Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

42
Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013

Transcript of Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Page 1: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Preliminary Detailed Design Review

Periodontal Measurement Test SystemFebruary 1, 2013

Page 2: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Periodontal Measurement Test System

• Current method• Goals of new method

Page 3: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Customer Needs

Page 4: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Specifications

Page 5: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Pugh Matrix

Page 6: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

PHANTOM TOOTH

Canadianawareness.org

Page 7: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Specifications

Spec Description CN

3.1 The phantom tooth shall be made of non-biological materials that mimic human tissue with respect to ultrasonic wave propagation properties. 8, 9, 10

3.2 The phantom tooth shall be able to be quickly replaced within the phantom holding fixture. 3, 11

3.3 The phantom tooth geometry shall allow for quick alterations of the physical relationship between the tooth and the gums. 1, 4

Page 8: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Phantom Design

Page 9: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Phantom Design

Page 10: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Relevant US Properties

• Density• Speed of sound• Acoustic impedance

Page 11: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Material Selection

Page 12: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Contacts

• Dr. Phillips – RIT bio medical engineering• Dr. Maria Helguera – RIT imaging science– Biomed ultrasound

• Eastman Dental Clinic – Dr. Caton• Bradley Rogge – RIT’s Pre-dental association• Professor Lamkin-Kennard – RIT Mechanical

Engineering• Hamad Ghazle, Jodie Crowley – RIT Ultrasound

Page 13: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

FIXTURE DESIGN

Page 14: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

SpecificationsSpecification Description Customer

Needs1.0 Test Probe Fixture 1.1 The test probe fixture must be able to move through 3 axis. 51.2 The test probe fixture shall be compact and portable. 3, 4, 11, 12

1.3The test probe fixture shall be simply constructed to allow for maintenance to be performed by untrained personnel relying only on a maintenance manual (to be provided). 3, 4, 11, 12

1.4

The test probe fixture shall allow for position changes on or about any axis to a programmatically specified point to within 0.1 mm for linear movement or 1° for angular movement. 2, 5, 6

1.5The test probe fixture shall hae minimal impact on the environment and noise levels within the surrounding area while active.

5, 7, 8, 9, 10, 11

2.0 Phantom Holding Fixture 2.1 The phantom holding fixture must be able to move through 2 axis. 5

2.2The phantom holding fixture shall be simply constructed to allow for maintenance to be performed by untrained personnel relying only on a maintenance manual (to be provided). 3, 4, 11, 12

2.3

The phantom holding fixture shall allow for position changes on or about any axis to a programmatically specified point to within 0.1 mm for linear movement or 1° for angular movement. 2, 5, 6

2.4The phantom holding fixture shall allow for the phantom tooth to be replaced by a human mandible. 11

2.5The phantom holding fixture shall be simply constructed to allow for maintenance to be performed by untrained personnel relying only on a maintenance manual (to be provided). 3, 4, 11, 12

Page 15: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Positioning System Model

Page 16: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Power Screw

Page 17: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Turn Table

Page 18: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Assembly Drawing

Page 19: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Pitch Movement

Page 20: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Power Screw Assembly

Page 21: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Power Screw Specs Power Screw Specs Nut Specs

Material 303 Stainless Self Lubricating Polyacetal

Total Length (in) 24 1.9

Pitch (in) 0.25 0.25

Diameter (in) 0.375 0.375

Supplier Haydon Kirk Haydon Kirk

Price 62.83 29.87

Efficiency (%) 70 70

Thread Direction Right Right

Dynamic Load - 10

Drag Torque (oz-in) - 1 to 5

Page 22: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Nut Detailed Drawing

Page 23: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Turn Table Drawing

Page 24: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Pitch Axis Drawing

Page 25: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Mounting Table Drawing

Page 26: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Guide Rail Clamp

Page 27: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Pitch Selection Calculations

Maximum Screw Pitch to Meet Accuracy Req's

Expected Stepper Motor Speeds

w/o Micro-Stepping (rev/sec) 16

w/ Micro-Stepping (rev/sec) 2 Pmax (mm) 36

Motor Angular Accuracy (deg) 1 Pmin (in) 1.417323

Desired Linear Accuracy (+/- mm) 0.1 Maximum Screw Pitch to Meet Travel Time Req's

Estimated Screw Length (in) 10

Estimated Screw Length (mm) 254

Estimated Time of Travel (s, full L) 12 Pmax (mm) 10.58333

Chosen Screw Pitch (in) 0.25 Pmax (in) 0.416667

Page 28: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Torque Requirement CalculationsProperty Symbol Units Value Required Torque to Overcome FrictionScrew Pitch P rev/in 4Load on Screw W oz 16Screw Efficiency e 0.7 Tfric (oz-in) 0.036378

Static Friction Coeff-nut on screw μs 0.04

Accel of Gravity G in/sec2 384 Jload 0.02533Time to reach constant Vel t sec 0.1Density of Screw ρ oz/in3 4.624 Jscrew 0.089772Length of Screw L in 10Radius of Screw R in 0.1875Load on Screw W oz 16

Ang velocity of motor ω rad/sec 12.5663706 Taccel (oz-in) 0.395513Screw Pitch P rev/in 4Motor Inertia Jmotor oz-in^2 1.0934959

Total Torque (oz-in) 0.431891

Note: All values for 1lb load on horizontal screw

Page 29: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Bearing Selection

• Very low radial loads– Guide rail support

• Small axial loads– Low friction nut– Guide rail support

• Basic radial bearing should handle loads

Page 30: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Bearing Specs

Bearing Specs

Style Double Sealed

ID (in) 0.375

OD (in) 0.6875

Width (in) 0.3125

Dyn. Load Cap (lbs) 255

Price (ea) 8.38

Quantity 6

Page 31: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Couplings

Coupling Specs

Material Steel

ID (in) 0.25

OD (in) 0.5

Length (in) 0.75

Set Screw Size 10-32

Max RPM 3450

Max Torque 59

Price 6.93

Page 32: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Collars

Collar Specs

Material Steel

ID (in) 0.25

OD (in) 0.5

Length (in) 0.28125

Set Screw Size 10-32

Price 0.63

Page 33: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

ELECTRICAL

Page 34: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Specifications

• 4.0 – Proof of concept put together, proved we could attain appropriate control required for final system

• 5.0 – Unable to test, waiting on probe, able to communicate with an oscilloscope, but no data has been attained from probe to date due to lack of probe

• 6.0 – Have data logging already written in code, working on limit comparison programming

Page 35: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

BLOCK DIAGRAM

Page 36: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

WIRING DIAGRAM

Page 37: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

PROGRAMMING FLOWCHART

Page 38: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

ELECTRICAL BOM AND COST

Preferred List

Item Part Number Manufacturer Lead Time Cost Qty Extended Cost

Arduino UNO UNO R3 Arduino 3-5 days $ 24.95 1 $ 24.95

Servo HS-225MG Hitec 3-5 days $ 35.99 2 $ 71.98

Stepper RB-SOY-14 SOYO 3-5 days $ 44.28 3 $ 132.84

EasyDriver B004G4XR60 Sparkfun 3-5 days $ 14.95 6 $ 89.70

Misc Budget $ 100.00 1 $ 100.00

$ 419.47

Page 39: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Justification

Control System• Arduino is very compatible with

LabVIEW• Arduino is low cost and as a skill

is going to be part of the new Control System curriculum at RIT

• EasyDriver motor control boards are inexpensive and individually replaceable if damaged.

• EasyDriver allows us to do Full, stepping to increase control.

Motors• Servos were chosen for

combination of price, mobility, and durability.

• Metal servo with high torque rating to allow ease of use with system design

• Stepper motors were picked based off of Voltage, Current, torque, and Driver limitations

Page 40: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Risk Assessment

Page 41: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

High Risk Items

• Non-bio materials– Availability– Similarity to human tooth

• Formation of phantom– Processes available hinge on materials chosen

• Probe!

Page 42: Preliminary Detailed Design Review Periodontal Measurement Test System February 1, 2013.

Questions/Comments/Suggestions