Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE...

36
Creator: Eric Tischer Checked: Date: 3/28/20 2:00 PM/4/8/20 8:31 PM Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram No.: Page: 1 / 36 A Ax Ay Par B006 On =0 Off =0 Gain =1.0+ Offset =0 Point=0 A Ax Par B005 fb_PSI Gain =0.19+ Offset=-37 Point =2 A Ax Ay Par B008 On =0 Off =0 Gain =1.0+ Offset =0 Point=0 Ax En S1 Ax Par B031 PEAK PSI MON Rem = off Mode = 1 ResetMaxMin = false RECORD PEAK INHALE PRESSURE PAGE 1 - PRESSURE MEASUREMENT AI AI1 Pressure 0-10V 0-1.45psi Ax En S1 Ax Par B128 PSI PLATEAU Rem = off Mode = 0 ResetMaxMin = false RECORD PLATEAU PRESSURE Ax En S1 Ax Par B015 PEEP START P Rem = off Mode = 3 ResetMaxMin = false RECORD PEEP STARTING PRESSURE A Ax Ay Par B171 INH DETECTED On =B172 Off =0 Gain =1.0+ Offset =0 Point=0 A Ax Par B172 Gain =1.0+ Offset=0 Point =2 9/B163 PEEP PSI SP PEEP PRESSURE DROP TRIGGER PSI INHALE DETECTED A Ax Ay Par B201 On =0 Off =0 Gain =1.0+ Offset =0 Point=0 AQ NAQ5 VW5 6/B033 Fault 1/S 5/B089 STEP2A 5/B092 STEP5A 5/B104 CYCLE DONE/1 9/B167 WAITPLATEAU >1 B106 PEAK PSI MON 5/B095 STEP2B 10/B011 v_HiLimPSI 10/B013 PK_PL DELTA PRESSURE TOO HIGH A En Par B233 V1=B31+ V2=B128 V3=0 V4=0 Point=0 ((B31-B128)+0)+0 M M24 DELTA PEAK TOO HIGH Trg R T B234 Rem = off 00:01s+ & B235 & B236 Delta PK PSI 6/B034 Fault 2/S Trg R T B022 Rem = off 00:01s+ PLATEAU TOO HIGH & B200 R Cnt Dir Par B237 Rem = off On=0+ Off=0 Start=0 prevent race condoition, may not be needed AQ NAQ2 VW0 X X12 10/B232 X X13

Transcript of Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE...

Page 1: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 1 / 36

AAxAyPar

B006

On =0Off =0Gain =1.0+Offset =0Point=0

AAx

Par

B005 fb_PSI

Gain =0.19+Offset=-37Point =2

AAxAyPar

B008

On =0Off =0Gain =1.0+Offset =0Point=0

Ax

EnS1AxPar

B031 PEAK PSI MON

Rem = offMode = 1ResetMaxMin = false

RECORD PEAK INHALE PRESSURE

PAGE 1 - PRESSURE MEASUREMENT

AIAI1

Pressure 0-10V 0-1.45psi

AxEnS1AxPar

B128 PSI PLATEAU

Rem = offMode = 0ResetMaxMin = false

RECORD PLATEAU PRESSURE

AxEnS1AxPar

B015 PEEP START P

Rem = offMode = 3ResetMaxMin = false

RECORD PEEP STARTING PRESSURE

AAxAyPar

B171 INH DETECTED

On =B172Off =0Gain =1.0+Offset =0Point=0

AAx

Par

B172

Gain =1.0+Offset=0Point =2

9/B163 PEEP PSI SPPEEP PRESSURE DROP TRIGGER PSI

INHALE DETECTED

AAxAyPar

B201

On =0Off =0Gain =1.0+Offset =0Point=0

AQ

NAQ5 VW5

6/B033 Fault 1/S

5/B089 STEP2A

5/B092 STEP5A

5/B104 CYCLE DONE/1

9/B167 WAITPLATEAU

>1B106 PEAK PSI MON

5/B095 STEP2B

10/B011 v_HiLimPSI

10/B013 PK_PL DELTA

PRESSURE TOO HIGH

AEn

Par

B233

V1=B31+V2=B128V3=0V4=0Point=0((B31-B128)+0)+0

MM24

DELTA PEAK TOO HIGH

TrgRT

B234

Rem = off00:01s+

&B235

&B236 Delta PK PSI

6/B034 Fault 2/S

TrgRT

B022

Rem = off00:01s+

PLATEAU TOO HIGH

&B200

RCntDirPar

B237

Rem = offOn=0+Off=0Start=0

prevent race condoition, may notbe needed

AQ

NAQ2 VW0

XX12

10/B232

XX13

Page 2: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 2 / 36

AAx

Par

B007

Gain =1.0+Offset=0Point =0

AAxAyPar

B009

On =0Off =0Gain =1.0+Offset =0Point=0

AAxAyPar

B010

On =0Off =0Gain =1.0+Offset =0Point=0

AEn

Par

B012 v_HiLimAngle

V1=999+V2=0V3=0V4=0Point=0((999+0)+0)+0

AEn

Par

B014 v_LoLimAngle

V1=1+V2=0V3=0V4=0Point=0((1+0)+0)+0

Trg

T

B023 ARM OVERTRAV

Rem = off00:00s+

Trg

T

B024 ARM OVERTRAV

Rem = off00:00s+

ARM ANGLE TOO HIGH ALARM DELAY

ARM ANGLE TOO LOW ALARM DELAY

PAGE 3 - ARM ANGLE MEASUREMENT

AAx

Par

B046 v_OpenAngl

Gain =1.0+Offset=0Point =0

AAx

Par

B048 v_ClosdAngl

Gain =1.0+Offset=0Point =0

AAxAyPar

B049

On =0Off =0Gain =1.0+Offset =0Point=0

AAxAyPar

B051

On =0Off =0Gain =1.0+Offset =0Point=0

&B060

&B061

RSSR

Par

B062 ARM IS OPEN

Rem = off

RSSR

Par

B063 ARM IS CLOSD

Rem = off

>1B076 clrOpenStat

>1B077 clrClosdStat

>1B078 clrArmStatus

3/Q2

3/Q1

3/B069/1

3/B071/1

AIAI2

ARM ANGLE 0-5V = 0-180 DEG

ARM IS OPEN

ARM IS CLOSED

3/B028/2

3/B127/2

9/B191

9/B193

6/B036 Fault 4/S

6/B035 Fault 3/S

5/B075 HOMING DONE/1

5/B080 DONE2A/1

5/B052/3

5/B121 DONE2B/1

&B047

&B050

hiHigh

5/B109 WAIT FOR BPM/1

Page 3: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 3 / 36

PAGE 5 ARM CONTROL

QQ1

QQ2

OPEN ARM

CLOSE ARMRS

SR

Par

B053

Rem = off

RSSR

Par

B054

Rem = off

I

I12

II13

>1B065 mOpenRequest

>1B069

>1B071

>1B073 mClseRequest

&B074 IN_MAN_MODE

2/B076 clrOpenStat/4

2/B077 clrClosdStat/1

2/B062 ARM IS OPEN

2/B063 ARM IS CLOSD

1B085 STOP REQ

>1B067 OpenRequest

>1B079 CloseRequest

&B055 IN_MAN_MODE

>1B056 OPEN CMD

>1B057 CLOSE CMD

>1B058 STOP

&

B066 STOP CMD

II6

(Red PB)

Red PB

2/B062 ARM IS OPEN

2/B063 ARM IS CLOSD

&B110 START CMD

5/B088 STEP1/S

5/B125 ABORT SEQ/1

&B120 ABORT CMD

&B127

&B028

MANUAL BUTTONS ONLY WORK IN MANUAL MODEREQUESTS FROM STEP SEQUENCER ONLY WORK IN AUTO MODE

ARM CLOSE TIMEOUT FAULT

ARM OPEN TIMEOUT FAULT

&B016 IN AUTO MODE

&B019 IN AUTO MODE

5/B099 ACTION1

5/B100 ACTION2A

5/B103 ACTION4A

5/B105 ACTION2B

4/B112/En

4/B114/1

MM15

4/B123

10/B029 almOpenTOD/Trg

10/B059 almClsTOD/Trg

5/B239 ACTION4B

Page 4: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 4 / 36

II3

(Run AC mode)

RUN AC Mode

PAGE 9

II4

(Run BC Mode)

RUN BC Mode

PAGE 7 TICK TOCK BIT

Trg

T

B134

Rem = off00:01s+

Trg

T

B135

Rem = off00:01s+

hi

High

&B136

MM2

TICKTOCK

9/B166/1

10/B192/1

8/B202/4

8/B207/4

>1B086 Mode REQ AC

5/B070 AC START/1

>1B102 Mode REQ BC

5/B101 BC START/2

RCntDirPar

B094 Total cycles

Rem = offOn=999999+Off=0Start=0

B094 Total cycles/Cnt

M2 RCntDirPar

B098

Rem = offOn=0+Off=0Start=0

MM14

REn

RalPar

B112

MI= 0h 0m+OT= 0h 0mQ->0=R+En

3/M15

REn

RalPar

B113

MI= 0h 0m+OT= 0h 0mQ->0=R+En

&

B114

3/M15

XX2

XX3

XX5

TOTAL HOURS IN ASSIST CONTROL (AC) MODE

TOTAL HOURS RUNNING (AC AND BC)

TOTAL ARM CYCLES

&B115

>1B123

3/M15/1

&B003

5/B075 HOMING DONE/4

Page 5: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 5 / 36

PAGE 9RS

SR

Par

B088 STEP1

Rem = off

RSSR

Par

B089 STEP2A

Rem = off

RSSR

Par

B090 STEP3A

Rem = off

RSSR

Par

B091 STEP4A

Rem = off

RSSR

Par

B092 STEP5A

Rem = off

RSSR

Par

B093 STEP6A

Rem = off

RSSR

Par

B095 STEP2B

Rem = off

RSSR

Par

B096 STEP3B

Rem = off

RSSR

Par

B097 STEP4B

Rem = off

&B099 ACTION1

&B100 ACTION2A

&B103 ACTION4A

&B075 HOMING DONE

&B080 DONE2A

&B081 DONE3A

&B082 DONE4A

&B083 DONE5A

&B121 DONE2B

&B122 DONE3B

&B125 ABORT SEQ

&B126 CYCLE DONE

B097 STEP4B/S

B122 DONE3B

B121 DONE2B

B096 STEP3B/S

B083 DONE5A

B093 STEP6A/SB082 DONE4A

B092 STEP5A/S

B081 DONE3A

B091 STEP4A/S

B080 DONE2A

B090 STEP3A/S

B118 CYCLE START/1

B075 HOMING DONE

3/B110 START CMD

MM1

TrgRT

B117

Rem = off00:01s+

>1B118 CYCLE START

&B119 REPEAT

3/B058 STOP

>1B116 CYCLE RESET

STEP 4A - "EXHALE" OPEN ARMS

CYCLE COMPLETE,RESET ALL STEPSRESTART CYCLE

3/B067 OpenRequest/12/B062 ARM IS OPEN

3/B079 CloseRequest/1

2/B063 ARM IS CLOSD1/B106 PEAK PSI MON/1

STEP 1 - HOME ARM

STEP 2A - START BPM TIMER, "INHALE" CLOSE ARMS, TRACK PEAK PRESSURE

STEP 3A - WAIT, THEN MEASURE PLATEAU PRESSURE

3/B067 OpenRequest/2

2/B062 ARM IS OPEN

1/B015 PEEP START P/En STEP 5A - SAVE PEEP START PRESSURETrg

T

B072

Rem = off00:01s+

>1B104 CYCLE DONE

1/B171 INH DETECTED

9/B001 BPM timer

9/B107 WAIT PLATEAU/1

STEP 6A - WAIT FOR PEEP PRESSURE DROPOR WAIT FOR TIMER TO TIME OUTBACKLIGHT LCD AMBER.

&B070 AC START

4/B086 Mode REQ AC

&B101 BC START

4/B102 Mode REQ BC STEP 2B - START BPM TIMER, "INHALE"CLOSE ARMS, TRACK PEAK PRESSURE&

B105 ACTION2B

3/B079 CloseRequest/2 2/B063 ARM IS CLOSD

1/B106 PEAK PSI MON/3

9/B167 WAITPLATEAU

9/B107 WAIT PLATEAU/4 STEP 3B - WAIT, THEN MEASUREPLATEAU PRESSURE

9/B167 WAITPLATEAU

9/B164 START BMP TM/4

9/B001 BPM timer

STEP 4B - WAIT FOR BPM TIMERBACKLIGHT LCD AMBER &

B109 WAIT FOR BPM

7/B108 AMBER LCD /3

--BREATH CONTROL SEQUENCE--

--ASSIST CONTROL SEQUENCE--

9/B164 START BMP TM/1

4/B123

&B052

>1B238

MM32

B238/2

M32

M1

B238/1

&B239 ACTION4B

3/B065 mOpenRequest/22/B062 ARM IS OPEN

7/B108 AMBER LCD /1

Page 6: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 6 / 36

&B032

RS

SR

Par

B033 Fault 1

Rem = off

RSSR

Par

B034 Fault 2

Rem = off

RSSR

Par

B035 Fault 3

Rem = off

RSSR

Par

B036 Fault 4

Rem = off

RSSR

Par

B037 Fault 5

Rem = off

RSSR

Par

B038 Fault 6

Rem = off

RSSR

Par

B039 Fault 7

Rem = off

RSSR

Par

B040 Fault 8

Rem = off

>1B041

>1B042

>1B043

MM29

(LOGO! displays red backlight)

PAGE 11 ALARMS

10/B029 almOpenTOD

10/B059 almClsTOD

EnP

Par

B020

Prio = 100Quit = offText1: enabledText2: disabled

EnP

Par

B030

Prio = 99Quit = offText1: enabledText2: disabled

XX11

1/B006

2/B024 ARM OVERTRAV

2/B023 ARM OVERTRAV

RSSR

Par

B026

Rem = off

RSSR

Par

B044

Rem = off

>1B064

>1B068

II5

(Joystick PB)

Joystick PB

PAGE 13 DISPLAY SCREENS

XX14

loLow

1/B236 Delta PK PSI

Trg

T

B137

Rem = on00:05s+

9/B001 BPM timer

DISABLE ALARMSTEMPORARY

ALARM SCREEN 2 OF 2

ALARM SCREEN 1 OF 2

BPM TIMER EXPIRED BEFORE ARM WAS OPEN. NEED TO INCR.SPEED OF ARM FOR BPM TO BEACCURATE

Page 7: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 7 / 36

EnP

Par

B045

Prio = 83Quit = offText1: enabledText2: disabled

MM25

(LOGO! displays white backlight)

MM28

(LOGO! displays the amber backlight)

EnP

Par

B018 ASSIST CNTRL

Prio = 40Quit = offText1: enabledText2: disabled

hiHigh

EnP

Par

B017

Prio = 80Quit = offText1: enabledText2: disabled

EnP

Par

B129

Prio = 81Quit = offText1: enabledText2: disabled

EnP

Par

B027

Prio = 102Quit = offText1: enabledText2: disabled

EnP

Par

B194

Prio = 112Quit = offText1: enabledText2: disabled

POPUP VARIABLE PAGE 6 OF 6

XX4

XX9

XX22

XX37

EnP

Par

B025

Prio = 84Quit = offText1: enabledText2: disabled

>1B108 AMBER LCD

5/B093 STEP6A XX1

EnP

Par

B002 BREATH CNTRL

Prio = 41Quit = offText1: enabledText2: disabled

EnP

Par

B004 MON INPUTS1

Prio = 91Quit = offText1: enabledText2: disabled

En

PPar

B084 MON INPUTS 2

Prio = 90Quit = offText1: enabledText2: disabled

EnP

Par

B111 STATS

Prio = 10Quit = offText1: enabledText2: disabled

PRESSURE FDBKPRESSURE ALARM LIMITS

STATSSEQUENCE BREATH CONTROL SEQUENCE ASSIST CONTROL

ARM FDBKARM ALARM LIMITS

9/B198 SHOW SCR 1

10/B199En

PPar

B225

Prio = 106Quit = offText1: enabledText2: disabled

SET MAX PRESSURE ALARM 3 OF6

EnP

Par

B226

Prio = 104Quit = offText1: enabledText2: disabled

9/B165

POPUP VARIABLE PAGE 2 OF 6

POPUP VARIABLE PAGE 1 OF 6En

PPar

B168

Prio = 108Quit = offText1: enabledText2: disabled

EnP

Par

B169

Prio = 110Quit = offText1: enabledText2: disabled

SET PEAK-PLATEAU ALARM 4 OF 6

SET MAX PLATEAU PRESSURE 5 OF 6

PAGE 13 DISPLAY SCREENS

10/B229

10/B230

10/B231

XX6

XX7

XX8

XX10

5/B097 STEP4B

MONITOR INPUTS 1 MONITOR INPUTS 2

Page 8: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 8 / 36

AIAI4

MOVE CURSOR TO NEXT VAR

AIAI3

INCREASE/DECREASEVAR VALUE

AAxAyPar

B130 PbCursor+

On =200Off =150Gain =-1.0+Offset =0Point=0

PAGE 15 4 WAY JOYSTICK CONTROL FOR MOVING CURSOR AND INCREASING DECREASING VARIABLE VALUES

AAxAyPar

B131 PbCursor-

On =200Off =150Gain =1.0+Offset =0Point=0

AAxAyPar

B132 PbIncrease

On =200Off =150Gain =-1.0+Offset =0Point=0

AAxAyPar

B133 PbDecrease

On =200Off =150Gain =1.0+Offset =0Point=0

RCntDirPar

B138

Rem = onOn=11-Off=0Start=0

>1B141

AAx

Par

B142

Gain =0.0+Offset=500Point =0

MM3

>1B143 COUNT PULSE

hiHigh

AEnS1S2Par

B146 ENUM

V1 =B138+V2=0V3=0V4=0Point =0

PAGE 17. TRIM ADJUSTMENTS TO VOLUME, PRESSURE, TIMERS... USING CURSOR LOGIC ON PREV. PAGE

9/B150/Ax

9/B149 DIRECTION/1 9/B153 INCREMENT/1

5V = JOY STICK CENTER = NO ACTION<5V = DECREASE; >5V = INCREASE

SEL. NEXT VAR

SEL. PREV. VAR

INCREASE VALUE

DECREASE VALUE

VARIABLE SELECTED ENUM

TrgRT

B087 ShowVarPages

Rem = off10:00s+

>1B144 BUTTON PRESS

10/B188/Ax

9/B198 SHOW SCR 1/4

10/B199/4Trg

T

B204

Rem = off01:00s+

=1B205

=1B203

&B202

4/M2M

M12

Trg

T

B206

Rem = off01:00s+

&B207

MM13

4/M2

RAPID INCR WHEN PB HELD

RAPID DECR WHEN PB HELD

COUNTDIRECTION

SHOW "MODIFY VAR" PAGES

AxEnS1AxPar

B124

Rem = offMode = 3ResetMaxMin = true

DISABLE CHANGES IFSCREEN NOT VISIBLE

9/B165/4

10/B229/4

10/B230/4

10/B231/4

Page 9: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 9 / 36

RCntDirPar

B139

Rem = onOn=200+Off=5Start=16

&

B140

MM4

PAGE 17. TRIM ADJUSTMENTS TO VOLUME, PRESSURE, TIMERS... USING CURSOR LOGIC ON PREV. PAGE

AAx

Par

B150

Gain =1.0+Offset=0Point =0

AAxAyPar

B147

On =1Off =2Gain =1.0+Offset =0Point=0

AAxAyPar

B148

On =2Off =3Gain =1.0+Offset =0Point=0

AAxAyPar

B151

On =3Off =4Gain =1.0+Offset =0Point=0

AAxAyPar

B152

On =4Off =5Gain =1.0+Offset =0Point=0

8/B124

>1B149 DIRECTION

8/B203

>1B153 INCREMENT

8/B143 COUNT PULSE

&B154

&B155

&B156

AAxAyPar

B157

On =5Off =6Gain =1.0+Offset =-4Point=0

&

B158

RCntDirPar

B159

Rem = onOn=500+Off=0Start=50

RCntDirPar

B160

Rem = onOn=500+Off=0Start=5

RCntDirPar

B161

Rem = onOn=1000+Off=0Start=400

RCntDirPar

B162

Rem = onOn=1000+Off=0Start=10

MM5

MM6

MM7

MM9

AEn

Par

B163 PEEP PSI SP

V1=B160+V2=0V3=0V4=0Point=2((B160+0)+0)+0

>1B166

4/M2

Trg

T

B167 WAITPLATEAU

Rem = offB159tx+

Trg

T

B001 BPM timer

Rem = offB186ms+

2/B048 v_ClosdAngl/Ax

1/B172/Ax

AEn

Par

B191

V1=B161+V2=0V3=0V4=0Point=0((B161+0)+0)+0

AEn

Par

B193

V1=B162+V2=0V3=0V4=0Point=0((B162+0)+0)+0

2/B046 v_OpenAngl/Ax

&B198 SHOW SCR 1

8/B087 ShowVarPages

5/B104 CYCLE DONE/3

5/B090 STEP3A

1/M24/1

5/B081 DONE3A/1>1B107 WAIT PLATEAU

5/B096 STEP3B

5/B122 DONE3B/1

5/B089 STEP2A 5/B109 WAIT FOR BPM/2

7/B027/En

SHOW SCREEN 1 OF 6

10/B189/2 10/B190/1

AEn

Par

B186

V1=600+V2=B139V3=100V4=0Point=2((600/B139)*100)+0

>1B164 START BMP TM

5/B095 STEP2B

&B165

8/B087 ShowVarPages7/B226/En

SHOW SCREEN 2 OF 6

>1B145

>1B196

AQ

NAQ1 VW0

6/B137/Trg

Page 10: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 10 / 36

AAx

Par

B011 v_HiLimPSI

Gain =1.0+Offset=0Point =0

AAx

Par

B013 PK_PL DELTA

Gain =1.0+Offset=1Point =0

Trg

T

B029 almOpenTOD

Rem = offB187ms+

Trg

T

B059 almClsTOD

Rem = offB21ms+

6/B038 Fault 6/S

6/B037 Fault 5/S

AAxAyPar

B173

On =6Off =7Gain =1.0+Offset =0Point=0

AAxAyPar

B174

On =7Off =8Gain =1.0+Offset =0Point=0

AAxAyPar

B175

On =8Off =9Gain =1.0+Offset =0Point=0

AAxAyPar

B176

On =9Off =10Gain =1.0+Offset =0Point=0

AAxAyPar

B177

On =10Off =11Gain =1.0+Offset =0Point=0

&B178

&B179

&B181

&B182

RCntDirPar

B183

Rem = onOn=500+Off=0Start=0

RCntDirPar

B184

Rem = onOn=500+Off=0Start=0

MM10

MM11

AAx

Par

B188

Gain =1.0+Offset=0Point =0

8/B124

>1B189

>1

B190

>1B192

4/M2

&B199

8/B087 ShowVarPages

PAGE 10. MORE TRIM ADJUSTMENTS. JOG ARM OPEN/CLOSE

7/B194/En

SHOW SCREEN 6 OF 6

9/B153 INCREMENT

9/B149 DIRECTION

1/B006/Ay

1/B008/Ay

AEn

Par

B180

V1=B183+V2=4V3=0V4=0Point=2((B183/4)+0)+0

AEn

Par

B185

V1=B184+V2=4V3=0V4=0Point=2((B184/4)+0)+0

RCntDirPar

B197

Rem = onOn=500+Off=0Start=0

MM21

&B227

&B229

&B230

&B231

8/B087 ShowVarPages

8/B087 ShowVarPages

8/B087 ShowVarPages

7/B225/En

7/B168/En

7/B169/En

SHOW SCREEN 3 OF 6

SHOW SCREEN 4 OF 6

SHOW SCREEN 5 OF 6

3/B028

3/B127

RCntDirPar

B021

Rem = offOn=60+Off=0Start=0

RCntDirPar

B187

Rem = offOn=60+Off=0Start=0

MM22

MM23

>1B195

AEn

Par

B228

V1=B197+V2=0V3=0V4=0Point=0((B197+0)+0)+0

AAx

Par

B232

Gain =1.0+Offset=0Point =0

1/B201/Ay

Page 11: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 11 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

AI1(Analog input) : Pressure 0-10V 0-1.45psi

AI2(Analog input) : ARM ANGLE 0-5V = 0-180 DEG

AI3(Analog input) : INCREASE/DECREASEVAR VALUE

AI4(Analog input) : MOVE CURSOR TO NEXT VAR

B001 BPM timer(On-Delay) : Rem = offB186ms+

Page 12: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 12 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B002 BREATH CNTRL(Message texts) : SEQUENCE BREATH CONTROL

Prio = 41Quit = offText1: enabledText2: disabled

- - B C - - 1:SEQUENCE0:nothomed

B095 STEP... - -

1:STEP 2B*0:step 2b

B095 STEP2B... I N H A L E

1:STEP 3B*0:step 3b

B096 STEP3B... M E A S U R E

1:STEP 4B*0:step 4b

B096 STEP3B... H O M E A R M

& W A I T F O R

B P M T I M E R 00:00 Se...B001 BPM ti...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line1.7 IOStatus: B095 STEP2B;Off="nothomed";On="SEQUENCE"Line2.1 IOStatus: B095 STEP2B;Off="step 2b";On="STEP 2B*"Line3.1 IOStatus: B096 STEP3B;Off="step 3b";On="STEP 3B*"Line4.1 IOStatus: B096 STEP3B;Off="step 4b";On="STEP 4B*"Line6.11 B001 BPM timer-TRest

Page 13: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 13 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B004 MON INPUTS1(Message texts) : MONITOR INPUTS 1

Prio = 91Quit = offText1: enabledText2: disabled

M O N I T O R - I N P U T S - 1

A 1 P R E S S U R E 0.0AI1

V

A 2 A R M P O S N 0.0AI2

V

D 1 R U N A C 1:ON 0:OFF

I3

D 2 R U N B C 1:ON 0:OFF

I4

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line2.12 AI: AI1Line3.12 AI: AI2Line4.12 IOStatus: I3;Off="OFF";On="ON"Line5.12 IOStatus: I4;Off="OFF";On="ON"

B005 fb_PSI(Analog Amplifier) : Gain =0.19+Offset=-37Point =2

B006(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B007(Analog Amplifier) : Gain =1.0+Offset=0Point =0

Page 14: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 14 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B008(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B009(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B010(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B011 v_HiLimPSI(Analog Amplifier) : Gain =1.0+Offset=0Point =0

B012 v_HiLimAngle(Mathematic instruction) : V1=999+V2=0V3=0V4=0Point=0((999+0)+0)+0

B013 PK_PL DELTA(Analog Amplifier) : Gain =1.0+Offset=1Point =0

B014 v_LoLimAngle(Mathematic instruction) : V1=1+V2=0V3=0V4=0Point=0((1+0)+0)+0

Page 15: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 15 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B015 PEEP START P(Max/Min) : RECORD PEEP STARTING PRESSURE

Rem = offMode = 3ResetMaxMin = false

B017(Message texts) : PRESSURE FDBK

Prio = 80Quit = offText1: enabledText2: disabled

- P R E S S U R E - F D B K -

A C T . L I V E -0.37B005 fb_PSI - ...

P E A K 0B031 PEAK PSI ...

P L A T E A U 0B128 PSI PLAT...

P E E P S T A R T 0.00B172 - Ax, am...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.10 B005 fb_PSI-AxLine4.10 B031 PEAK PSI MON-MaxLine5.10 B128 PSI PLATEAU-AqLine6.10 B172-Ax

Page 16: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 16 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B018 ASSIST CNTRL(Message texts) : SEQUENCE ASSIST CONTROL

Prio = 40Quit = offText1: enabledText2: disabled

- - A C - - 1:SEQUENCE0:NotHomed

B089 STEP... - -

1:STEP 2A*0:step 2a

B090 STEP3... I N H A L E

1:STEP 3A*0:step 3a

B091 STEP4... M E A S U R E

1:STEP 4a*0:step 4a

B092 STEP5A... H O M E A R M

1:STEP 5A*0:step 5a

B093 STEP6... P E E P P K

B P M T I M E R : 00:00 Se...B001 BPM ti...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line1.7 IOStatus: B089 STEP2A;Off="NotHomed";On="SEQUENCE"Line2.1 IOStatus: B090 STEP3A;Off="step 2a";On="STEP 2A*"Line3.1 IOStatus: B091 STEP4A;Off="step 3a";On="STEP 3A*"Line4.1 IOStatus: B092 STEP5A;Off="step 4a";On="STEP 4a*"Line5.1 IOStatus: B093 STEP6A;Off="step 5a";On="STEP 5A*"Line6.11 B001 BPM timer-TRest

Page 17: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 17 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B020(Message texts) : ALARM SCREEN 1 OF 2

Prio = 100Quit = offText1: enabledText2: disabled

- A L A R M - S C R E E N - 1 -

P S I T O O H I G H 1:NOK 0:OK

B033 Fault 1 [...

P S I T O O L O W 1:NOK 0:OK

B034 Fault 2 [...

+ O V E R T R A V E L 1:NOK 0:OK

B035 Fault 3 [...

- O V E R T R A V E L 1:NOK 0:OK

B036 Fault 4 [...

C L O S E T I M E O U T 1:NOK 0:OK

B037 Fault 5 [...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line2.14 IOStatus: B033 Fault 1;Off="OK";On="NOK"Line3.14 IOStatus: B034 Fault 2;Off="OK";On="NOK"Line4.14 IOStatus: B035 Fault 3;Off="OK";On="NOK"Line5.14 IOStatus: B036 Fault 4;Off="OK";On="NOK"Line6.14 IOStatus: B037 Fault 5;Off="OK";On="NOK"

B021(Up/Down counter) : Rem = offOn=60+Off=0Start=0

B022(Off-Delay) : Rem = off00:01s+

B023 ARM OVERTRAV(On-Delay) : ARM ANGLE TOO HIGH ALARM DELAY

Rem = off00:00s+

B024 ARM OVERTRAV(On-Delay) : ARM ANGLE TOO LOW ALARM DELAY

Rem = off00:00s+

Page 18: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 18 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B025(Message texts) : ARM ALARM LIMITS

Prio = 84Quit = offText1: enabledText2: disabled

A R M - A L A R M - L I M I T S

L I M I T S B E L O W A R E

H A R D C O D E D & R E Q

S T O P A N D E X I T

+ O V R T R V E L 0B012 v_HiLimA...

- O V R T R V E L 0B014 v_LoLimA...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line5.10 B012 v_HiLimAngle-AqLine6.10 B014 v_LoLimAngle-Aq

B026(Latching Relay) : Rem = off

Page 19: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 19 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B027(Message texts) : Prio = 102Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 1 / 6

B R E A T H S P E R

M I N U T E : 16B139 - Cou...

1:* 0:

B147 [Analog C...

S E C ( 1 / 1 0 0 ) A F T E R

P E A K T O M E A S U R E

P L A T E A U : 50B159 - Cou...

1:* 0:

B148 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.10 B139-CntLine3.16 IOStatus: B147;Off="";On="*"Line6.10 B159-CntLine6.16 IOStatus: B148;Off="";On="*"

B029 almOpenTOD(On-Delay) : Rem = offB187ms+

Page 20: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 20 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B030(Message texts) : ALARM SCREEN 2 OF 2

Prio = 99Quit = offText1: enabledText2: disabled

- A L A R M - S C R E E N - 2 -

O P E N T I M E O U T 1:NOK 0:OK

B038 Fault 6 [...

A R M T O O S L O W 1:NOK 0:OK

B039 Fault 7 [...

F A U L T 8 1:NOK 0:OK

B040 Fault 8 [...

F A U L T 9 1:NOK 0:OK

B026 [Latchin...

F A U L T 1 0 1:NOK 0:OK

B044 [Latchin...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line2.14 IOStatus: B038 Fault 6;Off="OK";On="NOK"Line3.14 IOStatus: B039 Fault 7;Off="OK";On="NOK"Line4.14 IOStatus: B040 Fault 8;Off="OK";On="NOK"Line5.14 IOStatus: B026;Off="OK";On="NOK"Line6.14 IOStatus: B044;Off="OK";On="NOK"

B031 PEAK PSI MON(Max/Min) : RECORD PEAK INHALE PRESSURE

Rem = offMode = 1ResetMaxMin = false

B032(AND (Edge)) : PAGE 11 ALARMS

B033 Fault 1(Latching Relay) : Rem = off

B034 Fault 2(Latching Relay) : Rem = off

B035 Fault 3(Latching Relay) : Rem = off

B036 Fault 4(Latching Relay) : Rem = off

Page 21: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 21 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B037 Fault 5(Latching Relay) : Rem = off

B038 Fault 6(Latching Relay) : Rem = off

B039 Fault 7(Latching Relay) : Rem = off

B040 Fault 8(Latching Relay) : Rem = off

B044(Latching Relay) : Rem = off

B045(Message texts) : ARM FDBK

Prio = 83Quit = offText1: enabledText2: disabled

A R M - P O S I T O N - F D B K

1:IS CLOSD0: ?

B063 ARM IS... 1:IS OPEN 0: ?

B062 ARM IS ...

1:CLOSING 0:q2-off

Q2 1:OPENING 0: q1-off

Q1

P O S I T I O N : 0B007 - Ax, am...

P O S . O P E N 0B046 v_OpenA...

P O S . C L S D 0B048 v_ClosdA...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line2.1 IOStatus: B063 ARM IS CLOSD;Off="?";On="IS CLOSD"Line2.9 IOStatus: B062 ARM IS OPEN;Off="?";On="IS OPEN"Line3.1 IOStatus: Q2;Off="q2-off";On="CLOSING"Line3.9 IOStatus: Q1;Off="q1-off";On="OPENING"Line4.10 B007-AxLine5.10 B046 v_OpenAngl-AxLine6.10 B048 v_ClosdAngl-Ax

B046 v_OpenAngl(Analog Amplifier) : Gain =1.0+Offset=0Point =0

Page 22: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 22 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B048 v_ClosdAngl(Analog Amplifier) : Gain =1.0+Offset=0Point =0

B049(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B051(Analog Comparator) : On =0Off =0Gain =1.0+Offset =0Point=0

B053(Latching Relay) : Rem = off

B054(Latching Relay) : Rem = off

B059 almClsTOD(On-Delay) : Rem = offB21ms+

B062 ARM IS OPEN(Latching Relay) : ARM IS OPEN

Rem = off

B063 ARM IS CLOSD(Latching Relay) : ARM IS CLOSED

Rem = off

B072(On-Delay) : Rem = off00:01s+

Page 23: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 23 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B084 MON INPUTS 2(Message texts) : MONITOR INPUTS 2

Prio = 90Quit = offText1: enabledText2: disabled

M O N I T O R - I N P U T S - 2

A 3 J O Y S T I C K 0.0AI3

X

A 4 J O Y S T I C K 0.0AI4

Y

D 5 J O Y S T K P B 1:ON 0:OFF

I5

D 6 S P A R E 1:ON 0:OFF

I6

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line2.12 AI: AI3Line3.12 AI: AI4Line4.14 IOStatus: I5;Off="OFF";On="ON"Line5.14 IOStatus: I6;Off="OFF";On="ON"

B087 ShowVarPages(Off-Delay) : Rem = off10:00s+

B088 STEP1(Latching Relay) : Rem = off

B089 STEP2A(Latching Relay) : Rem = off

B090 STEP3A(Latching Relay) : Rem = off

B091 STEP4A(Latching Relay) : Rem = off

B092 STEP5A(Latching Relay) : Rem = off

B093 STEP6A(Latching Relay) : Rem = off

Page 24: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 24 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B094 Total cycles(Up/Down counter) : TOTAL ARM CYCLES

Rem = offOn=999999+Off=0Start=0

B095 STEP2B(Latching Relay) : Rem = off

B096 STEP3B(Latching Relay) : Rem = off

B097 STEP4B(Latching Relay) : Rem = off

B098(Up/Down counter) : Rem = offOn=0+Off=0Start=0

B111 STATS(Message texts) : STATS

Prio = 10Quit = offText1: enabledText2: disabled

- S T A T I S T I C S - - - - -

T O T A L A R M C Y C L E S

0B098 - Cou...

, 0B094 Total ...

T O T A L H O U R S . . .

R U N N I N G : 0:0 Hours (...B112 - MN - hour:...

A C M O D E : 0:0 Hours (...B113 - MN - hour:...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.1 B098-CntLine3.8 B094 Total cycles-CntLine5.9 B112-MN_h:mLine6.9 B113-MN_h:m

Page 25: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 25 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B112(Hours Counter) : TOTAL HOURS RUNNING (AC AND BC)

MI= 0h 0m+OT= 0h 0mQ->0=R+En

B113(Hours Counter) : TOTAL HOURS IN ASSIST CONTROL (AC) MODE

MI= 0h 0m+OT= 0h 0mQ->0=R+En

B117(Off-Delay) : Rem = off00:01s+

B124(Max/Min) : Rem = offMode = 3ResetMaxMin = true

B128 PSI PLATEAU(Max/Min) : RECORD PLATEAU PRESSURE

Rem = offMode = 0ResetMaxMin = false

B129(Message texts) : PRESSURE ALARM LIMITS

Prio = 81Quit = offText1: enabledText2: disabled

P R E S S U R E - A L M - L I M

H I A L M L I M

H I A L M D L Y

L O A L M L I M 1B013 PK_PL DE...

P E E P S T A R T

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line4.10 B013 PK_PL DELTA-Ax

Page 26: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 26 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B130 PbCursor+(Analog Comparator) : SEL. NEXT VAR

On =200Off =150Gain =-1.0+Offset =0Point=0

B131 PbCursor-(Analog Comparator) : SEL. PREV. VAR

On =200Off =150Gain =1.0+Offset =0Point=0

B132 PbIncrease(Analog Comparator) : INCREASE VALUE

On =200Off =150Gain =-1.0+Offset =0Point=0

B133 PbDecrease(Analog Comparator) : DECREASE VALUE

On =200Off =150Gain =1.0+Offset =0Point=0

B134(On-Delay) : Rem = off00:01s+

B135(On-Delay) : Rem = off00:01s+

B137(On-Delay) : Rem = on00:05s+

B138(Up/Down counter) : Rem = onOn=11-Off=0Start=0

Page 27: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 27 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B139(Up/Down counter) : Rem = onOn=200+Off=5Start=16

B142(Analog Amplifier) : Gain =0.0+Offset=500Point =0

B146 ENUM(Analog MUX) : VARIABLE SELECTED ENUM

V1 =B138+V2=0V3=0V4=0Point =0

B147(Analog Comparator) : On =1Off =2Gain =1.0+Offset =0Point=0

B148(Analog Comparator) : On =2Off =3Gain =1.0+Offset =0Point=0

B150(Analog Amplifier) : Gain =1.0+Offset=0Point =0

B151(Analog Comparator) : On =3Off =4Gain =1.0+Offset =0Point=0

Page 28: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 28 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B152(Analog Comparator) : On =4Off =5Gain =1.0+Offset =0Point=0

B157(Analog Comparator) : On =5Off =6Gain =1.0+Offset =-4Point=0

B159(Up/Down counter) : Rem = onOn=500+Off=0Start=50

B160(Up/Down counter) : Rem = onOn=500+Off=0Start=5

B161(Up/Down counter) : Rem = onOn=1000+Off=0Start=400

B162(Up/Down counter) : Rem = onOn=1000+Off=0Start=10

B163 PEEP PSI SP(Mathematic instruction) : V1=B160+V2=0V3=0V4=0Point=2((B160+0)+0)+0

B167 WAITPLATEAU(On-Delay) : Rem = offB159tx+

Page 29: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 29 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B168(Message texts) : SET PEAK-PLATEAU ALARM 4 OF 6

Prio = 108Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 4 / 6

P E A K - P L E A T E A U

P S I I S :

S E T P K - P L E A T E A U

M A X A L A R M L I M I T

T O : 0.00B185 - Aq amp...

1:* 0:

B174 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line6.7 B185-AqLine6.16 IOStatus: B174;Off="";On="*"

Page 30: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 30 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B169(Message texts) : SET MAX PLATEAU PRESSURE 5 OF 6

Prio = 110Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 5 / 6

P L A T E A U P R E S S U R E

I S : 0B128 PSI PLAT...

S E T M A X P L E A T E A U

P R E S S U R E A L A R M

T O : 1:* 0:

B175 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.5 B128 PSI PLATEAU-AqLine6.16 IOStatus: B175;Off="";On="*"

B171 INH DETECTED(Analog Comparator) : INHALE DETECTED

On =B172Off =0Gain =1.0+Offset =0Point=0

B172(Analog Amplifier) : PEEP PRESSURE DROP TRIGGER PSI

Gain =1.0+Offset=0Point =2

B173(Analog Comparator) : On =6Off =7Gain =1.0+Offset =0Point=0

Page 31: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 31 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B174(Analog Comparator) : On =7Off =8Gain =1.0+Offset =0Point=0

B175(Analog Comparator) : On =8Off =9Gain =1.0+Offset =0Point=0

B176(Analog Comparator) : On =9Off =10Gain =1.0+Offset =0Point=0

B177(Analog Comparator) : On =10Off =11Gain =1.0+Offset =0Point=0

B180(Mathematic instruction) : V1=B183+V2=4V3=0V4=0Point=2((B183/4)+0)+0

B183(Up/Down counter) : Rem = onOn=500+Off=0Start=0

B184(Up/Down counter) : Rem = onOn=500+Off=0Start=0

Page 32: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 32 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B185(Mathematic instruction) : V1=B184+V2=4V3=0V4=0Point=2((B184/4)+0)+0

B186(Mathematic instruction) : V1=600+V2=B139V3=100V4=0Point=2((600/B139)*100)+0

B187(Up/Down counter) : Rem = offOn=60+Off=0Start=0

B188(Analog Amplifier) : Gain =1.0+Offset=0Point =0

B191(Mathematic instruction) : V1=B161+V2=0V3=0V4=0Point=0((B161+0)+0)+0

B193(Mathematic instruction) : V1=B162+V2=0V3=0V4=0Point=0((B162+0)+0)+0

Page 33: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 33 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B194(Message texts) : POPUP VARIABLE PAGE 6 OF 6

Prio = 112Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 6 / 6

S E T # S E C S T I L L

A L A R M O C C U R S

W H E N A R M S A R E

C L O S I N G : 0B021 - Cou...

1:* 0:

B176 [Analog C...

O P E N I N G : 0B187 - Cou...

1:* 0:

B177 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line5.10 B021-CntLine5.16 IOStatus: B176;Off="";On="*"Line6.10 B187-CntLine6.16 IOStatus: B177;Off="";On="*"

B197(Up/Down counter) : Rem = onOn=500+Off=0Start=0

B201(Analog Comparator) : PLATEAU TOO HIGH

On =0Off =0Gain =1.0+Offset =0Point=0

B204(On-Delay) : Rem = off01:00s+

B206(On-Delay) : Rem = off01:00s+

Page 34: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 34 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B225(Message texts) : SET MAX PRESSURE ALARM 3 OF6

Prio = 106Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 3 / 6

A C T M A X P R E S S U R E

I S : 0B031 PEAK PSI ...

S E T M A X P R E S S U R E

A L A R M L I M I T

T O : 0.00B180 - Aq amp...

1:* 0:

B173 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.7 B031 PEAK PSI MON-AqLine6.7 B180-AqLine6.16 IOStatus: B173;Off="";On="*"

Page 35: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 35 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B226(Message texts) : POPUP VARIABLE PAGE 2 OF 6

Prio = 104Quit = offText1: enabledText2: disabled

- - S E T P O I N T S - - 2 / 6

P E E P T R I G G E R

P R E S S U R E : 5B160 - Cou...

1:* 0:

B151 [Analog C...

V O L U M E - A R M P O S N -

O P E N P O S N 400B161 - Cou...

1:* 0:

B152 [Analog C...

C L S E P O S N 10B162 - Cou...

1:* 0:

B157 [Analog C...

--> Ticker setting - CBC - Line1: N - Line2: N - Line3: N - Line4: N - Line5: N - Line6: NMessage Destination - Both

Line3.10 B160-CntLine3.16 IOStatus: B151;Off="";On="*"Line5.10 B161-CntLine5.16 IOStatus: B152;Off="";On="*"Line6.10 B162-CntLine6.16 IOStatus: B157;Off="";On="*"

B228(Mathematic instruction) : V1=B197+V2=0V3=0V4=0Point=0((B197+0)+0)+0

B232(Analog Amplifier) : Gain =1.0+Offset=0Point =0

Page 36: Pressure 0-10V 0-1.45psi AI A PAGE 1 - PRESSURE ...etischer.com/ventilator/files/siemens/history/Ventilator...Project: DIY Ventilator Installation: File: Customer: OpenSource Diagram

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page: 36 / 36

Creator: Eric TischerChecked:Date: 3/28/20 2:00 PM/4/8/20 8:31 PM

Project: DIY VentilatorInstallation:File:

Customer: OpenSourceDiagram No.:Page:

Block Number (Type) Parameter

B233(Mathematic instruction) : V1=B31+V2=B128V3=0V4=0Point=0((B31-B128)+0)+0

B234(Off-Delay) : Rem = off00:01s+

B237(Up/Down counter) : Rem = offOn=0+Off=0Start=0

I3(Input) : RUN AC Mode

I4(Input) : RUN BC Mode

I5(Input) : Joystick PB

I6(Input) : Red PB

M2(Flag) : TICKTOCK

Q1(Output) : OPEN ARM

Q2(Output) : CLOSE ARM

X22(Open connector) : POPUP VARIABLE PAGE 1 OF 6