PRINCIPLES OF ELECTROSURGERY Muneeb Shah, DO PGY-2 Campbell University Sampson Regional Medical Center

Transcript of Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn...

Page 1: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

PRINCIPLES OF ELECTROSURGERYMuneeb Shah, DO PGY-2

Campbell University

Sampson Regional Medical Center

Page 2: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

DISCLOSURES

I have no financial disclosures that would be a potential conflict of interest with this presentation.

Page 3: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

OUTLINE

Definitions

Mechanism

Clinical uses

Cautions

Complications

Page 4: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

Electrocautery

• *Not electrosurgery

• A metal wire is heated by resistance to flow of direct current

• The tip of the wire is hot

• Advantages:

• Does not interfere with pacemakers / AICDs

• Can achieve hemostasis in a wet field

• Disadvantages:

• May lead to third-degree burns

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 5: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

Electrosurgery(synonym: radiofrequency surgery)

• High-frequency alternating current is passed via a

cold tipped electrode

• Neuromuscular stimulation becomes negligible

as frequency increases

• Tissue resistance to the passage of current converts

electrical energy into heat

• Examples include electrocoagulation,

electrodessication, electrosection, and

electrofulguration

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby and Covidien AG Boulder, Colorado.

Page 6: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTROSURGICAL UNIT

• Electrosurgical unit (ESU) is the source of

electron flow and voltage

• The ESU takes a frequency of 60 Hz

(standard outlet output) and converts it to

over 300 kHz

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 7: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

CIRCUITCircuit is composed of:

Generator

Active electrode

Patient

Return electrode*

Tissue provides resistance or impedance, generating heat

Images from Covidien AG. Boulder, Colorado.

Page 8: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

Monopolar: Electrode with 1 tip

Bipolar: Electrode with 2 tips; Active electrode and return electrode

Monoterminal: 1 electrode, no grounding electrode

Most frequently used in dermatology

Biterminal: 2 electrodes are used; Treating electrode and grounding electrode

-POLAR AND -TERMINAL

Page 9: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

OUTLINE

Definitions

Mechanism

Clinical uses

Cautions

Complications

Page 10: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

CURRENT WAVEFORMSCurrent can be continuous or discontinuous

Continuous current

Cutting mode produces heat very rapidly leading to vaporization

Discontinuous current

“On” time is reduced

Instead of vaporization, a coagulum is produced

Coagulation mode

Fulguration mode

Fulguration mode has the highest peak voltage

Images from Taheri, A., et al. Electrosurgery. Part I. Basics and Princsples. JAAD. Volume 70, Number 4. 2014

Page 11: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

CURRENT WAVEFORMS

Waveforms can be damped or undamped

Undamped waveforms remain unchanged in amplitude throughout the sine wave

Increased cutting effect

Damped waveforms decrease in amplitude with time and eventually approach zero

Increased coagulation effect

Undamped

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 12: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

TISSUE EFFECT

Tissue effect (e.g., coagulation, cutting) is determined by the rate at which heat is produced

High heat produced rapidly Vaporization

Low heat produced slowly Coagulation

Any waveform can accomplish both tasks

Electrocoagulation Slow heating below the boiling point Thermal denaturation of blood products

Electrodessication Slow heating above the boiling point Tissue drying

Electrosection (cutting) Rapid heating above the boiling point Explosive vaporization of water content in tissue and tissue fragmentation

Page 13: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTROFULGURATION↑High voltage, ↓Low amperage

Monoterminal circuit; Monopolar electrode

Discontinuous, damped waveform is applied

Active electrode is held a few millimeters above the tissue

An electrical discharge arc (spark) bridges the gap of air between the electrode and the tissue

Each spark acts as a very fine electrode

Allows for rapid coagulation over a larger area, when compared to contact electrocoagulation

Tissue destruction and coagulation is limited to superficial layer of tissue due to surface carbonization

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 14: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTRODESSICATION

↑High voltage, ↓Low amperage

Monoterminal circuit; Monopolar electrode

Discontinuous, damped waveform is applied

Electrode is in direct contact with the tissue causing superficial ablation

Less heat is generated and no cutting effect occurs

Tissue is heated until the stage of tissue drying

A popping sound will occur with desiccation

Page 15: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTROCOAGULATION

↓ Low voltage, ↑ High amperage

Biterminal circuit; Monopolar or bipolar electrode

Damped waveform is applied

Electrode is brought into direct contact with the tissue being treated

Slow cellular heating leads to fluid evaporation, protein denaturation, and coagulation

Higher amperage allows current to penetrate deeper than electrodessication

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 16: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTROSECTION↓ Low voltage, ↑ High amperage

Biterminal circuit; Monopolar electrode

Continuous, undamped waveform produces a pure cutting effect

Slightly damped, blended current can achieve cutting and hemostasis at the same time

Sudden increase in temperature above the boiling point Explosive vaporization of water content in tissue and tissue fragmentation

Useful for achieving relatively bloodless excisions

Images from Bolognia, J. L. (2012). Dermatology. London: Mosby.

Page 17: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

SUMMARY

Method Voltage Amperage Waveform Circuit

Electrofulguration High Low Discontinuous

Damped

Monopolar

Monoterminal

Electrodessication High Low Discontinuous

Damped

Monopolar

Monoterminal

Electrocoagulation Low High Variable

Damped

Variable

Biterminal

Electrosection Low High Continuous

Variable

Monopolar

Biterminal

Page 18: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

OUTLINE

Definitions

Mechanism

Clinical uses

Cautions

Complications

Page 19: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

ELECTRODESSICATION AND CURETTAGE

Electrosurgery and Curettage is useful for treating superficial malignancies

Preferred for treating small uncomplicated primary basal cell carcinoma and squamous cell carcinoma

Provides 90-95% cure rates

Area is scraped with a curette in all directions and then charred with the electrosurgery device

Can be repeated 2 or more times to remove any residual tumor

Wounds are allowed to heal by secondary intention

Scarring should be anticipated and discussed with the patient prior to the procedure

Images from Jere Mammino, DO.

Page 20: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

OUTLINE

Definitions

Mechanism

Clinical uses

Cautions

Complications

Page 21: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

CAUTIONS

AICDs/Pacemakers

Can theoretically lead to skipped beats, reprogramming of a pacemaker, or firing of an ICD

To minimize risk:

Electrocautery No risk

Bipolar forceps Minimizes risk

Use short bursts of energy (< 5 seconds)

Avoid cutting currents (highest risk with electrosection)

Avoid use on skin around device

Hold magnet over device to avoid electrical interference

Hyfrecators did not interfere with defibrillators and affected pacemakers only within 3 cm of the device (Weyer et al.)

Weyer, C., et al. Investigation of Hyfrecators and Their In Vitro Interference with Implantable Cardiac Devices. Dermatol Surg 2012; 38:1843-1848

Page 22: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

OUTLINE

Definitions

Mechanism

Clinical uses

Cautions

Complications

Page 23: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

COMPLICATIONS

Fire or explosion

Alcohol, oxygen, and bowel gases are flammable

Note: Aluminum chloride solutions contain over 90% alcohol

Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area

Thermoelectric burns

Burns can occur where the current exits the patient’s body

Keep metal objects away from the patient's body (E.g., ECG electrodes)

Microorganism transmission

Can be transmitted by the electrode or smoke plume inhalation

Use a new sterile tip for each patient

Use an N95 mask and smoke evacuation system

Image from Bisinotto, et al. Burns related to electrosurgery – Report of two cases. Rev Bras Anestesiol. Sep 2017: 527-534

Page 24: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

Katoch, S., Mysore, V. Surgical Smoke in Dermatology: Its Hazards and Management. J cutan Aesthet Surg Jan – Mar; 12(1):1-7

RISKS OF SURGICAL SMOKESurgical smoke is comprised of 95% water and 5% particulate matter

Surgical masks filter particles larger than 5 micrometers

77% of particulate matter is less than 1.1 micrometer

Particles less than 2 micrometers can deposit in alveoli

Transmission of infection

HPV, HIV, Staphylococcus, Corynebacterium, Neisseria

Risk of mutagenesis

Chronic inflammation

Transmission of HPV infection

Viable cancer cells in smoke

Page 25: Principles of Electrosurgery · 2019. 9. 12. · Avoid electrosurgery until alcohol has dried, turn off sources of oxygen, and exhibit precaution in the perianal area Thermoelectric

MINIMIZING RISK OF SURGICAL SMOKE

High filtration masks (e.g., N95)

Smoke evacuation system

Katoch, S., Mysore, V. Surgical Smoke in Dermatology: Its Hazards and Management. J cutan Aesthet Surg Jan – Mar; 12(1):1-7