Product Certifications 00825-0200-4140, Rev AC July 2017 Product Certifications Rosemount 2140 Level Detector Vibrating Fork

Transcript of Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME...

Page 1: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product Certifications00825-0200-4140, Rev AC

July 2017

Rosemount™ 2140 Level DetectorVibrating Fork

Product Certifications
Page 2: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

Failure to follow safe installation guidelines could result in death or serious injury.

This document contains information required to safely install, connect, commission, operate and maintain, and is required for certification compliance.

Use the level detector only as specified in this document and the product reference manual. Refer to the Rosemount 2140 Reference Manual for more instructions.

The level detector must be installed, connected, commissioned, operated, and maintained by suitably qualified personnel only, observing national and local requirements that may apply.

The weight of a level detector with a heavy flange and extended fork length may exceed 37 lb. (18 kg).A risk assessment is required before carrying, lifting, and installing the level detector.

Explosions could result in death or serious injury.

Please review all sections of this manual for any restrictions associated with an installation.

Electrical shock could cause death or serious injury.

Make sure that the level detector is not powered when removing the terminal cover and making terminal connections.

If the level detector is installed in a high voltage environment and a fault condition or installation error occurs, high voltage may be present on leads and terminals.

Use extreme caution when making contact with the leads and terminals.

Make sure that power to the level detector is off while making connections.

External surface may be hot.

Care must be taken to avoid possible burns.

Refer to manual

This document contains information required to safely install, connect, commission, operate

and maintain, and is required for certification compliance.

Hot surfaces

The flange and process seal may be hot at high process temperatures.

Allow to cool before servicing.

European directive information . . . . . . . . . . . . 3Ordinary locations certifications . . . . . . . . . . . . 3Canadian Registration Number . . . . . . . . . . . . 3Hazardous locations certifications . . . . . . . . . . 4Combinations of approvals . . . . . . . . . . . . . . . 10

Instructions for hazardous area installations 10Temperature charts . . . . . . . . . . . . . . . . . . . . . 12Intrinsic Safety Installation Drawing . . . . . . . 17EC Declaration of Conformity . . . . . . . . . . . . . 22

2

Page 3: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Product Certifications

1.0 European directive informationThe EU Declaration of Conformity for all applicable European directives for this product can be found on page 22 and at Emerson.com/Rosemount.

2.0 Ordinary locations certificationsG5 USA Ordinary Location

Certificate number: 16 CSA 70098390Standard:UL 61010-1: 2012The level detector has been examined and tested to determine that the design meets basic electrical, mechanical, and fire protection requirements by CSA, a nationally recognized testing laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA). Type 4X.

G6 Canada Ordinary LocationCertificate number: 16 CSA 70098390Standards:CAN/CSA C22.2 No 61010-1-12ANSI/ISA-12.27.01:2011The level detector has been examined and tested to determine that the design meets basic electrical, mechanical, and fire protection requirements by CSA, a nationally recognized testing laboratory as accredited by the Standards Council of Canada (SCC). Type 4X. Single Seal.

Special conditions for use (G5 and G6)

1. To be supplied from a Class 2 or Limited Energy Source in accordance withCAN/CSA C22.2 No. 61010-1-12.

3.0 Canadian Registration NumberCertificate number: CRN 0F04227.2

Standards:ASME B31.3:2014ASME B16.5:2013

The requirements of CRN are met when a Rosemount 2140 Level Detector is configured with 316/316L stainless steel (1.4401/1.4404) process-wetted parts and either NPT threaded or 2 to 8-in. ASME B16.5 flanged process connections.

3

Page 4: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

4.0 Hazardous locations certifications

4.1 USA and CanadaExplosion-proof approvals

E5 USA Explosion-proof and Division 2 (XP)Certificate: CSA 16CA70098390X Standards:FM Class 3600 - 2011FM Class 3615 - 2015UL 61010-1: 2012Markings:Class I Groups B, C and D, T6...T2Class I Division 2 Groups A, B, C and D, Type 4XClass I, Zone 1, AEx db IIC T6…T2 GbAmbient and process temperatures:See “Flameproof and explosion-proof approval temperatures” on page 14.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

E6 Canada Explosion-proof and Division 2 (XP)Certificate: CSA 16CA70098390X Standards:ANSI/ISA 12.27.01:2011CSA Std. C22.2 No. 30 -M1986CSA Std. C22.2 No.60079-0-15CSA Std. C22.2. No.60079-1-16CSA Std. C22.2 No. 61010-1-12CSA Std. C22.2 No.94-M91CSA Std. C22.2 No. 213-2016Markings:Class I Groups B, C and D, T6...T2Class I Division 2 Groups A, B, C and D, Type 4XEx db IIC T6…T2 Gb, single sealAmbient and process temperatures:See “Flameproof and explosion-proof approval temperatures” on page 14.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

Special conditions for use (X) (E5 and E6)

1. The user is to ensure the probe assembly is installed in such a way to prevent any damage due to impact or ignition source due to friction.

2. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

3. Flameproof joints are not intended for repair.

4. To be supplied from a Class 2 or Limited Energy Source in accordance withCAN/CSA C22.2 No. 61010-1-12.

4

Page 5: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Intrinsic safety and non-incendive approvals

I5 USA Intrinsic Safety (IS) and Non-Incendive (NI)Certificate: CSA 16CA70098390X Standards:FM Class 3600:2011FM Class 3610:2015FM Class 3611:2004Markings:Class I Groups A, B, C and D, T5…T2Class I, Division 2, Groups A, B, C, and D, Type 4XClass I, Zone 0, AEx ia IIC T5…T2 Gawhen connected using installation drawing 71097/1387 (Figure 7 on page 17).Ambient and process temperatures:See “Intrinsic safety/intrinsically safe approval temperatures” on page 12.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

I6 Canada Intrinsic Safety and Non-IncendiveCertificate: CSA 16CA70098390X Standards:ANSI/ISA 12.27.01:2011CSA Std. C22.2 No. 157 -92CSA Std. C22.2 No. 60079-0-15CSA Std. C22.2. No. 60079-11-14Markings:Class I Groups A, B, C and D, T5…T2Class I, Division 2, Groups A, B, C, and D, Type 4XEx ia IIC T5…T2 Ga, single sealwhen connected using installation drawing 71097/1387 (Figure 7 on page 17).Ambient and process temperatures:See “Intrinsic safety/intrinsically safe approval temperatures” on page 12.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

Special conditions for use (X) (I5 and I6)

1. The equipment, when fitted with the transient suppression terminal block, is not capable of withstanding the 500 volt insulation test. This must be taken into account when installing the equipment.

2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish. However, care should be taken to protect it from impact or abrasion if located in Zone 0.

3. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

5

Page 6: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

4.2 European approvalsATEX flameproof approval

E1 ATEX FlameproofCertificate: Dekra 16ATEX0082X Standards:EN60079-0:2012+A11:2013EN60079-1:2014EN60079-26:2015Markings:

II 1/2 G, Ex db IIC T6...T2 Ga/GbAmbient and process temperatures:See “Flameproof and explosion-proof approval temperatures” on page 14.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

ND ATEX DustCertificate: Baseefa 16ATEX0137XStandards:EN60079-0:2012+A11:2013EN60079-31:2014Markings:

II 1 D, Ex ta IIIC (T92°C…T272°C) (T500100°C…T500280°C) DaAmbient and process temperatures:See “Dust approval temperatures” on page 15.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

E8 Combines E1 and NDAmbient and process temperatures:See “Flameproof and explosion-proof approval temperatures” on page 14.See “Dust approval temperatures” on page 15.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

ATEX intrinsic safety approval

I1 ATEX Intrinsically Safe and Dust (Zone 0, 20)Certificates:Baseefa 16ATEX0136X and Baseefa 16ATEX0137XStandards:EN60079-0:2012+A11:2013EN60079-11:2012EN60079-26:2015EN60079-31:2014Markings:

II 1 G, Ex ia IIC T5...T2 Ga II 1 D, Ex ta IIIC (T92°C...T272°C) (T500100°C...T500280°C) Da

Input parameters:Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 F, and Li = 0 mH

6

Page 7: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Ambient and process temperatures:See “Intrinsic safety/intrinsically safe approval temperatures” on page 12.See “Dust approval temperatures” on page 15.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

(I1 includes the ND approval)

I8 ATEX Intrinsic Safety (Zone 1)Certificate: Baseefa 16ATEX0136XStandards:EN60079-0:2012+A11:2013EN60079-11:2012EN60079-26:2015Markings:

II 1/2 G, Ex ib IIC T5...T2 Ga/GbInput parameters:Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 F, and Li = 0 mHAmbient and process temperatures:See “Intrinsic safety/intrinsically safe approval temperatures” on page 12.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

Special conditions for use (X) (E1 and E8)

1. The user is to ensure the probe assembly is installed in such a way to prevent any damage due to impact or ignition source due to friction.

2. Non-standard paint options may cause risk from electrostatic discharge. Avoid installations that could cause electrostatic build-up of painted surfaces, and only clean the painted surfaces with a damp cloth. If paint is ordered through a special option code, contact the manufacturer for more information.

3. Flameproof joints are not intended for repair.

Special conditions for use (X) (I1 and I8)

1. The equipment, when fitted with the transient suppression terminal block, is not capable of withstanding the 500 volt insulation test. This must be taken into account when installing the equipment.

2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish. However, care should be taken to protect it from impact or abrasion if located in Zone 0.

3. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

Special conditions for use (X) (ND, E8, and I1)

1. Cable entries must maintain the ingress protection rating of the enclosure to at least IP66.

2. Unused cable entries must be filled with suitable blanking plugs which maintain the ingress protection of the enclosure to at least IP66.

7

Page 8: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

3. Cable entries and blanking plugs must be suitable for the ambient temperature of the equipment and be capable of withstanding a 7J impact test.

4. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

4.3 International approvalsInternational Electrotechnical Commission (IEC)flameproof approval

E7 IECEx Flameproof and DustCertificates: IECEx DEK 16.0040X and IECEx BAS 16.0106XStandards:IEC60079-0:2011IEC60079-1:2014IEC60079-26:2014IEC60079-31:2013Markings:Ex db IIC T6...T2 Ga/GbEx ta IIIC (T92°C…T272°C) (T500100°C…T500280°C) DaAmbient and process temperatures:See “Flameproof and explosion-proof approval temperatures” on page 14.See “Dust approval temperatures” on page 15.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

(E7 also includes the NK approval)

International Electrotechnical Commission (IEC)intrinsically safe approval

I7 IECEx Intrinsically SafeCertificate: IECEx BAS 16.0105XStandards:IEC60079-0:2011IEC60079-11:2011Markings: Ex ia IIC T5...T2 GaInput parameters:Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 F, and Li = 0 mHAmbient and process temperatures:See “Intrinsic safety/intrinsically safe approval temperatures” on page 12.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

8

Page 9: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

International Electrotechnical Commission (IEC)dust approval

NK IECEx DustCertificate: IECEx BAS 16.0106XStandards:IEC60079-0:2011IEC60079-31:2013Markings:Ex ta IIIC (T92°C…T272°C) (T500100°C…T500280°C) DaAmbient and process temperatures:See “Dust approval temperatures” on page 15.Hazardous area installation instructions:See “Instructions for hazardous area installations” on page 10.

Special conditions for use (X) (E7)

1. The user is to ensure the probe assembly is installed in such a way to prevent any damage due to impact or ignition source due to friction.

2. Non-standard paint options may cause risk from electrostatic discharge. Avoid installations that could cause electrostatic build-up of painted surfaces, and only clean the painted surfaces with a damp cloth. If paint is ordered through a special option code, contact the manufacturer for more information.

3. Flameproof joints are not intended for repair.

Special conditions for use (X) (I7)

1. The equipment, when fitted with the transient suppression terminal block, is not capable of withstanding the 500 volt insulation test. This must be taken into account when installing the equipment.

2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish. However, care should be taken to protect it from impact or abrasion if located in Zone 0.

3. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

Special conditions for use (X) (NK, E8)

1. Cable entries must maintain the ingress protection rating of the enclosure toat least IP66.

2. Unused cable entries must be filled with suitable blanking plugs which maintain the ingress protection of the enclosure to at least IP66.

3. Cable entries and blanking plugs must be suitable for the ambient temperature of the equipment and be capable of withstanding a 7J impact test.

4. The enclosure may be given a non-standard paint finish which may constitute a potential electrostatic ignition risk. Care should be taken to protect it from external conditions conducive to the build-up of electrostatic charge on such surfaces. The enclosure must not be rubbed or cleaned with a dry cloth.

9

Page 10: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

5.0 Combinations of approvalsK1 Combines I1 and E1K5 Combines I5 and E5KB Combines I5, I6, E5, and E6KZ Combines G5 and G6

6.0 Instructions for hazardous area installations

6.1 General1. Installation of this equipment shall be carried out by suitably trained personnel, in

accordance with the applicable code of practice.

2. Inspection and maintenance of this equipment shall be carried out by suitably trained personnel, in accordance with the applicable code of practice.

3. The user should not repair this equipment

4. The certification of this equipment relies upon the following materials used in its construction:

Housing and cover:Aluminum alloy ASTM B85 A360.0 or stainless steel 316C12

Probe (partition wall):Stainless steel 316/316L, or UNS N10276, or UNS N10002, or UNS N30002

Seals: Silicone.

5. If the equipment is likely to come into contact with aggressive substances, it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected, thus ensuring that the type of protection is not compromised.

Aggressive substances

e.g. acidic liquids or gases that may attack metals or solvents that may affect polymeric materials.

Suitable precautions

e.g. regular checks as part of routine inspections or establishing from the material's data sheet that it is resistant to specific chemicals.

6. It is the responsibility of the user to ensure:a. The voltage and current limits for this equipment are not exceeded.b. The protective earth terminal is connected to an external protective earthing

system.c. That the joint requirements between the probe and the vessel tank are

compatible with the process media.d. That the joint tightness is correct for the joint material used.e. Safe working practices for the media and process concerned are followed during

the installation and maintenance of the equipment.

7. The probe fork is subjected to small vibration stresses as part of its normal function. As this provides a partition wall, it is recommended that the fork should be inspected every two years for signs of defects.

8. Technical data

10

Page 11: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

a. Pressure must not exceed the rating of the coupling/flange fitted.b. For material details and pressure ratings, refer to the Rosemount 2140

Reference Manual.c. Year of manufacture is marked on the product label.

6.2 ATEX specific instructions1. The equipment has not been assessed as a safety related device (as referred to by

directive 2014/34/EU Annex II, clause 1.5).

6.3 Intrinsic safety specific instructions1. The equipment may be used with flammable gases and vapors with apparatus

groups IIA, IIB and IIC, and with temperature classes T1, T2, T3, T4 and T5. The temperature class of the installation will be determined from the higher of the process or ambient temperature.

2. The equipment with Product Certifications code “I8” is suitable for installation across the boundary between an area that specifically requires Equipment Protection Level Ga (Zone 0) and an area that specifically requires Equipment Protection Level Gb (Zone 1). The probe forks (and extension tube) are only to be installed in Zone 0.

3. When fitted with the Transient Protection Terminal Block (option code T1), the unit does not meet the requirements of clause 6.3.13 (Isolation of circuits from earth or frame) in EN 60079-11:2012 (IEC 60079-11:2011).

4. Technical dataa. Input parameters: Ui: 30 V, Ii: 100 mA, Pi: 0.9 W, Ci: 0.012 F, Li: 0

6.4 Flameproof and explosion-proof specific instructions1. The equipment may be used with flammable gases and vapors with apparatus

groups IIA, IIB and IIC, and with temperature classes T1, T2, T3, T4, T5 and T6. The temperature class of the installation will be determined from the higher of the process or ambient temperature.

2. The equipment is suitable for installation across the boundary between an area that specifically requires Equipment Protection Level Ga (Zone 0) and an area that specifically requires Equipment Protection Level Gb (Zone 1). The probe forks (and extension tube) are only to be installed in Zone 0.

3. Repair or modification of flame-paths is not permitted.

4. It is the responsibility of the user to ensure:a. The housing anti-rotation grub screw is screwed in fully.b. The housing cover locking screws are unscrewed tightly against the covers.c. That only suitably certified cable entry devices will be utilized when connecting

this equipment. The cable entry temperature may exceed 70 °C.d. That suitable temperature rated cable is used. For connections in an ambient in

excess of 60 °C, use cable rated to at least 90 °C.e. That any unused cable entries are sealed with suitably certified stopping plugs.

5. Only screws and fittings provided by the manufacturer can be attached to the housing.

6. Technical dataa. Electrical: Vmax = 42.4 Vdc, Imax =23.5 mA.

11

Page 12: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

6.5 Dust (protection by enclosure) specific instructions1. The equipment may be used in a hazardous area with explosive dusts with

apparatus groups IIIC, IIIB, and IIIA. The maximum surface temperature of the installation will be determined from the higher of the process or ambient temperature.

2. It is the responsibility of the user to ensure:a. The housing anti-rotation grub screw is screwed in fully.b. The housing cover locking screws are unscrewed tightly against the covers.

c. That only suitably certified cable entry devices will be utilized when connecting this equipment. The cable entry temperature may exceed 70 °C.

d. That suitable temperature rated cable is used. For connections in an ambient in excess of 60 °C, use cable rated to at least 90 °C.

7.0 Temperature chartsThe graphs in Figure 1 on page 13 through Figure 6 on page 16 show:

the maximum Ambient Temperature (Ta) permitted for a particular Process Temperature (Tp).

the permitted Ambient (Ta) and Process (Tp) Temperature ranges for a particular gas Temperature Class.

7.1 Intrinsic safety/intrinsically safe approval temperatures

2140****M*T5: ambient air temperature (Ta) = -60 °C to +40 °C

process temperature (Tp) = -40 °C to +95 °CT4…T1: see Figure 1 on page 13.

2140****E*T5: ambient air temperature (Ta) = -60 °C to +40 °C

process temperature (Tp) = -70 °C to +95 °CT4…T1: see Figure 2 on page 13.

12

Page 13: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Figure 1. 2140****M* Temperature De-rating (Intrinsically Safe)

Figure 2. 2140****E* Temperature De-rating (Intrinsically Safe)

Am

bie

nt

Tem

per

atur

e (°

C)

70

60.7

53

40

0

-60-40 0 70 95 130 180

Process Temperature (°C)

T4...T1

T5...T1T3...T1

Am

bie

nt

Tem

per

atu

re (°

C)

Process Temperature (°C)

0

40

-60-70 0 70 95 130 195 260

T4...T1

T5...T1

T3...T1 T2...T1

7066.863.460

13

Page 14: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

7.2 Flameproof and explosion-proof approval temperatures

2140****M*T6: ambient air temperature (Ta) = -40 °C to +65 °C

process temperature (Tp) = -40 °C to +80 °CT5…T1: see Figure 3 on page 14

2140****E*T6: ambient air temperature (Ta) = -40 °C to +65 °C

process temperature (Tp) = -70 °C to +80 °C

T5…T1: see Figure 4 on page 15

Figure 3. 2140****M* Temperature De-rating (Flameproof/Explosion-proof)

T6...T1

T5...T1

T4...T1 T3...T1

Am

bie

nt

Tem

per

atu

re (°

C)

Process Temperature (°C)

-40-40 0 80 95 130 180

0

8077706560

14

Page 15: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Figure 4. 2140****E* Temperature De-rating (Flameproof/Explosion-proof)

7.3 Dust approval temperatures

2140****M*Minimum ambient air temperature (Ta) = -20°C

Maximum ambient air temperature (Ta) = See Figure 5

Maximum Surface Temperature (T**°C) = See Figure 5

Maximum Surface Temperature under 500 mm layer (T500***°C) = See Figure 5

2140****E*Minimum ambient air temperature (Ta) = -20°C

Maximum ambient air temperature (Ta) = See Figure 6

Maximum Surface Temperature (T**°C) = See Figure 6

Maximum Surface Temperature under 500 mm layer (T500***°C) = See Figure 6

Am

bie

nt T

emp

erat

ure

(°C

)

Process Temperature (°C)

80

-40-70

7073.57779

65

0

0 80 95 130 195 260

T6...T1

T5...T1

T4...T1 T3...T1 T2...T1

15

Page 16: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

Figure 5. 2140****M* Maximum Surface Temperature

Figure 6. 2140****E* Maximum Surface Temperature

Surf

ace

Tem

per

atu

re (°

C)

Max

imu

m A

mb

ien

t Te

mp

erat

ure

(°C

)

Process Temperature (°C)

Ambient

T500***°C

T***°C

-400

0 80 180

92100

200192

80

60

Surf

ace

Tem

per

atu

re (°

C)

Ambient

T500***°C

T***°C

Am

bie

nt

Tem

per

atu

re (°

C)

Process Temperature (°C)

2600 80-700

92100

280272

80

70

16

Page 17: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

8.0 Intrinsic Safety Installation Drawing

Figure 7. USA and Canada Installation drawing 71097/1387 (Page 1)

APPROVAL

TITLE

APPR.DRG. I.S. & NONINCENDIVE

CONTROL DRAWING FOR ROSEMOUNT 2140

DOCUMENT NUMBER: -

71097/1387Page 1 of 5

AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16REVISION DATE ECO No. NAME APPROVED SEE ECO

L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.

CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.

GENERAL NOTES:

1. ASSOCIATED APPARATUS MANUFACTURER'S INSTALLATION DRAWINGMUST BE FOLLOWED WHEN INSTALLING THIS EQUIPMENT.

2. CONTROL EQUIPMENT CONNECTED TO BARRIER MUST NOTUSE OR GENERATE MORE THAN 250 Vrms or Vdc,

3. RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND ANDEARTH GROUND MUST BE LESS THAN 1 OHM,

4. INSTALLATION SHOULD BE IN ACCORDANCE WITH APPLICABLELAWS/REGULATIONS AND CODE OR PRACTICE. I.E. FOR CANADA,THE CANADIAN ELECTRICAL CODE (CSA C22.1);FOR AMERICA, THE NATIONAL ELECTRICAL CODE (ANSI/NFPA 70)AND ANSI/ISA-RP12.6 “INSTALLATION OF INTRINSICALLY SAFESYSTEMS FOR HAZARDOUS (CLASSIFIED) LOCATIONS”.

5. THE ASSOCIATED APPARATUS, BARRIER OR ISOLATOR MUST BE APPROVED.FOR CANADA, TO CANADIAN STANDARDS BY AN NRTL ACCREDITED BY THESTANDARDS COUNCIL OF CANADA (SCC).FOR AMERICA, TO AMERICAN STANDARDS BY AN NRTL ACCREDITED BYTHE OCCUPATIONAL SAFETY AND HEALTH ADMINISTRATION (OSHA).

6. WARNING -SUBSTITUTION OF COMPONENTS MAY IMPAIRINTRINSIC AND NON-INCENDIVE SAFETY, AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PEUTCOMPROMETTRE LA SÉCURITÉ INTRINSÈQUE ET LA SÉCURITÉNON INCENDIAIRIES

7. ASSOCIATED APPARATUS MUST MEET THE FOLLOWING PARAMETERS:Uo or Voc or Vt LESS THAN or EQUAL TO Ui (Vmax)Io or Isc or It LESS THAN or EQUAL TO Ii (Imax)Po or Pmax LESS THAN or EQUAL TO Pi (Pmax)Ca IS GREATER THAN or EQUAL THE SUM OF ALL Ci's PLUS CcableLa IS GREATER THAN or EQUAL THE SUM OF ALL Li's PLUS Lcable

8. THE ASSOCIATED APPARATUS MUST BE A RESISTIVELY LIMITED SINGLE OR MULTIPLE CHANNEL APPROVED BARRIER HAVING PARAMETERS LESS THAN THOSEQUOTED, AND FOR WHICH THE OUTPUT AND THE COMBINATIONS OF OUTPUTS ISNON-IGNITION CAPABLE FOR THE CLASS, DIVISION AND GROUP OF USE,

9. FIELD WIRING SHOULD BE RATED TO 70°C MINIMUM,

17

Page 18: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

Figure 8. USA and Canada Installation drawing 71097/1387 (Page 2)

APPROVAL

TITLE

APPR.DRG. I.S. & NONINCENDIVE

CONTROL DRAWING FOR ROSEMOUNT 2140

DOCUMENT NUMBER: -

71097/1387Page 2 of 5

AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16REVISION DATE ECO No. NAME APPROVED SEE ECO

L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.

CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.

DIV 1 INSTALLATION OPTIONS

THE ROSEMOUNT TRANSMITTER IS APPROVED AS INTRINSICALLY SAFE WHEN USED IN CIRCUIT WITH APPROVEDBARRIERS WHICH MEET THE ENTITY PARAMETERS LISTED IN THE CLASS I. DIVISION 1 GROUPS INDICATED.ADDITIONALLY, THE ROSEMOUNT 751 FIELD SIGNAL INDICATOR IS APPROVED AS INTRINSICALLY SAFE WHEN CONNECTED IN CIRCUIT WITH ROSEMOUNT TRANSMITTERS AND APPROVED BARRIERS WHICH MEET THE ENTITY PARAMETERS LISTED FOR CLASS I, DIVISION 1, GROUPS INDICATED.

TO ASSURE AN INTRINSICALLY SAFE SYSTEM. THE TRANSMITTER AND BARRIER MUST BE WIRED IN ACCORDANCEWITH THE BARRIER MANUFACTURER'S FIELD WIRING INSTRUCTIONS AND THE APPLICABLE CIRCUIT DIAGRAM.

18

Page 19: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Figure 9. USA and Canada Installation drawing 71097/1387 (Page 3)

APPROVAL

TITLE

APPR.DRG. I.S. & NONINCENDIVE

CONTROL DRAWING FOR ROSEMOUNT 2140

DOCUMENT NUMBER: -

71097/1387Page 3 of 5

AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16REVISION DATE ECO No. NAME APPROVED SEE ECO

L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.

CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.

ENTITY CONCEPT APPROVALSTHE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS TO ASSOCIATED APPARATUS NOT SPECIFICALLY EXAMINED IN COMBINATION AS A SYSTEM. THE APPROVED VALUES OF MAX. OPEN CIRCUIT VOLTAGE (Voc OR Vt) AND MAX. SHORT CIRCUIT CURRENT (Isc OR It) AND MAX,POWER (Voc X Isc/4) OR (Vt X It/4), FOR THE ASSOCIATED APPARATUS MUST BE LESS THAN OR EQUAL TO THE MAXIMUM SAFE INPUT VOLTAGE (Vmax), MAXIMUM SAFE INPUT CURRENT (Imax), AND MAXIMUM SAFE INPUT POWER (Pmax) OF THE INTRINSICALLY SAFE APPARATUS. IN ADDITION, THE APPROVED MAX. ALLOWABLECONNECTED CAPACITANCE (Ca) OF THE ASSOCIATED APPARATUS MUST BE GREATER THAN THE SUM OF THE INTERCONNECTING CABLE CAPACITANCE AND THE UNPROTECTED INTERNAL CAPACITANCE (Ci) OF THE INTRINSICALLY SAFE APPARATUS, AND THE APPROVED MAX. ALLOWABLE CONNECTED INDUCTANCE (La) OF THE ASSOCIATED APPARATUS MUST BE GREATER THAN THE SUM OF THE INTERCONNECTING CABLE INDUCTANCE AND THE UNPROTECTED INTERNAL INDUCTANCE (Li) OF THE INTRINSICALLY SAFE APPARATUS,

NOTE: ENTITY PARAMETERS LISTED APPLY ONLY TO ASSOCIATED APPARATUS WITH LINEAR OUTPUT.

Vmax= 30V Vt OR Voc IS LESS THAN OR EQUAL TO 30V

Imax = 100mA It OR Isc IS LESS THAN OR EQUAL TO 100mA

Pmax = 0.9 WATT ( ) OR ( ) IS LESS THAN OR EQUAL TO 0.9 WATT

Ci = 0.012µF Ca IS GREATER THAN 0.012µF

Li = 0mH La IS GREATER THAN 0mH

CLASS I, DIV 1, GROUPS A AND B

CLASS I, DIV 1, GROUPS C AND D

Vmax= 30V Vt OR Voc IS LESS THAN OR EQUAL TO 30V

Imax = 100mA It OR Isc IS LESS THAN OR EQUAL TO 100mA

Pmax = 0.9 WATT ( ) OR ( ) IS LESS THAN OR EQUAL TO 0.9 WATT

Ci = 0.012µF Ca IS GREATER THAN 0.012µF

Li = 0mH La IS GREATER THAN 0mH

DIV 2 INSTALLATION OPTIONS

19

Page 20: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

Figure 10. USA and Canada Installation drawing 71097/1387 (Page 4)

APPROVAL

TITLE

APPR.DRG. I.S. & NONINCENDIVE

CONTROL DRAWING FOR ROSEMOUNT 2140

DOCUMENT NUMBER: -

71097/1387Page 4 of 5

AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16REVISION DATE ECO No. NAME APPROVED SEE ECO

L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.

CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.

CLASS I, DIV.2 HAZARDOUS (CLASSIFIED) LOCATION

APPROVEDASSOCIATED

NONINCENDIVEAPPARATUS [1]

UNSPECIFIEDEQUIPMENT [3]

UNCLASSIFIEDLOCATION

MECHANICALLYPROTECTEDWIRE PER APPLICABLEELECTRICAL CODE

NOTES:

[1] ASSOCIATED NON-INCENDIVE APPARATUS PARAMETERS SHALL BE THE SAME AS THOSE SHOWN ON PAGE 3

[2] MUST BE INSTALLED IN ACCORDANCE WITH THE APPLICABLE ELECTRIC CODE FOR WIRING IN DIVISION 2 HAZARDOUS (CLASSIFIED) LOCATIONS.

[3] SUPPLIED BY A CLASS 2 OR LIMITED ENERGY SOURCE IN ACCORDANCE WITH CAN/CSA-C22.2 No. 61010-1-12

20

Page 21: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Figure 11. USA and Canada Installation drawing 71097/1387 (Page 5)

APPROVAL

TITLE

APPR.DRG. I.S. & NONINCENDIVE

CONTROL DRAWING FOR ROSEMOUNT 2140

DOCUMENT NUMBER: -

71097/1387Page 5 of 5

AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16REVISION DATE ECO No. NAME APPROVED SEE ECO

L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.

CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.

NONINCENDIVE FIELD CIRCUITCLASS I. DIV. 2 LOCATIONS

NON-HAZARDOUS DIVISION 2 HAZARDOUS (CLASSIFIED) LOCATIONLOCATION Vmax1 Vmax2 Vmax3 VmaxN

Ci1 Ci2 Ci3 CiNLi1 Li2 Li3 LiN

Imax1 Imax2 Imax3 ImaxN

APPROVEDNONINCENDIVE

SUPPLY

WIRING PER APPLICABLEELECTRICAL CODE

IN NORMAL OPERATIONDEVICES CONTROL THROUGH-CURRENT

4-20mA/HART Vmax 42.4VMaxImum normal 23.5mAoperatIng current Ca 0.012µF La 0µH

ROSEMOUNT 2140 TRANSMITTERS ARE CURRENT CONTROLLERSON INDIVIDUAL PARALLEL BRANCHES WITH RESPECT TO THEPOWER SUPPLY. IN NONINCENDIVE INSTALLATIONS THE ImaxFOR EACH TRANSMITTER IS NOT RELATED TO THE MAXIMUMCURRENT OF THE POWER SUPPLY (Isc) lN THE SAME MANNERAS FOR TRANSMITTER INSTALLED PER I.S. REQUIREMENTS.BECAUSE NONINCENDIVE REQUIREMENTS INCLUDE ONLYNORMAL OPERATING CONDITIONS.IT SHALL BE SUPPLIED BY A CLASS 2 OR LIMITED ENERGYSOURCE IN ACCORDANCE WITH CAN/CSA22.2 No. 61010-1-12

PARAMETERS DEVICEVoc = MInImum of (Vmax1, Vmax2, ...., VmaxN) Imax1 >= Iq1 + Isignal1 Imax2 >= Iq2 + Isignal2 , , , ImaxN >= IqN + IsignalN

Ca <= Ci1 + Ci2 + ... + CiN + CcableLa <= Li1 + Li2 + ... + LiN + Lcable

Imax for an Individual device = Iq + Isignal

Iq = Quiescent current through devlce (Maximum quiescent current for the device)Isignal = Signaling current through device (Protocol may limit signaling to one device at a time)

OperatIng Imax = Iq1 + Iq2 + ... + IqN + Isignal max

Isignal max = Max. of (lsignal1, Isignal2 ..... IsignalN)

D D D D

MECANICALLY PROTECTEDWIRE PER APPLICABLE

ELECTRICAL CODE

21

Page 22: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

9.0 EU Declaration of Conformity

Figure 12. EU Declaration of Conformity for Rosemount 2140 (Page 1)

22

Page 23: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

Figure 13. EU Declaration of Conformity for Rosemount 2140 (Page 2)

23

Page 24: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

July 2017Product Certifications

Figure 14. EU Declaration of Conformity for Rosemount 2140 (Page 3)

24

Page 25: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product CertificationsJuly 2017

China RoHS Rosemount 2140List of Rosemount 2140 Parts with China RoHS Concentration above MCVs

Part Name

/ Hazardous Substances

Lead(Pb)

Mercury(Hg)

Cadmium(Cd)

Hexavalent Chromium

(Cr +6)

Polybrominated biphenyls

(PBB)

Polybrominated diphenyl ethers

(PBDE)

Electronics Assembly

X O O O O O

Housing Assembly

O O O X O O

Sensor Assembly

X O O O O

SJ/T11364This table is proposed in accordance with the provision of SJ/T11364.

O: GB/T 26572O: Indicate that said hazardous substance in all of the homogeneous materials for this part is below the limit requirement ofGB/T 26572.

X: GB/T 26572X: Indicate that said hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement of GB/T 26572.

X

25

Page 26: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

February 2017

26

Product Certifications

Page 27: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Product Certifications

27

February 2017

Page 28: Product Certifications: Rosemount™ 2140 Level Detector ... Rosemount Documen… · ASME B31.3:2014 ASME B16.5:2013 The requirements of CRN are met when a Rosemount 2140 Level Detector

Global HeadquartersEmerson Automation Solutions 6021 Innovation Blvd.Shakopee, MN 55379, USA

+1 800 999 9307 or +1 952 906 8888+1 952 949 7001 [email protected]

North America Regional OfficeEmerson Automation Solutions8200 Market Blvd.Chanhassen, MN 55317, USA

+1 800 999 9307 or +1 952 906 8888

+1 952 949 7001

[email protected]

Latin America Regional OfficeEmerson Automation Solutions 1300 Concord Terrace, Suite 400Sunrise, FL 33323, USA

+1 954 846 5030

+1 954 846 5121

[email protected]/company/Emerson-Automation-Solutions

Twitter.com/Rosemount_News

Facebook.com/Rosemount

Youtube.com/user/RosemountMeasurement

Google.com/+RosemountMeasurement

Standard Terms and Conditions of Sale can be found at www.Emerson.com/en-us/Terms-of-UseThe Emerson logo is a trademark and service mark of Emerson Electric Co.Rosemount, and Rosemount logotype are trademarks of Emerson.HART is a registered trade mark of FieldComm Group.All other marks are the property of their respective owners.© 2017 Emerson. All rights reserved.

Europe Regional OfficeEmerson Automation Solutions Europe GmbHNeuhofstrasse 19a P.O. Box 1046CH 6340 BaarSwitzerland

+41 (0) 41 768 6111

+41 (0) 41 768 6300

[email protected]

Asia Pacific Regional OfficeEmerson Automation Solutions Asia Pacific Pte Ltd1 Pandan CrescentSingapore 128461

+65 6777 8211

+65 6777 0947 [email protected]

Middle East and Africa Regional OfficeEmerson Automation Solutions Emerson FZE P.O. Box 17033Jebel Ali Free Zone - South 2Dubai, United Arab Emirates

+971 4 8118100

+971 4 [email protected]

Product Certifications00825-0200-4140, Rev AC

July 2017