Manual speed variator: accessory for releasing angle
changing
NOTE: If automatic hammer recovery is available, the automatic speed variator will be shown instead of the manual one.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 33
2.6.3.2 THERMOSTATIC CELL CONFIGURATION
To configure the Thermostatic Cell or the External
Conditioning Unit parameters double-click on the
Thermostatic Cell label.
Set the following parameters according to the custom-
ers Thermostatic Cell or External Conditioning Unit
characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the
thermo-regulator or the external conditioning units is
connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween the software and the thermo-regulator or the ex-
ternal conditioning unit (default value is suggested).
Name: insert the customized name for the thermostat-
ic cell or external conditioning unit.
The Enable checkbox allows the user to enable/disable
the use of the Thermostatic Cell or External Condition-
ing Unit. Check to enable use.
NOTE: Thermostatic cell enabling can also be set in Test Execution. The enabling state set in configuration is used as a default opening Test Execution win-dow.
-
2 - INSTALLATION AND CONFIGURATION
34 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: If the Resilvis instrument is equipped with a Ge-fran made thermo-regulator, then after Visua-lIMPACT software installation the user have to copy the device calibration file provided by the manufacturer in the cfg folder overwriting the existing one.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 35
2.6.3.3 MOVEMENT SYSTEM CONFIGURATION
To configure the Specimen Movement System parame-
ters double-click on the Movement System label.
NOTE: The Movement System component is only availa-ble for the Resilvis system.
Set the following parameters according to customers
Specimen Movement System characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the
movement control module is connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and movement control module (default
value is suggested).
Name: insert customized name for the movement con-
trol module.
The Enable checkbox allows the user to enable/disable
the Specimens Movement System use. Check to enable
use.
NOTE: The Movement enabling can also be set in Test Execution. The enabling state set in configuration
-
2 - INSTALLATION AND CONFIGURATION
36 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
is used as a default opening Test Execution win-dow.
Whit the Movement System disabled, the Resilvis be-
comes a standard Pendulum and only the infrared and
vision system additional options are still available.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 37
2.6.3.4 MEASURING DEVICE CONFIGURATION
To configure the Measuring Device parameters,
double-click on the Measuring Device label.
Set the following parameters according to customers
Measuring Device characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where measur-
ing device is connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and measuring device (default value is
suggested).
Name: insert the customized name for the measuring
device.
The Enable checkbox allows the user to enable/disable
the use of the Measuring Device. Check to enable use.
NOTE: Measuring device enabling can also be set in Test Execution. The enabling state set in configuration is used as a default opening Test Execution win-dow.
-
2 - INSTALLATION AND CONFIGURATION
38 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.5 DAS CONFIGURATION
To configure the DAS parameters, double-click on the
DAS label.
Set the following parameters according to the custom-
ers DAS characteristics:
Instrument type: default field with the component
name.
Connection: automatically selected by the software on
the USB communication port.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and DAS (default value is suggested).
Name: insert the customized name for the DAS.
The Enable checkbox allows the user to enable/disable
the use of the DAS. Check to enable use.
NOTE: DAS enabling can also be set in Test Execution. The enabling state set in configuration is used as a default opening Test Execution window.
Move the cursor over the DAS image and select the
model in use from a list of Visual Impact-compatible
CEAST DAS.
DAS model can be taken from the sticker on the in-
strument case.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 39
NOTE: The DAS model choice will filter to the channel type list showing only those compatibles. When the DAS Junior model is selected, only channel 6 is available.
Select the channel type for each acquisition board in-
stalled on your DAS.
The channel number has to be the same as the acquisi-
tion board position on the DAS.
The available options depend on the DAS model in-
stalled on the Visual Impact system.
DAS 8000 JUNIOR
If the DAS model installed is the 8000 Junior the
available options are:
Device not present: if no acquisition boards are in-
stalled at the selected channel position.
Strain gauge 8K: strain gauge channel with capability
of 8000 acquisition points.
Piezoelectric 8K: piezoelectric channel with capability
of 8000 acquisition points.
Only one option is configurable at a time on the 6th
channel.
DAS 8000
If the DAS model installed is the 8000 the available op-
tions are:
Device not present: if no acquisition boards are in-
stalled at the selected channel position.
Strain gauge 8K: strain gauge channel with capability
of 8000 acquisition points.
Piezoelectric 8K: piezoelectric channel with capability
of 8000 acquisition points.
Both options can be active simultaneously on the DAS
and the acquisition channels must be selected accord-
ing to the DAS hardware configuration. The default
configuration foreseen is:
Ch1 = Strain Gauge 8K
Ch2 = Piezoelectric 8K
-
2 - INSTALLATION AND CONFIGURATION
40 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: Both channels (1 and 2) must be assigned to the corresponding sensor even if one (Strain Gauge or Piezoelectric) is not used.
DAS 16000
If the DAS model installed is the 16000 the available
options are:
Device not present: if no acquisition boards are in-
stalled at the selected channel position.
Strain gauge 16K: strain gauge channel with capabil-
ity of 16000 acquisition points.
Piezoelectric 16K: piezoelectric channel with capabil-
ity of 16000 acquisition points.
Fractovis Device: additional Fractovis communication
channel.
Universal 0-10 V: additional general purpose channel
for signals in the range of 1-10 V
All the options can be active simultaneously on the
DAS and the acquisition channels must be selected ac-
cording to the DAS hardware configuration. The de-
fault configuration foreseen is:
Ch1 = Strain Gauge 16K
Ch2 = Piezoelectric 16K
Ch3 = Fractovis device
Ch4 = Universal 0-10V
NOTE: Both channel (1 and 2) must be assigned to the corresponding sensor even if one (Strain Gauge or Piezoelectric) is not used.
NOTE: The Fractovis Device channel type is not neces-sary for Ceast 9350 and Fractovis Plus use.
For Pendulum, Generic Drop Tower and Generic pen-
dulum systems only, enable the Speed Detector op-
tion for impact speed direct measurement if an exter-
nal photocell is used.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 41
NOTE: The Speed Detector option is visible for the , and system only.
Moreover, for Generic Drop Tower and Generic pendu-
lum systems only you have to specify the Flag length
value as well.
-
2 - INSTALLATION AND CONFIGURATION
42 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.6 CEAST 9050/IMPACTOR II CONFIGURATION
To configure the CEAST 9050/Impactor II parameters,
double-click on the Impactor 2 label.
Set the following parameters according to the custom-
ers CEAST 9050/Impactor II characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the pen-
dulum has been connected.
NOTE: A warning message appears if two or more in-struments are configured to the same port, allow-ing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and pendulum (default value is sug-
gested).
Name: insert the customized name for your CEAST
9050/Impactor II instrument.
The Enable checkbox allows the user to enable/disable
the use of the CEAST 9050/Impactor II. Check it to en-
able use.
NOTE: CEAST 9050/Impactor II enabling can also be set in Test Execution. The enabling state set in confi-guration is used as a default opening Test Execu-tion window.
Move cursor over the CEAST 9050/Impactor II image
and select the instrument in use from a list of Visual
Impact-compatible CEAST CEAST 9050/Impactor II.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 43
The CEAST 9050/Impactor II code is available from the
instruments CE label.
According to the CEAST 9050/Impactor II type, the
software shows related characteristics:
Automatic hammer release: (read only value) moto-
rized system to release the hammer
Automatic brake: (read only value) pneumatic brake
on the hammer
Automatic recover: (read only value) motorized sys-
tem to recover the hammer in the rest position
Emergency button: (read only value) button in a suit-
able position on the instrument body to emergency
stop the instrument
Power restore: (read only value) is the power on but-
ton used to restore power to the instrument after an
emergency situation
Manostat: (read only value) is the pressure gauge that
controls the pneumatic brake and the hammer hook
Open door switch: (read only value) safety mechan-
ism to detect the instrument door opening
Hammer recognition: (read only value) automated
system to detect the hammer model using a detecting
cycle based on the 3 pins coding
Micrometer connection: (read only value) output
connection for the micrometer device signals
Trigger: (read only value) output connection for the
trigger signal
Instrumented IZOD Vice: Izod vice equipped with
load cell for clamping force measurement
Manual speed variator: accessory for releasing angle
changing
NOTE: If automatic hammer recovery is available, the automatic speed variator will be shown instead of the manual one.
-
2 - INSTALLATION AND CONFIGURATION
44 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.7 DOUBLE MEASURING DEVICE CONFIGURATION
If you are configuring an CEAST 9050/Impactor II in-
strument (both not instrumented or instrumented), a
second Measuring device, called Double Measuring
Device is available for configuration.
The configuration parameters for this device are the
same as those shown in the MEASURING DEVICE CONFIGU-
RATION.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 45
2.6.3.8 SPECIMEN DIGITAL MICROMETER CONFIGURATION
To configure the Specimen Digital Micrometer Device
parameters, double-click on the Specimen Digital
Micrometer label.
Set the following parameters according to the custom-
ers Specimen Digital Micrometer characteristics:
Instrument type: default field with the component
name.
Connection: select the communication port where the
Specimen Digital Micrometer is connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween the software and the Digital Micrometer (default
value is suggested).
Name: insert the customized name for the Digital
Micrometer.
NOTE: The configuration of a measuring device as Spe-cimen Digital Micrometer allows to link specimen dimensions directly in lots. To link specimen di-mensions during test execution it is necessary to configure the Digital Micrometer as system com-ponent (See MEASURING DEVICE CONFIGURATION para-graph).
-
2 - INSTALLATION AND CONFIGURATION
46 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.9 CEAST 9340 CONFIGURATION
To configure the Ceast 9340 parameters, double-click
on the Ceast 9340 label.
Set the following parameters according to the custom-
ers Ceast 9340 characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the Ceast
9340 has been connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and Ceast 9340 (default value is sug-
gested).
Name: insert the customized name for the Ceast 9340.
The Enable checkbox allows the user to enable/disable
the use of the Ceast 9340. Check to enable use.
NOTE: The Ceast 9340 enabling can also be set in Test Execution. The enabling state set in configuration is used as a default opening Test Execution win-dow.
According to the Ceast 9340 configuration select the
correct options from the following list:
Support type:
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 47
Fixed: suitable for fixed height support type or pi-
voting loader type installed.
Variable: suitable for variable height or tensile im-
pact support type installed.
Thermostatic Cell: for specimen conditioning
Antirebound System: avoids multiple impact on test
specimen in case of tup rebound.
-
2 - INSTALLATION AND CONFIGURATION
48 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.10 CEAST 9350 CONFIGURATION
To configure the Ceast 9350 parameters, double-click
on the Ceast 9350 label.
Set the following parameters according to the custom-
ers Ceast 9350 characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the Ceast
9350 is connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and Ceast 9350 (default value is sug-
gested).
Name: insert the customized name for the Ceast 9350.
The Enable checkbox allows the user to enable/disable
the use of the Ceast 9350. Check to enable use.
NOTE: The Ceast 9350 enabling can also be set in Test Execution. The enabling state set in configuration is used as a default opening Test Execution win-dow.
According to the Ceast 9350 configuration select the
correct options from the following list:
Support type:
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 49
Fixed: suitable for fixed height support type or pi-
voting loader type installed.
Variable: suitable for variable height or tensile im-
pact support type installed.
Automatic: suitable for automatic support type in-
stalled.
NOTE: If, on an automatic support type capable instru-ment, the automatic support is used as a fixed support, the Fixed value must be selected.
Lubrication System: lubricates the tup insert
Antirebound System: avoids multiple impact on test
specimen in case of tup rebound
Weight Measurement System: if enabled, the sys-
tem automatically verifies that actual drop tower will
match the weight setted in test parameters.
Thermostatic Cell: for specimen conditioning
High Energy System: additional energy system allows
to reach equivalent testing height higher than 1 [m]
Automatic loader: enable the automatic loader device
for specimens
Tup Cleaner: activate the cleaner system for the tup
insert.
2.6.3.11 FRACTOVIS PLUS CONFIGURATION
To configure the Fractovis Plus parameters, double-
click on the Fractovis Plus label.
The Fractovis Plus configuration is the same of that
available for the Ceast 9350 instrument, as illustrated
in the previous paragraph. The only difference is in the
instrument options list where a secondary photocell is
available.
Secondary photocell: calculates the impact absorbed
energy as difference of kinetic energies
-
2 - INSTALLATION AND CONFIGURATION
50 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.6.3.12 WIN9 WEIGHT DEVICE CONFIGURATION
To configure the Win9 Weight Device parameters,
double-click on the Win9 Weight Device label.
Set the following parameters according to the custom-
ers Win9 Weight Device characteristics:
Instrument type: default field with the component
name.
Connection: select the hardware port where the Win9
Weight Device is connected.
NOTE: A warning message will appear when two or more instruments are configured to the same port, al-lowing the user to proceed never the less.
Firmware version: specify the firmware version value
as reported on the Certificate of Quality and Origin
shipped with the device, or, after a firmware upgrade,
the version number of the last loaded firmware.
For devices having a display, the firmware version val-
ue can be read following the instructions reported on
the devices user manual.
If the value showed in the Firmware Version field is
not editable, the devices software and firmware are
fully compatible starting from the firmware version
showed in the field.
Polling time: set the communication time rate be-
tween software and Win9 Weight Device (default value
is suggested).
Name: insert the customized name for the Win9
Weight Device.
The Enable checkbox allows the user to enable/disable
the use of the Win9 Weight Device. Check to enable
use.
NOTE: The Win9 Weight Device enabling can also be set in Test Execution. The enabling state set in con-figuration is used as a default opening Test Ex-ecution window.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 51
2.6.4 UNITS OF MEASURE FOLDER
2.6.4.1 COLUMNS GLOSSARY
Line: Instrument lines (Impact, Rheological,).
Item: Physic quantities.
Unit: Units of measurement currently in use.
Decimals: With how many decimals places each units
value displays.
Last Version: Last saved units of measurement set.
Ceast SI: International System units set provided by
CEAST.
Ceast BU: British Units set provided by CEAST.
2.6.4.2 CUSTOMIZING UNITS
To customize units of measure:
Select an Item row from the list and click on the
Change button
A pop-up window specifies the selected item for which
to choose the unit of measure.
Clicking on the right side down arrow, a drop-down
menu lists the different units to choose from.
Select one and press the OK button.
The current selection will be displayed in the Unit col-
umn, from the main window.
Settings used by the program are shown in the Cur-
rent settings status line.
-
2 - INSTALLATION AND CONFIGURATION
52 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
SAVE A NEW CONFIGURATION
If the user customizes units of measurements, the new
unsaved configuration appears in the Units of Measure
folder status bar.
To save the new units configuration, once the units of
measure customization has been completed, press the
Last version button.
Press the Yes button in the next dialog to overwrite
the Last Version with the new units configuration.
The new saved configuration status appear on the
Units of Measure folder status bar and in the Last
version column of the units list.
RESTORE A CONFIGURATION
The user can load two default units of measurement
configuration sets: Ceast SI and Ceast BU. Or revert
the current unsaved configuration to the last saved
configuration.
To load a Ceast SI, Ceast BU, or the Last version confi-
guration, press Ceast SI, Ceast BU, or the Last ver-
sion buttons respectively.
Press the Yes button in the next dialog to overwrite
the Current Settings with the new configuration.
PRINT A CONFIGURATION
To print a report of a units of measurement configura-
tion, click on the Print button.
2.6.4.3 CHANGE DECIMAL PLACES VISUALIZATION
The user can customize with how many decimal places
each unit of measurement will be shown.
Select an item row from the units list and click on the
Change Decimal button
In the next dialog the user must specify the number of
decimal places by clicking on the right side down ar-
row, and selecting a number from the list.
Press the OK button to close the dialog.
The current value for the decimal places parameter
displays in the Decimals column in the units list.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 53
2.6.5 MISCELLANEOUS FOLDER
This folder allows the user to configure main Ceast-
VIEW software options.
- Process Area Display process window at start up: check to auto-
matically open the synoptic panel for test and instru-
ment control (See TEST EXECUTION chapter).
Auto-print after test end: check to send an automat-
ic report print after test end to the configured default
printer (See DATA PRINTOUT chapter).
Auto log-on as Operator: check to open CeastVIEW
software in Operator Mode (See OPERATING PROFILES
chapter)
Auto-Export after test end: check to automatically
export test results according to the export configura-
tion (See DATA EXPORT chapter).
Automatic Start Point Searching: this option is al-
ways checked and set not editable. The start point will
be searched according to the parameter indicated in
the POINTS FOLDER paragraph in the TEST PARAMETERS
chapter.
LOG Area Process LOG: write all messages exchanged between
software and instrument in a ASCII file under the
LOG folder.
Communication LOG: write all software operations
sequence items in a ASCII file under the LOG folder.
-
2 - INSTALLATION AND CONFIGURATION
54 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
Parameters Area No. of optional fields per page [0-5]: allows user to
specify how many optional fields will be shown in the
General and Material folders in the Edit parameters
set windows. User can specify up to 5 optional fields.
This option refers to the Test Parameters management
(See TEST PARAMETERS chapter for more information).
Test Name Options Area Append Date /Hour after Test Name: if checked,
the test timestamp will be added to the end of the test
file name. This option is useful when the user wants to
reuse an already configured loader for more than one
single test.
Database Area In the Database Configuration Area the
button allows the user to
set the folder on which all backup data will be stored
when the is pressed.
NOTE: If the Active backup Directory is not set, at every start CeastVIEW software will display a reminder popup window.
In the Database Actions Area the
button allows the user to
perform a backup of the entire stored dataset of the
CeastVIEW software.
NOTE: A backup dataset comprises both instrument set-tings and test data. All backup data are stored in the \Cfg and \Data folders.
In the Database Actions Area the
button allows the user to per-
form a copy of the entire stored dataset of the Ceast-
VIEW software.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 55
2.6.6 LANGUAGE FOLDER
Use this folder to configure CeastVIEW software lan-
guage.
Its possible for the user to check if the software lan-
guage is available for each instrument line (Base, Hdt,
Rheo, etc.) from this window.
Choose your favorite language, if supported for your
instrument line, and press the OK button.
NOTE: The language is modified when exiting from the System Configuration window. A pop-up window notifies the language modification.
-
2 - INSTALLATION AND CONFIGURATION
56 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
2.7 TUP/HAMMER/VICE CONFIGURATION
To proceed with tup, hammer or vice configurations:
From the main menu, click on
Management Striker Configuration
The Impact Tup/Hammer/Vice Manager window will
appears.
In the Impact Tup/Hammer/Vice Manager window
choose the desired system previously configured.
Choose the test type from those available.
NOTE: The number and type of available tests depend on the selected instrument.
Choose the Tup/Hammer/Vice type (Instru-
mented/Uninstrumented).
The previous points allow to filter the tup codes to the
underlying Tups/Hammers/Vices list.
Choose the tup, hammer or vice to be added to the list
of those configured from the List of the
tup/hammers/vices models and click on the Confi-
gure button.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 57
2.7.1 UNINSTRUMENTED TUP/HAMMER
Insert the serial number of the new tup or hammer, op-
tional notes and click on the OK button.
The tup is added to the List of configured tups.
2.7.2 INSTRUMENTED TUP/HAMMER/VICE
NOTE: Before selecting instrumented devices copy the calibration files into the C:\\CeastVIEW\CFG fold-er.
Insert the serial number of the new tup, hammer or
vice and optional notes.
Click on the Calibration File button and chose the
correct calibration file from windows explorer.
To show calibration file information click on the Open
File button.
A pop-up window appears with the file contents.
Calibration file information includes:
A Information related to D4EXTWIN software only
B Tup, hammer or vice range expressed in [daN] (de-
ca-Newton)
1 [daN] = 10 [N]
C Number of points acquired during the calibration
procedure
D Voltage values acquired during the calibration pro-
cedure; values expressed in [mV]
E Force values applied on the instrumented device
during calibration procedure; values expressed in
[daN]
F Proper instrumented device resonance frequency,
expressed in [kHz]
-
2 - INSTALLATION AND CONFIGURATION
58 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
G Main information including: Order number, Custom-
er, Tup code and Calibration date
H Mass and length of instrumented tup or hammer;
values expressed in [kg] and [m]
I Enabling of piezoelectric sensor:
TRUE = piezoelectric sensor
FALSE = strain gauge sensor
2.7.2.1 MODIFY TUP/HAMMER/VICE PROPERTIES
To modify the properties of a previously configured
tup, hammer or vice:
Select the tup, hammer or vice from the list of those
configured and press the Modify button.
Configure new tup/hammer/vice window will appear.
Fill in with custom information and click on the OK
button to confirm.
NOTE: Tups, hammers or vices already used in parame-ter sets cannot be modified.
2.7.2.2 SELECT NOT USED TUP/HAMMER/VICE
To reduce the list of configured tups/hammers /vices
click on the Select Not Used button.
The list of configured tups/hammers/vices will be fil-
tered showing and selecting the unused devices only.
NOTE: When Select Not Used function is enabled the Un-do button appears instead of the Select Not Used.
To restore the complete list of configured devices click
on the Undo button.
2.7.2.3 REMOVE CONFIGURED TUP/HAMMER/VICE
To eliminate tups, hammers or vices from the list, se-
lect the item and click on the Remove button.
The button Select Not Used allows the user to auto-
matically select the unused tups, hammers or vices
from the configured list.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 59
NOTE: Tups, hammers or vices already used in parame-ter sets cannot be removed.
The Remove command works directly on the
Tup/Hammer/Vice database.
Removed items cannot be restored.
To remove consecutive items simultaneously select the
first item from the list and then press SHIFT + or
SHIFT + (Arrows keys)on the keyboard.
To remove multiple items (not consecutive) press CTRL
and while the button is pressed select all the desired
items with the mouse.
Click on the Remove button when the multiple selec-
tion is completed.
2.7.2.4 DISABLE OBSOLETE TUP/HAMMER/VICE
When a tup, hammer or vice becomes obsolete but is
still used in old parameter configurations, it cannot be
removed.
If you want to eliminate it from the list of available tups
in test parameter configuration:
Select the tup from the list of configured tups, ham-
mers and vice.
Click on the Enable/Disble button to eliminate it from
actual visualization.
NOTE: Disabled tups, hammers and vices are not shown in parameters editing.
To enable/disable consecutive items simultaneously se-
lect the first item from the list and then press SHIFT +
or SHIFT + (Arrows keys) on keyboard.
To enable/disable multiple items (not consecutive)
press CTRL and while the button is pressed select all
the desired items with the mouse.
Press the Enable/Disable button when the multiple se-
lection is completed.
To view all configured tups, hammers or vices (enabled
and disabled) check the View disabled items checkbox.
-
2 - INSTALLATION AND CONFIGURATION
60 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: Disabled tups, hammers or vices are shown as unchecked.
To enable previously disabled tups, hammers or vices:
Select the disabled items from the complete list of con-
figured tups, hammers and vices.
Click on the Enable/Disable button again.
-
2 - INSTALLATION AND CONFIGURATION
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 61
2.8 IMPORT FUNCTION
The CeastVIEW software allows the user to import con-
figurations, parameters, test results and other types of
data in CeastVIEW software.
To activate import functions from the menu bar, under
the File menu, access to the Import submenu.
2.8.1.1 IMPORT CONFIGURATION
This import command allow the user to import an in-
strument configuration database. The name of the da-
tabase containing the data to be imported is
cwDAT.mdb and it is stored in the data folder under
the folder where the CeastVIEW software has been in-
stalled.
NOTE: This procedure also imports the Firmware Version and the System Version stored with the imported configuration.
2.8.1.2 IMPORT DATABASE
This import command allow the user to import test pa-
rameters and test results. The name of the database
containing the data to be imported is cwDAT.mdb and
it is stored in the data folder under the folder where
the Ceast VIEW software has been installed.
Click on the OK button to start the import database
process.
Click on the Cancel button to exit.
If the user import data from a previous version of the
CeastVIEW software database the following scenarios
happen:
-
2 - INSTALLATION AND CONFIGURATION
62 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
IMPORT FROM A VERSION PRIOR TO 5.80.13:
The import process loads, in the Parameter List, the
test parameters as Reference Templates. It does not
import parameters non involved in any test.
IMPORT FROM A VERSION STARTING FROM 5.80.13:
The import process loads, in the Parameter List, the
Templates and Reference Templates as in the original
database.
The test parameters, you can view in the View Parame-
ters from Test or Test List, maintain the Reference
Template name from which they have been executed.
Instead, if they have been executed starting from a
Template (not Reference) they adopt the name of the
test itself.
See the TEST PARAMETERS chapter for more information
about Templates and Reference Templates.
2.8.1.3 IMPORT DAS IMPACT TEST
This is a compatibility import function. Use this com-
mand to import instrumented impact tests saved using
a CeastVIEW software version older than 0700.xxx ver-
sion.
2.8.1.4 IMPORT IMPACT STANDARDS VISUALIZATION DEFAULTS (CUSTOM LEVEL)
This command allows the user to import in the Ceast-
VIEW software the visualization customization for a
test result, as saved using the button (See TEST
OPENING AND DATA ANALYSIS chapter for more details).
-
3 TEST PARAMETERS
CHAPTER 3
TEST PARAMETERS
-
3 TEST PARAMETERS
64 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
3.1 TEMPLATES MANAGEMENT
To open Template management window:
Click on Management menu Parameters com-
mand
or:
Open Text execution window and then click on the
button.
The Parameter window will appear. This window al-
lows user to manage test templates.
3.1.1.1 PARAMETERS WINDOW GLOSSARY
The Parameters windows can be divided in four dis-
tinct sections:
- Column Headers A template is a collection of characteristics. Tem-
plates are shown in the Parameters windows as
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 65
records (one template per row). Characteristics are
shown as fields (one characteristic per column). The
characteristic name is shown in the header of each
column.
User can order templates in ascending order by click-
ing on a header name one time, or in descending order
by clicking on the same header name a second time.
NOTE: When templates list is in a ordered status, a little triangle will appear on the right of the ordering characteristic.
= descending order
= ascending order
Fields can be drag among other columns to alter their
default order. To move a column in another position,
click on the field header and drag it to the new posi-
tion, then release the mouse left button.
- Query row User can filter templates writing filtering criteria in a
cell on the query row. Each field can be used to filter
the templates set shown in the Parameters window.
For numerical field (e.g. Date field) comparison opera-
tors , are allowed.
Example:
-
3 TEST PARAMETERS
66 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
- Templates Grid This area shows all Templates available depending on
active filters previously defined by the user.
NOTE: The characteristics shown in the templates table change according to the selected instrument type.
Double clicking on a row with a template will cause the
Edit parameter set window to appear in which user
can view or edit characteristics of that template.
Double clicking on a row without a template will cause
the No parameter selected error window to be dis-
played.
In the Templates Grid area a contextual menu is avail-
able. User can activate it clicking with the right mouse
button on the grid. That menu allows the user to acti-
vate the following commands:
Copy: Copy the whole content of the templates grid in
the Windows clipboard.
Copy selected: Copy the content of the template se-
lected through the multi selection (CTRL + row) in the
Windows clipboard.
NOTE: These two commands are useful to copy template values in external applications like spreadsheet or text editor.
Select all: Select all rows in the window.
Unselect all: Unselect all rows in the window.
Reverse selection: Toggle selected rows in unse-
lected and vice versa.
Grid Style: This is a modal command. When selected,
the template window appears like the working area of
a spreadsheet. When active, a check sign appears be-
fore the command name.
Alternate Row color style: This is a modal command.
When selected, the template rows appear in alternate
color style. When active, a check sign appears before
the command name. This command has no effect on se-
lected rows.
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 67
Select alt. row color1: Allows user to choice the color
to apply to even rows when Alternate color style com-
mand is active.
Select alt. row color2: Allows user to choice the color
to apply to odd rows when Alternate color style com-
mand is active.
Autosize cols header: Auto size templates window
columns to fit the width of the column header cells.
Some templates cell contents may appear truncated.
Autosize cols rows: Auto size templates window
columns to fit the width of the column contents cells.
Some header cell contents may appear truncated.
- Buttons Area In this area are located command buttons that allows
the user to perform main operations on single template
or set of templates.
All these commands are in depth explained on next pa-
ragraphs.
NOTE: The OK button is enabled only when templates are loaded from the Test execution window.
-
3 TEST PARAMETERS
68 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
3.2 OPERATIONS ON TEMPLATES
A template is a collection of physic, mechanical and
system set-up characteristics. The template scope is
to define the test characteristics in each of its details,
so that a test can be executed as foreseen by a stan-
dard or user customization.
At any time the user can edit a template changing any
of its characteristic. A template can be copied, re-
named or deleted.
Once a test has been executed using a template the
software will automatically make a copy of this tem-
plate (still modifiable), that is then called parameter,
which is linked to the test and not modifiable by the
user.
The parameters are not shown in the template man-
agement window, but can be recalled by the user from
the test analysis window. (See TEST OPENING AND DATA
ANALYSIS chapter for more details).
A template also can be set as reference. A Refer-
ence template becomes not editable, but only when
at least one test has been executed using this template.
The reference templates are not modifiable but still
shown in the templates management window.
In the Buttons area of the parameters windows are lo-
cated the commands that allow the user to perform
main operations on templates.
3.2.1 SET TEMPLATE AS DEFAULT
To set an existing template as default:
Select a template in the Parameter window.
Then press the Set as default button,
Templates created after this command will be created
cloning the characteristics available in the template
just set as default.
3.2.2 TOGGLE A TEMPLATE AS REFERENCE
To set an existing template as Reference Template:
Select a template in the Parameter window.
Then press the Toggle Reference button,
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 69
Reference Templates are templates set as Reference.
When a template assume this status became not edita-
ble, but only when at least one test based on it will be
executed.
NOTE: The user has to press the Toggle Reference button before to execute a test based on that template. Otherwise the test will not be based on that tem-plate.
NOTE: If a test has been executed starting from a Ref-erence Template, the parameters set associated to the test, you can display in View Parameters from Test or Test List, takes the name of the Ref-erence Template. Instead, if a test has been ex-ecuted from a Template (not Reference), the pa-rameters set associated to the test take the name of the test itself.
3.2.3 SELECT NOT USED TEMPLATE SETS
To reduce the list of template sets click on the Select
Not Used button.
The list of template sets will be filtered showing and
selecting those not used.
When Select Not Used function is enabled, the Undo
button appears instead of the Select Not Used
To restore the complete list of available template sets
press the Undo button.
3.2.4 DELETE TEMPLATE SETS
To delete a template set:
Select the desired template set then press the Remove
button.
A confirmation window will be shown.
Click on the OK button when the selection is com-
pleted.
A second confirmation window will be displayed.
Click on the OK button to delete the template.
-
3 TEST PARAMETERS
70 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
NOTE: Removed templates cannot be restored.
Clicking on Remove command before to select a tem-
plate will cause the No parameter selected error
window to be displayed.
Reference templates cannot be deleted if they have at
least a test that use their characteristics.
To remove consecutive templates simultaneously select
the first template from the list and then press and hold
SHIFT + Arrow keys or on the keyboard.
To remove non consecutive templates keep the CTRL
key pressed and click on the desired template sets with
the mouse. Click on the Remove button when the mul-
tiple selection is completed.
3.2.5 COPY TEMPLATE SETS
To copy a template set:
Select a template set from the list.
Click on the Copy button.
Insert a name for the copy.
Click on the OK button.
This procedure is suggested when modifying only a few
fields in the template set.
3.2.6 RENAME TEMPLATE SETS
To rename a template set:
Select a template set from the list.
Click on the Rename button.
Insert a new name for the set.
Click on the OK button
Clicking on Rename command before to select a tem-
plate will cause the No parameter selected error
window to be displayed.
3.2.7 VIEW OR MODIFY AN EXISTING TEMPLATE SET
To view details or to modify an existing template set:
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 71
Select a template set from the list.
Click on the View / Edit... button.
Edit parameter set window will be opened in edit
mode, or in read-only mode if the template set is a Ref-
erence template with at least one test associated or in
use on Text execution window.
3.2.8 ADD NEW TEMPLATE SETS
To create a new template:
Click on the Add new... button.
Edit parameter set window will be opened in edit
mode allowing user to create a new template.
-
3 TEST PARAMETERS
72 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
3.3 EDIT TEMPLATE SETS
To open Edit Template set window:
Click on the Add new... button to create a new tem-
plate set or on the View/Edit... button to modify an ex-
isting set.
The Edit parameter windows comprises eight folder.
3.3.1 GENERAL FOLDER
The General folder has 5 fields by default plus up to
five optional fields
Instrument: Show the instrument name as specified
during system configuration as system name. (See IN-
STRUMENT FOLDER paragraph on chapter 2).
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 73
Parameter name: input a name for the set to be
created; if the name already exists, the system will ask
that a new one be selected.
Operator, Laboratory, Comment: optional text
fields. They give further details that will be shown in
Parameters window and printed on test reports.
Optional fields: Optional fields are activated in the
System Configuration windows Miscellaneous fold-
er (See MISCELLANEOUS FOLDER paragraph on chapter 2)
and allow the user to set up to customized fields.
User can set a custom label for each optional field by
double-clicking on the optional field label using the
right mouse button (please note: right button and not
the left one), digit a new name and then press the OK
button to accept, or the C button to cancel.
-
3 TEST PARAMETERS
74 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
3.3.2 STANDARD FOLDER
The Standard folder consists of four sections:
Section 1: allows user to choose test type.
Section 2: allows user to choose international stan-
dards according to test type choice.
Section 3: shows the list of configured tups, hammers
or vices according to the selected test type and inter-
national standards.
Section 4: allows user to customize test templates ac-
cording to the configured instrument type.
- Test Type The test types list varies depending on the instrument
type selection. The following pictures show the list of
tests available for a pendulum instrument and for a
drop tower instrument.
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 75
List of available tests for a pendulum instrument:
List of available tests for drop tower instruments:
The Test Type selection allows user to choose a type
of test from the following:
Charpy
Izod
Tensile
Pipes
DynStat (for Pendulum only)
Biaxial (for Ceast drop towers only)
CAI- (Compression After Impact) Rebound (for Ceast
drop towers only)
Components (for Ceast drop towers only)
The selection of Test type automatically filters the con-
tent of the Standard List.
NOTE: A graphic representation shows the type of im-pact (in red) on the specimen (in yellow or grey).
- Standard List Shows the list of available international standard fil-
tered by selected test type.
It allows user to choose from the following:
AECMA
AIRBUS AITM
ASTM
-
3 TEST PARAMETERS
76 VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1
BOEING BSS
BS
DIN
ISO
NFT
SACMA SRM
Custom standard.
The selection of the Standard automatically filters to
the contents of tup/hammer/vice lists.
Only if the STANDARD CUSTOM option is chosen,
tup\hammer\vice lists will depend only on the selected
Test type.
NOTE: When the STANDARD CUSTOM option is selected, creating a customized name is possible.
- Tups List Configured tup\hammer\vice (See INSTRUMENT CONFIGU-
RATION paragraph on chapter 2) are shown in this list,
according to selected Test type and Standard.
For each tup, hammer or vice the table includes the
main characteristics useful to simplify the device
choice.
NOTE: Nominal Energy and Velocity for the Pendulum hammer are referred to 150 degrees impact an-gle.
NOTE: The Serial number and device code are usually marked on the tool.
For instrumented tensile impact test, two selections
are needed: one for uninstrumented hammer/tup and
one for instrumented vice.
Choose instrumented and uninstrumented devices from
the two lists available:
Uninstrumented hammer or tup list
Instrumented vices
-
3 TEST PARAMETERS
VISUAL IMPACT INSTRUCTION MANUAL 0710.680 ED. 7 REV. 1 77
If CEAST 9050/Impactor II is the current configured
instrument, the Hammer autodetection enabled
checkbox is also available. If selected, the configured
hammers list disappear from the window.
The Hammer autodetection system of the CEAST
9050/Impactor II instrument can recognize all ham-
mers complying with the 3 pins standard. The CEAST
9050/Imp